Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Signature:
Specification: BS 718
Specification Series: M5OSP
Scale Range: 0.9 to 0.95 g/mlat 15 degC
Serial Number: 193513
Manufacture: Amarell Precision
Method Used
The calibration method used to determine the density of the hydrometer is based on the F.W. Cuckow
method, Ref: J. Soc. Chem. 1948,68 44-49, for the calibration of reference standard hydrometers. The
hydrometer was suspended from a hydrostatic balance and immersed into in a temperature controlled bath
containing a liquid of known density.
Ambient Gonditions
Temperature: 20.9'C
Humidity: 50.8 %
Barometric Pressure: 999.90 mbar
This certificate is issued in accordance with the laboratory accreditation requirements of the United Kingdom Accreditation Service.
It provides traceability to recognised national standards, and to units of measurement realised at the National Physical Laboratory or
other recognised national standards laboratories. This certificate may not be reproduced other than in full, except with the prior written
Certificate of Calibration
UKAS Accredited Calibration Laboratory No. 0513 Certificate Number: Sample
Measurements
Nominal Surface
" Value Correction Tension
g/mlat 15 degC g/mlat 15 degC mN/m
Notes
coolidence of apprcrmatly 95%, The unc€rtairty E aluation has been caried out in accodance with UKAS requirenents.