Sei sulla pagina 1di 12

Agilent

8 Hints for Successful


Impedance Measurements
Application Note 346-4

Characterizing Electronic Components


to Achieve Designed Circuit Performance

<MEAS DISPLAY> SYS MENU


MEAS
FUNC : Z- θ RANGE : AUTO DISP
FREQ : 100.000 kHz BIAS : 0.000 V
LEVEL : 1.00 V INTEG : LONG BIN
No.

Z : 150.000 Ω BIN

θ : -90.000 deg
COUNT

LIST
SWEEP
Vm : 50.00mV Im : 10.74mA
CORR: OPEN, SHORT
Contents Impedance
HINT 1. Impedance Parameters
Measurements for
Engineers
HINT 2. Measurements Depend
Impedance is measured using a vari- The techniques employed by these
on Test Conditions
ety of techniques. A particular tech- instruments are independent of ana-
nique is selected according to the lyzer type, and can be RF-IV, IV or
HINT 3. Choose Appropriate test frequency, the impedance Auto-Balancing Bridge (depending
Instrument Display parameter to be measured and the on frequency).
Parameter preferred display parameters.
Engineers perform impedance meas-
HINT 4. A Measurement The Auto-Balancing Bridge urements for a variety of reasons. In
Technique has technique is exceptionally accurate a typical application, an electronic
Limitations over a broad impedance range (mΩ component used in a new circuit
to the order of 100 MΩ). The fre- design is characterized. Normally,
HINT 5. Perform Calibration quency range this technique can component manufacturers state only
be applied to is from a few Hz to nominal impedance values.
110 MHz.
HINT 6. Perform Compensation Design decisions, as well as deci-
The IV and RF-IV techniques are sions affecting the production of the
HINT 7. Understanding Phase also very accurate over a broad assembled product, depend to some
Shift and Port Extension impedance range (mΩ to MΩ). degree on the impedance values
Effects These techniques can be applied attributed to the product’s compo-
from 40 Hz to 3 GHz. nents. The performance and quality
HINT 8. Fixture and Connector of the final product are therefore
The Transmission/Reflection determined in part by the accuracy
Care
technique is applied over the and thoroughness with which its
broadest frequency range (5 Hz to components are characterized.
110 GHz). This technique delivers
exceptional accuracy near 50 Ω or This application note provides help-
75 Ω. ful information for using the Auto-
Balancing Bridge, IV and RF-IV tech-
LCR meters and impedance analyz- niques. Refer to Agilent Application
ers are differentiated primarily by Note 1291-1, 8 Hints for Making
display properties. An LCR meter Better Network Analyzer
displays numeric data, while an Measurements (literature number
impedance analyzer can display data 5965-8166E) for information on the
in either numeric or graphic formats. Transmission/Reflection technique.

Figure 0-1. Accuracy Profile

2
HINT 1.
Impedance
Parameters
Impedance is a parameter used to The Quality Factor (Q) and the
evaluate the characteristics of elec- Dissipation Factor (D) are also
tronic components. Impedance (Z) derived from resistance and reac-
is defined as the total opposition a tance. These parameters serve as
component offers to the flow of an measures of reactance purity. When
alternating current (AC) at a given Q is larger or D is smaller, the quali-
frequency. ty is better. Q is defined as the ratio
of the energy stored in a component
Impedance is represented as a com- to the energy dissipated by the com-
plex, vector quantity. A polar coordi- ponent. D is the inverse of Q. D is
nate system is used to map the also equal to “tan δ”, where δ is the
vector, where quadrants one and dielectric loss angle (δ is the com-
two correspond respectively to pas- plementary angle to θ, the phase
sive inductance and passive capaci- angle). Both D and Q are dimension-
tance. Quadrants three and four less quantities.
correspond to negative resistance.
The impedance vector consists of a Figure 1-2 describes the relationship
real part, resistance (R), and an between impedance and these
imaginary part, reactance (X). derived parameters.

Figure 1-1 shows the impedance


vector mapped in quadrant one of
the polar coordinate system.

Capacitance (C) and inductance (L)


are derived from resistance (R) and
reactance (X). The two forms of
reactance are inductive (XL) and
capacitive (XC).

+j
Z
jX L Inductor
R jX L XL
Q=
R
δ X L = 2πfL
θ = ωL
R
θ
Capacitor
R - jX C
δ R
D=
XC
1
jX C XC =
Z 2πfC
1
=
-j ωC

Figure 1-1. Impedance Vector Figure 1-2. Capacitor and Inductor Parameters

3
HINT 2.
Measurements
Depend on Test
Conditions
A manufacturer’s stated impedance The AC voltage across the compo-
values represent the performance of nent can be derived from the com-
a component under specific test ponent’s impedance, the source
conditions, as well as the tolerance resistance, and the signal source
permitted during manufacture. output (Figure 2-3).
When circuit performance requires
more accurate characterization of a An automatic level control (ALC)
component, it is necessary to verify function maintains a constant volt-
the stated values, or to evaluate age across the DUT (device under
component performance under oper- test). It is possible to write an ALC
ating conditions (usually different program for instruments that have a
than the manufacturer’s test condi- level monitor function, but not a
tions). Figure 2-1. Frequency Characteristics of a Capacitor built-in ALC.

Frequency dependency is common Signal level (AC) dependency is Control of measurement integration
to all components because of para- exhibited in the following ways time allows reduction of unwanted
sitic inductance, capacitance and (see Figure 2-2): signals. The averaging function is
resistance. used to reduce the effects of random
• Capacitance is dependent on AC noise. Increasing the integration
Figure 2-1 describes ideal and para- voltage level (dielectric constant, time or averaging allows improved
sitic frequency characteristics of a K, of the substrate). precision, but with slower measure-
typical capacitor. ment speed. Detailed explanations of
• Inductance is dependent on AC these test parameters can be found
current level (electromagnetic in the instrument operating manuals.
hysteresis of the core material).
Other physical and electrical factors
that affect measurement results
include DC bias, temperature,
humidity, magnetic fields, light,
atmosphere, vibration, and time.

Source Resistance V dut


Rs
Signal R X
Dielectric Source
Constant High Dielectric DUT I dut
Constant Medium Vo V A
C L
Dielectric
Constant Low (No Dependency)
0
0 (No Dependency) R2 + X 2
V dut = Vo *
(Rs+R) 2 + X 2
Feedback
AC Test Voltage AC Test Current
(a) AC Voltage Level Dependency of (b) AC Current Level Dependency of Vm
Ceramic Capacitor Measurement Core Inductor Measurement
DUT
Rs
A
˜ Vo

Figure 2-2. Signal Level Dependency Figure 2-3. Applied Signal and Constant Level Mechanism

4
HINT 3.
Choose Appropriate
Instrument Display
Parameter
Many modern impedance measuring No circuit components are purely
instruments measure the real and resistive or purely reactive. A typical
the imaginary parts of an impedance component contains many parasitic
vector and then convert them into elements. With the combination of
the desired parameters. primary and parasitic elements, a
component acts like a complex cir-
When a measurement is displayed as cuit.
impedance (Z) and phase (θ), the
primary element (R, C, or L) as well Recent, advanced impedance analyz-
as any parasitics are all represented ers have an Equivalent Circuit
in the |Z| and θ data. Analysis Function that allows analy-
sis of the measurement result in the
When parameters other than imped- form of three- or four-element cir-
ance and phase angle are displayed, cuit models (Figure 3-2). Use of this
a two-element model of the compo- function enables a more complete
nent is used. These two-element characterization of a component's
models are based on a series or par- complex residual elements.
allel circuit mode (Figure 3-1), and
are distinguished by the subscripts
“p” for parallel and “s” for series (Rp,
Rs, Cp, Cs, Lp, or Ls).

Figure 3-1. Measurement Circuit Mode Figure 3-2. Equivalent Circuit Analysis Function

5
HINT 4.
A Measurement
Technique Has
Limitations
The most frequently asked question Instrument accuracy specifications
in engineering and manufacturing is for D or Q measurements are usually
probably: “How accurate is the different than accuracy specifica-
data?” tions for other impedance terms.

Instrument accuracy depends on the In the case of a low-loss (low-


impedance values being measured, D/high-Q) component, the R-value is
as well as the measurement technol- very small relative to the X-value.
ogy employed (see Figure 0-1). Small changes in R result in large
changes in Q (Figure 4-2).
To determine the accuracy of a
measurement, compare the meas- The measurement error is on the
ured impedance value of the DUT to order of the measured R-value. This
the instrument accuracy for the can result in negative D or Q values.
applicable test conditions.
Be aware that measurement error
Figure 4-1 shows that a 1 nF capaci- includes error introduced by the
tor measured at 1 MHz exhibits an instrument as well as by the test fix-
impedance of 159 Ω. ture.

1 nF at 1 MHz
10 10
10 0 1 1
nF 1n pF 0pF 1pF 0fF 0fF 1f
F
10 F
10M 0n
F
1M 1u
F
Impedance value
10 of capacitor
100K u
10 F
0u
1
Impedance (Ω)

10K F
1m Z=
159 1K F ωC
10
100 10 mF
0m (ω = 2πf)
10 F
1
100m
1 10 100 1K 10K 100K 1M10M 100M 1G
Test frequency (Hz)

Figure 4-1. Capacitor Impedance and Test Frequency Figure 4-2. Concept of the Q Error

6
HINT 5.
Perform Calibration
Calibration is performed to define a Some instruments offer the choice of
reference plane where the measure- fixed-mode or user-mode calibration.
ment accuracy is specified (Figure Fixed-mode calibration measures
5-1). Normally, calibration is per- calibration standards at predeter-
formed at the instrument's test port. mined (fixed) frequencies. Calibra-
Corrections to raw data are based on tion data for frequencies between
calibration data. the fixed, calibrated points are inter-
polated.
A baseline calibration is performed
at service centers for Auto-Balancing Fixed-mode calibration sometimes
Bridge instruments such that the results in interpolation errors at fre-
specified accuracy can be realized quencies between the fixed, calibrat-
for a period of time (usually twelve ed points. At higher frequencies,
months) regardless of the instru- these errors can be substantial.
ment settings. With these instru-
ments, operators do not require User-mode calibration measures cali-
calibration standards. bration standards at the same fre-
quency points the user has selected
Baseline calibration for non-Auto- for a particular measurement. There
Balancing Bridge instruments can be no interpolation errors asso-
requires that a set of calibration ciated with user-mode calibration.
standards be used after instrument
initialization and setup. This hint It is important to recognize that the
provides information that may be operator-established calibration is
helpful when using calibration stan- valid only for the test conditions
dards to establish calibration for (instrument state) under which cali-
these instruments. bration is performed.

Calibration Defines a Reference Plane at


Which Measurement Accuracy is Specified.
Z Analyzer Calibration Standards
Low-Loss
Short Open Load Capacitor*

Z? OS 50 Ω –90 o
0Ω

*Agilent 4291B,
Calibration Plane 4286A and 4287A

Figure 5-1. Calibration Plane

7
HINT 6.
Perform
Compensation
Compensation is not the same as It is important to verify that error Perform load compensation when
calibration. The effectiveness of compensation is working properly. In the measurement port is extended a
compensation depends on the general, the impedance value for an non-standard distance, the configu-
instrument calibration accuracy, open condition should be greater ration uses additional passive cir-
therefore compensation must be than 100 times the impedance of the cuits/components (for example, a
performed after calibration has been DUT. In general, the impedance balun, attenuator, or filter), or when
completed. value for a short condition should be a scanner is used. The impedance
less than 1/100 the impedance of the value of the load must be accurately
When a device is directly connected DUT. known. A load should be selected
to the calibration plane, the instru- that is similar in impedance (at all
ment can measure within a specified Open compensation reduces or elim- test conditions) and form-factor to
measurement accuracy. Since a test inates stray capacitance, while short the DUT. Use a stable resistor or
fixture or adapter is usually connect- compensation reduces or eliminates capacitor as the LOAD device.
ed between the calibration plane and the unwanted resistance and induc-
the device, the residual impedance tance of fixturing. It is practical to measure a load
of the interface must be compensat- using open/short compensation and
ed for to perform accurate measure- When performing an open or a short a non-extended fixture to determine
ments. measurement, keep the distance the load impedance. The values
between the UNKNOWN terminals measured can then be input as com-
Additional measurement error intro- the same as when the DUT is pensation standard values.
duced by a test fixture or adapter contacted. This keeps parasitic
can be substantial. The total meas- impedance the same as when meas-
urement accuracy consists of the urements are performed.
instrument accuracy plus the error
from sources that exist between the
DUT and the calibration plane.

Figure 6-1. OPEN/SHORT Compensation

8
HINT 7.
Understanding Phase
Shift and Port
Extension Effects
Cable length correction, port exten- When working in the RF region, cali-
sion, or electrical delay is used to bration should be performed at the
extend or rotate the calibration end of an extension cable. If calibra-
plane to the end of a cable or the tion standards cannot be inserted,
surface of a fixture. This correction port extension can be used in this
reduces or eliminates phase shift region for short and well-character-
error in the measurement circuit. ized distances.

When the measurement port is When using the Auto-Balancing


extended away from the calibration Bridge technique with non-standard
plane (Figure 7-1), the electrical cables or extensions, open/short/load
characteristics of the cable affect the compensation should be performed
total measurement performance. To at the terminus of an extension or
reduce the resulting effects: fixture. Auto-Balancing Bridge prod-
ucts use cable length compensation
• Make measurement cables as for standardized test cables (1, 2, or
short as possible. 4 meters). At the terminus of the
standardized extension cable, the
• Use well-shielded, coaxial cables shields should normally be connect-
to prevent influence from ed together.
external noise.
Port extension in any form has limi-
• Use low-loss coaxial cables to tations. Since any extension will
keep from degrading accuracy, contribute to losses in the measure-
since the port extension method ment circuit and/or phase error, it is
assumes lossless cable. imperative that the limitations of the
measurement technique be fully
A phase-shift-induced error occurs understood prior to configuring an
due to the test fixture, which can extension.
not be reduced using OPEN/ SHORT
compensation.

Impedance
Measuring Device Under Test
Instrument

Extension Cable Test


Fixture

Figure 7-1. Measurement Port Extension

9
HINT 8.
Fixture and
Connector Care
High-quality electrical connections Technique
insure the capability to make precise The use of a torque wrench and
measurements. At every connection, common sense insures that damage
the characteristics of the mating sur- does not occur when making repeat-
faces vary with the quality of con- ed connections. Damage includes
nection. An impedance mismatch at scratching and deformation of the
mating surfaces will influence propa- mating surfaces.
gation of the test signal.
Maintenance
Attention should be paid to the mat- Many mating surfaces are designed
ing surfaces of test ports, adapters, to allow for the replacement of parts
calibration standards, fixture con- that degrade with use. If a mating
nectors, and test fixtures. The quali- surface cannot be repaired, regularly
ty of connections depends on the scheduled replacement is advised.
following:
Cleanliness
• composition The use of non-corrosive/non-
• technique destructive solvents (such as de-ion-
• maintenance ized water and pure isopropyl
• cleanliness alcohol) and lint-free wipes insures
• storage that the impedance at mating sur-
faces is not influenced by residual
Composition oils or other impurities. Note that
It has been said that a chain is as some plastics are denatured with the
strong as the weakest link. The same use of isopropyl alchohol.
is true for a measurement system. If
low-quality cables, adapters or fix- Storage
tures are used in a test system, the If a case is not provided with an
overall quality of the system is accessory, plastic caps should be
reduced to that of the lowest-quality used to cover and protect all mating
interface. surfaces when not in use.

10
Agilent Technologies’
Impedance Product
Lineup
Agilent offers the widest selection of programming, which ease evaluation Combination Analyzers
impedance measuring equipment for of measurement results. They also Combination analyzers from Agilent
your applications. An overview of have features that enable character- provide three capabilities—vector
these instruments is given below. istic evaluations for R&D, as well as network, spectrum and impedance
For more information, refer to the reliability evaluations (including measurements—in one box. These
product literature listed at the end temperature characteristics) for QA instruments deliver broad function-
of this note. purposes. ality to engineers over a wide range
of applications from circuit design to
LCR Meters component evaluation.
LCR meters can easily and accurate-
ly evaluate components such as
capacitors, inductors, transformers
and electromechanical devices. The
ability of these instruments to apply
specific measurement conditions
(such as test frequency and signal
level) is important in the R&D, pro-
duction test and QA environments.

Network Analyzers
Network analyzers allow impedance
measurements at RF and microwave
frequencies using the Transmission/
Reflection technique. Their graphical
displays have markers and program-
ming functions that simplify the
analysis of measurement results.
Agilent network analyzers are suit-
able for both R & D and QA use.

Impedance Analyzers
Agilent impedance analyzers can
measure characteristic changes in
component performance resulting
from changes in specific measure-
ment conditions. The characteristic
changes in performance can be dis-
played in a graphical format. These
analyzers provide sophisticated
functions, such as markers and

11
Product Literature For more assistance with your test &
measurement needs go to

1. LCR Meters, Impedance www.agilent.com/find/assist


Analyzers, and Test Fixtures Or contact the test and measurement
Selection Guide, literature experts at Agilent Technologies
number 5952-1430E. (During normal business hours)

United States:
2. RF Economy Network Analyzer, (tel) 1 800 452 4844
literature number 5967-6316E.
Canada:
(tel) 1 877 894 4414
Key Web Resourses (fax) (905) 206 4120

Europe:
(tel) (31 20) 547 2000
www.agilent.com/find/component_test
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840

Latin America:
(tel) (305) 267 4245
(fax) (305) 267 4286

Australia:
(tel) 1 800 629 485
(fax) (61 3) 9272 0749

New Zealand:
(tel) 0 800 738 378
(fax) 64 4 495 8950

Asia Pacific:
(tel) (852) 3197 7777
(fax) (852) 2506 9284

Product specifications and descriptions


in this document subject to change
without notice.

Copyright © 2000
Agilent Technologies
Printed in USA 06/00
5968-1947E

12

Potrebbero piacerti anche