XP Series PM Standard Procedure
Operation
Items to be checked Procedure
time
Go to [Menu] -> [Service] -> [Settings] -> [Print
Print Setting Value
Checking setting Settings] to print all setting values.
5 min.
values Go to [Menu] -> [Settings] -> [Host Output
HOST Output Setting OFF
Settings] to change "Connection" to "Disconnect".
Analyze normal sample or control material and
Data analysis Check background count level print average values or record them using stored
before 10 min. data/QC chart.
maintenance Check data using control material
Perform background check.
or samples
Go to [Menu] -> [Service] -> [Service Data]. Print
HGB background convert
out and check service data for the latest samples.
Check adjustment WBC clog level
values before RBC clog level
Go to [Menu] -> [Maint.] -> [Status Display] to
maintenance 0.05MPa
check pressure.
-0.0333MPa
Removing covers Remove front and side covers.
Disassemble and clean SRV Use CELLCLEAN and distilled water to perform
cleaning.
SRV Clean the tray 15 min. (Refer to Instructions for Use Chapter 12.)
Replace PIPETTE NO.57 using FITTING NO. 9-
Replace the aspiration pipette
A.
Use CELLCLEAN and distilled water to perform
Clean Rinse Cup 5 min. cleaning.
Rinse Cup (Refer to Instructions for Use Chapter 12.)
Replace Tube Pharmed for Waste Replace by using Tube Pharmed 1/4 IN X 1/8 IN.
5 min.
Line (Refer to Service Manual "Schematics".)
Replace Waste Chamber (Every 5
5 min. Replace WASTE CHAMBER ASSY NO.35.
years)
Disassemble and clean Waste Wipe the interior of the chamber and the float
Waste Chamber Chamber (when not replacing 5 min. switch using gauze impregnated with
Waste Chamber). CELLCLEAN.
Replace Waste Chamber Drain Replace silicon tube F739-8.
5 min.
Line silicon tubes. (Refer to Service Manual "Schematics".)
Wipe the unit and the needle tip using gauze
Disassemble and rinse Bellows Unit 5 min.
impregnated with ethanol.
Inject ethanol using an injection syringe to clean.
Bellows Unit If the tube is extremely dirty, replace the whole
Clean orifice tube 5 min. tube.
TUBE POLYURETHANE 1.8MMX3.4MM
(1400mm)
Rinse Trap Chamber Clean Trap Chamber using distilled water.
Replace Pneumatic Unit protection
Replace FILTER NO. 9.
Pneumatic Unit filter 30 min.
Replace by using KX-21 Pneumatic Unit (XP
Replace Compressor series).
(Refer to Service Manual Parts List.)
0.5kg/cm2(0.05MPa) Adjust pressure to 0.05Mpa using Regulator.
Pressure and 10 min.
Pressure adjustment (Refer to Instructions for Use Chapter 12.)
vacuum
250 mmHg (-0.0333MPa) Adjust vacuum to -0.0333MPa using the bellows
adjustment 5 min.
Vacuum adjustment unit. (Refer to Instructions for Use Chapter 12.)
Clean WBC transducer aperture Apply CELLCLEAN to a brush for clog removal
and clean the aperture. (Refer to Instructions for
Clean RBC transducer aperture Use Chapter 12.)
Transducer 10 min.
Remove dust, etc. and wipe with gauze
Clean inside of Transducer impregnated with ethanol. Then, wipe with dry
gauze.
(Version 2)
Operation
Items to be checked Procedure
time
Perform transducer cleaning Go to [Menu] -> [Maint.] -> [Clean Transducer] to
7 min.
HYDRAULIC sequence. perform the sequence.
UNIT Perform Waste Chamber Rinse Go to [Menu] -> [Maint.] -> [Clean W. Chamber] to
15 min.
Sequence perform the sequence.
Adjust VR 1 of PCB NO.20034/PCB. NO. 2135.
Adjust HGB background count level
(Refer to Service Manual "Adjustment".)
Other 5~ Adjust VR 6 of PCB NO.20034/PCB. NO. 2135.
Adjust WBC clog level
adjustments 10 min. (Refer to Service Manual "Adjustment".)
Adjust VR 7 of PCB NO.20034/PCB. NO. 2135.
Adjust RBC clog level
(Refer to Service Manual "Adjustment".)
Remove dust, etc. from inside of Main Unit.
Clean inside of Main Unit Use vacuum cleaner to remove dust, etc. from the
printer mechanism.
Main Unit, Interior 10 min.
Apply silicon grease to air cylinder shaft for
Lubricate mechanism driving SRV.
Apply BIRAL spray to linear slider for rinse cup.
Install instrument covers which were removed
Cover installation
previously.
Check background count level Check that the blank error has not occurred.
Check data 10 min. Analyze normal sample or control material (at
Check data using control material
least two times) and print average values or
or samples
record them using stored data/QC chart.
Remove dust, etc. and wipe with gauze
Clean Main Unit covers
impregnated with ethanol. Then, wipe with dry
Main Unit,
5 min. gauze.
Exterior Cleaning peripheral device (printer, Use vacuum cleaner to remove dust, etc. from the
etc.)
printer mechanism.
Go to [Menu] -> [Settings] -> [Host Output
Restore host output settings Settings] to restore "Connection" to the state
before maintenance is carried out.
Go to [Menu] -> [Maint.] -> [Status Display] to
Checking Check program version
check version.
performance after 15 min.
Go to [Menu] -> [Maint.] -> [Status Display] to
maintenance Cycle counts
check Total C.
Turn the power OFF and ON and start the
Final performance check instrument in the user mode to check sample
analysis is performed without errors.
* Operation time is an approximation and does not include removal of covers, etc.
ECR 312J003
(Version 2)
“Time for Replacement” is only a guideline and differs according to status of use. Time-based or cycled-based
maintenance selection should be determined by whichever comes first.