Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Burst pressures of occluded CVCs 95% CI 1521.4 to 1838.7; p = 0.51). New silicone CVCs had a
Twenty three polyurethane (12 new and 11 used) and 14 silicone higher burst pressure (296.3 kPa, 95% CI 267.2 to 325.4) than used
(11 new and three used) CVCs were tested. Catheters were silicone CVCs (199.7 kPa, 95% CI 45.5 to 353.9) (p = 0.008).
obstructed at the tip with a short length of wire glued with The nature of rupture was different for the two CVCs.
cyanacrylate for polyurethane CVCs and by a mechanical clamp Polyurethane CVCs showed coiling with increasing pressure
for silicone CVCs. CVCs were placed unrestrained in a normal followed by a localised longitudinal 1–3 mm split that occurred
saline bath at 37uC. Water coloured with methylene blue was in various places along the catheter length. Silicone CVCs
infused via the pump at a set rate of 3 ml/h to achieve a slowly developed a localised bleb either at the hub or within 2.5 cm of
ramped pressure increase. Rupture was indicated by methylene the obstructed tip before longitudinal splitting at this site of focal
blue efflux into the bath. Burst pressure was recorded, and the thinning. No catheters split transversely or fragmented at bursting.
sites and type of rupture were documented using a microscope.
Pressures during constant flow in patent CVCs
Pressures during constant flow in patent CVCs Figure 2 shows the flow–pressure relationships for the two
Twenty two polyurethane (11 new and 11 used) and 14 silicone catheter types. Intraluminal pressure was linearly related to
(11 new and three used) CVCs were tested without obstruction flow for both CVCs (p,0.001) with a statistically greater
with the same temperature and infusate as used above. increase in pressure per unit increase in flow for the
Incremental flow rates from 5 to 499 ml/h were applied and polyurethane CVCs compared to the silicone CVCs (1.64
pressures recorded after several seconds at each flow rate once (95% CI 1.57 to 1.71) vs 0.37 (95% CI 0.29 to 0.44); p,0.001).
stabilised. Each catheter was tested once. There was no statistically detectable effect of new versus used
CVCs when this variable was included in the model.
Pressure changes during manual flushing For polyurethane CVCs, the package insert recommended
Sixteen new polyurethane and six new silicone CVCs were tested safe pressure (150 kPa) was exceeded at 100 ml/h but remained
with a different operator (neonatal intensive care nurse or doctor) below our measured burst pressures at 499 ml/h. For silicone
flushing each CVC. A standard central line infusion set-up was CVCs, pressure exceeded the recommended safe pressure limit
assembled on a bench top. A manometer was connected at the (100 kPa) at 300 ml/h and reached our range of measured burst
proximal end of the CVC. The catheter and manometer were pressures at 400 ml/h.
hidden behind a drape to blind the operator to pressure readings
and the cause of obstruction. Two clinical scenarios were provided Manual flushing
to each operator. In the first ‘‘bolus’’ scenario, the operator was The peak pressures generated by flushing obstructed polyur-
asked to flush the CVC with normal saline to clear the line of ethane CVCs (348.9 kPa, 95% CI 295.1 to 403.3) were greater
medication using 1, 2 and 10 ml syringes. They were instructed to than for patent polyurethane CVCs (169.6 kPa, 95% CI 142.7 to
flush as they would in normal practice and to flush proximal to the 196.5) (p,0.001). Higher pressures were generated flushing the
in-line bacterial filter. In the second ‘‘obstruction’’ scenario, the obstructed polyurethane CVCs with a 1 ml syringe in comparison
same CVC was obstructed at its tip and the operator was asked to to a 10 ml syringe, with similar pressures otherwise noted
clear the line flushing distally to the filter with the same choice of between syringe sizes (1 ml syringe 451.6 kPa, 95% CI 365.4 to
syringes. Peak pressures were recorded for each flush episode. The 537.8; 2 ml syringe 333.7 kPa, 95% CI 242.0 to 425.4; 10 ml
limit of measurement of the digital manometer was 689.5 kPa; syringe 239.2 kPa, 95% CI 157.2 to 321.3; p = 0.001 1 ml vs 10 ml
readings above this limit were recorded as 689.5 kPa for analysis. using one-way ANOVA). In all cases pressures were below our
observed range of burst pressures and no polyurethane CVCs
Statistical methods ruptured. No statistical differences in pressures generated were
Flow–pressure relationships were analysed using a linear noted for different syringe sizes when flushing patent lines.
regression model with random slopes in order to account for For silicone CVCs, peak pressures generated were higher in
the correlated structure of the data across flow rate. the obstructed group (245.5 kPa, 95% CI 146.9 to 343.4) as
Comparison of normally distributed data used an independent compared to the patent group (36.5 kPa, 95% CI 18.6 to 55.2)
samples t test or one way ANOVA with the Tukey post hoc (p = 0.002) and similar to the pressures measured in obstructed
multiple comparison test as appropriate. Categorical data were polyurethane CVCs (p = 0.057). No differences in pressures
analysed using the x2 statistic. Statistical calculations used SPPS were noted between syringe sizes for flushing silicone CVCs for
15.0 for Windows 2006 (SPSS) and Stata v 10 (StataCorp). either patent or obstructed catheters. Rupture of obstructed
silicone CVCs was frequent during flushing (5/6 silicone CVCs
RESULTS ruptured compared to 0/16 polyurethane CVCs; p,0.001).
Rupture occurred with both 1 ml and 10 ml syringe sizes. No
Burst pressures of occluded CVCs
ruptures occurred when flushing patent silicone catheters.
Time to rupture was 400–500 s for all CVCs. Figure 1 shows the
burst pressures for occluded catheters. The mean burst pressure
for polyurethane CVCs (1730.8 kPa, 95% CI 1634.7 to 1826.8) DISCUSSION
was higher than for the silicone CVCs (275.6 kPa, 95% CI 240.4 These data document the pressures generated under different
to 310.8) (p,0.001). The mean and lowest burst pressures for circumstances in two commonly used CVCs and help to quantify
the polyurethane CVCs were 11.5 and 8.7 times, respectively, the risks of rupture. Product information gives a similar safe
above the safety limit of 150 kPa noted in the packaging insert. pressure limit for both CVCs (150 kPa for the 27-gauge
The mean and lowest burst pressures for the silicone CVCs polyurethane Premicath, and 100 kPa working pressure or
were 2.6 and 1.6 times, respectively, above the working safety 120 kPa bolus pressure for the 24-gauge silicone ECC).
limit of 100 kPa also noted in the packaging insert. However, our data show an approximately fivefold greater safety
New and used polyurethane CVCs had similar mean burst margin before burst for the polyurethane CVC in comparison to
pressures (1777.3 kPa, 95% CI 1645.4 to 1909.1 vs 1680.1 kPa, the silicone CVC. Silicone CVCs showed a narrow safety margin