Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
2. Both the test rule and the gauge block must be in temperature equilibrium
with the environment of the length laboratory.
3. Place the test rule and the gauge block on the length bench or similar flat
surface, parallel to one another using steel straight edge (not applicable to
flexible scale with stopper) in the reference point.
4. Observe readings from left to right sides of the corresponding graduation of the
standard rule and record the measurement value to the nearest 0.1.
5. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
2. Both the vernier caliper and the gauge block must be in temperature equilibrium
with the environment of the length laboratory.
3. Zero the Vernier caliper at the start and adjust as required by the manufacturers’
specifications. If you cannot zero it then mark it as a fail.
4. When testing the vernier caliper, one of the points must be near the lower limit,
middle part and near the upper limit that the instrument can measure.
5. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
STANDARD CALIBRATION FOR DEPTH GAUGE
2. Both the depth gauge and the gauge block must be in temperature equilibrium
with the environment of the length laboratory.
3. Zero the depth gauge at the start when placed directly on the steel straight edge
and in the gauge block. If you cannot zero it then mark it as a fail.
4. When testing the depth gauge, one of the points must be near the lower limit,
middle part and near the upper limit that the instrument can measure.
5. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
2. Both the outside micrometer and the gauge block must be in temperature
equilibrium with the environment of the length laboratory.
3. Zero the outside micrometer at the start and adjust as required by the
manufacturers’ specifications. If you cannot zero it then use a gauge block and
adjust the micrometer to that value.
4. When testing the outside micrometer, one of the points must be near the lower
limit that the instrument can measure, another somewhere in the middle, and the
third near the upper limit.
5. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
STANDARD CALIBRATION FOR INSIDE MICROMETER
2. Both the inside micrometer and gauge blocks must be in temperature equilibrium
with the environment of the length laboratory.
3. Set first the inside micrometer in zero (0) if not, use the ratchet to adjust it.
Gauge block is assembled using block clamp and jaws in the desired test gauge
dimension.
4. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
2. Both the dial gauge and the calibration tester must be in temperature equilibrium
with the environment of the length laboratory.
3. Set up the dial gauge in the calibration tester then set both instrument in zero as
its reference point. Then set the dial gauge in your preferred test gauge
dimension and read the measured value by the calibration tester.
4. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
STANDARD CALIBRATION FOR DIAL CALIPER GAUGE
2. Both the dial caliper gauge and the gauge blocks must be in temperature
equilibrium with the environment of the length laboratory.
3. Set first the dial caliper gauge to zero (0) measurement value and then position
the dial caliper gauge’s teeth correctly in the gauge block.
4. Repeat the measurement up to three (3) times; then get the average and the
difference of maximum and minimum.
STANDARD
CAPACITY TOLERANCE
TOOLS/EQUIPMENT METHOD/REFERENCE
(mm) (mm)
STANDARD
STEEL RULER &
500 max ±0.15
FLEXIBLE SCALE
over 500-
±0.25
1000 max
JIS B 7516
over 1000-
±0.25
1500 max
over 1500-
±0.30
2000 max
VERNIER CALIPER 0-150 ±0.05
0-200 ±0.05
0-300 ±0.08
JIS B 7507
DIGITAL CALIPER 0-150 ±0.02
0-200 ±0.02
0-300 ±0.03
DEPTH GAUGE 0-150 ±0.02
0-200 ±0.02 JIS B 7518
0-300 ±0.03
OUTSIDE MICROMETER 0-25 ±0.001
25-50 ±0.001
50-75 ±0.001
75-100 ±0.002
100-125 ±0.002 JIS B 7502
125-150 ±0.002
0-25
INSIDE MICROMETER micrometer ±0.002
head
DIAL GAUGE 0.002-0.28 ±0.002
0-0.8 ±0.002
JIS B 7503
0-3 ±0.006
0-10 ±0.013
DIAL CALIPER GAUGE 0-50 ±0.15
OZAKI MFG.
0-80 ±0.20