Balance & Weights

Q. 1. What are the parameters for calibration of analytical


balance?
A.1. Analytical Balance is calibrated for following five parameters:-
(a) Accuracy, (b) Linearity, (c) Uncertainty, (d) Eccentricity & (e)
Repeatability
Q. 2. What is the frequency of calibration for analytical balance?
A. 2. Analytical Balance is calibrated in following frequency:-
(a) Internal calibration Daily (If Balance is Digital)
(b) Accuracy Daily
(c) Linearity Monthly
(d) Uncertainty Monthly
(e) Eccentricity Monthly
(f) Repeatability Monthly
Q. 3. How the Accuracy of analytical balance is calibrated & what
are the acceptance criteria?
A. 3. The Accuracy of Analytical Balance is calibrated by taking weight of
50 mg, 100 mg, 500 mg, 1 gm, 10 gm, 100 gm, 110 gm, & 175 gm of
Certified Weights.
The acceptance criteria: ±0.2% of Certified Weights. (Rer: In-house but
IP says 0.1%)
Observed Weight Tolerance (g) (± 0.2 %
Weights (g)
of weight)
0.0500 0.0499 - 0.0501
0.1000 0.0980 – 0.1020
0.5000 0.4990 - 0.5010
1.0000 0.9980 - 1.0020
10.0000 9.9800 - 10.0200
110.0000 109.7800 - 110.2200
175.0000 174.6500 - 175.3500

Q. 4. How the Linearity of analytical balance is calibrated & what


are the acceptance criteria?
A. 4. The Linearity of Analytical Balance is calibrated by taking weight of
500 mg, 1 gm, 10 gm, 110 gm, & 175 gm of Certified Weights and plot the
graph between standard value and weights observed.
The acceptance criteria: The Graph shall be linear & Coefficient of
correlation: NLT 0.99.
Q. 5. How the Uncertainty of analytical balance is calibrated &
what are the acceptance criteria?
A. 5. The Uncertainty of Analytical Balance is calibrated by taking weight
of 500 mg of Certified Weights 10 times. Calculate the average and
standard deviation of all 10 measurements.
Replicates Observed weight (g)
1
2
3
4
5
6
7
8
9
10
Average
SD
3X Standard Deviation (SD)
Uncertainty = ------------------------------------- Acceptance criteria: NMT
0.001% (Ref: USP)
Value of Certified Weights
Q. 6. How the Eccentricity of analytical balance is calibrated &
what are the acceptance criteria?
A. 6. The Eccentricity of Analytical Balance is calibrated by taking weight
of 100 gm of Certified at five different place of weighing pan (centre &
edge of cross section of diameter).
Position Observed Deviation from centre
weight (g)
Centre of the pan (A) (A)
Top right corner (B) (B-A) x 1000 =
Bottom right corner ( (C-A) x 1000 =
C)
Bottom Left corner (D-A) x 1000 =
(D)
Top left corner (E) (E-A) x 1000 =
Acceptance criteria: Observed weight from each position should not
deviate from the weight observed at the centre of the pan by ± 0.2 mg.
Q. 7. How the Repeatability of analytical balance is calibrated &
what are the acceptance criteria?
A. 7. The Repeatability of Analytical Balance is calibrated by taking weight
of 100 gm of Certified Weights 10 times & calculates the minimum and
maximum weight deviated from the average weight.
S No Observed weight (g)
1
2
3
4
5
6
7
8
9
10
Average ( x)
Minimum weight deviated from average weight (Y) =
Deviation (mg) = (X-Y) x 1000 =
Maximum weight deviated from average weight (Z) =
Deviation (mg) = (X- Z) x 1000 =
Acceptance criteria: Observed weight should not be more than ± 0.2
mg from the average weight. (Ref: USP)
Q. 8. What are the parameters for calibration of single pan
balance?
A.8. Single pan Balance is calibrated for following four parameters:-
(a) Linearity A (b) Linearity B (c) Drift & (d) Uncertainty
Q. 9. What is the frequency of calibration for single pan balance?
A. 9. Single pan Balance is calibrated in following frequency:-
(a) Linearity A Daily (Accuracy)
(b) Linearity B Monthly
(c) Drift Monthly (Eccentricity)
(d) Uncertainty Monthly
Q. 10. How the Linearity A (Accuracy) of single pan balance is
calibrated & what are the acceptance criteria?
A. 10. The Linearity A (Accuracy) of Single Pan Balance is calibrated by
taking weight of (i) 100 times of least count, (ii) 5%, (iii) 20%, (iv) 50%
and (v) 80% of maximum capacity of balance by using Certified Weights.
The acceptance criteria: ±0.1% of Certified Weights. (Ref: IP)
Capacity Least Standard weight &acceptance criteria in kg
in kg count in 100 times 5% of 20% of 50% of 80%of
kg least max max max max
count capacity capacity capacity capacity

Note – Standard weight under 5%, 20%, 50%and 80% of capacity


of balance was rounded of nearest multiply by 5.
Q. 11. How the Linearity B of single pan balance is calibrated &
what are the acceptance criteria?
A. 11. The Linearity B of Single Pan Balance is calibrated by taking weight
of (i) 100 times of least count, (ii) 2.5% (iii) 5%, (iv) 20%, (v) 50% and (vi)
80% of maximum capacity of balance by using Certified Weights and plot
the graph between standard value and weights observed.
Note – Standard weight under 5%, 20%, 50%and 80% of capacity
of balance was rounded of nearest multiply by 5.
Capacity Least Standard weight &acceptance criteria in kg
in kg count in 100 2.5% of 5% of 20% of 50% of 80%of
kg times max max max max max
least capacity capacity capacity capacity capacity
count

The acceptance criteria: The Graph shall be linear & Coefficient of


correlation: NLT 0.99.
Q. 11. How the Drift (Eccentricity) of single pan balance is
calibrated & what are the acceptance criteria?
A. 11. The Drift (Eccentricity) of Single Pan Balance is calibrated by taking
weight of 5% of maximum capacity of balance by using Certified Weights
at five different place of weighing pan (centre & edge of cross section of
diameter).
Position Observed Deviation from centre
weight (g)
Centre of the pan (A) (A)
Top right corner (B) (B-A) x 1000 =
Bottom right corner ( (C-A) x 1000 =
C)
Bottom Left corner (D-A) x 1000 =
(D)
Top left corner (E) (E-A) x 1000 =
Acceptance criteria: Observed weight from each position should not
deviate from the weight observed at the centre of the pan by ± 0.1%
Q. 12. How the Uncertainty of single pan balance is calibrated &
what are the acceptance criteria?
A. 5. The Uncertainty of Single Pan Balance is calibrated by taking weight
of 5% of maximum capacity of balance by using Certified Weights 10
times. Calculate the average and standard deviation of all 10
measurements.
Replicates Observed weight (g)
1
2
3
4
5
6
7
8
9
10
Average
SD

3X Standard Deviation (SD)


Uncertainty = -------------------------------------
Value of Certified Weights
Acceptance criteria: The Uncertainty should not be more than 0.001%
(Ref: USP)