,,-;l~SMART

~,~ SENSOR® Introduction l. Product specifications


lntellilnstruments· Pro Model : AS51 O AS5 l O uses imported air pressure sensor, provide 1. 1 Technical specifications
high accuracy, temperature co mpensated far accurate
Measureing
measurement values . 0-100 hPa / 0-45.15 inH20
range
Differential Pressure Meter This unit can be app lied on chemical industry,
U ser Manual e lectr ical industry, nuclear industry to mea sure
Resolution o. 01 hPa / 0.01 inH20

the difference of a ir pressure. hPa, mbar , Pa, psi , mmH20,


Unit
mmHg ,inH20 ,inHG ,m/s,fpm

Safety notice ±0.03 hPa (0-0. 30 hPa)


±0 .05 hPa (0.31-1 . 00 hPa)
1. About the user manua l.
±(1.5%ofreadings+ O. 1 hPa)
a) Befare use thi s differential air pressure meter, remaining range
please carefully read through this user ma nual, Accuracy
±0. 01 inH20 (0-0 . 12 inH20)
and operate itas wel l.
"" ±0. 02 inH20 (0.13-0. 40 inH20)
b) Please pay attention when the warning symb ol ± (1.5% of readings + O. 04 inH20)
inno c
IUU..J 0 • " LÍ:,. " is appear on the LCD displ ay. remaining range

2. To preven! damage the unit and human body oftheuser. Measuring


speed 0 .5 sec
a) To use it w ithin the range ofthe sp ec ified value.
Don 't use too much force when insert the tube to 1.2 General specifications
the unit inlet.
Operating
b) Don 't place the unit under a cid solvent or other Oº C-50 º C / 32 º F -122 º F
environment
Corrosive solvent environment. Storage
-20ºC-60ºC /-4ºF-140ºF
c)P leas e follow what instruction manu al mentioned environment

method to clean and maintenance the unit. Operating


vo ltage 3x 1 .5V AAA batteries
d) For safety,please use original part and accessories .
50 hours
Battery lite
(without backlight)
S ize 124. 8mm*49 . 8mm*25. 6mm
We ight 7 1.5g

Version : AS510-1-1 2

2. Product outline CD Mea su re va I ue di sp I a y are a 4 .2.3 Selectabl e units.


~ Ba tt er y empty / fu 1 1 s ymb o 1 4.2.3. l A ir pre ss ure units :
@ Average symbo 1 hPa, mbar, Pa, mmH20, inHg, mmHG, psi, m/ s, fpm.
@ Mínimum symbo l 4.2.3 .2 A ir flo w unit:
@ Max i mum symbo 1 mis, fpm.
@ Value hold symbol
(J) Pressur e un i ts 4.2.3.3 Density units :
... 100% @
(J) @ A ir f I ow speed kg/m3 , lb.fh3.
4.3 Setup befare measure
® A i r Den s i t y un i t
®l Setup mode sy mbo 1 4 .3. l Setup the air pressure units :
~~- After turn on the in strument, press "UNIT" button
G) @ 4. DETAIL OPERATION DESCRIPTION once, instrument wi ll go to "units" selection interface,
Q) (2) 4.1 Check th e unit LCD will display " .&. "," +J "and blinking (hPa)
@ 4.1.1 Turning On and offthe unit: as figure. 2 showed (the blinkin g unit is the la st
Press" <!) " to turn on th e unit, then it goes to m eas u red unit or mass setup unit). Press" .&. "
measurement mode: If you want to turn it off, button to selected yo ur de sired air press ure unit,
press " <!) " until lcd no display, th en it " off'. then press [MODE] button to confirm the se lection,
CD Power on / off and escape button 4.1.2 Back light on/off: press" <!) " to cancel th e se lec tion.
~ Mode se I ect ion and conf i rm button Under m easurement mod e, press" <!) " once
@ Se I ect button will turn on th e back li g ht, ifno any opera tion of
@ Uni t selection button the keys, the backlight will turn off after 30 second .
@ LCD display 4.2 Preparation befare measu rement take.
@ Measure in I et 4.2.1 Slide offthe battery door, plac e AAA size
(J)Back mode I p I ate battery x 3 pi eces in corree! polarity.
@ Batt ery compartment
4 .2.2 Pres s" <!) " to turn on the unit, th en it goes to
® Battery door
®l Prote ct i on co ver measurem ent mode, th e lcd displa y showe d the
la st measured unit (e.g. hPa) , the mea sured
differential air pressure va lue and battery empty/full
3. LCD display
symbol , as followi ng fig. 1
fig. 2

Setting Indication
HoldMaxMinAvg illD Symbo ls
fpmm/s @

u,n1,1,n G)
4.3.2)S etup the air flow unit:
After turn on th~ in s lru1m,n l, press "UNIT" button
1CI Ct.Cl.O.O
hPa
once, instrument will go to " units" selection step,
LCD will display" .&. "," +J "and blinking
hPambar psimmH20 (hPa) as fig. 2 showed (the blinking unit is the
(J)
mmHg inH20inHg
fig . 1 last measured unit or mas s setup unit).
® Density(kg/m 3 )(1b/ft 3 )
Press " .&. " button to selected your desired air flow
.... Auto off +J Remarks: Ifyou found th e battery indication show
"D "or" =::) ", please change the batteries ,
unit, then press [MODE] button to confirm the
sel ection, press" <!) "to cancel the selection as
@) otherwise wi ll affect the accuracy ofth e unit. fig. 3 showed.

4 5 6
4.5 Av erage value ofair flow
4. 5. l At the air flow mea sure mode (m/s or fpm),
mis

Air Flow Unit


,,n pres s [MODE] button until LCD disp lay
blinking "Hold", "Avg" and the last average value,
as fig. 8 showed.
mi s I.CU
Hold Avg :llill fig. 8
Density(kg/ml) mis .
fig. 3 fig . 5 u,1-: ~
IU.0 Show last averag e
value , " -----" will
be shown when
fpm there is no averag e
va lu e.
A ir Flo w Uni t
fpm
4.5.2Atthat stage, press [MODE] button for2 seconds,
until LCD showed " Avg" symbo l and ai r fl ow
va lue as fig.9 showed.

fig . 4
Density

....
(lb/ft 3)

fig. 6 Avg:l!ID fig. 9


mis
After you selected the air flow uni t, ifyou want to 4.4 Ready for measure
setup the air density, please follow steps belo w. After finished ali setup step, connect the provided
silicone tube to the inl et ofth e in strument , then
4.3.3 Setup the density unit: you can start mea suring as fig. 7 sbowed.
a) After se lected air flow unit, pre ss [MODE] Avera ge Va lu e
button confirmed selection thenjump to
den sity setup interface.
b) Press" A " button to setup the density va lu e,
va lue from 0-9 in sequential, pre ss [UN IT]
button to change the digit po sition. 4.5 .3 Press (MODE] button to end th e measure of
.e) Jf you selected air flo w un-it in m/s, the ave rage , LCD display blinkin g "Ho ld", " Avg"
dens ity unit should be kg/m3 as fig. 5, ify ou Pa and calcu la ted average value. !fyou sti ll
se lected fpm air flow unit, the dens ity uni t need to measure the average valu e,
should be lb/ft3 as fig. 6 showed. repeat step "4.5.3".
4.5.4 Press [MO DE] button , return to norma l
fig . 7 measuring mode .

7 8 9

4.6 POWER OFF SETUP Max !!ID fig. 11


4.6. J At the power off stage, Press " (!) "button
until LCD showed setup interface, as fig. 1Oshowed.
In this time LCD display " A ", " +J ",
[Auto off] and blinking [ON] (or OFF).
4 .6.2 Pre ss " A " to select "ON" or " OFF" .
4.6.3 Pre ss [MODE] button to confirm this setup. hPa
----- ' ~
~

Resetting
max imum
value

fig. 10

n
ur,
Min !!ID fi g . 12

~
" ON" or " OFF"
will show when
setting. hPa
----- '
~
Re setting
minimum
A Auto off +J
value

4. 7 Pre- seta ble view manual:


Preset is rea l time measure reading,
"Hold", hold tbe last measure reading.
"Max" , ma ximum value , the maximum value of
& Instrument zero point calibration
last mea sure or last setup. Tbe location change of th e instrument may cause
"Min", mínimum value , the mínimum val ue of
measure error or inaccurate. After rest to zero ,
last measure or last setup.
"Avg" , average va lu e, the average value of the location ofthe instrument can 't be changed . Statement
last mea sure or last setup .
To compensa te the location change or long time zero a. We reserve the rights ofupgrading and
Press [MODE] button to se lect the de sired
viewmode. drift, the instrument must be res et to zero every time amending the design ofthe product
4.8 Reset the maximum ar mínimum valu e. as we ll as the manu al updating, and
before measure. zero ca li bration on ly works under
4 .8. 1 At measure mode, press [MODE] button , the product is subj ect to change without
change to "MAX" or "MIN" mode. 0-10% range . At the normal measuring mode,
4. 8.2) Press [MODE] button for 2 secs, unti l any further notification.
LCD showed "---- -" , to showed instrument has
press " A "button, start th e zero reset (instrument b. Dispose ofbattery should be in accordance
been re-seted. As fig. 11 and fig . 12 showed. must be reset zero within I 0hP a air pressure). with local laws and regulation s.

D.
lntell*Safe CE
1O 11 12