Specification For Digital Mass Flow Controllers.

1.Scope of supply
This specification is for the supply of Digital Mass Flow Controller For
Helium
This Unit is Intended to Be Used For Specified Flow Rates, Pressure,
Temperatures Only.

2.Make Of Instrument
Digital Mass Flow Controller.
Make:Precision Flow Control Instruments-Pune.
Email: info@massflow.in Phone Nos:7263089498

Technical Specification:-

Sr.No Parameters Value


1 Type Inline Mass Flow Controller With Digital Display
and Keypad.MOC-Suitable For Helium Gas
Applications.
2 Measuring & Control The Instrument Can Display The Flow Rates and
capability Can Control The Gas Flow Rate.
3 Media Helium Gas Flow
4 Material of Construction As Per Suitable For The Helium Media. Most Of
The Wetted Materials will be Stainless Steel
5 Minimum Control Maximum Flow Range
Range. Controllable
6 Flow Range 5% Of Max Flow rate 100% of Max Flow
= 25 Sccm Range = 500 Sccm
o
7 Process temperature 0 C 50 oC
8 Process pressure Min Of 5 Bar Maximum Of 5 Bar
Pressure Needs to be Pressure Will be
Applied To The Mass Operational Pressure For
Flow Controller Inlet The Mass Flow
Controller.
9 Flow accuracy ±0.5% of F.S. or better
10 Flow Meter/ Flow Sensor Flow Sensor/ Flow Meter Manufactured Solely
Independently By Thori India OPC Pvt Ltd-Pune
For M/s Precision Flow Control Instruments-Pune
11 Control Valve Manufactured or Made to Manufacture On
Precision Flow Control Instruments Name.by M/s
Thori India OPC Pvt Ltd-Pune.
12 Leak Integrity 10-8 mbar l/s or better
13 Process connection 1/4 Inch Thread Ends. As Per The Required
Standard Manufacturing Third Party Supplier.

14 Power Supply Required For 230 V AC 50 Hz


Display Unit.
15 Output 0 to 10 V DC Inbuilt Output. For The Integration
With The Control System ./ PLC.
16 Display Inline Display For Flow Rate, Keypad On The
Mass Flow Controller For Setting The Values.
17 Temperature compensation For 0 to 50 Degree Celcius
18 Calibration for Media Helium.