Sei sulla pagina 1di 5

Instituto Politécnico Nacional

Escuela Superior de Cómputo

Computer systems engineering

Instrumentation laboratory
Members:
Flores Aguilera Miguel Ángel
Ramírez Martínez Janet Naibi
Morales Garduño Brenda

Group: 3CV4

Professor: Hernández Aguilar José Luis

Date: Monday 30, October

Practice 5:
Conditioning of Generating Sensors with
Instrumentation Amplifier.
Index
Objective ............................................................................................................................................. 3
Equipment used: ................................................................................................................................. 3
Introduction. ....................................................................................................................................... 3
Development of the practice. ............................................................................................................. 3
1. Conditioning generator sensor with discrete Instrumentation Amplifier. ............................. 3
2. Conditioning generator sensor with the AD620, Al in Cl. ....................................................... 4
Questionnaire...................................................................................................................................... 5
Conclusions ......................................................................................................................................... 5
Bibliography ........................................................................................................................................ 5
Use of Generating Sensors

Objective
The student will learn to use the generator transducers, as well as to calibrate the different
components of a measurement system, in order to find the voltage value corresponding to
the variable under measurement.

Equipment used:
2 Multimeters 1 Protoboard.
1 Variable VCD source. 7 10KΩ resistors.
6 Banana-Cayman Tips. 2 0.01µF capacitors.
4 Cayman-Cayman Tips. 1 0.1µF capacitor.
1 10KΩ potentiometer.
2 LM358 or 3 LM741.
1 LM35.
1 AD620.

Introduction.
Insertar intro aquí

Development of the practice.


1. Conditioning generator sensor with discrete Instrumentation Amplifier.
Assemble the Instrumentation Amplifier shown in Figure 1, using 3 Operational Amplifiers.
Use the voltage source in bipolar mode to power the Op-Ams with ±12V.

Figure 1
To make the adjustment of the voltage gain in the Instrumentation Amplifier, consider that
Rf = 100Ω and that the potentiometer Rg must be manipulated, for this to make a voltage of
1V in terminal E2 and terminal E1 connect it to ground. Place the potentiometer in such a
way that the output has a voltage of 1V for a voltage gain Av=1; or manipulate the
potentiometer so that the output has a voltage of 2V for a gain Av=2, and so on. This
operation must be repeated each time the gain value requested in table 1 must be adjusted.
After adjusting each gain value indicated in table 1, connect the output of the LM35 sensor,
which is used in place of a generator sensor, to terminal E2 as shown in figure 2 and start
filling the data that are requested in that table.

Figure 2

Gain Av Sensor temperature E2 (Volts) Vsal (Volts)


(Fit RG) LM35 (°C) measured measured
25° 260.7 mV 260.7 mV
1 50° 512.4 mV 512.4 mV
75° 772.0 mV 772.0 mV
25° 260.7 mV 0.552 mV
2 50° 512.4 mV 1.10 V
75° 772.0 mV 1.68 V
25° 260.7 mV 8.60 mV
3 50° 512.4 mV 1.68 V
75° 772.0 mV 2.26 V
Table 1 Operation of the discrete instrumentation amplifier with generator sensor

2. Conditioning generator sensor with the AD620, Al in Cl.

Repeat the same procedure above but now with the AD620 Instrumentation Amplifier,
energizing with ±12V and connecting your terminal 2 and 5 to ground (refer to figure 3). I
varied the value of RG, between terminals 1 and 8, for the adjustment of voltage gain taking
into account that Rf = 24.7KΩ and fill the table 2 considering that again the LM35 will be
present with its output connected to terminal 3.

Gain Av Sensor temperature E2 (Volts) Vsal (Volts)


(Fit RG) LM35 (°C) measured measured
25° 1.38 mV
1 50° 1.58 mV
75° 1.80mV
25° 483.62 mV
2 50° 1.06 V
75° 1.45 V
25° 0.812 mV
3 50° 1.53 V
75° 2.25 V
Table 2 Operation of the AD620, instrumentation amplifier in Cl with generator sensor.

Figure 3

Questionnaire
1. What is the main characteristic of a generator sensor?
2. What is the Instrumentation Amplifier used for?
3. What are the characteristics of the Instrumentation Amplifier?
4. What is the mathematical expression to calculate the voltage gain in the used
Instrumentation Amplifiers?
5. To measure the temperature with a thermocouple, how should that sensor be used
with the Instrumentation Amplifier?

Conclusions

Bibliography

Potrebbero piacerti anche