Sei sulla pagina 1di 2

Issue 5

Setting Tolerances
for pipettes in the laboratory
“Fast is fine,

T
Users frequently find it difficult tips to be used when testing a
he performance to reproduce a manufacturer’s particular pipette. If the user
performance claims, for a num- chooses another type of tip
but accuracy of mechanical ber of reasons: (e.g., a filter or gel loading tip),
action pipettes or a lower quality tip, the
• There are no consistent stan- pipette can easily test outside
is everything” must be tested periodically to dards for how manufacturers of the manufacturer’s toler-
ensure accurate and precise liq- set their performance claims. ances.
uid delivery. The results of such These claims are often a trade-
testing then may be compared off between engineering judg- • Statistical factors such as the
Wyatt Earp
with pre-established tolerances, ment and marketing necessities. number of data points taken
and out-of-tolerance conditions impact the probability of inter-
• Pipette performance is influ- mittent or false failures. For
corrected. Tolerances that are too enced significantly by environ-
strict can cause a large number of example, four data points are
mental factors such as temper- sufficient for a “quick check”
so-called “false failures,” where a ature and humidity.1 This
pipette in good working order on accuracy and proper opera-
means that a pipette calibrated tion5, but ten data points are
produces test results that are out
at an environmentally con- recommended to ensure both
of tolerance. Tolerances that are
trolled facility may deliver accuracy and precision.
too broad can degrade the quality
incorrectly on the benchtop.2
of the laboratory’s analytical work.
Artel therefore recommends
Further complicating the situa-
testing pipette performance
Developing Achievable
tion is the fact that pipette per-
formance is dependent upon fac- under working conditions.3 Tolerance Limits
Based on our experience with
tors beyond the mechanical con- • The skill of the pipette opera- many different makes and mod-
dition of the pipette. Balancing tor plays a very important role els of pipettes, Artel recommends
this interplay of factors can make in the precision and accuracy the values in Table 1 as a starting
it a challenge to develop achiev- of the pipette.4 The choice of point for developing achievable
able tolerance limits for pipettes. pipetting technique (e.g., tolerance limits. These limits are
This Lab Report reviews the key reverse mode versus forward based on a simple guideline: Two
issues that should be considered mode) is also a source of vari- percent of full scale at all volume
when establishing tolerance lim- ability in pipetting results. settings.6 For example, the inac-
its for the working laboratory, Proper training can help curacy tolerance for a 100µL vari-
and recommends as a starting reduce false failures by ensur- able-volume pipette is 2.0µL (2%)
point a set of achievable tolerance ing that results are valid and at the 100µL setting, and 2.0µL
limits for pipettes of various sizes. can be reproduced across oper- (4%) at the 50µL setting.
ators. Artel’s guide, 10 Tips to This type of generalized toler-
Beware Manufacturers’ Improve Your Pipetting ance limit has been employed suc-
Specifications Technique, is available on cessfully in a number of other
Many laboratories attempt to request to help you establish fields, such as humidity measure-
apply pipette manufacturers’ good pipetting practices in ment and syringe calibration, where
specifications as the tolerance your laboratory. a fixed percentage of full scale
limits for their in-lab testing pro- reading is the customary means for
gram. However, a manufacturer’s • The type of pipette tip used in specifying performance. The ISO
specifications may not accurately the testing also affects results.1, 5 8655-2 standard for pipette con-
reflect the performance attained Most manufacturers and rep- formity testing 1 also uses a per-
in the working laboratory. utable calibration services centage of full scale approach.
19A3230C precisely define the type of
Table 1: Artel’s suggested initial tolerance limits
Using Table 1
Begin by choosing either the “Relative Error” or “Absolute Error” tolerance limit Pipette Volume, µL Relative Error Absolute Error
values. These tolerance limits reflect what is reasonably achievable in a working
Nominal Setting Inaccuracy CV Inaccuracy STD
laboratory. They presume that the pipette is calibrated and functioning properly,
±% % ± µL µL
is used with good quality tips, and is tested by a skilled operator. When these cri-
2.0 2.0 1.0
teria are met, most makes and models of pipettes should test within these toler- 2 1.0 4.0 2.0 0.04 0.02
ance limits unless they are mechanically defective. 0.2 20.0 10.0
2.5 2.0 1.0
For fixed-volume pipettes, the nominal value is the fixed volume. For variable-vol- 2.5 1.0 5.0 2.5 0.05 0.025
ume pipettes, the nominal value is the largest user-selectable volume setting; e.g., 0.2 25.0 12.5
a 10-100µL pipette has a nominal volume of 100µL. 10 2.0 1.0
10 5 4.0 2.0 0.20 0.10
The absolute error for the nominal volume applies to every selectable pipette vol- 1 20.0 10.0
20 2.0 1.0
ume; e.g., a 100µL nominal volume yields limits of ±2.0µL inaccuracy (mean 20 10 4.0 2.0 0.4 0.2
value) and ±1.0 imprecision (STD) for all volumes. The relative error varies 2 20.0 10.0
throughout the pipette range; e.g., for a 10-100µL pipette at 100µL the relative 50 2.0 1.0
inaccuracy is ±2.0%. However, at 10µL the relative inaccuracy is ±20.0%. 50 25 4.0 2.0 1.0 0.5
5 20.0 10.0
Notes 100 2.0 1.0
a) Inaccuracy is expressed as the deviation of the mean of ten samples from 100 50 4.0 2.0 2.0 1.0
the set point volume. Inaccuracy can be expressed in either absolute units 10 20.0 10.0
200 2.0 1.0
such as microliters, or relative units such as percent. Absolute imprecision 200 100 4.0 2.0 4.0 2.0
is expressed as the standard deviation (STD) of ten samples. Relative impre- 20 20.0 10.0
cision is expressed as the coefficient of variation (CV) of ten samples. 500 2.0 1.0
500 250 4.0 2.0 10.0 5.0
b) For pipettes with nominal volumes between those provided in this table, 50 20.0 10.0
absolute inaccuracy tolerance limits are equal to ±2.0% of the pipette’s nom- 1000 2.0 1.0
inal volume and the tolerance limit for absolute imprecision is 1% of the 1000 500 4.0 2.0 20.0 10.0
pipette’s nominal volume. 100 20.0 10.0
2000 2.0 1.0
2000 1000 4.0 2.0 40.0 20.0
c) Relative error tolerance limits at other volume settings can be calculated by 200 20.0 10.0
dividing the absolute tolerance limit (see Table 1 or Note b) by the set point 2500 2.0 1.0
volume. Multiply the result by 100 to convert it to a percentage. 2500 1000 5.0 2.5 50.0 25.0
500 10.0 5.0
5000 2.0 1.0
5000 2500 4.0 2.0 100.0 50.0
The allowable percentage that ISO Example 1: An analytical method 500 20.0 10.0
specifies is fairly constant for medi- requires dispensing a 100µL sample
um- to large-volume pipettes, but with 3% accuracy. The laboratory has
increases for the smallest volume been using a 200µL pipette set to References:
pipettes. This deviation is necessary 100µL for this purpose. Table 1 shows 1 Piston-Operated Volumetric Apparatus—Part 2: Piston Pipettes. ISO
when using gravimetric methods in the recommended tolerance to be 4%, 8655-2:2002.
order to accommodate the increased which is too liberal for the method. The
error inherent in gravimetry at small- 2. Bias, Uncertainty and Transferability in Standard Methods of Pipette
simplest and most reliable solution is to Calibration. Artel white paper, 2003.
er volumes.7 If a more exact method replace the 200µL pipette with a 100µL
such as ratiometric photometry is pipette. This pipette, when used at its 3. Rodrigues, George. “Put That in Your Pipette…and Calibrate It,”
used, then a constant percentage of full scale setting, can be tested against a Pharmaceutical Formulation and Quality, February/March, 2003: 64-65.
full scale may be applied to even the 2% tolerance. 4. Pentheny, G. “Effects of common pipette technique errors on accuracy
smallest volume pipettes. and precision,” American Clinical Laboratory (1997), 16(2): 10-12.
Example 2: An analytical method 5. Standard Specification for Piston or Plunger Operated Volumetric
Fine-Tuning Tolerance Limits requires 1% accuracy at a volume of Apparatus, Specification E 1154-89 (Re-approved 2003), American
The tolerance limits recommend- 1,000µL. This is a greater accuracy Society for Testing Materials (ASTM), p. 5, ¶ 11.2.2 and 11.2.3.
ed here are based on what is typical- than given for any pipette in Table 1. 6. Duff, L. David. Metrology A-Z, Inc., personal communication, 2003.
ly achievable when testing pipettes Pipette performance data are examined
in the absence of rigorous laboratory to determine whether this degree of 7. Clark. John P. and Shull, A. Harper. “Gravimetric & Spectrophoto-
environmental controls or special- metric Errors Impact on Pipette Calibration Certainty.” Cal. Lab., 10
accuracy can be attained. It is found (1): pp. 36-43, 2003.
ized training in pipetting technique. that two particular operators are regu-
These recommendations do not take larly attaining the desired level of per-
into account the more stringent data formance when using a particular make
quality requirements of a particularly and model of pipette, while other oper- lab report is published by Artel, Inc. for laboratory professionals
demanding analytical method. In ators are not. The superior pipette is and all pipette users.
such circumstances, laboratories specified in the procedure, and the For more information, please contact Artel, Inc.
should evaluate the results of past highly skilled operators are used as
testing to fine-tune the initial toler- benchmarks against which others may Tel: 207-854-0860 Email: post@artel-usa.com
ance limits relative to the require- be trained. The tolerance limit for this Fax: 207-854-0867 Web: www.artel-usa.com
ments of the method. The following pipette can then be tightened to 1% 25 Bradley Drive, Westbrook, Maine 04092-2013
examples illustrate solutions for without causing a large number of false
common problems encountered failures. © 2004 Artel, Inc. All Rights Reserved.
when establishing pipette tolerances.

Potrebbero piacerti anche