Inspection
425e | C01 Page 1 of 5
Prepared: DH | 11.10.2006
Tested: RM | 17.08.2010
Version 3.0
Client:
BioSealer® type:
Serial no:
Production date:
Technician
Name:
Signature:
Date:
Pos. 18 One tube of ordered type and wall-thickness are available. Tube Type and Size: Yes No
(this tube is used for the OQ too).
Comment
– Unit must operate automatically without the need for operator intervention.
– Three test seals of one ordered hose type and size type must be visually acceptable with no holes, bubbles or tears.
Tube type (brand) and inner|outer diameter as per Pos. 19 (page 4).
– Counter function is working. (See the counter status before and after the OQ).
Seal no: Successfully passed - visual check Counter functions work correctly
Confirm with
yes or no
Confirm with
yes or no
Pos. 9 QC conformity sticker is mounted in front of device (initials from Technical Engineer and date). Yes No
Comment