1 Introducing ARJUN 30
ARJUN 30 is India`s first ultra light palmtop ultrasonic flaw detector which weighs only
800grams along with the battery, It can be easily operated with one hand and its
portability and compactness makes it unique in its type. ARJUN 30 comes with 5 inch
color WVGA LCD/TFT display with wide viewing angles. It has a 7.4 VDC, 4AH
lithium ion battery, which gives four hours of continuous operation on a single charge
and also additional extra battery is provided to enable back to back 8 hours of operation.
500 trace patterns and 50 setups can be easily stored and recalled for the reference
purpose in the ARJUN 30, it has a micro SD Card storage up to 8GB and a USB
communication port for easy data transfer to the computer.

ARJUN 30
MODSONIC

2 How does an Ultrasonic Flaw detector Work


ARJUN 30 is a single channel Ultrasonic-testing instrument used for the inspection of
homogeneous material for the presence of inclusions, porosity and other discontinuities
that could affect the performance of material and components. It can also be used for
thickness gauging of homogeneous material, requiring access from only one side of the
test piece. High frequency sound (Ultrasonic sound) waves are introduced into the test
material/part from a transducer/probe that is usually coupled to the test piece by water or
other suitable coupling liquid. The transducer converts electric signals to Ultrasound and
vice versa. A short burst of Ultrasound is introduced into test material so some or all of
the energy is reflected by
discontinuities. The reflection of the ultrasound
energy is a function of the ratio between the
acoustic impedance of the discontinuity and the
base material. The greater the impedance ratio
the more sound energy will be reflected. The
principle of Ultrasonic testing is shown in fig. 1.
It shows the ultrasonic energy generated in the
test piece and resultant instrument display.
Thickness gauging with ARJUN 30
Operates on the principle of the time-of-flight
ARJUN 30
MODSO
NIC
measurement. This principle utilizes the precise timing of the transit time of a short burst
of ultrasound energy, through a material under test. The ultrasound waves travel to the
far side of the test piece and reflect back to the transducer/probe and a measurement is
obtained.
Basically three typed of transducers/probes are available for different types of
applications:
1> Straight beam Probe (Normal probe)
2> TR probe (Dual crystal probe)
3> Angle beam Probe
1> Straight beam Probe: This probe introduces ultrasound normal to the test piece
surface utilizing longitudinal or compression waves. Normal beam probe is used
mostly for flaw detection and thickness gauging
(refer figure 1.)
2> TR probe: This probe contains separate transmitting
and receiving elements as showing in figure 2
usually mounted on delay lines. This design
improves near surface resolution by separating
the initial pulse from the received echoes. TR
probe is suitable for the thickness gauging of pitting
and corrosion and also for better surface resolution.

ARJUN 30
MODSO
MODSONIC
NIC
3> Angle Beam Probe: This probe introduces ultrasound at an angle to the surface of
the test piece. In most angle beam probes,
the wave energy is mode converted from a
longitudinal wave to a shear wave by the
refraction principle. This probe is suitable for
inspection of the welds. The reason for this is
its ability to position the transducer/probe
away from the weld bead yet propagate energy
into the weld zone. Another reason to use angle
beam testing on the welds is to position the
sound beam more normal to the expected discontinuities since the flaws in welds are
usually perpendicular to the test piece (except porosity). Figure 3 shows the principle
of angled beam weld testing.

Arjun30
4
MODSONIC

3 Parts and Controls of ARJUN 30


10 31 21 11 12
9 13 15
14

6 7

16
4

3
17

1
20 19 18
Arjun30
5
MODSONIC

25

24

23
30

29

22
6

26 28
27
Arjun30
6
MODSONIC

1. POWER key : ARJUN 30 can be switched ON and OFF using this key.
2. MENU UP key: By pressing this key MENU changes in increasing order.
3. MENU DN key: By pressing this key MENU changes in decreasing order.
4. PERA SELECT: This key is used for navigation inside the particular MENU
DN key from one parameter to another.
5. PERA SELECT: This key is used for navigation inside the particular MENU
UP key from one parameter to another.
6. PERA DN key: This key is used for navigation inside the particular parameter
or selecting the action to be performed.

7. PERA UP key: This key is used for navigation inside the particular parameter
or one parameter to another.
8. GAIN DN key: This key is used for decreasing the gain.
9. GAIN UP key: This key is used for decreasing the gain.
10. LED INDICATOR: Led indication is displayed in two colors green when it is
Charging and Red when display is on.
11. LCD SCREEN: It displays the A-Scan trace patterns, menu parameters and
other details.
Arjun30
7
MODSONIC
12. USB DSK Indicator: This symbol is visible when USB DRIVE is connected to
ARJUN 30 or when ARJUN 30 is connected to PC.
13. SD card Indicator: This symbol is visible when SD card is inserted to
ARJUN 30.
14. REJECTION: This symbol is indicated when rejection percent is ON.
15. MODE Selection: This symbol indicates whether the Probe mode is dual
Crystal OR single crystal.
16. GATES: GATES are displayed when PLOGIC(red) and NLOGIC
(yellow) are selected from menu parameters.
17. MEASURE Details: When MEASURE is on details like sound path, depth path,
Surface distance and amplitude percent are displayed in
Bottom of screen.
18. BATTERY Indicator: Indicates battery level.
19. TIME AND DATE: Indicates present date and time.
20. MENU Nos.: Indicates the current menu number.
21. FREEZ: Current screen gets FREEZED by pressing HOT key/
STEP key twice.
22. CHARGER CONNECTOR: Connect the charger here to charge the battery.
Arjun30
8
MODSONIC

23. RX CONNECTOR: A probe is connected to the receiver of ARJUN 30


through this connector.( choice BNC and MINI LEMO).
24. TX CONNECTOR: A probe is connected to the transmitter of ARJUN 30
through this connector.( choice BNC and MINI LEMO).
25. BATTERY COMPARTMENT: Battery is fitted in this compartment.
26. USB SLOT: USB slot is provided for attaching USB drives and DSK.
27. SD CARD SLOT: SD card slot is provided for external storage and data transfer
28. I/O PORT: Encoder interface and Alarm signal.
29. PIN FOR HOLDING STRAP: Two Pins are provided for hand and neck
straps.
30. STAND: Support Stand is provided for loading of ARJUN 30 on a surface.
31. HOT/STEP KEY: It is used for rapid escalation and descend of the parameter
value.

Arjun30
9
MODSONIC

4 Operation detail
ARJUN 30 is user friendly, menu driven Ultrasonic Flaw Detector. To access any
parameter, various menus are available. Details of all menus and parameters are as per
below.
First menu is Basic menu. In this menu, Basic functions of an Ultrasonic
Flaw Detector (UFD) like Range, Fine Range, Delay and Gain functions are covered.

ZERO ZERO : It is to calibrate Artificial zero of the probe.


2200 RANGE : The required test range can be set directly in terms of mm/
RANGE Inch. The range available for steel is from 2.5mm to 10000
mm and rapid escalation or descend can be
100 MM done by pressing STEP/HOT key along with the GAIN
MTL VEL UP/DN.
5920M/S VELOCITY: The required velocity can be set directly in terms of
DELAY Met/Sec (Inch/sec) in MTL/VEL.
By pressing STEP/HOT key along with PERA UP/DN keys
0.00MM
it steps through preset value. In fine mode 1 met. /sec.
GAIN 1 (0.0001 In/us).
25.0 dB DELAY : Partial Range can be set in DELAY using
PERA UP/DN keys. Exact shift/delay in mm (Inch) is
displayed at the bottom.

Arjun30
10
MODSONIC

GAIN STEP: Gain step value can be set by using GAIN UP/DN key along with
STEP/HOT key, it follows dB in steps of 0.1, 0.5, 1, 2, 6, or 14 dB.
GAIN : Gain can be increased/decreased as per value of step by using GAIN
UP/DN keys.

Example: ARJUN 30 is to be calibrated for 100 mm Test Range using Single


crystal (Normal) probe.
 Using PERA SELECT UP/DN keys navigate to RANGE in MENU 1.
 By using PERA UP/DN keys set required range to 100mm.
 Couple the probe on 25mm thickness of V1 reference block, Navigate to ZERO in
MENU 1 using PERA SELECT UP/DN keys and set first echo at 25mm using
PERA UP/DN keys.
 Set 4th echo at 100 mm by changing velocity using PERA UP/DN keys in MTL
VEL, if required.
 Repeat above two steps until both the echoes are at their respective positions.
 Amplitudes of the echoes can be varied using GAIN keys.

Arjun30
11
MODSONIC
Second Menu covers MONITOR GATE a functions.

GATE a : Gate can be changed to PLOGIC, NLOGIC, CRV M,


GATEa
CRV-1, CRV-2, CRV+1, CRV+2 , EXPAND or OFF
OFF
STARTa PLOGIC: When PLOGIC is selected, the MONITOR is
30.0 mm triggered if an echo is present within the gate and its
height is equal to or crosses the set Threshold Level.
WIDTHa
40.0MM NLOGIC: When NLOGIC is selected, the MONITOR is
triggered if there is no echo within the gate or the
LEVELa
height is lower than the Threshold Level.
30 %
CRV M : When DAC is turned ON and CRV M is selected monitor
GAIN 1 Gets triggered if any Echo signal equals or crosses main
25.0 dB DAC Curve
CRV-1: If DAC is turned ON and if any echo signal cross DAC-1 dB
curve then monitor gets triggered.
CRV-2: If DAC is turned ON and if any echo signal cross DAC-2 dB
curve then monitor gets triggered.

Arjun30
12
MODSONIC

CRV+1: If DAC is turned ON and if any echo signal cross DAC+1 dB


curve then monitor gets triggered.

CRV+2: If DAC is turned ON and if any echo signal cross DAC+2 dB


curve then monitor gets triggered.

EXPAND: When EXPAND is selected then the Scale expansion is


activated and the GATEa width is expanded to Full-screen
width for a magnified view.

STARTa: Gate starting position can be set using START. By pressing


PERA UP/DN Keys Gate start can be directly set in terms
Of mm/inch.

WIDTHa: Gate width can be set using WIDTH. By pressing PERA


UP/DN keys end can be directly set in terms of mm/inch.

LEVELa: Gate Threshold Level can be directly set in terms of the


percentage of the Full Scale Height (FSH) by using LEVEL,
By pressing PERA UP/DN keys

Arjun30
13
MODSONIC
Example: The MONITOR to be triggered when any echo crosses the 50% FSH
between 35mm and 80mm.

 In the Gate setting MENU 2, select PLOGIC using PERA UP/DN keys.
 Set Gate START to 35mm using PERA UP/DN keys.
 Set Gate END to 80mm using PERA UP/DN keys in WIDTH.
 Set Gate LEVEL to 50% using PERA UP/DN keys.

Arjun30
14
MODSONIC

Third Menu covers MONITOR GATE b functions.


GATEb : Gate can be changed to PLOGIC, NLOGIC or OFF
GATEb
OFF PLOGIC: When PLOGIC is selected, the MONITOR is
triggered if an echo is present within the gate and its
STARTb height is equal to or crosses the set Threshold Level.
30.0 mm NLOGIC: When NLOGIC is selected, the MONITOR is
triggered if there is no echo within the gate or the
WIDTHb height is lower than the Threshold Level.
40.0MM STARTb: Gate starting position can be set using START. By
Pressing PERA UP/DN Keys Gate start can be directly set
LEVELb in erms of mm/inch.
30 % WIDTHb : Gate width can be set using WIDTH. By pressing
PERA UP/DN keys end can be directly set in terms of
mm/inch.
GAIN 1
LEVELb: Gate Threshold Level can be directly set in terms of the
25.0 dB percentage of the Full Scale Height (FSH) by using
LEVEL, By pressing PERA UP/DN keys.

Arjun30
15
MODSONIC
Forth Menu covers Memory storage/recall and Trace printout functions.

MEMORY : Memory function can be selected A SCAN or SET-UP


MEMORY using PERA UP/DN keys.
A SCAN MEM NO : Desired memory location can be selected from MEM NO
MEM NO* By PERA UP/DN keys. When changing Memory no If Data
is stored and valid then asterisk (“*”) will be displayed next
001
to “MEM NO”.
ACTION ACTION : Required functions like SAVE, DETAIL,DELETE,
COPY ALL, RECALL can be selected using PERA UP/DN
RECALL
Keys. When DETAIL is selected it shows Note data of
Stored A Scan it can be scroll by pressing PERA UP/DN
<NOTE in MEM NO keys.
ENTER>
ENTER :In NOTE ENTER by pressing PERA UP key, the ACTION
Selected is performed and on pressing PERA DN key Note
GAIN 1 menu will be displayed and Side text menu will be changed
25.0 dB for Text editing as shown. Edit/change Note detail if
required by pressing appropriate keys. For exit press press
PERA UP key.

Arjun30
16
MODSONIC

MEMORY
OPERATOR……….
A-SCAN
ABCDEFGHIJKLMNOPQRSTUVWXYZ
LINE
JOB DETAIL..........
UP/DN XXYYZZ
CURSOR DEFECT DETAIL.........
MOVE 00 LOC
EXIT DEFECT LOCATION………
ENTER> 005
REMARK………..
CHAR
TESTED OK
SCROLL

Arjun30
17
MODSONIC
Example: To SAVE current Calibration and A scan in Memory no 3.
 Get required pattern which to be stored. Press STEP KEY twice to freeze the
screen.
 Select A SCAN in Memory using PERA UP/DN keys.
 Set MEM NO to 3 using PERA UP/DN keys.
 Set ACTION to SAVE using PERA UP/DN keys.
 In NOTE/ENTER press PERA DN key so Note menu will be displayed. Edit it if
required and exit from note menu.
( To edit in side text menu- Using PERA SELECT UP/DN keys navigate to LINE
UP/DN then by using PERA UP/DN key select the line for entering text, and then
navigate to CURSOR MOVE and by using PERA UP/DN key and GAIN UP DN or
MENU UP/DN (CHAR SCROLL) keys to edit the text content, also to delete a
character press MENU UP key along with the HOT/STEP key, and to add a space
press MENU DN key along with the HOT/STEP key.)
 Then press PERA UP key in NOTE/ENTER to exit from side text menu. By
pressing PERA UP key again in NOTE/ENTER Current Calibration, A Scan
pattern and Note DATA will be stored at memory no. 3.

Note: If data is already stored at this location, then the message will be displayed on
LCD screen as “FOR MEMORY OVER WRITE PRESS HOT KEY AND ENTER
KEY”, so press HOT/STEP KEY and PERA UP key to over write the data.

Arjun30
18
MODSONIC

DATA TRANSFER INTERFACE:


 To save memory to the SD CARD or USB disk, Switch to MENU-4, select USB
DSK or SD CARD in MEM NO using PERA UP/DN keys (SD card and USB
DSK options are before internal memory numbers so press PERA DN key if you
are in internal memory numbers to come to SD card or USB DSK option) and
change ACTION to SAVE.
 Switch to MENU 30 and navigate to FL_NAME, by using PERA UP/DN keys
give the required filename.
 Again switch to MENU 4 and navigate to NOTE ENTER, then by pressing PERA
UP key memory can be saved. (NOTE: SD card or Pen drives should be inserted
to avail data transfer facility to PC)
 To view or edit details in SD CARD or USB DSK switch to MENU 4 select
destination that is SD CARD or USB DSK in MEM NO, navigate to ACTION
and select DETAIL, navigate to NOTE ENTER and press PARA DN key, side
text menu will appear And it can be scrolled by pressing PERA UP/DN keys in
FL_NAME in MENU 30.
 Saved memory in SD CARD or USB DSK can be recalled in computer.
 To copy all memories from ARJUN 30 to USB DSK or SD CARD switch to
MENU 4, In MEM NO select the destination SD CARD or USB DSK where you
want to copy the files.
 Navigate to ACTION and select CPY ALL, then navigate to NOTE ENTER and
PRESS PERA UP key all the files will be copied to the selected destination.
Arjun30
19
MODSONIC
TO TRANSFER DATA TO PC:

 Three options are provided for transferring or viewing internal memory in PC.
A) USING SD CARD.
 Insert a SD card inside ARJUN 30 save data you want to view or transfer in
PC from ARJUN 30 to SD card, eject the SD card from ARJUN 30 and
connect the SD card to the PC.
B) USING USB DSK.
 Data from SD card can be visible directly in PC without ejecting the SD card
from ARJUN 30, by connecting ARJUN 30 to PC via USB cable
C) USING EXTERNAL USB DRIVES.
 By Connecting a USB DRIVE to ARJUN 30 using a OTG cable, and
transferring or storing data to USB DSK in ARJUN 30 and then accessing
that PEN DRIVE in PC.
(NOTE: Internally stored memory cannot be directly transferred or accessed
in PC without inserting SD card or USB drives.)

Arjun30
20
MODSONIC

Fifth menu covers the required functions for DAC curve.


DAC : DAC setting can be switched to DRAW, ON
DAC or OFF using PERA UP/DN keys. In DRAW mode, peak
OFF points of echo signals are selected.
STRT a ON mode: The selected peak points are digitally connected to form a
030% DAC curve.
PRESS OFF mode: DAC curve will be not displayed on LCD.
ENTER STRT a : Start of the GATE can be set in START a using
PERA UP/DN keys.
POINT To select any peak of the echo signal, move the
ENTER Gate on the echo signal and then press PERA UP/DN
Key in ENTER. POINT value is automatically incremented to
GAIN 1
indicate next point.(If GATE 1 is OFF then turn to
25.0 dB PLOGIC).
PRESS : In ENTER press PERA UP/DN keys to accept the peak
points selected by the GATE a.
POINT : It indicates how many points are selected to draw
DAC curve.
POINT : Parameter will be changed to DAC dB to draw additional
offset DAC curve when DAC is ON.

Arjun30
21
MODSONIC
PRESS/ENTER : Parameter will be changed to “T-L” when DAC is ON
. Gain of machine can be change without affecting curve height.
It can be used for Transfer/Scanning gain correction.

Example: To digitally draw DAC curve to join four peaks of the echo signals using
25mm thick reference block and range of 100mm.

 Set Gain so that all four peaks are visible on the LCD screen above 50% FSH.
 Set DAC to DRAW using PERA UP/DN keys in DAC in MENU 5.
 Move cursor using PERA UP/DN keys in START a to cover the first echo. If
GATE 1 is OFF then switch to PLOGIC.
 Press PERA UP/DN key in ENTER to accept it and Point No. will be incremented
automatically.
 Repeat above two steps for second, third and forth echo.
 Set DAC to ON using PERA UP/DN key in DAC. Now DAC Curve will be
displayed on LCD.

Note: Cursor need not touch the echo for selection and acceptance of the peak point.

Arjun30
22
MODSONIC

AWS Function functions will be display as per selected in


AWS menu 10 CONFIG parameter.
OFF AWS : AWS setting can be switched to SET, MEASURE or OFF
START a using PERA UP/DN keys.
30.0mm SET mode: Make GATE 1 to PLOGIC then Bring reference echo of
SET-REF reflector in between 40 to 60% , Turn SET REF to ON
key so gain of reference echo is recorded and displayed as
OFF
REF GAIN. (RG)

MEASURE : Display “IL” (Indication level) dB value to bring echo


mode to reference level.
GAIN 1
25.0 dB AF : Display attenuation factor to correct for the depth of the
echo signal.

IR : Display indication rating as per AWS code. IR=IL-RG-AF.

Arjun30
23
MODSONIC

Example:
> Select MEASURE POSITION, MEAPOS1 and MEAPOS2 to ON.
> Set GATE 1 to “PLOGIC”
> Couple probe on reference defect and set gain value so echo height within 40 to 60% of
Screen.
> Set GATE 1 in such a way so echo signal within GATE.
> Then turn SET REF to ON so required gain to bring echo to 50% is recorded and
displayed as a REF Gain in GAIN box.
> Set AWS to MEASURE so it starts to display Indicating rating as per AWS code.
Indication rating is based below formula.
IR=IL-RG-AF
IL : Indicating Level… its gain value to bring echo height to reference level (50%).
RG : Reference Gain.. Its Gain value to bring reference echo height to reference
level(50%)
AF : Attenuation Factor… its attenuation of signal in material as per AWS standard.
Based on below formula.
AF= (Sound path (in Inch) -1) * 2
IR : Indicating Rating… Its different in dB between the indication signal and reference
signal with attenuation correction.

Arjun30
24
MODSONIC

DGS functions will be displayed as per selected in menu 11 EVALMOD parameter.

DGS DGS : DGS setting can be switched to SELECT, ON or OFF using


OFF PERA UP/DN keys. In SELECT mode probe type, reference
dia., reference type, required DGS curve dia. etc. can be
START a selected. Refer DGS selection menu. In ON mode DGS curve
30.0mm will be displayed on Screen as per what is selected in DGS
DGS-REF selection menu.
OFF STRT a : To select echo peak as reference move GATE a using
STRT a . If GATE a is OFF then select to
TL-CORR PLOGIC in Menu 2.
00.0dB
SET-REF :When switched to ON, reference echo within gate is
GAIN 1 registered and height of reference echo is brought to ref.
25.0 dB curve height. When DGS is turned to ON then SET-Ref will
automatically switched OFF.
TL-CORR : This correction is required if test object and reference block have different
surface qualities. By varying TL-CORR gain will change but DGS curve height remains
same.(It works when DGS is selected to ON).

Arjun30
25
MODSONIC

DGS Selection Menu : When DGS is selected to “SELECT” DGS selection menu will
DGS SCAL NO D EFFT
be displayed on Screen.
EXIT By pressing Menu UP/DN
SELECT 002 23.1 MM
keys & PERA SELECT
STRT a PROBE DGS-CRV ATT-REF
UP/DN Keys you can
30.0mm M2SN 10.0 MM 00.0 dB navigate inside the DGS
DGS-REF FREQ REFECHO ATT-OBJ SELECT MENU.
OFF 2.00MHZ FBH 00.0 dB Selected Parameter in the
DGS SELECT MENU
TL-CORR DEL-VEL REFSIZE AMPCOR
will be Highlighted.
00.0 dB 2500M/S 11.0 MM 00.0 dB
GAIN 1
25.0 dB

SCAL NO : It scrolls from 0 to 18. Select “0” for custom probe which one is not listed
between 1 to 18. For custom probe select appropriate probe frequency and
probe effective dia. For predefined probe, 1 to 18, it is not required.
PROBE : It displays Model name of predefined probe. (For scale no 1 to 18 but for
scale 0 it will display “CUSTOM”). With each Memory set one Custom
probe parameters will be stored.

Arjun30
26
MODSONIC

PRBFREQ : It displays probe frequency of predefined probe. For custom probe select
appropriate probe frequency by pressing appropriate keys.
DEL-VEL : It displays Ultrasound velocity of probe delay line. It is fixed for
predefined probes.
D-EFFT : It displays Effective diameter of probe crystals. It is fixed for predefined
probe. Select appropriate value for Custom probe.
DGS-CRV :Select appropriate value for which diameter DGS curve is required.
REF-ECHO :Select reference type weather it is FBH, BW (Back wall) OR SDH.
REF SIZE :Select appropriate value as per reference reflector size.
ATT-REF :Select Ultrasound Attenuation factor of Reference object. (dB/meter).
ATT-OBJ :Select Ultrasound Attenuation factor of test object. (dB/meter).
AMPLCOR :This is gain correction. When Angle probe is used and reference echo is
taken from 100 mm or 25 mm arc of V1/V2 reference block.

Example: To plot DGS for 11 mm size equivalent reflector for M2SN probe.
> Switch GATE a to PLOGIC. (NLOGIC will also work).
> Select DGS in menu 11 under EVALMOD. It will not turn to DGS if DAC is
selected to ON.
> Go to DGS menu 6.

Arjun30
27
MODSONIC
> Couple M2SN probe on 25 thick V1 block and calibrate for required range.
> Bring first echo within screen with height between 40 to 80 % FSH preferable.
> Set first back-wall echo signal within GATE a using START a.
> Change DGS to “SELECT” by pressing PERA UP/DN keys.
> Navigate to SCAL NO & Select SCAL NO to 2 by pressing corresponding UP/DN
keys.
> Navigate to DGS-CRV and Select DGS-CRV to 11.0 MM by pressing corresponding
UP/DN keys.
> Select REFECHO to BW by pressing corresponding UP/DN keys.
>ATT-REF and ATT-OBJ to be set as per appropriate value. Here it is set to 0.
> For exit Navigate to EXIT and Press PERA UP/DN keys.
>Select DGS-REF to ON so reference echo height is registered and brought to reference
curve. (make sure reference echo is within GATE 1 with 40 to 80% FSH preferable).
> Now navigate to DGS REF REC and Press PERA UP/DN key so it turns to ON and
appropriate DGS curve will be displayed.

Arjun30
28
MODSONIC
MODSO
NIC
Sixth Menu covers Rejection of unwanted signal, Set Reference, Video setting and
Testing mode selection function.

DAMPING DAMPING: Probe Damping HIGH/LOW can be selected using PERA


LOW UP/DN keys in DAMPING.
MODE : Test Mode for the use of SINGLE or DUAL Crystal probes
can be selected using PERA UP/DN keys in MODE.
MODE
PRF : Using PERA UP/DN keys PRF can be selected from
SINGLE High, Low and Medium. Low PRF value is useful
PRF MOD when ghost echo pattern appears when testing low attenuate
HIGH material.

GAIN 1
25.0 dB

Arjun30
29
MODSONIC

Seventh menu covers the required functions for Thickness/Distance measurements.


REJECT REJECT : Rejection level can be set as per requirement. For the
rejection of unwanted signals, Using PERA UP/DN keys
00% in Reject .It can be set directly in terms of percentage
SMOOTH of FSH. Activation of this function is indicated by REJ
OFF SYMBOL ®.
VIDEO SMOOTH : Smoothness of the echo can be set to LOW,MEDIUM ,
ENVELOP HIGH OR OFF using PERA UP/DN keys.
SET REF VIDEO : Echo signals can be selected to appear as line (ENVELOP)
OFF or solid (FILLED) using PERA UP/DN key in VIDEO.
In ENVELOP mode, echo signals appear same as in
GAIN 1
Analog Ultrasonic Flaw Detector. In FILLED mode, echo
25.0 dB signals looks solid for better visibility.
SET REF : To set any A-scan pattern as a reference, achieve
reference pattern, then set SET REF ON, so reference
pattern sets in background for easy comparison.
A-scan Pattern From Memory location can be recalled
and set as a reference.

Arjun30
30
MODSONIC

Eighth menu covers the required functions for Thickness/Distance measurements.

UNIT UNIT : Unit of measurement that is Metric/Inch is selectable


MM
COLOR COLOR : LCD Color combination can be selected using PERA
UP/DN keys in COLOR.
05
BRIGHT BRIGHT : Brightness of LCD can be adjusted for better visibility
50% using PERA UP/DN keys in BRIGHT. Set Low
brightness when used for indoor application for the longer
DISPLAY battery life.
RIGHT
GAIN 1 DISPLAY: Enables rotation of the display by pressing PERA UP/DN
Key in DISPLAY.
25.0 dB

Arjun30
31
MODSONIC
Example: To measure Beam path, Surface and Depth distance when 45 deg. probe is
used.

 Go to GATE a in MENU 2 and make Gate to PLOGIC/NLOGIC.


 Set Gate STARTa and Gate WIDTHa value in such a way so defect signal within
the GATE.
 Set Gain or Threshold so height of the echo/defect beyond the GATEa.
 Go to measure menu and MENU11 set MEAPOS1 to “ON G A”.
 Set angle value to 45 DEG in MENU9. Using PERA UP/DN keys in angle.
 Set appropriate THICK value in MENU9 so it can account multiple skips for
DEPTH distance calculations.
 Sound path, Depth, and Surface distance will be displayed on the bottom of the
screen.

Arjun30
32
MODSONIC

Ninth Menu covers miscellaneous functions, which are described as below.

X-OFF : Set X-OFF value to same as Angle probe index point


X-OFF To Front face distance so Surface Distance will be
0.00mm displayed. On LCD will be distance from angle probe
ANGLE front face to defect position.
0.0 DEG
ANGLE : Value to be set as per Angle of the probe. If angle
THICK value not zero and Measure is set to ON then
100 mm appropriate Surface and Depth of defect is calculated and
TRIGGER display on LCD.
FLANK THICK : Thickness value should be set as per the thickness of the
GAIN 1 test piece, so as to enable in calculation of
25.0 dB depth/distance accounting for multiple skips.

TRIGGER: Select Measurement point Weather trigger by FLANK or


on PEAK.

Arjun30
33
MODSONIC

Tenth menu covers the required functions for HORN and Key Beep Setting.
HORN : Horn can be switched ON or OFF using PERA UP/DN
HORN Keys in HORN. When HORN is switched ON, it
OFF gives out an audible alarm each time MONITOR is
BEEP triggered.
ON BEEP : Key press will be acknowledge by sound if BEEP is
CLOCK selected to ON. Otherwise it will not give beep sound
OFF when set to OFF.

FL_NAME CLOCK : Clock can be switched ON or OFF using PERA


UP/DN keys in CLOCK. If CLOCK is set to ON then
ARJ0001 date and time will be displayed at bottom of A-Scan.
GAIN 1 FL_NAME : It is the file name that is stored when MEM NO
25.0 dB is selected to SD CARD or USB DSK.

Arjun30
34
MODSONIC

Eleventh menu covers Selection of evaluation modes for inspection, And measurement
of parameters from respective gates
EVALMOD : Evaluation mode can be selected to DAC, AWS or
EVALMOD DGS. As per selection of evaluation mode parameters
in Menu 5, will turn to DAC, AWS or DGS.
DAC
MEAPOS1 MEAPOS1 : Measured Position 1, Measured values (SP) Sound path
ON G A (DP) Depth, (SD) Surface distance & (AMP) Amplitude
from GATE a OR GATE b will displayed at the bottom
MEAPOS2 of the screen.
ON G B
MEAPOS2 : Measured Position 2, Measured values (SP) Soundpath
(DP) Depth, (SD) Surface distance & (AMP) Amplitude
from GATE a OR GATE b will displayed at the bottom
of the screen.
GAIN 1
MEASURE setting can be switched to ON G A, ON G B
25.0 dB
ON G AB or OFF by pressing PERA UP/DN keys in MEASURE
MENU11.ON G A mode it will display thickness/distance of echo signal
which is in-side the gate A. In ON G B it will display thickness/distance
of echo signal which is inside gate B and in ON G AB it will display
thickness/distance of echo signal for both gates together. In OFF mode it
will not display thickness/distance.

Arjun30
35
MODSONIC

Twelveth menu covers the UT DATA recording and evaluation feature.

FL_NAME FL_NAME : It is the filename of the Recorded UT Data which is stored


LAS0001 Directly into the SD CARD. If Data is stored and valid
then asterisk (“*”) will be displayed next to FL_NAME.
ACTION
ACTION : Required functions like RECORD, PLAY, ANALYSE,
RECORD DETAIL, DELETE, CPY USB, CPY ALL can be selected
ENTER using PERA UP/DN Keys. When DETAIL is selected it
START shows Note data of Stored Data.
RECORD: To Start a new recording select RECORD in ACTION and
SCROLL then navigate ENTER and press PERA UP KEY, a new
00 recording will start.
GAIN 1 PLAY: To play a saved recording select PLAY in ACTION and
then navigate ENTER and press PERA UP KEY, a selected
25.0 dB recording will be played.
ANALYZE: To manually analyze a saved recording select ANALYSE in
ACTION and then navigate ENTER and press PERA UP
KEY, a selected recording will come on the screen, now
navigate to SCROLL and scroll the yellow cursor to analyze
the recorded file.

Arjun30
36
MODSONIC

CPY USB: Copy selected (Displayed under FL_NAME heading) file to


USB disk. Select CPY USB in ACTION and press PERA UP
KEY in ENTER so selected file will be copied in the
USB Disk.
CPY ALL: To transfer all the recorded files to external USB disk select
CPY ALL in ACTION and press PERA UP KEY in
ENTER, all files will be copied in the USB Disk.

ENTER/START: By pressing PERA UP key, the function selected in the


ACTION will perform.

SCROLL : When ACTION is selected to ANALYSE, manual


evaluation of the CB-SCAN and A-SCAN can be done on
the machine screen by scrolling the cursor point.

(NOTE: Recorded files will be stored directly in the SD CARD, if SD


CARD is not inserted it will display an error message “File Destination
Not Found” and files will not be saved.)

Arjun30
37
MODSONIC

Thirteenth menu covers the ENCODER Inputs and Outputs.

ENCODER ENCODER: When Encoder is set to ON, Encoder value will be


ON displayed at the bottom corner.
ON
CAL FAC CAL FAC : Calibration Factor value is needed to be entered to
1000 calibrate the encoder.
>> To reset encoder position to “0” Keep press Hot key and
turn ENCODER to ON so value will be set to “0”
Encoder Calibration:
>> Turn ENCODER to ON by keeping Hot key press so current
position value will set to “0”.
>> Move encoder for known distance say about 100 to 500 mm.
GAIN 1 >> Set Displayed position value to actual travel distance value by
25.0 dB adjusting “CAL FAC”. Check again by moving encoder for
known distance so it will display same value.

Arjun30
38
MODSONIC

5 Setting of time: Keep pressing PARA SELECT UP key and switch ON


FOR ARJUN 30. After display of time and date setting menu Parameter, release PARA
DATE SELECT UP key. Now set required Time/Data by using PERA UP/DN keys.
AND Press MENU DN key to store current time and date value to ARJUN 30.
TIME Press MENU UP key to exit without storing/setting
SETTIN Current time and date setting.
G
Factory Default Calibration set-up:
Last calibration setting remains stored in ARJUN 30. When ARJUN 30
starts again for the use after a switch off it will recall the last setting. But
if due to any reason stored calibration is corrupted then it will start with
factory default calibration set-up. If user wants to recall factory default
calibration set-up manually then one needs to press PARA SELECT DN
key and switch ON
ARJUN 30. After displaying menu parameter release PARA
SELECT DN key.
FOR
FACTORY
DEFAULT
CALIBRA
TION

Arjun30
39
MODSONIC

6 Power Supply
ARJUN 30 is powered by a Lithium-Ion battery pack MB-1, which is fitted in
ARJUN 30 using two half screws. It is electrically connected to ARJUN 30 via one
internal terminal contact point. With fully charged battery pack MB-1, ARJUN 30
can operate continuously for Eight hours. The battery pack MB-1 can be charged
inside ARJUN 30 by connecting Battery charger AC-1 to the connector (22). Also
dock station ABC 1 is provided for charging battery pack outside ARJUN 30 which
allows uninterrupted use of ARJUN 30.

Battery Charging
The battery Charger AC-1 supplied with ARJUN 30 is a special quick- charge
charger. The charging time is just 3 to 3 ½ hours. AC-1 battery charger is suitable for
use with mains supply from 100 to 240 volts AC/50 Hz. The LED on charger
illuminates to indicate power on. When the battery voltage is low (discharge
condition) it will indicate by beep Sound and blinking of battery Indicator. When the
battery is in charging condition LED Indicator on Arjun30 glows green, And in fully
charged condition green LED light shuts off.

Arjun30
40
MODSONIC

Battery Replacement

The battery can be easily replaced inside the machine by unlocking quick release
fasteners by using supplied KEY along with the standard Arjun KIT.

Arjun30
41
MODSONIC

7 Precaution
 Avoid shock to the ARJUN 30 and probes.

 Wipe the couplant off the probe and ARJUN 30 after use.

 When battery pack is out-of ARJUN 30, keep battery contacts away from
metal object

 Never heat the battery pack ATB 1 nor throw into a fire.

Arjun30
42
MODSONIC

8 Specification of ARJUN 30
Test Range : 2.5 mm to 10 meter (0.500 in to 400 in) (in steel). In Hot Key
mode it has 13 preset values. Fine mode adjustable in step of
1mm (0.2 in).
Velocity : 1000 met./sec to 15000 met./sec (0.0400 in/µs to 0.4000 in/µs).
In Hot Key mode it has 7 preset values. Fine Mode adjustable
in step of 1 met/sec (0.0001 in/µs).
Delay : 120 mm adjustable in 0.5mm or 5mm step.
Gain : 100dB calibrated gain adjustable in 0.1, 0.5, 1, 2, 6, or14 dB step.
Rejection : 0 to 125% FSH with LED indicator.
Rectification : Full-wave rectified with filtering.
Frequency : Broad Band amplifier 0.5 MHz. to 15MHz.
Linearity Deviation : Vertical: +3 %, Horizontal + 0.5%.
Test Modes : Pulse echo and Transmit/Receive.
Damping : Damping high/low is selectable.
Connectors : LEMO (Size 00) 2 Nos.

Arjun30
43
MODSONIC

Monitor : Dual gate adjustable in 1% of Screen width with


Positive/Negative Logic, Gate Expand modes.
Gate Expand : Expands Range to width of the gate.
A-Scan memory : 500 Trace Patterns can be stored (with Note/Detail), recalled,
or transferred to PC via USB port. A scan trace patterns can be
directly stored on SD card or USB disk.
USB and SD card : External USB disk and Micro SD card (up to 32GB) ports
provided for data storage.
Calibration Set-up : 50 different calibration set-ups can be Stored and Recalled.
Software : Arjunsoft Interface software for transferring A-Scan from
ARJUN 30 to PC is supplied.
Display : 5 Inch color wide VGA LCD/TFT display (800*480 pixels, size
108*64mm) with higher resolution LCD for better signal
representation and better visibility with different color sets.
Reference A-Scan : Reference A-Scan pattern of standard test object can saved and
recalled in the background for easy comparison during testing.
Echo Dynamics : Echo dynamic pattern display.

Arjun30
44
MODSONIC

DAC : Dynamic DAC curve can be digitally plotted (Smooth parabolic


curve) on screen with selectable offset curves from 0dB to 14dB
in 0.5 dB selectable steps. DAC curve can be set as flaw monitor
gate. Using minimum 2 to maximum 10 points.
DGS : Defect size evaluation using predefined probe settings and
custom probe set-ups, ERS (Equivalent Reflector Size) can also
be displayed on the screen.
AWS : Built in Software for Evaluation of defects in accordance with
AWS Specification D1.1.
Built-In UT Data : Carries out 100 Hours of continuous /Dynamic A-Scan and
Recording B-scan recording (in 8 GB SD Card) which can be replayed and
analyzed in Machine as well as PC.
Colour B-Scan : Real time Colour coded Un-Encoded B-Scan along with A-Scan
on the Screen which can be transferred to PC, Also analysis of
A-Scan and B-Scan can be done in the machine as well as PC
Digital Readout : Thickness/Depth can be displayed in digital readout when using
a normal probe and Sound path, Surface Distance and Depth are
directly displayed when angle probe is in use.
Measurement Unit : Metric or British unit of measurement is selectable.
Arjun30
45
MODSONIC
Update Rate : 50 Hz.
Power : Lithium-Ion Battery pack 7.2 Volts, 4AH, gives 8 hours
continuous operation from fully charged battery.
Charger : Input voltage 100 to 240 VAC. Charge status indicators provided.
Temperature : 0 to 55º C.
Size : 117 x 176 x 56 mm (HxWxD).
Weight : 800 gms.with Battery.

Arjun30
46
MODSONIC

9 Menu Tree
First Menu ZERO RANGE VEL DELAY
Second Menu GATE a START a WIDTH a LEVEL a
OFF
PLOGIC
NLOGIC
CRV M
CRV -1
CRV -2
CRV +1
CRV +2
Third Menu GATE b START b WIDTH b LEVEL b
OFF
PLOGIC
NLOGIC
Forth Menu MEMORY MEM NO ACTION NOTE
SET-UP RECALL ENTER
SAVE
DETAIL
DELETE
CPY ALL

Arjun30
47
MODSONIC
A SCAN RECALL
SAVE
DETAIL
DELETE
CPY ALL
Fifth Menu DAC START a PRESS POINT
(DAC)
OFF ENTER
DRAW HI
ON
(AWS) Menu OFF STARTa SET-REF RATING
SET START Ga ON
OFF
MEASURE IL ---- AF---- RATING
(DGS) Menu DGS STRT Ga DGS-REF TL-CORR
OFF OFF
SELECT ON
REF REC
ON

Arjun30
48
MODSONIC

Sixth Menu DAMPING MODE PRF MOD


LOW SINGLE MEDIUM
DUAL HI
HI LOW
Seventh REJECT SMOOTH VIDEO SET REF
Menu
OFF ENVELOP OFF
LOW FILLED ON
MEDIUM DYNAMIC
HIGH
Eighth Menu UNIT COLOR BRIGHT DISPLAY
MM RIGHT
LEFT
Ninth Menu X-OFF ANGLE THICK TRIGGER
FLANK
PEAK
Tenth Menu HORN BEEP CLOCK FL_NAME
OFF ON OFF
ON OFF ON

Arjun30
49
MODSONIC
Eleventh EVALMOD MEAPOS1 MEAPOS2
Menu
DAC OFF OFF
AWS ON G A ON G A
DGS ON G B ON G B
DAC/DGS DAC/DGS
Twelfth FL_NAME ACTION ENTER SCROLL
Menu
RECORD START
PLAY STOP
ANALYSE
DETAIL
DELETE
CPY USB
CPY ALL
Thirteenth ENCODER CAL FAC
Menu
OFF
ON

Arjun30
50
MODSONIC

All possible care has been taken in the preparation of this operation manual. If you find
any errors or would like to make suggestions for improvement to this operation manual,
then please write to:

MODSONIC INSTRUMENTS MFG. CO. (P) LTD.


Plot No.33, Phase III, GIDC Industrial Estate, Naroda, Ahmedabad – 382 330 INDIA.
Tel.:- +91 (079) 2281 1217, 2281 3131, 2284 1294 Fax.:- +91 (079) 2282 0012.
E-mail: modsonic@modsonic.com Web site: www.modsonic.com

MODSONIC reserves the rights to make changes without notice.

Arjun30
51
MODSONIC

ARJUN 30
Operating Manual

Arjun30
52

Potrebbero piacerti anche