Sei sulla pagina 1di 2

AFMOWMMIRKS

Certificate No . 17450/DBBPAM niie ;`„<rcdit~ : :i N,-.q


Date : May 30, 2019 SUCOFINOO

Issuing Office :
Jl . Arteri Tol Cibitung No . 1, Cibitung Bekasi 17520, Indonesia
Phone/Facs : +62 21 88321176188321166
E mail : jum .cbt@sucofindo .co .i d

CERTIFICATE OF CALIBRATION

CLIENT PT. ENERCO RPO INTERNASIONAL


ADDRESS JL . H . ANWAR HIPPY NO . 18 KAWASAN INDUSTRI
KEL . BATU BESAR, KEC . NONGSA, BATAM-29467

NAME OF EQUIPMENT THERMOMETER GLASS

SERIAL NO . / CODE : 0696165


I

TYPE /MODEL ASTM 34C - 86

MANUFACTURER : BRAND

INVENTORY NO .

The Attachment available is an integral part of this certificate .

This calibration result (s) related to the sample (s) submitted only and the certificate cannot be reproduced in any way, except in full context and with the prior approval in writing
from Sucofindo Laboratory .

This Certificatelreport is issued under our General Terms and Conditions, copy of which is available upon request or may be accessed at www.sucofindo.co.i d

SBU Laboratorium

CBT200144719

251167
CI-2007A

ArA R,W4 A&


Attachment
To Certificate No . 17450/DBBPAM IRINAKUWArvAr
W1kq WAI w
Date : May 30, 2019
9uucof4r,Vpo
ssn
Phone/Facs : +62 21 88321176/88321166
Page 2 of 2 JI . Arteri Tol Cibitung No . 1, Cibitung Bekasi 17520, Indonesia
E mail : jum.cbt@sucofindo .co .i d

CERTIFICATE OF CALIBRATION

TESTING CONDITION,
-TEMPERATURE 20 .2 ± 0 .4 ° C
- HUMIDITY 47 ± 3
STANDARD & EQUIPMENT PLATINUM RESISTANCE THERMOMETER
Serial No . 099/02/2018
Certificate of Calibration No . CE01807
MULTIFUNCTION PRECISION THERMOMETER
Serial No . 1AOOVFO0811
Certificate of Calibration No . CE01955
THERMOHYGROMETER
Serial No . 03838
Certificate of Calibration No . Cal . 6562018
The measurement is traceable to International System of Units (SI)
through P2M-LIPI & LK-160-IDN
CALIBRATION METHOD PO/SH/02
RECEIVED ON May 20, 2019
DATE OF CALIBRATION May 22, 2019
RESULTS
THERMOMETER GLASS SERIAL
NO ./CODE 0696165
RANGE ( - C) 25 105 / 0 .2
Uncertainty of Measurement
Range Equipment Reading Standard Reading Correction
k = 2 , at CL 95 %
°C °C °C °C °C
25 105 25,0 24,9119 -0,09 0,23
44,6 44,3961 -0,20 0,23
P

63,9 63,3388 -0,56 0,23


85,3 85,0636 -0,24 0,23
100,3 99,8512 -0,45 0,23

I;

1494347
LI 1
SCI-2007P

Potrebbero piacerti anche