Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
0 INTRODUCTION:
In our today’s country, radiation protection has turned out to be one of the most
alarming issues in majority of our radiological diagnostic centres. Though radiography is
gaining wide acceptance in nearly all part of the continent, it is equally important that
the radiation protection of staffs and patients is taken into consideration and this is the
sole responsibility of the radiographer operating the x-ray equipment. For this reason,
quality control was considered one of the tools in optimizing best practice in the area of
radiation protection of both the patient and the staff and also in maintaining the
production of consistently high-quality diagnostic radiographs. (Martin, 2007).
According to ICRP (1991), there are two basic principles of radiological
protection. There are; justification of the practice and optimization of protection. In the
area of optimization of protection, there is considerable scope for reducing doses to
patient without any loss of diagnostic information, but the extent to which the
measures available are used varies widely. Optimization of radiation protection does not
necessarily mean the reduction of doses to the patient or by operating in the absence
of a demonstrable threshold for stochastic effects but by trade-off between the benefits
of dose reduction and the costs of achieving these reductions. A number of factors
facilitate this trade-off. One of such factors is the quality control measurements and
practices of the department. (Saure and Hagemann, 1995).
This quality control as noted by Maccia and Moores (1997), involves a
quantitative and qualitative measurements and test of the performance of x-ray
equipments, programs and practices of that diagnostic centre. Hence, in instituting a
good quality control system in our x-ray department, a quality control program which
will monitor the basic components of the imaging process at a low cost through the use
of simple, inexpensive tools and minimal staff time must be put in place. This quality
control will now determine their adequacy in terms of production of high-quality
diagnostic radiographs and also evaluates their contribution to all the radiation
protection practices therefore outlining their role in the optimization of the radiation
protection of that centre.
1
2.0 DEFINITION OF TERMS
Quality Control (QC): These are specific actions designed to keep measurable
aspects of the process involved in manufacturing a product or providing a service within
specified limits. These actions typically involve measurement of a process variable,
checking the measured value against a limit, and performing corrective action if the
limit is exceeded. (CRCPD Pub., 2001).
Quality Assurance (QA): These are planned and systematic actions that provide
adequate confidence that a diagnostic x-ray facility will produce consistently high
quality images with minimum exposure of the patients and healing arts personnel. The
determination of what constitutes high quality will be made by the facility producing the
images. Quality assurance actions include both quality control techniques and quality
administration procedures. (CRCPD Pub., 2001).
Quality Control Program: allows a facility with limited resources and personnel to
monitor the basic components of the imaging process at a low cost through the use of
simple, inexpensive tools and minimal staff time.
Quality Assurance Program: Is an organized entity designed to provide quality
assurance for a diagnostic radiology facility. The nature and extent of this program will
vary with the size and type of the facility, the type of examinations conducted, and
other factors. (CRCPD Pub., 2001).
Optimization: Optimization in the field of diagnostic radiology simply means any
process or procedure which ensures that doses due to appropriate medical exposure for
radiological purposes are kept as low as reasonably achievable (ALARA) consistent with
obtaining the required diagnostic information, taking into account economic and social
factors. (IAEA Pub., 2004).
Cost-Effectiveness: Chesney (1981) defined Cost-Effectiveness as “the ratio of
spending to the efficiency of production that follows the result”. It compares the relative
expenditure (costs) and outcome (effects) of two or more courses of action.
2
3.0 AIMS AND OBJECTIVES:
The main aim and objective of this colloquium is to promote awareness creation
about the practical implementation of quality control protocols and image quality
evaluation by consistently implementing simple and inexpensive actions such as the use
of appropriate screen/film combination, use of secondary radiation grids when
necessary, etc.
It also aims to create pools of expertise in the area of radiation protection of
patients, hence alleviating the dangerous practices of unnecessary irradiation of our
patients.
A further objective of this research seminar is to offer assistance and guidance to
an imaging scientist implementing and operating a quality assurance program any in
diagnostic radiology department across the globe.
3
Some details which add up to improved cost-effectiveness includes;
a. The number of repeated radiographs is reduced.
b. The rate of flow of patients through the department is improved.
c. The department’s ability to meet the demands made upon its services is
raised.
d. Quality of radiograph produced is higher.
e. Standardization of the radiographic results is achieved and maintained.
f. The reliability efficiency of automatic processors and of x-ray equipment is
improved.
4
5.1 General Consideration:
An adequate quality control program for any individual facility will depend on a
number of factors which include, but may not necessarily be limited to, items such as
the type of procedures performed, type of equipment utilized, and patient workload.
This program is developed under the guidance and supervision of a medical physicist
qualified in this area of expertise by education, training, and experience.
5
2. Forms should be filed as part of the room log.
3. The charting of trend data is a recommended procedure which will allow easy
identifications of variation with time. This is of particular value in the case of
film processors.
6
5.5 The Radiograph As A Quality Control Tool:
Patient’s radiographs are also considered one of the quality control tools of which
they are being checked on periodic basis and should be factored into any
departmental evaluation program.
5.5.1 Rejected Films Analysis: Rogers (2008) defined Film Reject as “a film deemed
useless and discarded with another film being taken” and A Repeat Film as “a
film retaken to provide extra/missing diagnostic information sent with the original
for reporting”.
Film Reject Analysis according to Suleiman and Showalter (1984) are
periodic assessment and checks on rejected films as well as the accepted ones
usually on monthly intervals so as to identify the problem, determine it cause
and find solutions to it, all aiming at reduction of film reject rate.
The causes of rejection of films are analyzed according to the following;
- Too dark, too light (under/over exposure)
- Positioning/Collimation errors
- Patient movement
- Processing errors
- Others
Reference data should also be assigned to this analysis e.g. date/time, operator
ID code, room, exam type, reject or repeat, etc.
5.5.2 Accepted Films Analysis: Good practice should always question the adequacy
of radiographs of less than optimal quality for their acceptability in making a
diagnosis. Repeating a procedure to get a film of optimal quality is often not
necessary and should be evaluated in terms of the radiation exposure and cost
of the retake. Since one should expect to find films of less than optimal quality in
a departmental file, an analytic review of these films should be made on a
regular basis.
7
6.0 FACTORS IN OPTIMIZATION OF CONVENTIONAL RADIOGRAPHY
The formation of image of the body involves interplay between many different
factors. To achieve the correct balance between patient dose and image quality, it is
necessary to understand the way in which images are formed, and to know the factors
that influence the image quality and the radiation dose received by the patient, so that
appropriate options can be selected. These factors include;
8
6.2 Exposure Control:
To produce an image on a film with an acceptable level of contrast, the exposure
must be within a relatively narrow range of doses. This is to say that all exposures must
be as low as reasonably achievable. (ICRP Pub. 60, 1990).
Two major factors according to Martin and McKenzie (1993) are involved in the
quantity of radiation produced by the tube. These are; the tube potential difference
(kVp) and the beam filtration. They also noted that, the exposure factors used will be
optimized through the experience of the radiographers and exposure charts employed
for each X-ray unit. The charts provide a guide to the best factors for different
examinations for a patient of standard build. But however, adjustments will need to be
made for patients of different sizes.
9
iii. Increases the random background noise of the film and all these will
result in
iv. a reduction in the contrast of the film.
The amount of scattered radiation can be reduced by means of an anti-scatter
grid. (Bauer, 1998). The grid consist of a plate containing thin strips of lead lying
perpendicular to the plated surface, which are sandwiched between a low attenuation
inter-space material which are radiolucent materials made of either aluminium or
polyester. X-ray photons are more likely to be attenuated by the lead strips. (Sandborg
and Carisson, 1993).
Secondary radiation grids are not used for examination of the body parts but
only parts intended to produce scatter maybe as s result of increase in the tube
potential difference (kVp).
10
carefully optimized. A system of quality control that involves checking temperatures of
processing chemicals and carrying out sensitometry, involving development of a test
strip of film exposed to a range of light levels ensures optimal performance. (BIR Pub.
2001). These checks should be carried out daily to monitor performance in terms of film
density, contrast and background fog level. The performance levels of processors that
have a relatively low workload need to be monitored carefully.
Gray and Winkler (1983) noted that, film processing affects the film density;
therefore, it influences the speed index. Thus, the measurements of the characteristic
curve for a film will also reveal problems with processing. If a film is taken with
optimized processing, it can be considered the reference standard. Checks can then be
made by comparing future results with the reference standard to identify any
deterioration.
11
Required Equipment: Includes; a) Sensitometer b) Dedicated box of control film c)
Densitometer.
Steps:
1. Expose the control film with the sensitometer.
2. Develop the film.
3. Determine the average optical density of the mid-density step and record it on a
form.
4. Determine the average optical density difference and record.
5. Measure the background optical density (base + fog) and record. Verify that the
measured values are within a suggested optimal performance criteria.
Corrective Action:
The tests should be repeated if the values are outside the performance criteria. If, after
repeating, the results are still out of limits, look for processing problems and contact the
processor service supplier.
12
6. Record your results on the monthly quality control checklist.
Evaluation: Compare the current film with the comparison film. If the densities are not
within 1 step of the comparison film, constancy has not been maintained and clinical
images should not be taken until the problem has been identified and corrected.
Corrective Action: Repeat the test to confirm results. Verify that the processor is in
control. Contact your x-ray and processor service engineers.
13
dust may accumulate. Clean film hangers.
Corrective Action: If automatic processor can not be maintained at its optimal
operating temperature, call processor service supplier.
14
Determine if light is functioning and is clearly defined under normal operating
conditions, without visible dust or foreign matter shadows.
2. Collimator beam limiting devices (BLDs) available and used.
If unit provides variable collimation, determine that they are functioning correctly
and smoothly. If manual beam limiting devices are being used, assure they are
sufficient for confining the x-ray beam to the area of clinical interest. Assure that
both types are being used correctly.
3. Locks and detents operable.
Check to make sure all locks and detents are functioning as intended. Assure
that the x-ray tube maintains its position at the clinically used angles.
4. Boom smoothness of motion.
Determine if boom moves easily without catches or interruptions.
5. Grid condition and operation.
If grids are being used, check that grid lines, grid cutoff, or grid damage is not
visible on films. Assure that grid is properly positioned, centered to central ray
and if a focused grid is being utilized that the correct focal distance if being used.
6. Condition of cables.
Inspect all cables for frayed coverings, kinks, and determine that cables are free
from friction from other objects.
7. Tube or generator oil leakage.
Visually inspect areas around x-ray tube and generator for oil or abnormal
collection of dust attaching to oil leaks.
8. Cassettes and screens condition.
Cassettes and screens should be cleaned regularly. Check screen condition for
dust particles, scratches, and areas of discoloration. Assure screens are properly
fitted and attached to cassettes. Check cassette latches to make sure they are
functioning properly and are not broken. Cassettes and screens should be
replaced if necessary.
15
9. Loaded cassette storage.
Determine that loaded cassettes are stored in an area that is properly shielded
from radiation to prevent exposure. They should be stored off the ground and
kept free from dust.
10. Control panel indicators.
Assure all control panel switches, lights, and meters are functioning correctly.
11. Technique chart.
Make sure a technique chart is available, current, and appropriate for all
procedures normally performed.
12. Patient view ability.
Determine that means are provided to permit continuous observation of the
patient during the x-ray exposure.
13. Exposure switch placement.
Assure the exposure switch is mounted in such a way that exposure can only be
made with the operator in a protected area during the entire exposure. If unit is
portable or mobile without a portable protective barrier, assure cable on
exposure switch provides means for the operator to be at least nine feet from
the tube housing during the exposure.
14. Lead aprons, gloves, collars, etc.
Assure proper items are available and stored correctly without bends or folds. If
abnormal areas are found, complete procedure 13.
Corrective Action: Missing items from the room should be replaced as soon as
possible. Malfunctioning equipment should be reported to the x-ray service engineer for
repair or replacement as soon as possible.
16
facility should decide on its own, but should strive for a repeat rate of no greater than 5
to 7%.
Frequency:
1. Ongoing tracking of films
2. Quarterly data analysis
Steps:
1. Determine the reason for film repeat as compared to the categories listed on the
data sheet.
2. Record these numbers on the Repeat Analysis Form.
3. Determine the total number of repeated films and the total number of films
exposed. The overall repeat rate is the total of repeated films divided by the total
number of films exposed during the test period.
4. By dividing the number of repeats per category by the total number of repeated
films, a facility can determine the repeat rate per category.
Corrective Action: The percentage of repeats should guide the facility to focus their
efforts to those areas needing the most attention. For example, films that are too light
or too dark may be due to processing problems, equipment problems that require repair
or re-calibration, or technique charts may need updating.
17
Corrective Action: If storage conditions exceed manufacturer’s recommendations,
take the necessary steps to resolve the problem.
18
7.9 Intensifying Screen Cleaning Procedure:
Objective: To assure that screens and cassettes are free of dust and dirt particles that
may degrade image quality.
Suggested Performance Criteria: Minimize artifacts on films from screens or
cassettes.
Frequency:
1. Quarterly or semiannually (depending on workload and amount of dust in the
environment)
2. When a problem is noticed
Required Equipment:
1. Screen cleaner (as recommended by manufacturer)
2. Lint-free gauze pad or cloth, or camel’s hair brush.
3. Canned air (available from photographic supply store)
Steps:
1. Visually inspect the condition of the intensifying screen.
2. Dust the screen with the camel's hair brush and canned air.*
3. If foreign material (e.g. dirt, developer solution) cannot be readily removed
with the camel's hair brush, use liquid screen cleaner.
4. After cleaning with manufacturer approved cleaners, screens should be
allowed to air-dry, standing vertically, before returning the cassette to use.
5. Record results on the Quarterly QC Checklist.
Corrective Action: If the screen shows signs of cracking, fading, or discoloration it
should be evaluated for replacement.
Assure that the canned air used to clean the screens is "clean" air. If the air
contains moisture, oil, or other contaminants, you may be introducing artifacts or
damaging the screen.
19
7.10 Darkroom Integrity or Fog Test:
Objective: To determine and minimize the amount of darkroom fog.
Suggested Performance Criteria: An optical density increase of 0.05 or less.
Frequency:
1. Semiannually, with each type of film used clinically
2. After bulb or filter replacement
3. After changing or adding types of film
Required Equipment: Includes; opaque material (manila folder), watch or timer,
attenuation block (aluminum step wedge, phantom, acrylic block) to create a medium,
optical density of about 1.0 on the film, densitometer.
Steps:
1. Load a cassette with film and place on a flat surface.
2. Center the attenuation block and expose the film using an x-ray technique that
will result in an optical density of about 1.0 after the film is processed.
3. With the safelights on, place the exposed film on the work area in the darkroom.
Cover half the film with opaque material, bisecting the latent image parallel to
the long axis of the film.
4. Leave exposed film on the counter for 2 minutes, then process as usual.
5. While waiting 2 minutes for darkroom fog test, look for any sources of
extraneous light. Any light leaks identified should be repaired as soon as
possible.
6. Inspect the processed film. If there is no discernible delineation between the
shielded and unshielded sides of the film, there is no fog problem.
7. If a line is evident, measure the optical densities of both sides of the line with
the densitometer. If the density difference is greater than 0.05, corrective action
should be taken.
8. Record results on the Semiannual QC Checklist.
Corrective Action: Repeat the test with the safelight off. If the results remain the
same, the problem may be caused by a light leak or extraneous light. If the fog level
20
disappears, the fog was due to the safelight and remedial action must be taken to
correct the problem.
Possible Sources Of Darkroom Fog:
- Safelight filters (old or compromised) - Radios
- Safelight housing - Fluorescent light afterglow
- Safelight too close to work area - Light leaks
- Light bulb of incorrect wattage or type - Suspended ceilings
- Ancillary indicator lights on processor - Timers
- Any place there is a hole cut in the wall
- Excessive ambient light through the tinted viewing windows of daylight loading
systems.
21
3. Place the wire mesh on top of the cassette and expose the cassette. (Suggested
technique factors are: 5-10 mAs, 50 kVp; 2 mAs, 70 kVp; or 3-5 mAs, 60 kVp).
4. Process the film. The optical density of the area between the wires of the mesh
on the film should be between 1.5 and 2.0.
5. View the film on a in a room with low ambient lighting. Stand 6 to 8 feet away
from the illuminator to evaluate the film.
6. Areas of poor contact will appear as dark areas on the film.
7. Record results on the Annual QC Checklist.
Corrective Action:
Large areas (>2 cm in diameter) of poor contact may indicate the need for corrective
action. Clean the cassettes and retest. Areas of poor contact around the periphery of
the cassette may indicate faulty latches or worn seals on the cassettes. If the area of
poor contact is not eliminated by cleaning, consider replacing the cassette.
22
4. Expose (65 kVp, 4 mAs) and develop the film. If field edges are not well defined,
adjust techniques accordingly and repeat this step.
5. Measure the distances between the light (where the coins touch) and x-ray fields for
all coin locations.
6. Add differences for each set of coins along and across the film, and divide each set
of differences by the SID (Example: (1.5" along table / 28")(100) = 5.38% and (0.5"
across table / 28")(100) =1.79%).
7. Percentage differences greater than 2.0% in either direction should be corrected as
soon as possible.
8. Using the same exposed film, determine the center of the x-ray field (darkened
portion of film) using a straight edge.
9. In the same manner, determine the center of the film.
10. Measure distance between the two centers and calculate the difference as a
percentage of the SID. If the percentage difference is greater than 2.0%, corrective
action is necessary.
11. Measure the dimensions of the x-ray field on the film. If the difference between the
indicated and measured field size exceeds 2% of the SID, corrective action is
required.
12. Record on the Annual QC Checklist.
Corrective Action: Malfunctioning equipment should be reported to the x-ray service
engineer to correct the problem.
23
7.13 Protective Device Integrity Check:
Objective: To assure that the various protective devices such as lead aprons, gloves,
gonadal shields, and thyroid collars provide
optimal protection when positioned appropriately.
Suggested Performance Criteria: No breaks in lead lining of protective garments.
Frequency: Annually
Required Equipment: Lead aprons, gloves, gonadal, and thyroid shields
Steps:
Option 1: If an image intensified fluoroscopy unit is available, this is the preferred way
to inspect the aprons, gloves, and collars.
1. Lay out the item to be checked on the table.
2. Examine the entire item using the fluoroscope.
3. Record results on the Annual QC Checklist.
Option 2: If an image intensified fluoroscopy unit is not available:
1. Closely inspect each item for kinks and irregularities.
2. Take a radiograph of suspect areas.
3. Process the film and look for breaks in the lead lining.
4. Record results on the Annual QC Checklist
Corrective Action: Any item displaying breaks in the lead lining should be replaced.
24
5.0 CONCLUSION:
In conclusion, based on patient dose measurements, comparison with reference
values, assessment of image quality, the introduction of quality control and corrective
actions, if needed, and re-evaluation of patient doses and image quality, has
demonstrated its effectiveness for optimization of radiological protection program.
Finally, a ‘culture’ of regular patient dose measurements, film reject analysis, and
image quality assessment need to become part of diagnostic radiology.
6.0 RECOMMENDATION:
From all the information outlined in our colloquium, it has been proven beyond
reasonable doubt that quality control programs and protocols form an essential part of
the optimization process. Therefore, such programs covering physical and technical
parameters associated with the types of x-ray examination being carried out needs to
be instigated in every medical x-ray facility. For that reason, we strongly recommend
that all state radiation control personnel should be encouraged to promote quality
control as a proven means to reduce doses of exposure, increase and maintain
diagnostic image quality, and limit health care costs.
25
REFERENCES:
Bauer, B. (1998), Reference Doses and Quality in Medical Imaging – What the referring
Practioner and Directing Medical Staff should know: Report EUR 18532 EN;
Radiation Protection Dosimetry 80; Pg. 1-3.
Chesney DN., Muriel OC. (1981), Radiographic Imaging: 4th Edition. Pg. 328-29;
Blackwell Scientific Publications; Boston Melbourne.
Gray JE, Winkler NT, Stears JG, Frank ED. (1983), Technical Aspect of Screen-Film
Radiography, Film Processing and Quality Control. Aspen System Inc, Maryland.
Hart D, Wall BF, Shrimpton PC. (2000), Reference Dose and Patient Size in Paediatric
Radiology: National Radiological Protection Board Report NRPB – R318;
Chitton, UK.
26
Maccia C, Asriche – Cohen M, Wadeau X. (1995), The 1991 CEC Trial On Quality Criteria
for Diagnostic Radiographic Images. Radiation Protection Dosimetry; Pg 57: 111-
117, Luxembourg.
Maccia C, Moores BM, Wall BF. (1997), The 1991 ECE Trial On Quality Criteria for
Diagnostic Radiographic Images: Detailed Results and Findings; Report EUR 16635
EN. Office for OfficialPublications of the European Communities, Luxembourg.
Martin CJ. (2002), Interactions of Ionizing Radiations with Matter; Practical Radiation
Protection in Health Care: Pg. 13-26. Oxford University Press; Oxford, UK.
Martin CJ, Darragh CL, McKenzie GA. (1993), Implementation of a Program for
Reduction of Radiographic doses and results achieved through Increase In tube
potential. British Journal of radiology 66(783): Pg. 228-223.
Martin CJ, Sutton DG, Sharp PF. (1999), Balancing Dose and Imaging Quality: Appl
Radiat Isot. 50(1); Pg. 13-26.
Rogers A. (2008), Reject Analysis: A low Tech Solution; Nottingham City Hospital NHS
Trust: Report 6; Pg. 5-9; Paragon Press, Nottingham.
Sandborg M, Dance DR, Carisson GA (1993), Selection for Anti-Scatter Grids for
Different Tasks: the advantage of low atomic number cover and
interspace materials; British Journal of Radiology 69(820); Pg. 115-63.
Shrimpton PC, Wall BF, Jones DG. (1984), The Measurement of energy Imparted to
Patients during Diagnostic X-Ray Examination. Using the Dimentor Exposure – are
product metter;
Stewart CB. (1993), Radiologica Science for Technologist; Physics , Biology and
Protection: 5th Edition; Pg. 439-30. Mosby Year book, New York.
Suleiman OH, Showalter CK, Gross RE, Bunge RE. (1984), Radiographic Film Fog In
Darkroom: Pg 151:237-238. Neigman’s Unity Press; Toronto, USA.
27