Sei sulla pagina 1di 48

Odel s.r.l.

Via Lavoratori Autobianchi, 1 lotto 24/B


20832 Desio (MB) – ITALIA

Tel. (+39) 0362.30.10.64


Fax (+39) 0362.15.41.13.3
http://www.Odel
Odelxray.eu
info@Odelxray.eu
xray.eu
customer@OdelOdelxray.eu

Service Manual
R306.4x "Program H.F. US" mk3
series
40kW - 50kW – 65kW – 80kW - 100kW
n. 1044 ver ENG
DEFINITIONS IN THIS MANUAL

0051

While every care is taken to ensure the description and illustration accuracy in this publication, factory
reserves itself the right to make any change in apparatus design considered to be in the interest of our
customers, without notice it.

1
DEFINITIONS IN THIS MANUAL

1.1 Is defined MANUFACTURER here and in all the documentation the following
figure:

Odel s.r.l.
Via Lavoratori Autobianchi,1 Lotto24B
20832 Desio (MB) - Italia

1.2 Is defined address of SERVICE QUALITY MANUFACTURER here


and in all the documentation the following figure:

qualita@Odelxray.eu

1.3 Is defined address of CUSTOMER SERVICE MANUFACTURER


here and in all the documentation the following figure:

customer@Odelxray.eu

DEFINITIONS IN THIS MANUAL

2
PROPRIETARY STATEMENTS
The subject matter and contents of this document, together with any drawings therein, are
proprietary information of Odel X-ray and are not to be used other than according to the policy of
corporation and are not to be copied without prior authorization.
Equipment operation in conformity to sector normation is subjected to correct application of all the
guidelines contined in this manual.

RADIATION PRECAUTION
X-rays are hazardous to both operator and other unless radiation protection
procedure are esthabilished and strictly observed.
Odel X-ray recomends to follow these procedures:
 Only trained people should operate this unit
 Read and follow the instructions in the Operator Manual
 Read and follow all labels and the precaution statements affixed to the unit.
 Follow Your facilitie's Radiation Protection
 Never expose people if not necessary
 Keep all exposure to ioniziong radiation as reasonably achievable
 If any doubts would come, consult Your supervisor or other qualified expert
The above must not take place or supersede any procedure or precaution
which may already be established.

INSTALLATION PRECAUTION
Upon installation, a radiation protection survey must be made by a qualified expert according to the
most current protection law version.
A survey must be performed after each change in equipment, work load or operating conditions
which may increase the probability of persons receiving more than the maximum permissible dose.
Manufacturer is not responsable of wrong or missing application of guidelines contained in this
manual.
Manufacturer qualifies technician who can install and keep maintenance on the equipment.
Manufacturer is not responsable for installation or maintenance performed by unqualified
technician.
Manufactured is not rensponsable on the effects produced by interfacing the equipment to device
not included in the compatibility list contained in this manual.
Each modification on internal circuit must be analyzed and approved by manufacturer.

IMPORTANT NOTICE
The safety of both patient and operator was a major consideration in the design and manufacturing
of this equipment. The equipment will function reliably when operated, serviced, and repaired
according to the instructions in this manual.
PROPRIETARY STATEMENTS

Misuse, however, could result in hazard to patients, operator and/or equipment.


Odel X-ray cannot assume responsibility for any malfunction of this equipment resulting from
improper operation, installation, service, or repair, or from damage or modification of its
components.
The diagnostic facility and/or end user are responsible for performing all periodic checks described
in this manual.
Odel X-ray disclaims all responsability from any injury resulting from misuse of this equipment.

3
MECHANICAL SAFETY
All moveable assemblies and parts of the unit should be carefully operated and rutinely
inspected according to the Operator and Mainteinance manual.
Use appropriate counter balance weights during assembly and disassembly.
Follow all other user manteinance statements in this manual.

ELECTRICAL SAFETY
Only properly trained and qualified personel should have access to any internal components
Live electrical terminals are hazardous. Be sure that the line disconnect switch is opened and
other appropriate precautions are taken before opening access door and removing enclosure
panels
Do not remove the high tension cable from the X-ray tube housing or high tension generator
and/or the access covers from the generator until the main and auxiliary power supplies have
been disconnected
Electrical grounding must comply with the national electrical code and any local requirem
ents as necessary.
Follow all other mainteinance statements in this manual failure to comply with the above
may result in hazards to operator, patients, and/or equipment.

PROPRIETARY STATEMENTS

4
Ratings
LOADING FACTORS, INTERMITTENT MODE
 Model 40kW Type R306.44
 Model 50kW Type R306.45
Type And Model  Model 65kW Type R306.46
 Model 80kW Type R306.47
 Model 100kW Type R306.48
 Model 40kW 40kW
 Model 50kW 50kW
Generator Output Power  Model 65kW 65kW
 Model 80kW 80kW
 Model 100kW 100kW
TYPE B CLASS 1 CEI EN 60601-2-54:2011
Equipment Type and Class
GROUP 1 CLASS A CEI EN 60601-1-2:2010
IEC 60601-1-2:2007(III Ed.)
IEC 60601-1:2012
IEC 60601-2-54:2009 + A1:2015
Conform to IEC 60601-1-6:2010 (III Ed.) + A1:2013
IEC 62304:2006 + A1:2015
IEC 62366:2007+A1:2014
93/42/CEE Medical Device
Europe Directive Annex IX regola 10 punto 3.2
Date on Manufacturing see Rapporto di collaudo "Test Report"
Date of Test see Rapporto di collaudo "Test Report"
Original Manual Language Italian
Protection Index IPX0

POWER SUPPLY
Nominal Supply Voltage Current 400V~ (three-phase)
Supply Frequerncy 50/60 Hz
 Model 40kW 400V~ (three-phase) / 92Apk
 Model 50kW 400V~ (three-phase) / 108Apk
Nominal Supply Voltage  Model 65kW 400V~ (three-phase) / 141Apk
 Model 80kW 400V~ (three-phase) / 173Apk
 Model 100kW 400V~ (three-phase) / 207Apk
 Model 40kW Apparent 60 kVA Active 54 kW
 Model 50kW Apparent 75 kVA Active 63 kW
Mains Input Power
 Model 65kW Apparent 98 kVA Active 82 kW
Ratings

 Model 80kW Apparent 120 kVA Active 100 kW

5
 Model 100kW Apparent 150 kVA Active 125 kW
Nominal - 10% + 10%
Supply Tollerance Voltage

 Model 40kW 0,30 Ohm


 Model 50kW 0,20 Ohm
Max Supply Impedance  Model 65kW 0,15 Ohm
 Model 80kW 0,13 Ohm
 Model 100kW 0,10 Ohm

RATINGS ON INTERMITTENT MODE


Nominal Tube Voltage 150kV
 Model 40kW 500mA
 Model 50kW 630mA
Maximum X-ray Tube Current  Model 65kW 800mA
 Model 80kW 1000mA
 Model 100kW 1000mA
 Model 40kW 500mA @ 80kV – 400mA @ 100kV
 Model 50kW 630mA @ 80kV – 500mA @ 100kV
X-ray Tube Voltage and Current
 Model 65kW 800mA @ 80kV – 630mA @ 100kV
for highest electrical output
 Model 80kW 1000mA @ 80kV – 800mA @ 100kV
 Model 100kW 1000mA @ 100kV
 Model 40kW 100kV, 400mA, 0,1s
 Model 50kW 100kV, 500mA, 0,1s
Nominal X-ray Output Power  Model 65kW 100kV, 630mA, 0,1s
 Model 80kW 100kV, 800mA, 0,1s
 Model 100kW 100kV, 1000mA, 0,1s
Loading factors for High Voltage
From 40kV to 150kV in steps of 1kV
set by the user
From 10mA the maximum Current of High Voltage on scale
R10’
10mA, 13mA, 16mA, 20mA, 25mA, 32mA, 40mA, 50mA,
Loading factors for Current high 63mA, 80mA, 100mA, 125mA, 160mA, 200mA, 250mA,
voltage set by the user 320mA, 400mA, 500mA (tutti i mOdelli)
630mA (50kW, 65kW, 80kW, 100kW)
800mA (65kW, 80kW, 100kW)
1000mA (80kW, 100kW)

Loading factors for Time of From 0,001s to 6s on scale R10’ in compliance with the
Irradiation set by the user minimum product of High Voltage * Time of irradiation
Minimum Nominal time of
0,001s
irradiation
Ratings

Loading factors Product Current From 0,4mAs to 600mAs on scale R10’

6
High Voltage * Time of
irradiation
Minimum Product Current High
0,4mAs – 200mA, 0,002s
Voltage * Time of irradiation
Optional on all mOdels
Device Auto Exposure
Semiconductors chamber or ionization

RATINGS CONTINUOUS MODE


Nominal Tube Voltage 120kV
Highest X-ray Tube Current 8mA
Note: these ratings are for the X-ray High Voltage generator and are reduced by the ratings of the X-ray Tube when more
limiting.

TEMPERATURE REQUIREMENTS and HUMIDITY


ENVIRONMENT INSTALLATION
TEMPERATURE HUMIDITY PRECAUTIONS
No condensation
Control 10°C ~ 40°C 30% ~ 75%
should take place
No condensation
Rack 10°C ~ 40°C 30% ~ 75%
should take place
Note: when these specifications are not met, equipment functions and rating can be not guaranteed.

OPERATING ENVIRONMENT
TEMPERATURE HUMIDITY PRESSURE
Control 10°C ~ 40°C 0% ~ 75% 700hPA ~ 1060hPA
Rack 10°C ~ 40°C 0% ~ 75% 700hPA ~ 1060hPA

STORAGE CONDITIONS
TEMPERATURE HUMIDITY PRESSURE
Control 40°C ~ 70°C 10% ~ 95% 500hPA ~ 1060hPA
Rack 70°C ~ 70°C 10% ~ 95% 500hPA ~ 1060hPA

HEAT GENERATION
QUANTITY OF HEAT GENERATED
Control Approx 95kcal/h
Rack Approx 930kcal/h for standard use
Ventilation Absent, not necessary for standard use
Note: Contact the manufacturer in case of intensive use for more information
Ratings

7
EXTERNAL SAFETY DEVICES
Requirements
Device Three-pole differential switch
Type B
Load 63°
Protection 30mA
Requirements IEC328, 400V

DEFINITION OF X-RAY PARAMETERS


High voltage
Voltage value after the initial transient
Expressed in kV
Current High Voltage
Average value of current flowing between the cathode and anode during the Time of Irradiation
Expressed in mA
Time of Irradiation
Time when the high voltage (between anode and cathode) exceeds 75% of the kV peak
Expressed in s
Current product High Voltage * Irradiation time
In mAs

LOADING FACTORS MEASURING METODOLOGY


The generator has been tested in laboratories IMQ in compliance with the general and particular
industry, test instruments used are:
DYNALIZER III, is a system for measuring high voltage invasive that are in accordance with the
methodology of measurement required by IEC 601-2-7, for more information contact:
RadCal Corporation
426 west duarte road
Monrovia - California 91016 USA
818-357-7921
ASPN, measuring system of AIR KERMA with room 500 cc3.
BARRACUDA (RTI) measurement system SEMICONDUCTORS.
PROPORTIONAL DETECTOR (VACUTEC) is a system of detection dosimetry with ionization
technology.
Ratings

8
The acceptance test of the equipment at the manufacturer are carried out with the following
equipment, in test certificate "Rapporto di Collaudo" lists the serial tools and traceability to the
measurement system certificate
the measurement is carried out
DYNALIZER III in parallel to the circuit under
High Voltage Measurement + high voltage through a divider
DIGITAL DISPLAY and a high-voltage amplifier
configuration instrumental
the measurement is carried out
DYNALIZER III in series to the circuit of High
Measurement of the
+ Voltage on a measuring
Current High Voltage
DIGITAL DISPLAY resistor and an amplifier in
instrumental configuration
the measurement is carried out
in parallel to the circuit under
high voltage through a divider
DYNALIZER III and a high-voltage amplifier
Measurement of the
+ configuration instrumental, it
Time of Irradiation
DIGITAL DISPLAY is considered as irradiation
time exceeding the threshold
placed at 75% of the peak
value of High Voltage
The value of the dose depends
on the radiant circuit
introduced by the
manufacturer to run the tests
will acceptance of the system.
The measures are not given on
PROPORTIONAL DETECTOR certificates of inspection
+ depending on the irradiation
Measurement of the
ASPN device present at the
Dose Ionizing Radiation
+ laboratories of the
BARRACUDA manufacturer.
These measurements are made
to verify the continuity of the
system of measurement of the
primary (electric)

The value of Dose per Unit of


Area Radiated tests performed
Measurement of the during testing of the system
Chamber DAP (VACUTEC)
Dose per unit area Radiated validates the measurement
circuit through an ionization
chamber certified
Ratings

9
ACCURACY and PRECISION of the RADIOGRAPHIC
PARAMETERS
RANGE
ACCURACY
PARAMETER MEASURING UNIT RANGE CONTROL
DEFINED BY THE
STANDARD
High Voltage kV 40kV ~ 150kV ±5%
10mA ~ max mA
see table
RATINGS
High Voltage Current mA INTERMITTENT ±10%
MODE for definition of
maximum High
Voltage Current.
Irradiation Time S 0,001s ~ 6s ±1% + 1ms
0,4mAs ~ 600mAs
Current product High values outside this
Voltage * Time of mAs range for special ±10%
irradiation processes available out
of the standard

Ratings

10
LIQUID PRESENT IN THE SYSTEM
For maintenance checks ref.
Section description
High Voltage Transformer Shell Diala DX
Generator High Voltage
Transformer
Note for maintenance service: every 12 months check the oil level, if more than 30 mm from the top
of the cover to the oil level is found, refill in order to have 20 mm clearance only.

PRECAUTIONS
STORAGE TRANSPORTATION
Use a dry place, not subjected to do not turn during
great temperature changes, in order transportation, do
to avoid condensation; temperature not shake and deal it with
Transformer great care
range must be not less than 10 and
in order to avoid any oil
not more than 60 degrees. leakage
Use a dry place, not subjected to do not shake and deal it
great temperature changes, in order whit great
Power Rack to avoid condensation; temperature care, very fragile electronic
range must be not less than 10 and
not more than 60 degrees. components inside
Use a dry place, not subjected to
do not shake and deal it
great temperature changes, in order
whit great
Control Console to avoid condensation; temperature
care, very fragile electronic
range must be not less than 10 and
components inside
not more than 60 degrees.
Note: if what advised in this table is not attended, correct working of the unit cannot be guaranteed.

Ratings

11
ENVIRONMENTAL RISKS and RIDDING OF COMPONENTS
The Unit does not present for the utilized components no environmental risks.
This unit cannot be operated in presence of anesthetic mixture inflammable with air or
oxygen or nitrogen protoxide.
As the HV transformer is containing oil (specially formulated by SHELL for electrical
transformer) utilization and safety advice's reported on the A appendix of this manual must
be strictly observed do not waste oil in the environment, please attend to the local requirement for
getting rid of oil.
Please follow these rules to get rid of the different items:
for the transformer
  make the transformer empty of oil; divide the plastic and metallic parts from those oil
impregnated; the impregnated parts are the HT capacitor and the primary and HT
secondary winding; the primary and secondary can be in they turn dismounted up to
reaching the plastic components as the bobbins etc., metallic, as copper and iron, present
in the winding and in the core, the capacitor must be taken away from the plastic box
where it is contained.
  when separating the plastic and metallic parts from oil or from the oil impregnated parts,
you can proceed to deliver them to the exhaust oil receiving Companies enclosing the
relative form concerning the oil characteristics and annexed to the A appendix of this
manual.
  metallic and/or plastic parts must be rid as scrap iron or plastic.
  deliver the HT diodes stacks and the printed circuit placed on the cover of the HT
Transformer to specialized companies for getting rid of metal and plastic items
for the generator
  divide the metallic parts from: printed circuits, plastic parts, single components with huge
dimension as electrolytic and/or 600 Vdc film capacitors, IGBT power semiconductors and
three-phase bridge, line and breaker switches, fuses and transformers

ENVIRONMENTAL RISKS and RIDDING OF COMPONENTS


  get rid of the iron parts as iron scrap
  get rid of the plastic parts as plastic scrap
  the power semiconductors, line switches, huge dimension capacitors and transformers
must be released to specialized companies authorized to collect the iron/plastic material
with respect to the law in force.
  the printed circuits must be released to specialized companies authorized to collect them
with respect to the law in force.
Procedures above mentioned are only an advice to get rid of the generator manufactured by
Odel s.r.l., they cannot be intended as a substitution of rules and laws in force on the
territory but only as their complement, it is compulsory then to respect at first the law
in force in the Country and/or Region wher the prduct must be discarted.
Odel s.r.l. Company does not take any charge or responsibility of the getting rid of the
product discarded from its customer or of the products present into the market.

12
DECLARATION OF COMPLIANCE WITH THE
DIRECTIVE ON THE RESTRICTION OF THE USE OF
HAZARDOUS SUBSTANCES IN ELECTRICAL AND
ELECTRONIC EQUIPMENT (ROHS)
We declare that Odel s.r.l. production process and Odel in factory products are made under
compliance with the Directive on the restriction of the use of hazardous substances in electrical and
electronic equipment (Italian law no. 151 of 25 July 2005) since, pursuant to article 5, comma 2,
point a), referring to attachment 1A, category 8.

ENVIRONMENTAL RISKS and RIDDING OF COMPONENTS

13
RISKS ARRIVING FROM THE USE OF THE
EQUIPMENT
Responsibility Terms : the manufacturer ( Odel s.r.l.) does not take in any charge whatever
damage to persons or things due to a use different than the one the unit has been designed
for, or to a wrong and not perfect installation of the unit.
This is a HT generator that coupled to a X-ray tube, not manufactured by Odel s.r.l.,
can, under the operatively conditions described in the user manual, can produce X-ray
radiation with power and time previously set from the control console.

Warning: the correct use of the generator is user own responsibility, Odel S.r.l.
gives only the necessary information for its correct use and relative to the
dangers associated to the use of a High Tension generator, but does not take
any responsibility for the operations and safety and diagnostic techniques
adopted after the sale.
Odel S.r.l. does not take any responsibility for over exposures of users and/or
patients due to wrong procedures or operating techniques, and also if the
generator has not been installed and serviced in accordance to the procedures
described in this manual or for defects due to modification made by the
installer.

This equipment is sold knowing that the manufacturer and its representing agent do not take
any responsibility for over- exposure of patients or personnel.
There is any design technique that can make safe the operator or the service personnel who
take not attention to the safety rules. Only authorized people especially trained is in
position to work whit X-ray generators. The personnel of the technical service must be
conscious of the dangers associated to the servicing of the H T units and to the high X-ray
doses.

X-ray exposures can damage the health under cumulative effects that can be shown after
several mounts or years. Service and hospital personnel utilizing these equipments
must avoid with any care exposure to the primary beam and to be conscious of the
secondary beam and of the scattering radiation. The secondary beam is coming from
the contact of the primary beam with whatever thing and/or person being within the primary
beam and film.

RISKS ARRIVING FROM THE USE OF THE EQUIPMENT


Warning: do not over pass the data of the X-ray tube written on its own technical manual
. The life of the unit is guaranteed only if the generator is utilized respecting its technical
characteristics.
The generators must be utilized respecting the information contained in the technical
sheet

14
COMPATIBILITY LIST
The list above is a list of tubes tested in the laboratory of the manufacturer, however, the generator
can be interfaced, thanks to its versatility and precision in the measurement and control of the basic
parameters, tubes not expressly contained; the only requirements that are required are: X-ray tube
with rotating anode, bipolar, with a maximum of two burners with maximum current of ignition not
more than 8 amperes.
This genenerator was designed to respect its rating data while operating with a large fleet of X-ray
tubes and X-ray accessories. You can connect the following tubes
VARIAN
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
COMET
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
EUREKA
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
PHILIPS
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
IAE
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
CGR
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
DUNLEE
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
SIEMENS
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
TOSHIBA
All bifocal tubes from 0.3 to 1.2mm for small focus and from 0.6 to 2mm from
large focus with maximum filament current at 8A
COMPATIBILITY LIST

15
CHAMBERS FOR CONTROL OF EXPOSURE(AEC)
COMET
Semiconductor Chamber
ZIEHM
Semiconductor Chamber
GILARDONI
Semiconductor Chamber
AID
Ionization Chamber
VACUTEC
Ionization Chamber
MEDYS
Ionization Chamber

ACCESSORIES RADIOLOGICAL OF EXAMINATION


The generator can be interfaced to the usual accessories that are most commonly found on the
market, from remote controlled, the bucky horizontal and vertical as well as the interface to
scanners with optocouplers to 24 Vdc or 230 Vac.
Specifically they have been tried, and the manufacturer of the generator can ensure proper
operation, the tables examination of the following manufacturers (in alphabetical order):
ARCOM
CGR
IBIS
IMAGO
EUROCOLUMBUS
FUJI
KERMAT
MECALL
MT
PHILIPS
SIAS
VARIAN SIMULATORS
VILLA S.M.
Manufacturers may also provide advice for the installation of its generators on any examination
table
COMPATIBILITY LIST

16
Classic TV systems or with digital imaging
To the generator you can interface whatever type of TV Chain, this is due to the fact that it
can accept different signals arriving from the TV chain. For further informations about the
interfacing see the CONFIGURATION chapter.
Specifically the products made by the following manufacturers (listed in alphabetic order)
have been already tested and we can grant the correct working:
ALPHA TECHNOLOGIES
ATS
EIGEN
FIDELITY
INFIMED
NICAL
PHILIPS
PRECISE OPTICS
STALLION TECHNOLOGIES
THALES.

The manufacturer can also give assistance to the installation of our generators with whatever
other type of TV system or digital equipment.

Notes on the inflammability of the generator


The inflammable components contained into the generator can be:
Oil, to have information about the inflammability point of the oil contained in the HT
Transformer see at the appendix of this manual
  Cables, all the cables contained into the generator are NO 7V-K CEI 20-22 II type

  Printed circuits, the laminated material utilized for their production correspond to the FR 4

V0 grade of the NEMA/L1 Standard


  Transformers, they are all protected against the short circuit.

USE OF THE EQUIPMENT


The unit has been designed to work with a continuous or intermittent load, the power under
load is in function of the X-ray tube load and of the thermal load cumulated during the
working.
COMPATIBILITY LIST

17
EQUIPMENT INSTALLATION
Important Notice
THIS X-RAY EQUIPMENT ADOPTS AN INVERTER SYSTEM, ITS OSCILLATION FREQUENCY IS
AN AUDIO FREQUENCY RANGE OF SEVERAL Hz TO SEVERAL kHz AND NOISES ARE EMITTED
FROM THE CABINET DURING THE X-RAY EXPOSURE.
AS THE INVERTER OSCILLATES AT HIGH FREQUENCY, ELECTROMAGNETIC NOISES CAN BE
PRODUCED IN THE X-RAY MONITOR SYSTEM. THESE NOISES MAY BE PICKED UP BY A
CAMERA CABLE AND THEN APPEAR ON THE MONITOS. TO AVOID IT USE SOME PRECAUTION.
BEFORE THE INSTALLATION OF THE X-RAY POWER CABINET FOLLOW THESE RULES :
  MAKE SURE THAT THE INSTALLATION ROOM IS FAR FROM POWER PLANTS OR POWER

MACHINERY OR MECHANICAL MACHINERY.(AIR CONDITIONNING SYSTEMS,


ELECTROSTATICS FILTERS, LIQUID PUMP STATIONS, ETC)
  DURING THE ROOOM DESIGN PLACE THE POWER REMOTE TILTING TABLE RACK AS FAR

AS POSSIBLE FROM THE X-RAY GENERATOR.


  MAKE SURE THAT ALL THE CABLES OF THE TV CHAIN HAVE DIFFERENT PATH THAN THE

HV GENERATOR ONES AND THAT NO ONE SERVICE OF THE TV CHAIN SHOULD BE FED
(POWERED) WITH THE SUPPLIES PRESENT INTO GENERATOR. IN FACT THEY CAN BE
SUBJECTED TO THE INVERTER NOISES.
  POSITIONING OF THE POWER RACK IS VERY IMPORTANT TO MAKE THE SERVICE EASY.

THEREFORE CABINET MUST BE MOVABLE FROM THE WALL OF ABOUT 60 CM ALL AROUND
EVERY SIDE AND ALL LATERAL PARTS MUST BE QUICKLY AND EASILY ACCESSIBLE.

18
ELECTROMAGNETIC COMPATIBILITY
The equipment complies with the standards applicable to the product with regard to electromagnetic
compatibility EN 60601-1-2: 2007.
The equipment must be installed and put into service according to EMC information provided in
this section.
The equipment may be affected by portable and mobile communications equipment.
The equipment can't be provided with cables other than those specified by the manufacturer even if
sold as replacement parts for internal components, as this condition may' result in increased
emissions or decreased immunity of the EMC.
The apparatus should not be used near or stacked with other equipment in order to avoid
interference in the normal operating condition.
It does not therefore risks arising from the coupling of the electronic device with other electronic
devices, verification conducted within the limits dictated by the standard.
Test Report performed by IMQ testing laboratory.
Annex A (Table 1)

The [EQUIPMENT Series R306.4x] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT Series R306.4x] should assure that it is used in
such an environment.
Emissions test Compliance Electromagnetic Environment
RF emissions Group 1 This [EQUIPMENT Series R306.4x] uses RF energy only
for its internal function. Therefore, the RF emission is very low
CISPR 11 and not likely to cause any interference in nearby electronic
equipment.
RF emissions Class [A] This [EQUIPMENT Series R306.4x] is suitable for use in
CISPR 11 all environments other than domestic and those directly
connected to mains power to the public low-voltage which
Harmonic emissions [Not supplies buildings used for domestic purposes.
Applicable]
EN 61000-3-2
Voltage fluctuations/ [Not
flicker emissions Applicable]
EN 61000-3-3
CISPR 14 [Not Not Applicable
Applicable]
ELECTROMAGNETIC COMPATIBILITY

CISPR 15 [Not Not Applicable


Applicable]

19
Annex B (Table 2)

The [EQUIPMENT Series R306.4x] is intended for use in the electromagnetic environment specified
below. The customer or user of the [EQUIPMENT Series R306.4x] should assure that it is used in an
electromagnetic environment.
Immunity EN 60601-1-2 Compliance Electromagnetic Environment -
Test Test level level guidance

Electrostatic ± 6 kV contact EN 60601-1-2 Floors must be wood, concrete or


discharge (ESD) ± 8 kV air Test level ceramic tile. If floors are covered with
EN 61000-4-2 synthetic material, the relative
humidity must be at least 30 %.
Electrical fast ± 2 kV for power EN 60601-1-2 Mains power quality must be that of a
transient/burst supply lines Test level typical commercial or hospital
environment.
EN 61000-4-4
± 1 kV for input/output
lines > 3 m
Surge ± 1 kV differential mode EN 60601-1-2 Mains power quality must be that of a
EN 61000-4-5 Test level typical commercial or hospital
environment.
± 2 kV common mode
Voltage dips, short <5 % UT for 0,5 cycle EN 60601-1-2 Mains power quality must be that of a
interruptions (>95 % dip in UT) Test level typical commercial or hospital
and voltage environment.
40 % UT for 5 cycles
variations on power
supply input lines (60 % dip in UT)
EN 61000-4-11 70 % UT for 25 cycles
(30 % dip in UT)
<5 % UT for 5 s
(>95 % dip in UT)
Power frequency 3 A/m EN 60601-1-2 Power frequency magnetic fields
(50/60 Hz) magnetic Test level must be at levels characteristic of a
field typical location in a typical
ELECTROMAGNETIC COMPATIBILITY

EN 61000-4-8 commercial or hospital environment.

20
Annex C (Table 4)

non-LIFE SUPPORTING EQUIPMENT


The [EQUIPMENT Series R306.4x] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT Series R306.4x] should assure that it is used in
such an environment
Immunity EN 60601-1-2 Compliance Electromagnetic Environment
Test Test level level
Portable and mobile RF communications
3 V/m
Radiated RF 3 V/m equipment should be used no closer to any part of
EN 60601-1-2 the [EQUIPMENT OR SYSTEM], including
EN 61000-4-3 80 MHz to 2.5 GHz
Test level cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter
Recommended separation distance
d  1.2  P 80 MHz to 800MHz
d  2.3  P 800 MHz to 2.5GHz
Where P is the maximum output rating of the
transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended
3V separation distance in meters (m).
Conducted RF 3V
EN 60601-1-2 Field strengths for fixed RF transmitter, as
EN 61000-4-6 150 kHz to 80 MHz
Test level determined by an electromagnetic site survey,
should be less then the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structure, objects and
people.
non-LIFE SUPPORTING EQUIPMENT

a
Fields strengths from fixed transmitters, such as base stations for radio telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters , an
electromagnetic site survey should be considered. If the measured field strength outside the shielded location in which the [EQUIPMENT Series
R306.4x] is used exceeds [field strength] V/m, observe the [EQUIPMENT Series R306.4x] to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as relocating the [EQUIPMENT Series R306.4x] or using a shielded location with a higher RF
shielding effectiveness and filter attenuation
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

21
Annex D (Table 6)

Recommended Separation Distance for


non-LIFE SUPPORTING EQUIPMENT
The [EQUIPMENT Series R306.4x] is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the [EQUIPMENT Series R306.4x]
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitter) and the [EQUIPMENT Series R306.4x] as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output Separation distance according to frequency of transmitter
power of the
(m)
transmitter
(W) 80MHz to 800MHz 800MHz to 2.5GHz
d  1.2  P d  2.3  P
0.01 0.12 0.23
0.1 0.38 0.73
1 1.2 2.3
10 3.8 7.3

Recommended Separation Distance for non-LIFE SUPPORTING EQUIPMENT


100 12 23

For transmitters rated at the maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
(1) 800MHz, the separation distance for the higher frequency range applies
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people

22
MAINTENANCE - FAULTS
To ensure patient safety, the operator and other parts that might come into contact with the
equipment, we recommend performing a preventive maintenance program which consists of the
annual review of the main features of the generator, most service functions incidental.
More frequent checks will be required if the equipment works in conditions not specified in the
tables supplied by the manufacturer.
The tests to be performed are of two categories:
Functional tests, that can be executed by the operator
Technical service, which must be performed by specialized personnel, equipped with appropriate
instrumentation.
As a manufacturer and installer declining to liability on the operation of the equipment, if:
Installation, operation extensions, calibration, repairs are not performed by qualified personnel by
the manufacturer.
items related to safety circuits have not been modified and / or replaced with those in the spare part
list supplied by the manufacturer.
the electrical installation of the room where the equipment present, does not comply with the
current law and regulations and the IEC 601-2-7
misuse and / or abuse of the equipment

FUNCTIONAL TESTS
Before any test that requires the emission of ionization radiation, the wearing of the protections
including devices for personal dosimetry.

CONTROL AUTOMATIC EXPOSURE


Close the collimator. If you need to cover it with towels sealed. Perform a display by holding down
the rays; make sure the device before the AEC is active.
The yellow handwriting should be illuminated during the whole time of writing.
Warning! not consider the alarm at the end of handwriting, this indicates that the device is working
properly. Your handwriting is locked instantly, and This former correct.
Open the collimator.
Perform exposure. In this case the indicator of the radiation will have to remain switched on for a
time equal to 100 ms.

CONTROL DEVICE FOR AUTOMATIC ADJUSTMENT OF DOSE

Close the collimator. If you need to cover it.


Do one fluorine fluoroscopy device with ABS inserted. The yellow indicator fluoroscopy should be
illuminated during the whole time of exposure.
Warning! the dose, as reported on the appropriate value kV-mA display, should go up to the
maximum value.
MAINTENANCE - FAULTS

Stop fluoroscopy. Open the collimator.


Perform a new fluoroscopy. In this case, the indicator of the radiation will have to remain switched
on and the dose down to the lower value. Release the control of fluoroscopy.

23
CLEANING
The equipment must be turned off before any cleaning action.
Using clean cotton cloth and non-abrasive detergents. Do not use organic solvents or agents
containing solvents of any kind.
Do not spray the cleaner must not enter under any circumstances inside
of the apparatus.
To disinfect surfaces we recommend the use of aqueous solutions based on aldehyde.
Is recommended that no agents particularly aggressive based alcohol.

DAILY TEST
Before beginning the tests perform a visual inspection of the machine, to report visible damage.
During tests check that the yellow-ray fluoroscopy and handwriting are working properly.

WEEKLY TESTS
Check the integrity of cables and electrical connections.
Check oil leaks or unusual noises coming from the generator

MONTHLY TEST
A.E.C. and A.B.S. test, listed above.

TECHNICAL SERVICE
A correct service program follows those items
Inspection frequency Inspection activity
Radiological parameters inspection:
› High Voltage peak value
› Current*Time product accuracy
Every 6 month or every time a
component is substituted › In case Automatic Exposimeter verify integrated dose and
repeatibility
Every inspection required by country law where equipment is
installed
Verify functionality and clean High Voltage headings and put
Every 12 month
fresh insulating grease
Every 5 years Replace NVRam.
A correct and regular service program requires:
Integrity of cables, insulations, shields specially for live parts
Check for every bolt fixed with strenght.
Check the correct workflow of the generator (as explained in previous chapter).
MAINTENANCE - FAULTS

Leakage currents measure.


Heart resistance measure on every accessible metallic part.
Clean dust on every horizontal part inside the rack: dust, moisture and salted atmosphere, such as in
seaside locations can reduce ground insulation
Check, clean and use fresh insulating grease on High Voltage Cables headings

24
Check and Clean, wherever possible, relay and tele-controlled contactor’s contacts
Check dielectric oil level and transparency in High Voltage Transformer
Check alarms in alarm log
Inspect fuses integrity
Inspect also with an external meter, radiographical parameters and doses
Inspect tube calibration and all the test tah cvan grand the correct exposition

Cables and connections integrity


Switch off the gnerator, open the rack and visual inspect cables, than test with appropriate tools the
strenght of the connections:

Cables
Visual inpsects cables for breaks, loose connections and insulation integrity.

Prited Circuits Boards


Check for connectors insertions, remove and re-insert every connector to remove the possible
contects oxidation.
Do the same on every board, including the control console connection board.

Leakage Currents measure


Use appropriate tool to measure hearth leakage currents: refer to I.E.C. 601 -2 – 7 regulation.

Dust clean
Switch off the generator
Open the rack, use a dry brush to clean dust on surfaces, do not u s e l i q ui ds o n l i v e pa rt s , do
not clean when live mains is insert.
Clean also vertical PCBs.
Clean also the inner parts of the control console.

High Voltage connection clean


Switch off the generator
Wait 1 minute
Remove one H.V. cable by time
Make the three pin touch a grounded metallic surface
Inspect cable heading and pins to find black lines, smell also the cable to locate possible burn smells
Clean the headings
If necessary open the pins
Verify insulation between pins and ground
Replace the coumpound with new dielectric grease
Insert the heading and fix down using the gears screws
MAINTENANCE - FAULTS

Do it for every heading, also on tube side.

Contacts cleaning
Switch off the generator an all the generator’s accessories
Open the rack
Locate the relays, open the cap and inspect the contacts, in case clean using contacts clean spray.

25
Close the cap and repeat this procedure on every openable relay.
Open the frontal side of the big line contactor, inspect and clean the contacts: if seriously damaged
replace the contacts.

Dielectric oil inspection


Switch off the generator and open the rack.
Check oil level removing the black enclosure on the high voltage transformer.
Oilo level should be 1..1.5cm from the top, depending on oil temperature.
Sniff the oil to determinate if an acrid smell is inside the transformer.
Check with a visual inspection of the bottom of the cabinet to verify the absence of oil.
If oil is located outside of the transformer, check transformer to locate the point of origin.
In the event that a leak was localized at a critical point contact the manufacturer for provisions
about.
Clean dirty oil and, if necessary, top up oil. The oil usable for topping up must be treated with
microfilteringi and drying:
Use only oil supplied by the manufacturer.
Close the generator.

Alarm Check
Visual Inspection
On each card there are green LEDs that display the status of the power supply ALWAYS
PRESENT, check that the green LEDs are all switched on.
The forms are the yellow LEDs that display the status of the signals. Check through the table
attached to this manual, the LEDs on the status of the generator.
On some cards are red LEDs that display the status of selected alarm system, verify that the red
LEDs are switched off.

Electronics inspection
Turn off the generator.
Open the cabinet of the generator.
Chech cards inverter and Main Computer
Turn on the generator.
Warning!
The presence in the inverters of the electrolytic capacitors of high capacity, each time the generator
is switched on without the side walls, are avoided the contacts, an appropriate distance is
maintained between the terminals of the capacitors and the own body, or any thing that is not
isolated is handling.
Before touching the terminals to make sure that the power is disconnected and there is no voltage
on the capacitors.
MAINTENANCE - FAULTS

Open the front panel;


Connect the service program and enter into diagnostics.
Check for alarms stored.

Radiological Parameters inspection


Make an initial exposure to 80 kV, 100 mA, 0,50s.,

26
No alarm should trigger in this situation.
To check the accuracy of radiographic parameters is necessary to verify that the calibration of these
tubes is correct, while the mA depend on the type of tube, and then the calibration of the data
HardWare SoftWare but also by the setting of the tube.
kV peak are independent the type of tube installed, the parameters to be checked are: kV peak, mA
average spelling, handwriting time, mAs.
As regards the kV is possible to calibrate the measured value, the same applies to the mA, the time
does not require calibration of any kind, being controlled by a quartz, the mAs must be corrected if
the mA are within tolerance.
The adjustments of the circuits HW measurement of kVp and mA are performed in the factory
during the final testing phase; the standard instrument used for this work is a tool produced by
INVASIVE RadCal Corp. and called Dynalyzer; it measures kVp, mA and time.
It is advisable, therefore, in the case in which is necessary a calibration of the part HW, to use this
type of tool or equivalent.
For the tests of accuracy and repeatability follow tables and following procedures:
Connect instruments for the measurement of kV and mA;
It should connect external devices to the generator for the measure, if it were not possible to connect
the oscilloscope to signal kV feedback on the Main Computer.
An usage misuse of these test points or the introduction of noise on this signal can lead to serious
consequences and a malfunction of the power circuit of the generator, mARadio on its test Point
depending on the desired rate.
It is clear that in this case it is only possible to control the waveforms of the X-ray parameters, it is
not possible to calibrate the channels HW relying on themselves, because it lacks the standard
instrument.
Turn on the generator and bring it into operational mode

ACCURACY TEST
An example of periodic tests is represented by the ratio of final testing issued with the generator.
Repeat exposures indicated in values and ways on the test report and verify that the results beds are
included in the range of acceptability for the parameters under consideration.
The deviation of the real parameters than the set must not be higher than 5% for kV and to 10% for
mA.
If they are experiencing considerable differences in kV, before retouching control potentiometer kV
contact the manufacturer.
If you are experiencing considerable differences in mA, before proceeding to a new setting, check
with the service program that the filaments are properly fed, the high voltage cables have pin heads
thoroughly clean and wide, and the tube filaments are turned on .

kV Accuracy
If the kV value is not within the limits given in the table you must make the appropriate settings on
the potentiometers and gain feedback on the card Computer Main.
MAINTENANCE - FAULTS

To adjust the value of real kV must act on the potentiometer, checking the answer through an
external kVoltmeter.
To adjust the value of kV read from the main computer and displayed on the cabinet, you must act
on the feedback potentiometer reading, looking at the real value of kV as per the previous point, and
the value read through the service program .

27
mA Accuracy
If the value of mA is not within the limits given in the table must be rebuilt calibrate the focus.
This is done through the service software.
If the uncalibrated issue is an interval inspect also the system remembering the way the geenrator
uses to measure and control the current, as described below:
There are three channels for the measurement of mA:
icl, channel fluoroscopy with high resolution and saturation channel 10mA
icm, channel X-ray with high resolution and saturation channel 150mA
ich, channel radiography with medium resolution and saturation channel to 1500mA

Exposure Time Accuracy


The tests are to confirm the accuracy of time measurement in fact there are no adjustments to be
done on this parameter, in case of doubt, verify that your instrument reference is used in the correct
manner, and marks the time to 75% of the value kVpk.

mAs Accuracy
Since the mAs are the product of mA and time, and being the time is not adjustable, the only
adjustment is that of the mA already seen previously.

CALIBRATION
Since the emission characteristics in terms of quality of the ionizing radiation degrades significantly
with the wear of the filaments of the X-ray tube, the emission curves of the generator should be re-
calibrated every year.
To do this, follow the instructions below in the description of the calibration process of the focuses
through the service program of the generator.

MAINTENANCE - FAULTS

28
INSTALLATION REPORT
MAILTO: [address of the service quality 'manufacturer marked on the front page]

DATE: __/__/____
SUBJECT: INSTALLATION REPORT
ATTENTION: QUALITY SERVICE

COMPLIANCE TO DIRECTIVE 93/42 / EEC IN THE FIELD OF RESPONSIBILITY OF


THE MANUFACTURER REQUIRES THE KNOWLEDGE OF AVAILABILITY FINAL
PRODUCT.
PLEASE COMPLETE THIS FORM AND THE MANUFACTURER forward AFTER
PROPER INSTALLATION OF ANY PRODUCT.
LOCATION
Place Dept.

Address

City

Country ZIP

LOCAL ASSISTANCE
Name Family name

Company

Tel. Mobile

Fax. E-mail

SYSTEM
Model
INSTALLATION REPORT

Power

Serial Number

29
REFERRING TO TECHNICAL MANUAL
Visual inspection on reception:
________________________________________________________________________________
________________________________________________________________________________
The installation site is designed according to current standards.  Yes  No
Note :
________________________________________________________________________________
________________________________________________________________________________
Note on packaging and shipping:
________________________________________________________________________________
________________________________________________________________________________

HIGH VOLTAGE TRANSFORMER INSPECTION


Oil level satisfy the limits 
External Connections cable has at least 1 meter of abundance 
Ground cable is connected 
kV/mA signal flat cable is connected (trasformatore di Alta Tensione a Computer Principale) 
Focus Supply cables are connected correctly 

MAINS CONNECTION
Line Power meets the specifications of the technical beginning of this manual 
The cables for connection of Food meet the standard requirements for facilities videnti 
A line adaptor is required YES  NO 

COLLEGAMENTO CAVI DI ALTA TENSIONE


In the case in which the high voltage cables are not supplied by the manufacturer of the generator
Cables High Voltage meet the CE / MDD 
The terminals of the high voltage cables comply with the interference receptacles 
The screens of the cables of high voltage are grounded by the
High voltage transformer (only) 

ROTATING ANODE’S STARTER CONNECTION


The Power Cables of the stators are connected according to the installation instructions 
In the case where the pipes are not supplied by the manufacturer of the generator
The X-ray tubes meet the CE / MDD industry 
The X-ray tubes are present in the list of compatibility YES  NO 

ACCESSORY CONNECTION
The Potter Bucky are connected as indicated for installation 
In the case where the Potter Bucky are not supplied by the manufacturer of the generator
INSTALLATION REPORT

Potter Bucky meet the CE / MDD industry 


Potter Bucky are present in the list of compatibility YES  NO 

HANDSWITCHES and FOOTSWITCHES


Installation includes external Commands Radiography YES  NO 

30
External radiography commands meet the standard requirements 
In the case where the External Command are not supplied by the manufacturer of the generator
External Command meet the CE / MDD industry 
External Command are present in the list of compatibility YES  NO 

EXTERNAL FLUOROSCOPY FOOTSWITCH


Installation includes external Fluoroscopy Footswitch YES  NO 
Fluoroscopy Footswitch meet the standard requirements 
In the case where the Fluoroscopy Footswitch are not supplied by the manufacturer of the generator
Fluoroscopy Footswitch meet the CE / MDD industry 
Fluoroscopy Footswitch are present in the list of compatibility YES  NO 

EXTERNAL ACCESSORIES
Installation includes external Accessories YES  NO 
external Accessories meet the standard requirements 
In the case where the external Accessories are not supplied by the manufacturer of the generator
external Accessories meet the CE / MDD industry 
external Accessories are present in the list of compatibility YES  NO 

MAINS REQUIREMENTS
Mains is protected (Fuses/Magnetic thermal switches) 
In case of single phase accessory not supplied by generator’s manufacturer:
power Supply and intserfaceare insulated by relays 
External Systems satisfies CE / MDD regulation 

GENERATOR CONFIGURATION
GCF file stored in a safe place or sent to manufacturer e-mail address 

-------- END OF TRANSMISSION --------

INSTALLATION REPORT

31
CRITICAL FAILURE COMMUNICATION
MAILTO: [address of the service quality 'manufacturer marked on the front page]

DATE: __/__/____
SUBJECT: INSTALLATION REPORT
ATTENTION: QUALITY SERVICE

COMPLIANCE 'DIRECTIVE 93/42 / EEC REQUIRES THE KNOWLEDGE BY THE


MANUFACTURER OF CASES FOR WHICH THE PRODUCT HAS CAUSED THE DEATH OR
SERIOUS DETERIORATION IN THE STATE OF HEALTH OF THE PATIENT OR A USER.
PLEASE COMPLETE THIS FORM AND THE MANUFACTURER forward OCCUR IF ONE OF
THE CASES MENTIONED ABOVE.

SYSTEM and INSTALLATION DATA


Place Dept.
Address
City
Country ZIP

LOCAL SERVICE

Name Family name


Company
Tel. Mobile
Fax. E-mail

SYSTEM DATA

Model
Power
Serial Number

FAILURE DESCRIPTION AND CONSEQUENCE CRITICAL FAILURE COMMUNICATION

-------- END OF TRANSMISSION --------

32
DOSE MEASURE
DOSE*AREA PRODUCT
Configuration
DAP chambers detectors are interfaced through RS422 (1 pulse per uGy*cm2 or 1pulse per
mGy*m2)
Interface is autosensing the presence of the detector.
Measure items follows:
 Measure chamber on tube side, inserted in collimetor rails
 Connection cable connected between chamber to main computer inside the generator
 An optional TM-U220B printer.
Each measurement chamber leaves the factory with a calibration coefficient (number between 0500
and 2000) who introduced the main computer through the program dis erviziocalibra the weight of
each pulse to measure the room campion made in the laboratories of the manufacturer.
The label provides the printer will be on adhesive paper and then glued on the radiograph.
The information contained in it are:
 The type of work or FLUORO PULSED or RADIO or CINE
 The tube in use T1 or T2 or T3
 the date DD / MM / YY
 hour HH: MM: SS
 the value of kV
 the value of mA
 the time of handwriting or the sum of the times of handwriting in the same preparation in
millisecond
 the dose expressed in mGray per square centimeter
 the total time of the RUN or fluoroscopy in seconds
 the number of exposures of handwriting or pulsed fluoroscopyKERMA

During the installation of the generator you can activate the pre / post indication of kerma for
intermittent and pulsed regimes.
When the 6 parameters of the kerma dose measurements are non-zero, the control panel enables the
display of these parameters.
The geometric parameterization and the presence of an additional filter occurs through the
measures, statically, in the phase of installation through the service program.
DOSE MEASURE

33
TECHNICAL DESCRIPTION
GENERATOR’S REGULATION LOOP
The logic diagram of adjustment is shown schematically in the figure below, in which the blocks
represent the logical division of the central and peripheral functions, control and power:
 Inverter interfacing with the line, where there are safety switches
 Three-phase bridge, capacitors, and inductors to power the DC-RAIL and allow the transfer
of power to the inverter.
 Power Inverter
 High Voltage Transformer
 Distribution of power generator
 Main Computer and Automatic Control of Exposure
 Terminal general interfacing with the outside.
 Filament Inverter service power and drive anode

HIGH VOLTAGE REGULATION LOOP


TECHNICAL DESCRIPTION

The regulation loop of kV is completely harware in a mixed analog/digital process, this means that
the uP does not directly enter into such an adjustment, what it knows is the real value of kV, and the
status of the generator thanks to digital signals that arrive from the inverter.
When it is requesting the issuance of ionizing radiation predisposes the main computer peripheral
signals to put the tube in a position to deliver the desired parameters (Control Safeties, rotating
anode drive, pre-ignition Fire selected.

34
Then prepares the signals that enable the inverter oscillation [NONx], then checks that the feedback
signal of the presence of voltage on the inverter is active [IPW] and the signal showing the state of
balance of the branches of the bridge is active [IUB].
Is then set to the reference value of kV, analog signal (OKV +).
The activation of the inverter start signals initiates the charging of high voltage capacitors inside the
High Voltage transformer.
Such charging is limited in current by a ramp of about one millisecond, the Start inverter to the
point of exceeding the 75% of the set voltage.
Here is a digital signal generator of the presence of the High Voltage [IRM].
From here microprocessor start counting the length of time of irradiation

HIGH VOLTAGE CURRENT REGULATION LOOP


The regulation loop of mA is completely controlled by the main computer through the modulation
ignition of the filament chosen, this means that in every moment of the issue, intermittent or
continuous, the computer knows the value of emission mA.
The main computer controls the emission value by changing the value of lighting the filament, if the
issue is in the range of acceptance of the parameters, otherwise generates error interrupting the
issue.
Therefore, the magnitudes involved in this control are:
• measurement circuit (ich), high current handwriting, (icm), current media
handwriting, and (IFL), current fluoroscopy; This signal is taken from the high
voltage transformer and led to the analog section of the main computer, here is
measured by a circuit and amplified (ich): 10 Vdc = 1500 mA; (Icm): 10 Vdc = 150
mA; (Icl): 10V = 10mA)
• The circuit generates the DAC reference to lighting the fire selected to the inverter
power filaments
• The inverter board ignition filaments that through the transformers of the filament,
isolated, produces ring closed, the current necessary for the correct emission.
To know the characteristic of emission of a tube, uP needs the CALIBRATION TUBE, in this
manner can calculate the value at which the filament must turn on for the pair kV-mA chosen by the
operator, for from issuance correct.
The uP needs this value only for the departure rays and this value should be as close as possible to
the value of mA chosen, after a few usecondi departure from the uP begins to read the mA emission
and consequently adjust the ignition.
If the value of mA ignition was too far away from the chosen one (25% for the low values, 12.5%
for high values), the main computer stops the spelling and sounds the alarm, because it decided that
it was no longer possible to adjust get the correct value of the issue and then get a good risultato.ù
You can have this warning for two reasons:
• tube uncalibrated
• incorrect reading of the mA output, something is not working in the loop
measurement mA or lighting adjustments
TECHNICAL DESCRIPTION

35
TECHNICAL SHEETS
DIELECTRIC TRANSFORMER OIL
As provided by producer, Also available on produce site: http://www.shellitalia.it)

TECHNICAL SHEETS

36
TECHNICAL SHEETS

37
TECHNICAL SHEETS

38
TECHNICAL SHEETS

39
TECHNICAL SHEETS

40
TECHNICAL SHEETS

41
TECHNICAL SHEETS

42
TECHNICAL SHEETS

43
TECHNICAL SHEETS

44
TECHNICAL SHEETS

45
Symbol Description
Electrical

3Ph~ Alternate Current, Three-phase system

I System ON

O System OFF

L1, L2, L3 Mains Phases

High Voltage Hazard

Do not touch when mains is applied

Ground Protection

general

European Community markage

Type B equipment under IEC60601 regulation

Warning: read annexed documentation

Intermittent mode emission

Continuous mode emission

X-RAY emission

Other symbols used in User manual are described in User Manual


Symbol Description

46
Labelling
Power rack label

MANUFACTURER’s DATA

SYSTEM MODEL

SYSTEM SERIAL NUMBER

PRODUCTION DATE

EUROPEAN COMMUNITY COMPLIANCE LABEL


SYMBOLS
Labelling

47
MAINS REQUIREMENT

CONTINUOUS LOAD

MAXIMUM INTERMITTENT LOAD WHEN EMISSION IS


REQUIRED

OEM DATA

Labelling

48

Potrebbero piacerti anche