Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Introduction
The atom cloud required for atomic absorption measurements is produced by supplying
enough thermal energy to the sample to dissociate the chemical compounds into free atoms.
Aspirating a solution of the sample into a flame aligned in the light beam serves this
purpose. Under the proper flame conditions, most of the atoms will remain in the ground
state form and are capable of absorbing light at the analytical wavelength from a source
lamp. The ease and speed at which precise and accurate determinations can be made with
this technique have made atomic absorption one of the most popular methods for the
determination of metals.
Instrument
Materials
2) Unknown solution
Every absorption spectrometer must have components which fulfill the three basic
requirements shown in Figure 1. There must be:
1. A light source
2. A sample cell
3. A means of specific light measurement.
1. The light source that emits the spectrum of the element of interest
2. An "absorption cell" in which atoms of the sample are produced (flame, graphite
furnace, MHS cell, FIAS cell, FIMS cell)
4. A detector, which measures the light intensity and amplifies the signal
5. A display that shows the reading after it has been processed by the instrument
electronics
Single-Beam Photometers
Double-Beam Photometers
In the double-beam system as in Figure 6, additional optics is used to divide the light
from the lamp into a ‘‘sample beam’’ which directed through the sample cell and a
‘‘reference beam’’ which directed around the sample cell. The reference beam
serves as a monitor of lamp intensity and the response characteristics of common
electronic circuitry.
Figure 4 A double-beam AA spectrometer.
LIGHT SOURCES
An atom absorbs light at discrete wavelengths. In order to measure this narrow light
absorption with maximum sensitivity, it is necessary to use a line source, which emits
the specific wavelengths which can be absorbed by the atom. Narrow line sources
not only provide high sensitivity, but also make atomic absorption a very specific
analytical technique with little spectral interference. The two most common line
sources used in atomic absorption are the hollow cathode lamp and the
electrodeless discharge lamp.
In AAS, burner system functions as (i) sample holder, (ii) desolvation source, and (iii)
volatilization source. The dual option burner system can be operated either with or
without a flow spoiler for optimum operation under different analytical conditions.
However, for routine operation it is recommended that the burner system be
operated with a flow spoiler. The flow spoiler helps to remove large droplets from the
nebulizer aerosol and thus minimizes chemical interferences. A diagram of the dual
option burner system is shown in Figure 7.
Detector
Photomultiplier tubes
The dilution calculations of each working solution
8 ppm: 10 ppm:
MssVss = M2V2 MssVss = M2V2
(100)Vss = (8)(100) (100)Vss = (10)(100)
Vss = 8 ml Vss = 10 ml
𝑌2 − 𝑌1
m=
𝑋2 − 𝑋1
0.82−0.47
m=
4−1
m = 0.1167
Concentration of unknown, X:
b x Cstd
Cx =
𝑚 𝑥 𝑉𝑥
0.35 x 100
Cx =
0.1167 𝑥 5
Cx = 59.98ppm
Next is the standard addition method. It is a common method that had been used by
adding one or more increment of standard solution to sample aliquots of the same
size. This is also known as spiking the sample. A series of standard with 1, 2, 3, 4
and 5ppm were added to the 15mL of unknown in separate volumetric flask and was
diluted. The standard addition graph was plotted and the concentration of calcium in
the unknown sample, X, is determined which is 19.84ppm. Thus, the amount of
calcium in the unknown is 1.984mg
During the experiment was conducted, there could be errors that had occur. To avoid
errors, several precautions should be taken by students. To make sure that the
concentration of the standards are accurate during preparation of the solution, we
have to make sure that our eye is parallel with the volumetric flask mark and to use a
dropper when the solution almost reach the mark. This is so that we do not exceed
the mark. Besides that, when using the instrument for analysis, we must set the auto
sampler with correct location of vials so that there will be no incorrect sampling.
Conclusion:
Based on this experiment, we can conclude that the concentration of calcium in the
tap water and in the unknown sample is 11.90ppm and 19.84ppm respectively. The
concentration of calcium in tap water is determined by using the calibration curve
whereas the concentration of calcium in the unknown sample is determined by using
the standard addition method. Meanwhile, the amount of calcium in the tap water
and in the unknown sample is 1.18mg and 1.984mg respectively.
Reference
(n.d.). Retrieved November 11, 2017, from Chapter 9: Atomic Absorption Spectrometry:
http://www2.chemistry.msu.edu/courses/cem434/chap9atomabsspec.pdf
(1996). Analytical Methods for Atomic Absorption Spectroscopy. United States of America: The
Perkin-Elmer Corporation.
Beaty, R. D., & Kerber, J. D. (1993). Concepts, Instrumentation and Techniques in Atomic Absorption
Spectrophotometry. Norwalk, CT, U.S.A: The Perkin-Elmer Corporation.
Levinson, R. (n.d.). Royal Society of Chemistry. Burlington House, Piccadilly, London : The Education
Department, The Royal Society of Chemistry.
The calibration curve of tap water and tap water:distilled water
The calibration curve of standard solution
The calibration curve of the unknown solution