Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
MOBILETT XP
WIRING DIAGRAMS
SPR8-230.844.01.10.02
Siemens - MED PROD & I+D SPR8-230.844.01.10.02 MOBILETT XP
C.T.S. Getafe Madrid, Spain Page 1 of 11
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
D908
L1
R3-R6
REAR ZOOM
SPR8-230.844.01.10.02
D927 K1
C2
D952
D972
U1
U2
K22
HW Overview
RIGHT SIDE ZOOM
MOBILETT XP
MOBILETT
Page 2 of 11
XP
HW OVERVIEW
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
UNIT KEYBOARD
EXPOSURE
REMOTE CONTROL RELEASE USER CONTROL
SWITCH KV, mAs, Col
H1 SINGLE TANK
X28 X15
D991
D908 DISPLAY K3 DAP
X1 CHAMBER
& mA
Pressure
Measurement
sensors
X8 X1
X20 K2
X2 X10 X11 K1
FIL X-RAY
D927 X7 TRAFO TUBE
POWER BOARD 11
X5 12
0
ANODE HT
I
X6 STATOR TRAFO
II
U1 V1
Functional Blocks
X20
D952 D972 D962
X20 CAPACITOR CAPACITOR L1
CHARGER BANK INVERTER
MOBILETT XP
Page 3 of 11
C1 PE
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
K22
1 2
3 4
7 8
X1A 5 6
1 2 4
X14
D927
X1B
Z1 4 F8 2
3 F7 10A 1
1 10A
X8 0
3 K1
4 I
1
PE A1
RC1
X16A 2 A2
1 K7 K22
X9
XRAY EXT LAMP 2 8 7
2 4 4 3
3
1
X17
1 K8 +24V X4 K2 K1
3 6 U2 1
SPR8-230.844.01.10.02
K1 Collimator
X7
1 1
1
F6 K2 X15
2
2 20A X15 K1
3 1 5 D916
3 U1 2 4
4 3
1 1 +5VDC 5 6 CPU
±15VDC 6 7
Heating 1
D907 Inverter
X5
1 X10
11 X10
3 24
24
5 K7 +V
12
D952 D972 D962
X3 X3
3 F13 F14
7
Anode 9 20A 100A
Rotation 9 F1
20A
X6 ... F12
5 7 3
II C1
3 1 1 ... C12
0
1 -V
I
X9B 1 2 X9A 1 2
C1
Connections
R1
1 2
XCAP 61 71 X52
MOBILETT XP
C2 R3 PE L1
K22
Page 4 of 11
R5
R2
R4
62 72
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
UNIT KEYBOARD
D908
+ - + -
mAs BUZZER kV mAs
kV X15
READY
COL EXP
X12
PREPARATION +V EXPOSURE
D916 REQUEST RELEASE
X27 SWITCH
SPR8-230.844.01.10.02
RESET BUZZER 3
SW3 µ CONTROLLER DATA BUS
EXPOSURE +V
REQUEST
RUN C167
+5 2
MODE RAM EPROM
SW2
SERVICE 1
+5 H1
X8 K2 p
60ºC
8 1
X
TEMP OR OIL PRESURE
PRESSURE X>Y Y LIMIT OIL TEMPERATURE
9 3
ERROR
(ERR 3)
MOBILETT XP
Page 5 of 11
CPU
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
D972 D962 H1
F14 U+
100A
R1
MEASUREMENT
CURRENT
U- 0
R+ S R-
D907
t (µs)
30 µs 30 µs
KV
MEASURE
RESONANT CURRENT AT MAIN INVERTER
SHORT EXPOSURE 70kV 5mAs
CIRCUIT
DETECTION
* WAVEFORM ONLY VALID FOR FIRST PULSE TRIGGER
+ - + - + - + -
V (VDC) FC2
SPR8-230.844.01.10.02
GND
+12
GND
+5
3,25
t (µs)
22 21 20 µs
20 19 14 13 18 17 16 15 23 24 5 3,6 4
X20
X1
SET SERVICE PROGRAM 7 AND PRESS
EXPOSURE SWITCH TO PREPARATION
* ADJUSTED BY P4 AT CPU D916
D927 X10 D916
5,7
K4
KVC
SCM
13,14 ZERO CROSS X8
+15V
KV VALUE 1
1
SET BY CPU
KVS X=Y KV 3
1/3
KV REF X Y 1
2
P4
FREQUENCY Y TEST POINT DESCRIPTION SCALE
SETTING
(50kHz) Y>X
KVN NEGATIVE BRANCH KV 1V ≡ 10kV
X KVP MAX REFERENCE
KVP 1V ≡ 10kV
Main Inverter
POSITIVE BRANCH KV
1V ≡ 30kV
Y KV
MOBILETT XP
KVM 1 TOTAL KV
Y>X X KVN MAX REFERENCE KVS KV SET VALUE 1V ≡ 30kV
Page 6 of 11
KVC
FC2
XGND
KV
KVP
KVN
GND
SCM
KVM
KVC
6A
X-RAY
X5 TUBE
1 11
450VDC
! 5 12
PE
3
1 1
P N
PE D907
SHORT X-RAY TUBE
CIRCUIT CURRENT
DETECTION MEASURE
FILAMENT
CURRENT
MEASUREMENT
12 14 13 16 1 2 2 1
X1
SPR8-230.844.01.10.02
X11
D916
JR X
MAX JR VALUE
Y X>Y 1
X8
IH X 1 1
6
1
Y X>Y
MAX IH VALUE 5
JIM SCH I
SW
4
DISABLE
HEATING RMS
1
DISABLE PWM CONTROL
GAIN ADJUSTMENT
G P5
ZERO CROSS
IHS
IH JR
FILAMENT SET VALUE
JRS I SET I REAL 1
X-RAY CURRENT SET VALUE X=Y
X Y
Filament Circuit
SWITCH X-RAY/FILAMENT CURRENT TEST
FC1 DESCRIPTION SCALE
POINT
REGULATION
P3
mAs REACHED JR X-RAY TUBE CURRENT 1V ≡ 100mA
I FILAMENT CURRENT 1V ≡ 5A
MOBILETT XP
1V ≡ 1,2A
FREQUENCY
SETTING IH FILAMENT CURRENT (RMS)
SWT SWITCH CURRENT X-RAY/FIL ACTIVE: HIGH
1V ≡ 100 mA
(80kHz) X=Y F1
JRS X-RAY CURRENT SET
1V ≡ 1,2A
Page 7 of 11
JIM
GND
JR
I
F1
JRS
XGND
IHS
FC1
mAs MEASUREMENT
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
R2
C2 7 µ F
400VAC
XCAP
1 2
START PHASE 2,5 seg BRAKE PHASE 10 seg
D952 D927
X3 X3
3 7 F5
START BRAKE
6A
H1
X6
1 I EXPOSURE TIME
V40 TF1 5 II M
450VDC
1~
TF2 3 0
!
V (VDC)
V41
1 1
3,8
CH2
6
SHORT
ACTIVE: CIRCUIT
HIGH DETECTOR 14
CH1
SPR8-230.844.01.10.02
t (ms)
+ - + - CH1: 5V / DIV I1
CH2: 1V / DIV I2
STARTING
ACTIVE: HIGH
+5
+5 +5
7 8 10 19 4 3 5 6
X11
V (VDC)
D916
CH2
ERAR
6
1,5
1 1
ENABLE ERROR CH1
H PULSES CONTROL
L PULSES CONTROL t (ms)
POWER
I2 I1 CH1: 1V / DIV I1
CIRCUIT
I MAIN CH2: 1V / DIV I2
BRAKE
I AUX
PWM GENERATION
MOBILETT XP
POINT
CURRENT AT MAIN WINDING 1V ≡ 0,25A
GND
ERAR
XGND
Starter
I1
I2
I1
I2 CURRENT AT AVX WINDING 1V ≡ 0,25A
ERAR SHORT CIRCUIT DETECTED HIGH: ERROR
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
H1: SINGLE TANK
L1: COLLIMATOR
D952 D927 H1 L1 tª
X3 X3 X7 THERMAL PROTECTION
3 7 F5 +24
1 1
6A
P1 24V DC K1
2 2 -24
450VDC LAMP: 24V 100W
!
1 1 P2
PWM CONTROLLER
ENABLING P1
4 2
K2
SPR8-230.844.01.10.02
+15
10
X10
X8
9 7
D916
+5
IR REMOTE
CONTROL
RECEIVER
20" XRMT
& 3
MAXIMUM 1≤ ACT
COLLIMATOR 14
TIME
DATA BUS
DRIVER
X12
TO DATA BUS
Collimator
COLIMATOR
USER 1
Page 9 of 11
CONTROL
DISPLAY
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
D927 POWER SUPPLY TO HEATING & COLIMATOR & STARTER
F5 V40
6A
V41
+V
D952 X3 7 1 D972
X3
3
1
X3 F1 F13 F14
B1 PIN1 TO MAIN
7
(+) INVERTER
15A 20A 100A
D962
V1 20A
! F1 ... F12
B2
180÷350VDC
(-)
C1... C12
12x10mF
9
SPR8-230.844.01.10.02
X9A
X52 DISCHARGE RESISTORS NIN1
2
DISABLE
R3 R4 R5
1
1 X -V
X9B
2 71 72
X>Y Y OVER VOLTAGE VALUE
X TEMP SENSOR 1
61 62
+15V
Y 1
-15V
K22
SHORT
BE SURE CAPACITORS ARE
DISCHARGED BEFORE HANDLE
! THESE ELEMENTS
PE
200 µs
3, 4, 6, 8, 10, 12, 21
1 11 22 3 7 9 V (VDC)
X20
440V
D916
Capacitor Charger
CPU SW VC
SW ENABLING
POWER ERROR
SUPPLY HANDLE
TEST POINT DESCRIPTION SCALE
VC CAPACITORS VOLTAGE 1V ≡ 100V
t (s)
MOBILETT XP
STARTUP 3s
Page 10 of 11
X1
D991
XDAP
L1
H1
K3
I1
DAP
I2
XDAP
2
3
6
+15V
RESET
CPU
D916
Siemens - MED PROD & I+D SPR8-230.844.01.10.02 MOBILETT XP
C.T.S. Getafe Madrid, Spain Page 11 of 11