Calibration
4. search units
Frequanze size shape bam angle and propagating wave
Only single element type of search units shall be used. The following combination of
search unit properties and their expected propagating wave in the material being
examined shall confirm to the following table.
4. Operating temperature
Any search unit utilized under this procedure shall be used on surface temperature
not to exceed the temperature limitation as specified by their manufacturer
4.3 search unit box
The cable to be use used under this procedure shall be the co axial type. For
practicial purpose the leight of the cable – should not be less than 1.5 meters
Couplant
The couplant including additives, shall not be detrimental to the material being
examed. Couplant having goods wetting characteristic such as sae 30 mechene oil,
metylan, glycerine, cellulose gum, may be used. For ultrasonic examination on
material with temperature above 65 degress celcius, special couplants shall be used
to prevent excecive evaporation from the examknation surface, the couplant used
for calibration shall be of the same type as the one to be used on the examination of
the production parts. All couplant used shall be controlled from contaminants as
follows:
a. Couplants used on nickel base alloys shall not contain more than 250 ppm of
sulfur
b. Couplants used on austentic stainless stell or titanium shall not contain more
than 250 ppm of halides (chlorides plus flourideas)
6. Calibration reference block
7. The following block shall be used for calibration of equiptmrnt calibration and
examination system calibraton :
II V1 and /or V2 Block
Non piping basic calibration block
Calibration
Ultrasonic instrument qualification procedure
The linearity check as define in par 5.1.1 and 5.1.2 shall be met at intervals not to
exceed three monts or prior to first use thereafter. If the instrument is suspected of
being damanged or has gone throught major repair then linearity checks shall be
done prior to use. If the instrument has not been in used in three months or more,
calibration shall be done prior to use. These qualification results shall be recorded in
appropriate log ( see exhibit A) ath the minimum, the last valid qualification record
shall be maintened on file.
6.1. screen height linearity
Place an angle beam search unit, as shown in fig. 2, so that indication from booth ½
and ¾ t hole give a 2:1 ration of amplitudes between the two indications. Other
convenient reflector from any calibration block may be used. Adjust the sensitivity
gain so that the larger indication is set at 100 % of full screen height (FSH) without
moving the search unit , adjust gain sensitivity to sucsessively set the larger
indication from 100% to 20% FSH, in 2 db steps and read the smaller indication at
each setting. The readings must be 50 % of the large amplitude, within 5% of full
screen height. The setting ad reading must be estmed to e nearest 1 & of full screen.