Balance/Emotion/Esprit
CT

Troubleshooting Guide
System

Troubleshooting Guide System

This document is valid for:


SOMATOM Balance
SOMATOM Emotion/ Emotion Duo
SOMATOM Esprit/ Esprit+
03812950
03810160
03810152
03810137
03810079
© Siemens AG 2002
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: CT02-019.840.01.01.02 English


Replaces: C2-019.840.09.01.02 Doc. Gen. Date: 12.02
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 2 of 32 Siemens AG


12.02 CSCT Medical Solutions
Table of Contents 3
1- 0Table of Contents

1 _______ General ________________________________________________________ 5

Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Components in the gantry rear side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Components in the gantry front side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

2 _______ Generator (XRS)________________________________________________ 10

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Generator Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
HV_Drop Explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Important Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
UTact and ITact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
X-Ray timeout Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Rotating Anode Test RAC01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Filament Test FiL01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
High Voltage Test HVC01 (without rotation). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Prerequisite. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
High Voltage Test HVC02 (with rotation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
High Voltage Test HVC03 (Single Puls Test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Single pulse example: Balance / Emotion /Esprit+ . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Single pulse example: Esprit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3 _______ Data Measurement System (DMS) _________________________________ 21

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Data Measurement System (DMS) Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Disc Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Offset Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Mean and Standard Deviation Measurement without Xray . . . . . . . . . . . . . . . . . . . . . . 24
Measurement without Xray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Siemens AG CT02-019.840.01.01.02 Page 3 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
4 Table of Contents

Analyze function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Preparing of the raw data file for Analyze . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Calculation of the Mean and Standard Deviation (Noise) values . . . . . . . . . . . . . . . 26
Evaluation of the Mean and the Standard Deviation (Noise) values . . . . . . . . . . . . . 27

4 _______ Slipring System (SRS) ___________________________________________ 29

Slipring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Slipring cleaning procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

5 _______ Changes to previous version _____________________________________ 31

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 4 of 32 Siemens AG


12.02 CSCT Medical Solutions
General 5
2-
Safety Information
1General

General 0

Safety relevant aspects are decribed in

"General Safety Notes" print no. TD00-000.860.01.xx.02


"Product-specific Safety Notes” print no. CT02-019.860.01.xx.02
(CT02-019.860.01)

and within this section/chapter. This information applies to all service activities related to
the system and all delivered system components.The safety information provided in this
document (and those listed above) must be adhered to.
The installation, service and maintenance of equipment described herein is to be per-
formed by qualified personnel who are employed by Siemens or one of its affiliates, or
who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

WARNING Risk of accident and injury!


Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.
¹ The guidelines contained in this document must be
strictly followed.

High voltage

WARNING Risk of electric shock!


Dangerous voltages may be present.
¹ Switch off the main power supply before working on the
system and secure the system against unattended
switch on.
¹ Allow at least two minutes discharge time after switching
off the system. Prior to starting work on the components
indicated below. Measure the voltage level in the PDS
(right gantry stand) between D450 X7 and X9. The voltage
must be below 40 VDC.
¹ Connect the stationary and rotating parts of the gantry
with a protective conductor prior to working in the gan-
try.

Siemens AG CT02-019.840.01.01.02 Page 5 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
6 General

• Hazardous voltage levels may be present ( 565 V DC ) at the following parts even if the
system is switched off:
- DCC - Power D450
- Slip rings
- Carbon brushes
- Generator electronics

Working in the gantry


• Prior to starting work in the gantry:
- Wait until rotation has stopped
- Disable gantry rotation via service switch S5 on MAS D301
- Use the mechanical brake to prevent gantry rotation while working in the gantry.

NOTE Enable gantry rotation after finishing the work.

- Connect the stationary and rotating parts of the gantry with a protective conductor pri-
or to working in the gantry.

NOTE Remove the cable prior to rotating the gantry.

Automatic procedures

WARNING Risk of injury!


When a program is running with automatic movements or auto-
matic radiation.
¹ Do not leave the system unattended

Radiation protection

WARNING Risk of exposure to X-rays!


Certain tests and procedures use X-rays.

¹ Observe radiation protection regulations.

• Observe the following protective measures when working on the CT unit with radiation
"ON":
- use a radiation shield
- maintain a safe distance from the source, the farther the better
- use the minimum dose, smallest slice-thickness
- remove the phantom ( if possible ).

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 6 of 32 Siemens AG


12.02 CSCT Medical Solutions
General 7

Bloodborne pathogens

WARNING Infection caused by bloodborne pathogens:


Handling parts of the system that may have come into contact with
patients may lead to infections caused by bloodborne pathogens.
¹ Take appropriate precautions against exposure to blood-
borne pathogens!

Siemens AG CT02-019.840.01.01.02 Page 7 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
8 General

System components 0

Components in the gantry rear side 0

1 7

2
8

4 9

10
5
6 11

Fig. 1: Gantry rear side


Pos. 1 Power Distribution Rotating
Pos. 2 Power Supply Rotating
Pos. 3 Detector Power Supply
Pos. 4 Data brushes & holder
Pos. 5 Oil expansion vessel
Pos. 6 Power brushes & holder
Pos. 7 Generator Electronic Box
Pos. 8 Power Box
Pos. 9 HV Tank
Pos. 10 Tube
Pos. 11 Rotation Motor

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 8 of 32 Siemens AG


12.02 CSCT Medical Solutions
General 9

Components in the gantry front side 0

1 4

3 6

Fig. 2: Gantry front side


Pos. 1 Cooling unit
Pos. 2 MAR & FOC Board
Pos. 3 Belt Tension
Pos. 4 Tube
Pos. 5 Collimator Box
Pos. 6 HV Tank
Pos. 7 DMS

Siemens AG CT02-019.840.01.01.02 Page 9 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
10 Generator (XRS)
3-
General
2Generator (XRS)

Safety 0

WARNING Risk of accident and injury!


Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.

¹ The guidelines contained in this document must be


strictly followed (Safety Information / p. 5).

Generator Overview 0

For detailed information (e.g. diagrams, test points, LED’s, fuses), please refer to the
actual function description of the corresponding system.

Fig. 3: Generator Overview

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 10 of 32 Siemens AG


12.02 CSCT Medical Solutions
Generator (XRS) 11

HV_Drop Explanation 0

The actual high tension (P_UTact and N_UTact ) is monitored by the board D400. If either
positive, negative or both are arcing (high tension breakdown greater than 30 kV) the
D400 will send the HV_drop signal. In this case, the main inverter will be stopped and
restarted within 2 ms to prevent a high tension error.
Example: If during a 120 kV exposure, the 60 kV negative voltage side would "drop", the
60 KV positive voltage side would try to compensate the voltage and try to raise to 120
kV. If the positive side were allowed to rise to 120 kV, a safety scan abort would be per-
formed when the HT reaches 82 kV.

NOTE The HV_drop is recorded in the tube history but the HV_drop alone
CANNOT be used to evaluate tube performance.
The HV_drop is transmitted to the MAR. The MAR will mark the
readings in which the HV_drop was active to allow the IRS to inter-
polate these readings.
Also the firmware counts the number of HV_drops within 300 ms.
If there are to many HV_drops within 300 ms, an error is generated.

The oscillogram shows the generation of the HV_drop signal. The key point is where
HV_drop is active for 1.25 msec, the high voltage drops and is restored, allowing the scan
to continue without errors.

UTact

HV_drop

Fig. 4: Channel 1 = UTact, Channel 2 = HV_drop

Siemens AG CT02-019.840.01.01.02 Page 11 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
12 Generator (XRS)

Important Signals 0

The following oscillograms of signals represent key test points for the XRS. They contain
a description of the test point along with suggestions for evaluating the waveforms. Where
applicable, the scan parameters are recorded.

UTact and ITact 0

Test points on Main Control Board D400:


UTact X210, ANA_GND X210, ITact 100 mA X230, GND X270

UTact

ITact

Fig. 5: UTact and ITact for a 130kV 150 mA scan

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 12 of 32 Siemens AG


12.02 CSCT Medical Solutions
Generator (XRS) 13

X-Ray timeout Test 0

General 0

The timeout-test is used to check the automatic supervision of the total radiation duration
in the customer modes TOPO, SEQUENCE and SPIRAL. A longer radiation duration than
given by the rated value of the loaded mode is attempted. The scan has to be canceled
after 110% of the rated value and the scanner has to switch to state ABORT.
For modes TOPO and SEQUENCE, all selectable scantimes can be checked. For mode
SPIRAL, a rotation time and three different lengths of spiral ("min", "medium" and "max")
can be selected.

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select Controller
• select XRS
• select Xray timeout
• select Topo, Sequence or Spiral
(for each mode, different parameter selections are possible)
• select Go
• press START
An abort is expected when the radiation duration supervision is active.
The test is finished if the message "XRS Xray timout test completed" appears

• a System Continue must be performed after the timeout test

Siemens AG CT02-019.840.01.01.02 Page 13 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
14 Generator (XRS)

Rotating Anode Test RAC01 0

General 0

RAC01 tests the rotating anode. The user can select a rotation frequency of 60Hz or
150Hz. When the test is finished, the measured values for the rotation anode are dis-
played.
Example:
IRAC last acceleration 26A, IRAC actual 6A
(retrieved from table T50)

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select Controller
• select XRS
• select RAC01
• select 150 Hz (or 60 Hz)
• select Go
• press START
The test is finished if the message "XRS RAC01 Test completed" appears

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 14 of 32 Siemens AG


12.02 CSCT Medical Solutions
Generator (XRS) 15

Filament Test FiL01 0

General 0

This test checks the function of the filament current regulator with standby and exposure
heating (7000 mA). No radiation is released during this test procedure. When the scan is
finished, the measured value for the filament current is displayed.
Example: Filament current of last FiL01 Test Scan: 6997mA (retrieved from table T5E)
Scope example:
Test points on Main Control Board D400:
I_Fact X110, ANA_GND X110, Trigger: Start_Inv X260, GND X270

I_Fact

Fig. 6: Filament test

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select Controller
• select XRS
• select Fil01
• select Go
• press START
The test is finished if the message "XRS FiL01 Test completed" appears

Siemens AG CT02-019.840.01.01.02 Page 15 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
16 Generator (XRS)

High Voltage Test HVC01 (without rotation) 0

General 0

The test HVC01 checks the high tension path without tube current and without rotation. To
isolate a high tension fault to either the tube or HV tank, the test plugs (16 21 791) must
be used. When the scan is finished, the measured values are displayed.
Example:
Results of last HVC01 Test Scan: UT_act:130.000KV, IT_act: 0 mA
(retrieved from table T5E)

NOTICE The HVC01 test should not be performed with the tube connected,
because this can lead to misdiagnosis of failures.
Before replacing the Tube, try gettering to reestablish stable oper-
ation!
¹

dummy
P_UTact
plugs
XRS
Generator Electronic HV-Tank tube
box dummy
N_UTact
plugs

Fig. 7: High Voltage Test

Prerequisite 0

The dummy plugs (16 21 791) must be connected to the HV-Tank.

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select Controller
• select XRS
• select HVC01
• select 130 kV (range: 120 kV - 130 kV available)
• select Go
• press START
(The scan takes approximately 80 s)
The test is finished if the message "XRS HVC01 Test completed" appears

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 16 of 32 Siemens AG


12.02 CSCT Medical Solutions
Generator (XRS) 17

High Voltage Test HVC02 (with rotation) 0

This test is similar to the HCV01 test with the exception of rotation. Therefore small
testplugs have to be used.

NOTE Be careful of the disconnected HV cables during rotation!

Siemens AG CT02-019.840.01.01.02 Page 17 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
18 Generator (XRS)

High Voltage Test HVC03 (Single Puls Test) 0

General 0

The Single Pulse test allows to separate a fault between the Inverter (Electronic Box) and
the HV_Tank, by triggering the inverters for a Single Pulse.
Please use an oscilloscope to measure the output of the single pulse test. Both the nega-
tive and the positive pulse should be evaluated.
Expected valuas:
Tab. 1 Single Puls Test

Balance/Emotion/Esprit+ Esprit
Single puls time 9 (±20%) 9 (±20%)
(microseconds)
Current peak value 150 (±20%) 130 (±20%)
(amperes)

Single pulse example: Balance / Emotion /Esprit+ 0

Test points on the Main Control Board D400:


I_load X230, ANA_GND X210, Trigger: Start_Inv X260, GND X270

Positive single puls:

I_load

Start_Inv

Fig. 8: Single pulse test


The single pulse width is 10.4 microseconds, the current peak is approximately 150 A.

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 18 of 32 Siemens AG


12.02 CSCT Medical Solutions
Generator (XRS) 19

Negative single puls:

I_load

Start_Inv

Fig. 9: Single pulse test


The single pulse width is 10.4 microseconds, the current peak is approximately 140 A.

Single pulse example: Esprit 0

Test points on the Main Control Board D400: I_load X230, ANA_GND X210, Trigger:
Start_Inv X260, GND X270

Positive single pulse:

I_load

Start_Inv

Fig. 10: Single pulse test


The single pulse width is 8.92 microseconds, the current peak is approximately 130 A.

Siemens AG CT02-019.840.01.01.02 Page 19 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
20 Generator (XRS)

Negative single pulse:

Fig. 11: Single pulse test


The single pulse width is 9.28 microseconds, the current peak is approximately 130 A.

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 20 of 32 Siemens AG


12.02 CSCT Medical Solutions
Data Measurement System (DMS) 21
4-
General
3Data Measurement System (DMS)

Safety 0

WARNING Risk of accident and injury!


Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.

¹ The guidelines contained in this document must be


strictly followed (Safety Information / p. 5).

Data Measurement System (DMS) Overview 0

For detailed information (e.g. diagrams, test points, LED’s, fuses), please refer to the
actual function description of the corresponding system.

Fig. 12: DMS for Balance/Emotion/Esprit+

Siemens AG CT02-019.840.01.01.02 Page 21 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
22 Data Measurement System (DMS)

Disc Test 0

General 0

The disc test checks the data transfer from the transmitter to the receiver and the image
processing. The "syntetic" disc raw data are stored in the transmitter and in the receiver.
For trouble shooting, the data source of the disc raw data can be selected

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select IRS/ICS
• select IRS Disc Test
• select Data Source:
Transmitter with rotation
or
Transmitter without rotation
or
Receiver
• select Go
• press START

Example: Disc-Image

Fig. 13: Disc-Image

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 22 of 32 Siemens AG


12.02 CSCT Medical Solutions
Data Measurement System (DMS) 23

Offset Image 0

General 0

With the offset image, the image relevance of artefacts due to electronic, electro-mechan-
ical and data transmission defects can be seen.

Procedure 0

• change to service mode Options / Local Service


• select TestTools
• select DMS
• select Offset Image
• select Rotation on
• select Scantime 1 (Esprit: select 1.5 s)
• select Go
• press START
Evaluation:
The calculated offset image must be free of rings or stripe artefacts.

Example: Offset-Image

Fig. 14: Offset-Image

Siemens AG CT02-019.840.01.01.02 Page 23 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
24 Data Measurement System (DMS)

Mean and Standard Deviation Measurement without Xray 0

Measurement without Xray 0

Loading the Static Mode:


• change to service mode Options / Local Service
• select TestTools
• select Stat-Modes
• select mode parameter:
- Time: 1.000 s
- Store Step: FPA decoding
- Test Pattern: Data without correction
(all other parameters are not relevant)
• select Go

Switch off XRAY and Anode Rotation:


• select Add. Window (located in the upper selection bar of the TestTools window)
• select Control
• select General
• select XRAY Off and Anode Rotation Off
• select Go (Do not close the Control window)

Start the Scan without Xray:


• press START

Analyze function 0

• select Add. Window (located in the upper selection bar) and


• select Analyze (it is recommended to enlarge the Analyze window)

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 24 of 32 Siemens AG


12.02 CSCT Medical Solutions
Data Measurement System (DMS) 25

Example:

Fig. 15: Analyze window with a piot of mean values

Preparing of the raw data file for Analyze 0

Load the raw data file:


• select Load / ICS Filesystem / Raw Data
• select Go
• select rotstatic.raw
• select Go
Message: "Primary file successfully loaded. Data type: REAL" appears

Region Selection:
• select Row/Reading: All
• select Column/Channel: All

Siemens AG CT02-019.840.01.01.02 Page 25 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
26 Data Measurement System (DMS)

• select detectorLine: 1
Only for Emotion Duo select:
either detectorLine: 1 or detectorLine: 2 (don’t select both)
Recommendation:
Perform first the calculation of the Mean and Standard Deviation of the detectorLine 1,
then repeat the complete procedure with detectorLine 2.
• select Go Message: "Region selection has been performed!" appears

Calculation of the Mean and Standard Deviation (Noise) values 0

Calculation of the Mean values:


• select Statistic: Mean
• select Projection
• select Go Message: "Click on ’GO’ to store the computed data on ICS file system!" ap-
pears
• select Go Message: "Input a filename or select a file from the list. Click on ’GO’ ........."
appears
• In the line Selected File, the file name of the mean value can be changed.
Recommendation:
Only for Emotion Duo, the default Selected File name: Mean.out should be
changed:
If detectorLine: 1 has been selected: Change the Selected File name to Mean1.out
If detectorLine: 2 has been selected: Change the Selected File name to Mean2.out
• select Go Message:
"The computed data has been stored on the ICS file system!" appears

Calculation of the Standard Deviation (Noise) values:


• select Statistic: StdDev
• select Projection
• select Go
Message: "Click on ’GO’ to store the computed data on ICS file system!" appears
• select Go
Message: "Input a filename or select a file from the list. Click on ’GO’ .........!" appears
• In the line Selected File, the file name of the mean value can be changed.
Recommendation: Only for Emotion Duo, the default Selected File name:
StdDev.out should be changed:
If detectorLine: 1 has been selected: Change the Selected File name to StdDev1.out
If detectorLine: 2 has been selected: Change the Selected File name to StdDev2.out
• select Go
Message: "The computed data has been stored on the ICS file system!" appears

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 26 of 32 Siemens AG


12.02 CSCT Medical Solutions
Data Measurement System (DMS) 27

(Emotion Duo: For the second detector line, repeat the procedure (Preparing of the raw
data file for Analyze / p. 25)

Evaluation of the Mean and the Standard Deviation (Noise) values 0

Evaluation of the Mean values:


• select Common: Load
• select ICS Filesystem / Analyze Data
• select Go
Message: "Select a file from the list and click on ’GO’ to load!" appears
• select a "Mean" file from the list
Example: Mean.out or Mean1.out or Mean2.out
• select Go
Message: "Primary file successfully loaded. Data type: TABLE_DATA." appears
• select Go
Message: "Region selection has been performed!" appears
• select View: Plot
(The plot of the mean values, the minimum value and the maximum appear)

Esprit Balance/Emotion/Esprit+ Emotion Duo


Tolerance of Mean 24000 - 54000 10000 - 60000 1000 - 80000
Values

NOTE To get the list of the mean values select: List / Dec / Go).

Evaluation of the Standard Deviation (Noise) values:


• select Common: Load
• select ICS Filesystem / Analyze Data
• select Go
Message: "Select a file from the list and click on ’GO’ to load!" appears
• select a "Standard Deviation" file from the list
Example: StdDev.out or StdDev1.out or StdDev2.out
• select Go
Message: "Primary file successfully loaded. Data type: TABLE_DATA." appears
• select Go
Message: "Region selection has been performed!" appears

Siemens AG CT02-019.840.01.01.02 Page 27 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
28 Data Measurement System (DMS)

• select View: Plot


(The plot of the standard deviation values, the minimum value and the maximum value
appear)

Balance / Emotion / Emotion Duo / Esprit / Esprit+


Tolerance of standard devi-
2 - 40
ation values

NOTE To get the list of the standard deviation values select: List / Dec /
Go).

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 28 of 32 Siemens AG


12.02 CSCT Medical Solutions
Slipring System (SRS) 29
5-
Slipring system
4Slipring System (SRS)

Safety 0

WARNING Risk of accident and injury!


Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.

¹ The guidelines contained in this document must be


strictly followed (Safety Information / p. 5).

Slipring cleaning procedure 0

Troubleshooting hint in case of HW - MUX Errors


If one of the following errors occurs sporadically:
• CT_MAR_16, CT_MAR_41, CT_MAR_42
• CT_MAS_26, CT_MAS_27, CT_MAS_46
• CT_PHS_42
a possible cause is oil on the slipring.

Cleaning the slip ring system


• Remove the carbon dust from the slip ring using a vacuum cleaner while rotating the
gantry step by step manually counter clockwise.
• Clean the inner two slip rings (control data slip rings) first with a dry paper towel and
then with an alcohol-soaked paper towel.

Cleaning the data slip rings with the slip ring cleaner, part no. 73 95 440

• Press the slip ring cleaner onto the sur-


face of the control data slip rings and ro-
tate the gantry manually counter
clockwise by 360 degree.

Fig. 16: Data slip ring

Siemens AG CT02-019.840.01.01.02 Page 29 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
30 Slipring System (SRS)

Checking the system for oil leaks


Possibilities: HV connector on the tube, tube housing or HV tank
If there is any oil, clean first with a dry paper towel and then with an alcohol-soaked paper
towel.

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 30 of 32 Siemens AG


12.02 CSCT Medical Solutions
Changes to previous version 31
6-
5Changes to previous version

Chapter Change Reason


Editorial changes:
All The FrameMaker document was converted to CTTU-02-0012
the SGML database (Astoria)

Siemens AG CT02-019.840.01.01.02 Page 31 of 32 Balance/Emotion/Esprit


Medical Solutions 12.02 CSCT
32 Changes to previous version

Balance/Emotion/Esprit CT02-019.840.01.01.02 Page 32 of 32 Siemens AG


12.02 CSCT Medical Solutions