Liquichek Cardiac Markers Plus Control LT


Levels 1, 2, 3, 1A, 1B and 1C
Level 1 23661
146 Level 1 6 x 3 mL 146X Level 1 MiniPak 1 x 3 mL
Level 2 23662
147 Level 2 6 x 3 mL 147X Level 2 MiniPak 1 x 3 mL
Level 3 23663
148 Level 3 6 x 3 mL 148X Level 3 MiniPak 1 x 3 mL 2020-01-31 23660 Level 1A 23664
27104 Level 1A 6 x 3 mL 27104X Level 1A MiniPak 1 x 3 mL
Level 1B 23665
27105 Level 1B 6 x 3 mL 27105X Level 1B MiniPak 1 x 3 mL
Level 1C 23666
27106 Level 1C 6 x 3 mL 27106X Level 1C MiniPak 1 x 3 mL

http://www.myeinserts.com/23660

INTENDED USE LIMITATIONS


Liquichek Cardiac Markers Plus Control LT is intended for use as an assayed quality 1. This product should not be used past the expiration date.
control serum to monitor the precision of laboratory testing procedures for the 2. If there is evidence of microbial contamination or excessive turbidity in the
analytes listed in this package insert. product, discard the vial.
SUMMARY AND PRINCIPLE 3. This product is not intended for use as a standard.
The use of quality control materials is indicated as an objective assessment of the WARNING
precision of methods and techniques in use and is an integral part of good laboratory
practices. Multiple levels of control are available to allow monitoring of the test Biological source material. Treat as potentially infectious.
system’s reliability. Each human donor unit used to manufacture this product was tested by FDA
For customers in Germany: Quality control materials are required for assessment of accepted methods and found non-reactive for Hepatitis B Surface Antigen (HBsAg),
laboratory performance as described in the “Guideline for Quality Assurance of antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2. This product may also
Medical Laboratory Examinations following the German Medical Association” contain other human source materials for which there are no approved tests. In
(Rili-BÄK regulation). accordance with good laboratory practice, all human source material should be
REAGENT considered potentially infectious and handled with the same precautions used with
This product is prepared from human serum with added constituents of human and patient specimens.
animal origin, stabilizers and preservatives. This product is provided in liquid form Safety Data Sheet (SDS) available for professional users on www.bio-rad.com.
for convenience. SPECIFIC PERFORMANCE CHARACTERISTICS
STORAGE AND STABILITY This product is a stabilized liquid product manufactured under rigid quality control
This product will be stable until the expiration date when stored unopened standards. To obtain consistent assay values, the control requires proper storage and
at -20 to -70°C. Do not store this product in a frost-free freezer. handling as described.
Thawed Opened: Once thawed, opened, and stored tightly capped at 2 to 8°C, this ASSIGNMENT OF VALUES
product will be stable as follows: The mean values and corresponding ±3SD ranges in the Assignment of Values Data
- All analytes: 20 days Charts (available separately) were derived from replicate analyses and are specific for
Except: this lot of product. Data from Unity™ Interlaboratory Program are included in the
- Troponin I: 10 days determination of some ranges. The tests listed were performed by the manufacturer
- N-terminal pro-Brain Natriuretic Peptide (NT-proBNP) and B-type Natriuretic Peptide and/or independent laboratories using manufacturer supported reagents and a
(BNP): 8 days representative sampling of this lot of product. It is recommended that each laboratory
- Troponin T: 4 days establish its own acceptable ranges and use those provided only as guides.
Aliquot-Frozen: Once thawed, opened, and stored in tightly capped aliquot vials Laboratory established ranges may vary from those listed during the life of this
at -20 to -70°C, this product will be stable as follows: control. [Customers in Germany have to follow the requirements as described in the
- All analytes: 30 days Rili-BÄK regulation.] Variations over time and between laboratories may be caused by
Discard the remaining material. differences in laboratory technique, instrumentation and reagents, or by
This product is shipped under frozen conditions. manufacturer test method modifications.
PROCEDURE
This product should be treated the same as patient specimens and run in accordance Unity™ Interlaboratory Program is a proprietary Bio-Rad software program holding
with the instructions accompanying the instrument, kit, or reagent being used. more than 2 billion QC data points from thousands of laboratories.
Allow the frozen product to thaw at room temperature (18 to 25°C) for approximately INSTRUCTIONS FOR OBTAINING THE DATA CHARTS
30 minutes or until completely thawed prior to use. Gently swirl the vial several times The Data Charts are available through the Internet, at www.myeinserts.com/23660.
to ensure homogeneity. After each use promptly replace the stopper and return to Follow the directions at the website to receive email notifications of insert updates.
2 to 8°C storage. Alternate methods for receiving data charts are available by contacting your local
If the product has been stored refrigerated, allow it to reach room temperature Bio-Rad Laboratories Office.
(18 to 25°C) before use. Before sampling, gently swirl the vial several times to
ensure homogeneity.
To prepare aliquot samples, allow the control vial to thaw as described above. Gently
swirl the vial several times to ensure homogeneity, then dispense the control material
into aliquot vials and store at -20 to -70°C. If sampling aliquot, use it only once.
Discard the remaining material.
Dispose of any discarded materials in accordance with the requirements of your local
waste management authorities. In the event of damage to packaging, contact the
local Bio-Rad Laboratories Sales Office or Bio-Rad Laboratories Technical Services.

Page 1 English • 2017-04 • 16000202-00


TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C


ANALYTES
B-type Natriuretic Peptide (BNP) Creatine Kinase (CK) Myoglobin Troponin I
CK-MB Isoenzyme Digitoxin N-terminal pro-Brain Natriuretic Troponin T
C-Reactive Protein (CRP) Homocysteine ❑ Peptide (NT-proBNP)

NOTES
❑ No claim is made for expected value or stability of this analyte.

UNITED STATES, Bio-Rad Laboratories


9500 Jeronimo Road, Irvine, CA 92618

Bio-Rad FRANCE, Bio-Rad


3 boulevard Raymond Poincare, 92430 Marnes-la-Coquette
Laboratories Phone:(33) 1-4795-6000 / Fax:(33) 1-4741-9133

Australia, Bio-Rad Laboratories Pty. Ltd., Level 5, 446 Victoria Road, Gladesville NSW 2111 • Phone 61-2-9914-2800 • Telefax 61-2-9914-2888
Clinical Austria, Bio-Rad Laboratories Ges.m.b.H., Hummelgasse 88/3-6, A-1130 Vienna • Phone 43-1-877-8901 • Telefax 43-1-876-5629
Belgium, Bio-Rad S.A.-N.V. Winninglaan 3, BE-9140 Temse • Phone +32 (3)710-53-00 • Telefax +32 (3)710-53-01
Diagnostics Group Brazil, Bio-Rad Laboratórios Brasil Ltda, Rua Alfredo Albano da Costa, 100, Lagoa Santa - MG, CEP: 33400-000 • Phone +55 (31)3689-6600 • Telefax +55 (31)3689-6611
Canada, Bio-Rad Laboratories, Ltd., 2403 Guénette Street, Montréal, Québec H4R 2E9 • Phone 1-514-334-4372 • Telefax 1-514-334-4415
China, Bio-Rad Laboratories Shanghai Ltd., 3rd Floor, #18 Dong Fang Road, Bldg E, Poly Plaza, Pudong, Shanghai, PRC 200120 • Phone 86-21-61698500 • Telefax 86-21-61698599
9500 Jeronimo Road Czech Republic, Bio-Rad spol. s r.o., Nad ostrovem 1119/7, 147 00 Prague 4 • Phone 420-241-430-532 • Telefax 420-241-431-642
Irvine, California 92618 Denmark, Bio-Rad Laboratories, Symbion Science Park, Fruebjergvej 3, DK-2100 Copenhagen East • Phone +45-4452-1000 • Telefax +45-4452-1001
(800) 854-6737 Finland, Bio-Rad Laboratories, Linnanherrankuja 16, FIN-00950 Helsinki • Phone 358-9-804-22-00 • Telefax 358-9-7597-5010
France, Bio-Rad, 3 boulevard Raymond Poincaré, 92430 Marnes-la-Coquette • Phone 33-1-47-95-60-00 • Telefax 33-1-47-41-91-33
FAX (949) 598-1550 Germany, Bio-Rad Laboratories GmbH, Heidemannstrasse 164, D-80939 Munich • Phone +49 (0)89-318-840 • Telefax +49 (0)89-318-84100
bio-rad.com/qualitycontrol Greece, Bio-Rad Laboratories M E.P.E, 2-4 Mesogeion Street, Fourth Floor 115 27 Athens • Phone 30-210-7774396 • Telefax 30-210-7774376
Hong Kong, Bio-Rad Pacific Ltd., Unit 1101, 11/F DCH Commercial Centre, 25 Westlands Road, Quarry Bay • Phone 852-2789-3300 • Telefax 852-2789-1290
Hungary, Bio-Rad Hungary Ltd., H-1082 Budapest, Futo street 47-53, Hungary • Phone +36-1-459-6100 • Telefax +36-1-459-6101
Technical Service: India, Bio-Rad Laboratories (India) Pvt. Ltd., Bio-Rad House, 86-87, Udyog Vihar Phase IV, Gurgaon, Haryana 122 015 • Phone 1800-180-1224 • Telefax 91-124-2398115
Israel, Bio-Rad Laboratories Ltd., 14 Homa Street, New Industrial Area, Rishon Le Zion 75655 • Phone 972-3-9636050 • Telefax 972-3-9514129
(800) 854-6737
Italy, Bio-Rad Laboratories S.r.l., Via Cellini 18/A, 20090 Segrate, Milan • Phone +39-02-216091 • Telefax +39-02-21609553
Japan, Bio-Rad Laboratories K.K., Tennoz Central Tower 20F, 2-2-24 Higashi-Shinagawa, Shinagawa-ku, Tokyo 140-0002 • Phone 81-3-6361-7070 • Telefax 81-3-5463-8481
Korea, Bio-Rad Korea Ltd., 10th Floor, Hyunjuk Building, 832-41, Gangnam-gu, Seoul 135-080 • Phone 82-2-3473-4460 • Telefax 82-2-3472-7003
Mexico, Bio-Rad, S.A., Avenida Eugenia 197, Piso 10-A, Col. Narvarte, C.P. 03020 Mexico, D.F. • Phone +52 (55)5488-7670 • Telefax +52 (55)1107-7246
The Netherlands, Bio-Rad Laboratories B.V., Fokkerstraat 2-8, 3905 KV Veenendaal • Phone +31-318-540666 • Telefax +31-318-542216
New Zealand, Bio-Rad New Zealand, 189 Bush Road Unit B, Albany, Auckland • Phone 64-9-415-2280 • Telefax 64-9-415-2284
Norway, Bio-Rad Laboratories, Nydalsveien 33, 0484 Oslo • Phone +47-23-38-41-30 • Telefax +46(0)8-5551-2780
Poland, Bio-Rad Polska Sp. z o.o., Nakielska Str. 3, 01-106 Warsaw • Phone 48-22-3319999 • Telefax 48-22-3319988
Portugal, Bio-Rad Laboratories, Lda., Edificio Prime, Ave. Quinta Grande, 53 – Fracção 3B Alfragide 26114-521 Amadora • Phone 351-21-472-7700 • Telefax 351-21-472-7777
Russia, Bio-Rad Laboratorii, 117105, Russian Federation, Moscow, Varshavskoe sh., 9, Bldg., 1B • Phone +7-495-721-1404 • Telefax +7-495-721-1412
Singapore, Bio-Rad Laboratories (Singapore) Pte. Ltd., 27 International Business Park, #01-02 iQuest @IBP, Singapore 609924 • Phone 65-6415-3170 • Telefax 65-6415-3189
South Africa, Bio-Rad Laboratories (Pty) Ltd., 34 Bolton Road, Parkwood, Johannesburg 2193 • Phone 27-11-442-85-08 • Telefax 27-11-442-85-25
Spain, Bio-Rad Laboratories, S.A., C/ Caléndula, 95, Edificio M. Miniparc II, El Soto de la Moraleja, 28109 Madrid • Phone 34-91-590-5200 • Telefax 34-91-590-5211
Sweden, Bio-Rad Laboratories A.B., Box 1097, Solna Strandväg 3, SE-171 54, Solna • Phone +46-8-555-127-00 • Telefax +46-8-555-127-80
Switzerland, Bio-Rad Laboratories AG, Pra Rond 23, CH-1785 Cressier • Phone +41 (0)26-674-55-05/06 • Telefax +41 (0)26-674-52-19
Taiwan, Bio-Rad Laboratories Taiwan Ltd., 14F-B, No. 126 Nan-King East Road, Sec. 4, Taipei, Taiwan 10546 R.O.C. • Phone 886-2-2578-7189 • Telefax 886-2-2578-6890
Thailand, Bio-Rad Laboratories Ltd., 1st & 2nd Floor, Lumpini I Bldg., 239/2 Rajdamri Rd., Lumpini, Pathumwan, Bangkok 10330 • Phone 662-651-8311 • Telefax 662-651-8312
United Kingdom, Bio-Rad Laboratories Ltd., Bio-Rad House, Maxted Road, Hemel Hempstead, Herts HP2 7DX • Phone +44 (0)20-8328-2000 • Telefax +44 (0)20-8328-2550 10/14

© 2017 Bio-Rad Laboratories Inc. Printed in the USA Page 2 English • 2017-04 • 16000202-00
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 1 - 23661
Units Mean Range
CK-MB ISOENZYME (MASS)
Abbott ARCHITECT iSystems ng/mL 3.10 2.41 – 3.79
Abbott i-STAT ng/mL 3.50 2.40 – 4.60
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i ng/mL 3.29 2.60 – 3.97
Beckman Coulter UniCel DxI ng/mL 3.69 2.81 – 4.56
bioMerieux VIDAS (3) ng/mL 5.50 3.50 – 7.50
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF0031-K/PF1121-K) ng/mL 2.36 1.42 – 3.30
LSI MEDIENCE PATHFAST (PF1031-K) ng/mL 4.00 2.40 – 5.60
Monobind AccuBind ELISA ng/mL
Monobind AccuLite CLIA ng/mL
Ortho Vitros MicroWell Series (3) ng/mL 2.46 1.76 – 3.16
Radiometer (AQT90 FLEX) (3) ng/mL 4.90 3.91 – 5.88
Roche Elecsys / E170 / cobas e Systems ng/mL
Roche Elecsys / E170 / cobas e Systems (STAT) ng/mL 2.86 2.12 – 3.60
Siemens ADVIA Centaur CP (CKMB) ng/mL 3.06 2.32 – 3.81
Siemens ADVIA Centaur CP (CKMBJ) ng/mL 6.74 5.23 – 8.24
Siemens ADVIA Centaur Systems (CKMB) ng/mL 3.87 2.95 – 4.79
Siemens ADVIA Centaur Systems (CKMBJ) ng/mL 8.27 6.51 – 10.0
Siemens Dimension Series (MMB) ng/mL 2.32 1.27 – 3.36
Siemens Dimension Vista Systems (MMB) ng/mL 3.04 2.23 – 3.85
Siemens IMMULITE / IMMULITE 1000 ng/mL 4.61 3.42 – 5.80
Siemens IMMULITE 2000 / 2000 XPi ng/mL 4.40 3.42 – 5.37
Siemens Stratus CS ng/mL 3.32 2.70 – 3.93
TOSOH AIA Systems (3) ng/mL 3.92 2.94 – 4.90
MYOGLOBIN
Abbott ARCHITECT iSystems ng/mL 76.5 60.4 – 92.6
Beckman Coulter Access/2/2i ng/mL 57.5 48.4 – 66.7
Beckman Coulter AU Systems ng/mL 104 92.8 – 115
Beckman Coulter UniCel DxI ng/mL 61.3 46.2 – 76.5
Biokit ng/mL §
bioMérieux VIDAS (3) ng/mL 48.0 34.0 – 62.0
LSI MEDIENCE PATHFAST (PF0021-K/PF1111-K) ng/mL 37.4 26.2 – 48.6
LSI MEDIENCE PATHFAST (PF1021-K) ng/mL 25.4 17.8 – 33.1
Monobind AccuBind ELISA (3) ng/mL 92.5 62.0 – 123
Monobind AccuLite CLIA (3) ng/mL 93.0 62.3 – 124
Ortho Vitros MicroWell Series (3) ng/mL 103 78.0 – 128
Radiometer (AQT90 FLEX) (3) ng/mL 84.6 67.6 – 102
Roche / Hitachi cobas c Systems ng/mL 63.9 45.0 – 82.7
Roche Elecsys / E170 / cobas e Systems ng/mL 55.3 39.7 – 70.9
Roche Elecsys / E170 / cobas e Systems (STAT) (3) ng/mL 58.5 41.0 – 76.1
Siemens ADVIA Centaur CP (MYO) (Cal. U Lots 65 and higher) ng/mL 83.2 64.4 – 102
Siemens ADVIA Centaur Systems (MYO) (Cal. U Lots 65 and higher) ng/mL 81.7 57.6 – 106
Siemens ADVIA Chemistry Systems (MYO) ng/mL 72.9 50.7 – 95.1
Siemens ADVIA Chemistry Systems (Sentinel) (MYO) ng/mL
Siemens BN Series Nephelometers (MYO) ng/mL 68.1 54.5 – 81.7
Siemens Dimension Series ng/mL 82.7 68.4 – 96.9
Siemens Dimension Vista Systems ng/mL 77.3 67.3 – 87.2

Page 3 English • 2017-04 • 16000202-00

8
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 1 - 23661
Units Mean Range
MYOGLOBIN (continued)
Siemens IMMULITE / IMMULITE 1000 ng/mL 52.8 42.3 – 63.2
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 57.8 45.0 – 70.5
Siemens IMMULITE 2000 / 2000 XPi ng/mL 71.2 54.0 – 88.3
Siemens Stratus CS ng/mL 59.7 51.8 – 67.6
TOSOH AIA Systems (3) ng/mL 54.3 40.7 – 67.9
TROPONIN I
Abbott ARCHITECT iSystems (STAT) (Troponin I hs) (REF 3P25) pg/mL 50.0 23.1 – 76.9
Abbott ARCHITECT iSystems (STAT) (Troponin I) ng/mL 0.249 0.131 – 0.366
Abbott i-STAT ng/mL 0.338 0.235 – 0.440
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i (AccuTnI+3 REF A98143) (TnIA2) ng/mL 0.033 <0.020 – 0.048
Beckman Coulter UniCel DxI Series (AccuTnI+3 REF A98264) (TnIDxI) ng/mL 0.045 <0.030 – 0.083
bioMerieux VIDAS (TNHS) (3) ng/L 82.5 53.3 – 112
bioMerieux VIDAS (TNIU) (3) ng/mL 0.050 <0.010 – 0.110
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF1011-K/PF1101-K/PF0011-K) ng/mL (4)
Monobind AccuBind ELISA (3) ng/mL <0.300
Monobind AccuLite CLIA (3) ng/mL <0.200
Ortho Vitros MicroWell Series (3) ng/mL 0.138 0.090 – 0.186
Radiometer (AQT90 FLEX) (3) ng/mL <0.010
Roche Elecsys / E170 / cobas e Systems (e411) (STAT) ng/mL §
Roche Elecsys / E170 / cobas e Systems (e601, e602) ng/mL <0.300
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 0.072 0.040 – 0.105
Siemens ADVIA Centaur Systems (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 0.079 0.047 – 0.111
Siemens Dimension EXL (LOCI) ng/mL 0.059 0.035 – 0.083
Siemens Dimension Series (CTNI) ng/mL 0.036 <0.040 – 0.066
Siemens Dimension Vista Systems (CTNI) (LOCI) (7) ng/mL 0.063 0.030 – 0.095
Siemens IMMULITE / IMMULITE 1000 ng/mL 0.211 <0.200 – 0.414
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 0.297 <0.200 – 0.501
Siemens IMMULITE 2000 / 2000 XPi ng/mL <0.200
Siemens Stratus CS ng/mL <0.030
TOSOH AIA Systems (cTnI2) ng/mL 0.430 0.300 – 0.560
TOSOH AIA Systems (cTnI3) (3) ng/mL 0.280 0.170 – 0.390

METHOD (1)
Level 2 - 23662
Units Mean Range
CK-MB ISOENZYME (MASS)
Abbott ARCHITECT iSystems ng/mL 13.2 10.5 – 15.9
Abbott i-STAT ng/mL 14.9 11.3 – 18.6
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i ng/mL 13.8 11.4 – 16.2
Beckman Coulter UniCel DxI ng/mL 15.7 12.9 – 18.5
bioMerieux VIDAS (3) ng/mL 20.4 16.3 – 24.5
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF0031-K/PF1121-K) ng/mL 10.2 7.14 – 13.3

Page 4 English • 2017-04 • 16000202-00

8
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 2 - 23662
Units Mean Range
CK-MB ISOENZYME (MASS) (continued)
LSI MEDIENCE PATHFAST (PF1031-K) ng/mL 17.3 12.1 – 22.5
Monobind AccuBind ELISA ng/mL
Monobind AccuLite CLIA ng/mL
Ortho Vitros MicroWell Series (3) ng/mL 9.53 6.81 – 12.3
Radiometer (AQT90 FLEX) (3) ng/mL 20.6 17.5 – 23.7
Roche Elecsys / E170 / cobas e Systems ng/mL
Roche Elecsys / E170 / cobas e Systems (STAT) ng/mL 10.7 8.45 – 13.0
Siemens ADVIA Centaur CP (CKMB) ng/mL 11.8 9.34 – 14.3
Siemens ADVIA Centaur CP (CKMBJ) ng/mL 23.8 18.9 – 28.6
Siemens ADVIA Centaur Systems (CKMB) ng/mL 13.3 10.7 – 16.0
Siemens ADVIA Centaur Systems (CKMBJ) ng/mL 26.3 21.2 – 31.4
Siemens Dimension Series (MMB) ng/mL 10.8 8.35 – 13.2
Siemens Dimension Vista Systems (MMB) ng/mL 12.8 10.7 – 14.9
Siemens IMMULITE / IMMULITE 1000 ng/mL 15.7 12.5 – 19.0
Siemens IMMULITE 2000 / 2000 XPi ng/mL 15.1 11.9 – 18.3
Siemens Stratus CS ng/mL 13.7 11.6 – 15.8
TOSOH AIA Systems (3) ng/mL 15.8 11.8 – 19.7
MYOGLOBIN
Abbott ARCHITECT iSystems ng/mL 138 111 – 164
Beckman Coulter Access/2/2i ng/mL 104 87.2 – 120
Beckman Coulter AU Systems ng/mL 161 144 – 178
Beckman Coulter UniCel DxI ng/mL 110 85.5 – 135
Biokit ng/mL §
bioMérieux VIDAS (3) ng/mL 87.0 67.0 – 107
LSI MEDIENCE PATHFAST (PF0021-K/PF1111-K) ng/mL 81.4 57.0 – 106
LSI MEDIENCE PATHFAST (PF1021-K) ng/mL 55.4 38.8 – 72.0
Monobind AccuBind ELISA (3) ng/mL 153 103 – 204
Monobind AccuLite CLIA (3) ng/mL 158 106 – 210
Ortho Vitros MicroWell Series (3) ng/mL 181 137 – 225
Radiometer (AQT90 FLEX) (3) ng/mL 153 130 – 177
Roche / Hitachi cobas c Systems ng/mL 107 88.2 – 125
Roche Elecsys / E170 / cobas e Systems ng/mL 101 76.8 – 125
Roche Elecsys / E170 / cobas e Systems (STAT) (3) ng/mL 108 75.6 – 140
Siemens ADVIA Centaur CP (MYO) (Cal. U Lots 65 and higher) ng/mL 150 123 – 178
Siemens ADVIA Centaur Systems (MYO) (Cal. U Lots 65 and higher) ng/mL 144 111 – 176
Siemens ADVIA Chemistry Systems (MYO) ng/mL 131 108 – 153
Siemens ADVIA Chemistry Systems (Sentinel) (MYO) ng/mL
Siemens BN Series Nephelometers (MYO) ng/mL 115 97.4 – 133
Siemens Dimension Series ng/mL 149 126 – 171
Siemens Dimension Vista Systems ng/mL 138 121 – 156
Siemens IMMULITE / IMMULITE 1000 ng/mL 90.3 74.5 – 106
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 102 82.3 – 122
Siemens IMMULITE 2000 / 2000 XPi ng/mL 136 105 – 167
Siemens Stratus CS ng/mL 103 90.4 – 115
TOSOH AIA Systems (3) ng/mL 101 80.9 – 121
TROPONIN I
Abbott ARCHITECT iSystems (STAT) (Troponin I hs) (REF 3P25) pg/mL 1657 1211 – 2104
Abbott ARCHITECT iSystems (STAT) (Troponin I) ng/mL 7.76 3.51 – 12.0

Page 5 English • 2017-04 • 16000202-00

8
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 2 - 23662
Units Mean Range
TROPONIN I (continued)
Abbott i-STAT ng/mL 6.45 5.73 – 7.18
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i (AccuTnI+3 REF A98143) (TnIA2) ng/mL 0.691 0.475 – 0.907
Beckman Coulter UniCel DxI Series (AccuTnI+3 REF A98264) (TnIDxI) ng/mL 0.724 0.413 – 1.03
bioMerieux VIDAS (TNHS) (3) ng/L 3034 1960 – 4108
bioMerieux VIDAS (TNIU) (3) ng/mL 1.16 0.880 – 1.44
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF1011-K/PF1101-K/PF0011-K) ng/mL (4)
Monobind AccuBind ELISA (3) ng/mL 1.34 0.900 – 1.79
Monobind AccuLite CLIA (3) ng/mL 1.45 0.970 – 1.92
Ortho Vitros MicroWell Series (3) ng/mL 3.43 2.47 – 4.39
Radiometer (AQT90 FLEX) (3) ng/mL 0.345 0.275 – 0.414
Roche Elecsys / E170 / cobas e Systems (e411) (STAT) ng/mL §
Roche Elecsys / E170 / cobas e Systems (e601, e602) ng/mL 0.406 <0.300 – 0.527
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 2.82 2.17 – 3.46
Siemens ADVIA Centaur Systems (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 3.19 2.64 – 3.74
Siemens Dimension EXL (LOCI) ng/mL 1.14 0.797 – 1.49
Siemens Dimension Series (CTNI) ng/mL 0.683 0.470 – 0.895
Siemens Dimension Vista Systems (CTNI) (LOCI) (7) ng/mL 1.11 0.848 – 1.37
Siemens IMMULITE / IMMULITE 1000 ng/mL 3.60 2.20 – 5.01
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 4.31 2.72 – 5.91
Siemens IMMULITE 2000 / 2000 XPi ng/mL 3.21 2.37 – 4.06
Siemens Stratus CS ng/mL 1.11 0.883 – 1.33
TOSOH AIA Systems (cTnI2) ng/mL 8.03 5.62 – 10.4
TOSOH AIA Systems (cTnI3) (3) ng/mL 9.05 6.79 – 11.3

METHOD (1)
Level 3 - 23663
Units Mean Range
CK-MB ISOENZYME (MASS)
Abbott ARCHITECT iSystems ng/mL 63.5 50.7 – 76.2
Abbott i-STAT ng/mL 72.3 58.3 – 86.3
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i ng/mL 72.1 60.9 – 83.2
Beckman Coulter UniCel DxI ng/mL 78.2 64.6 – 91.8
bioMerieux VIDAS (3) ng/mL 89.0 72.0 – 106
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF0031-K/PF1121-K) ng/mL 51.4 36.0 – 66.8
LSI MEDIENCE PATHFAST (PF1031-K) ng/mL 87.1 61.0 – 113
Monobind AccuBind ELISA ng/mL
Monobind AccuLite CLIA ng/mL
Ortho Vitros MicroWell Series (3) ng/mL 47.8 35.5 – 60.1
Radiometer (AQT90 FLEX) (3) ng/mL 100 85.2 – 116
Roche Elecsys / E170 / cobas e Systems ng/mL
Roche Elecsys / E170 / cobas e Systems (STAT) ng/mL 51.4 41.2 – 61.5
Siemens ADVIA Centaur CP (CKMB) ng/mL 54.0 43.0 – 65.0

Page 6 English • 2017-04 • 16000202-00

8
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 3 - 23663
Units Mean Range
CK-MB ISOENZYME (MASS) (continued)
Siemens ADVIA Centaur CP (CKMBJ) ng/mL 124 111 – 137
Siemens ADVIA Centaur Systems (CKMB) ng/mL 58.3 47.1 – 69.4
Siemens ADVIA Centaur Systems (CKMBJ) ng/mL 125 113 – 137
Siemens Dimension Series (MMB) ng/mL 70.2 56.5 – 83.8
Siemens Dimension Vista Systems (MMB) ng/mL 63.1 55.4 – 70.7
Siemens IMMULITE / IMMULITE 1000 ng/mL 83.3 68.4 – 98.2
Siemens IMMULITE 2000 / 2000 XPi ng/mL 77.5 61.3 – 93.6
Siemens Stratus CS ng/mL 71.3 61.4 – 81.2
TOSOH AIA Systems (3) ng/mL 82.1 65.6 – 98.5
MYOGLOBIN
Abbott ARCHITECT iSystems ng/mL 313 250 – 377
Beckman Coulter Access/2/2i ng/mL 224 189 – 258
Beckman Coulter AU Systems ng/mL 329 296 – 363
Beckman Coulter UniCel DxI ng/mL 237 184 – 290
Biokit ng/mL §
bioMérieux VIDAS (3) ng/mL 179 150 – 208
LSI MEDIENCE PATHFAST (PF0021-K/PF1111-K) ng/mL 204 143 – 266
LSI MEDIENCE PATHFAST (PF1021-K) ng/mL 139 97.3 – 181
Monobind AccuBind ELISA (3) ng/mL 321 215 – >400
Monobind AccuLite CLIA (3) ng/mL 299 200 – 398
Ortho Vitros MicroWell Series (3) ng/mL 380 288 – 472
Radiometer (AQT90 FLEX) (3) ng/mL 341 289 – 392
Roche / Hitachi cobas c Systems ng/mL 223 196 – 250
Roche Elecsys / E170 / cobas e Systems ng/mL 222 178 – 267
Roche Elecsys / E170 / cobas e Systems (STAT) (3) ng/mL 240 168 – 312
Siemens ADVIA Centaur CP (MYO) (Cal. U Lots 65 and higher) ng/mL 353 289 – 416
Siemens ADVIA Centaur Systems (MYO) (Cal. U Lots 65 and higher) ng/mL 322 260 – 383
Siemens ADVIA Chemistry Systems (MYO) ng/mL 287 261 – 312
Siemens ADVIA Chemistry Systems (Sentinel) (MYO) ng/mL
Siemens BN Series Nephelometers (MYO) ng/mL 269 229 – 308
Siemens Dimension Series ng/mL 327 280 – 375
Siemens Dimension Vista Systems ng/mL 309 276 – 343
Siemens IMMULITE / IMMULITE 1000 ng/mL 196 163 – 229
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 222 181 – 262
Siemens IMMULITE 2000 / 2000 XPi ng/mL 299 218 – 380
Siemens Stratus CS ng/mL 223 198 – 247
TOSOH AIA Systems (3) ng/mL 234 199 – 269
TROPONIN I
Abbott ARCHITECT iSystems (STAT) (Troponin I hs) (REF 3P25) pg/mL 6195 4569 – 7822
Abbott ARCHITECT iSystems (STAT) (Troponin I) ng/mL 44.9 31.0 – 58.7
Abbott i-STAT ng/mL 33.5 31.2 – 35.9
Alere Triage (Cardiac Panel) ng/mL §
Beckman Coulter Access/2/2i (AccuTnI+3 REF A98143) (TnIA2) ng/mL 3.33 2.70 – 3.97
Beckman Coulter UniCel DxI Series (AccuTnI+3 REF A98264) (TnIDxI) ng/mL 3.56 2.11 – 5.02
bioMerieux VIDAS (TNHS) (3) ng/L 29459 19031 – 39887
bioMerieux VIDAS (TNIU) (3) ng/mL 5.80 4.78 – 6.82
DiaSorin LIAISON ng/mL §
LSI MEDIENCE PATHFAST (PF1011-K/PF1101-K/PF0011-K) ng/mL (4)

Page 7 English • 2017-04 • 16000202-00

8
TM

Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C

METHOD (1)
Level 3 - 23663
Units Mean Range
TROPONIN I (continued)
Monobind AccuBind ELISA (3) ng/mL 7.35 4.92 – 9.78
Monobind AccuLite CLIA (3) ng/mL 7.30 4.89 – 9.71
Ortho Vitros MicroWell Series (3) ng/mL 23.4 18.2 – 28.6
Radiometer (AQT90 FLEX) (3) ng/mL 1.55 1.31 – 1.79
Roche Elecsys / E170 / cobas e Systems (e411) (STAT) ng/mL §
Roche Elecsys / E170 / cobas e Systems (e601, e602) ng/mL 1.90 1.43 – 2.37
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 20.5 16.5 – 24.4
Siemens ADVIA Centaur Systems (TnI Ultra) (Rgt. Lots 010110 and above) ng/mL 23.5 19.7 – 27.2
Siemens Dimension EXL (LOCI) ng/mL 4.61 3.95 – 5.28
Siemens Dimension Series (CTNI) ng/mL 2.92 2.31 – 3.53
Siemens Dimension Vista Systems (CTNI) (LOCI) (7) ng/mL 4.71 3.51 – 5.90
Siemens IMMULITE / IMMULITE 1000 ng/mL 21.1 13.9 – 28.2
Siemens IMMULITE / IMMULITE 1000 (Turbo) ng/mL 28.6 21.6 – 35.6
Siemens IMMULITE 2000 / 2000 XPi ng/mL 19.1 15.4 – 22.7
Siemens Stratus CS ng/mL 5.11 4.21 – 6.00
TOSOH AIA Systems (cTnI2) ng/mL 49.2 34.4 – 64.0
TOSOH AIA Systems (cTnI3) (3) ng/mL 46.9 35.1 – >50.0

FOOTNOTES
(1) All footnotes may not apply to your custom selected data chart.
(2) Values were obtained at 37°C.
(3) The assigned values were determined using the reagent and/or instrument manufacturer’s protocol and may not represent ±3SD ranges.
(4) The reagent kit manufacturer requests that you refer to LSI MEDIENCE PATHFAST reagent kit’s CONTROL DATA SHEETS for assignment of values.
(5) This product should be run with automatic dilution, the same as patient samples.
(6) A dilution should be made prior to running those tests where the value listed exceeds the highest standard.
(7) Analyte may recover outside the assay analytical range.
Data is not available at this time. Please inquire.
§ The data required to establish the means and acceptable ranges for this assay were not obtained due to limited assignment participation. If your facility is
interested in participating in the Value Assignment Program for this assay, please contact your local Bio-Rad office.
Not applicable

Page 8 English • 2017-04 • 16000202-00