Sei sulla pagina 1di 4

ASC 2000™ AMBULATORY

SURGICAL CARE TABLE

APPLICATION
The ASC 2000 Ambulatory Surgical Care Table is
STERIS’s premier surgical table for the Ambulatory
Market. This table is designed to provide complete
and flexible positioning for outpatient procedures.
The unique, offset table column design facilitates
surgical team access and comfort while providing
superior C-Arm access. The ASC 2000 Surgical
Table is designed with the fast paced Ambulatory
Surgery Center in mind.

DESCRIPTION
The ASC 2000 Ambulatory Surgical Table is
comprised of a six-section tabletop attached to a
single column and mobile base assembly. This
table features Lateral Tilt, Trendelenburg/Reverse (Typical only - some details may vary.)
Trendelenburg, Back, Leg, and adjustable height
functions as described below.
Overall size (W x L x H): • Standard CSA-C22.2 No. 601.1-M90, Standard for
22 x 79 x 27.5 to 41" (547 x 2009 x 699 to 1045 mm) Electro-medical Equipment, as certified by ETL Testing
Laboratories, Inc.
The standard table configuration includes:
• EN 60601-1: 1993, Amendment 2, 1995, Electrical
• Latex-free tabletop pad
Safety, as certified by ETL Testing Laboratories, Inc.
• Bakelite tabletop
• IEC 601-1-2: 2001, Electromagnetic Compatibility, as
• Power adjustable height, back (upper/lower), certified by ETL Testing Laboratories, Inc.
Trendelenburg and lateral tilt functions
• EN 55011: Group I, Class B, Emissions Testing, as
• Manual, removable, split leg and headrest sections certified by ETL Testing Laboratories, Inc.
• Pendant hand control • EN60601-2-46 – 1st ED., as certified by ETL Testing
• Mobile base with manually operated floor locks Laboratories, Inc.

• 400 pound (181 kg) patient weight capacity • CE Marked to Medical Device Directive, 93/42/EEC

• Up to two weeks operation on internal battery power • FDA Good Manufacturing Practices Regulations
21CFR820 for Medical Devices
• Class 1 Equipment
STANDARDS
• Type B Equipment
ASC 2000 tables are in compliance with the following
standards: • Suitable for continuous operation with short-time
loading (15% duty cycle)
• Underwriters Laboratories (UL) Standard 2601-1 - 1st
ED., as certified by ETL Testing Laboratories, Inc. • IPX-4

The Selections Checked Below Apply To This Equipment

VOLTAGE PAD CONFIGURATION


❏ 100 VAC, Single-Phase, 50/60 Hz ❏ Latex-Free Pads (standard)
❏ 120 VAC, Single-Phase, 50/60 Hz
❏ 220 VAC, Single-Phase, 50/60 Hz ACCESSORY PACKAGES Item ____________________
❏ 230/240 VAC, Single-Phase, 50/60 Hz ❏ Standard Accessory Package Location(s) ______________
POWER SOURCE
❏ Battery/Electric-Powered Table

SD801 (09/01/06)
FEATURES ing the ON switch on
the Hand Control.
Tabletop is constructed of six-sections and is 79" (2009
Activation of the Battery
mm) long. Unobstructed Imaging Length of 71" (1803 mm)
System permits the
and 85% C-Arm access without table movement is
table to be articulated
available. For attaching accessories, the tabletop sections
as necessary for
include a standard stainless steel side rail on both sides -
approximately two
located where they will not obstruct the imaging area.
weeks (without facility
Hook fastener strips on the tabletop sections permit instant
AC power). The
application and removal of the 2" (51 mm) thick, latex-free
batteries are continu-
mattress pad. The radiolucent tabletop sections enable
ously charging as long
viewing of the entire anatomy (see illustration on next
as the table is supplied
page). The X-Ray top design enables cassettes to be
with the appropriate AC
loaded from the head, foot, or either side for a full range of
voltage and the
exposure angles.
BATTERY SWITCH is in
Column supports tabletop and includes lift cylinder, the ON position.
bearings, hydraulic piping, hydraulic actuators for
Pendant Hand Control
Trendelenburg, and electrical wiring. These components
is constructed of
are enclosed by two telescoping stainless column shrouds.
injection-molded black
The stainless shrouds are of two-piece construction for
plastic (two-piece,
service accessibility. Hand Control connects at top of
sealed) and is the
column.
primary interface for
Base structure is painted cast aluminum. The base cover table operation. It has a
is a one-section, welded, stainless-steel enclosure. Two coiled 3' (0.9 m), 12'
swivel casters inside the base cover facilitate table (3.7 m) extended, long
relocation and movement. Two mechanical, manually cord. The hand control
KEY:
operated, floor-locks are supplied. The microprocessor, plugs into a receptacle
power supply assembly, and floor lock actuator pedal are 1. AC Power Indicator
located at the top of the 2. Battery Power Indicator
also within the base. Table power cord and the optional column. Membrane 3. Control Power OFF
foot control can be plugged into receptacles centered on touch switches provide, 4. Trendelenburg
5. Height DOWN
the base head-end cover. Auxiliary Control System through a slave CPU, 6. Left Tilt
Controls are located beneath a cover plate in this same input signals to activate 7. Lower Back Section DOWN
8. Upper Back Section DOWN
area (see illustration on last page). Included are: REV./ table functions and 9. Low Battery Indicator
TREND., HEIGHT UP/DOWN, SIDE TILT RIGHT/LEFT, articulations. LED 10. Power Indicator
11. Control Power ON
LOWER BACK and UPPER BACK UP/DOWN, and POWER indicators provide the 12. Reverse Trendelenburg
TRIGGER switch. following table operation 13. Height UP
14. Right Tilt
Electric controlled system provides powered tabletop information: 15. Lower Back Section UP
positioning. Hydraulic actuators perform all powered table • AC Power 16. Upper Back Section UP
motions. The primary control system is a master computer • Battery Power
located in the table base. This computer selects which Hand Control
• Low Battery
outputs are to be actuated based on inputs from a slave
CPU in the pendant hand control and optional foot control. • ON (Power Indicator)
The hand control is a tethered pendant that hangs from Touch Pads include:
standard side rails. It provides user inputs (from touch
• OFF/ON
pad switches via a slave CPU) to the master computer. It
also includes status LEDs. An auxiliary control, located on • TREND./REV.
the table base, enables all table actuations. • HEIGHT DOWN/HEIGHT UP
Electrical System, the input electrical power, is fed by a • TILT LEFT/TILT RIGHT
detachable three-wire grounded power cord into an • LOWER BACK DOWN/UP
isolation transformer. The power is reduced to 24 V, and is • UPPER BACK DOWN/UP
rectified to DC. The operating system is powered by the 24
Auxiliary Control System can be actuated at any time and
VDC. Fuses are used for protection of the system. The
will allow table operation in the event of primary control
control operates on 24 VDC. The hand control (a slave
malfunction. A row of toggle switches located on the front
CPU) and optional foot control consist of switches that
of table base behind a cover panel (see illustration on last
open or close to signal the computer in the table base. The
page) is used for table movements as long as electrical
hand control includes feedback LEDs. If AC main power
power is still available. Functions included are REV./
fails, the battery system can be used to power the table.
TREND., HEIGHT UP/DOWN, SIDE TILT RIGHT/LEFT,
The Battery System is activated by ensuring the BATTERY
LOWER and UPPER BACK UP/DOWN, and POWER
SWITCH on the front of the table base is ON and depress-
TRIGGER switch.

2
Dimensions
Letter Inches mm
A 13 330
N B 11.4 290
O
A B C C 10.6 268
D 0.8 20
E 3.9 98
F 8.1 205
M
D G L G 1.8 45
H 21.7 550
E H J 31.7 805
K 29.9 760
F J K L 2.2 55
M 3.9 100
N 10.8 275
ASC 2000 Table Image Amplification Coverage O 15.4 390

TABLE OPERATION • Consult with


The ASC 2000 tabletop powered articulations are posi- STERIS Market-
tioned/articulated by pressing the desired position touch ing concerning
pad on the hand pendant or optional foot control. standard STERIS
The following pre-operative actions must be completed accessories for
before the table can be used: use with this
table
• Power activation: Plug the table power cord into the
head of the table base or toggle BATTERY SWITCH to
ON for battery power. PREVENTIVE
MAINTENANCE
• Turn the table on: Press Power ON button.
A global network of
• Lock the table: Press the Floor Lock Pedal down into
skilled service
the locked position to lock the floor locks.
specialists can
provide periodic KEY:
TABLE MOTION inspections and 1. Headrest
2. Upper Back Section
• HEIGHT DOWN/HEIGHT UP Range: 27.5 to 41" (699 to adjustments to help 3. Lower Back Section
ensure low-cost 4. Seat Section
1045 mm) from floor 5. Leg Section
peak performance. 6. Table Top
• TREND. Range: 0-45°±1° STERIS representa- 7. Side Rail
8. Fixed Handle For Leg Section
• REV. Range: 0-20°±1° tives can provide 9. Pull Rod
information 10. Lock Pedal
• TILT LEFT/TILT RIGHT Range: 0-20°±1° 11. Emergency Backup Switches
regarding annual
12. Hand Control
• LOWER BACK DOWN/UP Motion: UP: 75°±1° maintenance 13. Thumbscrew For Head Section
DOWN: 15°±1° agreements.
• UPPER BACK DOWN/UP Motion: UP: 40°±1° ASC 2000 Table Components
DOWN: 55°±1°
NOTES
• Head Motion (Manual): UP: 33°±1°; DOWN: 87°±1° Approximate Operating Weight: 485 lbs (220 kg)

• Leg Motion (Manual): UP: 0°; DOWN: 90°±1° Patient Weight Capacity: 400 lbs (181 kg); 300 lbs (136 kg)
(Swivel 90° toward outside) when using the U-Shaped Head Rest
ASC 2000 Table is manufactured for STERIS by
MEDILAND of Taiwan.
ASC 2000 SPECIFIC ACCESSORIES
• Foot Rest (Left/Right)
UTILITY REQUIREMENTS
• Drain Pan 100 V, 50/60 Hz, 2.10 Amp (nominal); 120 V, 50/60 Hz, 2.0
Amp (nominal); 220 V, 50/60 Hz, 1.17 Amp (nominal); or
• U-Shaped Head Rest 230/240 V, 50/60 Hz, 1.10 Amp (nominal)
• Cassette Top
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
• Foot Controller APPLICABLE LOCAL AND NATIONAL CODES AND
• Opthalmic Headrest (with Surgeon Wrist Support) REGULATIONS.

3
Reference the following equipment drawing for installation details.
Equip. Dwg. No. Equipment Drawing Title
150830-460 ASC 2000 Ambulatory Surgical Table
Dimensions are typical –
drawing is not to scale. FOOT END
(Rear End) Dimensions
D J Letter Inches mm
E F G H K A 21.5 547
B 19.7 500
C 15.7 400
A B C D 10.2 260
HEAD END
E 0.8 20
(Front End)
F 11.6 295
G 11.8 300
H 15.0 380
P J 2.0 50.5
K 26.8 680
L 27.6 700
ML M 41.1 1045
N 24.0 609
O 39.5 1004
P 79.1 2009
Q Q 5.1 130
R R 21.0 533
S O N S 22.8 578

VIEW A

WARNING – EXPLOSION HAZARD: Table must not be used in the presence of flammable anesthetics.

VIEW A
(Not To Scale)

For further information, please contact:


STERIS Corporation,
Montgomery, Alabama, is an ISO 9001
STERIS Corporation
and ISO 13485 certified facility.
5960 Heisley Road
Mentor, OH 44060-1834 • USA The base language of this document is
440-354-2600 • 800-548-4873 ENGLISH. Any translations must be
www.steris.com made from the base language document.

This data is intended for the exclusive use of STERIS customers, including architects or
SD801 ©2006, STERIS Corporation. All rights reserved. (09/01/06) designers. Reproduction in whole or in part by any party other than a customer is prohibited.

Potrebbero piacerti anche