Sei sulla pagina 1di 26

Laboratory Perfect Calibration Centre Pvt. Ltd., No.

84 First Floor, Times Corner,


Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 1 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

1. Temperature
Simulation$
(Temperature Indicator /
Controller / Recorder/
Scanner)
RTD Type (-) 200°C to 800°C 0.56°C Using Process Calibrator
‘J’ Type Thermocouple (-) 200°C to 1200°C 0.8°C By Direct Method
‘K’ Type Thermocouple (-) 200°C to 1350°C 0.81°C
100°C to 1750°C 1.14°C
‘R’ Type Thermocouple 100°C to 200°C 1.1°C
200°C to 1750°C 0.82°C
‘S’ Type Thermocouple 600°C to 1820°C 0.94°C
B’ Type Thermocouple (-)200°C to 400°C 0.59°C
T’ Type Thermocouple (-) 200°C to 0°C 0.43°C
0°C to 1300°C 0.37°C
‘N’ Type Thermocouple (-) 200°C to 1000°C 0.29°C
‘E’ Type Thermocouple

2. Time Interval (Measure) 1 sec to 1 min 0.08 to 0.13 sec Using Digital Timer
1 min to 24 hr 0.13 to 22 sec by Comparison Method

3. Temperature
Simulation♣
(Temperature Indicator /
Controller / Recorder/
Scanner )
RTD’ Type -200°C to 800°C 0.82°C Using Multifunction
‘J’ Type Thermocouple -200°C to 1200°C 1.45°C Calibrator by Direct
‘K’ Type Thermocouple -200°C to 1350°C 0.96°C Method
‘R’ Type Thermocouple 0°C to 1750° 2.81°C
‘S’ Type Thermocouple 0°C to 1750°C 2.32°C

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 2 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

‘B’ Type Thermocouple 600°C to 1820°C 2.88°C


‘T’ Type Thermocouple -200°C to 400°C 0.69°C
‘N’ Type Thermocouple -200°C to 1300°C 0.52°C
‘E’ Type Thermocouple -200°C to 1000°C 0.40°C

4. Time Interval (Measure) 1 sec to 1 min 0.2 to 0.23 sec Using Digital Timer
1 min to 24 hr 0.23 to 22 sec by Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 3 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

FLUID FLOW CALIBRATION

1. Air Flow 0.4 to 20 m/s 7.2%rdg Using Anemometer,


(Anemometer Vane Variable Speed Blower for
Type) Air Supply by Comparison
Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 4 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

MECHANICAL CALIBRATION

I. DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)

1. Calipers$
(Vernier/Dial/Digital)
L.C:0.01mm 0 to 450mm 18.2 µm Using Slip gauge Set
450 mm to 600mm 18.5 µm Grade ‘0’, Caliper
600 mm to 1000mm 19.5 µm Checker, Length bar &
1000 mm to 1500mm 21.3 µm Slip gauge Accessories.
Comparison Method
L.C:0.02mm 0 to 2000mm 20.2 µm IS :3651(Part 1 & 2)

2. External Micrometer/
Ball Micrometer/
Blade Micrometer/
Disc
Micrometer/Flange
Micrometer/Groove
Micrometer/
Spline Micrometer/
Sheet Metal
Micrometer/
Point
Micrometer/Pitch
Micrometer$
L.C.:0.001mm 0 to 25mm 2.0 µm Using Slip Gauge Set
25 mm to 150mm 5.8 µm Grade ‘0’, Optical Flat &
150 mm to 300mm 6.2 µm Parallels, length bar.
300 mm to 600mm 8.5 µm Comparison Method
600 mm to 1000mm 9.2 µm IS:2967

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 5 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

3. Depth Micrometer$
L.C:0.01mm 0 to 150mm 5.82 µm Using Slip Gauge Set
150 mm to 300 mm 8.2 µm Grade ‘0’, Caliper
Checker & Gauge Block
Accessories by
Comparison Method
BIS:6468

4. Internal Micrometer/
Stick Micrometer$
L.C:0.01mm 0 to 600mm 6.7 µm Using Slip Gauge Set
600 mm to 1000mm 8.5 µm Grade ‘0’, Gauge Block
1000 mm to 2000mm 20.0 µm Accessories Set & Length
bar by Comparison
Method as per IS:2966

5. Plunger Type Dial


Indicator$
L.C:0.001mm 0 to 25mm 2.1 µm Using Length Measuring
25 mm to 50mm 2.1 µm Machine by
Comparison Method as
L.C:0.0005mm ± 0.025mm 2.1 µm per IS:2092

6. Lever Type Dial


Indicators$
L.C:0.01mm 0 to 1.6 mm 5.8 µm Using Length Measuring
L.C:0.001mm 0 to 0.6 mm 2.1 µm Machine by Comparison
L.C:0.002mm 0 to 0.2 mm 2.1 µm Method as per IS:11498

7. Plain Plug Gauge/ Up to 100mm 2.1 µm Using Length Measuring


Width Gauge$ 100 mm to 200mm 2.8 µm Machine by
Comparison Method as
per IS:3455 & IS:7859

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 6 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

8. Snap 3 mm to 200mm 1.4 µm Using Slip gauge Set


Gauge/Adjustable 200 mm to 400mm 1.94 µm Grade ‘0’ by
Snap Gauge, Knife Comparison method as
Edge Gauge, Length per IS:3455, IS:7859 &
Gauge, Height IS:3477
Gauge, Master
Setting Gauge, Gap
Gauge, Distance
Gauge, Depth Gauge,
Bridge Gauge$

9. Dial Depth Gauge$


L.C:0.01mm 0 to 10mm 6.0 µm Using Slip gauge Set
0 to 100mm 6.0 µm Grade ‘0’ by
Comparison method

10. Feeler Gauge$ Upto 1.0mm 2.9 µm Using Digital Micrometer


by Comparison method
as per IS:3179

11. Bore Dial Gauge$


(Only Transmission
movement)
L.C:0.001mm Upto 3mm 2.1 µm Using Length Measuring
Machine by
Comparison Method

12. Dial Thickness


Gauge$ 0 to 50mm 5.9 µm Using Slip gauge Set
L.C:0.01mm Grade ‘0’ by
Comparison method as
perIS:2092

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 7 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

13. Height Gauge$


(Vernier/Digital/Dial)
L.C: 0.01mm 0 to 300mm 14.8 µm Using Caliper Checker by
0 to 1000mm 15.2 µm Comparison method
IS:2921

14 Pistol Caliper$
L.C:0.1mm 0 to 100mm 58.0 µm Using Slip gauge Set
Grade ‘0’ by
Comparison method
IS:2092

15. Depth Caliper $


(Vernier/Dial/Digital)
L.C:0.01mm 0 to 300 mm 9.1 µm Using Slip gauge Set
0 to 600 mm 11.1 µm Grade ‘0’, Gauge Block
Accessories & Caliper
Checker by
Comparison Method
IS:4213

16. Dial Caliper


Gauge/Groove Dial/
Inside Caliper
Gauge$ Using Slip gauge Set
L.C:0.01mm 10 mm to 150mm 5.8 µm Grade ‘0’, & Slip gauge
Accessories by
Comparison Method

17. Cylindrical Upto 25mm 2.1 µm Using Length Measuring


Measuring Pins$ Machine by
(Only Diameter) Comparison Method
IS:11103

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 8 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

18. Cylindrical Upto 100mm 2.1 µm Using Length Measuring


Measuring master$ Machine by
(Only Diameter) Comparison Method
IS:4349

19. Thread Measuring 0.17 mm to 6.35mm 2.1 µm Using Length Measuring


Wires Machine by
Comparison Method
IS:6311

20. Thread Plug gauge$ M2 to M100 3.6 µm Using Length Measuring


M100 to M200 3.6 µm Machine

M2 to M100 5.6 µm Using FCDM, Thread


M100 to M200 5.6 µm Measuring Wire &
Cylindrical Setting Master
by Comparison Method
as per IS 2334, IS 4218
(Part I to IV) ,IS 10216,
IS 8999, ANSI/ASME
B.20.1-1983

21. Taper Thread Ring M2 to M50 5.1 µm Using Length Measuring


Gauge$ M50 to M100 5.1 µm Machine by
Comparison Method
IS 2334, IS 4218 (Part I to
IV) , IS 10216, IS 8999 ,
ANSI/ASME B.20.1-1983

22. Thread Ring Gauge$ M3 to M100 3.6 µm Using Length Measuring


M100 to M 200 3.6 µm Machine & Master Setting
Ring by Comparison
Method as per
IS 4218,IS 2334 ,
ANSI/ASME B.20.1-1983

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 9 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

23. Taper Thread Plug M2 to M100 7.5 µm Using FCDM, Thread


Gauge$ Measuring Wire &
Cylindrical Setting Master
by Comparison Method as
per IS 8999, ANSI/ASME
B.20.1-1983

24. Plain Ring Gauge$ Up to 50mm 3.3 µm Using Length Measuring


50 mm to 100mm 3.3 µm Machine & Master Setting
100 mm to 200mm 3.3 µm Ring by Comparison
200 mm to 300mm 5.7 µm Method as per
IS: 3485

25. Flush Pin Gauge$ Up to 200 mm 2.8 µm Using Slip Gauge Set
Grade ‘0’ , Digital
Micrometer and Plunger
Dial gauge by
Comparison Method

26. Coating Thickness 0.01mm to 2mm 1.79 µm Using Standard Foils by


Gauge$ Comparison Method

27. Foils$ Up to 250 m 1.2 µm Using Length Measuring


0 to 5mm Machine by
Comparison Method

28. Setting Rod / Setting Up to 50mm 3.5 µm Using Slip Gauge Set
Piece$ 50 mm to 100mm 3.5 µm Grade ‘0’, Length bar &
100 mm to 200mm 3.5 µm Electronic comparator by
200 mm to 1000mm 4.7 µm Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 10 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

29. Dial Snap Gauge /


Dial Snap Meter$
L.C:0.001mm 0 to 100mm 0.8 µm Using Slip Gauge Set
(Parallelism of Grade ‘ 0’ by
Measuring faces) Comparison Method as
per IS 3455, IS 7859 &
IS 3477

30. V - Block$ Using Lever Dial gauge,


(Parallelism of (250x 200x180)mm Slip Gauge set Grade “0”,
seating faces, 4.2 µm Bevel Protractor by
Perpendicularity, 3.4 µm Comparison Method as
Angle, 14 min per IS 2949
Axis of the shaft) 1.4 µm

31. Ultrasonic
Thickness gauge$
L.C.: 0.1mm 0 to 100mm 70.2 µm Using slip gauges set
grade“0” by
Comparison Method

32. Comparator Stand$ 300 x 300mm 1.7 µm Using Dial gauge by


( Flatness Only) Comparison method as
per IS 6973

33. Straight edge$ Up to 1000mm 6.0 µm Using Electronic Level &


Dial Gauge by
Comparison Method as
per IS :2220

34. Tri Square$ Up to 500 x 500mm 4.4 µm Using Slip Gauge Set
(Squareness) Up to 500 x 500mm 4.0 µm Grade ‘ 0’, Master
(Parallelism) Square, Plunger dial,
Surface plate by
Comparison Method as
per IS :2103

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 11 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

35. LVDT Probe with


DRO$
L.C:0.1 µm 0 to 10mm 2.5 µm Using Slip Gauge Set
Grade ‘ 0’ by
Comparison Method

36. V- Anvil Micrometer$


L.C: 0.001mm 0 to 100mm 2.1 µm Using Cylindrical setting
master by Comparison
Method

37. Mandrel 100 mm dia 13.4 µm Using Bench centre with


(Run out only) 200 mm Length dial as per IS 2063,
ISO230-1

38. LVDT Probe with


DRO♣
L.C:0.001mm 0 to 25mm 2.5 m Using Slip gauge set
Grade’0’ by
Comparison Method

39. Surface plate♣ 6000 mm X 3000 mm 1.32


√L+W
m Using Electronic level as
150 per IS:7327,IS:2285,
IS:12937

40. Height Measuring


System♣
L.C:0.001 mm 0 to 600 mm 2.8 m Using Slip gauge Set
Grade ‘0’, Length bar by
Comparison Method

41. Microscope♣
Linear
L.C: 0.0001mm 0 to 300 mm 1.1 m Using Graticules scale by
Angle 0 to 360° 3’ Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 12 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

42. Bench Centre Centre height:300 mm 6.0 m Using Mandrel & Dial
Parallelity Admit between Indicator as per
Co- axiality of Centre:750mm IS:5980:2000
centre♣

43. Profile projector /


Video Measuring
Machine♣
Linear
L.C:0.001mm 0 to 300 mm 17.75 m Using Graticules scale ,
Angle L.C:36” 0 to 360 ° 2’ 16” ‘0’ Grade Slip gauge set
Magnification 5X to 100X 2.5% by Comparison Method

44. Gauge Block


Measuring Fixture
with Electronic
System♣
L.C:0.01 m 0 to 100mm 0.09 m Using Grade ‘K’ Pair
Gauge Block by
Comparison Method

45. Tape and Scale


Calibrator♣
L.C:0.001mm 0 to 1000 mm 18.7 m Using Slip Gauge Set
Grade ‘0’, Length Bar,
Lever Dial Gauge by
Comparison Method

II. ACCELERATION & SPEED

1. Tachometer / RPM$
Contact Type 150 rpm to 2500 rpm 0.82% Using RPM Indicator with
Non contact type 150 rpm to 50,000 rpm 0.87% Electric Motor by
Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 13 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

2. Indicator of
Centrifuge / RPM♣
Meter 60 to 25,000 rpm 0.06% of rdg Using Digital Tachometer
Comparison Method

III. ACCOUSTICS

1. Sound Level Meter$ 1kHz Using Sound Level


94 dB & 114 dB 0.8 dB Calibrator by Direct
Method as per
IS:9779:1981

IV. WEIGHTS

1. Mass$ 1 mg 0.013 mg Using E1 Class Standard


(Class F1 accuracy 2 mg 0.013 mg Weights and Electronic
& Coarser) 5 mg 0.013 mg Weighing Balance
10 mg 0.013 mg (Readability : 0.01mg)
20 mg 0.013 mg Substitution method of
weighing and ABBA
weighing cycle based on
OIML R111-1

50 mg 0.02 mg Using E1 Class Standard


100 mg 0.02 mg Weights and Electronic
200 mg 0.02 mg Weighing Balance
500 mg 0.02 mg (Readability: 0.01mg)
1g 0.03 mg Substitution method of
2g 0.04 mg weighing and ABBA
5g 0.04mg weighing cycle based on
10 g 0.05 mg OIML R111-1
20 g 0.06 mg

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 14 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Class F1/M1 50 g 0.09 mg Using E1& F1 Class


accuracy & Coarser 100 g 0.10 mg Standard Weights and
200 g 0.15 mg Electronic Weighing
500 g 0.013 g Balance
1 kg 0.08 g (Readability:
2 kg 0.09 g 0.1mg/10mg)
Substitution method of
weighing and ABBA
weighing cycle based on
OIML R111
5 kg
0.49 g Using F1 Class Standard
10 kg
0.49 g Weights and Electronic
20 kg
1.0 g Weighing Balance
(Readability : 500mg)
Substitution method of
weighing and ABBA
weighing cycle based on
OIML R 111

V. WEIGHING SCALE AND BALANCE

1. Weighing Balance
Calibration of
Electronic Weighing
Balance$
Class II and coarser
Readability : 0.01mg Up to 200 g 0.04 mg Using E1 Class standard
weights, as per OIML R
76-1

Readability : 10 mg Up to 4 kg 0.09 g Using E1 and F1 Class


Standard Weights
as per OIML R 76-1

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 15 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Readability : 500 mg Up to 30 kg 1.30 g Using F1 & M1 Class


Standard Weights as per
OIML R 76-1

2. Calibration of
Electronic Weighing
Balance♣
Class I and coarser
Readability : 0.001 Up to 6.1 g 0.01 mg Using E1 Class Standard
mg Weights as per OIML R
76-1
Readability : 20 g Up to 200 kg 41.0 g

Readability : 50 g Up to 300 kg 50.0 g

VI. VOLUME

1. Micro Pipettes$ 10 l to 50 l 0.11 l Using Precision balance


Readability : 50 l to 100 l 0.11 l and Distilled Water Of
0.01/0.1 mg 100 l to 1000 l 0.27 l Known Density By
1 ml to 5 ml 0.14 l Gravimetric Method as
per ISO 8655-6 & ISO
4787

2. Burette/ Pipette$ Using Precision balance


Readability : 0.01/0.1 1ml to 50 ml 0.14 ml and Distilled Water Of
mg 50 ml to 100 ml 0.14 ml Known Density By
Gravimetric Method as
per ISO 4787

3. Measuring Cylinder/ 100 ml to 200 ml 0.19 ml Using Precision balance


Graduated Jar$ 200 ml to 500 ml 0.63 ml and Distilled Water Of
Readability : 10 mg 500 ml to 1000 ml 0.73 ml Known Density By
& 100 mg 1000 ml to 2000 ml 1.32 ml Gravimetric Method as
2000 ml to 5000 ml 3.1 ml per ISO 4787

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 16 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

4. Volumetric Flask$ 10 ml to 100 ml 0.14 ml Using Precision balance


Readability 100 ml to 200 ml 0.14 ml and Distilled Water By
:0.01mg/ 10 mg/ 200 ml to 500 ml 0.63 ml Gravimetric Method as per
100 mg 500 ml to 1000 ml 0.73 ml ISO 4787
1000 ml to 2000 ml 1.31 ml
2000 ml to 5000 ml 3.1 ml

VI. TORQUE GENERATING DEVICES

1. Torque Wrench / 2 Nm to 20 Nm 2.91% Torque transducer with


Torque Screw 20 Nm to 200 Nm 2.61% Indicator & Fixtures
Driver/Torque 200 Nm to 2000 Nm 1.32% IS/ISO 6789:2003
Spanner
(Type-I & II All
Classes)

VII. UTM, TENSION CREEP AND TORSION TESTING MACHINE

1. Uni- Axial Testing


Machines♣
Compression 5 N to 300 kN 0.8% Using Load cell with
Tension 5 N to 300 kN 0.8% Indicator based on
IS:1828

VIII. HARDNESS TESTING MACHINES

1. Shore Hardness
Tester$
Shore A 0-100 Shore A 0.6 shore A Using Load cell with
Shore D 0-100 Shore D 0.6 shore D Indicator based on ASTM
D 2240-05

2. Verification of HRA 1.0 HRA Using Standard Test


Rockwell Hardness blocks as per IS:1586-
HRB 1.0 HRB
Tester♣ 2:2012 in- by Indirect
HRC 1.0 HRC Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 17 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

3. Verification of HBW 2.5/187.5 1.2% Using Standard Test


Brinell Hardness 1.3% blocks as per
HBW 5/750
Tester♣ 1.2% IS:1500-2:2013 IN by
HBW 10/3000
Indirect method

4. Verification of HV 5 1.6% Using Standard Test


Vickers Hardness HV 10 1.6% blocks as per
Tester♣ HV 20 1.6% IS:1501-2:2013 in by
HV 30 1.6%
Indirect Method
HV 50 1.6%

IX. PRESSURE INDICATING DEVICES

1. Pneumatic 0 to 20 bar 0.06%rdg Using Digital Pressure


Hydraulic$ Calibrator DPI 610 and
0 to 700 bar 0.18%rdg
(Pressure Gauge/ External transducer by
Pressure Indicator/ Comparison Method as
Switch/ per DKD-R6-2
Transmitters/
Manometer/
Compound Gauge/
Transducers With
Indicators)

2. Vacuum (-)0.1 bar to (-)0.9 bar 0.7%rdg Using Digital Pressure


Gauge/Switch/ Calibrator DPI 610 by
Transmitters/ Comparison Method as
Transducer With per DKD-R6-1
Indicators/
Compound Gauge$

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 18 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

3. Pressure (-)350 mbar to 350 mbar 0.24 % rdg Using Digital low pressure
Magnehelic Calibrator by Comparison
Gauges / method as per DKD-R6-1
Differential & DKD-R6-2
Pressure Gauge#

4. Pressure Gauge/ 0 to 20 bar 0.33 % rdg Using Digital Pressure


Pressure Indicator / Calibrator by Comparison
0 to 700 bar 0.20 % rdg
Switch/Transmitters Method as per DKD-R6-1
/Manometer/
Compound Gauge/
Transducers with
Indicators♣

5. Vacuum Gauge / (-) 0.1bar to (-) 0.85 bar 0.95 % rdg Using Digital Pressure
Vacuum Indicator/ Calibrator by Comparison
Switch / Method as per DKD-R6-1
Compound Gauge / & ISO: 3567
Transmitters and
Transducer with
Indicator♣

MECHANICAL CALIBRATION – MOBILE

1. Calipers:
(Venire / Dial /
Digital)
L.C:0.01mm 0 to 1000 mm 15.0 µm Using Slip gauge Set
grade ‘0’, Caliper
Checker, Length bar &
Slip gauge Accessories
IS:3651(Part I & II)

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 19 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

2. External
Micrometer/ Ball
Micrometer/ Blade
Micrometer /Disc
Micrometer /Flange
Micrometer/
Groove
Micrometer/Spline
Micrometer /Sheet
Metal
Micrometer/Point
Micrometer/Pitch
Micrometer
L.C:0.001mm 0 to 100mm 1.0 µm Using Slip Gauge Set
100 mm to 600mm 4.5 µm Grade ‘0’, Optical Flat &
Parallels, Length Bar,
Gauge Block Accessories
IS:2967

3. Depth Micrometer
L.C:0.01mm 0 to 300 mm 11.15 µm Using Slip Gauge Set
Grade ‘0’, Caliper
Checker, Gauge Block
Accessories
BIS:6468

4. Internal
Micrometer/ Stick
Micrometer
L.C:0.01 mm 0 to 600 mm 7.1 µm Using Slip Gauge Set
Grade ‘0’, Length Bar &
Gauge Block Accessories
IS:2966

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 20 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

5. Plunger Type Dial


Indicator
L.C:0.001 mm 0 to 25 mm 4.12 µm Using Dial Calibration
Tester IS:2092

6. Lever Type Dial


Indicator
L.C:0.01 mm 0 to 1.6mm 4.86 µm Using Dial Calibration
L.C:0.001mm 0 to 0.6 mm 4.17 µm Tester IS:11498

7. Snap 3 mm to 200 mm 1.74 µm Using Slip Gauge Set


Gauge/Adjustable 200 mm to 400 mm 3.7 µm Grade ‘0’, Slip Gauge
Snap Gauge, Knife Accessories IS: 3455,
Edge Gauge, IS:7859 & IS 3477
Length Gauge,
Height Gauge,
Master Setting
Gauge, Distance
Gauge, Depth
Gauge, Bridge
Gauge

8. Feeler Gauge Up to 1.0 mm 2.78 m Using Digital Micrometer


IS 3179

9. Bore Dial Gauge


( Only
Transmission
Movement)
L.C:0.001mm Up to 2 mm 4.27 m Using Dial Calibration
Tester by Comparison
Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 21 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

10. Dial Thickness


Gauge
L.C.: 0.01 mm 0 to 50 mm 5.79 m Using Slip Gauges Set
Grade ‘0’ as per IS 2092

11. Height Gauges


(Vernier/ Digital/
Dial)
L.C:0.01mm 0 to 600 mm 11.1 m Using Caliper Checker
IS 2921

12. Pistol Caliper


L.C.: 0.01 mm 0 to 100 mm 3.7 µm Using Slip Gauge Set
Grade ‘0’ as per IS 2092

13. Depth Vernier


Caliper
L.C.: 0.01 mm 0 to 300 mm 11.6 m Using Slip Gauge Set
Grade ‘ 0’, Gauge Block
Accessories & Caliper
checker as per IS 4213

14. Dial Caliper Gauge/


Groove Dial / Inside
Caliper Gauge
L.C.: 0.01mm 10 mm to 150 mm 3.62 m Using Slip Gauge Set
Grade ‘0’ & Slip Gauge
Accessories by
Comparison Method

15. Flush Pin Gauge Up to 200 mm 2.89 m Using Slip Gauge Set
Grade ‘0’ , Digital
micrometer & Plunger dial
by Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 22 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

16. Coating Thickness 0.01mm to 2 mm 1.79 m Using Standard Foils by


Gauge Comparison Method

17. Foils 0 to 5 mm 1.59 m Using Digital Micrometer


by Comparison Method

18. Measuring Pin 1mm to 20mm 2.83 m Using Slip Gauge Set
Grade ‘ 0’ Electronic
Comparator Stand
IS 11103

19. Dial Snap Gauge /


Dial Snap Meter
(Parallelism of
Measuring faces )
L.C: 0.001mm 0 to 100 mm 0.7 m Using Slip Gauge Set
Grade ‘ 0’ as pwe
IS 3455, IS 7859 & IS
3477

20. Plain Plug Gauge Up to 50 mm 2.83 m Using Slip Gauge Set


Grade ‘ 0’ Electronic
Comparator Stand
IS 3455 & IS 7859

21. Setting Rod Up to 50 mm 0.5 m Using Slip Gauge Set


50 mm to 100 mm 1.19 m Grade ‘ 0’, Length bar &
100 mm to 200 mm 1.32 m Electronic Comparator
200 mm to 600 mm 4.7 m Stand by
Comparison Method

22. Three Point 6 mm to 200mm 2.76 m Using Setting Ring Gauge


Micrometer 200 mm to 250mm 5.95 m by Comparison Method

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 23 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

23. V- Block Parallelism 4.2 m Using Slip Gauge Set


(Parallelism of Perpendicular 3.4 m Grade ‘ 0’, Lever dial
seating faces, Angle 15’ 51” Gauge, Bevel Protractor
Perpendicularity, Axis of Shaft 1.4 m As per IS 2949
Angle, Axis of the
shaft)

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 24 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

THERMAL CALIBRATION

1. TEMPERATURE

1. Liquid in Glass (-) 20 to 150°C 0.20°C Using SPRT with Digital


Thermometers$ Indicator and Low
Temperature Bath.
By Comparison Method
Confirming to IS:6274

2. RTD’s, Thermocouple , (-)80 to 100°C 0.24°C Using SPRT with


Digital Thermometer, 100 °C to 600°C 1.42°C Indicator,
Temperature Indicator 600°C to 1200°C 1.96°C S-type Thermocouple
With Sensor, Baths, with Indicator and Low
Temperature Controllers Temperature Bath/ Dry
With Sensor, Data Black Calibrator & Data
Logger, Transmitters logger by Comparison
With Sensors, Method confirming to
Temperature Gauges & EURAMET cg-8 Ver. 2.1
Temperature Switches$

3. Non – Contact 50 °C to 300°C 2.55°C Using IR Thermometer /


Temperature Calibration 300 °C to 1200°C 3.75°C Pyrometer with Digital
Infrared Thermometer & Indicator Black Body
Pyrometer$ Source by Comparison
Method MSL Technical
Guide & OEM Application
note

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 25 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

4. RTD’s, (-) 80°C to 100°C 0.5°C Using SPRT with Digital


Thermocouple , 100°C to 600°C 1.6°C Indicator and
Digital 600°C to 1200°C 2.1°C S-type Thermocouple
Thermometer, with Indicator and Low
Temperature Temperature Bath/ Dry
Indicator With Black Calibrator & Data
Sensor, Baths, logger By Comparison
Temperature Method confirming to
Controllers With DKD-R-5-1,
Sensor, Data EURAMET
Logger, cg-8,Ver 2.1
Transmitters With
Sensors,
Temperature
Gauges &
Temperature
Switches♣

5. Environment (-)80°C to 100°C 0.9°C Using SPRT with Indicator


Chamber, Cold 100°C to 300°C 1.5°C By Comparison Method,
Chamber, Freezer, based on DKD-R-5-7
Incubator,
Autoclave, Oven♣
( Single Point )

6. Environment (-)80°C to 250°C 0.8 to 1.4 °C Using RTD sensor with


Chamber, Cold Data logger (Multi sensor)
Chamber, Freezer, By Comparison Method,
Incubator, based on DKD-R-5-7.
Autoclave, Oven♣
( Multi Point )

Vishal Shukla Avijit Das


Convenor Program Director
Laboratory Perfect Calibration Centre Pvt. Ltd., No.84 First Floor, Times Corner,
Gengu Reddy Road, Egmore, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2207 (in lieu of C-0399, C-0243,C-0446) Page 26 of 26
Validity 18.06.2017 to 17.06.2019 Last Amended on 01.08.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

7. Furnace♣ 100°C to 600°C 1.39 °C Using SPRT with Indicator


(Single Point) 600°C to 1200°C 1.89 °C R-type Thermocouple
1200°C to1500°C 2.95 °C with Indicator
By Comparison Method,
based on AMS2750E

8. Furnace♣ 100°C to 1200°C 2.79°C Using N - Type


(Multi Point) Thermocouple with Data
logger By Comparison
Method, based on
AMS2750E

II. SPECIFIC HEAT AND HUMIDITY

1. Humidity Indicator 15% to 95% @ 25°C 1.4%RH Using Single Point Probe
of Humidity with Digital Indicator
Chamber / By Comparison Method,
Environment as per DKD – R-5-7
Chamber♣
(Single point)

2. Humidity Chamber / 15% to 95% @ 30°C 1.75%RH to 2.83%RH Using Humidity Sensors
Environment with Digital Indicator
Chamber♣ (Multi point)
(Multi point) By Comparison Method,
as per DKD- R-5-7

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.
Φ
Laboratory can also calibrate instruments/devices of coarser resolution / least count within the
accredited range using same reference standard/ master equipment under the scope of accreditation.

Vishal Shukla Avijit Das


Convenor Program Director

Potrebbero piacerti anche