MAMMOMAT Inspiration

Software
System
MammoUnit Firmware Installation

Answers for life.


2 Document Version / Disclaimer / Copyright

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed. Please contact your local Siemens
office to order a current version or refer to our website http://www.health‐
care.siemens.com.

Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens, 2013” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 2 of 18 © Siemens, 2013


06.15 CS SD XP
Table of Contents 3

1 General 4

1.1 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 General remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Component-specific remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.4 Conventions and abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Required aids and tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.3 Documents which may be helpful. . . . . . . . . . . . . . . . . . . . . . . . . 7

2 MammoUnit firmware 8

2.1 Backing up and restoring the MammoUnit settings . . . . . . . . . . . . . . . . . . . 8


2.1.1 Backing up the MammoUnit settings . . . . . . . . . . . . . . . . . . . . . . 8
2.1.2 Restoring the MammoUnit settings . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 Updating the MammoUnit firmware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

3 Changes to previous version 16

4 List of Hazard IDs 17

© Siemens, 2013 XPW7-330.816.04.02.02 Page 3 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
4 1 General

1.1 Notes
This document basically describes the firmware installation for the MammoUnit system
component.
Only qualified and system-trained service staff is allowed to perform MAMMOMAT Inspi‐
ration system installation, service, maintenance, and quality assurance. Ensure that the
most recent version of the service documentation is on hand.

1.1.1 Safety Information

 To avoid any risk of injury to persons and/or damage to the system read and observe the
General Safety Notes (TD00-000.860.01.xx.02).

 Adhere to the safety guidelines in the document “Mammography; Safety Information;


System; Safety Information including IEC 62353” (SP00-000.860.03.xx).

1.1.1.1 Icons

X Warning about ionizing radiation or radioactive substances. Tests and adjustments that
must be performed with the radiation switched on are indicated by this radiation warning
icon.

1.1.2 General remarks

NOTICE
The screen shots in this document are used for illustration and orientation purpo‐
ses.
 For this reason, the actual screen may show some differences.

NOTICE
The installation of software/firmware not released or validated for the system is
strictly prohibited.
 Otherwise, the syngo software or one or more of its applications may mal‐
function.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 4 of 18 © Siemens, 2013


06.15 CS SD XP
General 1 5

1.1.3 Component-specific remarks


Installing the firmware of the MammoUnit will be performed via the service software or
from the OS of the WH AWS.

1.1.3.1 Network connectivity information


Installing the MammoUnit firmware will be performed via the network connection WH
AWS - MammoUnit.
n Standard IP address configuration:

Component Default addresses of the spe‐


cific network adapter
MammoUnit 192.168.0.100
WH AWS 192.168.0.2

1.1.3.2 Condition
Before the firmware can be updated, the MammoUnit must be in good condition and
functioning properly, or must be brought into good condition.

1.1.3.3 Firmware versions information


MammoUnit firmware versions are only allowed in the combination with specific WH AWS
software versions. Refer to the latest version of the document “List of Compatible Parts,
System, Compatibility List, XPW7-330.840.01.xx.02.

1.1.3.4 Acceptance or Constancy Tests


If reinstallaing the same MammoUnit firmware, it is recommended to perform a constan‐
cy test prior and after the firmware update to verify that the Glandular dose and the mAs
values for a certain object , such as 40 mm PMMA for example, in mode AEC are still con‐
stant.
Performing the acceptance test of the MAMMOMAT Inspiration is recommended whenev‐
er changes that might affect performance, such as different MammoUnit firmware, have
been made to an existing system.
Tests from the Quality Control Manual have to be performed then.
Depending on the country where the system is installed, additional tests have to be per‐
formed.
n In Germany, tests specified by §16 of the X-ray Ordinance must be performed.
n In the USA, tests specified by DHHS regulations must be performed.

1.1.3.5 Passwords
The required passwords are not included in this document. You will find them in the CS
Knowledge Base on the intranet or you can obtain them from the relevant USC or RSC.

© Siemens, 2013 XPW7-330.816.04.02.02 Page 5 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
6 1 General

1.1.4 Conventions and abbreviations

Indicator Explanation
CBA Control board arm
CBS Control board stand
CS Customer Service
FW Firmware
HSC Headquarters Support Center
IP address Internet Protocol address
MU MammoUnit
OS Windows Operating System
RSC Regional Support Center
SW Software
TCP/IP Transmission Control Protocol / Internet Protocol
UI User interface
USC Uptime Service Center
WH AWS Acquisition workstation for MAMMOMAT Inspiration

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 6 of 18 © Siemens, 2013


06.15 CS SD XP
General 1 7

1.2 Requirements

1.2.1 Required aids and tools


No special tools or aids are required. Installing the MammoUnit firmware requires access
to the User Interface of the WH AWS.
- The MammoUnit is delivered on CD or can be found under C:\AWS\MU_Firmware.

1.2.2 Backup
n Although loading the MammoUnit firmware will not affect the configuration parame‐
ter or settings of the MammoUnit, performing a backup is recommended prior to in‐
stalling or updating the firmware. MammoUnit settings or parameters are stored on
CBS.

1.2.3 Documents which may be helpful

Document Print No.


General Safety information TD00-000.860.01.xx.02
User Instructions SPB7-330.620.01.xx.02
Configuration SPB7-330.843.02.xx.02
Quality Control SPB7-330.640.01.xx.02
Quality Control, USA SPB7-330.640.50.xx.24
Quality Assurance SPB7-330.820.01.xx.02

© Siemens, 2013 XPW7-330.816.04.02.02 Page 7 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
8 2 MammoUnit firmware

2.1 Backing up and restoring the MammoUnit settings

2.1.1 Backing up the MammoUnit settings

1. Log in to “Local Service” and select “Acquisition System / Web Service Interface”.
 The Web service interface opens.
2. From the main menu, select “Backup/Restore” and then scroll down and click “Back‐
up” in the window.
 A window pops up asking you to select the destination path of the backup.
Fig. 1: File download

3. Click “Save”.
 The “Choose File” box opens.
4. Select the destination path “C:\AWS\MammoUnit”, enter the file name in this format:
“MU_Config_YYYY_MM_DD.mcf” (for example, MU_Config_2008_06_09.mcf), and
then click “Save” to start the backup procedure.
 The folder C:\AWS\MammoUnit will be part of the WH AWS backup procedure.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 8 of 18 © Siemens, 2013


06.15 CS SD XP
MammoUnit firmware 2 9

Fig. 2: Configuration file name

5. Confirm the message box with “Close” when the file download is complete.
Fig. 3: Download complete

© Siemens, 2013 XPW7-330.816.04.02.02 Page 9 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
10 2 MammoUnit firmware

6. From the MammoUnit main menu, select “Set Time and IP Address” and then “Con‐
fig IP”.
Fig. 4: Web IP Config

7. Write down the network configuration data:

Address names Default Current value


IP config (IP address of the Mam‐ 192.168.0.100 _____________
moUnit)
Network mask config (subnet mask 255.255.255.0 _____________
of the MammoUnit)
MAC config (MAC address of the n.a. _____________
MammoUnit)
AWS IP config (IP address of the WH 192.168.0.2 _____________
AWS)
AWS MAC config (MAC address of n.a. _____________
the WH AWS)

8. Close the MammoUnit Web Service Interface.

2.1.2 Restoring the MammoUnit settings


Prerequisites:
A valid backup file (*.mcf) of the MammoUnit must be on hand, for example in “C:\AWS
\MammoUnit”.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 10 of 18 © Siemens, 2013


06.15 CS SD XP
MammoUnit firmware 2 11

Procedure:
1. Switch on the WH AWS.
 The MammoUnit stays off!
2. On the WH AWS, perform “Restart Application,” and then log in as administrator.
 Find the standard password in the knowledge base.

 Make sure to shut down the syngo application before beginning the MammoUnit config‐
uration restore procedure!

3. Switch on the MammoUnit using the ON button on the control box.


4. Open Internet Explorer and enter “192.168.0.100” in the URL line.
 The “Web Service Interface” opens.
 If the default IP address of the MammoUnit has been changed, enter the ap‐
propriate MammoUnit IP address into the URL line.
5. From the main menu, select “Backup/Restore.”
6. In the menu, click “Browse...”, select the valid MammoUnit backup file (the filename
format will be “MU_Config_YYYY_MM_DD.mcf”) in the directory C:\AWS\MammoUnit
(example location; the actual location may vary), and click “Open”.
7. Click “Restore” to transfer the backup data to the MammoUnit.
8. Confirm the message box and wait until the MammoUnit has rebooted - this will take
about one minute.
 Wait until the MammoUnit is up and running again.
9. Refresh the Web browser by pressing the “F5” function key, and then select “Back to
the Main Menu”.
10. Within the main menu of the Web Service Interface, select “Configuration data” and
make sure that all states are set to “valid”.
 If any parameters need to be changed, reboot the MammoUnit after making
these changes, and then check to make sure that all parameters are set to “val‐
id”.
11. Restart the complete system.

© Siemens, 2013 XPW7-330.816.04.02.02 Page 11 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
12 2 MammoUnit firmware

2.2 Updating the MammoUnit firmware


Installing the MammoUnit firmware can be performed from the Service UI (Acquisition
System/ MammoUnit/ Update Firmware) or from the OS (logging in as meduser or admin‐
istrator would be suitable).
Because it is more common, the following describes installing the MammoUnit firmware
as administrator.

 If an error occurs, try to download the firmware again.

1. From the UI log off the user and log in as administrator.


 Password required!
 The UI shuts down, access to the OS is granted.
2. Insert the CD with the corresponding flashmaster “VXxxx.exe” file into the DVD drive.
 Each system version of the MAMMOMAT Inspiration includes a specific Mam‐
moUnit firmware version in combination with a WH AWS software version.
Combinations others than those released are not allowed.
3. Open Windows Explorer and copy the flashmaster file from the CD into the folder C:
\AWS\MU_Firmware.
4. Double-click VXxxx.exe to run the flashmaster program from the path C:\AWS
\MU_Firmware.
 The program starts and a dialog is displayed on the computer monitor.
 The flashmaster program states the firmware version
 The corresponding IP address is displayed (default address is “192.168.0.100”)
Fig. 5: Flash master dialog

 The stand display shows the configured IP address, MammoUnit firmware ver‐
sion, and the MammoUnit status, in this example shown with firmware VA00D.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 12 of 18 © Siemens, 2013


06.15 CS SD XP
MammoUnit firmware 2 13

Fig. 6: MU - IP address

5. Click the “Check” button first.


 Wait until “Please press start” is highlighted in the dialog .
Fig. 7: "Check" button

6. When the MammoUnit is ready for download, click “Start” to initiate the firmware up‐
date for the CBS and CBA.
Fig. 8: Start button and version

 The update starts and a progress bar is displayed to indicate that the update is
“Preparing...”.
 The complete update time is 5 to 10 minutes.

© Siemens, 2013 XPW7-330.816.04.02.02 Page 13 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
14 2 MammoUnit firmware

Fig. 9: Status Preparing

Fig. 10: Status update

Fig. 11: Status update

7. When the update is finished, a dialog is displayed confirming that the update was suc‐
cessful. Click “OK” to close the program.
 The MammoUnit will restart itself once the update is completed.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 14 of 18 © Siemens, 2013


06.15 CS SD XP
MammoUnit firmware 2 15

Fig. 12: Update successfully finished

© Siemens, 2013 XPW7-330.816.04.02.02 Page 15 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
16 3 Changes to previous version

Backup/restore procedures added.

MAMMOMAT Inspiration XPW7-330.816.04.02.02 Page 16 of 18 © Siemens, 2013


06.15 CS SD XP
List of Hazard IDs 4 17

There are no Hazard IDs in this document.

© Siemens, 2013 XPW7-330.816.04.02.02 Page 17 of 18 MAMMOMAT Inspiration


06.15 CS SD XP
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

healthcare.siemens.com/services

Siemens Healthcare Headquarters


Siemens Healthcare GmbH
Henkestrasse 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens.com/healthcare

Print No.: XPW7-330.816.04.02.02 | Replaces: XPW7-330.816.04.01.02