INSTRUCTIONS for AUTO calibration. A <circle A> icon will 1.

Remove meter end-cap & press [ON/OFF]


be displayed when in AUTO-CAL mode. key, to turn meter ON.
CHEKMITE pH-10 3. Press the [ON/OFF] key. 2. Immerse tip of sensor in solution to be
Optional Accessories:
measured, to a depth of approx. 1”. 473040 Replacement Sensor.
The Chekmite pH-10 is a portable pH meter in Automatic Calibration (<circle A> icon 473676 Buffer Sachet Assortment
3. Press [READ] key a “beep” will be heard;
a waterproof and chemical resistant case. The displayed) 4, 7, 10pH (32/pk).
display will change towards pH value of
meter features LCD Readout & Automatic 478574 Buffer Rainbow Pk. - 4, 7, 10pH
1. Remove meter end-cap & press [ON/OFF] sample. When reading stabilizes, press
Calibration. The replaceable epoxy body (2 x 500mL bottles each/case).
key, to turn meter ON. [READ] key again, a “beep” will be heard
sensor is 8mm in diameter, which enables 470530 Replacement Batteries (2/pk).
2. Immerse tip of sensor in pH7.00 buffer to and the display will “freeze”; note value.
testing in slim or narrow vessels.
a depth of approx. 1”. 4. Rinse sensor tip with distilled or de-ionized
Specifications: 3. Press [CAL] key; a “beep” will be heard water & blot to remove excess water. Warranty:
and the <CAL> icon will flash while 5. Repeat steps 2, 3 & 4 for further samples. NOVA-pH warrants this product to be free from
Range: 0.00 to 14.00 pH endpoint is being tracked. At endpoint the Note: To conserve battery life, meter will defects in materials and workmanship for a
Resolution: 0.01 pH display will “freeze” to 7.00, a “beep’ will automatically turn off after 8 minutes if no key is period of six months from the date of purchase.
Accuracy: ± 0.2 pH be heard and a <square 7> icon will be
Environment: 0 to 50°C pressed, but calibration data is retained in THIS WARRANTY IS MADE IN LIEU OF ALL OTHER
displayed, indicating auto-calibration to memory.
Power Supply: 2 x 3volt Lithium Cells. (CR- pH7.00. WARRANTIES EXPRESSED OR IMPLIED
2032) INCLUDING THE WARRANTIES OF
4. Rinse sensor tip with distilled or de-ionized Care and Maintenance Tips: MERCHANTABILITY AND FITNESS FOR A
Battery Life: 200 Hrs. (Approx.)
Auto-OFF: 8 minutes.
water and blot to remove excess water. • A Test Plug is provided in the meter kit to PARTICULAR PURPOSE. NOVA-pH SHALL NOT
5. For two-point calibration repeat steps 2 assist in testing of the electronic meter BE LIABLE FOR ANY LOSS OR DAMAGES ARISING
Dimensions (in): 9.3Lx1.8Wx1.3D through 4 using pH4.00 or 10.01 buffer. FROM THE USE OF THESE PRODUCTS NOR FOR
Weight: 4 Ounces module in cases where error codes are
The readout will “freeze” to the selected CONSEQUENTIAL DAMAGES OF ANY KIND.
Compliant with CE directives EN-55011 & EN- observed. To use..
buffer value and a <square 4> or <square 1. Remove the batteries for several In the event that this product fails under normal
50082-1 and waterproof to IP66+. 10> icon will display alongside the seconds, then re-install. This will clear old laboratory conditions within the specified period
Unpacking & Conditioning: <square 7> icon indicating auto-calibration calibration information. because of a defect in material or workmanship,
to a second point. 2. Remove electrode by un-screwing NOVA-pH will, at its option, repair or replace the
1. Remove CHEKMITE from packaging. Manual Calibration (<circle A> icon NOT product. Contact NOVA-pH Customer Service for
2. Locate batteries. To install, remove the retaining ring and pulling electrode return authorization and shipping instructions at 1 866
displayed). straight out of connector.
two small Phillips screws retaining panel 1. Remove meter end-cap & press [ON/OFF] 664 6682.
on rear of unit. Ensuring that the positive 3. Replace with Test Plug.
key, to turn meter ON. 4. Press [ON/OFF] key. NOVA-pH, WOBURN MA 01801
(+) pole is facing up for each cell, slide the 2. Immerse tip of sensor in pH7.00 buffer to ©2006
two batteries into the battery compartment. 5. Press [CAL] key. Display should indicate
a depth of approx. 1”. 7.00 if all is OK.
Replace rear panel, being careful to keep 3. Press [CAL] key a “beep” will be heard;
the waterproofing sealing gasket in place. 6. Replace electrode; if problem still persists
wait for readout to stabilize, then press the a replacement electrode is required.
3. Remove meter end cap. Carefully remove [READ] key. A “beep” will be heard, the
adhesive tape and small Wetting Cap, display will “freeze” to 7.00 and a <square • For storage, moisten sponge in the bottom
containing KCl, from sensing tip of 7> icon will be displayed, indicating of the end cap with a small amount of 7.00
electrode. Retain this Cap for future long- calibration to pH7.00. buffer. NEVER use distilled or de-ionized
term storage. Moisten the sponge, 4. Rinse sensor tip with distilled or de-ionized water in the cap.
located inside the end cap, with pH-7 water and blot to remove excess water. • If the tip of the sensor is left dry, it must be
Buffer. This will ensure the pH-sensing 5. For two-point calibration repeat steps 2 rejuvenated. Place sensor tip in 7.00 or
bulb is kept moist during routine storage. through 4 using pH4.00 or 10.01 buffer. 4.00 buffer for 1 hour.
4. For optimum performance always
condition the electrode prior to use. To
When pressing the [READ] key the • Samples high in protein fats or oils will
readout will “freeze” to the selected coat the sensor measuring surfaces. Keep
condition, remove end cap and soak the buffer value and a <square 4> or the sensor clean by rinsing with a solution
sensing tip of the electrode in pH7.0 buffer <square 10> icon will display alongside of half alcohol and half water after each
for ten (10) minutes. the <square 7> icon, to indicate sample. DO NOT allow the sensor to
calibration to a second point. soak in this solution.
Calibration: Daily calibration is recommended.
Either a 1 or 2-point calibration is possible Error Codes:
using 7.00 & 4.00, or 7.00 & 10.01 pH buffers. Electrode Replacement:
E1 & E2. Wrong buffer or unfit sensor during
In automatic mode, end-point detection calibration procedure. Change buffer • Un-screw & remove the retaining nut at
software will detect the sensor response and/or clean sensor. Repeat conditioning the top of the electrode body.
plateau. To select calibration mode… procedure.
Er Calibration Buffer or Sample beyond
• Unplug electrode from CHEKMITE body
1. Press the [ON/OFF] key. by pulling straight out.
2. Press the [CAL] key. With the <CAL> icon instrument range.
flashing, press & HOLD the [READ] key • Replace with new electrode and refit
for >3seconds to select either MANUAL or Sample Measurement Procedure: retaining nut, being careful to include “O”
rings as originally positioned.