Symptoms/observations:
Problem Cause
Certain coils give a malfunction error when they detect the -5V detune voltage to be to low.
Known coils are Sense NV16 coil and Sense Spine-15 coil from Invivo, detection electronics are
out-of-spec.
About 20% of the installations are creating persistent problems that are not resolved by
exchange of coil or voltage supply parts.
The coils do not operate on the specified voltage range of -3.2V to -5.5V – specified on the
ODU connector.
Problem Solution
Occurrence of the problem can be reduced by adjust the -5VB of the FEPSU power supply -
located in SFB - according to next procedure.
Pre-conditions
http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 2 of 4
1. Measure the -5V voltage on the ODU connector, pin 1-2, see figure 1
http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 3 of 4
http://incenter.medical.philips.com/doclib/fetch/2000/4504/4396/896875/3643606/364377... 22/7/2010
Intermittent “unknown coil”, “coil disconnected”, ”coil discontinuity error” or “coilmalfu... Page 4 of 4
Sense NV16 coil and Sense Spine-15 coil that still fail, will be improved by INVIVO. They must
be sent back for repair.