USER MANUAL Dear Customer, FRONT PANEL & DISPLAY TECHNICAL DATA

Thank you for choosing an Adwa product.


Please read carefully this manual before Range -2.0 to 16.0 pH (AD11)
starting operations.
-2.00 to 16.00 pH (AD12)
These instruments are in compliance with the
AD11 & AD12 EMC Directive 89/336/EEC and Low Voltage
-5.0 to 60.0 °C / 23.0 to 140.0 °F
Waterproof pH Testers Directive 73/23/EEC for electrical equipments. Resolution 0.1 / 0.01 pH
For additional technical information, please 0.1 °C / 0.1 °F
e-mail us at sales@adwainstruments.com. Accuracy (@25 °C/77 °F)
±0.1 / ±0.01 pH
±0.5 °C / ±1 °F
pH Calibration
Automatic, 1 or 2 point
with 2 sets of memorized buffers
(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)
INTRODUCTION pH Electrode AD11P (included)
Temperature Compensation
AD11 and AD12 are waterproof pH and Automatic
temperature testers. The housing has been Battery Type 4 x 1.5V button type
1. Dual line LCD
completely sealed against humidity.
2. ON/OFF/ MODE button Battery Life Approx. 300 h of use
All pH readings are automatically temperature Auto-off After 8 minutes of non-use
compensated (ATC), and temperature values 3. pH electrode & temperature sensor
Environment
can be displayed in °C or °F units. 4. Electrode body
-5 to 50 °C (23 to122 °F); RH 100%
The meter can be calibrated at one or two 5. Battery compartment (inside)
Dimensions 175.5 x 39 x 23 mm
points for pH with auto-buffer recognition 6. SET/HOLD button Weight 100 g
and against five memorized buffer values.
7. Clip holder
Measurements are highly accurate with a
unique stability indicator right on the LCD.
The models are also provided with a low
battery symbol which warns the user when
the batteries need to be replaced. ELECTRODES & SOLUTIONS
The AD11P pH electrode, supplied with the
meter, is interchangeable and can be easily AD11P Spare pH electrode
replaced by the user. AD70004P pH 4.01 buffer solution, 20 ml, 25 pcs.
The encapsulated temperature sensor allows AD7004 pH 4.01 buffer solution, 230 ml bottle
fast and accurate temperature measurement 1. Stability indicator (hourglass symbol) AD70007P pH 7.01 buffer solution, 20 ml, 25 pcs.
and compensation. 2. Low battery warning indicator AD7007 pH 7.01 buffer solution, 230 ml bottle
3. Calibration mode / calibrated meter indication AD70010P pH 10.01 buffer solution, 20 ml, 25 pcs.
Each meter is supplied complete with:
4. ATC (Automatic Temperature Compensation) AD7010 pH 10.01 buffer solution, 230 ml bottle
• AD11P pH electrode indicator
AD7061 Cleaning solution, 230 ml
• 4 x 1.5V batteries, button type 5. Secondary LCD level
09/14

AD70300 Storage solution, 230 ml


• User manual 6. Primary LCD level
ISTAD12

w w w. a d w a i n s t r u m e n t s . c o m 7. Measurement unit for primary level


OPERATIONAL GUIDE a) The pH electrode is replaced. 9.18) is detected, then the meter completes ELECTRODE MAINTENANCE
b) After testing aggressive chemicals. the calibration procedure. The LCD shows
c) Where high accuracy is required. the accepted value with the “OK 2” message,
Turn the meter on and then the meter returns to the normal mode. When not in use, rinse the electrode with
d) At least once a month.
• Press and hold the ON/OFF/MODE button water and store it with a few drops of
Calibration Procedure Note: When the calibration procedure is
until the LCD lights up. All the used completed, the CAL tag is turned on. AD70300 storage solution in the protective
segments will be visible for one second • From normal measuring mode, press and cap.
hold the ON/OFF/MODE button until OFF To quit calibration & reset to default values
(or as long as the button is pressed). • After entering the calibration mode and NEVER STORE THE ELECTRODE IN
on the secondary LCD is replaced by CAL.
Freeze the display before the first point is accepted, you can DISTILLED OR DEIONIZED WATER!
• Release the button. The LCD enters the
• While in measurement mode, press the quit the procedure and return to the last • If the electrode has been left dry, soak the
calibration mode displaying “pH 7.01
SET/HOLD button. The reading will be calibration data by pressing the ON/OFF/ tip in storage solution for at least one hour
USE” (or “pH 6.86 USE” if the NIST buffer
frozen on the LCD. Press any button to MODE button. The secondary LCD displays to reactivate it.
set was selected).
return to normal mode. “ESC”for one second and then the meter • To prolong the electrode life, it is recommended
• The meter automatically recognizes buffers:
Turn the meter off returns to the normal measuring mode. to clean it monthly by immersing the tip
if a valid buffer is detected then its value
• While in measurement mode, press the • To reset to the default calibration values, press in AD7061 general cleaning solution for
is shown on the primary display and REC
ON/OFF/MODE button. OFF will appear SET/HOLD after entering the calibration 30 minutes. Afterwards, rinse it thoroughly
appears on the secondary LCD level. If no
on the secondary display. Release the mode and before the first point is accepted. with tap water and recalibrate the meter.
valid buffer is detected, the meter keeps
button. The secondary LCD displays “CLR” for • The electrode can be easily replaced by
the USE indication active for 12 seconds,
one second, the meter resets to the default unscrewing the body as shown below.
Note: If measurements are taken in different followed by WRNG, indicating the buffer
calibration and the CAL tag disappears.
samples successively, rinse the being measured is not a valid calibration value.
probe thoroughly to eliminate Single-Point Calibration
cross-contamination. After cleaning, • For a single-point calibration with buffers
rinse the probe with some of the pH 4.01, 9.18 or 10.01, the meter automatically SETUP
sample to be measured. accepts the calibration when the reading
is stable. The accepted buffer is displayed Setup mode allows the selection of temperature
together with the message “OK 1” for one unit and pH buffer set.
pH MEASUREMENT & BATTERY REPLACEMENT
second, then the meter automatically returns To enter the setup mode, press the ON/OFF/
CALIBRATION to the normal measuring mode. MODE button until CAL on the secondary
• If a single-point calibration with buffer pH display is replaced by TEMP and the current
When the batteries become weak, the battery
Taking Measurement 7.01 (or pH 6.86) is desired, then after the temperature unit (e.g. TEMP °C). Then:
symbol on the LCD lights up to indicate a
• Submerge the electrode in the solution to calibration point has been accepted, press For °C/°F selection: low battery condition. Batteries should be
be tested while stirring it gently. the ON/OFF/MODE button to return to • Use the SET/HOLD button. After the replaced soon.
normal mode. The meter shows “7.01” temperature unit has been selected, press
• Measurements should be taken when the To change the batteries, unscrew and release
(or“6.86”) and “OK 1” for one second, then the ON/OFF/MODE button to enter the
stability indicator (hourglass) disappears. the electrode body.
it automatically returns to the normal mode. buffer set selection mode.
• The pH value automatically compensated Take out the battery
Note: For better accuracy, it is recommended • Press the ON/OFF/MODE button again to
for temperature is shown on the primary compartment and
to perform a 2-point calibration. return to the normal measuring mode.
LCD level while the secondary one shows carefully replace all
the sample temperature. Two-Point Calibration four batteries while
To change the calibration buffer set:
• For a two-point calibration, place the electrode paying attention to
Note: Before taking any pH measurement, • After setting the temperature unit, the meter
in pH 7.01 (or pH 6.86) buffer. After the their polarity.
make sure the meter has been calibrated will show the current buffer set: “pH 7.01
first calibration point has been accepted, Reattach and tighten
(CAL tag is displayed). BUFF” (for 4.01/7.01/10.01) or “pH 6.86
the “pH 4.01 USE” message appears. The the electrode body
pH Calibration BUFF” (for NIST 4.01/6.86/*9.18).
message is held for 12 seconds, unless a properly to ensure a
For better accuracy, frequent calibration of valid buffer is recognized. If no valid buffer • Change the set with the SET/HOLD button,
then press ON/OFF/MODE to return to the watertight seal.
the instrument is recommended. In addition, is recognized, then the WRNG message is
the instrument must be recalibrated whenever: shown. If a valid buffer (pH 4.01, 10.01, or normal measurement mode.