Physical Property Analyzer PPA4100 Series

On-Line RVP Analyzer


PPA4100 Series

More accurate, on-line measurement of Reid Vapor Pressure will save millions of dollars
in gasoline blending... the PPA4100 is more than an analyzer... it’s a profit-maker.

The PPA4100 Reid Vapor Pressure In addition to the measuring cell, The ASTM D323 Method also
(RVP) Analyzer from ABB offers the three solenoid valves control the requires the measured temperature
first new on-line instrument design filling and emptying of the cell. to be 100 °F +- 0.2 °F. The PPA4100’s
in 25 years for accurately measuring All timing functions are provided cell temperature is controlled to
RVP, throughout the gasoline range by the PPA4100 microprocessor 100 °F +- 0.1 °F.
of 0-20 psia. on the logic board.
The 4:1 fixed ratio and more
Its design allows refineries to eco- Analysis results are obtained in reliably controlled temperature
nomically comply with demanding approximately 8.5 minutes. During results in RVP measurements that
volatility regulations for gasoline this time the cell is purged, zeroed exceed the ASTM D323 manual
blending; established by the Clean and the sample filled. After sample method, with no correction factors
Air Act and its amendments and equilibration, the final pressure or instrument calibration required.
enforced by the Environmental Pro- reading is obtained. Throughout the
tection Agency (EPA). The PPA4100 cycle, the instrument performs self- The PPA4100 also gives equivalent
may also be effectively applied for diagnostics for temperature, purge results to the ASTM D5191 mini-
RVP compliance in custody transfer pressure, sample-filling and cell- method, when run in an off-line
of gasoline product at pipelines leaking, as well as over pressure mode. The primary difference
and terminals. protection. between the PPA4100 and the
devices listed in D5191 is that the
The PPA4100 offers a simple, The PPA4100’s cell is designed so PPA4100 operates in an atmospher-
reliable, compact design that that the total cell volume and the ic basis. The devices described in
requires minimal operator interface amount of sample introduced are D5191 operate on a vacuum basis.
and minimal maintenance. fixed; resulting in a vapor-to-liquid
ratio of 4:1. This 4:1 ratio is a require- When run in its normal on-line
This physical property analyzer ment defined by the ASTM D323 mode, the PPA4100’s results
incorporates a single measuring manual method of measuring RVP, exceed the D5191 method.
cell with a pressure transducer. which was first published in 1930.
In 1993 the PPA4100 became ASTM
Method D5428-93 – Standard Test
Method for Vapor Pressure of Petro-
leum Products (Mini-Method –
Atmospheric).*
50CC @ 15 PSI N2 Liquid Sample
Volume 2.5 ml
Sample Purge

450CC @ 15 PSI N2 Cell Volume 12.5 ml

Sample In

Measuring
Cell

Cell Temperature
100 °F +- 1 °F

Flameproof
Breather

Cell Purge
Nitrogen

Flameproof
Breather

* This method applies only when


Sample Drain
(Must Vent To Atmosphere)
operated in the off-line mode and
when the optional user coefficients
are not used.
PPA4100 Series

The PPA4100 allows refineries to use more economical blending additives,


at their most cost-efficient levels, to achieve required reductions in volatility.

For years, refineries used butanes The challenge is to measure RVP as Service, Support and
for blending. Butanes are readily accurately as possible, so that refiner- Maintenance
available, economical, and increase ies can meet both their compliance
fuel vapor pressure as well as and performance objectives. Refinery The PPA4100 is designed to minimize
octane. Their use was and is industry sources have projected that the need for maintenance, operator
directly related to refinery profits. for every 0.1 psi in increased RVP training and other support services.
accuracy, the typical refinery will However, if the client should require
The EPA enforces volatility regu- save from 0.5 to 1 million dollars or request, ABB Analytical will pro-
lations for gasoline. Generally, annually. vide start-up services, maintenance
gasoline vapor pressure should services, training services, recondi-
be less than 9 psi in the summer With the PPA4100, refineries have tioning, repair and replacement parts
months and greater than 9 psi in the an on-line instrument that will service.
winter months. The requirements precisely measure Reid Vapor
for individual states are summarized Pressure; allowing the refinery to Training services are available for
in ASTM D4818, Table 4. effectively control the volatility of virtually every aspect of operating
blended gasoline, optimize blend- and maintaining ABB Analytical
This regulation, along with the ing processes and reduce blending analyzers. Training may be arranged
addition of oxygenates and other costs to their minimum levels. on-site or at any of our training
compliance requirements, have centers.
redefined gasoline blending rules Pipeline and terminal operators will
and left refineries vulnerable to also find that the PPA4100 is an Maintenance services are available
hundreds of millions of dollars economical and effective instrument on an unscheduled, as-needed
in additional costs. for RVP compliance, with either basis, or by way of long-term,
on-line or off-line application. scheduled maintenance agreements.

For off-line applications and verification


of on-line results, a specialized syringe
is available that provides a convenient
method to draw an air-free sample,
with sufficient pressure to introduce the
sample to the measuring cell. Syringe
volume is approximately 95 milliliters.
PPA4100 Series

RVP Analyzer designated as ASTM Method D5482-93 in the off-line mode.

Specifications
Performance Specifications General Installation Requirements
Cycle Time: 8.5 min. (4-20 ma output is held at 1/4 inch tube connections for:
the current value and updated after N2 Supply
each cycle). Sample In
Repeatability: 0.05 psia. Low Pressure Return
Reproducibility: 0.2 psia. Cell Drain
Operating Range: 0.20 psia. Seals: Standard seal material is Kalrez.
Pressure Transducer: High performance, Electrical Power:
accuracy 0.5 % of full scale. VAC, 50/60 Hz, 2 amps.
Outputs: 4-20 ma isolated into 600 OHM max. Protect the analyzer from direct sunlight and rain.
Common alarm output N.O. or N.C. Ambient temperature not to exceed 90 °F.
28 VDC at 0.1 amps max. End-of-cycle Cell drain must be unrestricted vent to atmosphere.
contact – voltage less contact closure Cell drain spent sample approximately 1000 cc/day.
capable of switching 28 VDC at 0.1 Must have 50-100 micron filtration up-stream of the
amps maximum. Contact is energized analyzer (provided by user).
for approximately 10 seconds at the Minimum sample inlet operating pressure 35 psia.
end of each update. Maximum low pressure return 30 psia (based on
sample inlet pressure).
Dimensions Sample temperature not to exceed 90 °F.
Weight: Approximately 100 lbs. (45.36 kg). Nitrogen supply (instrument grade).
Height: 54 inches (1371.6 mm).
Depth: 9.56 inches (242.9 mm). Note: It is the customer’s responsibility to provide a SIN-
Width: 30 inches (762.0 mm). GLE-PHASE SAMPLE for this system. The sample
CANNOT be a two-phase sample containing undis-
solved water and its associated compounds.

Area Classification
Class 1, Groups B, C, D Division 1. T6
ATEX: II2G EExdIIB+H2 T6

9.75 (247.7)
30.00 (762.0) 1.625 (41.3)

33.75 (857.3)

54.00 (1371.6)

1.00 (25.4) Typical 1.00 (25.4) Typical 8.75 (222.3) Specifications Subject To
Change Without Notice.
PPA4100 Series

The only on-line RVP analyzer Minimal operator interface.


designated as an ASTM Method.* Requires no calibration.
Simple, single-cell design… Small compact design.
virtually maintenance free.
Microprocessor-based.
Validates every analysis via
self-diagnostics.

Controller

Controller Housing
– Logic Board
– Power Supply Board
– Display Board
Hinged Door – Opens 180°

Analyzer

Explosion-Proof Housing
Measuring Cell
Sample Drain Solenoid

Sample System

The PPA4100’s fast loop liquid sample


system is comprised of the necessary
analytical components to control pres-
sure, filtration, and flow direction for
the N2 purge gas, the on-line sample,
the cell drain, the low pressure return,
and the standard sample introduction.

Nitrogen Inlet
Low Pressure Process Return
Sample Inlet
Syringe In Mounting Assembly

Fast Loop Return To Process

* ASTM Method D5482-93 –


When operated in off-line mode.
PPA4100 Series

Physical Property Analyzer PPA4100


On-line RVP Analyzer

Options
PPA4101 and PPA4102 SwirlKleanTM Filter Recovery Systems
The PPA4101, High Reid Vapor Available for streams that require The PPA4100’s design requires that
Pressure analyzer provides a removal of free water. (Optional) the cell drain be an unrestricted
measuring range of 0-90 psia. vent to atmosphere. This optional
Vortex Cooler recovery system provides the
The PPA4102, LPG analyzer provides Cools the measuring cell housing needed ventilation, automatically
a pressure range of 0-225 psia. when the ambient temperature monitors the level of spent sample
exceeds 90 °F. Includes the vortex and pumps the sample to the
Pressure Transducer Lookup cooling plate, an air pressure gauge, process return at up to 40 psi. The
Table temperature controlled air valve and system features a fiber optic
Enables new and replacement cell a manual shut-off valve. A minimum monitor, a one-gallon capacity, an
transducers to be linearized. It is a of 80 psia, at a flow rate of 12 scfm, explosion-proof pump, and an air
factory adjustment only. (Standard) allows the analyzer temperature to eductor to ensure that no sample
operate in a maxiumum ambient vapors escape the tank into the
Off-Line Diagnostics temperature of 104 °F. (Optional) operating area.
Requires approximately 11 minutes
to run. Checks for leaks into and out Serial Output
of the measuring cell. (Standard) Allows data transmission only from
RVP analyzer to customer’s serial
User Coefficients device. Output is an RS-232 format,
Span and zero correlation coeffi- with one start bit, eight data bits,
cients, that are determined by the one stop bit and no parity. Data is
user, can be used to alter the RVP transmitted as ASCII characters. Can
output to match another external support baud rates of 300, 2400 or
method. (Optional) 4800. Baud rates are factory set and
cannot be changed by the end user.
Diagnostic Test for N2 Purge (Optional)
Ensures that the N2 purge is within
specified limits. (Standard)

www.abb.com/analytical

ABB Inc. Analytical Products


843 N. Jefferson Street • Lewisburg, WV 24901 USA
Office: 304.647.4358 • Fax: 304.645.4236 © 2005 ABB Inc. BUASC4-1-405