Sei sulla pagina 1di 8

Clin. Lab.

2016;62:49-56
©Copyright
ORIGINAL ARTICLE

Urinalysis: The Automated Versus Manual Techniques;


Is It Time To Change?
Asmaa Ismail Ahmed, Heba Baz, Sarah Lotfy
Clinical and Chemical Pathology Department, Faculty of Medicine, Cairo University, Kasr Al Aini Street, Cairo, Egypt

SUMMARY

Background: Urinalysis is the third major test in clinical laboratory. Manual technique imprecision urges the need
for a rapid reliable automated test. We evaluated the H800-FUS100 automatic urine sediment analyzer and com-
pared it to the manual urinalysis technique to determine if it may be a competitive substitute in laboratories of
central hospitals.
Methods: 1000 urine samples were examined by the two methods in parallel. Agreement, precision, carry-over,
drift, sensitivity, specificity, and practicability criteria were tested.
Results: Agreement ranged from excellent to good for all urine semi-quantitative components (K > 0.4, p = 0.000),
except for granular casts (K = 0.317, p = 0.000). Specific gravity results correlated well between the two methods
(r = 0.884, p = 0.000). RBCS and WBCs showed moderate correlation (r = 0.42, p = 0.000) and (r = 0.44,
p = 0.000), respectively. The auto-analyzer's within-run precision was ≥ 75% for all semi-quantitative components
except for proteins (50% precision). This finding in addition to the granular casts poor agreement indicate the
necessity of operator interference at the critical cutoff values. As regards quantitative contents, RBCs showed a
mean of 69.8 ± 3.95, C.V. = 5.7, WBCs showed a mean of 38.9 ± 1.9, C.V. = 4.9). Specific gravity, pH, mi-
croalbumin, and creatinine also showed good precision results with C.Vs of 0.000, 2.6, 9.1, and 0.00 respectively.
In the between run precision, positive control showed good precision (C.V. = 2.9), while negative control's C.V.
was strikingly high (C.V. = 127). Carryover and drift studies were satisfactory. Manual examination of inter-
observer results showed major discrepancies (< 60% similar readings), while intra-observer's results correlated
well with each other (r = 0.99, p = 0.000).
Conclusions: Automation of urinalysis decreases observer-associated variation and offers prompt competitive re-
sults when standardized for screening away from the borderline cutoffs.
(Clin. Lab. 2016;62:49-56. DOI: 10.7754/Clin.Lab.2015.150520)

Correspondence: KEY WORDS


Asmaa Ismail Ahmed
Clinical and Chemical Pathology Department urine, automation, H800-FUS100, evaluation
Faculty of Medicine
Cairo University
Kasr Al Aini Street
Cairo, Egypt INTRODUCTION
Phone: +20 1093115275
Email: asmaa.foraida@yahoo.com Manual urine testing remains the most widely used
method for assessment of urine sediment. However,
urine testing protocols are not in consensus [1]. In the
manual urinalysis technique, microscopic examination
is used to check for the presence of formed elements.
Although the centrifugation step to concentrate the sam-
_____________________________________________ ple is necessary, it remains a major source of errors, as
Manuscript accepted June 16, 2015 it leads to a variable loss of erythrocytes and leukocytes.

Clin. Lab. 1+2/2016 49


Asmaa Ismail Ahmed et al.

Routine sediment analysis can have intra-assay coeffi- Microscopic examination


cient of variation as high as 100% when centrifugation 12 mL of the sample were centrifuged at 2000 rpm for
and residual sediment volume are taken into consider- 15 minutes. The supernatant was discarded, and the sed-
ation. On the other hand, neglecting centrifugation low- iment was re-suspended in 1 mL of residual urine vol-
ers the analytical sensitivity of the test. Strict standardi- ume. The sediment was examined microscopically at
zation of all procedures for collection, transport, sample 10X and 40X magnification power for the presence of
preparation, and analysis are mandatory as the basis of cells, crystals, casts, and others. Overall, the quantita-
an effective diagnostic strategy for urine analysis [2]. tive variables like RBCs, WBCs, and specific gravity
Different automated analyzers are currently used to per- (SG) were expressed numerically. Semi-quantitative
form the task of automated urine analysis [3]. These an- values like pH, glucose, albumin, ketones, bilirubin,
alyzers adopt two methodologies for urine sediment urobilinogen, epithelial cells, crystals, casts, sperm, and
analysis, one based on electrical impedance, and the bacteria were expressed in pluses.
other depends on image-based analysis systems that sort Furthermore, inter-observer precision (20 different ex-
particles according to predetermined particle dimen- aminers examined the same sample microscopically at
sions [3,4]. the same time) and intra-observer precision (a single ex-
The automated urine analyzer H800-Fus100 comprises aminer examined 12 samples microscopically before
two subunits. The H800 unit represents the automated and after rearrangement) were conducted for the manual
strip analyzer which aspirates and drops urine onto each technique.
reaction. The second unit is the microscopic examina-
tion unit FUS-100. It is based on the automatic particle Automated urine analysis
recognition system, which utilizes un-centrifuged urine Analysis was performed on the second aliquot using the
to capture and analyze 820 photos from each sample FUS-100/H-800.
(digital imaging). Artificial intelligence identification
technology is then used in order to auto-classify the Chemical examination
isolated images of the urine sediment constituents [5]. A DIRUI H14-Ca (H-800) (Dirui Industrial Co. Ltd.,
The aim of this study was to perform a comparison of China) recorded chemical results for the same parame-
methods between the automatic urine sediment analyzer ters of the manual method.
-FUS-100/H-800 and the manual technique for the ex- Microscopic examination relied on the automated plat-
amination of urinary formed elements and to judge its form and its integrated software as well as the examin-
eligibility to widely replace the manual standard meth- er's editing upon identification of unclassified and im-
od. properly classified particles as recommended by the
manufacturer.
Also between-run precision (positive and negative con-
MATERIALS AND METHODS trols), within run precision (pooled urine divided into 20
aliquots), carry over (three normal samples assayed be-
Subjects fore and after 3 high samples), and drift (difference be-
A cohort of 1000 urine samples from patients attending tween results of the same sample when introduced at
Kasr Al Aini hospitals during the period from July 2013 early, middle, and end of the run) were performed for
to March 2014 were enrolled in the study. A volume of the automated technique.
at least 15 mL urine in a clean cup, reaching the lab Other points of judgment were practicability criteria
within two hours of voiding was considered acceptable. such as economic considerations (cost per test, running
Urine samples which failed to fulfill the above criteria cost, and overhead cost) and operational considerations
were excluded from the study (turnaround time, friendliness of the software, degree of
The study was approved by the ethical committee of the training required to operate the procedure).
faculty of medicine, Cairo University and in accordance
with the code of ethics of the world medical association Statistical methods
(Declaration of Helsinki). Data were statistically described in terms of mean
 standard deviation ( SD), CV, and range, or frequen-
Methods cies and percentages when appropriate. Comparison of
All collected samples were examined within one hour of numerical variables between manual and automated
arrival, mixed properly and divided into two aliquots. techniques was done using paired t-test for matched
samples. For comparing categorical data, the McNemar
Manual method test was performed. Agreement was tested using kappa
Chemical analysis was done using Combur 10 Test UX statistics, and values were interpreted as follows:
urine strips (Roche Diagnostics, United Kingdom) for < 0.4 = poor agreement, 0.4 - 0.7 = good agreement,
the assessment of the following: pH, specific gravity and > 0.7 = excellent agreement. P-values less than 0.05
(SG), albumin, bilirubin, urobilinogen, glucose, and ac- were considered statistically significant. Statistical cal-
etone. Manufacturer’s protocol was rigorously follow- culations were done using the computer program SPSS
ed. (Statistical Package for the Social Science; SPSS Inc.,

50 Clin. Lab. 1+2/2016


Urinalysis, Automation Versus Manual

Table 1. Showing results of the manual and automated techniques for the tested parameters, with the corresponding method of
comparison results.

Manual Automated r, K *, †
Specific gravity 1019.8 (± 5.7) 1019 (± 7.2) r: 0.884,
[mean (± SD)] C.V. = 0.6 C.V. = 0.7 p = 0.00
pH Acidic Neutral Alkaline Acidic Neutral Alkaline
K: 0.914
(%) 89.9 2.0 8.1 89.9 3.6 6.5
Albumin Nil + ++ +++ ++++ Nil + ++ +++ ++++
K: 0.695
(%) 62.8 7.2 11.1 9.9 9.0 62.8 7.6 9.7 8.9 11.0
Glucose Nil + ++ +++ ++++ Nil + ++ +++ ++++
K: 0.78
(%) 88.0 3.5 4.1 4.2 0.2 87.8 3.2 2.4 6.2 0.4
Ketone Negative Positive Negative Positive
K:1
(%) 98.6 1.4 98.6 1.4
Nitrite Negative Positive Negative Positive
K:1
(%) 98.2 1.8 98.2 1.8
RBCs [median r: 0.417
1.5 (0.5 - 5) 4 (1.75 - 16)
(25th - 75th percentiles)] p = 0.00
WBCs [median r: 0.442,
1.5 (0.5 - 9) 2 (0.75 - 12)
(25th - 75th percentiles)] p = 0.00
Epithelial cells Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K: 0.765
(%) 59.0 12.0 17.4 7.9 2.2 1.5 58.0 15.3 14.8 7.6 2.8 1.5
Calcium Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K: 0.728
Oxalate 90.1 2.2 3.7 0.9 1.3 1.8 90.3 2.7 3.4 0.8 1.1 1.7
Crystals Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
Uric acid K: 0.65
(%) 93.5 1.6 2.2 1.2 0.8 0.7 94.4 1.9 1.7 0.8 0.5 0.7
Triple Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K **
phosphate 99.6 0.2 0.0 0.1 0.0 0.1 99.7 0.1 0.1 0.0 0.0 0.1
Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
Urate K: 0.71
Amorphous 97.8 0.1 1.0 0.5 0.2 0.4 98.0 0.2 0.6 0.4 0.1 0.7
(%) Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
Phosphate K **
99.4 0.0 0.0 0.2 0.3 0.1 99.3 0.1 0.0 0.2 0.2 0.2
Nil Few + ++ Nil Few + ++
Hyaline K **
97.9 0.9 1.2 0.0 97.2 2.3 0.1 0.0
Casts Nil Few + Nil Few +
Granular K: 0.317
(%) 97.4 1.7 0.9 99.2 1.7 0.1
Nil Few Nil Few
Waxy K **
99.9 0.1 100.0 0.0
Bacteria Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K: 0.74
(%) 72.9 4.9 7.3 4.1 3.2 7.6 72.1 7.6 5.9 3.9 3.3 7.2
Fungi Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K: 0.57
(%) 92.1 2.6 2.4 0.9 1.0 1.0 88.2 6.5 2.4 0.9 1.3 0.7
Mucous Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K: 0.559
(%) 83.5 0.4 8.4 4.6 2.2 0.9 76.1 7.7 9.9 3.4 2.1 0.8
Sperm Nil Few + ++ +++ ++++ Nil Few + ++ +++ ++++
K **
(%) 99.3 0.4 0.2 0.0 0.0 0.1 99.1 0.5 0.1 0.2 0.0 0.1

* Agreement was calculated for qualitative variables expressed in K, correlation was calculated for quantitative variables expressed in r.
** Kappa statistics could not be performed as it requires a symmetric 2-way table in which values of the first valuable must match the values
of the second variable. † The p-value of the kappa test was < 0.05 for all shown parameters denoting sound statistical calculations.

Clin. Lab. 1+2/2016 51


Asmaa Ismail Ahmed et al.

Table 2. Performance characteristics of the H-800 versus the manual technique in diagnosis of hematuraia and pyuria.

Performance characteristics Hematuria: cutoff ≤ 5/HPF Pyuria: cutoff ≥ 10/HPF


Sensitivity% (95th CI) 98.8 (96.5 - 99.7) 98.7 (96.3 - 99.7)
Specificity% (95th CI) 76.7 (73.5 - 79.7) 94.7 (92.8 - 96.1)
PPV% (95th CI) 58.5 (53.6 - 63.2) 84.9 (80.1 - 88.9
NPV% (95th CI) 99.48 (98.5 - 99.89). 99.6 (98.8 - 99.9)
DOR (95th CI) 1.7 (1.5 - 1.78) 1.2 (1.12 - 1.25)

Table 3. Results of the within-run precision of the automated method.

Variable (n = 20) Mean (± SD) CV%


Erythrocytes 69.80 (± 3.955) 5.7
Leucocytes 38.90 (± 1.889) 4.9
Specific gravity 1,021.30 (± 0.470) 0.0
pH 5.95 (± 0.154) 2.6
Microalbumin 147.00 (± 13.416) 9.1
Creatinine 17.700 (± 0.00) 0.0

Table 4. Showing the descriptive statistics of the semiquantitative variables of the inter-observer study.

Number of observers (%)


(n = 20)
+ ++ +++ Few
Epithelial cells
9 (45) 7 (35) 1 (5.0) 3 (15)
Granular Hyaline
Nil
Casts Few + Few +
2 (10) 4 (20) 1 (5) 1 (5) 12 (60)
Calcium oxalate Uric acid
Nil
Crystals Few + ++ Few +
5 (25) 2 (10) 1 (5) 1 (5) 4 (20) 7 (35)
Nil Few +
Amorphous
12 (60) 4 (20) 4 (20)
Nil Few + ++
Bacteria
12 (60) 1 (5) 6 (30) 1 (5)
Nil
Fungi
20 (100)
Nil +
Mucus
19 (95) 1 (5)

Chicago, IL, USA) release 15 for Microsoft Windows RESULTS


(2006). Diagnostic test performance was performed us-
ing the following online calculator: All hospital departments were effectively represented as
http://araw.mede.uic.edu/cgi-bin/testcalc.pl. follows: outpatient clinic represented 22.5%, uro-sur-
gery 11.9%, rheumatology 10%, tropical medicine

52 Clin. Lab. 1+2/2016


Urinalysis, Automation Versus Manual

Table 5. The descriptive statistics of the quantitative variables of the inter-observer study.

Variable N Mean Std. Deviation CV%


Erythrocytes 20 5.60 4.584 81.9
Leucocytes 20 4.43 1.794 40.5

Table 6. Practicability criteria of the manual technique versus the H800-FUS100.

Criterion Manual method H800-FUS100


It is user friendly and requires a well-
User friendly and need for personal
A skilled examiner is needed trained operator (not totally dedicated)
skills
especially for the editing process
Required sample type Centrifuged urine sample Uncentrifuged urine sample
A minimum of 3 mL. (Aspiration volume of
Required sample size 5 - 12 mL
each of H800 and FUS100 is 1 mL)
30 samples/hour because of the
60 samples/hour (not including time needed
need for:
for the editing process) if analysis is applied
Throughput time -manual use of test strips
using the 2 partitions (H800 chemistry and
-centrifugation
FUS100 microscopic)
-manual recording of results
Reagents including: QC, focus, sheath,
Need for reagents and its stability No need for reagents detergent, and diluent are needed and
generally they are stable until expiry date
No need for calibration, external
According to manufacturer's instructions
QC programs are inapplicable.
monthly calibration and daily running of
Calibration and QC frequency Internal QC is possible through
FUSs focus, positive and negative controls
repeating random samples by a
are recommended
different observer
Depends on the applied system in Results storage is available where enough
each laboratory: either daily memory is available:
Results storage results are not stored or stored; -10,000 routine results
in lab log-books or on the Lab -5000 emergency results
information system -1000 QC results
On application of bar coding, should not
Possibility of mixing patients' results Can happen.
happen
Cost depends on the price of materials used:
special tube, specific strips, focus, sheath,
Cost depends on the price of used
Cost/test detergent, diluent. The overall running cost
test strips
is overall comparable between the two
methods

9.8%, general medicine 8.4%, dermatology 7.2%, neu- H800-FUS100 are shown in Table 2. Cutoff points were
rology 5.9%, ICUs collectively 6.6%, cardiology 3.7%. considered > 5 cells/HPF for the diagnosis of hematuria
The remaining 15% were patients of general and special [5] and ≥ 10 for leukocytes as it is more relevant to the
surgeries, psychiatry, and occupational health. Males reflex culture cutoff [6].
represented 48% of the cohort, while 52% were fe- A second cutoff for hematuria was searched. It was des-
males. ignated by the researchers to be ≥ 10/HPF for the cases
Agreement and correlation studies’ results are shown in tested by automation, and the specificity was recalculat-
Table 1. Figures 1 and 2 show how H800-FUS100 strat- ed. Specificity improved to 88.9%, sensitivity remained
ified the studied patients as regards the diagnosis of he- acceptable at 98.8%, with DOR at 1.3 (CI; 1.22 - 1.36).
maturia and pyuria in comparison to the manual meth- Within-run precision testing for the semi-quantitative
od. The derived performance characteristics of the variables by FUS-100/H-800 showed exactly the same

Clin. Lab. 1+2/2016 53


Asmaa Ismail Ahmed et al.

Figure 1. How the FUS100H800 stratified hematuria patients at cutoff more than 5HPF.

Figure 2. How the FUS100H800 stratified pyuria patients at cutoff equal or more than 10HPF.

results for 20 successive times for casts, bacteria, fungi, with a C.V. = 127%. Median negative control was 0.75
urobilinogen, bilirubin, ketone, occult blood, glucose, with (25th - 75th% = 0 - 2).
WBCs, and nitrite. Results of crystals, mucus, and epi- No carryover contaminated the three negative samples
thelial cells showed 95%, 80%, and 75% concordance run after three positive samples. No drift was observed.
respectively. Protein content precision was the least For the manual technique, Tables 4 and 5 show the de-
(50% ++, 45% +, and 5% nil). Quantitative variables of scriptive statistics of the semi-quantitative and quantita-
within run precision are shown in Table 3. tive variables of the inter-observer assay. In the intra-
In the between-run precision study, the positive control observer study, a strong positive correlation was found
showed a mean ± SD of 1050 ± 29.5, C.V. = 2.9%. between the two runs of samples read by the same ob-
Negative control repeats had a mean ± SD of 1.04 ± 1.3 server with r = 0.999 for both erythrocytes and leuco-

54 Clin. Lab. 1+2/2016


Urinalysis, Automation Versus Manual

cytes and a p = 0.000. Practicability criteria of both further study relating the lab diagnosis to the clinical
methods are shown (Table 6). outcome is mandatory to resolve this debate, and per-
haps new cutoffs for hematuria and pyuria will be des-
ignated for the automated count. Furthermore, regarding
DISCUSSION the manual method, at the high level counts, erythro-
cytes and leucocytes are expressed as > 100/HPF. Auto-
Urine analysis is the third major diagnostic screening mated methods, on the other hand, can read figures up
test in the clinical laboratory, only preceded by serum/ to orders of thousands. This means that automated
plasma chemistry profiles and complete blood count methods in general will yield higher readings when
analysis [2]. A correct urine analysis result offers a di- compared to manual results especially at higher levels
rect indication of the state of the patient's renal and gen- of RBCs or WBCs, which is not necessarily clinically
itourinary system and a monitoring of other body sys- significant. This is comparable to previously shown re-
tems [5]. sults [4] on a different instrument (IQ200 analyzer) in
In this study, the H800-FUS100 was assessed as an al- which more RBCs and WBCs were obtained with the
ternative to the manual urine analysis testing procedure. automated than with the manual microscopy.
In contrast to other published studies; this study includ- The fully automated urine devices classify some cells as
ed a large sample size, with a single observer perform- “unclassified” and allow the trained operator to correct
ing both the manual microscopic examination and auto- the designations of the cells. The device accuracy in-
mated visual classification of unclassified particles. The creases when the operator examines these cells and
effectiveness of the manual technique as a standard makes the correct designations [5].
method was also explored. Other performance criteria for the H800-FUS100
The H800-FUS100 agreed well with the manual method showed a good within-run precision for either semi-
in ketone (K = 1), nitrite (K = 1), pH (k = 0.914), glu- quantitative or quantitative parameters except for the
cose (k = 0.78), epithelial cells (k = 0.765), calcium ox- protein content of the sample. Casts, bacteria, fungi,
alate crystals (k = 0.728), bacteria (k = 0.74), amor- urobilinogen, bilirubin, ketone, occult blood, glucose,
phous urates (k = 0.71), albumin (k = 0.695), uric acid and nitrite showed exactly the same results for 20 re-
crystals (k = 0.65), mucous (k = 0.559), and fungi peated times. Calcium oxalate crystals, epithelial cells,
(k = 0.57) with a p-value of (0.000) for all. Granular and mucus showed > 75% precision. Only the protein
cast was the only item that showed poor agreement be- content of the sample failed to show a good precision
tween the two methods (k = 0.317) (p = 0.0000). with 50% of the times (10 times ) showing two pluses
A good correlation was found between the two methods (++), 45% showing one plus (+), and one time showing
as regards the specific gravity (r = 0.884, p = 0.000). a negative result.
Also a moderate correlation was found for RBCs Red blood cells content showed a mean of 69.8 ± 3.95
(r = 0.42, p = 0.000) and WBCs (r = 0.44, p = 0.000). with a C.V. of 5.7. WBCs showed a mean of 38.9 ± 1.9,
Yuksel et al. [5] did a study on the same apparatus and C.V. = 4.9. Specific gravity, pH, microalbumin, and
found that the strip tests for SG and pH and the micro- creatinine also showed good precision results with C.Vs
scopic tests for leukocyte and epithelial cell counts of 0.000, 2.6, 9.1, and 0.00 respectively.
showed good correlations, while protein, glucose, ke- In the between-run precision, good results were found
tones, and erythrocytes showed moderate correlations. for level II (positive) control, with a mean of 1050
Due to the small number of positive specimens for uro- + 29.5, C.V. = 2.9. For level I (negative) control the
bilinogen and bilirubin, it was difficult to find concor- C.V. was strikingly high (C.V. = 127). However, this
dance either in our study or in theirs. However, nitrite can be explained by the very low values of the negative
positive cases were prevalent enough in the current control (0 - 4), which are below the critical medical de-
study and thus agreement could be established. cision level of “5 particles/HPF”.
The H800-FUS100 showed acceptable performance In Yuksel et al. [5] study, the between- and within-pre-
characteristics; however, there was a high false positive cision were lower for the specimens with fewer cells,
rate noted especially in erythrocytes, which means that while for the specimens containing a large number of
the H800-FUS100 diagnoses 41.2% more cases of he- cells, the reproducibility was much better, which is sim-
maturia than the manual method does. Another cutoff ilar to our study and to the previous studies [7].
was set for hematuria for the H800-FUS100 at ≥ 10HPF Also, Mayo et al. [3] stated that precision studies pro-
and as expected the specificity (88.9%) was increased vided adequate variation coefficient values except for
with no major effect on the sensitivity (98.8%). This the samples with poor cellularity.
means that the H800-FUS100 has more tendency to Lamchiagdhase et al. [8] also reported that between-run
classify results in the range of (> 5 - ≥ 10HPF) as a CVs of RBC for negative and positive control were
round figure, a range at which manual microscopic ob- 61.6% and 6.4%, respectively.
servers may designate the sample as normal. It is not The results of the carryover study were consistent with
certain whether this is due to automation overestimation all urine contents of the three negative samples remain-
or to manual technique underestimation (cellular loss ing the same, even when run after three positive sam-
during centrifugation or a manual counting error). A ples. This ensures the efficiency of the apparatus in ana-

Clin. Lab. 1+2/2016 55


Asmaa Ismail Ahmed et al.

lyzing each sample separately and eliminates the fear of Support:


positive contamination of negative samples from pro- This work was supported by a fund from the Faculty of
ceeding positive samples. Medicine, Cairo University.
Carryover studies on other devices (UF-100, iQ200, and
sediMAX) showed no significant increase in the mean
values of the negative pool samples following the path- Declaration of Interest:
ological urine samples (p > 0.05), except for WBC mea- The authors do not have any conflict of interest.
sured by the iQ200 analyzer [3,9].
To study the stability of the apparatus, drift was tested
and showed acceptable stability in spite of the diversity References:
of the samples which arrived from all departments of
the hospital. This can be considered an additional point 1. European Confederation of Laboratory Medicine: European uri-
nalysis guidelines. Scand J Clin Lab Investig 2000;60:1-96.
of strength to the excellent carry over results. This was
a weak point in Yuksel et al. [5] study because the ma- 2. Delanghe J, Speeckaert M. Preanalytical requirements of urinaly-
jority of their specimens were from women and chil- sis. Biochemia Medica 2014;24(1):89-104.
dren, and they were outpatients, so their pathological
3. Mayo S, Acevedo D, Quiñones-Torrelo C, Canós L, Sancho M.
specimens were few. Clinical Laboratory Automated Urinalysis: Comparison among
When comparing the inter-observer results of the manu- automated microscopy, flow cytometry, two test strips analyzers,
al examination method, major discrepancies were and manual microscopic examination of the urine sediments. J
found, with only mucus and fungal contents reaching Clin Lab Anal 2008;22:262-70.
95% and 100% agreement between observers, which is 4. Chien TI, Kao JT, Liu HL, et al. Urine sediment examination: A
probably because they were absent in all samples. For comparison of automated urinalysis systems and manual micros-
all other contents, only ≤ 60% of observers had similar copy. Clin Chim Acta 2007;384:28-34.
readings. For example, only 60% of observers reported
5. Yuksel H, Kilic E, Ekinci A, Evliyaoglu O. Comparison of Fully
casts as (Nil) and less percentages of observers reported Automated Urine Sediment Analyzers H800-FUS100 and Labu-
other options (few, +, ++). Only 35% of observers re- mat-Urised with Manual Microscopy. J Clin Lab Anal 2013;27:
ported crystals as (Nil) with less percentages reporting 312-6.
the presence of Ca oxalate or uric acid crystals at differ-
6. Hwang JH, Park HC, Jeong JC, et al. Chronic asymptomatic py-
ent quantities. uria precedes overt urinary tract infection and deterioration of
This huge unacceptable variability in inter-observer renal function in autosomal dominant polycystic kidney disease.
findings is one of the most important causes for devia- BMC Nephrol 2013,14:1 doi:10.1186/1471-2369-14-1.
tion towards automation of urine analysis so as to mini-
7. Akin OK, Serdar MA, Cizmeci Z, Genc O, Aydin S. Comparison
mize the subjective effect of the lab technologist on pa- of LabUMat-with-UriSed and iQ200 fully automatic urine sedi-
tient results. This effect was proposed to be because ment analyzers with manual urine analysis. Biotechnol Appl Bio-
manual microscopic examination of centrifuged urine chem 2009;53:139-44.
has several variables associated with sample prepara-
8. Lamchiagdhase P, Preechaborisutkul K, Lomsomboon P, et al.
tion, centrifugation, and counting variations [8]. Urine sediment examination: A comparison between the manual
In the intra-observer study of the manual technique, a method and the iQ200 automated urine microscopy analyzer. Clin
strong positive correlation was found between the two Chim Acta 2005;358:167-74.
runs of samples read by the same observer with r equal
9. Zaman Z, Fogazzi G, Garigali G, Croci M D, Bayer G, Kránicz
to 0.999 for both RBCs and WBCs and a p = 0.000. T. Urine sediment analysis: Analytical and diagnostic perfor-
mance of sediMAX® - A new automated microscopy image-based
urine sediment analyser. Clin Chim Acta 2010;411:147-54.
CONCLUSION

The automation of urine examination may decrease ob-


server-associated variation and offer results comparable
to those of manual microscopy in a rapid, reproducible,
and accurate way. This is provided the reliability of the
analyzer is periodically reviewed and the apparatus is
used for screening away from the borderline cutoffs.
Revision either by the manual microscopic examination
or through visual revision of the apparatus images may
be needed to confirm positive readings, and revision of
the cutoffs themselves might be explored to enhance
performance characteristics of the H800-FUS100. It is
also recommended to standardize all pre-analytical pro-
cedures to reduce the huge percentage of fallacies in re-
sults.

56 Clin. Lab. 1+2/2016

Potrebbero piacerti anche