Ultrasonic Meter Calibration Report

Pipe Diameter DN300 Date 2018.05.22

Verification Nº JJG 1030-2018<Ultrasonic flowmeter verification procedures> Series Number Q81800105

Ambient Temperature 25℃ Measured Medium Water

Standard Device Before


Normal
Test

Standard Device After Test Normal Test Result Qualified

Static volumetric method/Standard Meter Method Water Flow


Standard Device Name Accuracy 1%
Standard Device

Standard Device Accuracy 0.2% Signal Strength UP : 92 DOWN: 92

Test Standard Meter Flow TemperaturePressure Tested Meter Flow Basic Error Repeatability

Point m³/h ℃ MPa m³/h % %

438.24 15.0 0.300 438.65 0.09


Point One 438.54 438.38 15.0 0.300 438.74 438.69 0.05 0.007
438.35 15.0 0.300 438.68 0.08
175.26 15.0 0.300 175.64 0.22
Point Two 175.64 175.43 15.0 0.300 175.76 175.74 0.07 0.78 0.034 0.175
175.39 15.0 0.300 175.82 0.25
25.46 15.0 0.300 25.63 0.67
Point Three 25.52 25.50 15.0 0.300 25.72 25.67 0.78 0.175
25.52 15.0 0.300 25.67 0.59
Scale Kactor=1

Inspector Mr.Wu Reviewer: Mr.Wei