O P E R AT O R S M A N U A L
Contents
1 The SAS-1 Plus
1.1 Introduction 3
1.2 Informational symbols 3
1.3 Precautions and Limitations 3
1.4 Safety Considerations 4
2 Installation
2.1 Instructions for installation 5
3 Controls
3.1 Mains Supply 6
3.2 Front Panel 6
3.3 Application Drawer 6
4 Operating Instructions
4.1 Power up 7
4.2 Program selection 7
4.3 Prepare for electrophoresis 8
4.4 Start an application 9
4.5 Create or Edit a program 11
4.6 Edit or Change a program name 12
4.7 View process parameters during a program run 13
4.8 Abort a program run 14
5 Maintenance
5.1 Heat sink 15
5.2 Carbon electrodes 15
5.3 Antisera template 15
5.4 Sample tray 15
5.5 Cleaning spills 15
5.6 Replace fuse(s) 15
6 Troubleshooting
6.1 Problem, Cause and Solution 16
7 Performance Specification
7.1 Specifications and parameters 17
Warranty Report 18
2 English
SAS-1 Plus Operators Manual
This manual provides all necessary information for the setup and operation of the SAS-1 Plus
including important information regarding safety and maintenance of the unit. To ensure safe
and reliable operation, always operate the SAS-1 Plus in accordance with the operator
manual.
Helena BioSciences Europe is not responsible for injuries or damages caused by improper
use.
Warning
The operators manual should be consulted to avoid possible personal injury
or instrument damage.
To isolate the equipment, remove the plug from the socket. The mains switch is considered
to be the primary disconnect device and should be easily accessible by the operator.
> The entire operators manual should be read and understood before attempting
instrument operation.
> This instrument is designed for use with Helena BioSciences Europe products only.
> Always check the parameters set in the SAS-1 Plus program before starting.
The parameters may be different than those prescribed in the Instructions For Use.
The parameters can be accessed via the edit-function.
> Ensure the fan at the rear of the unit is not covered and has adequate air circulation.
> Ensure that the electrodes and the sample tray are removed from the instrument prior
to any transportation.
3 English
SAS-1 Plus Operators Manual
> This piece of equipment must not be used in areas of high electromagnetic
phenomenon as temporary loss of function may occur. Contact Helena Biosciences
Europe for advice.
The SAS-1 Plus is fully overload protected. However, the following guidelines should be
observed to ensure safe operation of the unit.
> The SAS-1 Plus must only be operated when connected to an earthed supply of the
correct voltage and frequency as indicated on the serial number label.
> For emergency shutdown, use either the ON/OFF (switch on the rear panel of the
device), or switch off at the mains supply.
> Wait at least 5 seconds after stopping a run before handling output leads or connected
apparatus.
> Always make sure your work area is clean and dry before operating the SAS-1 Plus.
> If the equipment is used in a manner not specified by the manufacturer the protection
provided by the equipment may be impaired.
4 English
SAS-1 Plus Operators Manual
2 Installation
Please verify upon receiving the SAS-1 Plus that no damage has occurred during transit and
confirm that the following components and accessories are included:
> Place the SAS-1 Plus unit on a level bench. Keep the area around the instrument clear to
ensure correct ventilation.
> Select an environment free from excessive humidity, dust and large temperature
fluctuations. Allow the instrument to acclimatise to room temperature for a period of
at least 2 hours before operating.
> The power cord attachment module is located at the rear of the unit. Position the unit
so this is easily accessible.
> Remove the transport pin and plug from the drawer, located on the bottom side of the
instrument. Store in a safe place for possible re-use should the device require further
transportation.
> Connect the unit to a suitable earthed mains outlet of the correct voltage and
frequency. This unit MUST be earthed.
> Turn on the main power switch located on the left side of the instrument.
5 English
SAS-1 Plus Operators Manual
3 Controls
3.1 Mains Supply
ON/OFF Switch This is a rocker-type switch, which controls the mains input to the unit
and is situated on the rear of the instrument.
Location mount The location mount is for the positioning of the sample tray.
Electrode Posts The electrode posts carry the voltage from the power supply to the
electrodes. The posts are also magnetic to hold the electrodes in place.
The positive electrode posts are located at the front of the drawer,
the negative at the rear.
Heat sink The heat sink carries heat away from the gel and is also used to position
the gel. The heat sink is adapted with a peltier device that controls the
temperature between 15°C and 45°C.
6 English
SAS-1 Plus Operators Manual
4 Operating Instructions
4.1 Power up
Connect the unit to an appropriate mains outlet and switch on by using the power switch
located on the rear of the instrument. The display will prompt:
Use the ‘+’ and the ‘-’ keys to scroll between programs.
SAS program
Press + to Start
and
SAS program
Press - to Cancel
and
7 English
SAS-1 Plus Operators Manual
SAS program
Press ENTER to edit
> Press ‘-’ to cancel and return to the program selection prompt.
> Press ENTER to edit or view the parameters of the selected program.
> Press ‘+’ to start the application.
NOTE: Always check the parameters in the SAS-1 Plus program before starting.
The parameters may differ to those prescribed in the Instructions For Use, supplied
with the electrophoresis kit.
The SAS-1 Plus must be prepared for the required application before the program
commences. Refer to the Instructions For Use supplied with each kit for full operator
instructions.
NOTE: Always put the Sample in the Sample row as displayed in figure 1 below. The Dilution
cups are for manual use only, the SAS-1plus is not able to reach the Dilution rows
Fig.1
8 English
SAS-1 Plus Operators Manual
Before starting the electrophoresis, verify the following. The sample gels are orientated
correctly and the sample cups are installed correctly and fully seated in the sample tray.
The applicator blades are properly installed and seated in the slots.
After selecting a program, as described in section 4.2, press ‘+’ to start. The drawer opens
and the display will prompt:
SAS program
Press key to close
NOTE: Refer to the Instructions For Use for specific gel preparation.
Place the sample tray into the instrument, apply REP Prep, place the gel into position and blot,
then place the electrodes into the correct position. Place the gel cover over the gel and
electrodes, press for 5 seconds to ensure good electrode contact.
Applicator Blades
Gel Cover
Electrodes
Gel
9 English
SAS-1 Plus Operators Manual
Electr. completed
Press key to open
When incubation is programmed, the SAS-1 Plus will sound and alternate between the
displays shown below:
Incubation step 1
Press + to open
and
Incubation step 1
Press - to continue
Follow the Instructions For Use for reagent preparation. When ready Press ANY KEY to
commence incubation.
Process completed
Press key to open
10 English
SAS-1 Plus Operators Manual
SAS program
Press ENTER to edit
At this cursor position it is possible to use the +/- keys to scroll through the following
program parameters:
* if this parameter is set to OFF, no incubation programming or use of the Add Reagent
function is possible.
** up to a maximum of 9 steps.
Press the Enter key to edit a parameter The cursor will automatically go to the setting, which
can be altered with the +/- keys, within the limits that are specified above. When the
parameter is set, press enter to leave parameter editing and return to scrolling between the
parameters.
11 English
SAS-1 Plus Operators Manual
To edit the program name, use the - key until you reach the following prompt:
When pressing ENTER the cursor will go to the first digit of the name, which can be changed
with the +/- keys. When the digit is correct press ENTER and the cursor will go to the next
digit. All digits need to be confirmed before the cursor returns to its start position.
Use the STOP key to leave the edit-page. The prompt will read and alternate between:
and:
After pressing the + or - key the prompt will return to start position and the operater can
choose to start, cancel or edit the program.
12 English
SAS-1 Plus Operators Manual
During a program run it is possible to view the process parameters on the display.
The standard display states the time left.
Using the +/- keys it is possible to scroll through different values and show the following
screens.
The second line of the display will alternate between the actual program step
(electrophoresis or incubation) and “press STOP to abort”.
13 English
SAS-1 Plus Operators Manual
To abort a program run at any stage, the STOP key should be pressed, NOTE: it is not
possible to pause any stages of the program.
The drawer will open and the gel, electrodes and sample tray can be removed. The display
will prompt:
Electro. aborted
Press key to close
SAS program
Press + to start
14 English
SAS-1 Plus Operators Manual
5 Maintenance
This section describes routine operators maintenance procedures.
For instrument calibration or for maintenance not described in this manual, call Helena
BioSciences Europe for assistance.
15 English
SAS-1 Plus Operators Manual
6 Troubleshooting
NOTE: If problem persists or any other messages are displayed on the screen, contact
Helena BioSciences Europe for assistance.
16 English
SAS-1 Plus Operators Manual
7 Performance Specifications
Input Power: 230 Vac (Cat. No. 1531)
115 Vac (Cat. No. 1532)
Frequency 50/60 Hz, 225 watts
Voltage is factory set (see Serial No. Label).
Keypad: 4 keys
Operating Modes: Standby, Program selection, Program editing and Program run.
Safety features: Safety Interlock breaks electrophoresis circuit when applicator drawer
door is open.
17 English
SAS-1 Plus Operators Manual
Warranty
Please complete and return to Helena BioSciences Europe, by email, fax or post.
Instrument Name:
Date Received: Catalogue Number:
Serial Number:
Name: Department:
Institution:
Address:
Post Code:
Telephone Number:
Ext. (where applicable):
Email:
Helena warrants each unit to be free from defects in material and workmanship. Helena’s liability under this warranty
is limited to servicing or adjusting any unit and to replacing any defective part. This warranty is effective for 1 year
commencing from the date of shipment to the original purchaser, and does not cover faults caused by misuse, abnormal
conditions, damage incurred in shipment, or consumable items such as lamps. This warranty is made in lieu of all other
warranties, expressed or implied, and is limited in any sale to the repair or replacement of the affected component.
18 English
SAS-1 Plus Operators Manual
Yes [ ] No [ ]
Yes [ ] No [ ]
Comments:
Helena BioSciences Europe is committed to providing you with the finest equipment and supplies possible.
If you need assistance please call our Sales / Service Department:
tel: +44 (0) 191 549 6064
fax: +44 (0) 191 549 6271
email: info@helena-biosciences.com
www.helena-biosiences.com
19 English