Original Author:
Linda Barrow
Updated by:
Rachel Burton (Infection Control Link Nurse)
Judith Simpson (Infection Control Link Sister)
Ambubags
Single patient use. Keep dust free by keeping in dedicated plastic bag.
Baby baths
Clean with Clinell universal sanitising wipes. Store separately and inverted.
Bilibed
Dedicated sleep suit to be washed as per instructions for general baby
linen.
Bed to be cleaned with Clinell universal sanitising wipes.
Bili blanket
Clean with Clinell universal sanitising wipes.
Blood warmer
Clean with Clinell universal sanitising wipes.
Cots
When in use, clean daily with Clinell universal sanitising wipes. After transfer /
discharge, completely dismantle the cot and clean thoroughly with Clinell
universal sanitising wipes.
CPAP machines
All tubing and humidity chamber are disposable and renewed every 7
days (remember to label the humidifier with the date that it is due to be
changed).
Clean machine with Clinell universal sanitising wipes.
Clipboards
Clean with Clinell universal sanitising wipes.
Dummies
Single patient use. Rinse with hot water and place in steam / microwave
steriliser.
Filters
Depending on item, either changed by technician or included in service by
company. (At PRH filters are changed by the staff on a routine maintenance
programme).
Head boxes
Clean with Clinell universal sanitising wipes.
Heatshields
Clean with Clinell universal sanitising wipes.
Incubators
Airshields Isolette T1500/500 (transport incubators); Ohmeda Giraffe and
Drager Babytherm 8010 (Open Incubator):
When in use damp dust daily with Clinell universal sanitising wipes.
Change weekly (when possible) and completely dismantle and clean
the incubator with Clinell universal sanitising wipes. If the baby has
had an infection, clean incubator as above but any concerns should be
discussed with the Infection Control Link nurse or sister, or the
Infection Control Department (ext 4595).
Humidification chambers are non-disposable and are changed every 7
days with the incubators.
Store covered with a plastic sheet.
Remember that alcohol wipes should never be used on incubators as
they will damage the perspex.
IVAC pumps
Wipe thoroughly with Clinell universal sanitising wipes. Pay particular
attention to the grooves and recesses of the pump.
Laryngoscope blades
Single patient use and disposable.
M3/M4 monitor
(These are the small cardiac monitors e.g. transport incubator).
Monitor, SaO2 probe, ECG cable, pressure transducer, temperature probe and
NBP cable can all be cleaned with Clinell universal sanitising wipes. Take
care with the screens as they are more sensitive to touch.
Masks
Ambubag / neopuff masks are non-disposable and should be cleaned
with Clinell universal sanitising wipes.
Nebuliser masks are single patient use and should be wiped clean with
Clinell universal sanitising wipes and disposed of when no longer
required by the patient.
Nebulisers
Single patient use; clean with general purpose detergent and hot water, rinse
and dry. (Difficult to dry areas - leave to air dry as per manufacturers’
instructions)
Optiflow
All tubing and humidity chamber are disposable and renewable every 7 days
(remember a label should be attached to the chamber stating when the
change is next due). Clean machine with Clinell universal sanitising wipes.
Oxygen funnels
Clean with Clinell universal sanitising wipes. Please note if the funnel is in a
poor condition, it should be discarded and new one obtained.
Oxygen tubing
Used from equipment to patient needs to be discarded after patient
discharge / transfer.
Used from equipment to equipment can be kept unless contaminated,
damaged or dirty.
If baby infected then all tubing discarded.
Phototherapy light
Clean with Clinell universal sanitising wipes (remember to clean the base).
Phototherapy shield
Clean with Clinell universal sanitising wipes.
Razors
Single patient use and are disposable. (The use of razors is soon to be
reviewed and probably discontinued in favour of surgical clippers; be aware of
updates in the near future).
Radiant heater
Clean with Clinell universal sanitising wipes (remember to clean the base).
Saturation monitors
Clean monitor with Clinell universal sanitising wipes.
3 types of probe:
Disposable – single patient use.
Semi-disposable and re-usable - clean with Clinell universal sanitising
wipes.
Spotlight
Clean with Clinell universal sanitising wipes (remember to clean the base).
Stethoscopes
Checked / cleaned daily using Clinell universal sanitising wipes.
Remove ear pieces only. Leave chest piece intact unless visibly dirty.
Suction tubing
Patient to reservoir - change daily and between patients.
Wall to reservoir - date and change monthly.
Suction catheter
Do not store opened in paper sheaths, too many become contaminated or fall
out; also do not overstock the drawers.
TCM monitoring
The module, leads and probes can be cleaned with Clinell universal sanitising
wipes. The probe should be stored covered with the small green protective
cap.
Teats
See Baby bottles and teats.
Tecotherm Neo
Clean before and after use with Clinell universal sanitising wipes.
Clean device bottom side (ventilation holes) generally once every 1 – 2
months and before treatment as part of the system preparation.
Inspect every 3 months that air ventilation holes in the device bottom
side are not covered and free from dust. If so then inform the
technician.
The mattress should be cleaned with Clinell universal sanitising wipes.
Store covered with a plastic sheet.
Thermometers (electronic)
Probe covers are single patient use. Dispose of when patient
discharged or transferred.
Clean case with Clinell universal sanitising wipes.
Trolleys
Silver trolleys should be cleaned before and after use with Clinell
universal sanitising wipes.
Equipment trolleys (such as resuscitation and cannulation) should be
damp dusted daily and thoroughly cleaned weekly with Clinell universal
sanitising wipes.
Bedside trolleys in the nurseries (cot space) should be damp dusted
daily and thoroughly cleaned weekly with Clinell universal sanitising
wipes; remembering to clean the drawer runners and trays.
Notes trolleys and “COWs” (computers on wheels) should be cleaned
with Clinell universal sanitising wipes.
VENTILATOR CLEANING
All models can be cleaned with Clinell universal sanitising wipes.
AVEA VENTS
SLE (PRH)
STEPHANIE VENT
2. DISPOSABLE CIRCUITS
Clear breathing circuit including NO t-piece; red striped P-Sensor
tubing & double lumen red striped pressure tubing are disposable
Everything else apart from grey T-sensor wire is returned to CSSD for
processing