Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
N
Parameter Spesifikasi
o
1. Down Flow Velocity 0.25 – 0.45 m/s
2. In Flow Velocity 0.35 – 0.55 m/s
1. Upstream aerosol concentration must be
greater than 10 µg/L
2. The scanning rate shall be maintained at a
Filter Integrity Testing
3. speed approximately 5 cm/s, and at 1 inch
(Tes Integritas Filter)
below the filter area
3. The percentage of penetration shall not
exceed 0.01%
1. Average of light intensity when the lamp on ≥
Lighting Intensity 450 Lux
4.
(Intensitas Cahaya) 2. Average of background light intensity ≤ 150
Lux
Maximum permitted number of particles at rest
Particle Count condition are:
5.
(Bilangan Partikel) 1. ≥0.5 µ = NMT 3,500,000 particles/m3
2. ≥5 µ = NMT 20,000 particles/m3
Noise Level
6. The noise level should be ≤ 85 dB
(Tingkat Kebisingan)
Airflow Pattern by
7.
Smoke Test
Microbial contamination - Air Sampling: ≤100 CFU/m3
8. by air sampling and - Sedimentation Plate: ≤50 CFU/4h
sedimentation plate - Rodac Plate: ≤300 CFU/25m2
PT Aventis Pharma
Quality Operations Department
Testing 1
Procedure Appropriate
Test Objective
Equipment operates successfully as per the instruction in the SOP written for
operation
Test Description
1. Read thoroughly the SOP for operation of the equipment
2. Operate the equipment as per the instruction in the SOP for operation
3. Verify that the operation procedure is correct and match as routine
procedure
4. Examine the system based on the procedure and mark-up the procedure as
follows:
Yellow highlight : to confirm the appropriated procedure
Blue overwrite : to indicate the additional procedure/comments
Red strikethrough : to indicate the procedure is not appropriate with actual
procedure
5. Record the SOP name with number and effective date
Sign and date the SOP as “Verified in Qualification”, attach the marked up
document to this test sheet
Acceptance Criteria
The procedure is appropriate with actual procedure
Result
Testing 2
Preventive Maintenance System and Calibration
Test Objective
To ensure that the preventive maintance measures and calibration are entered in
e-MMS
Test Description
Check whether preventive maintance measures and calibration are entered in e-
MMS
Acceptance Criteria
The preventive maintance measures are entered in e-MMS
Result
Page 2 of 6
PT Aventis Pharma
Quality Operations Department
Testing 3
Downflow Velocity
Test Objective
To ensure the downflow velocity is met with the requirements
Test Equipment
Calibrated Anemometer
Test Description
1. Operate the BioSafety Cabinet as per SOP
2. Downflow is measured at 6 different locations (100 mm from the perimeter
wall) front, middle and back
3. Measure and record the downflow velocity using calibrated anemometer at the
location as per Appendix 1
Acceptance Criteria
The downflow velocity is met with the requirements (0.25 – 0.45 m/s)
Result
Page 3 of 6
PT Aventis Pharma
Quality Operations Department
Testing 5
Filter Integrity Test
Test Objective
To verify the absence of bypass leakage (between filter frame and ceiling grid
system) in the installation of the filters and verify that the installed filters are
free of defects and small leaks
Test Equipment
Page 4 of 6
PT Aventis Pharma
Quality Operations Department
11. The first scanning is at the edge of the heap filter then to the side of the
HEPA filter
12.Identify the suspected leaks and record. Retest is conducted after remedial
action has been taken and recorded. When filter area beyond repair and
replacement are made, a record will also be made to reflect the
replacement
Acceptance Criteria
1. Upstream aerosol concentration must be greater than 10 µg/L
2. The scanning rate shall be maintained at a speed approximately 5 cm/s, and
at 1 inch below the filter area
3. The percentage of penetration shall not exceed 0.01%
Result
Page 5 of 6
PT Aventis Pharma
Quality Operations Department
Page 6 of 6
PT Aventis Pharma
Quality Operations Department
Page 7 of 6
PT Aventis Pharma
Quality Operations Department
Page 8 of 6
PT Aventis Pharma
Quality Operations Department
Testing 7
Light Intensity
Test Objective
To ensure the lighting intensity is met with the requirements
Test Equipment
Calibrated light intensity meter (Lux meter)
Test Description
1. Press Light ON/OFF button
2. Measure and record the lighting intensity on background and light on using
calibrated light intensity meter at the location as per Appendix 3 (6 point)
Acceptance Criteria
3. Average of light intensity when the lamp on ≥ 450 Lux
4. Average of background light intensity ≤ 150 Lux
Result
Page 9 of 6
PT Aventis Pharma
Quality Operations Department
Testing 8
Noise Level
Test Objective
To ensure the Noise Level is met with the requirements
Test Equipment
Calibrated Sound Level Meter
Test Description
1. Operate biosafety cabinet as per SOP
2. Measure and record the noise using calibrated Sound Level Meter at left, right
and and center of the working desk area at the location as per appendix 4 (3
point)
Acceptance Criteria
The noise level should be ≤ 85 dBA
Result
Page 10 of 6
PT Aventis Pharma
Quality Operations Department
Testing 9
Particle Count for 0.5 µ and 5 µ at Rest Condition
Test Objective
To Ensure the particle count for 0.5 µ and 5 µ are meet with specification at
Rest Condition
Test Equipment
Calibrated Air Borne Particle Counter
Test Description
Acceptance Criteria
Maximum permitted number of particles at rest condition are:
3. ≥0.5 µ = NMT 3,500,000 particles/m3
4. ≥5 µ = NMT 20,000 particles/m3
Result
Page 11 of 6
PT Aventis Pharma
Quality Operations Department
Comment
Page 12 of 6
PT Aventis Pharma
Quality Operations Department
Test Objective
Test Equipment
Tryptic Soya Agar (TSA) setting plate, and Calibrated MAS 100 Air Sampler
Test Description
A. By Air Sampling
1. Petridish containing TSA put in the MAS 100 Air Sampler
2. Run the MAS 100 Air Sampler for 10 minutes (1 m3)
3. Take out the petridish from MAS 100 Air Sampler
4. Incubate the petridish in incubator at 30-35°C for 3-5 days
5. Calculate the colonies
6. Record and attach the result
B. By Sedimentation Plate
1. Open Petridish containing TSA for 3 hours
2. Close the petridish
3. Incubate the petridish in incubator at 30-35°C for 3-5 days
4. Calculate the colonies
5. Record and attach the result
C. By Rodac Plate
1. Open Petridish containing TSA and Polysorbate 80%
2. Placed the petridish on the surface and slightly press for approximately 2
days
3. Remove the plate and shut immediately
4. Clean the examined surface carefully and disinfect
5. Incubate the petridish in incubator at 30-35°C for 3-5 days and then 2 days at
20-25°C
6. Calculate the colonies
7. Record and attach the result
Acceptance Criteria
Page 13 of 6
PT Aventis Pharma
Quality Operations Department
Comment
Verified by :____________________________________________
Page 14 of 6
PT CS2 Pola Sehat
Quality Assurance Section
Appendix 1
Downflow Velocity
Acceptance Criteria:
Downflow Velocity : 0.25 – 0.45 m/s
5 6
Back
3 4
Middle
1 2
Front
Verified by :____________________________________________
Appendix 2
Inflow Velocity
Acceptance Criteria:
Page 1 of 6
PT Aventis Pharma
Quality Operations Department
2 5 8
1 11
1 1
3 6 9
1 1 1
1 4 10
7
1 1 1
Verified by :____________________________________________
Appendix 3
HEPA Integrity
Page 2 of 6
PT Aventis Pharma
Quality Operations Department
Page 3 of 6
PT Aventis Pharma
Quality Operations Department
Acceptance Criteria:
Average of light intensity when the lamp on ≥ 450 Lux
Average of background light intensity ≤ 150 Lux
Verified by :____________________________________________
Page 4 of 6
PT Aventis Pharma
Quality Operations Department
Appendix 4
Noise Level
Acceptance Criteria:
The noise level should be ≤ 85 dBA
Page 5 of 6
PT Aventis Pharma
Quality Operations Department
Verified by :____________________________________________
Page 6 of 6