Christian Ullner
Federal Institute for Materials Research and Testing (BAM), Berlin, Germany
Abstract Based on the estimation of uncertainty the TABLE I. Requirements for the verification and calibration of
present requirements of the standard are discussed. testing machines (ISO 14577 part 2) and for the reference materials
Additional requirements needed for improving the reliability (ISO 14577 part 3).
and the reproducibility are proposed. Especially the Range Part 2 Part 3
compliance of the machine affects strongly the parameters
Tolerance
EIT and HMs in the macro range.
Force Macro 1.0% 0.25%
Force Micro 1.5% 0.5%
Keywords: Indentation, standard, uncertainty Maximum
permissible error
1. INTRODUCTION Displacement Macro 1% of h 0.5% of h or 30 nm
Displacement Micro 1% of h 1% of h or 5 nm
The ISO standard on the instrumented indentation test Apex angle Macro 0.3° 0.15°
for hardness and other materials parameter was published in Apex angle Micro 0.3° 0.15°
October 2002 [1]. Additional information on the history and Repeatability
the metrological problems of the standard are given in [2,3]. Hardness HM, HIT h>1µm 2% 2%
Appropriate reference materials needed for the verification Indentation
h>1µm 5% 5%
of the testing machine is the object of the contributions on modulus EIT
the micro range [4] and on the macro range [5]. Although
the ISO 14577 represents a great progress in the local To find further requirements needed for the best
elastic-plastic characterisation of materials some points for reliability and reproducibility of the instrumented
improving the standard have been appeared since the indentation test the uncertainty of the parameters calculated
publication. New ideas were given on the determination of from the indentation curve is estimated.
the contact point [6], on the force calibration [7], and on the
determination of the machine compliance [8]. Based on the 3. FUNCTIONAL RELATIONSHIPS
estimation of uncertainty in the instrumented indentation
testing [9] the present paper is aimed to improve the To start with, the materials parameters of the instrumented
verification and calibration of testing machines as well as indentation test are introduced. Martens hardness HM is
the calibration of reference materials. directly calculated from the measured test force F and
indentation depth h = he + hc
2. METROLOGICAL REQUIREMENTS F
HM (2)
The main requirements concern the uncertainty of the As (h)
test force F and the displacement h because the procedure As(h) is the surface of the indenter from the tip up to the
of the instrumented indentation test is primary aimed to the indentation depth h=he+hc, especially As(h)=G1*h2 for
measurement of the indentation curve (for instance Fig. 1). pyramids and h>6µm. Using the slope of the indentation
However, the raw data of the displacement are not the curve in the plot square root of force against depth the
indentation depth h. It is needed to determine the zero depth Martens hardness HMs reads
h0 (contact point) and to take into account the compliance .
of the machine Cm. The indentation depth h reads F 2
( ) (3)
HM s h
h h' h0 ' C m F (1)
G1
While the ISO 14577 defines an upper limit of the The initial unloading slope can also be used for separating
uncertainty of h0 (u(h0) < 0.01hmax) the requirement for the the elastic part he of the indentation depth h=hc+he. In this
uncertainty of the machine compliance Cm is missing. The way the indentation hardness HIT can be calculated as
requirements given in the standard are collected in Table I. follows
Fmax Fmax Inserting (12) and (13) into (6) the effect of compliance on
H IT with hc hmax (4) the resulting modulus Er is given instead of (6)
AP ( hc ) S
Ap(hc) is perpendicular to the test force and is named (CT C m )(h'max h0 (1 ) Fmax Cm
projected area at the plastic indentation depth hc. In analogy 2 E r G2
to As, Ap = G2*hc2 for pyramids and h>6µm. The constant * Fmax * CT ) (14)
is dependent on the geometry of the indenter ( =0.75 for
Vickers and Berkovich indenter).
The initial unloading slope S (stiffness)
4. SOURCES OF UNCERTAINTY
S ( dF
dh ) h hmax (5)
The macro range, F>2N, is the main object of this paper.
allows to calculate the elastic property. The indentation Some effective sources of uncertainty are illustrated in
modulu s EIT reads Figure 1. Because the indentation curve is a second order
2
polynomial in the macro range (geometrical similarity of the
1 (6) pyramidal indentation) the determination of the zero point h0
EIT s
2 (contact point) can lead to an important source of
1 1 i uncertainty [6, 11]. The compliance of the machine affects
Er Ei not only the indentation depth but also the initial unloading
with slope. Additional problems concerning the calculation of
the initial unloading slope (specific choose of fitting
S* (7) function and fitting range) should be noted.
Er 14
2 * hc * G2
Compliance of machine
12
Ei is the Youngs modulus of the indenter, i and s are the
Poisson ratio of the indenter and the specimen. 10 h max h max'
Three interrelations for calculating the uncertainty must
be added to (3), (4), and (7). The uncertainty of the initial 8
unloading slope is traced to the uncertainty of the slopes in
6
the calibration of force and displacement.
dh'
u ( dF
dh ) dF 2 dF dh 2 (8) 4 CT ( )F Fmax
( ) 2
(u ( )) (u ( St )) 2 (u ( )) dF
dF
dh dFSt dhSt dhSt 2 Zero point
Fst and hst are the standards in force and displacement. The 0
uncertainty of the slope d F/dh is traced to the uncertainty -2 0 2 4 6
h8 c 10 12 14 16
of dF/dh. Displacement
dh dh' Fig. 1. Additional sources of uncertainty for the materials
2 F (9) parameters of the instrumented indentation test
d F dF
Finally, in all equations the indentation depth h must be The sources of uncertainty needed for estimating the
inserted as a function of the raw displacement h after (1) to combined uncertainty of the materials parameters can be
take into account the correction by the machine compliance concluded from the equations of the last caption. As can be
Cm. Using the compliance C instead of the stiffness S=1/C seen the additional input values are the uncertainties of the
the derivates becomes slopes in the calibration of force and displacement for
dh dh' (10)
increasing and decreasing quantities as well as the
( ) ( ) Cm uncertainty of the compliance. The input values for the
dF dF calculation of the combined uncertainty are collected in
and Table II. For simplification, all distributions are assumed to
dh dh' be normally. For instance, the use of rectangular
2Cm F (11) distributions for the force and the displacement do not
d F d F
change the concluded proposals in caption 6.
The reciprocal initial unloading slope reads The uncertainty of the slopes of increasing force and
dh dh' decreasing force were determined by an extended
( )F ( )F C m CT Cm (12) verification of an available testing machine. Regarding the
Fmax Fmax
dF dF calibration of the displacement at the same machine the
and the plastic indentation depth hc after (7) reads results have agreed with the requirements of ISO 14577 part
3 (Table I). The determination of the machine compliance is
hc h'max h0 ' (1 ) Fmax C m Fmax CT (13) a separate topic of a paper [8]. As a result the chosen value
in Table II, Cm = 0.0005 µm/N, is the smallest one of the
realistic range. The uncertainty of the geometrical factors G1 (6). The more detailed description of the calculation is given
and G2 (for the indenter area functions As(h) and Ap(hc) of in [9].
the ideal Vickers pyramid, h>6µm) was evaluated by the The results of the combined uncertainty based on the
required uncertainty of the apex angle. The uncertainty of sources according to Table II is demonstrated as a function
in Table II represents the reliability of the used model for of the test force in Fig. 2 for the four materials parameters
separating the elastic part of the indentation. HM, HMs, HIT, and EIT. The jumps at 2 N are caused by the
The uncertainty of the contact point is formed by the different input values of the micro and macro range. Fig. 2
density of the data acquisition, the knowledge of the type of shows that the relative uncertainty of the materials
function for the extrapolation, and the fitting range [11]. For parameter is partially greater than the repeatability required
simplification, the uncertainty in Table II is the above limit by ISO 14577 parts 2 and 3 (Table I) although the
required in ISO 14577 part 1. experimental scatter has not been involved in the
To take into account the behaviour in the micro range the calculation. To reduce the uncertainty in the future the
different uncertainties of that range are also collected in knowledge of the individual contributions to the combined
Table II. The values for the force and the displacement are uncertainty of the materials parameter is of great interest.
given by the requirements of the standard (Table I). Because Obviously, the amount of the contributions are dependent on
the force generation is based on electromagnetic actors no the material, especially of the indentation hardness and the
hysteresis in the force measurement is assumed. Most of the indentation modulus.
machines for the nano and micro range are designed in such
a way that a larger correction of the machine compliance 7,0
(including frame compliance) is needed. Therefore a higher HM
uncertainty of the correction (about 5%) must be considered. 6,0 2N HMs
An important source of uncertainty is introduced by the 5,0
HIT
use and determination of the specific indenter area function EIT
needed for h<6µm. Dependent on the direct or indirect 4,0
procedure for the determination of the indenter area function
the uncertainty can vary strongly. Besides of the scattering 3,0
area, the real indentation process is more complex than it is 2,0
represented by the simplificated hardness definition, force
per area. For all of that reasons the twice uncertainty is 1,0
assumed for the micro range but a more detailed study
seems to be needed urgently. 0,0
0,1 1 10 100 1000
TABLE II. Input values (u absolute uncertainty, w relative Test force (N )
uncertainty) for the calculation of uncertainty (k=1, normal
distribution) Fig. 2. Relative combined uncertainty of the materials parameter
as a function of the test force for a material with HM=1000 N/mm2
Uncertainty Makro Mikro and EIT=200000 N/mm2.
w(force, increasing) % 0.25 0.5
w(force slope, increasing) % 0.15 0.3 Four representative materials according to Table III have
w(force slope, decreasing) % 0.3 0.3 been used. The materials are ordered by the increasing ratio
u(displacement, increasing) µm 0.03 0.005
of HIT/EIT. The curves in Fig. 2 are calculated for the
u(displacement slope, increasing) % 0.5 1
material like WC.
W(displacement slope, decreasing) % 0,5 1
u(Cm) µm/N 0,0005 0.005
TABLE III. Four typical combinations of Martens hardness
u(G1, G2) - 0.35 0.7
and indentation modulus in the order of HIT/EIT .
u( ) - 0.01 0.01
w(h0) % 1 1 HM N/mm2 1000 700 16000 4200
w(dF/dh) % 2 2 EIT N/mm2 200000 70000 400000 80000
w(Ei) % 10 10 HIT/EIT - 0.005 0.011 0.064 0.092
w( i) % 10 10 HIT/EIT µm/ N 0.005 0.013 0.013 0.034
w( s) % 5 5 Example
Fe Al WC Glass
material