Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity


error” or “coilmalfunction error” with Sense NV16 coil and Sense
Spine-15 coil.
Applies To

Magnetic Resonance Magnetic Resonance, Achieva, Achieva XR

Problem Description Details

Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfunction


error” can occur when Sense NV16 coil and Sense Spine-15 coil (together or stand-alone) are
connected to the system.

Symptoms/observations:

- Scan stops during preparation phase or halfway a scan with messages:


• Coil A Malfunction
• Coil B Malfunction
• Unknown coil or coil with malfunction at conn:
• Interlock (CoilMalfunction)
- Power supply -5VB not between -5V and -4.9V, measured at PICU or PFEI.

Problem Cause

Certain coils give a malfunction error when they detect the -5V detune voltage to be to low.
Known coils are Sense NV16 coil and Sense Spine-15 coil from Invivo, detection electronics are
out-of-spec.
About 20% of the installations are creating persistent problems that are not resolved by
exchange of coil or voltage supply parts.

The coils do not operate on the specified voltage range of -3.2V to -5.5V – specified on the
ODU connector.
Problem Solution

Occurrence of the problem can be reduced by adjust the -5VB of the FEPSU power supply -
located in SFB - according to next procedure.

Pre-conditions

http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 2 of 4

• No coils connected to system, only QBC


• No tests/scans are running (system in stand-by state)
• Voltages are measured at ODU connector

1. Measure the -5V voltage on the ODU connector, pin 1-2, see figure 1

Figure 1 – Overview ODU connector

2. If voltage is within specification ( spec = - 4.95V ± 0.05V ), leave as it is


3. If voltage is out of specification (spec = - 4.95V ± 0.05V ) adjust FEPSU power supply,
for adjustment point see figure 2.

Figure 2 – Two types of possible FEPSU power supplies

http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 3 of 4

http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 4 of 4

Sense NV16 coil and Sense Spine-15 coil that still fail, will be improved by INVIVO. They must
be sent back for repair.

!!! Include problem description when returning the coil. !!!


Terms and Conditions for use of the Eureka Fault/Fix Solution Knowledgebase:
You may not use or access this knowledgebase without prior acceptance of these terms and conditions. Use of this knowledgebase constitutes your
acceptance of these terms and conditions. You may only access this knowledgebase as an authorized user. You must treat all materials displayed in this
knowledgebase as company confidential. All materials displayed in this knowledgebase are the property of Philips Medical Systems B.V. and are protected
by copyright, trademarks, or other intellectual property rights afforded by law. You agree not to copy, reproduce, re -transmit, distribute, disseminate, sell,
publish, broadcast, circulate, or otherwise publish any such information in this knowledgebase to any third party. Transferring and/or copying in whole or in
part of the knowledgebase to other databases is only allowed after prior written consent of Philips Medical Systems B.V. As an authorized user of the
knowledgebase you are allowed to download the content and other related information for use within Philips Medical Systems provided that you treat them as
Company Confidential. Knowledgebase information may be changed or updated by PMS B.V. without notice. Please note that information content may not
apply to all system configurations due to variation in product or product upgrades. Use this information only in conjunction with your professional knowledge
and judgement. Copies of any kind of this information are uncontrolled. After being registered as an authorized user, you are requested to enter only once
per browser session, your user name and password to get access tothe knowledgebase information.

http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010

Potrebbero piacerti anche