Sei sulla pagina 1di 192

ACUSON X300

US

Replacements of Parts
System
Replacement of Parts

This document is valid for use with the following products:


© Siemens, 2011

ACUSON X300 and SONOVISTA X300 and ACUSON


X300 Premium Edition
10348533
10348532
10348531
10132987
10038837
10037409

© Siemens, 2011
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
BJ
Siemens
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: US11-104.841.01.09.02 English


Replaces: US11-104.841.01.08.02 Doc. Gen. Date: 02.11
n.a.

Part No.: 2011

Ultrasound
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2011“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

ACUSON X300 US11-104.841.01.09.02 Page 2 of 192 © Siemens, 2011


02.11 Ultrasound
Table of Contents 3
0Table of Contents

1________ Removing and Replacing FRUs_____________________________________ 8

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2________ System Covers _________________________________________________ 10

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Required Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Tray Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Removing the Tray Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Left Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Removing the Left Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Replacing the Left Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Right Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Removing the Right Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Replacing the Right Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Removing the Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Replacing the Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Front Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Removing the Front Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Replacing the Front Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Removing the Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Replacing the Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
CP Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Removing the CP Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Replacing the CP Rear Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

3________ Replacing the 15 inch Monitor_____________________________________ 25

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Removing the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Replacing the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

4________ Replacing the 17 inch Monitor_____________________________________ 34

© Siemens, 2011 US11-104.841.01.09.02 Page 3 of 192 ACUSON X300


02.11 Ultrasound
4 Table of Contents

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Removing the Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Replacing the Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

5 _______ Monitor Arm Covers ____________________________________________ 43

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Removing the Monitor Arm Covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Replacing the Monitor Arm Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

6 _______ FPD Arm and Control Panel Neck _________________________________ 52

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Replace the Arm Lock Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Remove the Arm Lock Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Replace the Arm Lock Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Removing the Control Panel Neck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Replacing the Control Panel Neck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

7 _______ Control Panel __________________________________________________ 61

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Removing the Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Replace the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Removing the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Replace the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Power Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Removing the Power Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
DGC Slide Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

ACUSON X300 US11-104.841.01.09.02 Page 4 of 192 © Siemens, 2011


02.11 Ultrasound
Table of Contents 5

Removing the DGC Slide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70


Trackball . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Removing the Trackball (Mechanical, ACUSON X300) . . . . . . . . . . . . . . . . . . . . . . . 73
Removing the Trackball (Optical, ACUSON X300 PE). . . . . . . . . . . . . . . . . . . . . . . . 74
Control Panel Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Removing the Control Panel Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Replacing the Control Panel and Control Panel Sub-assemblies . . . . . . . . . . . . . . . . . . 79
Control Panel Base Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Removing the Control Panel Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Replacing the Control Panel Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

8________ Gas Cylinder Replacement _______________________________________ 86

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
System Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Removing the Gas Cylinder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Replacing the Gas Cylinder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

9________ Up Down Roller Assembly Replacement ____________________________ 92

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Removing the Up Down Roller Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Replacing the Up Down Roller Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

10_______ Modules and Board Replacement _________________________________ 100

TI Module Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100


TI Daughter Board (TI DL3 Board: ACUSON X300 PE) . . . . . . . . . . . . . . . . . . . . . 100
TI Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Replacement of the Wobbler Boards Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Electronic Modules and Boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Electronic Module and Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
ETX Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
RC Daughter Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Removing and Replacing the Hard Disk Drive (HDD) . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Removing the HDD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Replacing the HDD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Verifying the Operation of the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131

© Siemens, 2011 US11-104.841.01.09.02 Page 5 of 192 ACUSON X300


02.11 Ultrasound
6 Table of Contents

I/O Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132


Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Removing the I/O Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Replacing the I/O Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Replacing of the I/O Sub Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Replacing Backplane with Chassis Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Removing the Backplane with Chassis Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . 141
Replacing the Backplane with Chassis Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . 143

11 ______ CD/DVD-R/RW and ECG ________________________________________ 144

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Removing the CD/DVD-R/RW and ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Replacing the CD/DVD-R/RW and ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151

12 ______ Air Filter _____________________________________________________ 152

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Removing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Replacing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154

13 ______ AC/DC Power Supply ___________________________________________ 155

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Removing the AC/DC Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Replacing the AC/DC Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Verifying the Operation of the AC/DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . 160

14 ______ Replace the Battery Pack _______________________________________ 161

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Removing the Battery Pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Replacing the Battery Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
Verifying the Operation of the AC/DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . 166

ACUSON X300 US11-104.841.01.09.02 Page 6 of 192 © Siemens, 2011


02.11 Ultrasound
Table of Contents 7

15_______ DC/DC Power Supply ___________________________________________ 167

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Removing the DC/DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Replacing the DC/DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Verifying the Operation of the DC/DC Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . 171

16_______ AC/DC Power Fan ______________________________________________ 172

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Removing the AC/DC Power Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Replacing the AC/DC Power Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175
Verifying the Operation of the AC/DC Power Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176

17_______ Transducer Holders ____________________________________________ 177

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Removing the Transducer Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Replacing the Transducer Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179

18_______ Cables Replacement____________________________________________ 180

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Time Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Removing the Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Replacing the System Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190

19_______ Changes to Previous Version ____________________________________ 192

Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192

© Siemens, 2011 US11-104.841.01.09.02 Page 7 of 192 ACUSON X300


02.11 Ultrasound
8 Removing and Replacing FRUs
1-
Introduction
1Removing and Replacing FRUs

The removal and replacement procedures included in this section share several character-
istics: The operating voltages and the weight of the monitor represent specific hazards.

WARNING The system contains hazardous voltages that can range from
90 VAC to 264 VAC and cause serious injury or death.
¹ Only authorized service personnel are permitted to
remove protective covers and access the energized
components of this system.

WARNING The weight of the ACUSON X300™ ultrasound systems


approximately (98 kg; 216 lb.) and the weight of the monitor
represent lifting hazards.
The use of improper lifting methods or lifting in an awkward
manner can cause serious injury.
¹ The US National Institute for Occupational Safety and
Health publishes design guidelines that indicate that
items weighing greater than 15.9 kg (35 lb.) must be
lifted by two people or mechanical lifting aids. Items
weighing greater than 34 kg (75 lb.) must be lifted only
with mechanical lifting equipment. The transport
instructions for the system, when packaged in its ship-
ping container, are provided on the container. The sys-
tem, once removed from its shipping container, is not
designed to be lifted. Transport the system by moving
it on its casters.

A Spare Part Catalog (SPC) has been established on the Internet at the following URL:
(Spare Parts Catalog, http://cs.med.siemens.de/).
The installation of a spare part employs the reverse sequence provided for the disassembly
and removal procedure.

CAUTION Use fasteners of the correct length.


¹ Failure to use fasteners of correct may cause damage
to underlying equipment.

ACUSON X300 US11-104.841.01.09.02 Page 8 of 192 © Siemens, 2011


02.11 Ultrasound
Removing and Replacing FRUs 9

CAUTION The system contains components that are sensitive to dam-


age by electrostatic discharge (ESD).
¹ Observe all ESD protective procedures and use ESD
protective equipment including an ESD wrist strap
when the protective covers are removed from the sys-
tem.

NOTE Before proceeding with these instructions, make sure the


system is positioned on a flat surface with the system wheels
securely locked in place and that there is sufficient work
space, more than one meter, around the system.

NOTE The ACUSON X300 Premium Edition system has a different


color look than the ACUSON X300 system.

© Siemens, 2011 US11-104.841.01.09.02 Page 9 of 192 ACUSON X300


02.11 Ultrasound
10 System Covers
2-
Overview
2System Covers

This procedure describes the removal and installation of the ACUSON X300™ ultrasound
system covers.

Workflow 0

1. Remove/replace the Tray Box


2. Remove/replace the Left cover
3. Remove/replace the Right cover
4. Remove/replace the Rear cover
5. Remove/replace the Front cover
6. Remove/replace the Top cover

Required Tools 0

#2 Phillips screwdriver

Required Time 0

Replacing the Covers can take up to 30 minutes.

System Preparations 0

1. Press the Power button on the Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the to Off.
3. Disconnect the system power cord from the external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 10 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 11

Tray Box 2.1

Removing the Tray Box 0

Perform the following steps to remove the tray box from the system.

1. Remove the contents from the tray box.

Fig. 1: Remove the contents from the tray box


2. Loosen the four (4) PWH screws from the tray box.

Fig. 2: Loosen the screws fon the tray box


3. Remove the tray box.

© Siemens, 2011 US11-104.841.01.09.02 Page 11 of 192 ACUSON X300


02.11 Ultrasound
12 System Covers

Left Cover 2.2

Removing the Left Cover 0

Perform the following steps to remove the left cover of the system.

1. Remove the air filter from the left side of the system.

Fig. 3: Remove the air filter on the left side of the system

ACUSON X300 US11-104.841.01.09.02 Page 12 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 13

2. Loosen the two (2) TH half screws from the left cover.

Fig. 4: Loosen the screws from the left cover


3. Lift the left cover up to release it from the hooks.

Fig. 5: Lift up the left cover

© Siemens, 2011 US11-104.841.01.09.02 Page 13 of 192 ACUSON X300


02.11 Ultrasound
14 System Covers

4. Pull the left cover out away from the system.

Fig. 6: Pull out the left cover

Replacing the Left Cover 0

Perform the steps in reverse sequence to replace the left cover.

ACUSON X300 US11-104.841.01.09.02 Page 14 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 15

Right Cover 2.3

Removing the Right Cover 0

Perform the following steps to remove the right cover of the system.

1. Loosen the two (2) TH half screws at the bottom of the right cover.

Fig. 7: Loosen the two TH screws from the right cover

© Siemens, 2011 US11-104.841.01.09.02 Page 15 of 192 ACUSON X300


02.11 Ultrasound
16 System Covers

2. Lift the right cover up to release it from the hooks.

Fig. 8: Lift up the right cover


3. Pull the right cover out away from the system.

Fig. 9: Pull out the right cover

Replacing the Right Cover 0

Perform the steps in reverse sequence to replace the right cover.

ACUSON X300 US11-104.841.01.09.02 Page 16 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 17

Rear Cover 2.4

Removing the Rear Cover 0

Perform the following steps to remove the rear cover of the system.

1. Remove the right and left covers, see (Removing the Left Cover / p. 12) and (Remov-
ing the Right Cover / p. 15).
2. Loosen the six (6) TH washer screws.
- There are three (3) screws on each side of the rear cover.

Fig. 10: Loosen the six screws from the rear cover - three on each side

© Siemens, 2011 US11-104.841.01.09.02 Page 17 of 192 ACUSON X300


02.11 Ultrasound
18 System Covers

3. Pull out the rear cover away from the system.

Fig. 11: Remove the rear cover

Replacing the Rear Cover 0

Perform the steps in reverse sequence to replace the rear cover.

ACUSON X300 US11-104.841.01.09.02 Page 18 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 19

Front Cover 2.5

Removing the Front Cover 0

Perform the following steps to remove the front cover of the system.

1. Remove the left and right covers, see (Removing the Left Cover / p. 12) and (Remov-
ing the Right Cover / p. 15).
2. Loosen the six (6) TH washer screws.
- There are three (3) screws on each side of the front cover.

Fig. 12: Loosen the six screws from the front cover - three on each side

© Siemens, 2011 US11-104.841.01.09.02 Page 19 of 192 ACUSON X300


02.11 Ultrasound
20 System Covers

3. Pull the front cover off the system.

Fig. 13: Remove the front cover

Replacing the Front Cover 0

Perform the steps in reverse sequence to replace the front cover.

ACUSON X300 US11-104.841.01.09.02 Page 20 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 21

Top Cover 2.6

Removing the Top Cover 0

Perform the following steps to remove the top cover of the system.

NOTE If a peripheral (color, BW printer, VCR or DVD) or Pocket


accessory option (ACUSON X300 PE) is installed on the sys-
tem, remove them first before removing the top cover.

1. Loosen the four (4) TH screws at the edge of the top cover.

Fig. 14: Loosen the four screws from the top cover

© Siemens, 2011 US11-104.841.01.09.02 Page 21 of 192 ACUSON X300


02.11 Ultrasound
22 System Covers

2. Pull the top cover off the system.

Fig. 15: Remove the top cover

Replacing the Top Cover 0

Perform the steps in reverse sequence to replace the top cover.

ACUSON X300 US11-104.841.01.09.02 Page 22 of 192 © Siemens, 2011


02.11 Ultrasound
System Covers 23

CP Rear Cover 2.7

Removing the CP Rear Cover 0

Perform the following steps to remove the CP rear cover of the system.

1. Loosen the four (4) TH washer screws.


- There are two (2) screws on each side of the CP rear cover.

Fig. 16: Loosen the four screws from the CP rear cover - two on each side

© Siemens, 2011 US11-104.841.01.09.02 Page 23 of 192 ACUSON X300


02.11 Ultrasound
24 System Covers

2. Pull the CP rear cover out, away from the system.

Fig. 17: Remove the CP rear cover

Replacing the CP Rear Cover 0

Perform the steps in reverse sequence to replace the CP rear cover.

ACUSON X300 US11-104.841.01.09.02 Page 24 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 15 inch Monitor 25
3-
Overview
3Replacing the 15 inch Monitor

Use the following instructions to remove and replace the 15 inch monitor. There are 4 types
of 15 inch monitors and 1 type of 17 inch monitor available for the ACUSON X300™ ultra-
sound system. Determine whether the system has a 15 inch, or a 17 inch monitor, and use
the appropriate procedure.

Workflow 0

1. Remove the monitor


2. Replace the monitor

Requirements 0

Required Materials
Tab. 1 ACUSON X300 15 inch Monitor and Software Versions

Software Version Monitor Part Number


R1.0 and 2.0 10133128 or 10131649
R3.0 10348372
R3.0, R4.0, R5.0 and 10427540
R6.0

Required Tools 0

• #2 Phillips screwdriver
• Loctite 243
• 5mm Allen wrench
• 6mm Allen wrench

Time Required 0

Approximately 30 minutes.

© Siemens, 2011 US11-104.841.01.09.02 Page 25 of 192 ACUSON X300


02.11 Ultrasound
26 Replacing the 15 inch Monitor

System Preparations 0

1. Press the Power button on the Control Panel to power off the system.
2. Switch the Main Circuit Breaker to Off.
3. Disconnect the system power cord from the external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 26 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 15 inch Monitor 27

Removing the Monitor 3.1

1. Lock the monitor by pushing and turning the arm lock knob clockwise.

NOTE If a peripheral, (color, BW printer, VCR or DVD) is installed on


the system, remove the peripheral to gain access to the mon-
itor and the rear of the monitor.

Fig. 18: Direction for Locking the Monitor Arm Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 27 of 192 ACUSON X300


02.11 Ultrasound
28 Replacing the 15 inch Monitor

2. Loosen the three (3) TH thumbscrews attached to the pinch cover.

Fig. 19: Three Cover Screws Shown


3. Remove the pinch cover.

Fig. 20: Shown with Pinch Cover Removed

ACUSON X300 US11-104.841.01.09.02 Page 28 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 15 inch Monitor 29

4. Remove the four (4) PWH screws that attach the monitor cables.

Fig. 21: 15 inch Monitor Cable Securing Screws Shown


5. Disconnect the three (3) cables from the monitor.

Fig. 22: Monitor Cables Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 29 of 192 ACUSON X300


02.11 Ultrasound
30 Replacing the 15 inch Monitor

6. Unfasten the two (2) PWH screws and remove the monitor cable cover.

Fig. 23: Remove the Monitor Cable Cover


With the Monitor cable cover removed, the system should look like the following:

Fig. 24: Cable Cover Removed from Monitor

ACUSON X300 US11-104.841.01.09.02 Page 30 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 15 inch Monitor 31

7. Unlock the monitor by pushing and turning the lever counterclockwise.

Fig. 25: Direction to Unlock Monitor Shown


8. Remove the SBH bolt, use the 5 mm and 6 mm Allen wrenches, located at the bottom of
the monitor rotation part.

Fig. 26: Removing Monitor with Wrenches


9. Pull the monitor up and removed it from the system.

© Siemens, 2011 US11-104.841.01.09.02 Page 31 of 192 ACUSON X300


02.11 Ultrasound
32 Replacing the 15 inch Monitor

NOTE While removing the monitor, keep the shaft monitor rotation
level with the front of the monitor arm.

Fig. 27: Direction to Lift Monitor Shown

ACUSON X300 US11-104.841.01.09.02 Page 32 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 15 inch Monitor 33

Replacing the Monitor 3.2

Replace the monitor using the reverse sequence of the procedure provided for the removal.

NOTE Use Loctite 243 for holding the SBH bolt when replacing the
monitor.

© Siemens, 2011 US11-104.841.01.09.02 Page 33 of 192 ACUSON X300


02.11 Ultrasound
34 Replacing the 17 inch Monitor
4-
Overview
4Replacing the 17 inch Monitor

Use the following instructions to remove and replace the 17 inch monitor. There are 4 types
of 15 inch monitors. and 1 type 17 inch monitor available for the ACUSON X300™ ultra-
sound system. Determine whether the system has a 15 inch, or a 17 inch monitor, and use
the appropriate procedure.

Workflow 0

1. Remove the monitor


2. Replace the monitor

Requirements 0

Required Materials
17 inch monitor P/N 10566101

Required Tools 0

• #2 Phillips screwdriver
• Loctite 243
• 5mm Allen wrench
• 6mm Allen wrench

Time Required 0

Approximately 30 minutes.

System Preparations 0

1. Press the Power button on the Control Panel to power off the system.
2. Switch the Main Circuit Breaker to Off.
3. Disconnect the system power cord from the external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 34 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 17 inch Monitor 35

Removing the Monitor 4.1

1. Lock the monitor by pushing and turning the arm lock knob clockwise.

NOTE If a peripheral, (color, BW printer, VCR or DVD) is installed on


the system, remove the peripheral to gain access to the mon-
itor and the rear of the monitor.

Fig. 28: Direction for Locking the Monitor Arm Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 35 of 192 ACUSON X300


02.11 Ultrasound
36 Replacing the 17 inch Monitor

2. Loosen the two (2) screws attached to the pinch cover.

Fig. 29: 17 inch Monitor Pinch Cover Shown


3. Remove the pinch cover.

Fig. 30: Shown with Pinch Cover Removed

ACUSON X300 US11-104.841.01.09.02 Page 36 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 17 inch Monitor 37

4. Remove the four (4) PWH screws that attach the monitor cables.

Fig. 31: 17 inch Monitor Cable Securing Screws


5. Disconnect the three (3) cables from the monitor.

Fig. 32: Monitor Cables Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 37 of 192 ACUSON X300


02.11 Ultrasound
38 Replacing the 17 inch Monitor

6. Unlock the monitor by pushing and turning the lever counterclockwise.

Fig. 33: Direction to Unlock Monitor Shown


7. Remove the SBH bolt, use the 5 mm and 6 mm Allen wrenches, located at the bottom of
the monitor rotation part.

Fig. 34: Removing the Monitor with Wrenches


8. Pull the monitor up and removed it from the system.

ACUSON X300 US11-104.841.01.09.02 Page 38 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 17 inch Monitor 39

NOTE While removing the monitor, keep the shaft monitor rotation
level with the front of the monitor arm.

Fig. 35: Removal of Monitor Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 39 of 192 ACUSON X300


02.11 Ultrasound
40 Replacing the 17 inch Monitor

Replacing the Monitor 4.2

This section describes the steps required to replace the 17 inch monitor.
1. Have someone hold the monitor while inserting the SBH bolt in place. Apply Loctite 243
on thread ends of bolt. Tighten the bolt.

Fig. 36: Removing the Monitor with Wrenches

NOTE Use Loctite 243 for holding the SBH bolt when replacing the
monitor.

ACUSON X300 US11-104.841.01.09.02 Page 40 of 192 © Siemens, 2011


02.11 Ultrasound
Replacing the 17 inch Monitor 41

2. Connect the three (3) cables from the monitor as shown below.

Fig. 37: Monitor Cables Shown


3. Install the four (4) PWH screws that attach the monitor cables.

Fig. 38: 17 inch Monitor Cable Securing Screws

© Siemens, 2011 US11-104.841.01.09.02 Page 41 of 192 ACUSON X300


02.11 Ultrasound
42 Replacing the 17 inch Monitor

4. Install the pinch cover and install the two (2) screws to secure the pinch cover in place.

Fig. 39: 17 inch Monitor Pinch Cover Shown


5. Unlock the monitor by pushing and turning the lever counterclockwise.

Fig. 40: Direction to Unlock Monitor Shown

ACUSON X300 US11-104.841.01.09.02 Page 42 of 192 © Siemens, 2011


02.11 Ultrasound
Monitor Arm Covers 43
5-
Overview
5Monitor Arm Covers

Use the procedure provided in this section to remove and replace the Monitor Arm covers.

Workflow 0

1. Remove Monitor Arm covers


2. Replace Monitor Arm covers

Tools Required 0

#2 Phillips screwdriver

Time Required 0

Approximately 10 minutes.

System Preparations 0

1. Press the Power button on the system Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the systme to Off.
3. Disconnect the system power cord from the system external power source.
4. Maintain a minimum of 3.5 feet (1.70 meters) clearance while working on the system.

© Siemens, 2011 US11-104.841.01.09.02 Page 43 of 192 ACUSON X300


02.11 Ultrasound
44 Monitor Arm Covers

Removing the Monitor Arm Covers 5.1

1. Straighten the monitor to its fullest vertical position.

Fig. 41: Straighten the monitor to its full vertical position

ACUSON X300 US11-104.841.01.09.02 Page 44 of 192 © Siemens, 2011


02.11 Ultrasound
Monitor Arm Covers 45

2. Remove the (1) PWH screw.

Fig. 42: Remove the PWH screw from the arm cable cover
3. Remove the four (4) PH screws from the bottom of the arm cover.

Fig. 43: Remove the screws from the bottom of the arm cover

© Siemens, 2011 US11-104.841.01.09.02 Page 45 of 192 ACUSON X300


02.11 Ultrasound
46 Monitor Arm Covers

- Remove the arm top rear cover and the two (2) arm side covers.

Fig. 44: Remove the top rear and side arm covers

Fig. 45: Remove the bottom and side arm covers

ACUSON X300 US11-104.841.01.09.02 Page 46 of 192 © Siemens, 2011


02.11 Ultrasound
Monitor Arm Covers 47

4. Remove the four (4) PH screws from the bottom of the arm cover.

Fig. 46: Remove the four screws from the bottom of the Arm cover
- Remove the arm bottom cover and the two (2) arm side covers.

Fig. 47: Remove the bottom and side arm covers

© Siemens, 2011 US11-104.841.01.09.02 Page 47 of 192 ACUSON X300


02.11 Ultrasound
48 Monitor Arm Covers

5. Fold the monitor arm down.

Fig. 48: Fold the monitor arm down

ACUSON X300 US11-104.841.01.09.02 Page 48 of 192 © Siemens, 2011


02.11 Ultrasound
Monitor Arm Covers 49

- Remove the arm top front cover and two (2) arm side covers.

Fig. 49: Remove the top front and side covers

© Siemens, 2011 US11-104.841.01.09.02 Page 49 of 192 ACUSON X300


02.11 Ultrasound
50 Monitor Arm Covers

When the monitor arm covers are removed, the system should look like the following:

Fig. 50: Monitor Arm covers removed

ACUSON X300 US11-104.841.01.09.02 Page 50 of 192 © Siemens, 2011


02.11 Ultrasound
Monitor Arm Covers 51

Replacing the Monitor Arm Covers 5.2

To replace the Monitor Arm covers, perform the instructions in reverse. After replacing the
Monitor Arm covers, replace the Monitor Cables before cable installation.

© Siemens, 2011 US11-104.841.01.09.02 Page 51 of 192 ACUSON X300


02.11 Ultrasound
52 FPD Arm and Control Panel Neck
6-
Overview
6FPD Arm and Control Panel Neck

Use the instructions provided in this section to remove and replace the control panel neck
on the system.

Workflow 0

• Remove Arm Lock Assembly


• Replace the Arm Lock Assembly
• Remove the Control Panel Neck
• Replace the Control Panel Neck

Tools Required 0

#2 Phillips screwdriver
• Number 2 Phillips screwdriver
• 6mm Allen wrench
• 1.5mm Allen wrench

Time Required 0

Approximately 1 hour.

System Preparations 0

1. Press the Power button on the system Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the system to Off.
3. Disconnect the system power cord from the system external power source.
4. Maintain a minimum of 3.5 feet (1.70 meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 52 of 192 © Siemens, 2011


02.11 Ultrasound
FPD Arm and Control Panel Neck 53

Replace the Arm Lock Assembly 6.1

NOTE Ensure that the Control Panel is fully raised before attempt-
ing to remove or replace the Arm Lock Assembly. This helps
to gain access to the bottom of the Control Panel Neck.

Remove the Arm Lock Assembly 0

This procedure describes the removal and replacement of the Arm Lock Assembly.
1. Rotate the Knob of the Arm Lock Assembly to the Unlock position.

Fig. 51: Arm Lock Lock/Unlock Knob Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 53 of 192 ACUSON X300


02.11 Ultrasound
54 FPD Arm and Control Panel Neck

2. Remove the (5) screws located on the bottom of the CP neck and remove the CP Neck
cover.

Fig. 52: Control Panel Neck Cover with Screws Shown


3. While pressing the knob inward, remove the Set screw using a 1.5mm Allen wrench.

Fig. 53: Details of Knob Removal Shown

ACUSON X300 US11-104.841.01.09.02 Page 54 of 192 © Siemens, 2011


02.11 Ultrasound
FPD Arm and Control Panel Neck 55

4. Remove the (4) screws securing the Arm Lock Assembly to the Control Panel Neck and
pull down the Arm Lock Assembly to separate it from the Control Panel.

Fig. 54: Arm Lock Assembly - Screws Shown

Fig. 55: Arm Lock Assembly Removed

© Siemens, 2011 US11-104.841.01.09.02 Page 55 of 192 ACUSON X300


02.11 Ultrasound
56 FPD Arm and Control Panel Neck

Replace the Arm Lock Assembly 0

1. Install the Arm Lock Assembly into the Control Panel Neck.

NOTE Before installing the Arm Lock Assembly, extend the Shaft
Lock upward for ease of assembly. Refer to figure below.

Fig. 56: Arm Lock Assembly - Installation Postion


2. Install the (4) screws to the Arm Lock Assembly. Refer to(Fig. 54 / p. 55).
3. While pressing the knob inward, install the Set screw and tighten using a 1.5mm Allen
wrench. Refer to (Fig. 53 / p. 54).
4. Install Neck Cover and the (5) screws that hold the CP Neck cover. Refer to
(Fig. 52 / p. 54).
5. Verify that the knob functions properly by placing it in both the lock and unlock positions.

ACUSON X300 US11-104.841.01.09.02 Page 56 of 192 © Siemens, 2011


02.11 Ultrasound
FPD Arm and Control Panel Neck 57

Removing the Control Panel Neck 6.2

Use the following instructions to remove the control panel neck.

1. Power off the system and switch the MAIN switch to OFF; leave the power cord con-
nected to the wall outlet.

WARNING The system contains hazardous voltages, 90 VAC to 264


VAC, that can result in serious injury or death.
¹ Only authorized service personnel are permitted to
remove the protective covers to access the energized
components of the system.

2. Remove the (5) screws from the CP neck cover.


3. Remove the cover of the CP neck from the system.

Fig. 57: Remove the five TH screws and control panel neck

© Siemens, 2011 US11-104.841.01.09.02 Page 57 of 192 ACUSON X300


02.11 Ultrasound
58 FPD Arm and Control Panel Neck

4. Cut the cable tie at the location marked.

Fig. 58: Cut the cable tie


5. Remove the (3) HFH bolts and disassemble the Flat Panel DIsplay (FPD) arm.

Fig. 59: Remove the three HFH bolts

ACUSON X300 US11-104.841.01.09.02 Page 58 of 192 © Siemens, 2011


02.11 Ultrasound
FPD Arm and Control Panel Neck 59

6. Remove the (3) HSH bolts to disassemble the control panel neck from the control panel
base.

Fig. 60: Remove the screws on the control panel base

© Siemens, 2011 US11-104.841.01.09.02 Page 59 of 192 ACUSON X300


02.11 Ultrasound
60 FPD Arm and Control Panel Neck

Replacing the Control Panel Neck 6.3

Replace the control panel neck using the reverse sequence of the procedures provided for
the removal.

ACUSON X300 US11-104.841.01.09.02 Page 60 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 61
7-
Overview
7Control Panel

Use the instructions provided in this section to remove and replace the control panel and
its’ sub-assemblies on the system. The sub-assembly components of the Control Panel
include:
• Keyboard
• Power Button
• DGC Slide Assembly
• Trackball
• Control Panel Main PCB

Workflow 0

Remove Control Panel


Remove/replace the Keyboard
Remove/replace the Power Button
Remove/replace the DGC Slide assembly
Remove/replace the Trackball
Remove/replace Control Panel Main PCB
Remove/replace Control Panel Base

Tools Required 0

Number 2 Phillips screwdriver


Loctite 271
Metric wrench set
Tie-wraps
Diagnoial cutters

Time Required 0

Each sub-assembly procedure takes approximately 15 minutes, with the exception of the
Control Panel Main PCB board removal which takes approximately 45 minutes, and the
Control Panel Base procedure which takes approximately 3 hours.

© Siemens, 2011 US11-104.841.01.09.02 Page 61 of 192 ACUSON X300


02.11 Ultrasound
62 Control Panel

System Preparations 0

1. Press the Power button on the system Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the system to Off.
3. Disconnect the system power cord from the system external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 62 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 63

Removing the Control Panel 7.1

Use the following instructions to remove the control panel.

1. Remove the (6) TH screws located under the control panel.

Fig. 61: Remove the screws under the control panel


2. Pull up on the control panel and disconnect the cables that are attached.

© Siemens, 2011 US11-104.841.01.09.02 Page 63 of 192 ACUSON X300


02.11 Ultrasound
64 Control Panel

- Disconnect the USB Frame cable

Fig. 62: Disconnect the USB cable from the control panel base
- Disconnect the Signal Frame cable

Fig. 63: Disconnect the Signal Frame cable from the control panel base

ACUSON X300 US11-104.841.01.09.02 Page 64 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 65

The control panel base looks like the following after removing the control panel and
the rear side of the control panel.

Fig. 64: Control panel base

Fig. 65: The rear of the control panel

© Siemens, 2011 US11-104.841.01.09.02 Page 65 of 192 ACUSON X300


02.11 Ultrasound
66 Control Panel

Replace the Keyboard 7.2

Perform the following to remove and replace the keyboard

Removing the Keyboard 0

1. Remove the control panel from the system, see (Control Panel / p. 61).
2. Remove the signal cable and power cable from the back side of the control panel.

Fig. 66: Remove the power and signal cables

ACUSON X300 US11-104.841.01.09.02 Page 66 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 67

3. Remove the (6) PWH tapping screws that attach the keyboard to the control panel.

Fig. 67: Remove the six screws


The keyboard looks like the following once it is removed from the control panel:

Fig. 68: The keyboard removed from the control panel

Replace the Keyboard 0

To replace the keyboard follow the steps in (Removing the Keyboard / p. 66).

© Siemens, 2011 US11-104.841.01.09.02 Page 67 of 192 ACUSON X300


02.11 Ultrasound
68 Control Panel

Power Button 7.3

Use the following procedure to remove and replace the power button assembly from the
control panel.

Removing the Power Button 0

1. Remove the control panel from the system, see (Control Panel / p. 61).
2. Remove the power cable from the back of the control panel.

Fig. 69: Remove the power cable from the rear of the control panel

ACUSON X300 US11-104.841.01.09.02 Page 68 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 69

3. Remove the three (3) PWH tapping screws.

Fig. 70: Remove the three screws


4. Remove the power button from the control panel.

Fig. 71: The power button removed from the control panel

© Siemens, 2011 US11-104.841.01.09.02 Page 69 of 192 ACUSON X300


02.11 Ultrasound
70 Control Panel

DGC Slide Assembly 7.4

Perform the following to remove and replace the DCG Slide Assembly.

Removing the DGC Slide 0

1. Remove the control panel from the system, see (Control Panel / p. 61).
2. Remove the power cable from the rear of the control panel.

Fig. 72: Remove the signal cable from the rear of the control panel
3. Remove the six (6) PWH tapping screws that attach the DGC slide to the control panel.

ACUSON X300 US11-104.841.01.09.02 Page 70 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 71

NOTE Before removing DGC slide, separate all the DGC knobs from
the DGC slide.

Fig. 73: Remove the six screws on the DGC slide

© Siemens, 2011 US11-104.841.01.09.02 Page 71 of 192 ACUSON X300


02.11 Ultrasound
72 Control Panel

4. Remove the DGC slide from the control panel.

Fig. 74: Remove the DGC slide from the control panel

ACUSON X300 US11-104.841.01.09.02 Page 72 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 73

Trackball 7.5

Perform the following to remove and replace the trackball from the control panel.

Removing the Trackball (Mechanical, ACUSON X300) 0

1. Remove the Control Panel from the system, see (Control Panel / p. 61).
2. Remove the trackball 4p cable and trackball 6p cable from the rear of the control panel.

Fig. 75: Remove the signal cables from the trackball assembly

© Siemens, 2011 US11-104.841.01.09.02 Page 73 of 192 ACUSON X300


02.11 Ultrasound
74 Control Panel

3. Remove the (4) PWH tapping screws.

Fig. 76: Remove the screws holding the trackball assembly


4. Remove the trackball assembly from the control panel.

Fig. 77: Remove the trackball assembly from the control panel

Removing the Trackball (Optical, ACUSON X300 PE) 0

Perform the following to remove and replace the trackball from the control panel.

ACUSON X300 US11-104.841.01.09.02 Page 74 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 75

1. Remove the Control Panel from the system, see (Control Panel / p. 61).
2. Remove the (2) trackball 4P cable and 10P cable from the rear of the control panel.

Fig. 78: Remove the signal cables from the trackball assembly
3. Remove the (4) PWH tapping screws.

Fig. 79: Remove the screws holding the trackball assembly

© Siemens, 2011 US11-104.841.01.09.02 Page 75 of 192 ACUSON X300


02.11 Ultrasound
76 Control Panel

4. Remove the Trackball assembly from the control panel.

Fig. 80: Remove the trackball assembly from the control panel

ACUSON X300 US11-104.841.01.09.02 Page 76 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 77

Control Panel Main PCB 7.6

Perform the following to remove and replace the Control Panel Main PCB.

Removing the Control Panel Main PCB 0

The following pictures show the Control Panel Main PCBs.

Fig. 81: Main PCB of the control panel assembly (ACUSON X300)

© Siemens, 2011 US11-104.841.01.09.02 Page 77 of 192 ACUSON X300


02.11 Ultrasound
78 Control Panel

Fig. 82: Main PCB of the control panel assembly. (ACUSON X300 PE)

ACUSON X300 US11-104.841.01.09.02 Page 78 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 79

Replacing the Control Panel and Control Panel Sub-assemblies 7.7

Replace the control panel and its sub-assemblies by using the reverse sequence of each
of the procedures.

© Siemens, 2011 US11-104.841.01.09.02 Page 79 of 192 ACUSON X300


02.11 Ultrasound
80 Control Panel

Control Panel Base Replacement 7.8

Overview 0

Use the instructions in this section to remove and replace the control panel base on the
ACUSON X300 system.

System Preparations 0

1. Press the Power button on the system Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the system to Off.
3. Disconnect the system power cord from the system external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

Removing the Control Panel Base 0

1. Remove the monitor; see (Removing the Monitor / p. 27).


2. Remove the monitor arm covers; see (Removing the Monitor Arm Covers / p. 44).
3. Remove the control panel; see(Removing the Control Panel / p. 63).
4. Cut the cable ties at the marked location.

Fig. 83: Cable Ties Shown

ACUSON X300 US11-104.841.01.09.02 Page 80 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 81

5. Remove the five (5) screws beneath the CP neck to remove the CP neck cover from the
CP neck.

Fig. 84: Remove the CP neck cover


6. Pull out the three (3) cables through the Hole 1 beneath the CP neck and the Hole 2 at
the front of the CP neck.

Fig. 85: Calbe Ties Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 81 of 192 ACUSON X300


02.11 Ultrasound
82 Control Panel

7. Cut the cable ties at the marked location and disconnect the gas cylinder cable from the
gas cylinder.

Fig. 86: Disconnect the cables


8. Remove the seven (7) screws to remove the up-down button assembly from the control
panel base.

Fig. 87: Remove the up-down button assembly

ACUSON X300 US11-104.841.01.09.02 Page 82 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 83

9. Remove the four (4) HSH bolts on the control panel base using a 6mm Allen wrench.

Fig. 88: Remove the bolts on the control panel base


10. Lift the control panel base and pull out the cables through the control panel hole to sep-
arate it from the main frame.

Fig. 89: Control Panel Base and Main Frame Separated

© Siemens, 2011 US11-104.841.01.09.02 Page 83 of 192 ACUSON X300


02.11 Ultrasound
84 Control Panel

11. Remove the three (3) HSH screws located on the back side of the control panel base to
remove the CP neck from the control panel base using a 5mm Allen wrench.

Fig. 90: Remove the control panel neck


12. Grasp the rear side of the transducer holder and pull upward to remove the transducer
holders and remove the screws on the cable hooks to remove the cable hooks from the
control panel base.

Fig. 91: Remove the cable hooks, cup probes

ACUSON X300 US11-104.841.01.09.02 Page 84 of 192 © Siemens, 2011


02.11 Ultrasound
Control Panel 85

Replacing the Control Panel Base 0

NOTE Use Loctite 271 for holding the HSH bolts when replacing the
control panel base.

Perform the reverse sequence of the procedure provided for the removal. Refer to the fol-
lowing figure for the cable routing and tying.

Fig. 92: Cable Ties Routing Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 85 of 192 ACUSON X300


02.11 Ultrasound
86 Gas Cylinder Replacement
8-
Overview
8Gas Cylinder Replacement

Use the instructions in this section to remove and replace the gas cylinder on the ACUSON
X300 system.

Workflow 0

Remove Gas Cylinder


Replace Gas Cylinder

Tool Requirements 0

# 2 Phillips screwdriver
Loctite 271
Metric wrench set
Tie-wraps
Wire cutters

Time Required 0

Replacement of the Gas Cylinder requires approximately 3 hours.

System Preparations 0

1. Press the Power button on the system Control Panel to power off the system.
2. Switch the Main Circuit Breaker of the system to Off.
3. Disconnect the system power cord from the system external power source.
4. Maintain a minimum of 3.5 feet (1.70 Meters) clearance while working on the system.

ACUSON X300 US11-104.841.01.09.02 Page 86 of 192 © Siemens, 2011


02.11 Ultrasound
Gas Cylinder Replacement 87

Removing the Gas Cylinder 8.1

NOTE Mark all the locations of cable ties with a pen during removal
of up down roller assembly, for quick and easy replacement.

1. Remove the monitor; see (Removing the Monitor / p. 27).


2. Remove the monitor arm covers; see (Removing the Monitor Arm Covers / p. 44).
3. Remove the control panel; see (Removing the Control Panel / p. 63).
4. Remove the right, left and front covers; see (Removing the Right Cover / p. 15),
(Removing the Left Cover / p. 12), and (Removing the Front Cover / p. 19).
5. Cut the cable ties at the marked locations.

Fig. 93: Cable Ties Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 87 of 192 ACUSON X300


02.11 Ultrasound
88 Gas Cylinder Replacement

6. Remove the (5) screws beneath the CP neck to remove the CP neck cover from the CP
neck.

Fig. 94: Remove the CP neck cover


7. Pull out the (3) cables through the Hole 1 beneath the CP neck and the Hole 2 at the
front of the CP neck.

Fig. 95: Calbe Ties Shown

ACUSON X300 US11-104.841.01.09.02 Page 88 of 192 © Siemens, 2011


02.11 Ultrasound
Gas Cylinder Replacement 89

8. Cut the cable ties at the marked locations.

Fig. 96: Cut the Cable Ties


9. Remove one of the (2) E rings on each side of the shaft using a pair of pincers. And then
pull the shaft outward to separate the gas cylinder from the main frame.

Fig. 97: Remove the E ring and Shaft

© Siemens, 2011 US11-104.841.01.09.02 Page 89 of 192 ACUSON X300


02.11 Ultrasound
90 Gas Cylinder Replacement

10. Disconnect the gas cylinder cable from the gas cylinder. Then remove the (4) HSH bolts
on the control panel base using a 6mm Allen wrench.

Fig. 98: Gas Cylinder Cable and Bolts Shown


11. Lift the control panel base and remove the gas cylinder from the main frame. Use cau-
tion when lifting the control panel.

Fig. 99: Remove the gas cylinder from the main frame

ACUSON X300 US11-104.841.01.09.02 Page 90 of 192 © Siemens, 2011


02.11 Ultrasound
Gas Cylinder Replacement 91

Replacing the Gas Cylinder 8.2

NOTE Use Loctite 271 for holding the HSH bolts when replacing the
Gas Cylinder.

Perform the reverse sequence of the procedure provided for the removal. Refer to the fol-
lowing figures for the cable routing and tying.

NOTE Orient the gas cylinder toward the gas cylinder cable for the
connection.

Fig. 100: Cable Ties Routing Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 91 of 192 ACUSON X300


02.11 Ultrasound
92 Up Down Roller Assembly Replacement
9-
Overview
9Up Down Roller Assembly Replacement

Use the instructions in this section to remove and replace the Up Down Roller Assembly
Replacement on the ACUSON X300 system.

Tool Requirements 0

• No. 2 Phillips screwdriver


• Loctite 271
• 5mm Allen wrench
• 6mm Allen wrench
• Cable Ties ( 5 EA )
• Wire cutters

Time Required 0

Replacement of the cables requires approximately 3 hours.

ACUSON X300 US11-104.841.01.09.02 Page 92 of 192 © Siemens, 2011


02.11 Ultrasound
Up Down Roller Assembly Replacement 93

Removing the Up Down Roller Assembly 9.1

NOTE Mark all the locations of cable ties with a pen during removal
of up down roller assembly, for quick and easy replacement.

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the monitor; see (Removing the Monitor / p. 27).
3. Remove the monitor arm covers; see (Removing the Monitor Arm Covers / p. 44).
4. Remove the control panel; see (Removing the Control Panel / p. 63).
5. Remove the right, left and front covers; see (Removing the Right Cover / p. 15),
(Removing the Left Cover / p. 12), and (Removing the Front Cover / p. 19).
6. Cut the cable ties at the marked locations.

Fig. 101: Cable Ties Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 93 of 192 ACUSON X300


02.11 Ultrasound
94 Up Down Roller Assembly Replacement

7. Remove the five (5) screws beneath the CP neck to remove the CP neck cover from the
CP neck.

Fig. 102: Remove the CP neck cover


8. Pull out the three (3) cables through the hole1 beneath the CP neck and the hole2 at the
front of the CP neck.

Fig. 103: Calbe Ties Shown

ACUSON X300 US11-104.841.01.09.02 Page 94 of 192 © Siemens, 2011


02.11 Ultrasound
Up Down Roller Assembly Replacement 95

9. Cut the cable ties at the marked locations.

Fig. 104: Cut the Cable Ties


10. Disconnect the gas cylinder cable from the gas cylinder. Then, remove the four (4)
HSH bolts on the control panel base, using a 6mm Allen wrench.

Fig. 105: Gas Cylinder Cable and Bolts Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 95 of 192 ACUSON X300


02.11 Ultrasound
96 Up Down Roller Assembly Replacement

11. Lift the control panel base to separate it from the main frame.

Fig. 106: Control Panel Base and Main Frame Separated


12. Remove the TI module; see (Removing the TI Module / p. 103).
13. Remove the four (4) HSH bolts beneath the up down roller assembly and the two (2)
bolts on the right and left side of the up-down roller assembly, using a 6mm Allen
wrench.

Fig. 107: Bolt Removal Shown

ACUSON X300 US11-104.841.01.09.02 Page 96 of 192 © Siemens, 2011


02.11 Ultrasound
Up Down Roller Assembly Replacement 97

14. Remove the roller outer by lifting it and cut the cable tie on the roller inner.

Fig. 108: Cut the cable ties on the Roller Inner


15. Remove the roller inner and pull the cables downward.

Fig. 109: Remove the roller inner and pull the cables downward

© Siemens, 2011 US11-104.841.01.09.02 Page 97 of 192 ACUSON X300


02.11 Ultrasound
98 Up Down Roller Assembly Replacement

Replacing the Up Down Roller Assembly 9.2

NOTE Use Loctite 271 for holding the HSH bolts when replacing the
System Cables.

Perform the reverse sequence of the procedure provided for the removal. Refer to the fol-
lowing figures for the cable routing and tying.

1. Tie the cables on the roller inner. Ensure that there is adequate cable length for the fully
raised control panel base.

Fig. 110: Tie the cables on the roller inner

ACUSON X300 US11-104.841.01.09.02 Page 98 of 192 © Siemens, 2011


02.11 Ultrasound
Up Down Roller Assembly Replacement 99

2. Route and tie the cables on the control panel base.

Fig. 111: Cable Ties Routing Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 99 of 192 ACUSON X300


02.11 Ultrasound
100 Modules and Board Replacement
10-
TI Module Assembly
10Modules and Board Replacement

TI Daughter Board (TI DL3 Board: ACUSON X300 PE) 0

Overview

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of each module requires approximately 20 minutes.

NOTE The system license information will be corrupted if multiple


boards/modules are replaced or interchanged at the same
time. A corrupt license file will result in the loss of optional
features and system imaging functionality. To minimize this
risk, only replace one board or module at a time and perform
a system reboot each time.

Removing the TI Daughter Board

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Before removing the TI Daughter board, remove all connected transducers.
3. Remove the front cover, see (Removing the Front Cover / p. 19).

ACUSON X300 US11-104.841.01.09.02 Page 100 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 101

4. Remove the (4) PH screws on the TI Daughter board.

Fig. 112: Remove the four PH screws on the TI Daughter Board


5. Pull out the TI Daughter board from the TI Module using the TI block handle.

© Siemens, 2011 US11-104.841.01.09.02 Page 101 of 192 ACUSON X300


02.11 Ultrasound
102 Modules and Board Replacement

NOTE Grasp the lower side of the TI Daughter Board at the same
time.

Fig. 113: Carefully pull the TI Daughter Board out using the TI block handle

Replacing the TI Daughter Board


Replace the TI Daughter board by following the procedure in reverse sequence.

CAUTION CAUTION: When installing the TI Daughter Board, ensure


that the module aligns properly.
¹ The connector pins may bend if the TI Daughter Board
is inserted improperly into the system.

ACUSON X300 US11-104.841.01.09.02 Page 102 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 103

TI Module 0

Overview

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of this module requires approximately 20 minutes.

NOTE In case of ACUSON X300 PE, check if the fourSight 4D Imag-


ing option is installed on the system before removing the TI
module.

Removing the TI Module

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the right and front cover, see (Removing the Front Cover / p. 19) and (Remov-
ing the Right Cover / p. 15).

© Siemens, 2011 US11-104.841.01.09.02 Page 103 of 192 ACUSON X300


02.11 Ultrasound
104 Modules and Board Replacement

3. Disconnect the frame cable from the TI module or the right side of the system.

Fig. 114: Disconnect the Frame Cable from the TI Module, from the right side of the system

ACUSON X300 US11-104.841.01.09.02 Page 104 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 105

4. Remove the (3) PWH and (2) TH screws that secure the TI module to the bracket.

Fig. 115: Remove the five screws from the TI Module

© Siemens, 2011 US11-104.841.01.09.02 Page 105 of 192 ACUSON X300


02.11 Ultrasound
106 Modules and Board Replacement

5. Hold the handle located on the edge of the TI Module and pull the board out away from
the system.

Fig. 116: Remove the TI Module carefully using the handle on the side

In Case of the FourSight 4D Imaging Option Installed System

NOTE Follow this procedure if the fourSight 4D Imaging option


installed system needs to replace only the TI module without
the replacement of Wobbler Boards Assembly.

ACUSON X300 US11-104.841.01.09.02 Page 106 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 107

1. Remove the (5) screws on the TI front cover and remove the TI front cover from the TI
module.

Fig. 117: TI Front Cover with 5 Screws Shown


2. Remove the (6) screws on the Wobbler boards assembly.

Fig. 118: Wobbler Board Assembly with Screws Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 107 of 192 ACUSON X300


02.11 Ultrasound
108 Modules and Board Replacement

3. Disconnect the Wobbler boards assembly from the connector located on the TI main
board.

Fig. 119: Wobbler Boards Assembly with Connector Shown


4. Remove the (5) screws from the cover of the new TI module and remove the front
cover.

ACUSON X300 US11-104.841.01.09.02 Page 108 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 109

5. Place the existing Wobbler boards assembly on the upper side of the new TI module
and connect it to the connector of the TI main board.

Fig. 120: Wobbler Boards Assembly in Place on TI Module


6. Install and tighten the (6) screws on the Wobbler boards assembly.

Fig. 121: Wobbler Board Assembly with Screws Shown


7. Place the TI front cover on the TI module and install and tighten the 5 screws.

© Siemens, 2011 US11-104.841.01.09.02 Page 109 of 192 ACUSON X300


02.11 Ultrasound
110 Modules and Board Replacement

CAUTION When installing the TI front cover, ensure that the thermal
pads and heatsinks are attached properly on the back side of
the TI front cover.
¹ Always check to ensure the thermal pads and heat-
sinks are properly attached for installing the front
cover.

Fig. 122: Front (right) and Back Side (left) of TI Front Cover Shown

Replacing the TI Module


Replace the TI Module using the reverse sequence of the procedures provided for removal.

CAUTION CAUTION: When installing the TI module, ensure that the


module aligns properly.
¹ The connector pins may bend if the TI Module is
inserted improperly into the system.

Verifying the Operation of the TI Module

1. Connect a transducer to the system.


2. Set the system MAIN switch to ON.
3. Power on the system, and wait for it to boot up completely and without errors.
4. Verify a transducer works in the right transducer port.
a) Connect a transducer to the right transducer port.
b) Verify the image updates and the transducer selection displays on the monitor.
5. Press Freeze.

ACUSON X300 US11-104.841.01.09.02 Page 110 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 111

6. Verify a transducer works in the center transducer port.


a) Disconnect the transducer from the center transducer port and connect it to the right
transducer port.
b) Verify the image updates and the transducer selection displays on the monitor
7. Press Freeze.
8. Verify a transducer works in the left transducer port.
a) Disconnect the transducer from the center transducer port and connect it to the left
transducer port.
b) Verify the image updates and the transducer selection displays on the monitor.
9. Press the Preset key to display the Preset Main Menu Screen.
10. Select Service.
11. Enter the Service password and select GO.
12. Select Configuration.
13. Select Hardware from the list.
14. Verify the board name and system serial number from the Configuration screen.

Fig. 123: Verify the board name and system serial number from the Configuration screen

© Siemens, 2011 US11-104.841.01.09.02 Page 111 of 192 ACUSON X300


02.11 Ultrasound
112 Modules and Board Replacement

Replacement of the Wobbler Boards Assembly 0

Overview

Tool Requirements
Number 2 Phillips screwdriver

Time Required
Replacement of this module requires approximately 30 minutes.

Removing the Wobbler Boards Assembly

NOTE Follow this procedure if the fourSight 4D Imaging option


installed system needs to replace only the Wobbler Boards
Assembly without the replacement of the whole TI module.

1. Power off the system and set the MAIN switch to OFF. Leave the power cord con-
nected to the wall outlet.
2. Remove the TI module, refer to (Removing the TI Module / p. 103).
3. Remove the (5) screws on the TI front cover, and remove the TI front cover from the TI
module.

Fig. 124: TI Front Cover with 5 Screws Shown

ACUSON X300 US11-104.841.01.09.02 Page 112 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 113

4. Remove the (6) screws on the Wobbler boards assembly.

Fig. 125: Wobbler Board Assembly with Screws Shown


5. Disconnect Wobbler boards assembly from the connector located on the TI main board.

Fig. 126: Wobbler Boards Assembly with Connector Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 113 of 192 ACUSON X300


02.11 Ultrasound
114 Modules and Board Replacement

Replacing the Wobbler Boards Assembly


Replace the Wobbler Boards Assembly using the reverse sequence of
1. Before install the new Wobbler boards assembly, remove both liners on the thermal
pads located on the bottom side of the Wobbler boards assembly.

Fig. 127: Thermal Pad Liners on Wobbler Boards Assy. Shown

ACUSON X300 US11-104.841.01.09.02 Page 114 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 115

CAUTION When installing the TI front cover, ensure that the thermal
pads and heatsinks are attached properly on the back side of
the TI front cover.
¹ Always check to ensure the thermal pads and heat-
sinks are properly attached for installing the front
cover.

Fig. 128: Heatsinks and Thermal Pads on TI Front Cover

© Siemens, 2011 US11-104.841.01.09.02 Page 115 of 192 ACUSON X300


02.11 Ultrasound
116 Modules and Board Replacement

Electronic Modules and Boards 10.1

Electronic Module and Boards 0

Overview

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of each module requires approximately 20 minutes.

NOTE The system license information will be corrupted if multiple


boards/modules are replaced or interchanged at the same
time. A corrupt license file will result in the loss of optional
features and system imaging functionality. To minimize this
risk, only replace one board or module at a time and perform
a system reboot each time.

Removing the Electronic Module and Boards

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the right cover, see (Removing the Right Cover / p. 15).
3. Remove the (2) TH and PWH screws to remove the lower bracket securing the elec-
tronic module to the main frame.

Fig. 129: Remove the screws and remove the lower bracket from the electronic module

ACUSON X300 US11-104.841.01.09.02 Page 116 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 117

4. Remove the (2) PWH screws on the electronic module cover.

Fig. 130: Remove the two PWH screws on the electronic module cover

© Siemens, 2011 US11-104.841.01.09.02 Page 117 of 192 ACUSON X300


02.11 Ultrasound
118 Modules and Board Replacement

5. Grasp the handle and pull the cover out away from the system.

Fig. 131: Remove the electronic module cover by pulling the handle out away from the system
The Electronic Module contain the following boards:
- TR Board: located in the left-side of Electronic Module
- RC Board: located in the center of Electronic Module
- BE Board: located in the right-side of Electronic Module

ACUSON X300 US11-104.841.01.09.02 Page 118 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 119

NOTE BE board contains a CR2032 3V Battery. It is recommended


to replace the BE coin battery every 2 years during system
preventive maintenance.

Fig. 132: Board location in the Electronic Module

© Siemens, 2011 US11-104.841.01.09.02 Page 119 of 192 ACUSON X300


02.11 Ultrasound
120 Modules and Board Replacement

6. Lift up on the ejectors and carefully pull out the board.

Fig. 133: Lift up on the ejectors and carefully pull the board out

Replacing the Electronic Module and Boards

1. Replace the boards and the Electronic Module using the reverse sequence of the pro-
cedures provided for board removal.
2. During the installation, ensure the board fits to the two brackets located on the top and
the bottom of the card cage. Also, make sure to fix the boards into the rail inside of Elec-
tronic Module.
3. Follow the same sequence for installing each Electronic Module; TR board, RC board
and BE board.

Verifying the Operation of the Electronic Module

1. Connect a transducer to the system.


2. Set the MAIN switch to ON.
3. Power on the system, and wait for it to boot up completely.
4. Verify an image is displayed on the monitor.
5. Press Presets to display the Preset Main Menu Screen.
6. Select Service.

ACUSON X300 US11-104.841.01.09.02 Page 120 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 121

7. Enter the Service password and select GO.


8. Select Configuration.
9. Select Hardware from the list.
10. Verify the board name and serial number on the screen.

ETX Board 0

Overview

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of each module requires approximately 20 minutes.

NOTE After ETX board replacement the IP address can be lost from
SW version 4.0. If no IP address is set, enter 192.168.0.1
(default) or the current IP address into the system manually.
Refer to (US11-102.843.01 / Configuring Network Devices).

Removing the ETX Board

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the right cover, see (Removing the Right Cover / p. 15).
3. Remove the electronic module cover from the system, see (Removing the Electronic
Module and Boards / p. 116).
4. Remove the BE board, see (Removing the Electronic Module and Boards / p. 116).

© Siemens, 2011 US11-104.841.01.09.02 Page 121 of 192 ACUSON X300


02.11 Ultrasound
122 Modules and Board Replacement

5. On the rear of the BE board, remove the (4) screws that attach the ETX board to the BE
board.

Fig. 134: Remove the four screws on the ETX board


6. Turn the BE board over and disconnect the cable from the ETX board.

Fig. 135: Disconnect the cable from the ETX board


7. Remove the ETX board from the BE board.

ACUSON X300 US11-104.841.01.09.02 Page 122 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 123

NOTE When removing the ETX board, be careful of the four connec-
tors.

Fig. 136: Remove the ETX board from the BE board

Replacing the ETX Board

1. Replace the ETX board using the reverse sequence of the procedures provided for
board removal.

NOTE Ensure that the board seats and aligns correctly with the four
connectors on the bottom of the BE board.

Verifying the Operation of the ETX Board

1. Connect a transducer to the system.


2. Set the MAIN switch to ON.
3. Power on the system, and wait for it to boot up completely.
4. Verify an image is displayed on the monitor.
5. Press Presets to display the Preset Main Menu Screen.
6. Select Service.
7. Enter the Service password and select GO.

© Siemens, 2011 US11-104.841.01.09.02 Page 123 of 192 ACUSON X300


02.11 Ultrasound
124 Modules and Board Replacement

8. Select Configuration.
9. Select Hardware from the list.
10. Verify the board name and serial number on the screen.

RC Daughter Board 0

Overview

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of this module requires approximately 20 minutes.

Removing the RC Daughter Board

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the right and electronic module covers, see (Removing the Right
Cover / p. 15) and (Removing the Electronic Module and Boards / p. 116).
3. Remove the RC board from the system, see instructions (Removing the Electronic
Module and Boards / p. 116).

ACUSON X300 US11-104.841.01.09.02 Page 124 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 125

4. Remove the (2) screws that attach the RC daughter board to the RC board.

Fig. 137: Remove the two screws on the RC board (ACUSON X300)

Fig. 138: Remove the two screws on the RC board (ACUSON X300 PE)

© Siemens, 2011 US11-104.841.01.09.02 Page 125 of 192 ACUSON X300


02.11 Ultrasound
126 Modules and Board Replacement

5. Carefully remove the RC daughter board from the RC board.

Fig. 139: Remove the RC daughter board from the RC Board

Replacing the RC Daughter Board

1. Replace the board using the reverse sequence of the procedures provided for board
removal.

CAUTION CAUTION: Be cautious when you install the RC daughter


board into the system.
¹ The connector pins may bend if the board is inserted
improperly into the system.

ACUSON X300 US11-104.841.01.09.02 Page 126 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 127

Removing and Replacing the Hard Disk Drive (HDD) 10.2

Overview 0

Tool Requirements
Number 2 Phillips screwdriver

Time Required
Replacement of the HDD requires approximately 30 minutes.

Removing the HDD 0

Follow the procedure described below for removing the HDD from the system.

1. Power off the system and set the MAINS switch to OFF; leave the power cord con-
nected to the wall outlet.

WARNING The ACUSON X300 system contains hazardous voltages that


can range from 90 VAC to 246 VAC and cause serous injury.
¹ Only authorized service personnel are permitted to
remove protective covers and access the energized
components of this system.

2. Remove right-side panel, left-side panel, and rear panel, see (System Covers / p. 10).

© Siemens, 2011 US11-104.841.01.09.02 Page 127 of 192 ACUSON X300


02.11 Ultrasound
128 Modules and Board Replacement

3. Remove the HDD cover by loosening the (4) screws.

Fig. 140: Removing the HDD Cover

ACUSON X300 US11-104.841.01.09.02 Page 128 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 129

4. Disconnect the (2) cables that tie the HDD into the I/O module.

Fig. 141: Disconnecting the two cables (E-IDE HDD, ACUSON X300)

© Siemens, 2011 US11-104.841.01.09.02 Page 129 of 192 ACUSON X300


02.11 Ultrasound
130 Modules and Board Replacement

Fig. 142: Disconnecting the cables. (S-ATA HDD, ACUSON X300 PE)

ACUSON X300 US11-104.841.01.09.02 Page 130 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 131

5. Remove the (4) screws that secure the HDD cover into the HDD.

Fig. 143: Removing the screws from the HDD cover

Replacing the HDD 0

Replace the HDD using the reverse sequence of steps for removing the HDD.

Verifying the Operation of the System 0

1. Connect power cord to system.


2. Power on the system and verify the system is operational.

NOTE After replacing of the HDD, the system needs to have a fac-
tory installation. Refer to (US11-101.816.01 / Installing the
System Software).

© Siemens, 2011 US11-104.841.01.09.02 Page 131 of 192 ACUSON X300


02.11 Ultrasound
132 Modules and Board Replacement

I/O Module 10.3

Overview 0

Tool Requirements
• Number 2 Phillips screwdriver

Time Required
Replacement of each module requires approximately 30 minutes.

Removing the I/O Module 0

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the left, right and rear covers, see (Removing the Left Cover / p. 12), (Remov-
ing the Right Cover / p. 15) and (Removing the Rear Cover / p. 17).

ACUSON X300 US11-104.841.01.09.02 Page 132 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 133

3. Remove the cables from the I/O Module on the rear of the system.

Fig. 144: Disconnect the cables from the I/O Module

© Siemens, 2011 US11-104.841.01.09.02 Page 133 of 192 ACUSON X300


02.11 Ultrasound
134 Modules and Board Replacement

Fig. 145: Remove the cables connected to the I/O module.

NOTE In 2.0 systems and over, the I/O module has an additional
USB port for the connection of BW PRT USB frame cable.

ACUSON X300 US11-104.841.01.09.02 Page 134 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 135

4. Remove the (3) cables from the I/O Module on the right side of the system.

Fig. 146: Disconnect the three cables from the I/O Module from the right side of the system

© Siemens, 2011 US11-104.841.01.09.02 Page 135 of 192 ACUSON X300


02.11 Ultrasound
136 Modules and Board Replacement

Fig. 147: Disconnect and remove the three Frame connector cables

ACUSON X300 US11-104.841.01.09.02 Page 136 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 137

5. Remove the (2) PWH screws that attach the I/O module to the electronic module.

Fig. 148: Remove the two PWH screws that connect the electronic module to the I/O module
6. Remove the screws to remove the lower bracket that secure the Electronic Module and
the I/O Module to the frame.

Fig. 149: Remove the four screws from the lower bracket
7. Using the handle, pull the I/O Module outward and remove the Module from the system.

© Siemens, 2011 US11-104.841.01.09.02 Page 137 of 192 ACUSON X300


02.11 Ultrasound
138 Modules and Board Replacement

CAUTION WARNING: Do not pinch any cables while pulling the I/O
module out from the system.
¹ Pinching a cable may cause damage to the system or
create operational errors.

Fig. 150: Pull the I/O module out carefully and remove it from the system

Replacing the I/O Module 0

1. Replace I/O Module using the reverse sequence of the procedures provided for
removal.

CAUTION CAUTION: Be cautious when you install I/O module into the
system.
¹ The connector pins may bend if the I/O Module is
inserted improperly into the system.

ACUSON X300 US11-104.841.01.09.02 Page 138 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 139

2. Make sure that the connectors attached on the I/O module fit tightly to the connectors
attached on the system main frame.

Replacing of the I/O Sub Board 0

Overview

Tool Requirement
Number 2 Phillips screwdriver
American standard screwdriver

Time Required
Replacement of each module requires approximately 40 minutes.

Removing the I/O Sub Board

1. Power off the system and set MAIN switch to OFF, leave the power cord connected to
the wall outlet.
2. Remove the I/O module from the system, Refer to (Removing the I/O Module / p. 132).
3. Remove the (6) screws on the I/O sub board panel.

Fig. 151: I/O Sub Board with Screws Shown


4. Remove the I/O sub board form the I/O module.

© Siemens, 2011 US11-104.841.01.09.02 Page 139 of 192 ACUSON X300


02.11 Ultrasound
140 Modules and Board Replacement

NOTE Use an American standard shape screwdriver to life up the


I/O sub board from the I/O module through the holes located
on each side of the I/O sub board panel.

Fig. 152: Lift Areas for Removal of I/O Sub Board from I/O Module

Replacing the I/O Sub Board


Replace I/O sub board using the reverse sequence of the procedures provided for removal.

CAUTION Use caution when installing the I/O sub board into the I/O
module.
¹ The connector pins may become bent if the I/O sub
board is inserted improperly into the I/O module.

ACUSON X300 US11-104.841.01.09.02 Page 140 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 141

Replacing Backplane with Chassis Assembly 10.4

Overview 0

Tool Requirements
Number 2 Phillips screwdriver

Time Required
Replacement of this module requires approximately 30 minutes.

Removing the Backplane with Chassis Assembly 0

1. Power off the system and set the MAIN switch to OFF. Leave the power cord con-
nected to the wall outlet.

WARNING Shock Hazard


The X300 systems contain hazardous voltages that can rang
from 90VAC to 264VAC and can cause serious injury or
death.
¹ Always ensure the main switch is in the OFF position
before working on the systems.

2. Remove all connected transducers from the system.


3. Remove the TI daughter board and TI module. Refer to (Removing the TI Daughter
Board / p. 100) and (Removing the TI Module / p. 103).
4. Remove the DC/DC power supply, refer to (Removing the DC/DC Power
Supply / p. 168).
5. Remove the I/O module, refer to (Removing the I/O Module / p. 132).
6. Remove the TR, RC and BE boards, refer to (Removing the Electronic Module and
Boards / p. 116).

© Siemens, 2011 US11-104.841.01.09.02 Page 141 of 192 ACUSON X300


02.11 Ultrasound
142 Modules and Board Replacement

7. Remove the (4) screws located on the bottom side of the chassis.

Fig. 153: Front Side of Chassis Rear Side of Chassis


8. Remove the (2) screws located on the upper side of the chassis.

Fig. 154: Upper Side of Chassis with Screws Shown

ACUSON X300 US11-104.841.01.09.02 Page 142 of 192 © Siemens, 2011


02.11 Ultrasound
Modules and Board Replacement 143

9. Remove the backplane with chassis assembly from the main frame by pulling it out
toward the left side.

Fig. 155: Removal Direction of Backplane Assembly Shown

Replacing the Backplane with Chassis Assembly 0

• Replace the backplane with chassis assembly using the reverse sequence of the
removal procedure, refer to (Removing the Backplane with Chassis Assembly / p. 141).

CAUTION Possible Equipment Damage


The connector pins may get bent if the backplane with chas-
sis assembly is inserted improperly into the system.
¹ When installing the backplane with chassis assembly,
ensure that the module aligns properly.

© Siemens, 2011 US11-104.841.01.09.02 Page 143 of 192 ACUSON X300


02.11 Ultrasound
144 CD/DVD-R/RW and ECG
11-
Overview
11CD/DVD-R/RW and ECG

Tool Requirements 0

Number 2 Phillips screwdriver

Time Required 0

Replacement of CD/DVD-R/RW and ECG requires approximately 20 minutes.

ACUSON X300 US11-104.841.01.09.02 Page 144 of 192 © Siemens, 2011


02.11 Ultrasound
CD/DVD-R/RW and ECG 145

Removing the CD/DVD-R/RW and ECG 11.1

NOTE The ECG module is the same as Physio module.

1. Power off the system and switch the MAIN switch to OFF; leave the power cord con-
nected to the wall outlet.
2. Remove the air filter from the left cover, see (Fig. 3 / p. 12).
3. Remove the left, right and rear covers (see (Removing the Left Cover / p. 12), (Remov-
ing the Right Cover / p. 15), (Removing the Rear Cover / p. 17)).
4. From the right side of the system, disconnect the signal and power cables attached to
the CD-R/RW and ECG module.

Fig. 156: Disconnecting the Cables

© Siemens, 2011 US11-104.841.01.09.02 Page 145 of 192 ACUSON X300


02.11 Ultrasound
146 CD/DVD-R/RW and ECG

5. Remove the (4) PWH screws from the rear side of the system.

Fig. 157: Remove the PHW screws on the rear side of the system
6. Remove the (2) PHW screws from the CD/DVD-R/RW and ECG.

Fig. 158: Remove the screws from the CD/DVD-R/RW and ECG

ACUSON X300 US11-104.841.01.09.02 Page 146 of 192 © Siemens, 2011


02.11 Ultrasound
CD/DVD-R/RW and ECG 147

7. Remove the CD/DVD-R/RW and ECG assembly.

Fig. 159: Remove the CD/DVD-R/RW and ECG assembly from the system

NOTE Steps 8 through 10 apply to the ACUSON X300 PE r4.0 and


below systems only.

8. Disconnect the signal cable and power cable that connect the CD/DVD-R/RW and
ECG.
9. Disconnect the cables from the CD/DVD-R/RW interboard.

Fig. 160: Disconnect the power and signal cables between the CD/DVD-R/RW and ECG

© Siemens, 2011 US11-104.841.01.09.02 Page 147 of 192 ACUSON X300


02.11 Ultrasound
148 CD/DVD-R/RW and ECG

10. Remove the four PH screws and remove the ECG module.

Fig. 161: Remove the ECG module


The disassembled ECG module should look like the following.

Fig. 162: ECG Module

NOTE Steps 11 through 14 apply to ACUSON X300 PE r5.0 and


higher systems only.

ACUSON X300 US11-104.841.01.09.02 Page 148 of 192 © Siemens, 2011


02.11 Ultrasound
CD/DVD-R/RW and ECG 149

11. Disconnect the signal and power cables from the CD/DVD-R/RW.

Fig. 163: Drive Signal and Power Cables Shown


12. Disconnect the Physio Inner Cable from the Advanced ECG Module.

Fig. 164: Advanced ECG Module with Physio Inner Cable Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 149 of 192 ACUSON X300


02.11 Ultrasound
150 CD/DVD-R/RW and ECG

13. Remove the (4) screws the hold the bottom of the ECG Module bracket.

Fig. 165: ECG Module Bracket - Screws Shown


14. Remove the advanced ECG Module from the bracket.

Fig. 166: ECG Module Removed from Bracket

ACUSON X300 US11-104.841.01.09.02 Page 150 of 192 © Siemens, 2011


02.11 Ultrasound
CD/DVD-R/RW and ECG 151

Replacing the CD/DVD-R/RW and ECG 11.2

To replace the CD/DVD-R/RW and ECG, perform the steps of removing in reverse
sequence.

CAUTION CAUTION: Use fasteners of the correct length.


¹ Failure to do so may cause damage to underlying
equipment.

NOTE Before the replacement, make sure that the jumper is set to
SLAVE mode in ACUSON X300 system and MASTER mode in
ACUSON X300 PE system. This is available at the rear side of
the CD/DVD-R/RW.

Fig. 167: Jumper Setting

© Siemens, 2011 US11-104.841.01.09.02 Page 151 of 192 ACUSON X300


02.11 Ultrasound
152 Air Filter
12-
Overview
12Air Filter

Use the following instructions to remove and replace the air filter of the system.

Tool Requirements 0

• None

Time Required 0

Replacement of the Air Filter takes approximately 5 minutes.

ACUSON X300 US11-104.841.01.09.02 Page 152 of 192 © Siemens, 2011


02.11 Ultrasound
Air Filter 153

Removing the Air Filter 12.1

1. Remove the air filter from the left side of the system.

Fig. 168: Remove the air filter on the left side of the system

NOTE Make sure the air filter is cleaned. To replace the air filter,
order the field replacement unit, Kit, Air Filter.

© Siemens, 2011 US11-104.841.01.09.02 Page 153 of 192 ACUSON X300


02.11 Ultrasound
154 Air Filter

Replacing the Air Filter 12.2

To replace the Air Filter; follow in reverse sequence the procedures for removal. The sys-
tem provides a warning message to replace the air filter every six months, see
(US11-101.831.01 / Planned Maintenance).

ACUSON X300 US11-104.841.01.09.02 Page 154 of 192 © Siemens, 2011


02.11 Ultrasound
AC/DC Power Supply 155
13-
Overview
13AC/DC Power Supply

Workflow 0

1. Remove the AC/DC Power Supply


2. Replace the AC/DC Power Supply
3. Verification

Tool Requirements 0

No. 2 Phillips screwdriver

Time Required 0

Replacement of the AC/DC Power Supply requires approximately 20 minutes.

© Siemens, 2011 US11-104.841.01.09.02 Page 155 of 192 ACUSON X300


02.11 Ultrasound
156 AC/DC Power Supply

Removing the AC/DC Power Supply 13.1

1. Power off the system and set the MAIN switch to OFF; unplug the power cord from the
wall outlet.
2. Remove the (4) PWH screws on the AC/DC connector cover, located on the front side
of the system.

Fig. 169: Remove the screws from the AC/DC connector cover
3. Remove the AC/DC connector cover.

Fig. 170: Remove the AC/DC connector cover

ACUSON X300 US11-104.841.01.09.02 Page 156 of 192 © Siemens, 2011


02.11 Ultrasound
AC/DC Power Supply 157

4. Disconnect the cables located on the AC/DC power supply.

Fig. 171: Disconnect the cables from the AC/DC power supply

NOTE NOTE: OEM outlet cable 1, OEM outlet cable 2, and monitor
outlet cable can be attached to any of the three connectors.

NOTE If QuickStart option is installed, disconnect the AC to Battery


power cable from BAT1 connector (In ACUSON X300 PE sys-
tem).

5. Remove the (4) PWH screws from the AC/DC power supply frame (two on each side).

Fig. 172: Remove the screws from the AC/DC power supply frame

© Siemens, 2011 US11-104.841.01.09.02 Page 157 of 192 ACUSON X300


02.11 Ultrasound
158 AC/DC Power Supply

6. Pull the AC/DC power supply out toward the rear of the system.

Fig. 173: Pull the AC/DC power supply out toward the rear of the system

ACUSON X300 US11-104.841.01.09.02 Page 158 of 192 © Siemens, 2011


02.11 Ultrasound
AC/DC Power Supply 159

Replacing the AC/DC Power Supply 13.2

WARNING Before installing the AC/DC power supply, move all cables
aside.
¹ Failure to do so may cause damage to the cables.

1. Configure the voltage setting. The AC power tray is pre-configured to 230 V. If you are
installing the AC power try on a system with different power requirements, such as 110
V or 115 V, reconfigure the voltage:
- Loosen one screw of the sheet metal cover, located at the rear of the AC Power Tray,
to get access to the voltage switch.
- Set the voltage switch position.
- Fasten the screw of the sheet metal cover to hide the voltage switch.
2. Replace the AC power supply. Use the revers procedures described in (Removing the
AC/DC Power Supply / p. 156). Refer to the following figure for the cable connections.

Fig. 174: Cable connection of the AC/DC power supply

NOTE If the QuickStart option is installed, connect the AC To Bat-


tery power cable to BAT1 connector in (ACUSON X300 PE
system).

© Siemens, 2011 US11-104.841.01.09.02 Page 159 of 192 ACUSON X300


02.11 Ultrasound
160 AC/DC Power Supply

Verifying the Operation of the AC/DC Power Supply 13.3

1. Connect a transducer to the system.


2. Plug the system power cord into the wall outlet and set the MAIN switch to ON.
3. Power on the system, and wait for it to boot up completely and without errors.
4. Verify that an image is displayed on the monitor.
5. Use the Diagnostic LED Indicator to verify operation.

ACUSON X300 US11-104.841.01.09.02 Page 160 of 192 © Siemens, 2011


02.11 Ultrasound
Replace the Battery Pack 161
14-
Overview
14Replace the Battery Pack

This procedure describes the steps required to remove and replace the battery pack on
system with the QuickStart option installed (ACUSON X300 PE system 6.0 and above.

Workflow 0

1. Remove the battery pack


2. Replace the battery pack
3. Verification

Tool Requirements 0

No. 2 Phillips screwdriver

Time Required 0

Replacement of the Battery Pack requires approximately 20 minutes.

© Siemens, 2011 US11-104.841.01.09.02 Page 161 of 192 ACUSON X300


02.11 Ultrasound
162 Replace the Battery Pack

Removing the Battery Pack 14.1

1. Power off the system and set the MAIN switch to OFF; unplug the power cord from the
wall outlet.
2. Remove the (4) PWH screws on the AC/DC connector cover, located on the front side
of the system.

Fig. 175: Remove the screws from the AC/DC connector cover
3. Remove the AC/DC connector cover.

Fig. 176: Remove the AC/DC connector cover

ACUSON X300 US11-104.841.01.09.02 Page 162 of 192 © Siemens, 2011


02.11 Ultrasound
Replace the Battery Pack 163

4. Disconnect the AC to Battery power cable from the BAT1 connector.

Fig. 177: Disconnect BAT1 Connection Shown

Fig. 178: Disconnect the cables from the AC/DC power supply

© Siemens, 2011 US11-104.841.01.09.02 Page 163 of 192 ACUSON X300


02.11 Ultrasound
164 Replace the Battery Pack

5. Remove the (4) PWH screws from the battery cover (two on each side).

Fig. 179: Battery Cover Screws Shown


6. Pull the Battery Pack out toward the rear of the system.

Fig. 180: Battery Pack Removed

ACUSON X300 US11-104.841.01.09.02 Page 164 of 192 © Siemens, 2011


02.11 Ultrasound
Replace the Battery Pack 165

Replacing the Battery Pack 14.2

WARNING Before installing the Battery Pack and/or AC/DC power sup-
ply, move all cables aside.
¹ Failure to do so may cause damage to the cables.

1. Configure the voltage setting. The AC power tray is pre-configured to 230 V. If you are
installing the AC power try on a system with different power requirements, such as 110
V or 115 V, reconfigure the voltage:
- Loosen one screw of the sheet metal cover, located at the rear of the AC Power Tray,
to get access to the voltage switch.
- Set the voltage switch position.
- Fasten the screw of the sheet metal cover to hide the voltage switch.
2. Replace the Battery Pack. Follow the steps for Removing the Battery Pack section in
reverse.

© Siemens, 2011 US11-104.841.01.09.02 Page 165 of 192 ACUSON X300


02.11 Ultrasound
166 Replace the Battery Pack

Verifying the Operation of the AC/DC Power Supply 14.3

1. Connect a transducer to the system.


2. Plug the system power cord into the wall outlet and set the MAIN switch to ON.
3. Power on the system, and wait for it to boot up completely and without errors.
4. Verify that an image is displayed on the monitor.
5. Use the Diagnostic LED Indicator to verify operation.

NOTE QuickStart will occur within 30 seconds from start up to live


imaging from sleep mode.

ACUSON X300 US11-104.841.01.09.02 Page 166 of 192 © Siemens, 2011


02.11 Ultrasound
DC/DC Power Supply 167
15-
Overview
15DC/DC Power Supply

Tool Requirements 0

No. 2 Phillips screwdriver

Time Required 0

Replacement of the DC/DC Power Supply requires approximately 20 minutes.

© Siemens, 2011 US11-104.841.01.09.02 Page 167 of 192 ACUSON X300


02.11 Ultrasound
168 DC/DC Power Supply

Removing the DC/DC Power Supply 15.1

1. Power off the system and set the MAIN switch to OFF; unplug power cord from the wall
outlet.
2. Remove the left cover, (Removing the Left Cover / p. 12).
3. Remove the six (6) PWH screws on the DC/DC power supply.

Fig. 181: Remove the screws from the DC/DC power supply
4. Disconnect the two (2) cables on the right side of the DC Power Supply.

Fig. 182: Disconnect the cables on the right side of the DC/DC power supply

ACUSON X300 US11-104.841.01.09.02 Page 168 of 192 © Siemens, 2011


02.11 Ultrasound
DC/DC Power Supply 169

5. Grasp the two (2) holders and pull outward to remove the DC/DC Power Supply from
the system.

Fig. 183: Remove the DC/DC power supply

© Siemens, 2011 US11-104.841.01.09.02 Page 169 of 192 ACUSON X300


02.11 Ultrasound
170 DC/DC Power Supply

Replacing the DC/DC Power Supply 15.2

1. Install the DC/DC Power Supply using the reverse sequence of the replacement proce-
dures.
2. Make sure that the DC/DC Power Supply connectors are tightly attached to the connec-
tor on Electronic module.

ACUSON X300 US11-104.841.01.09.02 Page 170 of 192 © Siemens, 2011


02.11 Ultrasound
DC/DC Power Supply 171

Verifying the Operation of the DC/DC Power Supply 15.3

1. Connect a transducer to the system.


2. Plug the system power cord into the wall outlet.
3. Set the system MAIN switch to ON.
4. Power on the system, and wait for it to boot up completely.
5. Verify an image is displayed on the monitor.
6. Check the diagnostic LEDs to verify the operation of the power supply.

© Siemens, 2011 US11-104.841.01.09.02 Page 171 of 192 ACUSON X300


02.11 Ultrasound
172 AC/DC Power Fan
16-
Overview
16AC/DC Power Fan

Workflow 0

1. Remove the AC/DC Power Fan


2. Replace the AC/DC Power Fan
3. Verification

Tool Requirements 0

• No. 2 Phillips screwdriver


• No. 4 Allen Wrench

Time Required 0

Replacement of the AC/DC Power Fan requires approximately 20 minutes.

ACUSON X300 US11-104.841.01.09.02 Page 172 of 192 © Siemens, 2011


02.11 Ultrasound
AC/DC Power Fan 173

Removing the AC/DC Power Fan 16.1

1. Power off the system and set the MAIN switch to OFF.
2. Unplug the power cord from the wall outlet.
3. Remove the AC/DC Power Supply, see (Removing the AC/DC Power Supply / p. 156).
4. Remove the (6) PH screws that attach the AC/DC power fan to the supply frame.
- There are three screws on each side of the AC/DC power fan.
- Use a 4 mm Allen wrench to remove the screws

WARNING When the (6) screws securing the frame are removed, the
frame can fall free. Securely hold the frame in place to pre-
vent the frame from falling and being damaged.
¹ Hold the frame securely in place.

Fig. 184: Remove the six screws from the AC/DC power supply frame - three on each side
5. Remove the cables and cable-tie from the AC Power Fan.

© Siemens, 2011 US11-104.841.01.09.02 Page 173 of 192 ACUSON X300


02.11 Ultrasound
174 AC/DC Power Fan

6. Remove the AC/DC power fan assembly from the system.

Fig. 185: Remove the AC/DC power fan assembly (ACUSON X300 system)

Fig. 186: Remove the AC/DC power fan assembly (ACUSON X300 PE system)

ACUSON X300 US11-104.841.01.09.02 Page 174 of 192 © Siemens, 2011


02.11 Ultrasound
AC/DC Power Fan 175

Replacing the AC/DC Power Fan 16.2

To replace the AC/DC Power Fan; follow in reverse sequence the procedures of removal.

© Siemens, 2011 US11-104.841.01.09.02 Page 175 of 192 ACUSON X300


02.11 Ultrasound
176 AC/DC Power Fan

Verifying the Operation of the AC/DC Power Fan 16.3

1. Connect a transducer to the system.


2. Set the system MAIN switch to ON.
3. Power on the system and wait for it to boot up completely without any error messages.
4. Verify that the fan is running.

ACUSON X300 US11-104.841.01.09.02 Page 176 of 192 © Siemens, 2011


02.11 Ultrasound
Transducer Holders 177
17-
Overview
17Transducer Holders

Tool Requirements 0

None

Time Required 0

Replacement of the transducer holders requires approximately 5 minutes.

© Siemens, 2011 US11-104.841.01.09.02 Page 177 of 192 ACUSON X300


02.11 Ultrasound
178 Transducer Holders

Removing the Transducer Holder 17.1

Grasp the rear side of the transducer holder, and pull downward to remove.

Fig. 187: Remove the transducer holders

ACUSON X300 US11-104.841.01.09.02 Page 178 of 192 © Siemens, 2011


02.11 Ultrasound
Transducer Holders 179

Replacing the Transducer Holder 17.2

1. Firmly press the transducer holder downward.


2. Verify the holder locks into position.

CAUTION Verify that the transducer holder locks into place on the side
of the Control Panel.
¹ Failure to do so may cause the transducer holder to
disengage from the system, dropping the transducer to
the floor and possibly damaging the transducer.

© Siemens, 2011 US11-104.841.01.09.02 Page 179 of 192 ACUSON X300


02.11 Ultrasound
180 Cables Replacement
18-
Overview
18Cables Replacement

Use the instructions in this section to remove and replace the cables on the system.

Tool Requirements 0

• No. 2 Phillips screwdriver


• Loctite 271
• 6mm Allen wrench
• 5mm Allen wrench
• Cable Ties (12 each)
• Wire cutters

Time Required 0

Replacement of the cables requires approximately 4 hours.

ACUSON X300 US11-104.841.01.09.02 Page 180 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 181

Removing the Cables 18.1

1. Power off the system and set the MAIN switch to OFF; leave the power cord connected
to the wall outlet.
2. Remove the monitor; see (Removing the Monitor / p. 27).
3. Remove the monitor arm covers; see (Removing the Monitor Arm Covers / p. 44).
4. Remove the control panel; see(Removing the Control Panel / p. 63).
5. Remove the right, left and front covers; see (Removing the Right Cover / p. 15),
(Removing the Left Cover / p. 12), and (Removing the Front Cover / p. 19).
6. Remove the rear cover, the top cover and the CP rear cover; see (Removing the Rear
Cover / p. 17), (Removing the Top Cover / p. 21), and (Removing the CP Rear
Cover / p. 23).
7. Cut the cable ties at the marked locations.

Fig. 188: Cable Ties Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 181 of 192 ACUSON X300


02.11 Ultrasound
182 Cables Replacement

8. Remove the five (5) screws beneath the CP neck to remove the CP neck cover from the
CP neck.

Fig. 189: Remove the CP neck cover


9. Pull out the three (3) cables through the hole1 beneath the CP neck and the hole2 at the
front of the CP neck.

Fig. 190: Calbe Ties Shown

ACUSON X300 US11-104.841.01.09.02 Page 182 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 183

10. Cut the cable ties at the marked locations.

Fig. 191: Cut the Cable Ties


11. Disconnect the gas cylinder cable from the gas cylinder. Then remove the four (4) HSH
bolts on the control panel base using a 6mm Allen wrench.

Fig. 192: Gas Cylinder Cable and Bolts Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 183 of 192 ACUSON X300


02.11 Ultrasound
184 Cables Replacement

12. Lift the control panel base to separate it from the main frame.

Fig. 193: Control Panel Base and Main Frame Separated


13. Remove the TI module; see (Removing the TI Module / p. 103).
14. Remove the four (4) HSH bolts beneath the up-down roller assembly and the two (2)
bolts on the right and left side of the up-down roller assembly using a 6mm Allen
wrench.

Fig. 194: Bolt Removal Shown

ACUSON X300 US11-104.841.01.09.02 Page 184 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 185

15. Lift the roller outer a little and cut the cable tie on the roller inner.

Fig. 195: Cut the cable ties on the front side


16. Cut the cable ties located on the left side of the main frame.

Fig. 196: Cut the cable ties on the left side

© Siemens, 2011 US11-104.841.01.09.02 Page 185 of 192 ACUSON X300


02.11 Ultrasound
186 Cables Replacement

17. Cut the cable ties located on the right side of the main frame.

Fig. 197: Cut the cable ties on the right side


18. Remove the IO module; see (Removing the I/O Module / p. 132).
19. Remove the DC/DC module; see (Removing the DC/DC Power Supply / p. 168).
20. Cut the cable tie at the marked location and pull out the cables through the hole1 and
hole2.

Fig. 198: Cut the cable tie and pull out the cable

ACUSON X300 US11-104.841.01.09.02 Page 186 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 187

21. Remove the two (2) screws on the electronic module.

Fig. 199: Remove the screws on the electronic module


22. Remove the four (4) screws on the bottom side of the electronic module.

Fig. 200: Remove the screws on the bottom side of the electronic module

© Siemens, 2011 US11-104.841.01.09.02 Page 187 of 192 ACUSON X300


02.11 Ultrasound
188 Cables Replacement

23. Pull the electronic module outward to remove the electronic module from the main
frame.

Fig. 201: Remove the electronic module


24. Cut the cable tie marked and disconnect all the cables from the AD/DC power supply.

Fig. 202: Disconnect Cables from AC/DC Power Supply


25. Remove AC/DC power supply, refer to (Removing the AC/DC Power Supply / p. 156).
26. Remove AC/DC power fan, refer to (Removing the AC/DC Power Fan / p. 173).

ACUSON X300 US11-104.841.01.09.02 Page 188 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 189

27. Pull out the three snap rivets located on the inner and the two on the outer of the front
main frame. Then remove the frame cable bracket from the main frame.

Fig. 203: Remove the snap rivets and Frame Cable Bracket
28. Pull out the cables through the hole located on the main frame.

Fig. 204: Main Frame with Cables Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 189 of 192 ACUSON X300


02.11 Ultrasound
190 Cables Replacement

Replacing the System Cables 18.2

NOTE Use Loctite 271 for holding the HSH bolts when replacing the
System Cables.

Perform the reverse sequence of the procedure provided for the removal. Refer to the fol-
lowing figures for the cable routing and tying.

1. Tie the cables on the roller inner. Ensure that there is adequate cable length for the fully
raised control panel base.

Fig. 205: Tie the cables on the roller inner

ACUSON X300 US11-104.841.01.09.02 Page 190 of 192 © Siemens, 2011


02.11 Ultrasound
Cables Replacement 191

2. Route and tie the cables on the control panel base.

Fig. 206: Cable Ties Routing Shown

© Siemens, 2011 US11-104.841.01.09.02 Page 191 of 192 ACUSON X300


02.11 Ultrasound
192 Changes to Previous Version
19-
Changes to Previous Version
19Changes to Previous Version

Change(s) Description
Revision 09 Added in 17inch monitor chapter (Replacing the
17 inch Monitor / p. 34). Performed general sys-
tem cleanup.
Revision 08 Minor edits throughout.
Revision 07 Added (Replace the Arm Lock Assembly / p. 53)
and modified the (Replace the Keyboard / p. 67)
sections. Added (Replace the Battery
Pack / p. 161).
Update for X300 PE Addition of AUCSON X300 Premium Edition to
various chapters, and the inclusion of new mod-
ule procedures.
Update Changes throughout to incorporate updates to
Version 3.0
Update Changes to RC Daughter board and CD / DVD
Drive replacement procedure.
Update Major changes throughout the document
New document Initial release for ACUSON X300.

ACUSON X300 US11-104.841.01.09.02 Page 192 of 192 © Siemens, 2011


02.11 Ultrasound

Potrebbero piacerti anche