TABmaster Model 6715

Instruction Manual

Be sure to read this manual thoroughly before using the instrument ,

fully understand and pay attention to each cautions mentioned.

Please well-keeping this manual for long time service reference.

01003
14.10
CONTENTS

Product Configuration..................................................................................................................... 3

Caution........................................................................................................................................... 5

1. Introduction ................................................................................................................................ 7

1.1. Features............................................................................................................................. 7

1.2. Specifications ..................................................................................................................... 7

2. Outlook & Structure .................................................................................................................... 8

2.1. Base structure .................................................................................................................... 8

2.2. Micromanometer ............................................................................................................... 9

3. Settubg up................................................................................................................................... 9

3.1. Capture Hood .................................................................................................................... 9

3.2. Velocity Grid.....................................................................................................................10

3.3. Pitot tube .......................................................................................................................... 11

3.4. Micromanometer .............................................................................................................. 11

3.5. Shoulder Strap.................................................................................................................. 11

4. Operating Instruction .................................................................................................................12

4.1. Power supply ....................................................................................................................12

4.2. Start On / Power Off .........................................................................................................12

4.3. Keypad operation..............................................................................................................13

5. Function test...............................................................................................................................14

5.1. User Interface ...................................................................................................................14

5.2. Single Mode ......................................................................................................................15

5.3. Average Mode ...................................................................................................................15

5.4. Back Pressure Mode ..........................................................................................................15

6. Menu setting...............................................................................................................................16

6.1. General setting..................................................................................................................17

6.1.1. Date ........................................................................................................................17

6.1.2. Time .......................................................................................................................17

6.1.3. Auto Off..................................................................................................................17

6.1.4. Backlighting............................................................................................................18

6.1.5. Communication.......................................................................................................18

6.2. Test setting........................................................................................................................19

1
6.2.1. Tools .......................................................................................................................19

6.2.2. ID application .........................................................................................................20

6.2.3. Test Mode ...............................................................................................................20

6.2.5. Data record .............................................................................................................21

6.2.6. Standard/Actual test ................................................................................................21

6.2.7. Compensation factor ...............................................................................................22

7 Error and Troubleshooting...........................................................................................................23

8 Warranty and Service ..................................................................................................................24

8.1. Product Warranty.............................................................................................................24

8.2. After service......................................................................................................................24

2
Product Configuration.
6715 Standard:

Item. Qt’y
Micromanometer Model 6700 1
Instrument Base 1
Fabric Hood: 2ft×2ft 1
Carrying case 1
Portable handle 1
Frames 1
Poles 4
Communication cable 1
Calibration certificate 1
CD-Rom(Manual and software inside) 1
Quick-Guide 1
6715 Optional Parts:

Item. Specifications
Spare hood 2x2ft
Spare hood 2x4ft
Spare hood 1x4ft
Spare hood 3x2ft
Spare hood 3x3ft
Spare hood 500x500mm
Poles
Capture Hood Stand Up to 3.4m with capture hood
Communication cable
Printer
AC adapter DC5V
International plug adapter
Carrying case
Velocity Grid
Pitot tube (2.3 x 200mm)
Pitot tube (4 x 300mm)
Pitot tube (8 x 500mm)
Pressure tubing 2*250cm

3
6700 Standard:
Item. Qt’y
Micromanometer Model 6700 1
Carrying case 1
Pressure tubing 2
Communication cable 1
Calibration certificate 1
CD-Rom(Manual and software inside) 1
Quick-Guide 1
6700 Optional Parts:

Item. Specifications
Velocity Grid
Pitot tube (2.3 x 200mm)
Pitot tube (4 x 300mm)
Pitot tube (8 x 500mm)
Pressure tubing 2*250cm
Shoulder Strap
Printer
AC adapter DC5V
International plug adapter
Communication cable

4
Caution.
The symbols for warning mentioned in this manual are defined below:

〔symbols classifications〕
Danger: To Prevent Serious Injury or Death
Wa rnings in this classifi cation indi cate a danger tha t ma y resul t in serious injury or dea th if not observed .

Caution: To Prevent Damage to the Product

Wa rnings in this classifica tion indi ca te a risk of damage to the product tha t ma y void the product wa rranty if not

observed.

〔 Description of Symbols〕
△ This s ymbol indi ca tes a condi tion tha t requi res caution (including danger). The subject of ea ch cauti on is
illustra ted inside the triangle .

This s ymbol indica tes a prohibi tion. Do not take the prohibi ted a ction shown inside or nea r this s ymbol .

● This symbol indicates a mandatory action. A specific action is given near the symbol.

WARNING
○ Never bring the fabri c hood to a flammable gas
a tmosphere.

Heated forbidden … otherwise, the hea t ma y cause a fi re or explosion.

○ Do not disassembl y or refi t the instrument.

Disassemble … otherwise, ma y cause the electri c shock or a fi re.


prohibition

○ Using properl y under the instruction manual.

… otherwise, ma y cause sensor da maged or an electri c shock even a fire.


Using properly

○ If abnormal smells, noises or s moke occur, or if liquid enters the ins trument, pull out
the AC adapter and remove the batteries immedia tel y. Then send i t to the
maintena nce Dept. of KANOMAX for after servi ce.
Using properly … or, there is possible of an electri c shock or a fi re or instrument malfunction.
○ Do not expose the fabri c hood, base and the ins trument
noumenon in the rain.
… otherwise, ma y cause an electri c shock, an fire and an
Forbidden person injure.

5
CAUTION
○ Alwa ys unplug when the instrument not in use.

… Failure to do so ma y cause an electri c shock, an fire or ci rcui t da mage.


Using properly

○ Remove the ba tteries when s tori ng the instrument for a l ong period. Do not lea ve
the exhaused batteries in the ba ttery compa rement and exchange the ba ttereis on
time.
… Failure to do so ma y cause the batteries leakage and the ins trument dama ge.
Using properly

○ Ma tching the correct AC adapter outwa rd as reqes ted.

… or ma y ca use the ins trument dama ge.


Using properly
○ Do not wipe the ins trument wi th a vola tile sol vent.
… the body ma y deform or deteri ora te. Use a soft dry cl oth to remove stains . If stains
persist, soak the cloth in a neutral detergent a nd wipe the i nstrument with the soft
Forbidden cloth. Never use vola tile sol vents such as thinner or benzene .
○ Do not use or lea ve the ins trument in a high
temperature, high humidi ty or dus ty envi ronment. Do
not lea ve the instrument under di rect sunlight.
… otherwise, the instrument ma y not function properl y
Forbidden out of the speci fied opera ting condi tions or the inside
components damaged.

○ Never dropping the uni t or pla cing hea vy objects on i t

… It ma y cause damage or malfunction to the ins trument


Forbidden

6
1. Introduction
TABmaster Model 6715 is a kind of intelligent test instrument with multi functional of airflow test, velocity

test and micro-Differential pressure test. It’s widely used in air flow and velocity testing of air - conditioning,

HAVC system and other places, especially for the high precision micro differential pressure test.

1.1. Features
 3.5 inch true color LCD.

 Available measuring the air flow, velocity, temperature and humidity at the same time.

 Two kinds of tests are available of in flowing and exhaust flowing.

 Back pressure compensation.

 Ultra-large Storage capacity: 8000 records

 Blue tooth communication makes remote monitoring and data transferring available.

 Light weight, compact configuration for easy carrying.

 Data printing or transferring to the computer.

1.2. Specifications
Items Specifications
Test range 40~4300 m3 /h
Air flow Accuracy ±3% of readings ± 8m3 /h(>85m3 /h)
3
Resolution 1 m /h
Test range 0.15~40 m/s(Pitot Tube)、0.15~15 m/s(Velocity Grid)
Air velocity Accuracy ±3% of readings ±0.05m/s(>0.25m/s)
Resolution 0.01m/s
Test range -2500~2500 Pa
Differential
Accuracy ±0.25% of readings ±1Pa
pressure
Resolution 0.001 Pa
Test range 0~60℃
Temperature Accuracy ±0.5℃
Resolution 0.1℃
Test range 0~100%RH
Humidity Accuracy ±3%RH(10~90%RH)
Resolution 0.1%RH
Test range 70~130 kPa
Absolute
Accuracy ±2% of readings
pressure
Resolution 0.1 kPa
Operation Temp. 0~60℃(non condensing )
Storage Temp. -20~70℃(non condensing)
Power Source 4 pcs of AA-size buttery ( approx. 9hrs available)or DC5V Adapter
Weight 3.6kg approx.

7
2. Outlook & Structure
Frame

Poles
Pi vot
Wi ndow

Fabri c
Hood

Ins trument Mi cromanometer


Base

2.1. Base structure


1.Outlook

Potable handle

Mi cromanometer
Ba ck pressure
ON/OFF

Adjus t handle Base buttom

Test handle

2.Internal view 3.Grid

Ba ck pressure plate
Testing

Poles point

Grid

8
2.2. Micromanometer

Keypad Displa y LCD Cover lock Ba ttery Compa rtment Cover

Front view Ba ck

Power on/off Ai r port (+) Ai r port (-) Mini USB Port

Top view Bottom view

3. Settubg up

3.1. Capture Hood


1.Fabric Hood & basement

(1) (2) (3) (4)

Note:

1) Make sure the elastic band of the fabric hood is placed exactly as shown (Figure(3)) to ensure a good fit.

The elastic band must lineup with the top lip of the base as shown. If you don’t line it up correctly, the

hood might slide off the base when you insert the poles.

2) Use the 4 positioning holes on the base stitching as a guide to ensure the hood fully wraps around the

base..

2.Frame Poles

9
Note: The poles always cross in an “X” shape when assembling. To remove the poles, simply reverse the steps.

3.Potable handle

① ②

Note: Release the poles and portable handle when packing in the carring case to avoid case damaged.

4.Micromanometer & basement

(+) (-)

Note: 1)Correct install the micromanometer to the base with the same S/N No.

2)Do not cross- insert the Air pines just refer to the figure as above.

3.2. Velocity Grid


Air velocity testing can be achieved when micromanometer work together with the velocity Grid.

10
Grid

Ha ndle
(+)(-)

Note: Mounting angle between the Grid and handle can be adjusted as request.

3.3. Pitot tube


Air flow and velocity testing can be achieved when micromanometer work together with Pitot tube.

Sta ti c pressure
port

Testing point

(+)(-)

3.4. Micromanometer
Differential pressure testing can be achieved by using the micromanometer.

Device
under test

(+)(-)

3.5. Shoulder Strap


When testing by Velocity Grid、Pitot tube、Micromanometer as the tool, for convenient view and easy
operation, the shoulder strap can be fit for using.

11
4. Operating Instruction

4.1. Power supply


1.Power by AC adapter. AC adapter will be as the priority power supply when the AC adapter and batteries are
all available using. The specification of the AC adapter is: I/P:AC 110-240V 50/60HZ O/P:DC 5V/2A。

1)AC adapter for the master 2)AC adapter for Capture Hood

2.Power by battery

 4 AA-size batteries can be used for a power supply.


 Press the compartment cover lock (refer to figure indicating ①/②), and then slide the cover off to open
the battery compartment。
 Put 4 AA-size batteries in and make certain the batteries are correctly oriented.
 New alkaline battery or rechargeable Ni-MH battery can be used. Never mix battery types, or battery
leakage or damage to the instrument may occur.
 Replace the compartment cover (refer to figure ③). The compartment cover will be automatically locked.

② ③

 When power is supplied with batteries, the current charge value will be displayed on the upper of the LCD.
When the batteries run out of power, the instrument will turn off.

Power off

Note:
1) Do not test with low battery power.
2) Do not mixing use full power battery with low power ones for ensure testing well.
3) Change the batteries in time when power shows low for ensure testing well.

4.2. Start On / Power Off

12
POWER 2014 / 03 / 20 13 : 10
ID:
01
Ready
m3/h Cycle:
Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode:
START SAVE MENU Single

1. Start On
Press button【POWER】for 2 seconds for start On, displaying “ KANOMAX” then enter into the testing
interface.
2. Power Off
Press【POWER】for 2 seconds, shut down and power off.

4.3. Keypad operation


Keypad panel includes control buttons, back-pressure switch and LCD adjusting keys.
1. Control buttons:

UP/SAVE

SET/MENU/OK

START/STOP/
BACK/EXIT/ESC
DOWN START/STOP SAVE

UP:Value increasing or Cursors upward moving.


SAVE:Save the current test results
SET:Store changes or enter into the selected option’s next level interface
MENU:On the main screen holding this down for 2 seconds will enter into the main “MENU”
OK:In the “General Setting ”, pressing for 2 seconds will save the date or time ;
While in the “Record Processing”, pressing for 2 seconds will delete saved data.
START:Begin a measurement in main menu
STOP:Stop a measurement in main menu
BACK:Cancel or end an operation or return to the previous screen
EXIT:Return to the main menu
ESC:Stop printing when printing
DOWN::Value decreasing or Cursors downward moving

2. Back pressure switch for on/off


Note :
1) If you are not in the back pressure mode, shut down the back pressure plate and lock the
switch to avoid testing errors.
2) If the airflow is greater than 1500m3 /h (883CFM), testing in back-pressure Mode is
recommended according to testing procedures.

13
Ba ck pressure
On

Ba ck pressure
Off

3.LCD adjusting

Note :
1)Please switching the handle slow gently during LCD adjusting.
2)Stop turning when reach to the bound, for avoiding over turning and handle falling down.

5. Function test
Besides the airflow test, if along with other optional products, test air velocity and micro -differential
pressure is available.

5.1. User Interface


Testing information, parameter settings and the testing conditions will be displayed on the main interface.

J
2014 / 03 / 20 13 : 10
A I
ID:
01 H
B Ready 3
m /h Cycle:
C
Atmos: 99.0 kPa 001 G
D 20.5 ℃ 30.7 %RH
Mode:
START SAVE MENU Single F
E

A. Date and time —— real-time display the current date and time
B. Testing data display —— airflow 、air velocity 、differential pressure and etc.
C. Tools icon —— working tools display
D. Testing condition —— temperature 、humidity or absolute pressure displaying according to the

14
working tools.
E. Keypad —— current status indicating and the grey state is not available.
F. Testing mode —— display the current testing mode
G. Cycle—— recorded Cycle quantity
H. ID—— current ID quantity
I. Power supply —— AC adapter or batteries
J. Blue tooth status —— when blue tooth working, the icon display or no icon display with closed.

5.2. Single Mode

2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10


ID: ID: ID:

Ready
m3/h
01
Waiting...
m3/h
01
32 m3/h
01

Cycle: Cycle: Cycle:


Atmos: 99.0 kPa 001S TART Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
START SAVE MENU Single STOP SAVE MENU Single START SAVE MENU Single

1. Ready for testing 2. Testing 3. tes t output


待测状态
Note:
1) START: Left-button for the Micromanometer or instrument base
2) After testing start for about 8s with stably measuring,output displaying the test data and testing wind
direction.
3) Testing stop when finish outputting.
4) Press SAVE for data record with adding 1 to the Cycle value.
5) Same testing method with the capture hood when using the Velocity Grid、Pitot、Micromanometer.

5.3. Average Mode


2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10
ID: ID: ID: ID:

Ready
m3/h
01
Waiting...
m3/h
01
32 m3/h
01 31 m3/h
01

Cycle: Cycle: Cycle: Cycle:


S TART Atmos: STOP
Atmos: 99.0 kPa 001 99.0 kPa 001 Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode: 20.5 ℃ 30.7 %RH Mode: 20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
START SAVE MENU Avg STOP SAVE MENU Avg STOP SAVE MENU Avg STOP SAVE MENU Avg

1.Ready for tes ting 2. Testing 3. test output 4. testing stop

Note:
1) START: Left-button for the Micromanometer or instrument base.
2) After testing start for about 8s with stably measuring , output displaying the test data for the first time.
3) Go on the testing with frequency of 1s for data updating.
4) The continuously outputting is as the average data and be related with the “AvgTime” set by user. More
larger of the value of“AvgTime” will get more stable of the testing data.
5) If auto-save function On, testing data will be recorded automatically according to the setting “AutoSave”
and value of “Cycle” will be added automatically.
6) Press “STOP”, testing stop. And “STOP” will be as the Left-button for the Micromanometer or instrument
base.
7) Same testing method with the capture hood when using the Velocity Grid、Pitot、Micromanometer.

5.4. Back Pressure Mode

15
In order to reduce the measure error of the pressure loss,which causing from the hood to the system and
get the more real air flow, we suggest testing under the Back Pressure Mode when air flow more than 1500
m3/h.

2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10


ID: ID: ID:
01 01 01
Ready Waiting... Open Flap !
m3/h m3/h m3/h Cycle:
Cycle: Cycle:
S TART
Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
START SAVE MENU B.P STOP SAVE MENU B.P START SAVE MENU B.P

1.Ready for tes ting 2. Testing 3. Ba ck-pressure On promoting

2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10
ID: ID:
01 01
Waiting... Ready
m3/h Cycle: m3/h Cycle:
S TART
Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode: 20.5 ℃ 30.7 %RH Mode:
STOP SAVE MENU B.P START SAVE MENU B.P
6. testing 5. testing ready 4. s wi tch on Back-Pressure Pla te

2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10
ID: ID:

2000 m3/h Cycle:


01 Ready
m3/h Cycle:
01

Atmos: 99.0 kPa 001


Atmos: 99.0 kPa 001

20.5 ℃ 30.7 %RH 20.5 ℃ 30.7 %RH Mode:


Mode:
START SAVE MENU B.P
START SAVE MENU B.P

7. test output 8. shut off the pla te 9. testing ready

Note:
1) START: Left-button for the Micromanometer or instrument base.
2) After testing start for about 8s , indicating “Open Flap”. It’s time for switching on the Back-Pressure Plate
as needed.
3) When Back-Pressure Plate opened On, tip of “Open Flap” disappear.
4) Press START for about 8s, testing stop, real airflow outputting with back pressure compensation.
5) Press SAVE for testing data record and the meanwhile, value Cycle added 1.
6) When select Velocity Grid、Pitot tube、Micromanometer for testing, Back-Pressure compensation will be
not available.

6. Menu setting
when stop for ready, press for 2s for enter into the menu.

2014 / 03 / 20 13 : 10 MENU
ID:
UP (1)2 opti onal settings :
01
Ready 2s General Setting
m3/h
DOWN 1)Basi c setting
Cycle:
Atmos: 99.0 kPa 001
SET
2)Test setting
20.5 ℃ 30.7 %RH
Mode:
Test Setting OK
3)Working tool displayi ng below
START SAVE MENU Avg EXIT

16
6.1. General setting

MENU General Setting (1)Basic setting:


UP 1. Date 2014 / 03 / 20 UP 1)Date
General Setting 2. Time 13 : 10
DOWN
3. Auto Shut Off DOWN 2)Time
4. Backlight High
SET
5. Communicate
SET 3)Auto Shut
Test Setting OK OK 4)Backlight
EXIT BACK 5)Communicate

6.1.1. Date

General Setting General Setting

1. Date 2014 / 03 / 20 UP 1. Date 2014 / 04 / 01 UP


(1) press or for data modify.
2. Time 13 : 10 2. Time 13 : 10
Auto Shut DOWN DOWN
3. Off 3. Auto Shut
4. Backlight
Off (2)press , for sele cting Year/Month/day.
High SET 4. Backlight High SET
5. Communicate 5. Communicate
OK OK
(3)Press for 2s for save settings.

BACK BACK (4)Press for quit.

2s
General Setting General Setting

1. Date 20 14 / 03 / 20 UP 1. Date 2014 / 04 / 01 UP

2. Time 13 : 10 2. Time 13 : 10
DOWN Auto Shut DOWN
3. Auto Shut Off 3. Off
4. Backlight High 4. Backlight High SET
SET
5. Communicate 5. Communicate
OK OK

BACK BACK

6.1.2. Time

General Setting General Setting

2014 / 03 / 20 UP 1. Data 2014 / 03 / 20 UP


(1)Press or for data modify.
1. Date
2. Time 13 : 10 2. Time 15 : 15
3. Auto Shut DOWN
3. Auto Shut Off
DOWN (2)Press for hour/minute selection.
Off
4. Backlight High 4. Backlight High SET
SET
5. Communicate 5. Communicate (3)Press for 2s for save settings.
OK OK
(4)Press for quit.
BACK BACK

2s
General Setting General Setting (1)按 或 键进行数值更改。
UP
1. Data 2014 / 03 / 20 UP 1. Date 2014 / 03 / 20 (2)按 键,时/分选项循环移动。
2. Time 13 : 10 2. Time 15 : 15
3. Auto Shut Off
DOWN
3. Auto Shut Off 2s DOWN
(3)长按 键 2s,保存设置。
4. Backlight High SET
4. Backlight High SET
5. Communicate 5. Communicate (4)按 键,退出设置。
OK OK

BACK BACK

6.1.3. Auto Off

When instrument not in use, it will be turn off automatically according to Auto-Off time setting to avoid
power waste.

17
General Setting General Setting

1. Date 2014 / 03 / 20 UP 1. Data 2014 / 03 / 20 UP

2. Time 13 : 10 2. Time 13 : 10 (1)Press or for options cha nge.


DOWN DOWN
3. Auto Shut Off 3. Auto Shut Off
4. Backlight High SET
4. Backlight 10 min SET
(2)Press for save settings .
5. Communicate 5. Communicate 20 min
OK 30 min OK (3)Press for quit.
60 min
BACK BACK
(3)按 键,退出设置。

General Setting General Setting

1. Data 2014 / 03 / 20 UP 1. Date 2014 / 03 / 20 UP

2. Time 13 : 10 2. Time 13 : 10
DOWN DOWN
3. Auto Shut Off 3. Auto Shut 10 min
4. Backlight 10 min 4. Backlight High
SET SET
5. Communicate 20 min 5. Communicate
30 min OK OK
60 min
BACK BACK

6.1.4. Backlighting

Adjust the LCD lighting according to the environmental brightness for the best viewing. And meanwhile, the
lower the brightness, the less battery power will be needed.

General Setting General Setting

1. Date 2014 / 03 / 20 UP 1. Date 2014 / 03 / 20 UP

2. Time 2. Time
(1)Press or for options cha nge.
13 : 10 13 : 10
DOWN DOWN
3. Auto Shut Off 3. Auto Shut Off
4. Backlight High 4. Backlight Low
(2)Press for save settings .
SET SET
5. Communicate 5. Communicate Normal
OK High OK
(3)Press for quit.
BACK BACK

General Setting General Setting

1. Date 2014 / 03 / 20 UP 1. Date 2014 / 03 / 20 UP

2. Time 13 : 10 2. Time 13 : 10
Off DOWN DOWN
3. Auto Shut 3. Auto Shut Off
4. Backlight Low SET
4. Backlight Low SET
5. Communicate Normal 5. Communicate
High OK OK

BACK BACK

6.1.5. Communication

Communicate with PC, printer and Bluetooth are available. Data transfer or remote monitoring is available
by selecting communicating way.
Note:

1)PC communication means to save the testing record to the computer. And the communication cable and
software reserved for data-fetching are necessary for using together. Refer to the << Manuel for Data
fetching Software>> .
2)For printing out the testing record, the optional printer and cables for special are necessary. Printer baud
rate is set to 19200 bps and Do not change it. Refer to the << Manuel for RD series Mini printer >> .
3)Bluetooth make the remote control and data monitoring available by Hand phone. And our APP software is
necessary for using together. Refer to the << Manuel for TAB Mobile>>.
4)Shut off the switch when not use the Bluetooth communication for improving battery life.

18
General Setting

1. Date 2014 / 03 / 20 UP

2. Time 13 : 10
DOWN
3. Auto Shut Off
4. Backlight High SET
5. Communicate
OK

BACK

Communicate Computer

1. Computer UP UP

2. Printer Comfirm the Connection


Off DOWN DOWN
3. Bluetooth
SET Waiting... SET

OK OK

BACK BACK

Communicate Print

1. Computer UP 1. Print By IDs UP

2. Printer 2. Print By Cycles


DOWN DOWN
3. Bluetooth Off
SET SET

OK OK

BACK BACK

Communicate Communicate

1. Computer UP 1. Computer UP

2. Printer 2. Printer
Off DOWN On DOWN
3. Bluetooth 3. Bluetooth
SET SET

OK OK

BACK BACK

5)If other settings need to be changed when Bluetooth successful communication, you need to shutdown
Bluetooth firstly then enter the menu to change other settings. Just as below shown:

2014 / 03 / 20 13 : 10 2014 / 03 / 20 13 : 10
ID: ID:

32 m3/h
01
32 m3/h
01

Cycle: 2s Cycle:
Atmos: 99.0 kPa 001 Atmos: 99.0 kPa 001
20.5 ℃ 30.7 %RH Mode:
20.5 ℃ 30.7 %RH Mode:
START SAVE OFF Avg START SAVE MENU Avg

6.2. Test setting

MENU Test Setting Test settings:


UP
UP 1. Select Tool 1)Select Tool
2. Test ID 2)Test ID
General Setting DOWN DOWN
3. Test Mode RunAvg
4. Unit 3)Test Mode
SET
SET
5. Data Record 4)Unit
OK 6. Std/Act Act OK 5)Data Record
Test Setting
7. K factor 1.000 6)Standard/Actual testing
EXIT BACK
7)Compensation factor

6.2.1. Tools

19
The TABmaster model 6715 is an instrument for airflow, air temperature, and relative humidity
measurements. The optional accessories (Velocity Grid, Pitot tube, and Micromanometer) expand the
parameters that the instrument can measure. Refer to the Chapter 3 and 5 for details.

Test Setting Select Tool

1. Select Tool UP 1. Capture Hood UP

2. Test ID (1)Press or for options cha nge.


DOWN DOWN
3. Test Mode RunAvg 2. Velocity Grid
4. Unit (2)Press for save settings .
SET SET
5. Data Record 3. Pitot
6. Std/Act Act OK OK (3)Press for quit.
7. K factor 1.000 4. Pressure P
BACK BACK

Select Tool Test Setting

1. Capture Hood UP 1. Select Tool UP Tools Test item


2. Test ID TABmaster Air flow / Temperature / RH
DOWN DOWN
2. Velocity Grid 3. Test Mode RunAvg Velocity Grid Air velocity
4. Unit SET
SET
3. Pitot 5. Data Record Pitot tube velocity
OK 6. Std/Act Act OK Micromanomete r Micro-differe ntial pressure
4. Pressure P 7. K factor 1.000
BACK BACK

6.2.2. ID application

Test Setting Test ID

1. Select Tool UP 1. Select ID 01 UP (1) Test ID is up to 20.


2. Test ID 2. Add New ID No
3. Test Mode RunAvg
DOWN DOWN (2) Set different Test ID for the di fferent tool .
4. Unit SET SET
5. Data Record (3) 100 data records can be stored in a single
6. Std/Act Act OK
Total ID: 20 OK
7. K factor 1.000
ID
BACK BACK

6.2.3. Test Mode

Test Setting Test Setting


(1)”Avg Ti me” is as a peri od for
UP
1. Select Tool UP
1. Select Tool
2. Test ID
2. Test ID testi ng; the longer the ti me peri od,
DOWN 3. Test Mode RunAvg DOWN
3. Test Mode Single
4. Unit SET
Avg Time: 03 Sec
SET
the more s table the testing value will
5. Data Record Auto Save: Off
6. Std/Act Act OK OK be as spikes in ai rflow will be avera ged
7. K factor 1.000 BACK
BACK
out over time.

Test Setting Test Setting


(2)“Auto Sa ve” is a time interval for
1. Select Tool UP
1. Select Tool UP
2. Test ID 2. Test ID automa tic da ta recording during testing.
3. Test Mode Single DOWN 3. Test Mode RunAvg DOWN
4. Unit RunAvg Avg Time: 03 Sec
SET SET
5. Data Record B.Prees Auto Save: Off
6. Std/Act Act OK
10 Sec OK
7. K factor 1.000 20 Sec
30 Sec BACK
BACK 60 Sec

Test Setting Test Setting


1. Select Tool UP 1. Select Tool UP
2. Test ID 2. Test ID
3. Test Mode Single DOWN DOWN
3. Test Mode RunAvg
4. Unit RunAvg 4. Unit
5. Data Record SET SET
B.Prees 5. Data Record
6. Std/Act Act OK OK
6. Std/Act Act
7. K factor 1.000
7. K factor 1.000
BACK BACK

20
6.2.4. Units

Test Setting Unit

1. Select Tool UP 1. Airflow m3/h UP

2. Test ID CFM
(1)Press or for options cha nge.
2. Atoms
DOWN DOWN
3. Test Mode RunAvg 3. Temperature l/s
4. Unit lbs (2)Press for save settings .
SET SET
5. Data Record
6. Std/Act Act OK OK (3)Press for quit.
7. K factor 1.000
BACK BACK

Unit Unit

1. Airflow CFM UP 1. Airflow m3/h UP

2. Atoms kPa 2. Atoms kPa


DOWN DOWN
3. Temperature ℃ 3. Temperature ℃
SET SET

OK OK

BACK BACK

6.2.5. Data record

Test Setting

1. Select Tool UP

2. Test ID
DOWN
3. Test Mode RunAvg
4. Unit SET
5. Data Record
6. Std/Act Act OK
7. K factor 1.000
BACK

Data Record Test ID: 01 Total: 20 Cycle: 01 Total: 01

UP UP AirFlow: 0 m3/h UP
1. Display the Record
Max: 0 m3/h Atmos : 101.3 kPa
2. Delete the Record
DOWN DOWN DOWN
Min : 0 m3/h Temp: 21.1 ℃
SET SET Humidity: 35.8 %RH SET
Avg: 0 m3/h Std/Act: Std
OK OK Time: OK
2014 / 03 / 18
Press [SET] See Cycle Data BACK
13 : 10 : 18 BACK
BACK

Data Record Delete


1. Display the Record UP 1. Delete All No UP
2. Delete the Record 2. Delete IDs
DOWN DOWN
Select ID: 01 No
SET
3. Delete Cycles SET
Select ID: 01
OK Select Cycle: 001 No OK

BACK BACK

Note:
1) Browsing or deleting test data in current testing tools is available for allowed.
2) Data exporting by communication function and deleting the non-valid data in time for the effective
storage space.
3) Pressing for 2 seconds will delete saved data.

6.2.6. Standard/Actual test

Test Setting Test Setting Test Setting

1. Select Tool UP 1. Select Tool UP 1. Select Tool UP

2. Test ID 2. Test ID 2. Test ID


DOWN DOWN DOWN
3. Test Mode RunAvg 3. Test Mode RunAvg 3. Test Mode RunAvg
4. Unit 4. Unit SET
4. Unit
SET SET
5. Data Record 5. Data Record 5. Data Record
6. Std/Act Act OK 6. Std/Act Act OK 6. Std/Act Std OK

7. K factor 1.000 7. K factor Std 7. K factor 1.000


BACK BACK BACK

21
Note:
1) Standard Test (Std): in the condition of atmosphere 101.3kPa and temperature 21.1℃.
2) Actual Test(Act): in actual condition with the compensation for the atmosphere and temperature.
3) No Std/Act selection when in Micromanometer testing.
4) When doing actual testing under Velocity Matrix and pitot tube, manually input the current
temperature is needed for the compensation. Refer to the figures as below:

Test Setting Test Setting Test Setting

UP 1. Select Tool UP 1. Select Tool UP


1. Select Tool
2. Test ID 2. Test ID 2. Test ID
DOWN DOWN DOWN
3. Test Mode RunAvg 3. Test Mode RunAvg 3. Test Mode RunAvg
4. Unit 4. Unit 4. Unit SET
SET SET
5. Data Record 5. Data Record 5. Data Record
6. Std/Act Act OK 6. Std/Act Act OK 6. Std/Act Act OK

7. K factor 1.000 7. K factor 1.000 7. K factor 1.000


BACK BACK BACK
8. Temperature 23.5 ℃ 8. Temperature 25.0 ℃ 8. Temperature 25.0 ℃

6.2.7. Compensation factor

Test Setting Test Setting

1. Select Tool UP 1. Select Tool UP


(1)Press or for options cha nge.
2. Test ID 2. Test ID
DOWN DOWN
3. Test Mode RunAvg 3. Test Mode RunAvg
4. Unit 4. Unit (2)Press for save settings .
SET SET
5. Data Record 5. Data Record
6. Std/Act Act OK 6. Std/Act Act OK (3)Press for quit.
7. K factor 1.000 7. K factor 1.010
BACK BACK

Test Setting Test Setting

1. Select Tool UP 1. Select Tool UP

2. Test ID 2. Test ID
DOWN DOWN
3. Test Mode RunAvg 3. Test Mode RunAvg
4. Unit 4. Unit SET
SET
5. Data Record 5. Data Record
6. Std/Act Act OK 6. Std/Act Act OK

7. K factor 1.000 7. K factor 1.010


BACK BACK

Note:
1) Under Capture Hood and Velocity Grid testing, K factor is used for testing data compensation in
order to get more accuracy result. And the testing range will be 0.500~1.500.
Note: K factor will affect the accuracy of measurement, so, it’s necessary for having the test bench for
standard reference and using with caution.
For example:
If the Capture Hood is as the selected testing tool, testing value by Model 6715 is 1000m3 /h,
the standard airflow is 1020m3 /h,at this point, K factor= standard value/testing value =1.020,input
this K factor to correct the accuracy of the measurement by Model 6715.
2) Under Pitot tube testing, K factor is the factory standard coefficient and each pitot with its own
independent factor.
3) No K factor setting under the micromanometer testing.

22
7 Error and Troubleshooting
Symptom Possible causes Corrective action
incorrect specification of the AC Check and refer to Chapter4.1
Adapter in Manuel
No display when power On Check and refer to Chapter4.1
Battery installation error
in Manuel
Low or dead battery power Replace the batteries
Connection cable loss with the
Display “Connect Error” Reseat the connect cable
base
Back-Pressure mode using
Display “DATA Error” Refer to Chapter 5.4 in Manuel
incorrect
Low battery charge Replace new batteries
Press “START” but no response
Back Pressure Plate opened On Shut off the plate
Over-limit Cycle quantity in Add new ID or delete some
Testing record can not be saved
current ID data
ID over-limit Delete some ID
New ID can not be added
Current ID is without “Cycle” Using the current ID
Using our special optional
Unmatched printer
printer supplied by us.
Printing unreadable code
Reset referring to the Printer
Wrong baud rate setting
Manuel

23
8 Warranty and Service

8.1. Product Warranty


The limited warranty set forth below is given by KANOMAX GROUP COMPANIES with respect to the
KANOMAX brand capture hood and other accessories (hereafter referred to as “PRODUCT”) purchased
directly from KANOMAX GROUP COMPANIES or from an authorized distributor.
Your PRODUCT, when delivered to you in new condition in its original container, is warranted against defects
in materials or workmanship as follows: for a period of one (1) year from the date of original purchase, defective
parts or a defective PRODUCT returned to KANOMAX GROUP COMPANIES, as applicable, and proven to be
defective upon inspection, will be exchanged for a new or comparable rebuilt parts, or a refurbished PRODUCT as
determined by KANOMAX GROUP COMPANIES. Warranty for such replacements s hall not extend the original
warranty period of the defective PRODUCT.

This limited warranty covers all defects encountered in normal use of the PRODUCT, and does not
apply in the following cases:
(1) Use of parts or supplies other than the PRODUCT sold by KANOMAX GROUP COMPANIES, which cause
damage to the PRODUCT or cause abnormally frequent service calls or service problems.
(2) If any PRODUCT has its serial number or date altered or removed.
(3) Loss of damage to the PRODUCT due to abuse, mishandling, alternation, improper packaging by the
owner, accident, natural disaster, electrical current fluctuations, failure to follow operation,
maintenance or environmental instructions prescribed in the PRODUCT's operation manual provided by
KANOMAX GROUP COMPANIES, or service performed by other than KANOMAX GROUP COMPANIES.
NO IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE, APPLIES TO THE PRODUCT AFTER THE APPLICABLE PERIOD OF THE EXPRESS LIMITED
WARRANTY STATED ABOVE, AND NO OTHER EXPRESS WARRANTY OR GUARANTY, EXCEPT AS MENTIONED
ABOVE, GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO THE PRODUCT SHALL BIND KANOMAX GROUP
COMPANIES. KANOMAX GROUP COMPANIES SHALL NOT BE LIABLE FOR LOSS OF STORAGE CHARGES, LOSS
OR CORRUPTION OF DATA, OR ANY OTHER SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY
THE USE OR MISUSE OF, OR INABILITY TO USE, THE PRODUCT, REGARDLESS OF THE LEGAL THEORY ON
WHICH THE CLAIM IS BASED, AND EVEN IF KANOMAX GROUP COMPANIES HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. IN NO EVENT SHALL RECOVERY OF ANY KIND AGAINST KANOMAX GROUP
COMPANIES BE GREATER IN AMOUNT THAN THE PURCHASE PRICE OF THE PRODUCT SOLD BY KANOMAX
GROUP COMPANIES AND CAUSING THE ALLEGED DAMAGE. WITHOUT LIMITING THE FOREGOING, THE
OWNER ASSUMES ALL RISK AND LIABILITY FOR LOSS, DAMAGE OF, OR INJURY TO THE OWNER AND THE
OWNER'S PROPERTY AND TO OTHERS AND THEIR PROPERTY ARISING OUT OF USE OR MISUSE OF, OR
INABILITY TO USE, THE PRODUCT NOT CAUSED DIRECTLY BY THE NEGLIGENCE OF KANOMAX G ROUP
COMPANIES. THIS LIMITED WARRANTY SHALL NOT EXTEND TO ANYONE OTHER THAN THE ORIGINAL
PURCHASER OF THE PRODUCT, OR THE PERSON FOR WHOM IT WAS PURCHASED AS A GIFT, AND STATES THE
PURCHASER'S EXCLUSIVE REMEDY.

8.2. After service


 When you have a problem with your instrument, please check out the “Common Trouble Shooting”
section first.
 If that does not help, please contact your local distributor, or contacts on the last page.
 During the warranty period, we will repair at no charge a product that proves to be defective due to
material or workmanship under normal use. (See Section 14.1 Warranty)
 All return shipping charges are the responsibility of the customer.

24
 Repair after warranty expiration:
 Upon request, we will repair the instrument at the customer’s expense, if the instrument’s
performance is found to be recoverable by providing the repair.
 Replacement parts are available for a minimum period of five (5) years after termination of production.
This storage period of replacement parts is considered as the period during which we can provide
repair service. For further information, please contact your local distributor, or contacts on the last page.

When making an inquiry, please provide the following information :


* PRODUCT NAME ----------
* Model No. ----------
* Serial No. ----------
* Description of the problem : ----------
* Data of Purchase: Day, Month and Year

25
Kanomax Group Companies

■ Americas, Europe, Mid-East, Africa, Oceania

KANOMAX US A, INC.

219 US Highway. 206, Andover, New Jers ey 07821

TEL: 1-800-247-8887(USA ) / 1-973-786-6386

FAX: 1-973-786-7586

URL: www.kanomax -usa.com

E-Mail: info@kanomax-usa.com

■ Japan & Asia

KANOMAX JAPAN, INC.

2-1 Shimizu Suit a City, Osaka Japan 565 -0805

TEL: 81-6-6877-0183

FAX: 81-6-6879-2080

URL: www.kanomax.co.jp

E-Mail: sales@kanomax.co.jp