Adjusting the meter includes two steps: 1. setting up parameters; It can be done before
on-site calibration . 2. on-site calibration.
Panel
Functions of keys:
【ON】: power on;
【OFF】: power off;
【ZERO】: display 0;
【TOTAL】: to display total weight, also used in report printing;
【REQ】: to recall weighing results, displaying in reversed turn;
【F1】 : auxiliary key;
【BLIGHT】: ON/OFF key of the LCD display backlight; it is economical to turn
it off in strong light situations;
【T.C】:TTL CLR, push this key 2 times and all the records are cleared;
【DEL】: push this key 2 times and the last weighing record is cleared;
【 SUM】 : push this key when the weighing is finished; the indicator will then
start printing out a summary report on weighing of the current
category.
【AUTO】: entry into auto the auto status; when the highest bit of the display is
A, the indicator will automatically put results of each weighing into memo and
print it out; to exit AUTO, push this key for a second time;
【C.N】: to recall the weighing times concerned in the memo;
【CLASS】: to set the category No.(0~255);
【T.S】:TARE SET, to set tare;
【DATE】: date-set;
【TIME】: time-set;
【STORE】: to store up the weighing result and print it out;
【CE/PRINT】:to clear the incorrect input
Parameter set-up
1. Entering parameter set-up
(1). Press【DATE】 . Input “660328”, then press【DATE】, it displays “ SEL”.
(2). The number on the menu are defined as follows:
( ) is an optional value and 〖 〗is pre-set by our factory.
【1】: Full range (1-9999) 〖2〗
【2】: Calibration parameter (16384-65000)
【3】:Wait for input data of standard weight for automatic calibration (applied in on-
site calibration)
【4】:Analog gain (4-7) 〖5〗
【6】:Graduation (1,2,5,10,20,50) 〖1〗
【7】:Decimal point position (0-3) 〖0〗
【8】:Unit of capacity and weight (0,1) 〖1〗
【9】:Filtering parameter (1-6) 〖5〗
Real weight
New calibrated parameter = ---------------------- * original calibrated parameter
Display weight
For example:
The real weight of goods is 5000kg, but the scale display 5004kg, the original
calibrated parameter is 42126, so the new calibrated parameter is:
(5000÷5004)×42126 = 42092, input “42092”, then press ”recall” for a while,
come back to weighing, then it display “5000”.