Sei sulla pagina 1di 109

User's

Manual
GC1000 Mark II
Process Gas Chromatograph
LCD Panel Operation Manual

IM 11B03A03-05E

IM 11B03A03-05E
2nd Edition
<Toc> <Ind> <Rev> <Introduction> i

◆ Notice
● Regarding This Manual
1. This Manual should be passed on to the end user.
2. Read this manual carefully and fully understand how to operate this product before
you start operation.
3. Yokogawa makes no warranty of any kind with regard to this material, but not limited
to, implied warranties of merchantability for particular purpose.
4. All rights reserved. No part of this manual may be reproduced in any form without
Yokogawa’s written permission.
5. Great effort has been expended to ensure that the descriptions in this manual are
correct. Should you, however, come across a questionable area or note an inconsis-
tency, a telephone call or letter to Yokogawa.co.,ltd. noting the questionable area
would be highly appreciated.
6. The contents of this manual are subject to change without prior notice.

● Regarding Protection, Safety, and Prohibition Against Unauthorized Modifica-


tion.
1. For the protection and safe use of the product and the system controlled by it, be sure
to follow the instructions on safety described in this manual when handling the prod-
uct. In addition, if you handle the product in contradiction to these instructions, our
company does not guarantee safety.
2. The following safety symbol marks are used on the product concerned or in this
Manual :

WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice, condition
or the like, which, if not correctly performed or adhered to, could result in injury or
death of personnel.

CAUTION
A CAUTION sign denotes a hazard. It calls attention to a procedure, practice, condi-
tion or the like, which, if not correctly performed or adhered to, could result in damage
to or destruction of part or all of the product.

IMPORTANT:
Indicates that operating the hardware or software in this manner may damage it or
lead to system failure.

NOTE:
Draws attention to information essential for understanding the operation and features.

TIP:
Gives information that complements the present topic.

Media No. IM 11B03A03-05E 2nd Edition : July. 2002 (YK) IM 11B03A03-05E 2nd Edition : July. 1. 2002-00
All Rights Reserved Copyright © 2001, Yokogawa Electric Corporation
<Toc> <Ind> <Rev> <Introduction> ii

See Also:
Gives the reference locations for further information on the topic.

Protective ground terminal:


In order to provide protection against electrical shock in case of a fault. This symbol
indicates that the terminal must be connected to ground prior to operation of equip-
ment.

Function ground terminal:


In order to provide protection against noise. This symbol indicates that the terminal
must be connected to ground prior to operation of equipment.

Alternating current
Indicates the power switch state “ON”.

Indicates the power switch state “Stand - by”.

Indicate the power switch state “OFF”.

3. If protection / safety circuits are to be used for the product or the system controlled by
it, they should be installed outside of the product.
4. When you replace parts or consumables of the product, use those specified by our
company.
5. Do not modify the product.

● Exemption from Responsibility


1. Yokogawa Electric Corporation does not make any warranties regarding the product
except those mentioned in the WARRANTY that is provided separately.
2. Yokogawa Electric Corporation assumes no liability to any party for any loss or dam-
age, direct or indirect, caused by the use or any unpredictable defect of the product.

● Regarding Software and set including Software Supplied by YOKOGAWA


1. Yokogawa makes no other warranties expressed or implied except as provided in its
warranty clause for software supplied Yokogawa.
2. Use this software with one specified computer only.
You must purchase another copy of the software for use with each additional com-
puter.
3. Copying this software for purposes other than backup is strictly prohibited.
4. Store the streamer or floppy disk (original medium) in a secure place.
5. Reverse engineering such as the disassembly of software is strictly prohibited.
6. No portion of the software supplied by Yokogawa may be transferred, exchanged, or
sublet or leased for use by any third party without prior permission by Yokogawa.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <Rev> <Introduction> iii

◆ Warning/Caution Labels
● To ensure safety operation of this equipment, warning/caution labels are at-
tached on the equipment as follows. Check these labels for your safety opera-
tion.

[Left side] [Front] [Right side] [Back]


F0001.EPS

A B

The inside of the enclosure is high temperature after turning off the power.
Don’t touch the inside of the enclosure or the compponents, the protective gas
should be suoolied for more than one hour after turning off the power.
D E

F0002.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <Rev> <Introduction> iv

◆ Introduction
Thank you for purchasing the GC1000 Mark II process gas chromatograph.
This manual describes the technical information on LCD panel operation of Model
GC1000D / GC1000S / GC1000T / GC1000E / GC1000W / GC1000C (Hereafter, it is
abbreviated as GC1000 Mark II) Process Gas Chromatograph.
Please lead the following respective documents before installing and using the GC1000
Mark II system.

● Documents Related to the GC1000 Mark II Process Gas Chromatograph


1. Instruction manuals
The product comes with the following instruction manuals.

■ Instruction manuals that do not depend upon the specifications of the product
(1) Overview (IM 11B03A03-01E)
(2) Basic Operation and Startup (IM 11B03A03-02E)
(3) Maintenance and Inspection Manual (IM 11B03A03-04E)
(4) LCD Panel Operation Manual (IM 11B03A03-05E)
(5) Alarm Message Manual (IM 11B03A03-06E)
and Installation Manual (TI 11B03A03-03E)

■ Instruction manuals that depend upon the specifications of the product


(1) GCMT Gas Chromatograph Maintenance Terminal Software Package Operation
Guide (IM 11B03G03-03E)
(2) Capture It! Manual (IM 11B3G1-02E)

■ Instruction manuals for related products


(1) Model GCCU Computing Unit for a gas chromatograph (IM11B03S03-01E)
(2) GCCU Utility Software Manual (IM11B3S2-02E)
(3) Analyzer Server User's Manual (IM11B5B1-01E)
(4) ASMT Analyzer Server Maintenance Terminal Operation Guide
(IM11B05A03-01E)
(5) GCIU Interface Unit (IM 11B5B2-01E)
and Hub GCHUB Installation Manual (TI 11B5C1-01E)

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <Rev> <Introduction> v
2. Operation Data
Operation data is supplied with the operation manuals in the delivered package and con-
tains the following required to use the gas chromatographs.
● Process conditions and measurement range
● Instrument specifications and operating conditions
● Standard sample for calibration
● Column system and column
● Miscellaneous data
Chromatogram, base line, repeatability, power supply voltage variation, etc.
● Analyzer flow diagram and installation
● Parts composition table
● General connection diagram
● Sampling system diagram (only if supplied by Yokogawa)

● Is the System Ready?


Before reading this manual, the following preparations must have been completed.
● The system must be unpacked and installed in the correct place.
● The piping for the utility gases such as carrier and calibration gases must be com-
pleted, followed by leak checking.
● The wiring for the power supply and others must be completed.

If these have not been completed yet, see the Installation Manual (TI 11B03A03-03E).
After completion, return to this manual and do the following:
● If the system power is on, turn off the power.
● Shut off all the gases at the flow control units.
Please read the following respective cautions (General Precautions, Caution of
using Explosion-Protection Instruments, on Piping Construction, and on Piping
Work) before installing and using the GC1000 system.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <Rev> <Introduction> vi
● General Precautions

WARNING
(1) In order to analyze gases, process gas chromatographs use a sample of the process
gas and utility gases.
Since these gases are typically combustible, combustion-sustaining, toxic,
odorous, resolvable, polymerizing, or corrosive, refer to the “Safety Informa-
tion” in our approval drawings and others to ensure safety thoroughly before
using these analyzers.
(2) Up to two protection systems, each of which weighs approximately 10 kg, are installed
on top of the GC1000. Therefore, the center of gravity is higher than the center of the
analyzer body.
Take great care when carrying and installing (piping- wiring) the GC1000. The
GC1000 must be carried and installed very carefully (including piping and
wiring) by more than one person (at least four people are recommended).
(3) Since the GC1000 are precision instruments, take care when handling not to jolt
of knock them.
(4) Use the GC1000 within the range of your purchase specifications.
Yokogawa assumes no responsibility for problems resulting from use by the customer
outside the purchase specifications.
If the GC1000 need to be modified or repaired, please contact your nearest Yokogawa
representative. Yokogawa assumes no responsibility for results where the customer or
any third party has attempted to modify or repair these products.
(5) When touching LCD Panel switches
When touching LCD Panel switches, first, discharge Electro Static Charge of the body.
Then, touch the LCD Panel switches. If not, LCD display may be changed by Electro
Static Discharge.

IMPORTANT:
(1) Read the attached instruction manual before operating the GC1000
(2) The instruments must be installed and operated according to the installation manual,
instruction manual, approval drawings, and operation data.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <Rev> <Introduction> vii
● CAUTIONS OF USING EXPLOSION-PROTECTION INSTRUMENTS
The GC1000 Process Gas Chromatographs are designed to protect against explosion.
When these analyzers are used in a hazardous area, observe the following precautions.
Since the applicable standard differs depending on the specifications of the analyzer to be
used, check the specifications of your analyzer.

(1) Kinds of explosion protection


To assure explosion protection, the GC1000 Process Gas Chromatographs have a pres-
surized and flameproof construction, or type X purging and explosionproof construction
meeting the following standards :

<GC1000D / GC1000S>
• JIS Expd IIB + H2 T1 (programmed-temperature oven 320˚C max., isothermal oven
225˚C max., liquid-sample valve 250˚C max.)
• JIS Expd IIB + H2 T2 (programmed-temperature oven 225˚C max., isothermal oven
225˚C max., liquid-sample valve 225˚C max.)
• JIS Expd IIB + H2 T3 (programmed-temperature oven 145˚C max., isothermal oven
145˚C max., liquid-sample valve 145˚C max.)
• JIS Expd IIB + H2 T4 (programmed-temperature oven 95˚C max., isothermal oven
95˚C max., liquid-sample valve 95˚C max.)

<GC1000W / GC1000C>
CENELEC (ATEX directive) certified : Group II Category 2G
• EEx pd II B +H2 T1 (programmed-temperature oven 320°C max, isothermal oven
225°C max., liquid-sample valve 250°C max.)
• EEx pd II B +H2 T2 (programmed-temperature oven 225°C max, isothermal oven
225°C max., liquid-sample valve 225°C max.)
• EEx pd II B +H2 T3 (programmed-temperature oven 145°C max, isothermal oven
145°C max., liquid-sample valve 145°C max.)
• EEx pd II B +H2 T4 (programmed-temperature oven 95°C max, isothermal oven 95°C
max., liquid-sample valve 95°C max.)

<GC1000T / GC1000E>
FM Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, NEMA3R.
Type Y purging and Type X purging for CLI, DIV1, GPS B, C & D, NEMA3R.
• T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid-
sample valve 250˚C max.)
• T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid-
sample valve 225˚C max.)
• T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid-
sample valve 145˚C max.)
• T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid-
sample valve 95˚C max.)

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00


<Toc> <Ind> <Rev> <Introduction> viii
CSA Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, TYPE3R.
Type Y purging and Type X purging for CLI, DIV1, GPS B, C & D, TYPE3R.
• T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max., liquid-
sample valve 250˚C max.)
• T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max., liquid-
sample valve 225˚C max.)
• T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max., liquid-
sample valve 145˚C max.)
• T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid-
sample valve 95˚C max.)

(2) Precautions for Explosionproof section (The analyzer with optional code
"FM/CSA Type Y purging" does not have the explosionproof section.)
When handling the screws on the cover of the Protection system, note the following to
avoid damaging the screws since they cannot be repaired.
(1) The enclosure is pressurized. Before removing the cover, reduce the internal pres-
sure by loosening the sealing plug for wiring on the enclosure or relevant means.
(2) When removing the cover, prevent any dirt or foreign matter from contaminating the
screw part.
(3) When installing the cover, tighten the screws by hand ; never use tools.
(4) Since the screws are coated with MOLYKOTE, do not lubricate them.

(3) Precautions when using hydrogen gas


When using hydrogen gas as the carrier gas, the FID or FPD combustion gas, to ensure
safety, install the analyzer in a location equipped with a ventilator or where there is suffi-
cient ventilation. Make sure there are no gas leaks from the pipe joints and inspect for
leaks.

(4) Installation site and environment


The analyzer specifications allow it to be used in hazardous areas as defined by Zone 1 IIB
+ H2T1, T2, T3, T4 (JIS / CENELEC) or DIV1, GPS B, C & D, T1, T2, T3, T4 (FM / CSA).
However, never install the analyzer in an area where the density of explosive gas persists
for a long time.

(5) Wiring works


Model GC1000D / GC1000S, analyzer obtains explosion proof authorization by the com-
plete set including metal fittings of the attachment.
When performing wiring, always use the attached sealing fittings and flameproof packing
adapter.

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00


<Toc> <Ind> <Rev> <Introduction> ix
(6) Maintenance and inspection
During usual maintenance and inspection, it is not necessary to check the explosion-
protected section.
Before opening the door of the explosion-protected section for maintenance and inspec-
tion, be sure to turn off the power. After completing maintenance and checks, close the
door completely then turn on the power after checking that the specified explosion protec-
tion performance is guaranteed. The parts to be checked are described in the Maintenance
and Inspection Manual (IM 11B03A03 - 04E).
If any of the following damage occurs, contact a Yokogawa sales representative or the
Yokogawa sales division
(1) If the screws securing the Protection System (explosionproof construction) are dam-
aged
(2) If the exterior or light transmission section of the enclosures is damaged
(3) If packings are cracked or conspicuously deformed

(7) Override function (The analyzer with optional code "FM/CSA Type Y purg
ing" does not have this function.)

WARNING
• When the override function is used, Analyzer becomes an ignition source and the high
temperature and the high voltage part will be exposed.
• Please confirm that in the ambient atmosphere, the concentration of explosive gases
is less than the allowable limit, by using a gas detector.

To return to the normal operation, turn off “the override switch” and then close the door as it
was before turning on power.
In this analyzer, if the pressure of the pressurized / type X purged enclosure system (oven,
electronic section) drops while the power is on, the pressurized explosion protection sec-
tion is activated to stop power supply. Therefore, in case of opening the door of the oven or
of the electronic section inadvertently, for maintenance, while the power is on, the protec-
tion system is activated to cut off the power.
The “override function” intensively releases this function of protection system.
The override switch is installed in section.

(8) Replacing parts


Always use parts specified by Yokogawa when replacing parts, for replacement, refer to the
Maintenance and Inspection Manual (IM 11B03A03 - 04E).

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00


<Toc> <Ind> <Rev> <Introduction> x
(9) Operation

WARNING
<CENELEC>
* Only trained persons may use this instrument in a hazardous location.
* Do not open when energized.

<FM>
For type X purging:
* This equipment contains components that operate at high temperature. The equip-
ment shall be deenergized for 60 minutes to permit those components to cool before
the enclosure is opened unless the area is demonstrated to be nonhazardous at the
time.
* Enclosure shall not be opened unless the area is known to be nonhazardous, or
unless all devices within have been de-energized.
* Power shall not be restored after enclosure has been opened until enclosure has
been purged for 12 minutes. (When the internal pressure is restored, the system
automatically purges over the 12 minutes, then turns on the power again.)
For explosionproof enclosure:
* Seal all conduits within 18 inches
* Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within have been de-energized. Power shall not be restored after
enclosure has been opened until enclosure has been purged for 12 minutes at speci-
fied pressure indicated by the pressure gauge labeled “EL.BOX” in the pressure and
flow control section.
* Alarm shall be provided and connected to alarm contact output.
(a) The alarm shall generate a visual or audible signal that attracts attention
(b) The alarm shall be located at constantly attended location.
(c) Electrical alarms shall be approved for the location in which they are installed.

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00


<Toc> <Ind> <Rev> <Introduction> xi
<CSA>
For type X purging:
* This equipment contains components that operate at high temperature. The equip-
ment shall be deenergized for 60 minutes to permit those components to cool before
the enclosure is opened unless the area is demonstrated to be nonhazardous at the
time.
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within the enclosure have been de-energized. Power must not be
restored after enclosure has been opened until enclosure has been purged for 12
minutes at a flow rate of 0.05m3/min.
* Power will automatically be removed when purge pressure falls below 40 mm (1.6 in)
of water column.
For explosionproof enclosure:
* A seal shall be installed within 50 cm of the enclosure.
* Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within the enclosure have been de-energized. Power must not be
restored after enclosure has been opened until enclosure has been purged for 12
minutes at a flow rate of 0.05m3/minute min.
* Remove power below 40mm (1.6in) of water column.

Take care not to generate mechanical spark when accessing to the instrument and periph-
eral devices in hazardous locations.
Do not press prick the keyboard of LCD panel (operation and display section) using such
as knives and sticks.

(10) Maintenance and Repair


The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the approval of Factory Mutual
Research Corporation and CSA certification and CENELEC certification .

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00


<Toc> <Ind> <Rev> <Introduction> xii
● Precautions Against Electrostatic Problems
The GC1000 system uses numerous IC components. When handling cards with IC
components mounted on them for maintenance or setting changes, take full precau-
tions against electrostatic problems.
These precautions are summarized below.
(a) When storing or carrying cards, enclose them in a conductive bag or antistatic
bag. (Cards as shipped by Yokogawa are enclosed in a conductive bag or
antistatic bag labeled with cautions against electrostatic problems.)
(b) Whenever mounting or demounting cards into or from a product, wear a wrist
strap grounded via a 1 MΩ resistance. Connect the wrist strap to any ground
terminal near the ground wire or to any unpainted part of the grounded frame.

Gard List strap


Sheet
List strap

1M‰

F0003.EPS

Using a wrist strap and conductive sheet Using a Conductive Sheet

(c) When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 MΩ resistance, wearing a wrist strap as in (2) above. Keep
easily-chargeable plastic materials away from the bench.
(d) Never touch components mounted on the cards, the pattern side, connectors,
pin components, etc. with bare hands, unless using a wrist strap and a conduc-
tive sheet.
(e) Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).

IM 11B03A03-05E
<Int> <Ind> <Rev> Toc-1

GC1000 Mark II Process Gas Chromatograph


LCD Panel Operation Manual

IM 11B03A03-05E 2nd Edition

CONTENTS
◆ Notice ............................................................................................................ i
◆ Warning/Caution Labels ............................................................................... iii
◆ Introduction.................................................................................................. iv
1. Basic Configuration of LCD Panel and Keys .............................................. 1-1
1.1 Panel Configuration .......................................................................................... 1-1
1.2 Panel Switching Key ......................................................................................... 1-2
1.3 Cursor Scroll Keys ............................................................................................ 1-2
1.4 Multipurpose Function Keys .............................................................................. 1-2
1.5 Setup Item Call Keys ......................................................................................... 1-2
1.6 Alpha-numeric and Backspace Keys ................................................................. 1-2
1.7 Cancel and Entry Keys ...................................................................................... 1-3
1.8 Page Keys ........................................................................................................ 1-3
1.9 Display Reset Keys ........................................................................................... 1-3
2. Panel Configuration Diagram .................................................................... 2-1
2.1 Operation Status Display Panel STATUS ........................................................... 2-1
2.2 Analysis Result Display Panel RESULT ............................................................. 2-2
2.3 Alarm Display Panel ALARM ............................................................................. 2-2
2.4 Temperature Status Display Panel TEMP ....................................................... 2-2
2.5 Chromatogram Display Panel CHROM ............................................................... 2-3
LOCAL/
2.6 Communication Mode Switching Display Panel REMOTE .................................... 2-3
2.7 Operation Panel OPERATION ................................................................................... 2-3
2.8 Parameter Setup Panel TABLE ........................................................................ 2-4
3. Basic Procedures for Viewing and Entry .................................................... 3-1
3.1 Panel Switching 1 ............................................................................................. 3-1
3.2 Panel Switching 2 ............................................................................................. 3-2
3.3 Setup Item Selection 1 ...................................................................................... 3-2
3.4 Setup Item Selection 2 ...................................................................................... 3-4
3.5 Numeric Entry ................................................................................................... 3-6
3.6 Character Entry ................................................................................................. 3-6
3.7 Entry Cancellation 1 .......................................................................................... 3-6
3.8 Entry Cancellation 2 .......................................................................................... 3-9
3.9 Display Page Switching ..................................................................................... 3-9
3.10 Scroll ............................................................................................................... 3-10

IM 11B03A03-05E 2nd Edition : July. 1. 2 002-00


<Int> <Ind> <Rev> Toc-2
4. Display Panels .......................................................................................... 4-1
4.1 User Levels ....................................................................................................... 4-1
4.2 Standard Configuration of the LCD Panel .......................................................... 4-2
4.3 Operation Status Display Panel STATUS ........................................................... 4-4
F1

4.3.1 Detector Signal Display Panel STATUS Signal


............................... 4-4
F2
4.3.2 AI Signal Status Display Panel STATUS AI .............................. 4-5
F3
4.3.3 AO Signal Status Display Panel STATUS AO ............................. 4-6
F4
4.3.4 DI Signal Status Display Panel STATUS DI .............................. 4-7
F5
4.3.5 DO Signal Status Display Panel STATUS DO
............................ 4-8
4.4 Analysis Result Display Panel RESULT .............................................................. 4-9
4.4.1 Latest Analysis Result Display Panel ................................................ 4-9
4.4.2 Concentration
F2
Analysis Historical Record Display Panel
RESULT Conc .............................................................................. 4-11
4.4.3 Retention TimeF3
of Historical Record Display Panel
RESULT R . Time ............................................................................. 4-12
F4

4.4.4 Laboratory Analysis Result Display Panel RESULT Lab ............ 4-13
F5
4.4.5 Standard B.P Data Table Display Panel RESULT Lab ............... 4-14
4.5 Alarm Display Panel ALARM ........................................................................... 4-15
4.5.1 Alarm Status Display Panel ............................................................. 4-15
F2
4.5.2 Alarm Historical Record Display Panel ALARM Histry ................ 4-16
4.6 Temperature Status Display Panel TEMP ..................................................... 4-17
4.6.1 Temperature Status ........................................................................ 4-17
4.6.2 Column Oven Temperature Display Panel
(only when EPC is installed) ........................................................... 4-20
4.7 Chromatogram Display Panel CHROM ............................................................. 4-21
4.7.1 Chromatogram Display Panel ......................................................... 4-21
LOCAL/
4.8 Communication Mode Switching Display Panel REMOTE .................................. 4-23
4.8.1 Communication Mode Switching Display Panel .............................. 4-23
4.9 Operation Panel OPERATION ................................................................................. 4-24
4.9.1 Operation Menu Panel .................................................................... 4-24
4.9.2 Stream Command Display Panel .................................................... 4-26
4.9.3 Calibration (Validation) Command Display Panel ............................ 4-27
4.9.4 Manual Operation Panel ................................................................. 4-28
4.9.5 Laboratory Operation Panel ............................................................ 4-30
4.9.6 Range Select Display Panel ............................................................ 4-31
4.10 Parameter List Setup Panel TABLE ............................................................... 4-32
4.10.1 System Set TABLE SYSTEM SET ........................................... 4-33
4.10.2 Operation Condition Set TABLE Operation Condition Set ............ 4-34
4.10.3 Temp Set TABLE Temp Set .................................................. 4-35
4.10.4 Detector Signal Settings TABLE Detector Signal Set ................. 4-37
4.10.5 Method Set TABLE Method Set .............................................. 4-38
Setup Item Description Set Minimum Maximum Unit Remarks
Stream Setram number s 1 31
Name of stream Stream name s Max 8 letters
s

4.10.6 Stream Set .............................................. 4-43


Stream valve number Stream valve number 1 31
Method number Method number s 1 4
Kind of stream The kind of stream s Measurement , Calibration ,
Validation",
"Calibration run"(only for distillation),
"Sample run"(only for distillation)
Total compens Totalized value correction s Executed
computation specified Not executed
Total value Total valve s 0.1 100.0 %
Base line corr Base line correction s Executed
calculation request Not executed
SIMDIS calib stream Distillation calibration stream s 0 31 Display
T041006.EPS

4.10.7 Stream Sequence Set TABLE Str Sequence Set ...................... 4-44

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Int> <Ind> <Rev> Toc-3
4.10.8 Peak Common Set TABLE Peak Common Set ......................... 4-45
4.10.9 Peak Individual Set TABLE Peak Individual Set ......................... 4-46
4.10.10 Calibration/Validation Set TABLE Cal/Val Set ......................... 4-55
4.10.11 Multi Range Set Display TABLE Multi Range Set ...................... 4-57
4.10.12 Lab Set Display TABLE Lab Set .......................................... 4-58
4.10.13 Alarm Setup TABLE Alarm Set ............................................. 4-60
4.10.14 Peak Composition Set Panel TABLE Peak Composition Set ...... 4-61
4.10.15 Communication Set Panel TABLE Communication Set .............. 4-62
4.10.16 DO Setup Panel TABLE DO Set .......................................... 4-64
4.10.17 DI Setup Panel TABLE DI Set ............................................ 4-65
4.10.18 AO Setup Panel TABLE AO Set .......................................... 4-66
4.10.19 AI Set Panel TABLE AI Set ................................................ 4-67
4.10.20 ANA-BUS Set panel TABLE ANABUS Set .............................. 4-68
4.10.21 GCCU Set panel TABLE GCCU Set ................................... 4-69
4.10.22 Distillation Analysis Set panel TABLE Distillation Analysis Set ..... 4-70
◆ Revision Record ............................................................................................ i

IM 11B03A03-05E 1st Edition : Oct.31,2 001-00


Blank Page
<Toc> <Ind> <1. Basic Configuration of LCD Panel and Keys > 1-1

1. Basic Configuration of LCD Panel and Keys

1.1 Panel Configuration


Figure 1.1 shows the configuration of the LCD panel and associated keys.

LED display area LCD display area Data set/entry keys

Mark II PROCESS GAS


CHROMATOGRAPH

A B C D E F G H I

ALARM LOCAL POWER 1 2 3


J K L M N O P Q R
93/08/28
00:22:38 Chromatogram 1 4 5 6
1 TOP-1 CO2 -119.1614mV S T U V W X Y Z ∗

7 8 9
( ) + # $ : < > %

0
/ SP ,

-240sec -120sec 0 BACK


CANCEL
93/08/28 00:18 264sec Manu: 1 SPACE

Det#2 Scale# Auto Again Mgain SET / ENT

F1 F2 F3 F4 F5 F6

STATUS RESULT ALARM TEMP CHROM

DISPLAY LOCAL/
OPERATION TABLE PREV NEXT
RESET REMOTE

Subsystem restart key Function keys Panel switching keys F0101.EPS

Figure 1.1 LCD Panel and Associated Keys

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <1. Basic Configuration of LCD Panel and Keys > 1-2

1.2 Panel Switching Key


Used to switch panels between major categories.

STATUS RESULT ALARM TEMP CHROM

LOCAL/
OPERATION TABLE
REMOTE
F0102.EPS

1.3 Cursor Scroll Keys


Used to move the cursor on the display panel. Also used for scrolling panels which are
larger than the display area.

F0103.EPS

1.4 Multipurpose Function Keys


Used to activate the functions specifically defined in each panel.

F1 F2 F3 F4 F5 F6
F0104.EPS

1.5 Setup Item Call Keys


Used to select an item on the display to call up the selection.

F0105.EPS

1.6 Alpha-numeric and Backspace Keys


Alpha-numeric keys: Used to enter alphanumeric characters.
Backspace key: Used to delete the last entered character. Valid only within the same
item being entered.

A B C D E F G H I J K L M N O P Q R

1 2 3 4 5 6
S T U V W X Y Z ∗ ( ) + # $ : < > %

7 8 9 0
/ SP ,

BACK
SPACE
F0106.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <1. Basic Configuration of LCD Panel and Keys > 1-3

1.7 Cancel and Entry Keys


Cancel key: Used to cancel the entry.
Entry key: Used to complete the entry of one item.

CANCEL SET / ENT


F0107.EPS

1.8 Page Keys


Used to call the previous page and next page of a panel comprising more than one page.

PREV NEXT
F0108.EPS

1.9 Display Reset Keys


Used to reset the display if the LCD display does not function normally due to any reason.
Press this key for two seconds.

DISPLAY
RESET
F0109.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


Blank Page
<Toc> <Ind> <2. Panel Configuration Diagram > 2-1

2. Panel Configuration Diagram


The panels displayed on the LCD of the GC1000 Mark II are divided into eight major
categories and many minor sub-categories as described below.

2.1 Operation Status Display Panel STATUS

Refer to Section 4.3.

Function keys
Major panel category
switching key F1
STATUS Signal Detector Signal

F2
AI Analog/Input Signal

F3
AO Analog/Output Signal
Minor panel category
F4
DI Digital/Onput Signal

F5
DO Digital/Output Signal

F6
Table
* Not-equipped signals are not indicated.
F0201.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <2. Panel Configuration Diagram > 2-2

2.2 Analysis Result Display Panel RESULT

Refer to Section 4.4.

Function keys

F1
Latest analysis
RESULT
results

F2
Concentration analysis
historical record

F3
Retention time
historical record

F4
Laboratory analysis
results

F5 *
Standard boiling point
data table * Available only for distillation analysis systems.
F0202.EPS

2.3 Alarm Display Panel ALARM

Refer to Section 4.5.

Function keys

F1
ALARM Alarm status

F2
Alarm history

F0203.EPS

2.4 Temperature Status Display Panel TEMP

Refer to Section 4.6.

Function keys

F1
TEMP Temperature status

F2 *
Pressure status

* Available only when the analyzer is equipped with EPC.


F0204.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <2. Panel Configuration Diagram > 2-3

2.5 Chromatogram Display Panel CHROM

Refer to Section 4.7.

CHROM Chromatogram 1

F1
Chromatogram 2

F0205.EPS

2.6 Communication Mode Switching Display Panel LOCAL/


REMOTE

Refer to Section 4.8.

LOCAL/ Communication mode


REMOTE switching

Before password input After password input


F0206.EPS

2.7 Operation Panel OPERATION

Refer to Section 4.9.

Function keys

F1
OPERATION Stream command

F2
Calibration (validation)
command

F3
Manual operation

F4
Laboratory operation

F5
Range select

F6
Method set

F0207.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <2. Panel Configuration Diagram > 2-4

2.8 Parameter Setup Panel TABLE

Refer to Section 4.10.

*
*
*
*

* Indicated only when available.


F0208_1.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <2. Panel Configuration Diagram > 2-5

TABLE Table menu

System set

Operation condition set


F3
Temp set Temperature status
F6
F3
Detector signal set Detector signal
F6

Operation Valve
F2 F2
F6 F1
F3
F3 gate
Method set Sequential F1
F4
F1 Stram valve
F5 F1
F4
Stream set F2 EPC DO
F3 F1
Stream sequense set Stream command
F6

Peak common set


F2 F3
Individual peak set Peak copy 1, 2
F2
F3 Cal (val) command
F2 Cal peak
F6
F5 F2
Cal/val set Calibration set F3
F4
Cal/val command
F6 F3
F4 validate set
F3
Multi range set Range select
F6
F2 Lab method copy
F2
F3
Lab set Laboratory operation
F6
Alarm set F4 Sequential

Peak composition set F4 PC


F4 F2
F5
Communication set DCS DCS
F5 F2
F3
DO set DO status
F5
F3
DI set DI status
F5
F4 GCCU set
F5
F3
AO set AO status
F5
F3
AI set AI status
F5
ANABUS set
F4
GCCU set AO set
F5
F6
Distillation analysis set SIMDIS calib data set
F5
F0208_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


Blank Page
<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-1

3. Basic Procedures for Viewing and Entry


Any operation on the display panel is performed as a combination of the following basic
procedures.

3.1 Panel Switching 1


Press the key corresponding to the desired display panel to view it.
Procedure 1: Press the corresponding key. (For the corresponding key, refer to the
specifications for each panel.)

Corresponding key

F0301_1.EPS

Examples of panel switching 1

[Example 1]Switching between major panel categories


Use the major category switching key to switch between the eight major panel categories.
This key operation is valid in any display panel. The panel (including the minor categories)
displayed when the desired major category was used last will be displayed.

1) STATUS key Operation status display panels

2) RESULT key Analysis results display panels

3) ALARM key Alam display panels

4) TEMP key Temperature status display panels

5) CHROM key Chromatogram display panels


LOCAL/
6) REMOTE
key Communication mode switching panels

7) OPERATION key Operation panels

8) TABLE key Parameter setup panels

Each Panel
STATUS
Operation

RESULT

ALARM

TEMP

CHROM

LOCAL/
REMOTE

OPERATION

TABLE
F0301_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-2
[Example 2]Switching to a minor panel category
Major categories 1) to 6): Switching to a minor category is executed by pressing the
defined function key.
Major categories 7) and 8): A menu panel lists its minor categories. Select a category as
described in the following (Section 3.3), “Setup Item Selection
1.”
Note: For details of switching to a minor panel category, see the detailed description for each major panel.

3.2 Panel Switching 2


Press the request key corresponding to the desired panel to set further display panel
specification details.
Procedure 1: Press the corresponding request key. (The request key is referred to the
specifications of each display panel.)
Procedure 2: Use the “setup item selection 2” procedure or numerical data entry.

Corresponding request key Setup item selection 2

Numeric data entry


F0302.EPS

3.3 Setup Item Selection 1


Use the cursor keys to move the cursor to select the desired item.
1: Use the cursor key to move the cursor to the desired item.
2: Press the SET / ENT key

SET / ENT

Cursor keys
F0303_1.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-3

● Example of setup item selection 1 TABLE

(a) (a) Use the cursor key, to move


>* System Set
* Operation Condition Set the cursor (>) to the item to be set.
* Detector Signal Set
* Operation Pattern Set
[Example] Press

,then
press the key
* System Set
>* Operation Condition Set
* Detector Signal Set
* Operation Pattern Set

* System Set
* Operation Condition Set
>* Detector Signal Set
* Operation Pattern Set

(b) (b) Press the entry key, SET / ENT


* System Set
>* Operation Condition Set
* Detector Signal Set
* Operation Pattern Set

SET / ENT

(c) Operation Condition Set (c) The selected panel is displayed.


(1/6)
1 ....
2 ....
3 ....
.
.
F0303_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-4

3.4 Setup Item Selection 2


Use the setup item call keys to increment or decrement the selected data in the display
field (at the bottom left of the panel together). Each time the key is pressed, the valid
selection data is incremented or decremented by one.
Procedure 1: Use the setup item call key(s) to call the desired selection data to the
display field.
Procedure 2: Press the SET / ENT key to confirm the selection. Or press the
CANCEL key to make changes or corrections.

SET / ENT

CANCEL

F0304_1.EPS
Setup item call keys

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-5

● Example of setup item selection 2 TABLE Operation Parameter Set

(a) Operation Condition Set(1/6) (a) Use the cursor key to move
>1 Type of Carrier1 N2 the cursor to the item to be set.
2 Carrier1 Press ... ¥ keys to select the item to
3 ....
be set. (Setup item selection 1)
¥ Press the SET / ENT key.
SET / ENT

(b) Operation Condition Set(1/6) (b) The selected item appears in the display field at the bottom left
of the panel. The selected setup listing (line) on the main
>1 Type of Carrier1 N2
2 Carrier1 Press ... display area is also displayed in reverse video.
. ¥ Use the key to increment or decrement the
.
Carrier 1-Gas : H2. valid selection for the cursor item displayed in the field at the
bottom to call the data to be set.
[Example]
Press
Operation Condition Set(1/6)

>1 Type of Carrier1 N2


2 Carrier1 Press ...
. , then
.
Carrier 1-Gas : He. press the SET / ENT key

Operation Condition Set(1/6)

>1 Type of Carrier1 N2


2 Carrier1 Press ...
.
.
Carrier 1-Gas : H2.

Operation Condition Set(1/6)

>1 Type of Carrier 1 N2


2 Carrier1 Press ...
.
.
Carrier 1-Gas : He.

SET / ENT

(c) Operation Condition Set(1/6) (c) The message in the bottom left display field disappears and
the data in the main display area returns to normal from
>1 Type of Carrier 1 He
2 Carrier 1 Press ... reverse video.
Note: If a major panel category switch key is pressed before
pressing the SET / ENT key, the intended entry will
be canceled and the panel will switch to that major
: Reverse video
category.
NOTE
keys differ from are keys for use.
keys should be used to select options.
F0304_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-6

3.5 Numeric Entry


Use the numeric keys to enter numerical data.
Procedure 1: Press the numeric key(s) to enter a decimal number.
Procedure 2: Press the SET / ENT key.

Numeric Numeric SET / ENT


key(s) key(s)

CANCEL
F0305.EPS

3.6 Character Entry


Use alpha-numeric keys to enter characters (including letters, numbers, and symbols).
Procedure 1: Press alphanumeric keys.
Procedure 2: Press the SET / ENT key.

Letter key SET / ENT

Numeric CANCEL
key
BACK
SPACE
F0306_1.EPS

How to entry the alphabet


Each key represents three letters; press the alphabet key after pressing the
keys to select the desired letter.

A B C
(To select A ) A
1

A B C
(To select B ) B
1

A B C
(To select C ) C
1
F0306_2.EPS

3.7 Entry Cancellation 1


Cancels an entered error just after a wrong character was pressed during alpha-numeric
entry.
Procedure 1: Press the backspace key.

BACK
SPACE
F0307_1.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-7

● Example of numeric entry and entry cancellation 1


TABLE OPERATION CONDITION SET

(a) Operation Condition Set(1/6) (a) Use the cursor keys to select
1 Type of Carrier1 He the item to be set. (Setup item selection 1)
>2 Carrier1 Press 123.4kPa ¥ Press the key to move the
3 :
cursor to the item to be set.
¥ Press the SET / ENT key.
SET / ENT
(b) The selected item appears with the data entry cursor at the
(b) Operation Condition Set(1/6)
bottom left of the display panel.
1 Type of Carrier1 He The selected listing in the main display area is also displayed
>2 Carrier1 Press 123.4kPa
3 : in reverse video.
Carrier 1-Pressure: ¥ Use numeric keys to enter the number.
[Example]
5 Press 5
5
5
8
8
Operation Condition Set(1/6) BACK
SPACE
1 Type of Carrier1 He
>2 Carrier1 Press 123.4kPa BACK
SPACE
3 :

Carrier 1-Pressure:s558 6
BACK 7
SPACE

BACK
SPACE

8 , then
Operation Condition Set(1/6)

1 Type of Carrier1 He
press the SET / ENT key
>2 Carrier1 Press 123.4kPa
3 :

Carrier 1-Pressure:5

6
7

8
Operation Condition Set(1/6)

1 Type of Carrier1 He
>2 Carrier1 Press 123.4kPa
3 :

Carrier 1-Pressure:567.8

SET / ENT

(c) Operation Condition Set(1/6) (c) The message at the bottom left of the panel disappears and
the listing in the main display area changes from reverse video
1 Type of Carrier1 He
>2 Carrier1 Press 123.4kPa to normal display.
3 : Note: If a major panel category switch key is pressed before
pressing the SET / ENT key, the intended entry
will be canceled and the panel will be switched to that
: Reverse video
major category.
In case of entry error, press the SET / ENT key
again to return to the beginning of Step (b). F0307_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-8

● Example of character entry and entry cancellation 1 TABLE SYSTEM SET

(a) System Set (a) Use the cursor keys to select


>1 Analyzer name ABCDE the item to be set. (Setup item selection 1)
2 Analyzer ID ¥ Press the key to move the
.
. cursor to the item to be set.
¥ Press the SET / ENT key.
SET / ENT

(b) System Set (b) The selected item appears with the data entry cursor at the
>1 Analyzer name ABCDE
bottom left of the display panel.
2 Analyzer ID The selected listing in the main display area is also displayed
.
. in reverse video.
Analyzer name: ¥ Use numeric keys to enter the number.
1 [Example]
Press 1
1
System Set 1
BACK
>1 Analyzer name ABCDE SPACE
2 Analyzer ID
. 5
.
Analyzer name:AB 4 , then
BACK
SPACE
press the SET / ENT key
5
4
System Set
>1 Analyzer name ABCDE
2 Analyzer ID
.
.
Analyzer name:A5J

SET / ENT

(c) System Set (c) The message at the bottom left of the panel disappears and
the listing in the main display area changes from reverse video
>1 Analyzer name ASJ
2 Analyzer ID to normal display.
. Note: If a major panel category switch key is pressed before
.
pressing the SET / ENT key, the intended entry
will be canceled and the panel will be switched to that
: Reverse video
major category. F0307_3.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-9

3.8 Entry Cancellation 2


Cancel the entry procedure itself while entering a data item.
Procedure 1: Press the cancel key.

CANCEL
F0308.EPS

3.9 Display Page Switching


Turn the page of the panel which comprises multiple pages.
Procedure 1: Press the page key (NEXT or PREV).

PREV key : Previous page

PREV NEXT key : Next page

NEXT

Page keys

● Example of page switching TABLE OPERATION CONDITION SET

Operation Condition Set(1/6) NEXT Operation Condition Set(2/6)


(a)
>1 .... >7 ....
2 .... 8 ....
. .
. .
PREV
(b)

(c) (c)
F1 (menu) F1 (menu)

Table Menu(1/2)
* ....
>* ....
* ....
.
.
F0309.EPS

(a): To display the next page, press the NEXT key.


(b): To display the previous page, press the PREV key.
(c): To return to the menu panel, press the F1 (menu) key.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <3. Basic Procedures for Viewing and Entry > 3-10

3.10 Scroll
Scroll the display to view the items which are not currently displayed on the screen.
Procedure: Press the scroll key.

Scroll keys

● Example of scroll RESULT

Latest Result Latest Result


Composition Conc. .... Conc. ....
>1 .... (c) >1 ....
2 .... 2 ....
3 .... 3 ....
. .
. .
(a) (b) (d)

Latest Result
Composition Conc. ....
>4 ....
5 ....
6 ....
.
.

F0310.EPS

(a): To scroll the page downward by the item, press the key.
(b): To scroll the page upward by the item, press the key.
(c): To scroll the page to the right by the item, press the key.
(d): To scroll the page to the left by the item,press the keys.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-1

4. Display Panels

4.1 User Levels


There are three user levels for the GC1000 Mark II : A, B, and C. To select a level, use the
LOCAL/
REMOTE key. For details, see Section “4.8 Communication Mode Switching Display Panel ”.

To enter level B or C, you must enter a password.

[1] User level A: Password is not necessary.


OPERATION key: You cannot access the Operation Panel. (See Section 4.9, “Operation
Panel”.).
TABLE key: You can only check the parameters on the Paramater Setup Panel. (See
Section 4.10, “Paramater List Setup Panel”.)
ALARM key: You cannot erase the Alarm Historical Record Display Panel. (See Section
4.5.2, “Alarm Historical Record Display Panel”.)

[2] User level B: Password [1192] is necessary.


OPERATION key: You can check and set the Operation Panel. (See Subsection 4.9, “Opera-
tion Panel”.)
TABLE key: You can only check the parameters on the Parameter Setup Panel. (See
Section 4.10, “Parameter List Setup Panel”.)
ALARM key: You cannot erase the Alarm HistoricalRecord Display Panel. (See Section
4.5.2, “Alarm Historical Display Record Panel”.)

[3] User level C: Password [1603] is necessary.


OPERATION key: You can operate the Operation Panel. (See Section 4.9, “Operation
Panel”.)
TABLE key: You can set the parameter on the Parameters Setup Panel. (See Section
4.10, “Parameter List Setup Panel”.)
ALARM key: You can erase the Alarm Historical Record Display Panel. (See Section
4.5.2, “Alarm Historical Record Display Panel”.)

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-2

4.2 Standard Configuration of the LCD Panel


The display panel consists of the following five areas:
(a): displays time.
(b): displays titles.
(c): displays various panels.
(d): displays messages, input prompts, or echoes for the input data.
(e): function key labels

(a) Time (b) Title

(c) Panel display

(d) Message, input prompt, echoes for the input data

(e) Function key labels

F0402_1.EPS

Common information

(a)
#1/#2/#3
#4:#5:#6
( →)
(b) (←→)
(#7/#8) or (← )
(Change the page using (Change the display using
the PREV NEXT key.) the scroll key.)
F0402_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-3
• List of Display Data
No. Description Setup Maximum Minimum Unit Remarks
#1 Current time : year 00 99 Year Last two digits
#2 Current time : month 1 12 Month
#3 Current time : day 1 31 Day
#4 Current time : hour 00 23 Hour
#5 Current time : minute 00 59 Minute
#6 Current time : second 00 59 Second
#7 Page number of the displayed panel
#8 Total page number
T0402.EPS

Note: 1. For data which is unknown or is outside the minimum and maximum values, all digits of related items are displayed
as “?.”
2. For data whose conditional values are unknown or are outside the minimum and maximum values, all digits of
related items are displayed as “?.”

<Display items on the next page>


• Displayed items vary depending on the setup status.
• Time, page number, and scroll arrows are not displayed. (See Section 4.2 on page 4-
2.)
• Only defined function keys, or function keys for which labels are assigned upper part
of F1 to F6 , are enabled. The number of assigned keys varies depending on the
setup.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-4

4.3 Operation Status Display Panel STATUS

F1

4.3.1 Detector Signal Display Panel STATUS Signal

(a) When TCD is installed (b) When FID, FPD, or FI-CONV is installed
STATUS STATUS

F040301.EPS

F2 (AI): AI signal status display panel


F3 (AO): AO signal status display panel
F4 (DI): DI signal status display panel
F5 (DO): DO signal status display panel
F6 (Table): Operation status display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Detecter-* Detector No. 1 2 Page 1 indicates detector 1 with 1.
Page 2 indicates detector 2 with 2.
Type Types of detector TCD detector : TCD
FID detector : FID
FID detector methane convertor: FI-CONV
FPD detector : FPD
Not installed : None
Input signal Input signal —1000.000 1000.000 mV
Filter signal Filtered signal —1000.000 1000.000 mV
Std deviation Standard deviation of 0.000 1000.000 mV
detector signal
Voltage TCD voltage 0.0 20.0 V Displayed only when TCD is installed:
Current TCD current 0 500 mA Displayed only when TCD is installed:
Flame signal FI/FPD flame detection —20.0 20.0 mV Displayed only when FID, FPD, or FI-CONV
signal level is installed.
Therm signal FI/FPD thermocouple —20.0 20.0 mV Actual data of flame detection signal.
signal Displayed only when FID, FPD, or FI-CONV
is installed.
Flame status FI/FPD flame detection Flame on
status Flame out
Displayed only when FID, FPD, or FI-CONV
is installed.
Voltage of Voltage of methane 0.0 20.0 V Displayed only when FI-CONV is installed.
metha-con convertor
T040301.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-5
F2
4.3.2 AI Signal Status Display Panel STATUS AI

F040302.EPS

F1 (Signal): Detector signal display panel


F3 (AO): AO signal status display panel
F4 (DI): DI signal status display panel
F5 (DO): DO signal status display panel
F6 (Table): AI set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
ch# AI contact number 1 4
Name AI signal name
Value AI signal value —1.000 1.000
T040302.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-6
F3
4.3.3 AO Signal Status Display Panel STATUS AO

F040303.EPS

F1 (Signal): Detector signal display panel


F2 (AI): AI signal status display panel
F4 (DI): DI signal status display panel
F5 (DO): DO signal status display panel
F6 (Table): AO set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Ch# AO contact number 1 *
Output item Peak name during AO output
Value of analysis result during 0.000 999.999 % % unit
AO output 0.000 9999.999 ppm ppm unit
0.000 9999.999 No unit
Unit of analysis result during "%", "ppm", or "None"
AO output
Stream number during AO 1 *
output
Peak number during AO output 1 255
Value AO signal value —1.000 1.000
T040303.EPS
* : Set number of AO contact

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-7
F4
4.3.4 DI Signal Status Display Panel STATUS DI

F040304.EPS

F1 (Signal): Detector signal display panel


F2 (AI): AI signal status display panel
F3 (AO): AO signal status display panel
F5 (DO): DO signal status display panel
F6 (Table): DI set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
ch# DI contact number 1 8
Name DI signal name
Value DI signal value On Open
Off Close
Processing DI processing name Items specified for DI processing:
name No assignment
Stream sequence
Stream (cont)
Stream (1cycle)
"Cal/Val"
"Ope mode change"
Alarm process
T040304.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-8
F5
4.3.5 DO Signal Status Display Panel STATUS DO

F040305.EPS

F1 (Signal): Detector signal display panel


F2 (AI): AI signal status display panel
F3 (AO): AO signal status display panel
F4 (DI): DI signal status display panel
F6 (Table): DO set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
ch# DO contact number 1 8
Output item Output item Any of the following items is displayed:
• No processing
• Stream sequence
(The selected numbers are displayed)
• Stream
(The selected number is displayed)
• Ope mode
(The selected mode is displayed)
• Alarm (The selected alarm level or
concentration is displayed)
• Timing
Value DO signal value "On"
"Off"

For output of the stream identification signal, DO numbers are assigned as follows.
(DO1 and DO2 are normal outputs, and DO3 is a stream identification flag.)

DO No.
Number DO4 DO5 DO6 DO7 DO8
of stream
Stream 3 or before Stream Stream Normal output Normal output Normal output

Stream 7 or before Stream Stream Normal output Normal output Normal output

Stream 15 or before Stream Stream Normal output Normal output Normal output

Stream 31 or before Stream Stream Normal output Normal output Normal output

(Example) The output of the 6th stream


DO4 DO5 DO6
0 1 1 T040305.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-9

4.4 Analysis Result Display Panel RESULT

4.4.1 Latest Analysis Result Display Panel


RESULT

Analysis start date and time


Total peak number, stream number and name

F040401.EPS

F2 (Conc): Concentration analysis historical record display panel


F3 (R. Time): Retention time of historical record display panel
F4 (Lab): Laboratory analysis result display panel
F5 (Dist): Standard boiling point data table display panel *
F6 (Strm #): Stream change
Pressing F6 displays the input prompt “Strm # (number of displayed
stream):” at the bottom left of the displayed panel. Enter the number of the
stream you wish to display referring Section 3.3 and 3.4, “Setup Item
Section 1 and 2”, or Section 3.5, “Numeric Entry”.
* Available only for distillation analysis systems.

CAUTION
• Pressing the or key moves the data up or down by
one item.
• Pressing the or key moves the data to the left or right by one item.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-10
• List of Displayed Data
Display Item Description Set Minimum Maximum Unit Remarks
# Continuous number of a 1 255
peak in displayed stream
Composition Peak name
Conc. Concentration/peak analysis 0.00 999.99 % When the result is displayed in %
result 0.000 9999.999 ppm When the result is displayed in ppm
0.000 9999.999 When the result is displayed without any unit
R.Time Retention time(Peak top time) 0.0 99999.9 Second
Det # Detector number 1 2

S.Time Peak start time 0.0 99999.9 Second


E.Time Peak end time 0.0 99999.9 Second
Gate on Gate on time 0.0 99999.9 Second When the peak is followed up, the time when
the followed result is turned on is displayed.
In standard mode, the time when the gate
standard is turned on is displayed.
Gate off Gate off time 0.0 99999.9 Second When the peak is followed, the time when
the followed result is turned off is displayed.
In standard mode, the time when the gate
standard is turned off is displayed.
Tracking Peak Tracking result Standard
Tracking
None
Change rate 0.000 1.000 Detector #1
Tracking intercept —99999.9 99999.9
Change rate 0.000 1.000 Detector #2
Tracking intercept —99999.9 99999.9
S.Level Peak start level —1000.000 1000.000 mV
T.Level Peak top level —1000.000 1000.000 mV
E.Level Peak end level —1000.000 1000.000 mV
Area Peak area 0.0 9999.999 mVsec
P/H Peak shape Display in the first digit
P .................Tangent correction reset
(Blank) .......Without tangent correction reset
Display in the second digit
(Blank) .......Skimming
H ................Vertical processing
H-width Peak half-width 0.0 999.9 Second
Tail Tailing factor 0.00 99.99
Varifactor Concentration Variation factor 0.000 1.999
Calc method Concentration calculation None"
method "Correction area(%)
Calibration
External 3rd
External linear
Indirect
Year : Displayed stream 0 99 Year Last two digits
analysis start
Month : Displayed stream 1 12 Month
analysis start
Day : Displayed stream 1 31 Day
analysis start
Hour : Displayed stream 0 23 Hour
analysis start
Minute : Displayed stream 0 59 Minute
analysis start
Total peak number in 1 255
displayed stream
Displayed stream number 1 31
Displayed stream name
T040401.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-11
F2
4.4.2 Concentration Analysis Historical Record Display Panel RESULT Conc

RESULT

F2

Analysis start date


Total peak number, stream number and name

Analysis start time


F040402.EPS

F1 (Last): Latest analyzed result display panel


F3 (R.Time): Retention time of historical record display panel
F4 (Lab): Laboratory analysis result display panel
F5 (Dist): Standard boiling point data table display panel *
F6 (Strm #): Stream change
Pressing F6 displays the input prompt “Strm # (number of displayed
stream):” at the bottom left of the displayed panel. Enter the number of the
stream you wish to display referring Section 3.3 and 3.4, “Setup Item
Section 1 and 2”, or Section 3.5, “Numeric Entry”.
* Available only for distillation analysis systems.

CAUTION
Pressing the or key moves the data up or down by one
item.

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
# Continuous peak 1 255
number in displayed stream
Comp. Peak name
Before * Concentration/peak 0.00 999.99 % When the result is displayed in % .
analysis result Note that unit is not displayed.
0.000 9999.999 ppm/— When the result is displayed in ppm or
without unit.
Note that unit is not displayed.
0 11 Times The latest result is 0 times ago.
ago The maximum value is decreased
depending on the peak setup value.
Total peak number in 1 255
displayed stream
Number of displayed stream 1 31
Name of displayed stream
Year : analysis start 0 99 Year Last two digits
Month : analysis start 1 12 Month Date of the latest data being displayed
Day : analysis start 1 31 Day
Hour : analysis start 0 23 Hour
Minute : analysis start 0 59 Minute
T040402.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-12
F3
4.4.3 Retention Time of Historical Record Display Panel RESULT .
R Time

RESULT

F3

Analysis start date


Total peak number, stream number and name

Analysis start time

F040403.EPS

F1 (Last): Latest analysis result display panel


F2 (Conc): Concentration analysis historical record display panel
F4 (Lab): Laboratory analysis result display panel
F5 (Dist): Standard boiling point data table display panel *
F6 (Strm #): Stream change
Pressing F6 displays the input prompt “Strm # (number of displayed
stream):” at the bottom left of the displayed panel. Enter the number of the
stream you wish to display referring Section 3.3 and 3.4, “Setup Item
Section 1 and 2”, or Section 3.5, “Numeric Entry”.
* Available only for distillation analysis systems.

CAUTION
Pressing the or key moves the data up or down by one
item.

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
# Continuous peak 1 255
number in displayed stream
Comp. Peak name
Before * Retention time 0.0 999.99 Second
0 11 Times The latest result is 0 times ago.
ago The maximum value is decreased depending
on the peak setup value.
Total peak number in 1 255
displayed stream
Number of displayed stream 1 31
Name of displayed stream
Year : analysis start 0 99 Year Last two digits
Month : analysis start 1 12 Month
Day : analysis start 1 31 Day Date of the latest data being displayed
Hour : analysis start 0 23 Hour
Minute : analysis start 0 59 Minute
T040403.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-13
F4
4.4.4 Laboratory Analysis Result Display Panel RESULT Lab

RESULT

F4

Analysis start date and time


Total peak number

F040404.EPS

F1 (Last): Latest analysis result display panel


F2 (Conc): Concentration analysis historical record display panel
F3 (R.Time): Retention time historical record display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Conc. Concentration 0.000000 100.000000 %
Det # Detector #1 1 2
S.Time Peak start time 0.0 99999.9 Second
R.Time Retention time (Peak top time) 0.0 99999.9 Second
E.Time Peak end time 0.0 99999.9 Second
S.Level Peak start level —1000.000 1000.000 mV
T.Level Peak top level —1000.000 1000.000 mV
E.Level Peak end level —1000.000 1000.000 mV
Area Peak area 0.0 9999.999 mVsec
P/H Peak shape Display in the first digit
P .................Tangent correction reset
(Blank) ........Without tangent correction reset
Display in the second digit
(Blank) ........Skimming
H .................Vertical processing
H-width Peak half-width 0.0 999.9 Second
Tail Tailing constant 0.00 99.99
Total peak number in 1 255
displayed task
Year : analysis start 0 99 Year Last two digits
Month : analysis start 1 12 Month
Day : analysis start 1 31 Day
Hour : analysis start 0 23 Hour
Minute : analysis start 0 59 Minute
T040404.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-14
F5
4.4.5 Standard B.P Data Table Display Panel RESULT Lab

This display is shown only when the analyzer is used in a distillation analysis system.

RESULT

F5

F040405.EPS

F1 (Last): Latest analysis result display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
% Fraction % 0.0 100.0 %
Elapsed time 0.0 99999.9 Second
B.P 0.0 999.9 According to temperature unit setup on the
Distillation Analysis Set panel.
T040405.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-15

4.5 Alarm Display Panel ALARM

4.5.1 Alarm Status Display Panel


ALARM

F040501.EPS

F2 (History): Alarm historical record display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Total Total alarm number issued
ALM # Alarm number 1 350
Contents Alarm description
Date Alarm time, Month 1 12 Month
Alarm time, Day 1 31 Day
Time Alarm time, Hour 00 23 Hour
Alarm time, Minute 00 59 Minute
T040501.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-16
F2
4.5.2 Alarm Historical Record Display Panel ALARM Histry

ALARM

F2

F040502.EPS

F1 (Alarm): Alarm status display panel


F3 (Auto): Returns to automatic update status.
F5 (Clean): Erases the alarm historical records and enters automatic update status.
Note that this key is valid only at user level C.

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Total Total alarm number in
the record table
Date Alarm time, Month 1 12 Month
Alarm time, Day 1 31 Day
Time Alarm time, Hour 00 23 Hour
Alarm time, Minute 00 59 Minute
ALM # Alarm on/off display On ..................... 1
Off ..................... 0
Contents Alarm number 1 350
Alarm description
Alarm data 1 0 999
Alarm data 2 0 999
T040502.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-17

4.6 Temperature Status Display Panel TEMP

4.6.1 Temperature Status


(a) For single oven (b) For dual oven

F040601.EPS

F2 (EPC): Pressure status display panel (only for EPC-equipped analyzer)


F6 (Table): Temperature set display panel

a) For GC1000S/GC1000E/GC1000C (Single oven)

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Isotherm oven Isothermal oven setup 55 See the °C Maximum : 225 for non explosion-protection
temperature (SV) remarks to 225 for explosion-protection T2
the right. 145 for explosion-protection T3
95 for explosion-protection T4
Isothermal oven present —10 999 °C
temperature (PV)
Isothermal oven comment None
LSV1 LSV1 setup temperature 60 See the °C Maximum : 250 for non explosion-protection
(SV) remarks to 225 for explosion-protection T2
the right. 145 for explosion-protection T3
95 for explosion-protection T4
LSV1 present —10 999 °C Heater On/Off status is displayed when LSV1
temperature (PV) is installed.
LSV1 comment On
Off
If LSV1 is not installed : ............ None
LSV2 LSV2 setup temperature 60 See the °C Maximum : 250 for non explosion-protection
(SV) remarks to 225 for explosion-protection T2
the right. 145 for explosion-protection T3
95 for explosion-protection T4
LSV2 present —10 999 °C Heater On/Off status is displayed when LSV2
temperature (PV) is installed.
LSV2 comment On
Off
If LSV2 is not installed : ............ None
FPD FPD setup temperature 0 60 °C Displayed only when FPD is installed.
(SV)
FPD present temperature —10 999 °C
(PV)
FPD comment Heater On/Off status and "FPD" are
displayed when FPD is installed.
If FPD is not installed : ............ None
T040601_1.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-18
b) For GC1000D/GC1000T/GC1000W (Dual oven)

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Column oven Column oven setup 5 See the °C Maximum : 320 for non explosion-protected
temperature (SV) remarks to 320 for explosion-protected T1
the right. 225 for explosion-protected T2
145 for explosion-protected T3
95 for explosion-protected T4
Column oven present —10 999 °C
temperature (PV)
Column oven comment Temp Up (Heating in progress by the heating
program)
Temp stable (Controlling at a constant
temperature)
Ready (Temperature control not yet completed)
If the temperature controller status is not yet
registered : ............ None
Isothermal Isothermal oven setup 55 See the °C Maximum : 225 for non explosion-protection
oven temperature (SV) remarks to 225 for explosion-protection T1
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
Isothermal oven —10 999 °C
present temperature (PV)
Isothermal oven Temp Up (Heating in progress by the heating
comment program)
Temp stable (Controlling at a constant
temperature)
Ready (Temperature control not yet completed)
If the temperature controller status is not yet
registered : ............ None
LSV1 LSV1 set temperature 60 See the °C Maximum : 250 for non explosion-protection
(SV) remarks to 250 for explosion-protection T1
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
LSV1 present —10 999 °C
temperature (PV)
LSV1 comment Heater On/Off status is displayed when LSV1
is installed.
On
Off
If LSV1 is not installed : ............ None
LSV2 LSV2 Set temperature 60 See the °C Maximum : 250 for non explosion-protection
(SV) remarks to 250 for explosion-protection T1
the right. 250 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
LSV2 present —10 999 °C
temperature (PV)
LSV2 comment Heater On/Off status is displayed when LSV2
is installed.
On
Off
If LSV2 is not installed : ............ None
FID1 FID 1 setup 60 See the °C Maximum : 320 for non explosion-protection
temperature (SV) remarks to 320 for explosion-protection T1
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
FID 1 present temperature —10 999 °C
(PV)
FID 1 Comment The type of detector installed and heater
On/Off status are displayed.
T040601_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-19
• List of Displayed Data (con’t)
Display Item Description Set Minimum Maximum Unit Remarks
FID 2 FID 2 setup 60 See the °C Maximum : 320 for non explosion-protection
temperature (SV) remarks on 320 for explosion-protection T1
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
FID 2 present temperature —10 999 °C
(PV)
FID 2 Comment The type of detector installed and heater
On/Off status are displayed.
T040601_3.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-20

4.6.2 Column Oven Temperature Display Panel (only when EPC is


installed)

F040602.EPS

F1 (Temp): Temperature status display panel


F6 (Table): Operation condition set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Carrier 1/2 Carrier 1/2 setup 0.0 999.9 See the According to pressure unit setup on the
pressure (SV) remarks Operation Condition Set display
Carrier 1/2 present 0.0 999.9 to kPa ............ 0
pressure (PV) the right. PSI ............ 1
Carrier 1/2 comment "Press up"
"Press stable"
"Ready"
Hydrogen 1/2 hydrogen 1/2 0.0 999.9 See the According to pressure unit setup on the
setup pressure (SV) remarks Operation Condition Set display
hydrogen 1/2 0.0 999.9 to kPa ............ 0
present pressure (PV) the right. PSI ............ 1
hydrogen 1/2 "Press up"
comment "Press stable"
"Ready"
Air 1/2 air 1/2 0.0 999.9 See the According to pressure unit setup on the
setup pressure (SV) remarks Operation Condition Set display
air 1/2 0.0 999.9 to kPa ............ 0
present pressure (PV) the right. PSI ............ 1
air 1/2 "Press up"
comment "Press stable"
"Ready"
T040602.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-21

4.7 Chromatogram Display Panel CHROM

4.7.1 Chromatogram Display Panel


CHROM
Stream number and name, peak name,
signal value (at standard time)

Displayed standard time


Analysis start date and time, analysis time,
displayed gain mode, displayed gain value
Off set value
F040701.EPS

F1 (Det #1/Det #2): Changes the display between detectors 1 and 2.


F2 (Scale #)): Changes the time axis scale.
Pressing F2 displays the input prompt “Scale # (number of displayed
scale):” at the bottom left of the displayed panel. Enter the number of the
stream you wish to display referring Section 3.3 and 3.4, “Setup Item
Section 1 and 2”, or Section 3.5, “Numeric Entry”.
(For details of the time axis scale, see the remarks of the list of displayed
data on the next page.)
F3 (Auto): Displays a chromatogram based on the gain setting specified in the peak
individual setting
F4 (A.Zero): When in Manual, pressing F4 allows the present signal value to be offset
value.
F5 (A gain): Displays a chromatogram based on the gain setting specified in the peak
individual setting.
F6 (M gain): Displays Manual Gain and Gain Change.
Change the gain in the same manner as F2 (Scale#). (For details of gain#,
see “Displayed gain value” in the list of displayed data.)

CAUTION
Initially the display panel appears in the Automatic Update Status Mode.
Pressing the or key stops the automatic update, and enable to change the
display standard time.
Pressing the or key enable to carry out the bias operation.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-22
• List of Displayed Data
Description Set Minimum Maximum Unit Remarks
Current stream number 1 31
Current stream name
Peak name of display standard time Displayed only in real time display mode.
Signal voltage of display standard time. —1000.000 1000.000 mV
Display time axis comment s Set the scale of the time axis (X axis) using
the F2 function key as shown below.
Scale# #1 #2
0 —30 —15
1 —60 —30
2 —120 —60
3 —240 —120
4 —480 —240
5 —960 —480
6 —1920 —960
7 —3840 —1920
Gate mark Gate on: A vertical line is displayed
downward by 6 dots.
Gate off: A vertical line is displayed upward
by 6 dots.
Peak mark Peak on: A vertical line is displayed
downward by 6 dots.
Peak off: A vertical line is displayed upward
by 6 dots.
Year : current stream start 00 99 Year Last two digits
Month : current stream start 1 12 Month
Day : current operating stream start 1 31 Day
Hour : current stream start 0 23 Hour
Minute : current stream start 0 59 Minute
Analysis time 0 99999 Second analysis time from start *
Displayed gain mode Automatic ............. Auto
Manual .................. Manu
Displayed gain value s 0 15 Signal voltage per 10 scales
1V at 0 .....................1 V
s 0.5V at 1 ..................0.5 V
0.25V at 2 ................0.25 V
0.125V at 3 ..............0.125 V
............
Displayed off set valve —1000.0000 1000.0000
Displayed detector number 1 2 2 for displaying detector 1
s 1 for displaying detector 2
s is displayed.
* After output of analysis results, "??/??/??" T040701.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-23

4.8 Communication Mode Switching Display Panel LOCAL/


REMOTE

4.8.1 Communication Mode Switching Display Panel

a) In Remote mode (User level A)


LOCAL/
REMOTE

F040801_A.EPS

This panel can be accessed in Remote mode.


Refer to Section 3.5, “Numeric Enter”, and enter the four-digit password.
(Completed by pressing the SET / ENT key)
Numbers which have already been entered are displayed in reverse video. When the
password is accepted, the mode enters Local mode.
[1] When the password [1192] for user level B is entered, the mode enters user level B.
[2] When the password [1603] for user level C is entered, the mode enters user level C.

b) In Local mode (User level B or C)


LOCAL/
REMOTE

F040801_B.EPS

This panel can be accessed in Local mode.


Press the SET / ENT key to enter the Remote mode for user level A.
* For user level, refer to Section 4.1, “User Levels”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-24

4.9 Operation Panel OPERATION

4.9.1 Operation Menu Panel


OPERATION

F040901.EPS

F1 (Stream): Stream command display panel (at user level B or higher)


F2 (Cal): Calibration (Validation) command display panel (at user level B or higher)
F3 (Manual): Manual operation display panel (at user level C)
F4 (Lab): Laboratory operation display panel (at user level C)
F5 (Range): Range select display panel (at user level B or higher)
F6 (Method): Method set display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-25
• List of Displayed Data
Display Item Description Set Minimum Maximum Unit Remarks
Ope mode Operation mode “Stop”, “Run”, "Pause"
Operation request mode s Operation request modes can be changed
as follows: Stop

Run Pause

When "Stop" or "Pause" command is the


next measurement after the operation mode
has been requested, the command can be
canceled.
Meas state Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Next meas Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Analysis time Analysis time 0 99999.9 Second
Method Method 1 4
Active stream Active stream 1 31
Stream number 1 31
Status of stream "Off" "On"
Next stream Next stream 1 31
Stream number 1 31
Status of stream "Off" "On"
Detector Types of detector TCD detector: ........... "TCD"
FID detector: ............ "FID"
FID detector with methane
convertor: ............ “FI-CONV”
FPD detector: ........... "FPD"
Not installed: ............ "None"
Operation status of Operating: ................ "On"
detectors Stopped: .................. "Off"
Not installed: ............ "Blank"
Valve Status of valve Operating: ................ "On"
Stopped: .................. "Off"
Not installed: ............ "None"
T040901.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-26

4.9.2 Stream Command Display Panel


OPERATION

F1

F040902.EPS

F2 (Ope): Operation menu display panel


F6 (Table): Stream sequence set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Meas state Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Next meas Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Analysis time Analysis time 0 99999.9 Second
Active stream Active stream 1 31
Stream number 1 31
Status of stream "Off" "On"
Ope mode Operation mode “Stop”, “Run”, "Pause"
Operation request mode s Operation request modes can be changed
as follows: Stop

Run Pause
When "Stop" or "Pause" command is the
next measurement after the operation mode
has been requested, the command can be
canceled.
Str seq select Stream seqence select s 1 4
Str (cont) select Stream (cont) select s 1 31
Str (1cycle) select Stream (1 cycle) select s 1 31
Selection cancel Selection cancel s Normally not indicated (indicated at user level C)
T040902.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-27

4.9.3 Calibration (Validation) Command Display Panel


OPERATION

F2

F040903.EPS

F2 (Ope): Operation menu display panel


F6 (Table): Calibration/Validation set (main) display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Meas state Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Next meas Status of measurment "(space)", "Str sequence", "Str (cont)",
"Str (1cycle)", "Calibration", "Validation"
Number • Stream sequence number for Stream
sequence
• Stream number for stream (1 cycle/
continuous)
• Calibration/Validation number for
Calibration/Validation
Analysis time Analysis time 0 99999.9 Second
Active stream Active stream 1 31
Stream number 1 31
Status of stream "Off" "On"
Ope mode Ooperation mode “Stop”, “Run”, "Pause"
Operation request mode s Operation request modes can be changed
as follows: Stop

Run Pause
When "Stop" or "Pause" command is the
next measurement after the operation mode
has been requested, the command can be
canceled.
Cal (Val) No. Calibration (Validation) s "Cal1", "Cal2", "Cal3", "Val1", "Val2", "Val3"
select select Normally not indicated
Selection cancel Selection cancel s Normally not indicated (indicated at user level C)
Cal (Val) start/ Calibration (Validation) s "Start", "Stop" are indicated only when in
stop start/stop Manual
Purge sample Purge sample "No", "Yes" can be selected only when in Manual
T040903.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-28

4.9.4 Manual Operation Panel


OPERATION

F3

F040904.EPS

F2 (Ope): Operation menu display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Ope mode Manual operation s Stop , Run
Analysis time Analysis time 0 99999.9 Second
Det1/2 Operation status of Operating: ................ "On"
operation detectors Stopped: .................. "Off"
Not installed: ............ "Blank"
Types of detector TCD detector: ........... "TCD"
FID detector: ............ "FID"
FID detector with methane
convertor: ............ FI-CONV
FPD detector: ........... "FPD"
Not installed: ............ "None"
Str valve Stream No. , status of s 1 31 Stream No. On/Off
stream valve
VLV#1-#8 Valve s Operating: ................ "On"
Stopped: .................. "Off"
T040904_1.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-29
(a)For Dual oven

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Column oven heater Activated ..................... On
Column oven
heater
operation status s Not activated ............... Off
Installation status Not installed ................ None
Installed .......................(Blank)
Isothermal oven heater Activated ..................... On
Isothermal
oven heater
operation status s Not activated ............... Off
Installation status Not installed ................ None
Installed .......................(Blank)
Liquid sampling valve 1 Activated ..................... On
LSV 1 heater
heater operation status s Not activated ............... Off
Installation status Not installed ................ None
Installed .......................(Blank)
Liquid sampling valve 2 Activated ..................... On
LSV 2 heater
heater operation status s Not activated ............... Off
Installation status Not installed ................ None
Installed .......................(Blank)
Detector 1 heater operation Activated ..................... On
FID 1 heater
status Installation status s Not activated ............... Off
Not installed ................ None
Detector 1 heater operation Activated ..................... On
FID 2 heater
status Installation status s Not activated ............... Off
Not installed ................ None
T040904_2.EPS

(b)For Single oven

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Activated ..................... On
Isothermal oven operation
Isothermal Not activated ............... Off
oven heater
status of heater Installation s Not installed ................ None
status
Installed .......................(Blank)
Activated ..................... On
Liquid sampling valve 1
LSV 1 heater Not activated ............... Off
operation status of heater s Not installed ................ None
Installation status
Installed .......................(Blank)
Activated ..................... On
Liquid sampling valve 2
LSV 2 heater Not activated ............... Off
operation status of heater s Not installed ................ None
Installation status
Installed .......................(Blank)
Activated ..................... On
operation status of FPD
FPD heater
heater s Not activated ............... Off
(indicated only when installed )
T040904_3.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-30

4.9.5 Laboratory Operation Panel


OPERATION

F4

F040905.EPS

F2 (Ope): Operation menu display panel


F6 (Table): Laboratory set (main) display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Ope mode Operation mode Stop , Run
Operation request mode s "Stop" command when operation mode is in "Run".
Analysis time Analysis time 0.0 99999.9 Second
Detecter Types of detector TCD detector: ........... "TCD"
FID detector: ............ "FID"
FID detector with methane
convertor: ............ FI-CONV
FPD detector: ........... "FPD"
Not installed: ............ "None"
Operation status of Operating: ................ "On"
detectors Stopped: .................. "Off"
Not installed: ............ "Blank"
Valve Status of valve Operating: ................ "On"
Stopped: .................. "Off"
T040905.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-31

4.9.6 Range Select Display Panel


OPERATION

F5

F040906.EPS

F2 (Ope): Operation menu display panel


F3 (Exec): Range select execution
F6 (Table): Multiple range set display panel

• List of Displayed Data


Display Item Description Set Minimum Maximum Unit Remarks
Stream number Range specified stream (0), 1 31 (0) represents no range select ** indication
Peak number Range specified peak (0), 1 255 (0) represents no range select ** indication
Range number Range number (0), 1 31 (0) represents no range select ** indication
T040906.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-32

4.10 Parameter List Setup Panel TABLE

In user level C, each item can be set.


In user level B, each item can only be checked.
The parameter setup panel provides the menu options in the same manner as “4.9 Opera-
tion Panel.” This section describes the setup items and function keys displayed in each
menu.
The setup items displayed on each menu panel vary depending on the setup conditions.
<Table Menu>
System Set
Operation Condition Set
Temperature Set
Detector Signal Setting
Method Set
Stream Set
Stream Sequence Set
Peak Common Set
Peak Individual Set
Calibration/Validation Set
Multi Range Set
Lab Set
Alarm Set
Peak Composition Set
Communication Set
DO Set
DI Set
AO Set
AI Set **
ANABUS Set **
GCCU Set **
Distillation Analysis Set *
* Availavle only for distillation analysis systems.
** Indicated only when installed.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-33

4.10.1 System Set TABLE SYSTEM SET

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Analyzer name Analyzer name s Maximum 8 letters
Analyzer ID s 1 240 Maximum 8 letters
Valve system Valve system name s Maximum 8 letters
Column system Column system name s Maximum 8 letters
Current date Setup date (year) s 1993 2200 Year
Setup date (month) s 01 12 Month
Setup date (day) s 01 31 Day
Current time Setup time (year) s 00 23 Hour
Setup time (month) s 00 59 Minute
Setup (day) s 00 59 Second
Auto start set s Yes No
Auto start time s 0 10000 Second
Language "Japanese" "English"
T041001.EPS

F1 (Menu): Table menu display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-34

4.10.2 Operation Condition Set TABLE Operation Condition Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks

Type of carrier 1 Type of carrier gas 1 s H2 , He , N2 , Ar , No use


Depends on set EPC is suffixed when EPC is
Carrier 1 pressure Pressure of carrier gas 1 s
gas pressure unit installed.
Combustion Pressure value of Depends on set EPC is suffixed when EPC is
s installed.
H2 1 pressure combustion hydrogen 1 gas pressure unit
Combustion air 1 Pressure value of Depends on set EPC is suffixed when EPC is
s installed.
pressure combustion air 1 gas pressure unit
Depends on set
Sample 1 volume s
gas pressure unit

Depends on set
Sample 1 flowrate s
vent flowrate unit
Depends on set
DET1 Flowrate s
vent flowrate unit
Depends on set
REF1 Flowrate s
vent flowrate unit
Depends on set
BE1 Flowrate s
vent flowrate unit
Depends on set
SPLIT1 Flowrate s
vent flowrate unit
Depends on set
FF1 Flowrate s
vent flowrate unit
Depends on set
FF3 Flowrate s
vent flowrate unit

Type of carrier 2 Type of carrier gas 2 s H2 , He , N2 , Ar , No use


Pressure value of Depends on set
Carrier 2 pressure s H2 , He , N2 , Ar , No use
carrier gas 2 gas pressure unit
Combustion Pressure value of Depends on set
s H2 , He , N2 , Ar , No use
H2 2 pressure combustion hydrogen 2 gas pressure unit
Supporting air 2 Pressure value of Depends on set
s H2 , He , N2 , Ar , No use
pressure supporting air 2 gas pressure unit
Depends on set
Sample 2 volume s
gas pressure unit

Depends on set
Sample 2 Flowrate s
vent flowrate unit
Depends on set
DET2 Flowrate s
vent flowrate unit
Depends on set
REF2 Flowrate s
vent flowrate unit
Depends on set
BF2 Flowrate s
vent flowrate unit
Depends on set
SPLIT2 Flowrate s
vent flowrate unit
Depends on set
FF2 Flowrate s
vent flowrate unit
Depends on set
FF4 Flowrate s
vent flowrate unit

Pressure unit Gas pressure unit setup s kPa , PSI


Sample volume
Sample volume unit setup s ml , µl
unit
Flowrate unit Vent flowrate unit setup s ml/min
T041002.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-35

4.10.3 Temp Set TABLE Temp Set

(a) For Dual Oven


• List of Displayed Data
Setup Item Description Set Minimum Maximum Unit Remarks
Max temp column Maximum temperature s 5 See the ¡C Maximum : 320 for non explosion-
oven column oven remarks to protection
the right. 320 for explosion-protection T1
225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
Set temp column s 5 Maximum ¡C
oven temperature in
programmed-
temperature
oven
Set temp isothermal s 55 See the ¡C Maximum : 225 for non explosion-
oven remarks to protection
the right. 225 for explosion-protection T1
225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
LSV1 Set Temp s 60 See the ¡C Maximum : 250 for non explosion-
remarks to protection.
the right. 250 for explosion-protection T1
LSV2 Set Temp s 60 See the ¡C 225 for explosion-protection T2
remarks to 145 for explosion-protection T3
the right. 95 for explosion-protection T4

Det 1 set temp s 60 See the ¡C Displayed only when FID, or


remarks to FI-CONV is installed.
the right. Maximum : 320 for non explosion-
Det 2 set temp s 60 See the ¡C protection
remarks to 320 for explosion-protection T1
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
T041003_A.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-36
(b)For Single Oven
• List of Displayed Data
Setup Item Description Set Minimum Maximum Unit Remarks
Max temp isothermal Maximum temperature s 55 See the ¡C Maximum : 225 for non explosion-
oven isothermal oven remarks on protection
the right. 225 for explosion-protection T2
145 for explosion-protection T3
95 for explosion-protection T4
Set temp isothermal s 55 Maximum ¡C
oven temperature
in constant-
temperature
oven
LSV1 set temp s 60 See the ¡C Maximum : 250 for non explosion-
remarks to the protection
right. 225 for explosion-protection T2
LSV2 set temp s 60 See the ¡C 145 for explosion-protection T3
remarks to 95 for explosion-protection T4
the right.
FPD set temp s 0 60 ¡C Displayed only when FPD is installed
T041003_B.EPS

F1 (Menu): Table menu display panel


F3 (Status): Temperature status display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-37

4.10.4 Detector Signal Settings TABLE Detector Signal Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Detector 1/2 Type of installed TCD detector: ........... "TCD"
detector 1/2 FID detector: ............ "FID"
FID detector with methane
convertor: ............ FI-CONV
FPD detector: ........... "FPD"
Not installed: ............ "None"
Filtering const Filtering constant of s 0.001 1.000
detector 1/2
Sample rate Sampling rate of s Ten types of rate:
detector 1/2 20ms x 2n (n=0 to 9)
Square root calc req Square calculation s Displayed only when FPD is installed.
request for detector 1/2 None ...Square computation
not specified
"Method-1" ...Square computation
method 1
"Method-2"...Square computation
method 2
Flame detect level Flame detection signal s -20 20 mV Displayed only when FID, FPD, or
level of detector 1/2 FI-CONV is installed.
Auto zero Auto zero s "Auto" "Manual"
Auto zero value Auto zero value s -1000.0000 1000.0000 mV
T041004.EPS

F1 (Menu): Table menu display panel


F3 (Status): Detector signal display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-38

4.10.5 Method Set TABLE Method Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream Stream number 1 31 Only for streams with assigned peaks
Method number Method number s 1 4
Analysis cycle Analysis cycle s 0.00.0 28800.0 Second
Warming up time Warming up time s 0.0 (Note) 99999.9 Second
End of peak detect End of peak detect s 0 Analyzing cycle Second
Pause time Pause time s 0 Analyzing cycle Second
Inlet control temp Inlet control temp s 5 Max. temp ¡C Only for dual oven
Inlet temp time Inlet temp time s 0.0 Analyzing cycle Second Only for dual oven
T04105_1.EPS

NOTE: The warming up time, which decides the minimum value of the entry range for the param-
eters, is actually used as a negative value to restrict the range because the time to start
analyzing is assumed to be 0.0 [sec].
(For example, if the warming up time is 100.0 [sec] and the analysis cycle is 1000.0, the
valid entry range of the on-off time of the stream valves is -100.0 to 1000.0 seconds.)

Time to start analyzing


Warming up time Analyzing cycle
100.0 sec 1000.0 sec

-100.0 sec 0.0 sec. 1000.0 sec

Valid entry range of the ON time


of the sample-line valves

(Minimum) (Maximum)
F041005.EPS

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
First temp gradient Rising speed to first temperature stage s 0.0 30.0 ¡C/sec Only
of column oven Dual oven
First control temp First temperature stage setpoint of s 5 High-limit temperature ¡C Only
column oven of column oven Dual oven
First temp time Duration of first temperature stage of s 0.0 Analysis cycle Second Only
column oven Dual oven
Second gradient temp Rising speed to second temperature s 0.0 30.0 ¡C/sec Only
stage of column oven Dual oven
Second control temp Second temperature stage setpoint of s 5 High-limit temperature ¡C Only
column oven of column oven Dual oven
Second temp time Duration of second temperature stage s 0.0 Analysis cycle Second Only
of column oven Dual oven
Third temp gradient Rising speed to third temperature s 0.0 30.0 ¡C/sec Only
stage of column oven Dual oven
Third control temp Third temperature stage setpoint of s 5 High-limit temperature ¡C Only
column oven of column oven Dual oven
Third temp time Duration of third temperature stage of s 0.0 Analysis cycle Second Only
column oven Dual oven
T041005_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-39
F1 (Menu): Table menu display panel
F3 (Cmd): Operation menu display panel
F4 (Seq): Sequential display panel
For details on the sequential display panel, see the following section.
F5 (EPC): EPC set display panel (indicated only when EPC is installed)
F6 (Strm#): Change the stream number.
Pressing F6 displays the input prompt of “ Sequence number: ” at the
bottom left of the panel. Enter the number of the stream you wish to display
referring to Section 3.5, “Numeric Entry”.

(a)Sequential Display

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream Stream number 1 31 512 steps max. per stream
Time Action time -99999.9 99999.9 Second
Contents of action Sequence name Strv on/off: Stream valve On/Off
Vm 1st on/off: 1st time of valve
number m On/Off
Vm 2nd on/off: 2ndtime of valve
number m On/Off
Vm 3rd on/off: 3rd time of valve
number m On/Off
Pn gate on/off: Gate of peak number
n On/Off
Pause time: Pause time
Peak fin time: Pealk fin time
DO on/off: DO On/Off
T04105_3.EPS

F1 (Main): Method set (main) display panel


F2 (Valve): Action set (Valve) display panel
F3 (Gate): Action set (Gate) display panel
F4 (StrV): Action set (StrV) display panel
F5 (DO): Action set (DO) display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-40
(i) Action Set Display (Valve)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Valve number Valve number 1 8
1st ON time Time to start first opening s -Warming Analysis Second
up time cycle-2
1st OFF time Time to start first closing s -Warming Analysis Second
up time cycle-2
2nd ON time Time to start second opening s -Warming Analysis Second
up time cycle-2
2nd OFF time Time to start second closing s -Warming Analysis Second
up time cycle-2
3rd ON time Time to start third opening s -Warming Analysis Second
up time cycle-2
3rd OFF time Time to start third closing s -Warming Analysis Second
up time cycle-2
T041005_4.EPS

F1 (Seq): Sequential display panel


F3 (Gate): Action set (Gate) display panel
F4 (StrV): Action set (StrV) display panel
F5 (DO): Action set (DO) display panel
F6 (Vlv#): Changes the valve number.
Pressing F6 displays the input prompt of “ Sequence number: ” at the
bottom left of the panel. Enter the number of the valve you wish to display
referring to Section 3.5, “Numeric Entry”.

(ii) Action Set Display (Gate)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Peak number Peak number 1 255
Gate ON time Time to start gate opening s 0.0 99999.9 Second
Gate OFF time Time to start gate closing s 0.0 99999.9 Second
T041005_5.EPS

F1 (Seq): Sequential display panel


F2 (Valve): Action set (Valve) display panel
F4 (StrV): Action set (StrV) display panel
F5 (DO): Action set (DO) display panel
F6 (Peak#): Changes the peak number.
Pressing F6 displays the input prompt of “ Sequence number: ” at the
bottom left of the panel. Enter the number of the peak you wish to display
referring to Section 3.5, “Numeric Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-41
(iii) Action Set Display (Str valve)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream valve ON time Time to start valve opening s -Warming Analysis Second
up time cycle-2
Stream valve OFF time Time to start valve closing s -Warming Analysis Second
up time cycle-2
T041005_6.EPS

F1 (Seq): Sequential display panel


F2 (Valve): Action set (Valve) display panel
F3 (Gate): Action set (Gate) display panel
F5 (DO): Action set (DO) display panel

(iv) Action Set Display (DO)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
DO ON time Time to start DO opening s -Warming Analysis Second
up time cycle-2
DO OFF time Time to start DO closing s -Warming Analysis Second
up time cycle-2
T041005_7.EPS

F1 (Seq): Sequential display panel


F2 (Valve): Action set (Valve) display panel
F3 (Gate): Action set (Gate) display panel
F4 (StrV): Action set (StrV) display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-42
(b)EPC Set Display (indicated only when EPC is installed)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream Stream number 1 31
Carrier gas Carrier gas number 1 2
Method number Method number s 1 4
Init control press Inlet control temp s kPa/PSI Set the unit on the Operation
Condition Set display.
Init press time Inlet temp time s Second

First press gradient Rising speed to first s 0.0 999.9 kPa/min


pressure 0.0 999.9 psg/min
First control press First control s 0 5000 kPa Set the unit on the Operation
pressure 0 72.5 PSI Condition Set display.
First control time First control time s 0 99999.9 Second
Second press gradient Rising speed to second s 0.0 999.9 kPa/min
pressure 0.0 999.9 psg/min
Second control press Second control s 0 5000 kPa Set the unit on the Operation
pressure 0 72.5 PSI Condition Set display.
Second control time Second control time s 0 99999.9 Second
Third press gradient Rising speed to third s 0.0 999.9 kPa/min
pressure 0.0 999.9 psg/min
Third control press Third control s 0 5000 kPa Set the unit on the Operation
pressure 0 72.5 PSI Condition Set display.
Third control time Third control time s 0 99999.9 Second
T04105_8.EPS

F1 (Menu): Table menu display panel


F2 (Method): Method set display panel
F5 (Gas#): Select the number of the carrier gas.
F6 (Strm#): Select the stream number.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-43

4.10.6 Stream Set TABLE Stream Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream Setram number s 1 31
Name of stream Stream name s Max 8 letters
Stream valve number Stream valve number s 1 31
Method number Method number s 1 4
Kind of stream The kind of stream s Measurement , Calibration ,
Validation",
"Calibration run"(only for distillation),
"Sample run"(only for distillation)
Total compens Totalized value correction s Executed
computation specified Not executed
Total value Total valve s 0.1 100.0 %
Base line corr Base line correction s Executed
calculation request Not executed
SIMDIS calib stream Distillation calibration stream s 0 31 Display
T041006.EPS

F1 (Menu): Table menu display panel


F6 (Strm#): Changes the distream number.
Pressing F6 displays the input prompt of “Stream number:” at the bottom
left of panel. Enter the number of the stream you wish to display referring to
Section 3.5, “Numeric Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-44

4.10.7 Stream Sequence Set TABLE Str Sequence Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream sequence Setram sequence number 1 4 maximum 31 streams can be
scheduled for one stream
sequence.
#01~31 Action stream Scheduled stream number s (0), 1 31 "0" cannot be set for action
stream #01.
Stream name Stream name
T041007.EPS

F1 (Menu): Table menu display panel


F3 (Cmd): Stream command display panel
F6 (Seq): Changes the displayed stream sequence.
Pressing F6 displays the input prompt of “ Sequence number: ” at the
bottom left of the panel. Enter the number of the stream you wish to display
referring to Section 3.5, “Numeric Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-45

4.10.8 Peak Common Set TABLE Peak Common Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Conc calc req s Standard
Corrected area normalization
Det slope ratio Detected slope end/start ratio s 0.00 9.99
Calib range s 0.000 1.000
Calib repeat range s 0.000 2.000
Calib coef renge s 0.000 4.000
Reserved stream s (0), 1 31 Number of stream to be saved
number as chromatogram for base line
correction.
A minimum value of 0 means
that there is no task to be
saved.
Mark output req 1 Chromatogram mark output s None
request 1 Start"
"Start-10s
Start-10s-60s
Start-60s
10s
60s
"10s-60s"
Mark output req 2 Chromatogram mark output s None
request 2 gate"
"peak
gate peak
T041008.EPS

F1 (Menu): Table menu display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-46

4.10.9 Peak Individual Set TABLE Peak Individual Set

• List of Displayed Data (1/7)


Panel Setup Item Description Set Minimum Maximum Unit Remarks
No.
Stream # Number of task being displayed 1 31
or set
Peak # Number of peak being displayed 1 255
or set
Using/Assigned Number of peak used 0 255
peak num Number of peak assigned 0 255
Process assign s External 3rd "External linear"
Indirect Base level
Signal level Noise level
Deviation calc Linear 1
Linear 2 Linear 3 Linear 4
Linear 5 Ratio
Separation capa "Divisor"
"A1-Calorific val"
"A1-Spec gravity"
1 "A1-Compress fact"
"A1-Wobbe index"
"A2-Heating v/vol"
"A2-Heating v/mas"
"A2-Relative dens"
"A2-Density"
"A2-Compress fact"
"A2-Wobbe index"
"J-Calorific val"
"J-Spec gravity"
"J-Compress fact"
"J-Wobbe index"
"I-Calorific V/mol"
"I-Calorific V/mas"
"I-Calorific V/vol"
"I-Relative dens"
"I-Density"
"I-Compress fact"
"I-Wobbe index"
"AI signal 1" "AI signal 2"
"AI signal 3" "AI signal 4"
"Std B.P proc" "RVP proc"
"Distill % proc"
Name Peak name s Max 8 letters
Execution Processing of the related peak s Execut No execut
Synchronize to Stream number of synchronized s (0), 1 31 Setting "0" in stream and peak
numbers disables synchronization,
Peak number of synchronized s (0), 1 255 resulting in independent peaks.
T041009_1.EPS

The layout of subsequent pages of this panel varies depending on specified processing in
the processing assignment item on panel No.1. (Page 1 of this panel, assigned to panel
No.1 is common irrespective of specified processing.)
Ths upper section of panel No.2 onward displays the title of the specified processing on the
right as well as the stream # and peak # specified on panel No.1.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-47
<Configuration of panels No.2 and onward>
Panel No.1 Item 4.10.10 Item 4.10.8 Page 2 Page 3 Page 4

External 3rd *1 "Calibration" "Standard" 2 3 4


"Corrected area ratio" - - -
"Validation" "Measuring" "Standard" 2 3 5
"Corrected area ratio" - - -
"External linear *1 "Calibration" "Standard" 2 3 6
"Corrected area ratio" - - -
"Validation" "Measuring" "Standard" 2 3 7
"Corrected area ratio" - - -
Indirect "Calibration" "Validation" "Standard" 2 8 9
"Measuring" *1 "Corrected area ratio" 2 3 10
"Cal run" *2 - 2 3 11
"Sample run" *2 - 2 3 10
"Base level "Calibration" "Validation" "Measuring" *1 - 12 - -
Signal level "Calibration" "Validation" "Measuring" *1 - - -
"Noise level" "Calibration" "Validation" "Measuring" *1 - - -
Deviation calc "Calibration" "Validation" "Measuring" *1 - 13 - -
Linear 1~5 "Calibration" "Validation" "Measuring" *1 - 14 - -
Ratio "Calibration" "Validation" "Measuring" *1 - 15 - -
Separation capa "Calibration" "Validation" "Measuring" *1 - 16 - -
"Divisor" "Validation" "Measuring" - 17 - -
"A1-Calorific val" *1 - - 19 - -
"A1-Spec gravity" *1 - - - -
"A1-Compress fact" *1 - - - -
"A1-Wobbe index" *1 - - - -
"A2-Heating v/vol" *1 - - 20 - -
"A2-Heating v/mas" *1 - - - -
"A2-Relative dens" *1 - - - -
"A2-Density" *1 - - - -
"A2-Compressfact" *1 - - - -
"A2-Wobbe index" *1 - - - -
"J-Calorific val" *1 - - 19 - -
"J-Spec gravity" *1 - - - -
"J-Compress fact" *1 - - - -
"J-Wobbe index" *1 - - - -
"I-Calorific V/mol" *1 - - 20 - -
"I-Calorific V/mas" *1 - - - -
"I-Calorific V/vol" *1 - - - -
"I-Relativedens" *1 - - 21 - -
"I-Density" *1 - - 22 - -
"I-Compress fact" *1 - - 20 - -
"I-Wobbe index" *1 - - 21 - -
"AI signal 1~4" (When AI - - 18 - -
signal) *1
"Std B.P proc" *2 "Sample run" - 23 - -
"RVP proc" *2 "Sample run" - 24 - -
"Distill % proc" *2 "Sample run" - 25 - -
*1: Indicated only when the system type is other than a distillation system.
*2: Indicated only when the system type is a distillation system.
T041009_2.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-48
• List of Displayed Data (2/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Detector number s 1 2
Measuring unit s % ppm No unit
Measuring range s 0.000 100.000 % For % unit
0.000 9999.999 ppm For ppm unit
0.000 9999.999 For no unit
Gate cut method s Time gate
Slope gate
2 "Slope/Time"
"Time/Slope"
"Zone"
Auto tracking req s Tracking exec
No tracking
Tracking std
Peak polarity s Positive Negative
invert
Detected slope s 0.000 9999.999 mV/sec
Detected level s —9999.999 9999.999 mV

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Peak std time s 0.0 99999.9 Second
Gate/ON time s 0.0 99999.9 Second
Gate/OFF time s 0.0 99999.9 Second
3 Integ method s Skimming Vert
method
A/H req s Area Height
Chrom gain Chromatogram gain s 0 15 Valid when auto gain
specification is set on the
chromatogram display.
Total compens s Indicated only when
totalized value
correction computation
is set to "Executed".
"Total compensation
disabled"
"Total compensation
enable"

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Factor K K of external standard s —9.999 9.999
Factor A A of external standard s —9.999 9.999
Factor B B of external standard s —9.999 9.999
4 Std conc Standard concentration of s 0.000 100.000 % For %
calibration peak 0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
Std area s 0.000 9999.999
Calib factor s 0.000 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Ref peak Reference peak : stream No. s 1 31
Reference peak : peak No. 1 255
Linear calc 1 factor s —9999.999 9999.999
5 Linear calc 2 factor s —9999.999 9999.999
Linear calc 3 factor s —9999.999 9999.999
Linear calc 4 factor s —9999.999 9999.999
Linear calc 5 factor s —9999.999 9999.999
T041009_3.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-49
• List of Displayed Data (3/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Std conc Standard concentration of s 0.000 100.000 % For %
calibration peak 0.000 9999.999 ppm For ppm
6 0.000 9999.999 For no unit
Std area s 0.000 9999.999
Calib factor s 0.000 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Ref peak 1 Reference peak 1 : stream No. s 1 31
Reference peak 1 : peak No. s 1 255
Ref peak 2 Reference peak 2 : stream No. s 1 31
Reference peak 1 : peak No. s 1 255
7 Linear calc 1 factor s —9999.999 9999.999
Linear calc 2 factor s —9999.999 9999.999
Linear calc 3 factor s —9999.999 9999.999
Linear calc 4 factor s —9999.999 9999.999
Linear calc 5 factor s —9999.999 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Peak std time s 0.0 99999.9 Second
Gate/ON time s 0.0 99999.9 Second
Gate/OFF time s 0.0 99999.9 Second
Integ method s Skimming Vert
8
method
A/H req s Area Height
Chrom gain Chromatogram gain s 0 15
specification
Total compens s "Total compensation
disabled"
"Total compensation
enable"
Sense compens coef s 0.000 999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Factor A s —9.999 9.999
Factor B s —9.999 9.999
Factor K f of corrected area ratio s —9.999 9.999
Std peak Reference peak : stream No. s 1 31
9 Reference peak : peak No. 1 255
Linear calc 1 factor s —9999.999 9999.999
Linear calc 2 factor s —9999.999 9999.999
Linear calc 3 factor s —9999.999 9999.999
Linear calc 4 factor s —9999.999 9999.999
Linear calc 5 factor s —9999.999 9999.999
T041009_4.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-50
• List of Displayed Data (4/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Factor f f of corrected area ratio s 0.000 9999.999
Linear calc 1 factor s —9999.999 9999.999
Linear calc 2 factor s —9999.999 9999.999
10
Linear calc 3 factor s —9999.999 9999.999
Linear calc 4 factor s —9999.999 9999.999
Linear calc 5 factor s —9999.999 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Std B.P f of corrected area ratio s -999.9 999.9
Set up Conc. s 0.00000 1.00000
11 Section set start s 0.0 99999.9
Section seet stop s 0.0 99999.9
Sense compens coef s —9999.999 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Detector number s 0 2
Measurement rang s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
12 0.000 9999.999 For no unit
Auto tracking reg s No tracking
Tracking exec
Tracking std
Action time s 0.0 99999.9 Second

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Result unit s % ppm No unit
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
13 Linear calc 1 factor s —9999.999 9999.999
Linear calc 2 factor s —9999.999 9999.999
Linear calc 3 factor s —9999.999 9999.999
Linear calc 4 factor s —9999.999 9999.999
Linear calc 5 factor s —9999.999 9999.999
T041009_5.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-51
• List of Displayed Data (5/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Result unit s % ppm No unit
Measuring range s 0.000 100.000 % For %
14 0.000 9999.999 ppm For ppm
a of external standard 0.000 9999.999 For no unit
Factor a b of external standard s —9.999 9.999
Factor b s —9.999 9.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
Ref peak nume Ratio (numerator) : s 1 31
15 stream number
Ratio (numerator) : s 1 255
peak number
Ref peak deno Ratio (denominator) : s 1 31
stream number
Ratio (denominator) : s 1 255
peak number

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
Ref peak l Separation capability s 1 31
16 (component m) : stream number
Separation capability s 1 255
(component l) : peak number
Ref peak m Separation capability s 1 31
(component m) : stream number
Separation capability s 1 255
(component m) : peak number

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Detector number s 0 2 "%" "ppm" "No unit"
17 Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Zero s -999.999 9999.999
18 Span s -999.999 9999.999
Action time s 0.000 99999.9 Second

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
19 0.000 9999.999 For no unit
H2 peak no s 1 20
T041009_6.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-52
• List of Displayed Data (6/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
20 0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
21 0.000 9999.999 For no unit
Compress fact.(air) s -9.9999 9.9999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
22 0.000 9999.999 For no unit
Temp. (deg.) s -999.99 999.99

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
Area% s 0.5 99.5 Except 0.5, 99.5
23 Linear calc 1 factor s -9999.999 9999.999
Linear calc 2 factor s -9999.999 9999.999
Linear calc 3 factor s -9999.999 9999.999
Linear calc 4 factor s -9999.999 9999.999
Linear calc 5 factor s -9999.999 9999.999

Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
RVP coef a s -9999.999 9999.999
24 RVP coef b s -9999.999 9999.999
Linear calc 1 factor s -9999.999 9999.999
Linear calc 2 factor s -9999.999 9999.999
Linear calc 3 factor s -9999.999 9999.999
Linear calc 4 factor s -9999.999 9999.999
Linear calc 5 factor s -9999.999 9999.999
T041009_7.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-53
• List of Displayed Data (7/7)
Panel No. Setup Item Description Set Minimum Maximum Unit Remarks
Measuring range s 0.000 100.000 % For %
0.000 9999.999 ppm For ppm
0.000 9999.999 For no unit
Std B.P s -999.999 999.999
25 Linear calc 1 factor s -9999.999 9999.999
Linear calc 2 factor s -9999.999 9999.999
Linear calc 3 factor s -9999.999 9999.999
Linear calc 4 factor s -9999.999 9999.999
Linear calc 5 factor s -9999.999 9999.999
T041009_8.EPS

F1 (Menu): Table menu display panel


F2 (Copy1): Peak copy I display panel
F3 (Copy2): Peak copy II display panel
F5 (Strm#): Changes the displayed stream number.
Pressing F5 displays the input prompt of “Strm #:” at the bottom left of the
panel. Enter the number of the stream you wish to display referring to
Section 3.5, “Numeric Entry”.
F6 (Peak#): Changes the displayed peak.
Pressing F6 displays the input prompt of “Peak #:” at the bottom left of the
panel. Enter the number of the peak you wish to display referring to Section
3.5, “Numeric Entry”.

(i) Peak Copy 1 Panel


You can copy the data of peak set between tasks.

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Total assign Total number of assigned peaks 0 255
Stream# Stream number 1 31
Stream name Stream name
Assign Total number of peaks assigned
to the stream of interest
Peak number Total number of peaks being
used in the stream of interest
T041009_9.EPS

F1 (Menu): Table menu display panel


F2 (PeakSet): Returns to the individual peak set panels.
F6 (Copy): By entering the stream numbers of the source and destination, you can
copy the stream from the former to the latter.
Pressing F6 displays the input prompt of stream numbers of the source and
destination at the bottom left of the panel. Enter the number of two streams
you wish to copy from and copy to, referring to Section 3.5, “Numeric
Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-54
(ii) Peak Copy 2 Panel

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Strm# Stream number 1 31
Use num Total number of peaks being 0 255
used in the stream of interest
Assign Total number of peaks assigned 0 255
to the related stream
Peak # Number of peak of the stream 1 255
of interest
Peak name Peak/processing name
Usage Peak usage status ON : .........Used
OFF : .......Unused
Process assign Type of processing specifications External 3rd "External
linear Indirect Base level
Signal level "Noise level"
Deviation calc Linear 1
Linear 2 Linear 3
Linear 4 Linear 5 Ratio
Separation capa "Divisor"
"A1-Calorific val" "A1-Spec
gravity" "A1-Compress fact"
"A1-Wobbe index"
"A2-Heating v/vol"
"A2-Heating v/mas"
"A2-Relative dens"
"A2-Density" "A2-Compress
fact" "A2-Wobbe index"
"J-Calorific val" J-Spec
gravity" "J-Compress fact"
"J-Wobbe index" "I-Calorific
V/mol" "I-Calorific V/mas"
"I-Calorific V/vol" "I-Relative
dens" "I-Density"
"I-Compress fact" "I-Wobbe
index" "AI signal 1"
"AI signal 2" "AI signal 3"
"AI signal 4" "Std B.P proc"
"RVP proc" "Distill % proc"
T041009_10.EPS

F1 (Menu): Table menu display panel


F2 (PeakSet): Returns to the individual peak set panels.
F6 (Copy): By entering the peak numbers of the source and destination, you can copy
the peak from the former to the latter.
Pressing F6 displays the input prompt of “Peak #:” of the source and
destination at the bottom left of the panel. Enter the number of the peaks
you wish to copy from and copy to, referring to Section 3.5, “Numeric
Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-55

4.10.10 Calibration/Validation Set TABLE Cal/Val Set

(i) Calibration/Validation Set Display (Main)

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Cal (Val) method Calibration (Validation) method s Manual" "Semi-auto" "Auto"
Auto start time Auto start time s 0 23 hour
0 59 min
Time interval Time interval s 0 255 day
0 23 hour
0 59 min
T041010_1.EPS

F1 (Menu): Table menu display panel


F3 (Cmd): Caibrationl(validation) command display panel
F5 (Cal): Ccalibration set display panel
F6 (Val): Validation set display panel

(ii) Calibration Set Display

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Calibration Calibration number 1 3
Cal stream Calibration stream number s 1 31
Cal time Calibration times s 1 30
Auto cal Auto calibration select s Yes" "No"
Val stream Validation stream number s 1 31
T041010_2.EPS

F1 (Menu): Table menu display panel


F2 (Peak): Calibration peak set display panel
F3 (Cmd): Caibrationl(validation) command display panel
F4 (Main): Calibration/validation set display panel
F6 (Cal#): Changes the calibration number.
Pressing F6 displays the input prompt of “ Cal number: ” at the bottom left
of the panel. Enter the number of the calibration you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-56
(iii) Calibration Peak Set Display

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Peak num Peak number 1 255
Peak name Peak name
Execution Execution s No execute" "Execute"
Measuring unit Measuring unit s "%", "ppm", or "No unit"
Measuring range Measuring range s 0.000 100.00 % % unit
0.000 9999.999 ppm ppm unit
0.000 9999.999 No unit
Std conc Standard concentration s 0.000 100.00 % % unit
0.000 9999.999 ppm ppm unit
0.000 9999.999 No unit
Std area Standard area s 0.000 9999.999
Calib factor Calibration factor s 0.000 9999.999
T041010_3.EPS

F2 (Cal): Calibration set display panel


F6 (Peak#): Changes the peak number.
Pressing F6 displays the input prompt of “ Peak number: ” at the bottom left
of the panel. Enter the number of the peak you wish to display.

(iv) Validation Set Display

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Validation Validation number 1 3
Val stream Validation stream number s 1 31
Val times Validation times s 1 30
Auto val Auto validation select s Yes" "No"
T041010_4.EPS

F1 (Menu): Table menu display panel


F3 (Cmd): Caibrationl(validation) command display panel
F4 (Main): Calibration/validation set display panel
F6 (Val#): Changes the validation number.
Pressing F6 displays the input prompt of “ Val number: ” at the bottom left of
the panel. Enter the number of the validation you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-57

4.10.11 Multi Range Set Display TABLE Multi Range Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Range number Range number 1 32 Range number
Measuring unit Measuring unit s "%", "ppm", or "No unit"
Measuring range Measuring range s 0.000 100.00 % % unit
0.000 9999.999 ppm ppm unit
0.000 9999.999 No unit
Ref peak number Reference stream number s 1 31
Reference peak number s 1 255
T041011.EPS

F1 (Menu): Table menu display panel


F3 (Cmd): Caibrationl(validation) command V
F6 (Range#): Changes the range number.
Pressing F6 displays the input prompt of “ Range number: ” at the bottom
left of the panel. Enter the number of the range contact you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-58

4.10.12 Lab Set Display TABLE Lab Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Stream number s 1 31
Det number s 0 2
Peak polarity s "Positive" "Negative"
Integ method s "Skimming" "Vert method"
Base line set Base line correction s "Ref invalid" "Ref" "Save"
Min peak width s 0.2 , 1.0 , 5.0 ,
30.0
Min peak height s mV 0.008 , 0.016 , 0.031 ,
0.063 , 0.125 , 0.250 ,
0.500 , 1.000 , 2.000 ,
4.000 , 8.000 , 16.00 ,
32.00 , 64.00 , 128.0
Min peak area s mVs 0.008 , 0.016 , 0.031 ,
0.063 , 0.125 , 0.250 ,
0.500 , 1.000 , 2.000 ,
4.000 , 8.000 , 16.00 ,
32.00 , 64.00 , 128.0

Repeat number s 1 65535


Analysis cycle s 0.0 28800.0 Second
Warming up time s 0.0 99999.9 Second
End of peak detect s 0 Analysis Second
cycle
Pause time s 0 Analysis Second
cycle
Init control temp s 5 High-limit temp ¡C Indicated only for
of column oven dual oven
Init temp time s 0.0 Analysis Second Indicated only for
cycle dual oven
T041012_1.EPS

F1 (Menu): Table menu display panel


F2 (Copy): Laboratory method copy display panel
F3 (Cmd): Laboratory operation display panel
F4 (Seq): Sequential display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-59
(a)When Dual oven is used

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
First temp gradient Rising speed to first temperature stage s 0.0 30.0 ¡C/sec
of column oven
First control temp First temperature stage setpoint of s 5 High-limit temperature ¡C
column oven of column oven
First temp time Duration of first temperature stage of s 0.0 Analysis cycle Second
column oven
Second temp gradient Rising speed to second temperature s 0.0 30.0 ¡C/sec
stage of column oven
Second control temp Second temperature stage setpoint of s 5 High-limit temperature ¡C
column oven of column oven
Second stage temp Duration of second temperature stage s 0.0 Analysis cycle Second
time of column oven
Third temp gradient Rising speed to Third temperature s 0.0 30.0 ¡C/sec
stage of column oven
Third control temp Third temperature stage setpoint of s 5 High-limit temperature ¡C
column oven of column oven
Third temp time Duration of Third temperature stage of s 0.0 Analysis cycle Second
column oven
T041012_2.EPS

F1 (Menu): Table menu display panel


F2 (Copy): Laboratory method copy display panel
F3 (Cmd): Laboratory operation display panel
F4 (Seq): Sequential display panel

(i) Laboratory Method Copy Display


The parameters defined for the lab setup items and the sequence setup items can be
copied by selecting the item you want to copy as described in Section 3.3, “Setup Item
Selection 1” .
F1 (Menu): Table menu display panel
F6 (Lab-Set): Laboratory set display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-60

4.10.13 Alarm Setup TABLE Alarm Set

[Calculation]

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Set number Setup number of currently 1 32
displayed/entered information
Stream number Stream number of alarm check s 1 31
Peak number Peak number of alarm check s 1 255
Check item Specifies item to be checked s Conc , R. time , Vari coeff ,
Tailing coeff
Maximum s
Minimum s
0.000 100.00 % When concentration [%] is selected
0.000 9999.999 ppmWhen concentration [ppm] is selected
-9999.999 9999.999 When concentration (no unit and
processing assign is AI signal) is selected
0.000 9999.999 When concentration (no unit and
processing assign is other than AI
signal) is selected
0.000 99999.900 Second When retention time [sec] is selected
0.000 1.999 When variation coefficient is selected
0.000 100.00 When tailing coefficient is selected
T041013.EPS

F1 (Menu): Table menu display display panel


F6 (Set#): Changes the setup number to be displayed.
Pressing F6 displays the input prompt of “ Setup No.: ” .
Enter the setup number you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-61

4.10.14 Peak Composition Set Panel TABLE Peak Composition Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Total assign Total number of peaks 0 255
assigned
Unassign Total number of peaks not 0 255
yet assigned
Stream # Stream number s 1 31
Name Stream name
Assign Number of peaks assigned s 0 255
to the stream
Peak num Number of peaks used by 0 255
the stream
T041014.EPS

F1 (Menu): Table menu display display panel


F4 (Exec): Modifies the peak table composotion.
F6 (Assign): Calls the input prompt for the stream number and the number assigned to
stream (can be set only in Stop in Measurement or Lab).
Pressing F6 displays the input prompts of “Stream No.:” and “Number
assigned:”.
Assign the peak table to the object stream. Then the display data shows
the entered values.
Note: Changing settings may affect the data of the peak individual set.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-62

4.10.15 Communication Set Panel TABLE Communication Set

• List of Displayed Data

(a)Communication Set Panel (PC)


Setup Item Description Set Minimum Maximum Unit Remarks
Trans speed Specifies the Transmission s 9600bps , 19200bps , "38400bps"
speed
T041015_1.EPS

F1 (Menu): Table menu display panel


F5 (DCS): Communication parameters setup display panel for the port DCS.

(b)Communication Set Panel (DCS)


Setup Item Description Set Minimum Maximum Unit Remarks
Protocol s GC1000/GC8", ""MODBUS", "GC6"
T041015_2.EPS

F1 (Menu): Table menu display panel


F2 (Detail): DCS communication detail display panel
F4 (PC): Communication parameters setup display panel for the port PC.

(i) Communication Detail Set Display (GC1000/GC8)


Setup Item Description Set Minimum Maximum Unit Remarks
Trans contents Specifies the transmission s Analyzed values , Alarm , All data
data
Recept contents Specifies the receiving data s None , Command , All data
or "GC8 type"
Trance speed Specifies the baud rate s 1200bps , 2400bps , 4800bps ,
9600bps , 19200bps
Procedure Specifies the type of s With procedure or Without
transmission format procedure
Parity Specifies the parity s "Even", "Odd" or "None"
T041015_3.EPS

(ii) Communication Detail Set Display (MODBUS)


Setup Item Description Set Minimum Maximum Unit Remarks
Trance speed Specifies the transmission s 1200bps , 2400bps , 4800bps ,
rate 9600bps
Com mode Specifies the communication s RTU mode" or "ASCII mode"
mode
Parity Specifies the parity s "Even", "Odd" or "None"
Scaling coef s "Real", "9999" or "65535"
T041015_4.EPS

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-63
(iii) Communication Detail Set Display (GC6)
Setup Item Description Set Minimum Maximum Unit Remarks
Trans contents Specifies the transmission s Analyzed values , Alarm , All data
data
Recept contents Specifies the receiving data s None , Command , All data
or "GC8 type"
Trance speed Specifies the baud rate s 1200bps , 2400bps , 4800bps ,
9600bps , 19200bps
Procedure Specifies the type of s With procedure or Without
transmission format procedure
Parity Specifies the parity s "Even", "Odd" or "None"
T041015_5.EPS

F1 (Menu): Table menu display panel


F2 (Format): Communication parameters setup display panel for the port DCS.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-64

4.10.16 DO Setup Panel TABLE DO Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Contact number Contact number of the DO 1 8
currently displayed/entered
Process assign Selects the processing s Stream sequence , Stream , Alarm ,
specification for the contact Ope mode or No processing
Stream Specifies the stream s 1 4 #1 , #2 , #3 , #4 , #1#2 , #1#3 ,
sequence sequence #1#4 , #1#2#3 , #1#2#4 ,
#1#2#3#4 , #2#3 , #2#4 , #2#3#4 ,
or #3#4
Stream Specifies the stream s 1 31
Ope mode Specifies the operation s Pause , Run or Stop
mode to be recognized
by the DO
Alarm level Specifies the alarm level s #1 , #2 , #1#2 or "Conc"
to be output
Method Request the sequence s 1 4 #1 , #2 , #3 or #4
number
T041016.EPS

F1 (Menu): Table menu display panel


F3 (Status): DO status display panel
F6 (DO#): Changes the contact number of the DO currently displayed.
Pressing F6 displays the input prompt of “Contact No.:” at the bottom left of
the panel. Enter the contact number of the DO you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-65

4.10.17 DI Setup Panel TABLE DI Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Contact number Contact number of the DI s 1 8
currently displayed/entered
Signal name Service of the contact s Max 8 letters
Process assign Selects the processing s Stream sequence , Stream (cont) ,
specification for the contact Stream (1cycle)", "Cal/Val",
"Ope mode change , Alarm process
or No Assignment
Stream Specifies the stream s 1 4
sequence sequence
Stream (cont) Specifies the stream s 1 31
(continue)
Stream (1cycle) Specifies the stream s 1 31
(1 cycle)
Cal/Val Specifies the method of s "Cal1", "Cal2", "Cal3", "Val1", "Val2"
calibration or "Val 3"
Operation mode Specifies the operation s Pause , Run or Stop
mode to be changed
Alarm level Output alarm level s 2 3 Displayed when Alarm processing
Output alarm number s is selected. The alarm number
displayed depends on the specified
alarm level and contact number.
Alarm level Alarm number
2 140 + Contact number
3 240 + Contact number
Message Message of alarm Max 22 letters
T041017.EPS

F1 (Menu): Table menu display panel


F3 (Status): DI status display panel
F6 (DI#): Changes the contact number of the DI currently displayed.
Pressing F6 displays the input prompt of “Contact No.:” at the bottom left of
the panel. Enter the contact number of the DI you wish to display.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-66

4.10.18 AO Setup Panel TABLE AO Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Ch number Channnel number of the AO s 1, (5) AO channel If the analog serial output
currently displayed/entered number is set to "enabled", this
setting should start on
channel 5.
#1 Output stream s (0), 1 31 (99) Entering the maximum
value (99) enables the
data of all streams to be
output.
#1 Output peak s (0), 1 255 Entering the minimum
value (0) disables the
corresponding peak to
be output.
#2 Output stream s (0), 1 31 (99) Entering the maximum
value (99) enables the
data of all streams to be
output.
#2 Output peak s (0), 1 255 Entering the minimum
value (0) disables the
corresponding peak to
be output.
#1 Partial hi s 0.001 1.000 To set the high limit for
the range of output
peak #1.
#1 Partial low s 0.000 0.999 To set the low limit for
the range of output
peak #1.
#2 Partial hi s 0.001 1.000 To set the high limit for
the range of output
peak #2.
#2 Partial low s 0.000 0.999 To set the low limit for
the range of output
peak #2.
T041018.EPS

F1 (Menu): Table menu display panel


F3 (Status): AO status display panel
F6 (AO#): Changes the contact number of the AO currently displayed.
Pressing F6 displays the input prompt of “Contact No.:” at the bottom left of
the panel. Enter the contact number of the AO you wish to display referring
to Section 3.5, “Numeric Entry”.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-67

4.10.19 AI Set Panel TABLE AI Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Ch#1 Signal name AI contact 1 signal name s Max 8 letters
Ch#2 Signal name AI contact 2 signal name s Max 8 letters
Ch#3 Signal name AI contact 3 signal name s Max 8 letters
Ch#4 Signal name AI contact 4 signal name s Max 8 letters

Ch#1 filter const Filter constant of AI contact 1 s 0.001 1.000


Ch#2 filter const Filter constant of AI contact 2 s 0.001 1.000
Ch#3 filter const Filter constant of AI contact 3 s 0.001 1.000
Ch#4 filter const Filter constant of AI contact 4 s 0.001 1.000
T041019.EPS

F1 (Menu): Table menu display panel


F3 (Status): AI status display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-68

4.10.20 ANA-BUS Set panel TABLE ANABUS Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
IP adress 240
ANABUS use Request using ANABUS s Off or On
#1 server Request Server address s 0 255 0 represents not specified.
address
#2 server Request Back up Server s 0 255 0 represents not specified.
address address
T041020.EPS

F1 (Menu): Table menu display panel


F6 (IP-Set): Set the IP adress.
Pressing F6 enables the last one octet of the IP adress to be set to the
analyzer ID.

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-69

4.10.21 GCCU Set panel TABLE GCCU Set

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Analog num s s 36 If analog serial output is set to "enabled",
add 4 to the number of outputs for setting.
Analog hold output s Enabled "Disabled"
Analog serial output s Enabled "Disabled"
Cannot be output with bit parallel
output simultaneously.
Bit parallel output s Enabled "Disabled"
Cannot be output with analog serial
output simultaneously.
Digital input s Enabled "Disabled"
Alarm contact s Abnormal close or "Normal close"
T041021.EPS

F1 (Menu): Table menu display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Toc> <Ind> <4. Display Panels > 4-70

4.10.22 Distillation Analysis Set panel TABLE Distillation Analysis Set

Only displayed

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Temperature unit s ¡C or ¡F
Det number SIMDIS s 1 2
Set time of calib curve s 0.0 99999.9 seocnd
Std B.P of calib curve s -999.9 999.9 ¡C/¡F
Calib curve coef a s -9.999 9.999
Calib curve coef b s -999.99 999.999
T041022_1.EPS

F1 (Menu): Table menu display panel


F6 (Ken): Distillation calibration curve data display panel

Calibration curve data

• List of Displayed Data


Setup Item Description Set Minimum Maximum Unit Remarks
Elapsed time 1 s 0.0 99999.9 seocnd
Std B.P 1 s -999.9 999.9 ¡C/¡F
Elapsed time 2 s 0.0 99999.9 seocnd
Std B.P 2 s -999.9 999.9 ¡C/¡F
Elapsed time 3 s 0.0 9.999 seocnd
Std B.P 3 s -999.99 999.999 ¡C/¡F
Elapsed time 4 s 0.0 9.999 seocnd
Std B.P 4 s -999.99 999.999 ¡C/¡F
Elapsed time 5 s 0.0 99999.9 seocnd
Std B.P 5 s -999.9 999.9 ¡C/¡F
Elapsed time 6 s 0.0 99999.9 seocnd
Std B.P 6 s -999.9 999.9 ¡C/¡F
Elapsed time 7 s 0.0 9.999 seocnd
Std B.P 7 s -999.99 999.999 ¡C/¡F
T041022_2.EPS

F1 (Menu): Table menu display panel


F5 (Dist): Distillation analysis set display panel

IM 11B03A03-05E 1st Edition : Oct.31,2001-00


<Int> <Toc> i

◆ Revision Record
● Manual No. : IM 11B03A03-05E
● Title : GC1000 Mark II Process Gas Chromatograph LCD Panel Operation Manual

Oct. 2001/1st Edition


Newly published

Introduction modified below;


Added “(5) When touching LCD Panel switches” to “WARNING”;
Corrected a part of “CENELEC (ATEX directive) certification”;
Corrected a part of “Installation site and environment”;
Added “Explanation for type Y purging” to “WARNING of explosionproof enclosure” of FM
and CSA certification;

IM 11B03A03-05E 2nd Edition : July. 1. 2002-00