Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User manual
G-30-1781-en
Version 4.0
2012-05-13
S100 / OPMI® pico
About this manual The instruction for use is part of the delivery package.
• Carefully read it before using the device.
• Keep it at the site of use of the device.
• Store it for the entire service life of the device.
• Pass it on to every subsequent owner or user of the device.
Orientation aids – The chapter overview at the beginning of these instructions for use pro-
vides a summary of all subjects.
– The contents of each chapter are specified in detail at the beginning of
each chapter.
– A list of abbreviations, key words and technical terms in the annex facili-
tates the search for specific terms.
Scope This user manual applies for the S100 / OPMI(r) pico with the following labe-
ling:
Version 4.0
Page 2 G-30-1781-en
S100 / OPMI® pico
Version 4.0
G-30-1781-en Page 3
S100 / OPMI® pico
Version 4.0
Page 4 G-30-1781-en
S100 / OPMI® pico
Chapter overview
Version 4.0
G-30-1781-en Page 5
S100 / OPMI® pico
Version 4.0
Page 6 G-30-1781-en
S100 / OPMI® pico Safety Measures
Safety Measures
Safety devices..................................................................................22
Version 4.0
G-30-1781-en Page 7
Safety Measures S100 / OPMI® pico
Version 4.0
Page 8 G-30-1781-en
S100 / OPMI® pico Safety Measures
Key to symbols
We would like to inform you about safety aspects which must be observed
when handling this device. This chapter contains a summary of the most im-
portant information concerning matters relevant to instrument safety.
Hazard symbols
The following safety information has been incorporated into the instructions
for use. Please note this information and be particularly careful in these cases.
WARNING Warning label, which may refer to fatal injuries or severe injuries if pre-
cautions are not followed.
CAUTION Indicating a hazard, which may lead to moderately severe injuries
if risks are not avoided.
NOTE Warning label, which may refer to minor injuries and property damages
if precautions are not followed.
Information symbols
The following information symbols are used in these instructions for use:
– Listing
Requirements for an action
• Prompt for action
Result of an action
Additional information and hints
Version 4.0
G-30-1781-en Page 9
Safety Measures S100 / OPMI® pico
Target group
Field of application
Intended use
OPMI pico is a surgical microscope intended for the illumination and magnifi-
cation of the surgical area and for the support of visualization in surgical pro-
cedures.
Normal use
Version 4.0
Page 10 G-30-1781-en
S100 / OPMI® pico Safety Measures
OPMI pico is a manually operated microscope that can be used whether the
patient is in a sitting or recumbent position.
Provision: OPMI pico offers in connection with different suspension systems (e.g. floor
stand, wall mount, ceiling mount) and flexible support arms an ergonomic
working position within the treatment environment or for the surgical use.
The system can be securely moved into the working position with the hand
grips without any problems.
Treatment / Surgery: The microscope is positioned and roughly focused by the physician or the sur-
geon.
The most frequently used functions as the setting of focus, magnification and
light intensity are manually controlled by the physician or the surgeon during
treatment or surgery.
A video device is connected to the system for learning purposes or in order to
visualize the treatment / surgery progress.
After the treatment / surgery The physician or the surgeon swings the surgical microscope from the treat-
ment or surgical area into its standby position.
It is possible to clean and sterilize the OPMI pico microscope.
Disposal: For the disposal of electrical and electronic devices, please observe the rele-
vant national legislation and other regulations applicable in your country.
The correct use of the device is absolutely vital for safe operation. Therefore,
please thoroughly familiarize yourself with the content of this user manual be-
fore starting up the device. Please also observe the user manuals of additional
instrument equipment. Further information is available from our service de-
partment or from authorized representatives.
Version 4.0
G-30-1781-en Page 11
Safety Measures S100 / OPMI® pico
• Only operate the device within the scope of its intended use.
• Only personnel who have undergone training and instruction are allowed
to use this instrument. It is the responsibility of the customer or institution
operating the equipment to train and instruct all staff using the equip-
ment.
• Please keep these instructions for use where there are easily accessible at
all times for the persons operating the system.
• Please observe all symbols and labels attached to the device! (see page 24)
• Please do only use approved original accessories or consumables.
Legal regulations • Observe the legal regulations for accident prevention and occupational
health and safety applicable in the country concerned.
• This system must not be modified without the manufacturer's approval. If
the system is modified, suitable inspections and testing must be complet-
ed to ensure that it can still be used safely.
The manufacturer is not liable for damage caused by unauthorized per-
sons tampering with the system. Furthermore, this will forfeit any rights
to claim under warranty.
• Modifications and repairs of this device or any equipment operated to-
gether with this device may only be performed by Carl Zeiss service staff
or other persons authorized by Carl Zeiss.
• If required by the regulations and directives applicable in the country of
use, connect the system to a special emergency backup power supply.
Ambient requirements • Do not operate the system contained in the delivery package
– in explosion-risk areas,
– if inflammable anesthetics or volatile solvents such as alcohol, benzine
or similar chemicals are present at a distance of less than 25 cm.
• Do not station or use the instrument in damp rooms. Do not expose the
instrument to water splashes, dripping water or sprayed water.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Make sure that the installation conditions and the use of the device meet
surgical requirements: See Ambient conditions on page 272
Electrical safety • Always switch off the system before disconnecting it from / connecting it
to line power, for cleaning its surface or if it will not be used for a pro-
longed period of time.
Version 4.0
Page 12 G-30-1781-en
S100 / OPMI® pico Safety Measures
• Switch off the unit at the power switch if you notice any smoke, sparks or
unusual noise. Do not use the unit until it has been repaired by our service
team.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live compo-
nents. If you remove the housing, you run the risk of electric shock.
• Do not touch the system if your body is electrostatically charged and the
system is not grounded.
• Please note the information on EMC (electromagnetic compatibility) in the
chapter "System data" on page 258.
Transport and service • The device shall only be transported over long distances (e.g. move, re-
turning for repair etc.) within its original packaging or in a special return
packaging. For details, please contact your dealer or the Carl Zeiss service
team.
• This system is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
• In order to prevent any impairment of the device's safety due to age,
wear, etc., the user must ensure that the device is subjected to the neces-
sary safety checks (see section "Care and maintenance").
Optics and light • Never directly look with binocular tube, object lens or an eyepiece into the
sun.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Adjust the brightness at the stand. Start with the lowest brightness and
slowly increase it to the required but not more than the allowed bright-
ness.
• With increasing age of the light source, the actual illumination intensity
delivered at the respective setting decreases (this is a normal property of
the system).
Additional devices that are connected to electronic medical devices shall
comply with the relevant standards and guidelines
(for example IEC 60950 for data processing devices).
Furthermore, any configurations shall comply with normative requirements
for electronic medical devices /please refer to IEC 60601-1-1:2000 or IEC
60601-1:2005, chapter 16).
The one who is connecting additional devices to electronic medical devices
acts as system configurer and is responsible for the system complying with the
Version 4.0
G-30-1781-en Page 13
Safety Measures S100 / OPMI® pico
Version 4.0
Page 14 G-30-1781-en
S100 / OPMI® pico Safety Measures
Carl Zeiss Service or a person authorized by Carl Zeiss will install the device.
Please make sure that the following requirements continue to be met for fur-
ther operation:
The connecting components have been properly connected. The screw
connections have been firmly tightened.
All cables and plugs are in perfect condition.
Determine the supply voltage at the place of installation and set the volt-
age of the device accordingly.
Check whether the device fuse used is the correct one as specified on the
fuse label.
More specific information about the floor stand:
The main plug shall only be connected to a socket with
a faultless protective ground conductor.
The device is connected with the power cord supplied.
Additional information about wall and ceiling mount:
The device may only be connected to a power supply network which has
a faultless protective ground conductor.
Version 4.0
G-30-1781-en Page 15
Safety Measures S100 / OPMI® pico
Version 4.0
Page 16 G-30-1781-en
S100 / OPMI® pico Safety Measures
During use
• Reduce the brightness and duration of light exposure of the surgical area
to the absolute minimum level required to avoid burn injuries.
• Never directly look into the light source, e.g. a microscope object lens or
a light guide to avoid phototoxic damages to the eyes.
• Never leave a device unattended with the light source still switched on.
• Not approved or defective accessories for the device may lead to elevated
leakage currents. Never connect the non approved or defective accesso-
ries to the device.
• Never touch video interfaces while in contact with the patient.
• If you change the lamp shortly after it has failed, the lamp will still be very
hot. Use heat-resistant protective gloves when replacing the lamp to pre-
vent burns.
• If the xenon lamp is used beyond its maximum service life of 500 hours,
sudden failure may occur and interrupt operation. Change to the backup
lamp or replace the lamp module in a timely manner.
• If the halogen lamp is used beyond its maximum service life of 50 hours,
sudden failure may occur and interrupt operation. Please replace the hal-
ogen lamp in due time.
• Do not pull at the power cord or any other connecting cables.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the system
stating it is out of order and contact the Carl Zeiss service staff or the au-
thorized partner.
• Never operate the system unattended.
• Always switch off the system
- before disconnecting it from / connecting it to line power
- if it will not be used for a prolonged period of time
- or for cleaning purposes.
• Always use the master switch to turn off the device if it is not in use.
Version 4.0
G-30-1781-en Page 17
Safety Measures S100 / OPMI® pico
Loss of warranty!
Modifications to this system are not permissible. The manufacturer is notli-
able for damages caused by an unauthorized use of the device. Furthermore,
this will forfeit any rights to claim under warranty.
Version 4.0
Page 18 G-30-1781-en
S100 / OPMI® pico Safety Measures
Version 4.0
G-30-1781-en Page 19
Safety Measures S100 / OPMI® pico
General
This device is optionally equipped with a powerful xenon light source.
Various factors contribute to the risk of burn injuries:
Device-related factors:
– With increasing service life of the lamp, the actual light intensity delivered
at the respective setting decreases. In case of a bulb exchange, illumina-
tion increases to the originally high value.
Surgery-related factors:
– The selected intensity of the light source is a major factor for the risk of
injury. It should always be set to the minimum required for the surgical
procedure to be performed.
– A long surgical procedure increases the risk of injury, in particular if a
standard procedure takes considerably longer than usual.
– Injuries in the peripheral area may be avoided by covering the respective
parts with wet, sterile gauze.
The gauze must be moistened at regular intervals to avoid that the area
dries out or heats up.
– Certain preparations of the surgical field, local vasoconstrictive medica-
tions and incision drapes may also result in a higher risk of injury (drapes
may heat up to varying degrees depending on their color and moisture
content).
– You should also take into account that some areas of the body may be
more sensitive than others.
Patient-related factors:
– The general condition of a patient's health may contribute to the risk of
injury.
Version 4.0
Page 20 G-30-1781-en
S100 / OPMI® pico Safety Measures
– The skin type may also play a major role for the risk of injury.
– Certain medications also affect the sensitivity to light.
– The interaction of heat and antimicrobial substances in incision foils may
lead to an increased reaction of the patient to these substances.
Recommendations
Due to the large number of different factors involved and the lack of scientific
publications on this topic, Carl Zeiss cannot provide guidance on acceptable
illumination intensities and exposure durations. However, this system has sev-
eral features that can help the user to reduce the risk of burns.
– The start value of the illumination intensity can be set to a low value.
– The risk of burns can be reduced by constantly by irrigating the illuminated
surgical field and by keeping it moist.
– If used, drapes should also be re-moistened at regular intervals in order to
prevent heat from accumulating underneath the drape.
– Switch off the light when the microscope is not used, and make sure that
it is not pointed at unprotected bare skin.
Please keep in mind that the most published cases with burns refer to the skin
around the incision. The most important measure to avoid burns is covering
the peripheral area with wet, sterile gauze. The area of the incision should be
constantly moistened.
Final remark
Carl Zeiss recommends to reduce illumination of the surgical field in that pos-
sible extend to ensure the patient's security and a good microscopic image.
Please refer to the warnings and precautions in this case.
Version 4.0
G-30-1781-en Page 21
Safety Measures S100 / OPMI® pico
Safety devices
Version 4.0
Page 22 G-30-1781-en
S100 / OPMI® pico Safety Measures
Version 4.0
G-30-1781-en Page 23
Safety Measures S100 / OPMI® pico
Microscope labels
1 Friction symbol
Explanation of the braking effect that is achieved by turning the friction
control.
- Turn left: Reducing the braking effect
- Turn right: Increasing the braking effect
2 Marking the magnification changer
Marks the magnification setting.
1
1.6 3 Labeling magnification changer
Displays the magnification value
2.5
0.4
0.6
1
Version 4.0
Page 24 G-30-1781-en
S100 / OPMI® pico Safety Measures
5 6
5
1 6
2
3
11
5 6
1
5
6
2
3
Version 4.0
G-30-1781-en Page 25
Safety Measures S100 / OPMI® pico
1 Friction symbol
Explanation of the braking effect that is achieved by turning the friction
control.
- Turn left: Reducing the braking effect
- Turn right: Increasing the braking effect
2 "Risk of crushing" warning label
Fingers may be crushed. Do not touch this area while moving the surgical
microscope or parts of it.
3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness
Version 4.0
Page 26 G-30-1781-en
S100 / OPMI® pico Safety Measures
2 1 2 3
4
2 5 1
2
Version 4.0
G-30-1781-en Page 27
Safety Measures S100 / OPMI® pico
6 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact data,
i.e. phone number, fax number and email address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your system. A product file
is maintained by Carl Zeiss for this SIP number.
7 System identification label
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
8 Year of manufacture
This label indicates the year of manufacture of the device.
Version 4.0
Page 28 G-30-1781-en
S100 / OPMI® pico Safety Measures
6
7
8
Version 4.0
G-30-1781-en Page 29
Safety Measures S100 / OPMI® pico
Version 4.0
Page 30 G-30-1781-en
S100 / OPMI® pico Safety Measures
10
11
12
12
Version 4.0
G-30-1781-en Page 31
Safety Measures S100 / OPMI® pico
1 Friction symbol
Explanation of the braking effect that is achieved by turning the friction
control.
- Turn left: Reducing the braking effect
- Turn right: Increasing the braking effect
2 "Risk of crushing" warning label
Fingers may be crushed. Do not touch this area while moving the surgical
microscope or parts of it.
3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness
Version 4.0
Page 32 G-30-1781-en
S100 / OPMI® pico Safety Measures
1 1
2 1 2 3
5
4
1
2
Version 4.0
G-30-1781-en Page 33
Safety Measures S100 / OPMI® pico
6 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact data,
i.e. phone number, fax number and email address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your system. A product file
is maintained by Carl Zeiss for this SIP number.
7 System identification label (S100 wall mount with long arm)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
8 Year of manufacture
This label indicates the year of manufacture of the device.
Version 4.0
Page 34 G-30-1781-en
S100 / OPMI® pico Safety Measures
7
8
9
Version 4.0
G-30-1781-en Page 35
Safety Measures S100 / OPMI® pico
1 Friction symbol
Explanation of the braking effect that is achieved by turning the friction
control.
- Turn left: Reducing the braking effect
- Turn right: Increasing the braking effect
2 "Risk of crushing" warning label
Fingers may be crushed. Do not touch this area while moving the surgical
microscope or parts of it.
3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness
Version 4.0
Page 36 G-30-1781-en
S100 / OPMI® pico Safety Measures
1 1
2
1 2 3
2
1 4
2
Version 4.0
G-30-1781-en Page 37
Safety Measures S100 / OPMI® pico
6 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Manufacturer's contact data,
i.e. phone number, fax number and email address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your system. A product file
is maintained by Carl Zeiss for this SIP number.
7 System identification label
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
8 Year of manufacture
This label indicates the year of manufacture of the device.
Version 4.0
Page 38 G-30-1781-en
S100 / OPMI® pico Safety Measures
7
8
9
Version 4.0
G-30-1781-en Page 39
Safety Measures S100 / OPMI® pico
3 Brightness setting
If the light source is turned on, the brightness can be continuously set by
turning the brightness control of the fiber illumination:
- Turn left: Minimum brightness
- Turn right: Maximum brightness
6 Year of manufacture
This label indicates the year of manufacture of the device.
Version 4.0
Page 40 G-30-1781-en
S100 / OPMI® pico Safety Measures
1 1
2
2 1 2 3
5
4
1
2
Version 4.0
G-30-1781-en Page 41
Safety Measures S100 / OPMI® pico
1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Contact possibilities to the device manufacturer,
i.e. phone number, fax number and e-mail address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device. A product file
is maintained by Carl Zeiss for this SIP number.
2 System identification label (Centro suspension system)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
Version 4.0
Page 42 G-30-1781-en
S100 / OPMI® pico Safety Measures
1
2
Version 4.0
G-30-1781-en Page 43
Safety Measures S100 / OPMI® pico
1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Contact possibilities to the device manufacturer,
i.e. phone number, fax number and e-mail address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device. A product file
is maintained by Carl Zeiss for this SIP number.
2 System identification label (S100 support unit)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CSA label
– CE label
3 System identification label
(S100 support unit for secondary light source)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / (OPMI pico)
– Reference number
– CE label
Version 4.0
Page 44 G-30-1781-en
S100 / OPMI® pico Safety Measures
1
2
1
3
Version 4.0
G-30-1781-en Page 45
Safety Measures S100 / OPMI® pico
1 SIP label
The SIP label contains the following information:
– Manufacturer (company name)
– Contact possibilities to the device manufacturer,
i.e. phone number, fax number and e-mail address of the local contact
of the national Carl Zeiss sales organization.
XXXX XXXXXX – SIP number
A unique identification number assigned to your device. A product file
is maintained by Carl Zeiss for this SIP number.
2 System identification label (S100 support unit 600)
The system identification label provides the following information:
– Manufacturing symbol
– Manufacturer (company name) Carl Zeiss Meditec AG
– Manufacturer's address Goeschwitzer Strasse 51-52
07745 Jena, Germany
– Serial number
– Project classification P30200001
– Device name S100 / OPMI pico
– Reference number
– CE label
Version 4.0
Page 46 G-30-1781-en
S100 / OPMI® pico Safety Measures
1
2
Version 4.0
G-30-1781-en Page 47
Safety Measures S100 / OPMI® pico
Version 4.0
Page 48 G-30-1781-en
S100 / OPMI® pico Safety Measures
Version 4.0
G-30-1781-en Page 49
Safety Measures S100 / OPMI® pico
3 Fuse label
This label contains the following information:
– Maximum voltage 115 V / 230 V
– Rated current 6.3 A-H / 3.15 A-H
– (Tripping) characteristics T
4 "Observe disposal regulations" label
Electrical or electronic devices must not be disposed of as normal domestic
waste. For more information on the disposal of electrical and electronic
devices, please see the chapter on "Maintenance and care."
Version 4.0
Page 50 G-30-1781-en
S100 / OPMI® pico Safety Measures
Version 4.0
G-30-1781-en Page 51
Safety Measures S100 / OPMI® pico
9 CAUTION label
Warning of a potential hazard
Version 4.0
Page 52 G-30-1781-en
S100 / OPMI® pico Safety Measures
6 7
9 8
Version 4.0
G-30-1781-en Page 53
Safety Measures S100 / OPMI® pico
Version 4.0
Page 54 G-30-1781-en
S100 / OPMI® pico Safety Measures
3 2
Version 4.0
G-30-1781-en Page 55
Safety Measures S100 / OPMI® pico
5 Warning label
The xenon light source shall not be used for ophthalmological surgeries.
• Make sure that no xenon light enters the patient's eyes.
6 "CAUTION" label
The lamp housing contains an integrated xenon light source.
7 Bulb exchange
The label graphically shows the bulb exchange in three steps.
A - Press the button
B - Pull out the lamp module
C - Change bulbs
Version 4.0
Page 56 G-30-1781-en
S100 / OPMI® pico Safety Measures
8 1 2
Version 4.0
G-30-1781-en Page 57
Safety Measures S100 / OPMI® pico
9 CAUTION label
Warning of a potential hazard
12 Rating "Xenon"
Version 4.0
Page 58 G-30-1781-en
S100 / OPMI® pico Safety Measures
8 9 10
11
Version 4.0
G-30-1781-en Page 59
Safety Measures S100 / OPMI® pico
Version 4.0
Page 60 G-30-1781-en
S100 / OPMI® pico Safety Measures
1 2
Version 4.0
G-30-1781-en Page 61
Safety Measures S100 / OPMI® pico
Version 4.0
Page 62 G-30-1781-en
S100 / OPMI® pico System Overview
System Overview
System overview..............................................................................64
OPMI pico on S100 floor stand..................................................................64
OPMI pico on S100 wall mount .................................................................66
OPMI pico on S100 ceiling mount .............................................................68
OPMI pico on the S100 Centro suspension system and
S100 support units ....................................................................................70
Version 4.0
G-30-1781-en Page 63
System Overview S100 / OPMI® pico
System overview
Surgical microscope The surgical microscope (4) enables the illumination and visual magnification
of the field of view. For visual magnification, the surgical microscope can be
equipped with different tubes, eyepieces and objective lenses.
Floor stand The floor stand is used to power and control the surgical microscope func-
tions. It consists of support arm (2), suspension arm (3), supply unit with light
source (1), stand column (5) and stand base (6). The suspension arm enables
rough positioning of the surgical microscope and moving it into its standby
position. The stand base is equipped with four steerable casters (7) which
make it easier for you to position the system in the OR. Arrestors at the casters
secure the system against accidental moving and rolling away.
Light source Depending on the order, the device may be equipped with a halogen or a
xenonlight source. The halogen or xenon light source is used to illuminate the
surgical area. It has a fiber illumination in the surgical microscope, i. e. the
light is directed from the light source via a light guide to the surgical micro-
scope. The halogen or xenon light source is situated in the lamp housing of
the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as a backup lamp which can be swung in in case the first lamp fails.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains two xenon lamps. The second lamp is used
as a backup lamp which can be swung in case the first lamp fails.
Version 4.0
Page 64 G-30-1781-en
S100 / OPMI® pico System Overview
1 2
5
7 6
Version 4.0
G-30-1781-en Page 65
System Overview S100 / OPMI® pico
Surgical microscope The surgical microscope (5) enables the illumination and visual magnification
of the field of view.
For visual magnification, the surgical microscope can be equipped with dif-
ferent tubes, eyepieces and objective lenses.
Wall mount The wall mount is used to power and control the surgical microscope func-
tions. It consists of support arm (3), suspension arm (4), supply unit with light
source (1) and wall flange (2). The wall mount is firmly mounted to the wall.
However, the support and suspension arms can be moved.
The suspension arm enables rough positioning of the surgical microscope and
moving it into its standby position.
Light source Depending on the order, the device may be equipped with a halogen or a
xenonlight source. The halogen or xenon light source is used to illuminate the
surgical area. It has a fiber illumination in the surgical microscope, i. e. the
light is directed from the light source via a light guide to the surgical micro-
scope. The halogen or xenon light source is situated in the lamp housing of
the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as a backup lamp which can be swung in in case the first lamp fails.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains two xenon lamps. The second lamp is used
as a backup lamp which can be swung in case the first lamp fails.
Version 4.0
Page 66 G-30-1781-en
S100 / OPMI® pico System Overview
Version 4.0
G-30-1781-en Page 67
System Overview S100 / OPMI® pico
Surgical microscope The surgical microscope (5) enables the illumination and visual magnification
of the field of view.
For visual magnification, the surgical microscope can be equipped with dif-
ferent tubes, eyepieces and objective lenses.
Ceiling mount The ceiling mount is used to power and control the surgical microscope func-
tions. It consists of support arm (3), suspension arm (4), supply unit with light
source (1) and ceiling flange with column (2). The ceiling mount is firmly
mounted to the ceiling. However, the support and suspension arms as well as
the illumination unit around the column can be moved.
The suspension arm (4) enables rough positioning of the surgical microscope
and moving it into its standby position.
Light source Depending on the order, the device may be equipped with a halogen or a
xenonlight source. The halogen or xenon light source is used to illuminate the
surgical area. It has a fiber illumination in the surgical microscope, i. e. the
light is directed from the light source via a light guide to the surgical micro-
scope. The halogen or xenon light source is situated in the lamp housing of
the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as a backup lamp which can be swung in in case the first lamp fails.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains two xenon lamps. The second lamp is used
as a backup lamp which can be swung in case the first lamp fails.
Version 4.0
Page 68 G-30-1781-en
S100 / OPMI® pico System Overview
2 3
Version 4.0
G-30-1781-en Page 69
System Overview S100 / OPMI® pico
Surgical microscope The surgical microscope (4) enables the illumination and visual magnification
of the field of view. For visual magnification, the surgical microscope can be
equipped with different tubes, eyepieces and objective lenses.
S100 Centro suspension sys- Centro suspension system and S100 support units are used to power and con-
tem and S100 support units trol the surgical microscope functions. It consists of support arm (2), suspen-
sion arm (3) and supply unit with light source (1) (optional).
The suspension arm (3) enables rough positioning of the surgical microscope
and moving it into its standby position.
The Centro suspension system and the S100 support units are installed by au-
thorized staff according to the installation instructions to your existing treat-
ment unit.
Light source Depending on the order, the device may be equipped with a halogen or a
xenonlight source. The halogen or xenon light source is used to illuminate the
surgical area. It has a fiber illumination in the surgical microscope, i. e. the
light is directed from the light source via a light guide to the surgical micro-
scope. The halogen or xenon light source is situated in the lamp housing of
the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as a backup lamp which can be swung in in case the first lamp fails.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion fil-
ters can therefore be used for photographic documentation.
The xenon light source contains two xenon lamps. The second lamp is used
as a backup lamp which can be swung in case the first lamp fails.
Version 4.0
Page 70 G-30-1781-en
S100 / OPMI® pico System Overview
1
3
Version 4.0
G-30-1781-en Page 71
System Overview S100 / OPMI® pico
OPMI pico
Version 4.0
Page 72 G-30-1781-en
S100 / OPMI® pico System Overview
9
8b
4 2
10
9
1
8a
2
4 3
4
5
4
6
Version 4.0
G-30-1781-en Page 73
System Overview S100 / OPMI® pico
Version 4.0
Page 74 G-30-1781-en
S100 / OPMI® pico System Overview
Fig. 31: Controls for the tubes 180° tiltable tube 45° inclined tube
1 2 1 2
Straight tube
1
Version 4.0
G-30-1781-en Page 75
System Overview S100 / OPMI® pico
2
3
Version 4.0
Page 76 G-30-1781-en
S100 / OPMI® pico System Overview
Version 4.0
G-30-1781-en Page 77
System Overview S100 / OPMI® pico
Use The MORA interface is intended exclusively for use on the OPMI pico surgical
microscope in dentistry.
Description The MORA interface increases the microscope's mobility about its lateral tilt
axis, without changing the user's upright, ergonomic seated posture.
This offers advantages for the treatment of relatively large fields of therapy,
e.g. when several teeth are being treated at once.
A manually operated annular diaphragm permits the depth of field to be
varied.
With maximum lateral tilt of the microscope body and high magnification,
slight vignetting of the field of view may occur.
Version 4.0
Page 78 G-30-1781-en
S100 / OPMI® pico System Overview
approx. ±25°
Version 4.0
G-30-1781-en Page 79
System Overview S100 / OPMI® pico
Version 4.0
Page 80 G-30-1781-en
S100 / OPMI® pico System Overview
1 2 3 4
Version 4.0
G-30-1781-en Page 81
System Overview S100 / OPMI® pico
Use The Endoport endoscope port is intended for use on OPMI pico microscopes
in ENT surgery.
Description The endoscope port of the OPMI pico Endoport is a standardized interface for
endoscope cameras. This port permits an external endoscope camera with
special adapter to be connected using the quick-action locking mechanism.
If you want to equip the surgical microscope with the Endoport accessory,
please observe the requirements for system expansions, see page 265.
Only attach external endoscope cameras including adapter with a maximum
weight of 1.0 kg to the endoscope port.
The endoscope port always produces an image at infinity; for this reason, the
external endoscope camera's focus must also be set at infinity. The procedure
of focusing the endoscope camera at infinity is a camera-specific procedure
and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect the
endoscope port.
Version 4.0
Page 82 G-30-1781-en
S100 / OPMI® pico System Overview
Version 4.0
G-30-1781-en Page 83
System Overview S100 / OPMI® pico
1 Support arm
2 Suspension arm
The suspension arm is provided with a switch. This switch turns on the
light source when the suspension arm swings into the working position.
The light source is switched off in the upper resting position.
3 Mobility adjustment screw
You are able to set the lifting mobility of the suspension arm with the ad-
justment screw after the successful balance setting of the microscope
equipment.
4 Mobility adjustment screw
Use this screw to set the mobility of the coupling's rotation.
5 Mobility adjustment screw
Use this screw to set the mobility of the surgical microscope's swing
movement.
6 120° coupling
The 120° coupling improves ergonomic handling. The use of other cou-
plings is also possible, please refer to order data (page 266).
7 Balance setting screw
8 Mobility adjustment screw
Use this screw to set to set the mobility of the suspension arm's rotation.
9 Mobility adjustment screw
Use this screw to set the friction of the carrier arm's rotation.
10 Supply unit with light source
The supply unit provides the connectors for the power cord, light guide,
camera and video as well as the power switch. The supply unit provides
the lamp housing with halogen or xenon light source.
11 Wall flange (wall mount only)
The wall mount differs from the ceiling mount by the type of mounting to
the building.
12 Ceiling flange (ceiling mount only)
The ceiling mount differs from the wall mount by the type of mounting to
the building.
13 Securing screws
Version 4.0
Page 84 G-30-1781-en
S100 / OPMI® pico System Overview
2 3 4 5
10 9
13
8 7 6
14
12 11
Version 4.0
G-30-1781-en Page 85
System Overview S100 / OPMI® pico
1 Air grid
Do not cover the air grid!
For example, drapes could be covering the grid. The light source may be
switched off by the built-in thermostatic switch.
2 Flap
The flap is the mechanical indicator for the operating status of the halogen
lamps.
– When the flap is closed, the main lamp is operative.
– The open flap indicates that the backup lamp is being used.
3 Manual switching to the backup lamp
By pressing this button, the backup lamp swings in (page 222)
4 Lamp module
5 Open lamp module
By pressing this button, the lamp module is partly ejected. Pull out the
lamp module all the way for exchanging the lamp.
6 Brightness setting
If the light source is switched on, the brightness can be continuously set
by turning the brightness control of the fiber illumination:
- Left position: Minimum brightness
- Right position: Maximum brightness
Version 4.0
Page 86 G-30-1781-en
S100 / OPMI® pico System Overview
3
4
5
Version 4.0
G-30-1781-en Page 87
System Overview S100 / OPMI® pico
Version 4.0
Page 88 G-30-1781-en
S100 / OPMI® pico System Overview
4
1
5
Version 4.0
G-30-1781-en Page 89
System Overview S100 / OPMI® pico
Version 4.0
Page 90 G-30-1781-en
S100 / OPMI® pico System Overview
2 4
3
Version 4.0
G-30-1781-en Page 91
System Overview S100 / OPMI® pico
Version 4.0
Page 92 G-30-1781-en
S100 / OPMI® pico System Overview
Version 4.0
G-30-1781-en Page 93
System Overview S100 / OPMI® pico
Depending on the integrated video camera selected, the supply unit contains
different video control units:
For the integrated SD video camera
– SD video control unit
MediLive video control unit
– SD video control unit with digital interface
MediLive video control unit digital
For the integrated HD video camera
– HD video 720p control unit
– HD video 1080p control unit
(full HD)
The area to the right of the dashed line is displayed with a 90° rotation. It is
located on the level of the connector panel.
1 config
The configuration menu permits you to set various functions to meet your
specific needs.
You access the configuration menu by pressing the "config" key.
The different setting options for the configuration of the user program are
described in the chapter "Preparations for use", see page 157.
The camera is delivered with a default configuration (factory settings).
2 Indicator light (green)
The indicator light (LED) is lit when the power switch has been turned on.
3 white bal
White balance (see page 154)
4 brightness
Use the , keys to adjust the brightness of the video image.The con-
figuration menu (page 155) gives you the choice between two different
modes: Automatic - Manual.
Version 4.0
Page 94 G-30-1781-en
S100 / OPMI® pico System Overview
1 2 3 4
1 2 3 4
Version 4.0
G-30-1781-en Page 95
System Overview S100 / OPMI® pico
5 Image
The <freeze> button allows you to freeze the current live video image and
to display it as a full-screen freeze frame on the video monitor. At the
same time, the text "freeze" is displayed in the left upper corner.
When you press the <live> key, the current live video image is displayed
again on the monitor. The text "freeze" disappears.
6 Record
Operation of the Image Capture functionality: Please refer to page 178.
(only SD video control unit with digital interface)
7 select cursor keys
These control keys permit you to navigate in the configuration menu
(page 157).
8 enter key
Press the <enter> key to confirm a selection or trigger a function. The dif-
ferent functions are described in the corresponding menu item of the con-
figuration menu.
In addition, textual information for the selected mode is displayed.
The control key <select> (7) and the key <enter> (8) are only active within
the sub-menus.
Version 4.0
Page 96 G-30-1781-en
S100 / OPMI® pico System Overview
5 6 7 8
5 7 8
Version 4.0
G-30-1781-en Page 97
System Overview S100 / OPMI® pico
Video connectors
9 Remote socket (MediLive Video Control Unit)
It is possible to connect a foot switch here (optional).
The foot switch can be used to execute the function "image".
10 Remote socket (MediLive Video Control Unit Digital)
It is possible to connect a foot switch here (optional).
You can configure the foot switch as described on page 168 .
11 Remote socket (HD)
It is possible to connect a foot switch here (optional).
You can configure the foot switch as described on page 203 .
12 Connector for the camera head
The camera installed in the surgical microscope is connected via the
system cable to the video unit.
Switch off the video unit before plugging in/unplugging the system cable
at the video unit. The camera may be damaged if the cable is plugged/un-
plugged while voltage is still being applied.
13 Video signal output Y/C
We recommend using video devices suitable for Y/C signals, if possible.
This standard provides higher video image quality than CVBS.
14 Video signal output BNC (VBS)
Suitable for video systems capable of processing FBAS/VBS signals, or if
video signals have to be transmitted over relatively long distances (e.g.
BNC lines already installed on site).
The connection to the HD video control unit enables the connection of a
SD suitable device.
We recommend: Connect each video device to be installed such as re-
corder, monitor, printer, etc. via a separate output and do not loop it via
several devices. In this case, only the terminating resistor (75 W) switch of
the connected video devices (if present) in the last device of the chain
must be set to the "ON" position.
Version 4.0
Page 98 G-30-1781-en
S100 / OPMI® pico System Overview
13
14
10 12
13
14
11
13
Version 4.0
G-30-1781-en Page 99
System Overview S100 / OPMI® pico
Version 4.0
Page 100 G-30-1781-en
S100 / OPMI® pico System Overview
16
15
15
17
18
19
Version 4.0
G-30-1781-en Page 101
System Overview S100 / OPMI® pico
Version 4.0
Page 102 G-30-1781-en
S100 / OPMI® pico System Overview
Version 4.0
G-30-1781-en Page 103
System Overview S100 / OPMI® pico
Version 4.0
Page 104 G-30-1781-en
S100 / OPMI® pico Preparations for use
Connections................................................................................... 124
Connecting the S100 floor stand .............................................................124
Connecting the integrated video camera .................................................126
Connect PC to SD video control unit with digital interface .......................130
Version 4.0
G-30-1781-en Page 105
Preparations for use S100 / OPMI® pico
Version 4.0
Page 106 G-30-1781-en
S100 / OPMI® pico Preparations for use
• Securing screw (2) contains a hex socket head screw (5) which is an addi-
tional securing screw. Firmly tighten hex socket head screw (5) using a
suitable hex tool.
• Insert shaft (6) in the mount of the coupling. Use a suitable tool (e.g. hex
key or screwdriver) to screw in and tighten securing screws (7) and (8).
• Securing screw (8) contains a hex socket head screw (9) which is an addi-
tional securing screw. Firmly tighten hex socket head screw (9) using a
suitable tool.
4 4
8 7
Version 4.0
G-30-1781-en Page 107
Preparations for use S100 / OPMI® pico
Version 4.0
Page 108 G-30-1781-en
S100 / OPMI® pico Preparations for use
4 4
8 7
Version 4.0
G-30-1781-en Page 109
Preparations for use S100 / OPMI® pico
1 2 3 4 5
Version 4.0
Page 110 G-30-1781-en
S100 / OPMI® pico Preparations for use
Four hand grips are available for the OPMI pico surgical microscope (see
page 74).
The hand grips can be positioned as required by reinserting (2) or pressing
them together (1).
• Select the hand grip suitable for your application.
• Mount the hand grip supplied with the instrument. Firmly tighten the
screw(s).
Version 4.0
G-30-1781-en Page 111
Preparations for use S100 / OPMI® pico
Make sure that the light guide is not stretched or bent when the micro-
scope is turned or tilted.
Version 4.0
Page 112 G-30-1781-en
S100 / OPMI® pico Preparations for use
5 6 7
approx. ± 25°
Version 4.0
G-30-1781-en Page 113
Preparations for use S100 / OPMI® pico
Requirement The MORA interface with lateral port for documentation accessories has
been mounted on the surgical microscope.
The method described below can also be used for other accessories.
The operating principle of the accessories is described in the relevant user
manuals.
Version 4.0
Page 114 G-30-1781-en
S100 / OPMI® pico Preparations for use
1 2 3 4
Version 4.0
G-30-1781-en Page 115
Preparations for use S100 / OPMI® pico
Prerequisites Your OPMI pico has been equipped with the "Endoport" accessory com-
ponent:
Your endoscope camera, including adapter, does not weigh more than
1.0 kg.
• Focus the endoscope camera at infinity.
The procedure of focusing the endoscope camera at infinity is a camera-
specific procedure and differs from camera type to camera type.
Microscope settings such as the focusing of the eyepieces do not affect
the endoscope port.
• Remove the cover from the endoscope port.
• Attach your endoscope camera to the endoscope port and allow the
quick-action lock to snap in.
• Connect the video cable of the endoscope camera to the endoscope sys-
tem. Please also observe the information contained in the user's manual
on the endoscope system.
• Adjust the orientation of the image on the video monitor. To achieve this,
turn the endoscope camera in the endoscope port as required.
Version 4.0
Page 116 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 117
Preparations for use S100 / OPMI® pico
Version 4.0
Page 118 G-30-1781-en
S100 / OPMI® pico Preparations for use
• Do not perform balance setting until the surgical microscope and all ac-
cessories have been mounted!
• The maximum load on the suspension arm must not exceed 7 kg.
• Loosen the friction adjustment screw (1) of the suspension arm.
• Insert an 8 mm hex key into opening (3) of joint cover (2).
• Use the hex key to turn balance setting screw (4) until balance has been
achieved.
• Check the balance setting in several positions.
• Use friction adjustment screw (1) to set the required friction of the up-
ward/downward movement of the suspension arm.
2
3
4
Version 4.0
G-30-1781-en Page 119
Preparations for use S100 / OPMI® pico
Notes on adjustment
– It is recommended to provide the interocular distance as well as the refrac-
tion values for the different users to adjust the surgical microscope more
quickly and to preset the values in the scope of the preparation.
– Do not wear multifocal or varifocal glasses during adjustment or during
the operation of the microscope.
Such glasses make a proper diopter adjustment ring setting impossible
and result in unsatisfactory imaging.
– Optimum image clarity in the eyepieces and at the optical port for docu-
mentation equipment (photo, video) can only be achieved through careful
focusing and precisely adjusting the binocular tube.
Interocular distance • Move the microscope to a vertical position above a flat object, e. g. a writ-
adjustment ten piece of paper.
• Adjust the distance of the eyepieces at the binocular tube to your interoc-
ular distance so that the two eyepiece images (object and edge of visual
field) merge into one.
Adjusting the eyepieces The following procedure must be performed separately for each eyepiece in
the order described.
• Set the microscope to minimum magnification.
• Set the diopter adjustment ring at the eyepiece to zero position.
• Look through the eyepiece and move the microscope body to focus the
image.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Then reset the microscope to minimum magnification without changing
the working distance.
• Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
• Repeat the entire procedure for the second eyepiece.
Version 4.0
Page 120 G-30-1781-en
S100 / OPMI® pico Preparations for use
Now, the microscope is adjusted in such a way that the image is sharply
defined over the complete magnification range without any renewed fo-
cusing during the magnification setting.
If this is not the case, repeat the procedure.
Adjusting the eyecups • Adjust the eyecups of the eyepieces in such a way that the entire visual
field can be seen.
– Viewing with eyeglasses: Rotate eyecups in.
– Viewing without eyeglasses: Rotate eyecups out.
Version 4.0
G-30-1781-en Page 121
Preparations for use S100 / OPMI® pico
Version 4.0
Page 122 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 123
Preparations for use S100 / OPMI® pico
Connections
• Make sure that the rated voltage indicated in the display window (3) cor-
responds to the rated voltage available on the site of installation.
• Make sure that the device has been turned off with the power switch be-
fore plugging in or unplugging the plugs at the power inlet socket (4).
• Make sure that power switch (1) is OFF.
• Plug the included power cord at the supply unit into the power inlet sock-
et (4).
Version 4.0
Page 124 G-30-1781-en
S100 / OPMI® pico Preparations for use
1 3 Xenon
4
5
Version 4.0
G-30-1781-en Page 125
Preparations for use S100 / OPMI® pico
Version 4.0
Page 126 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 127
Preparations for use S100 / OPMI® pico
Y/C Y/C
2 Y/C
VIDEO OUT 3 VIDEO IN
VIDEO OUT
Y/C
2 VIDEO OUT
2 FBAS
VIDEO OUT
The video signal is scaled down for SD devices at the HD video control unit at
the SD video output (FBAS VIDEO OUT).
Version 4.0
Page 128 G-30-1781-en
S100 / OPMI® pico Preparations for use
2 HDMI 2 HDMI
VIDEO OUT VIDEO OUT
2 YPbPr
VIDEO OUT
Version 4.0
G-30-1781-en Page 129
Preparations for use S100 / OPMI® pico
You can connect a PC to the SD video control unit with digital interface. Please
pay attention to the requirements for a system extension, refer to page 265.
The connecting cable is available as accessories, please refer to accessories.
Adapter and software are not included in our delivery package.
System requirements
PC with FireWire port
Processor > 3.2GHz
2 GB RAM
Video software
The 6 poles FireWire port (DV cable) is only for data transfer purposes and has
no power supply.
• Connect the PC to the corresponding DV cable using the FireWire inter-
face at the video unit (on the suspension system).
Version 4.0
Page 130 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 131
Preparations for use S100 / OPMI® pico
Instrument tray • Hold the instrument tray with one hand and unscrew the four locking
screws at the adapter (1) with a few turns using the hex key until the
adapter (1) can easily be moved.
• Position instrument tray (3) in the required height (max. 800 mm above
the floor) and uniformly tighten (2.1 Nm) all four locking screws of
adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.
Flat screen holder • Hold the flat screen holder with one hand and unscrew the four locking
screws at the adapter (1) with a few turns using the hex key until the
adapter (1) can easily be moved.
• Position the flat screen holder (2) in the required height (max. 1230 mm
above the floor) and uniformly tighten (2.1 Nm) all four locking screws of
adapter (1) using the hex key.
• When attaching adapter (1), make sure the two components of the
adapter (1) are separated by the same distance (4) from top to bottom.
Version 4.0
Page 132 G-30-1781-en
S100 / OPMI® pico Preparations for use
1
3 max.
1230 mm
max.
800 mm
1 4
4x
Version 4.0
G-30-1781-en Page 133
Preparations for use S100 / OPMI® pico
Version 4.0
Page 134 G-30-1781-en
S100 / OPMI® pico Preparations for use
1 1
4 5 6
Version 4.0
G-30-1781-en Page 135
Preparations for use S100 / OPMI® pico
Version 4.0
Page 136 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 137
Preparations for use S100 / OPMI® pico
After mounting the flat screen holder, you can adjust the friction of its joints
as required. The factory setting of the joints is suitable for a medium load ca-
pacity which meets the requirements of most users. If one joint is set too
heavy-going or too easy-going, it is possible to adjust friction.
Version 4.0
Page 138 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 139
Preparations for use S100 / OPMI® pico
1
2
Version 4.0
Page 140 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 141
Preparations for use S100 / OPMI® pico
Applying drapes
Version 4.0
Page 142 G-30-1781-en
S100 / OPMI® pico Preparations for use
Version 4.0
G-30-1781-en Page 143
Preparations for use S100 / OPMI® pico
Version 4.0
Page 144 G-30-1781-en
S100 / OPMI® pico Operation
Operation
Version 4.0
G-30-1781-en Page 145
Operation S100 / OPMI® pico
Prerequisites The correct rated voltage has been set on the suspension system (please
refer to page 124).
The included power cord (floor stand) is connected (please refer to
page 124).
The light guide is connected (please refer to page 112).
Proceed as follows to power the system up and down:
• Press the power switch (1) on the suspension system.
Before starting to work with the system, you have to perform a function test
(see next page). This permits you to check the system configuration and user-
specific settings and to reconfigure them, if necessary.
1
Version 4.0
Page 146 G-30-1781-en
S100 / OPMI® pico Operation
Version 4.0
G-30-1781-en Page 147
Operation S100 / OPMI® pico
The securing screws of the video system cable have been firmly tight-
ened.
The correct function of the other equipment has been checked with
the relevant user manual.
Version 4.0
Page 148 G-30-1781-en
S100 / OPMI® pico Operation
Main and backup lamp of the halogen light source are intact.
Main and backup lamp of the xenon light source are intact.
In case of exceeding the maximum service life of 500h of the main
lamp and in case of an operation of the backup lamp, have a backup
lamp module available as precaution. The order number can be found
in the device data.
It is possible to switch on or off the halogen or xenon light source by
moving up and down the suspension arm.
The light source is turned on within the preset working area.
The light source is turned off in the upper resting position.
Version 4.0
G-30-1781-en Page 149
Operation S100 / OPMI® pico
Typical procedure
Starting work
Version 4.0
Page 150 G-30-1781-en
S100 / OPMI® pico Operation
Ending work
Version 4.0
G-30-1781-en Page 151
Operation S100 / OPMI® pico
Menu structure
Test image
A color bar pattern is displayed if the camera head is not connected when you
switch on the system.
• Switch off the system again, connect the camera head and switch the sys-
tem back on.
enter key
Press this key to activate a selected submenu or an image in the thumbnail
index.
Confirm the selection of a menu option.
Version 4.0
Page 152 G-30-1781-en
S100 / OPMI® pico Operation
* Only available in connection with the SD video control unit with digital in-
terface (MediLive Video Control Unit Digital).
Version 4.0
G-30-1781-en Page 153
Operation S100 / OPMI® pico
During the white balancing procedure, the system adjusts the video signal in
such a way that white areas in the surgical field are also white on the video
image.
• Press the <start> key to activate the white balance procedure.
The following message is displayed:
"Please put a piece of white paper under the microscope"
• Place a suitable white object (e. g. a piece of white paper with a matte sur-
face) under the surgical microscope.
start
• Carry out the white balance procedure using a medium brightness level.
• Press the <start> key to start the white balance procedure.
During the white balance procedure (display: "White Balance in Progress")
the key functions are locked.
If the white balance procedure is successfully completed, the following will be
displayed: "White Balance OK". Otherwise, the following message is displayed
on the monitor: "White Balance Failed" Check the settings described above
and repeat the white balance procedure.
After the white balance procedure is completed, all keys are enabled
again. The determined values are stored and are available after every re-
start.
In case of the halogen light source the video image will be adapted to the
image in the surgical microscope (yellowness).
Version 4.0
Page 154 G-30-1781-en
S100 / OPMI® pico Operation
Setting brightness
Under brightness you can set the brightness of the video image between 0%
and 100% using the , buttons. The selected value is displayed at the
top left on the monitor.
The selected time disappears on the monitor if
or control key is not pressed for more than 5 seconds.
You can chose between two different modes in the configuration menu in the
brightness menu (page 160). Automatic - Manual.
Automatic In automatic mode, the camera automatically sets the brightness of the video
image to the value indicated on the brightness - , buttons. Exposure
control affects the exposure time of the camera.
In automatic mode, the target brightness is preset, the camera regulates ex-
posure time in a range between 1/50(60)s to 1/10000s around this target
value. The exposure time currently selected is not displayed on the monitor.
The target value can be set between 0 to 100%. Changes are made in 1% in-
crements.
Manual In the manual mode, the selected exposure time is displayed in percent (%) by
pressing the control keys. The setting range extends from 0% to 100%.
Changes are made in 1% increments.
The selected exposure time disappears again from the monitor if the key has
not been pressed for more than 5 seconds.
Brightness control notes
– In case of too less light, brightness control reaches its limits. The video im-
age is too dark. Increase the brightness of the surgical field illumination to
have sufficient light available.
– White balance will not be changed by setting the brightness of the video
image. Dark image areas, however, generally appear in paler colors.
Version 4.0
G-30-1781-en Page 155
Operation S100 / OPMI® pico
The <freeze> key allows you to freeze the current live video image and to dis-
play it as a full-screen freeze frame on the video monitor. At the same time,
the text "freeze" is displayed in the left upper corner.
When you press the <live> key, the current live video image is displayed again
on the monitor. The text "freeze" disappears.
In case the current live video image is displayed and the foot switch connected
to the "Remote" port is pressed, the live video image will be frozen and dis-
played as full-screen freeze frame on the monitor. This corresponds to the
"freeze" function. If you press the foot switch connected to the "Remote" port
while the freeze frame is being displayed, the live video image appears again
on the monitor. This corresponds to the "live" function.
You can configure the foot switch with the option "MediLive Video Control
Unit Digital" (page 170).
Version 4.0
Page 156 G-30-1781-en
S100 / OPMI® pico Operation
video You can set several modes for some functions in the sub menu by using the
the configuration menus. The user program is set to default values on delivery.
You can access the configuration menu by pressing the <config> key.
After pressing the <config> key, the video settings are displayed on the mon-
itor:
config Functions Sub menu
User: User 1 -5
Brightness Mode: Automatic - Manual
Lightmeter: Small Spot - Large Spot - Integral
Chroma: 0% -100%
Image Orientation: Standard - Flip - Mirror - Rotate
Factory Setting:
*Remote Control: Sound on/off - Freeze/Store
*Record Mode: Moving/Still - File Format
*Delete Mode: Confirm Deletion -
No Confirmation - Delete all
*Storage Space: 0
*Date and Time: yyyy-mm-dd/hh:min
*File Name:
Exit
The menus marked with * are only available for the SD video control unit with
digital interface.
Version 4.0
G-30-1781-en Page 157
Operation S100 / OPMI® pico
You can exit the configuration menu using the menu item "Exit" or by pressing
the <config> or <live> key.
• Use the control keys to select a function.
The selected function is displayed with an arrow.
• Use the control keys to change the settings.
If you need more information on a menu item, you can open the corre-
sponding sub menu by pressing the <enter> key. This sub menu provides
additional information on the setting options. The current setting is high-
lighted as the initial setting in this sub menu.
• You can exit the configuration menu by selecting menu item "Exit" or by
pressing the <config> or <live> key. After changes have been made, a
prompt as to adapt the settings is displayed before exiting the configura-
tion menus. Answer this with "YES" or "NO". The settings are individually
saved for each user.
Version 4.0
Page 158 G-30-1781-en
S100 / OPMI® pico Operation
• Proceed as follows to select a user and load and activate the respective
configuration:
• Press the <config> key to access the configuration menu.
• In the configuration menu, select the menu item "User" with the control
keys and press the <enter> key.
The following menu appears:
User Settings
User 1
User 2
User 3
User 4
User 5
Cancel
: Select item
Enter: Load User Settings
Version 4.0
G-30-1781-en Page 159
Operation S100 / OPMI® pico
Automatic In automatic mode, the camera automatically sets the brightness of the video
image to the value indicated on the brightness - , - keys. Exposure con-
trol affects the exposure time of the camera.
In automatic mode, the target brightness is preset, the camera regulates ex-
posure time in a range between 1/50(60)s to 1/10000s around this target
value. The exposure time currently selected is not displayed on the monitor.
The target value can be set between 0 to 100%. Changes are made in 1% in-
crements.
Manual In the manual mode, the selected exposure time is displayed in percent (%) by
pressing the control keys. The setting range extends from 0% to 100%.
Changes are made in 1% increments.
The selected exposure time disappears again from the monitor if the key has
not been pressed for more than 5 seconds. The values marked with * are only
example values.
Version 4.0
Page 160 G-30-1781-en
S100 / OPMI® pico Operation
Brightness Mode
Automatic mode 50%*
Manual mode 30 %*
Cancel changes
Version 4.0
G-30-1781-en Page 161
Operation S100 / OPMI® pico
Intergral: The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly lit.
large: The exposure is measured in an area in the image center. This measurement
is ideal for working with an eclipsed surgical field edge (resulting from a re-
duced illuminated field diameter).
small: The exposure is measured in a very small area in the image center. This meas-
urement is suitable for work in narrow channels when an extremely small illu-
minated field diameter is being used.
However, if the object of interest is not located in the image center, the small
setting usually does not provide the desired exposure result. In this case, select
either a larger area (integral or large) or use the manual exposure mode.
Version 4.0
Page 162 G-30-1781-en
S100 / OPMI® pico Operation
Proceed as follows to select a exposure measurement type and to load the re-
spective configuration:
• Press the <config> key to access the configuration menu.
• In the configuration menu, select the menu item "Lightmeter" with the
control keys and press the <enter> key.
The following menu appears
Lightmeter Mode
Small Spot
Large Spot
Integral
Cancel changes
Version 4.0
G-30-1781-en Page 163
Operation S100 / OPMI® pico
Chroma Adjust
60 % Intensity
Cancel
Version 4.0
Page 164 G-30-1781-en
S100 / OPMI® pico Operation
Image Orientation
Standard Standard
Flip Reflect image horizontally*
Mirror Reflect image vertically*
Rotate Rotate the image by 180°
Cancel changes
The text passages marked with * are used for information purposes only
and do not appear on the screen.
• Select the desired mode with the control keys.
The selected choice is displayed with an arrow.
• Press the <enter> key to save the changes.
The selection is saved and you return to the configuration menu.
• If you don't want to save the changes, you can go back to the configura-
tion menu via the menu item "Cancel changes". The monitor orientation
is set as it was when the sub menu was accessed.
Version 4.0
G-30-1781-en Page 165
Operation S100 / OPMI® pico
Factory setting
After the selection of this function, all settings of the selected user can be
reset to the factory settings.
• Press the "config" key to access the configuration menu.
• In the configuration menu, select the line "Factory settings" with the
control keys.
• After the selection of the line "Copy Factory Settings to User X" and by
pressing the <enter> key "Loading Factory settings please wait" will be
displayed.
All settings of the selected user are now reset to the factory settings.
You go back to the configuration menu after approx. 5 seconds or by
pressing any key.
Version 4.0
Page 166 G-30-1781-en
S100 / OPMI® pico Operation
Version 4.0
G-30-1781-en Page 167
Operation S100 / OPMI® pico
Version 4.0
Page 168 G-30-1781-en
S100 / OPMI® pico Operation
Remote Control
Remote Sound Sound on
Store Method Freeze/Store
Version 4.0
G-30-1781-en Page 169
Operation S100 / OPMI® pico
JPEG / TIFF Files You can save video images in the JPEG format or in the TIFF format. You need
more storage capacity in the TIFF format than in the JPEG format.
The default setting on delivery is storage in the JPEG format only.
Still / Moving Subject Do you want the system to store both fields of a video image or to store only
one field? Two modes are available.
Still Subject (High Res): In this mode, the system stores both fields of the video
signal for each video image. This mode provides the maximum resolution with
still images.
The default setting on delivery is "Still Subject (High Res)".
Moving Subject (Low Res): In this mode, the system stores only one field of
the video signal for each video image. The system writes the individual lines
twice below each other, thus replacing the missing field. This mode provides
the maximum resolution with moving images (compensates for the blur
caused by movement).
Even with freeze images, the system only records one field of the video signal,
computes the second field by doubling the lines and displays it on the video
monitor. The same applies to the storage process.
Version 4.0
Page 170 G-30-1781-en
S100 / OPMI® pico Operation
Record Mode
Moving/Still Still (High Res)
File Format JPEG
Cancel
Version 4.0
G-30-1781-en Page 171
Operation S100 / OPMI® pico
Delete all In this mode, the system deletes all memory contents of the USB stick.
After pressing the <enter> key, the system asks "Do you really want to de-
lete the memory?".
This inquiry can be answered with "Yes" or "No".
If the answer is "Yes", the message "Memory will be erased now" ap-
pears during the deletion procedure.
If the answer is "No", the message "Delete process abort" appears for approx.
2 seconds.
Version 4.0
Page 172 G-30-1781-en
S100 / OPMI® pico Operation
Delete mode
Confirm Delete
No Confirmation
Delete all
Cancel
Version 4.0
G-30-1781-en Page 173
Operation S100 / OPMI® pico
Enter: Execute
Version 4.0
Page 174 G-30-1781-en
S100 / OPMI® pico Operation
Version 4.0
G-30-1781-en Page 175
Operation S100 / OPMI® pico
1 2 3 4 5 6 7 8 9 0
q w a r t z u i s p
a s d f b p j k l +
y x c v b c m - _
OK Space Delete
The word "File Name" stands for the expected entry. The entered name is dis-
played in the input field under "File Name". The entered name is always sep-
arated by the " " character. If the maximum number of characters is
reached, the " " character will no more be displayed.
Version 4.0
Page 176 G-30-1781-en
S100 / OPMI® pico Operation
Use the cursor keys to select the characters. Each marked character is trans-
ferred one after another into the "File Name" entry field by pressing the
<enter> key.
The cursor keys have a repeat function. When the character at the top of a
column is reached, the cursor jumps to the character at the bottom of the
same column. The same applies to the character at the bottom and to the
right and left ends of a line.
• You can switch between upper case and lower case letters with the
" "key.
– Text is then written in upper case or lower case letters until you switch
over again.
– In case of a system restart, the upper case mode is automatically set.
• Always delete the last character of the entered file name with the <De-
lete> key of the virtual keypad.
• Go to the configuration menu with the <OK> key.
The entered file name is stored temporarily in this case.
It will be saved when you exit the configuration main menu, after an
appropriate inquiry. The file name is retained when you switch off the
system.
When you switch on the system, it starts with the file name saved last.
You can either save or delete the name.
Instead of a video image, the monitor displays the question:
– FileName: "Current Filename"“
– Preserve file name
– Delete filename
Preserve file name The current file name is retained, the system goes to the live mode.
Delete file name The name is deleted, the system goes to the live mode. If no filename is saved,
this inquiry menu will not be displayed.
You can always define again a new file name under "File Name" in the con-
figuration menu.
• Chose the corresponding answer with the cursor.
• Press the <enter> key to save and execute your selection.
Version 4.0
G-30-1781-en Page 177
Operation S100 / OPMI® pico
Index key(1)
You can go to the graphic table of contents with the <Index> key.
Saved images are displayed as graphic table of contents in the form of 12 dis-
tortion-free miniature images (thumbnails) per screen page.
Image name and, if available, expanded file name (please refer to the config-
uration menu: File Name) will be displayed below the image.
The thumbnails are arranged in the order of their generation. The most recent
thumbnail is shown at the bottom on the right. If fewer than 12 thumbnails
are available for a page, the other image positions will not be used.
Selecting thumbnails
• Use the control keys to select a thumbnail.
A blue frame marks each selected image.
• In case a thumbnail is selected, press the <enter> key.
The selected thumbnail is then displayed in the form of a full frame on
the video monitor.
• Select further images with the control keys .
When you have reached the first or last image, the note "Last Image"
is displayed.
You can go to the graphic table of contents by pressing the <enter>
key again.
You can exit the graphic table of contents and change to the live im-
age mode by pressing the <live> key.
If no images are available, the following error message appears: "No
Images - Check USB Stick".
Version 4.0
Page 178 G-30-1781-en
S100 / OPMI® pico Operation
Version 4.0
G-30-1781-en Page 179
Operation S100 / OPMI® pico
Version 4.0
Page 180 G-30-1781-en
S100 / OPMI® pico Operation
USB messages
Version 4.0
G-30-1781-en Page 181
Operation S100 / OPMI® pico
Delete images
• Select an image in the image overview by pressing the <delete> key.
– The image is immediately deleted or
– you have to confirm deletion with the <enter> key (please refer to De-
lete Mode page 172).
Rename images
Please refer to page 176.
Version 4.0
Page 182 G-30-1781-en
S100 / OPMI® pico Operation
Date
2007/07/25
15:37:56 Time
Patient 123
File Name
Version 4.0
G-30-1781-en Page 183
Operation S100 / OPMI® pico
Version 4.0
Page 184 G-30-1781-en
S100 / OPMI® pico Operation
Menu structure
Test image
If the camera head is disconnected when the system is operating, a color bar
is automatically displayed on the monitor. The following message is displayed
on the monitor: "CAMERA HEAD NOT CONNECTED". An active menu is im-
mediately closed when the camera head is unplugged. Changes made re-
cently are lost in the process. All keys, except for the Power button <on>, are
disabled. If the camera head is reconnected, the color bar and the error mes-
sage automatically disappear and the camera is ready for use again.
If a camera head that does not match is connected, a color bar and the fol-
lowing message are displayed on the monitor: "VERSION OF THE CAMERA
HEAD DOES NOT MATCH - THIS MAY CAUSE COLOR VARIATIONS OR
REDUCE FUNCTIONALITY - ENTER OK".
If an uncalibrated camera head is connected, the following message are dis-
played on the monitor: "MISSING OR CORRUPT COLOR CALIBRATION.
THIS MAY CAUSE COLOR VARIATIONS OR REDUCE PICTURE
QUALITY".
All keys, except for the Power on button, are disabled. If a matching camera
head is reconnected, the color bar and the error message automatically dis-
appear and the camera is ready for use again.
Menu display
– The menu text is displayed on the monitor in light green.
A light green arrow is displayed in front of a selected line.
Version 4.0
G-30-1781-en Page 185
Operation S100 / OPMI® pico
Version 4.0
Page 186 G-30-1781-en
S100 / OPMI® pico Operation
Version 4.0
G-30-1781-en Page 187
Operation S100 / OPMI® pico
During the initial start of the surgical microscope with integrated HD video
camera, the sub menu "APPLICATION VARIANT" will be displayed for the first
time (please refer to page 214).
The application is set in this menu.
Carl Zeiss Service adjusts this setting for you during initial start-up. You can
readjust this setting in the "INSTALLATION MENU". see “Opening the instal-
lation menu (config)” on page 206.
The configuration menu allows you to make user-specific settings, such as se-
lecting a user or setting the brightness / color saturation.
• Press the <config> key to access the configuration menu.
The following message is shown on the display:
USER User 1 - 4
BRIGHTNESS AUTOMATIC - LOW AUTO -
MANUAL
LIGHTMETER SMALL SPOT - LARGE SPOT -
INTEGRAL
ORIENTATION STANDARD - MIRROR - FLIP -
ROTATE
ADVANCED ...
SERVICE MENU ...
EXIT
Version 4.0
Page 188 G-30-1781-en
S100 / OPMI® pico Operation
• You can exit the configuration menu by selecting menu item "Exit" or by
pressing the <config> or <live> key. If changes have been made, a prompt
is displayed for each user on exit from the configuration menu as to
whether the settings should be saved. Answer this prompt with "YES" or
"NO".
An existing "freeze" condition is deleted in case of an activation of the
configuration menu.
Version 4.0
G-30-1781-en Page 189
Operation S100 / OPMI® pico
User submenu
You can select among four different user profiles in the "USER SELECTION"
submenu: USER 1, USER 2, USER 3, USER 4
The user names and their configurations depend on the selected application.
The application is freely selectable on the first startup of the system, and can
subsequently be changed at any time in the installation menu.
– The "OPHTHALMOLOGY" application features user "POSTERIOR" instead
of USER 1 and user "ANTERIOR" instead of USER 2.
– Instead of USER 1, the "DENTRISTRY" application has user
"ENDODONTY" and user "TREATMENT" instead of USER 2.
– The Anterior, Posterior, Endodonty and Treatment user have optimized
settings.
– The video image is displayed in its unaltered condition (standard) for the
Anterior user.
– The video image is rotated 180°(ROTATE) for the Posterior user.
– The video image is displayed in its unaltered condition (standard) for the
Endodonty user.
– The video image is displayed in its unaltered condition (standard) for the
Treatment user.
The white balance procedure is user-specific. This allows User 1, for example,
to save a different white balance than User 2. If the user is changed, the white
balance values are loaded for the new user. If the light source selection is not
changed at the same time, discoloration will occur. Therefore perform the
white balance procedure for each envisaged user using the light source to be
used for the respective user.
Version 4.0
Page 190 G-30-1781-en
S100 / OPMI® pico Operation
Proceed as follows to select a user and load and activate the relevant config-
uration:
• Press the "config" key to access the configuration menu.
• In the configuration menu, use the control keys to select menu item
"USER" and press the "enter" key.
The following menu appears:
USER SELECTION
USER 1
USER 2
USER 3
USER 4
CANCEL
Version 4.0
G-30-1781-en Page 191
Operation S100 / OPMI® pico
AUTOMATIC In this mode, the camera automatically controls the brightness of the video
image to maintain the value selected using the , <brightness> keys. Ex-
posure control affects the exposure time of the camera.
The nominal brightness is set in the AUTOMATIC mode. The camera controls
the exposure time in the range between 1/50(60)s and 1/10000s. The expo-
sure time currently selected is not displayed on the monitor. The nominal
brightness can be adjusted between 0 and 100%. Changes are made in 5%
increments.
A nominal value in the range between 50% and 70% usually provides op-
timum exposure results.
LOW AUTO This mode has the same function as AUTOMATIC mode, however, exposure
times longer than 1/50(60)s are allowed, which can result in blurred move-
ments.
MANUAL In this mode, the selected exposure time is displayed by pressing the con-
trol keys. The setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Large exposure times may result in blurred movements.
The selected exposure time disappears again from the monitor if the <bright-
ness> key or has not been pressed for more than 2 seconds.
Version 4.0
Page 192 G-30-1781-en
S100 / OPMI® pico Operation
Proceed as follows to select a light control and to load the respective config-
uration:
• Press the <config> key to access the configuration menu.
• Select the menu item "BRIGHTNESS" in the configuration menu using the
control keys and press the <enter> key.
The following menu appears:
BRIGHTNESS MODE
AUTOMATIC 80%*
LOW LIGHT AUTO 80%*
MANUAL 1/16*
CANCEL
Version 4.0
G-30-1781-en Page 193
Operation S100 / OPMI® pico
INTEGRAL: The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly lit.
LARGE SPOT: The exposure is measured in an area in the image center. This measurement
is ideal for working with an eclipsed surgical field edge (resulting from a re-
duced illuminated field diameter).
SMALL SPOT: The exposure is measured in a very small area in the image center. This meas-
urement is suitable for work in narrow channels when an extremely small illu-
minated field diameter is being used.
However, if the object of interest is not located in the image center, the
SMALL SPOT setting usually does not provide the desired exposure result. In
this case, select either a larger area (INTEGRAL or LARGE SPOT) or use the
manual exposure mode.
Version 4.0
Page 194 G-30-1781-en
S100 / OPMI® pico Operation
Proceed as follows to select a exposure measurement type and to load the re-
spective configuration:
• Press the <config> key to access the configuration menu.
• Select the menu item "LIGHTMETER" in the configuration menu using the
control keys and press the <enter> key.
The following menu appears:
LIGHTMETER
SMALL SPOT
LARGE SPOT
INTEGRAL
CANCEL
Version 4.0
G-30-1781-en Page 195
Operation S100 / OPMI® pico
Orientation submenu
The "ORIENTATION" submenu allows setting the screen orientation. To do
this, proceed as follows:
• Press the "config" key to access the configuration menu.
• In the configuration menu, use the control keys to select menu item
"ORIENTATION" and press the "enter" key.
The following menu appears:
ORIENTATION
STANDARD Standard
MIRROR Mirror the image vertically*
FLIP Mirror the image horizontally*
ROTATE Turn the image through 180°*
CANCEL
The comments marked with * are only provided for information purposes
and do not appear on the monitor.
• Use the control keys to select the required mode.
The selection is displayed with an arrow.
• Press the "enter" key to save the changes.
The selection is saved and you return to the configuration menu.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the configuration menu. As a result, the image orientation on the
screen will be reset to the setting present when the submenu was opened.
Version 4.0
Page 196 G-30-1781-en
S100 / OPMI® pico Operation
Advanced submenu
In the "ADVANCED" submenu, you can set the settings that are seldom used.
To change the settings, proceed as follows:
• Press the "config" key to access the configuration menu.
• Select menu item "ADVANCED" in the configuration menu using the
control keys and press the "enter" key.
The following menu appears:
RED ±0
BLUE ±0
CHROMA ±0
COLOR ANGLE ±0
BLACK LEVEL ±0
ZOOM LEVEL 1.0 x
PEAK/AVERAGE 0
BACK
Version 4.0
G-30-1781-en Page 197
Operation S100 / OPMI® pico
CHROMA The "Chroma" setting increases or reduces the color saturation of the camera.
COLOR ANGLE The "Color Angle" (or HUE) color setting can be used to change the color
angle and the hues of the camera.
BLACK LEVEL The "Black Level" setting allows setting the black level of the camera. If the
black level is reduced, the contrast will be higher. If the black level is in-
creased, dark details can be seen more clearly.
ZOOM LEVEL The "Zoom Level" setting allows enlarging the camera image electronically.
Version 4.0
Page 198 G-30-1781-en
S100 / OPMI® pico Operation
PEAK / AVERAGE The "Peak / Average" setting allows modifying the exposure control charac-
teristics. If a low value is set, exposure control will be based on peak values
and react precisely to changes in the object brightness. If a higher value is set,
exposure control will be based on average values and react less sensible to
changed in brightness.
RED submenu In this submenu, you can set the red value of the camera in a range from -20
to +20. To do this, proceed as follows:
• Select menu item "RED" in the "ADVANCED" menu using the control
keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
The selected choice is displayed with an arrow.
• Change the red value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The red value is set as it was when the
submenu was accessed.
The red value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on page 216.
BLUE submenu In this submenu, you can set the blue value of the camera in a range from
-20 to +20. To do this, proceed as follows:
• Select the "BLUE" menu item in the "USER MENU ADVANCED" menu us-
ing the control keys and press the "enter" key.
• Select the "SETTINGS" menu item with the control keys.
The selected choice is displayed with an arrow.
• Change the blue value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
• If you do not want to save your changes, return to the "ADVANCED USER
SETTINGS" menu via "CANCEL". The blue value is set as it was when the
submenu was accessed.
Version 4.0
G-30-1781-en Page 199
Operation S100 / OPMI® pico
The blue value setting also influences the camera's white balance. see “Per-
forming the white balance (white bal)” on page 216.
CHROMA submenu In this submenu, you can set the color saturation of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "CHROMA" menu item in the "USER MENU ADVANCED" menu
using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the color saturation using the control keys.
A chroma setting in the range of ± 0 usually provides the best color im-
pression. Chroma setting is most effective with strongly saturated (bright)
colors. The effect is low with weakly saturated colors.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, color saturation will be reset to the
setting present when the submenu was opened.
COLOR ANGLE submenu In this submenu, you can set the color angle (hue) of the camera in a range
from -10 to +10. To do this, proceed as follows:
• Select the "COLOR ANGLE" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the color angle using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". As a result, the color angle will be reset to the
setting present when the submenu was opened.
BLACK LEVEL submenu This submenu allows setting the black level of the camera in a range from -10
to +10. To do this, proceed as follows:
Version 4.0
Page 200 G-30-1781-en
S100 / OPMI® pico Operation
• Select the "BLACK LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the black level value using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The black level is set as it was when the submenu
was accessed.
ZOOM LEVEL submenu In this submenu, you can set the zoom level of the camera in a range from
1.0x to 2.0x. To do this, proceed as follows:
• Select the "ZOOM LEVEL" menu item in the "USER MENU ADVANCED"
menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the zoom level using the control keys.
• Press the "enter" key to save the changes.
The selection is saved and you return to the "ADVANCED USER SET-
TINGS" menu.
If you do not want to save your changes, return to the "ADVANCED USER SET-
TINGS" menu via "CANCEL". The zoom level is set as it was when the sub-
menu was accessed.
PEAK / AVERAGE submenu In this submenu, you can set the exposure control characteristics of the
camera in a range from 0 to +8. To do this, proceed as follows:
• Select menu item "PEAK / AVERAGE" from the "ADVANCED USER SET-
TINGS" menu using the control keys and press the "enter" key.
• Use the control keys to select "SETTINGS".
The selection is displayed with an arrow.
• Change the value using the control keys.
Version 4.0
G-30-1781-en Page 201
Operation S100 / OPMI® pico
Version 4.0
Page 202 G-30-1781-en
S100 / OPMI® pico Operation
Service submenu
The "SERVICE MENU" submenu allows you to make settings which are re-
quired less frequently. To change the settings, proceed as follows:
• Press the "config" key to access the configuration menu.
• In the configuration menu, use the control keys to select menu item
"SERVICE MENU" and press the "enter" key.
The following menu appears:
SERVICE MENU
BACK TO CONFIG
REMOTE SETUP You can configure the device in this sub menu so that a selected function
(please refer to REMOTE FUNCTION) is controllable via a foot switch or a
fundus imaging system. The sub menu contains the following areas:
– REMOTE FUNCTION
– SWITCHING MODE
– AUDIO CONFIRMATION
You can assign one of the functions described below to the integrated HD
video camera under "REMOTE FUNCTION". This function is then triggered
when the foot switch is activated or by swiveling in and out of a fundus im-
aging system.
The following options are available:
Version 4.0
G-30-1781-en Page 203
Operation S100 / OPMI® pico
– FREEZE / LIVE
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
– USER1 / USER2
(or Endodonty/Treatment or Anterior/Posterior)
You can switch between user 1 and user 2 by pressing the foot switch. If
the current user is whether user 1 nor user 2, user 1 is switched on with
the first activation of the foot switch.
– USER3 / USER4
You can switch between user 3 and user 4 by pressing the foot switch. If
the current user is whether user 3 nor user 4, user 1 is switched on with
the first activation of the foot switch.
– INVERTED / STANDARD (is only required with a special tube)
If the current live video image is displayed and the foot switch is pressed,
the live video image is frozen and displayed on the monitor as a freeze
frame. With renewed pressing of the foot switch the live video image is
displayed again.
You can select under "SWITCHING MODE" if a remote function shall be con-
trolled via a foot switch or a fundus imaging system. The following control
types are available:
– FOOTSWITCH / AUX
Function control is performed via a foot switch.
– FUNDUS SYSTEM
The function is controlled via a fundus imaging system. This setting is only
available under the "OPHTHALMOLOGY" application.
FACTORY SETTINGS This submenu allows resetting the user and camera settings to the factory set-
tings. To do this, proceed as follows:
• In the "SERVICE ", use the control keys to select menu item "FACTORY
SETTINGS" and press the "enter" key.
• Use the control keys to select one of the following items:
Version 4.0
Page 204 G-30-1781-en
S100 / OPMI® pico Operation
– RESET USER 1
– RESET ALL USERS
– RESET CAMERA
After selection of one of the menu items, a message informs you that
the stored data will be overwritten.
• Press the "enter" key to overwrite the settings.
The system returns to the SERVICE MENU after approx. 2 seconds.
• If you do not want to save any changes, select menu item "CANCEL" to
return to the SERVICE MENU.
VERSION INFORMATION This submenu shows the current versions of the hardware and software com-
ponents as well as the serial numbers of the camera and CCU. Proceed as fol-
lows:
• Select menu item "VERSION INFORMATION" and press the "enter" key.
The following information is displayed:
– SOFTWARE (software version of the CCU)
– S/N CCU (serial number of the CCU)
• Press the "enter" key to return to the SERVICE MENU.
Version 4.0
G-30-1781-en Page 205
Operation S100 / OPMI® pico
INSTALLATION MENU
Version 4.0
Page 206 G-30-1781-en
S100 / OPMI® pico Operation
720P Select this setting format if your display device is 720p compatible. This mode
is similar to the 1080p progressive standard, but its maximum resolution is
only 1280 x 720 pixels.
1080P* Select this setting if your display device is 1080p compatible. The image is set
up using real full frames. This prevents any interline flicker, and the image is
sharply defined and stable. The maximum resolution is 1920 x 1080 pixels.
1080I* Use the 1080i (interlace) mode if your display device supports the 1080i
standard. The image is set up by two consecutive half images. The delayed
image setup can cause slight interline flicker. The maximum resolution is 1920
x 1080 pixels.
As a requirement, you should have connected the HD compatible OPMI pico
with a 1080p version and a 1080p or 1080i compatible display device. We
recommend the use of a 1080p compatible display device.
Version 4.0
G-30-1781-en Page 207
Operation S100 / OPMI® pico
HDTV MODE
720P
1080P*
1080I*
CANCEL
Version 4.0
Page 208 G-30-1781-en
S100 / OPMI® pico Operation
When you chose "NO", the camera resolution is not changed and you
go back to the installation menu.
• If you do not want to save your changes, return to the installation menu
via "CANCEL". The video signal is set as it was when the sub menu was
accessed.
Version 4.0
G-30-1781-en Page 209
Operation S100 / OPMI® pico
CANCEL
Version 4.0
Page 210 G-30-1781-en
S100 / OPMI® pico Operation
LETTERBOX Use this setting for display devices that are not compatible with the 16:9 wide
screen format. A black bar will appear at the top and bottom of the screen
and the image will be displayed in the 16:9 format.
SIDE-CUT In this mode, the sides of the image will be "cut", leaving the center of the
image to be displayed with a 4:3 aspect ratio.
Proceed as follows to select a format and to load the respective configuration:
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "SDTV FORMAT MODE" in the installation menu using
the control keys and press the "enter" key.
The following menu appears:
SDTV FORMAT
LETTERBOX
SIDE-CUT
CANCEL
Version 4.0
G-30-1781-en Page 211
Operation S100 / OPMI® pico
PAL
NTSC
CANCEL
Version 4.0
Page 212 G-30-1781-en
S100 / OPMI® pico Operation
LIGHT SOURCE
In the "LIGHT SOURCE" submenu, you can select between LED / XENON and
HALOGEN.
• Press the "config" key for approx. 5 seconds to access the installation
menu.
• Select menu item "LIGHT SOURCE" from the installation menu using the
control keys and press the "enter" key.
The following menu appears:
LIGHT SOURCE
XENON / LED
HALOGEN
CANCEL
Version 4.0
G-30-1781-en Page 213
Operation S100 / OPMI® pico
OPHTHALMOLOGY This application is designed for displaying the image in the ophthalmologic
area. The "ANTERIOR" and "POSTERIOR" users are available for displaying the
best image in the anterior or posterior eye segment. Both users have fixed set-
tings for the brightness control. This is why the "BRIGHTNESS" and "LIGHT-
METER" menu items are not available to these users.
DENTISTRY This application is designed for displaying the best image in the lower or upper
tooth segment. The "ENDODONTY" and "TREATMENT" users have optimized
and defined settings for the brightness control. This is why the "BRIGHTNESS"
and "LIGHTMETER" menu items are not available to these users.
NEURO/SPINE This application is designed for displaying the image in the neurological area.
The difference refers to the camera settings as well as to the not selectable
image orientation.
ENT This application id for the imaging of the neck, nose, ear area. The differences
can be found in regard of the camera settings and the adjusted image section.
Version 4.0
Page 214 G-30-1781-en
S100 / OPMI® pico Operation
APPLICATION VARIANT
OPHTHALMOLOGY
DENTISTRY
NEURO/SPINE
ENT
CANCEL
Version 4.0
G-30-1781-en Page 215
Operation S100 / OPMI® pico
In the white balancing procedure, the system adjusts the video signal in such
a way that white areas in the surgical field are also white on the video image.
• Press the <start> key to activate the white balance procedure.
The following message is displayed:
"PLEASE PUT A PIECE OF WHITE PAPER UNDER THE MICROSCOPE"
• Place a suitable white object (e. g., a piece of white paper with a matte
surface) under the surgical microscope.
• Carry out the white balance procedure using a medium brightness level.
• Press the <start> key to start the white balance procedure.
During the white balance procedure (display: "WHITE BALANCE IN
PROGRESS") the key functions are locked.
If the white balance procedure is successfully completed, the following will be
displayed: "WHITE BALANCE SUCCESSFUL". Otherwise, the following mes-
sage is displayed on the monitor: "WHITE BALANCE FAILED". Check the set-
tings described above and repeat the white balance procedure.
The white balance is saved for the user currently selected. When the user is
changed, the white balance values pertaining to that user are loaded.
After the white balance procedure is completed, all keys are enabled again.
The determined values are saved and are available to all users each time the
device is restarted.
White balance is adjusted corresponding to the set light type. The adjustment
depends on the "LIGHT SOURCE" setting in the "INSTALLATION MENU" and
the BLUE and RED settings in the "USER MENU ADVANCED".
Version 4.0
Page 216 G-30-1781-en
S100 / OPMI® pico Operation
The RED and BLUE values have an influence on the camera colors even after
the white balance procedure.
Recommended settings for a halogen light source
"LIGHT SOURCE" - "HALOGEN"
RED BLUE
Adjustment for a weak to medium illumi- +4 -8
nation (default)
Adjustment for a bright halogen illumina- +2 -4
tion
Without halogen light adjustment 0 0
RED BLUE
No adjustment needed (every illumina- 0 0
tion level)
Version 4.0
G-30-1781-en Page 217
Operation S100 / OPMI® pico
Setting brightness
Under Brightness, you can continuously adjust the video image brightness in
5% increments between 0% and 100% using the and keys. The se-
lected value is displayed in the left upper corner of the monitor.
In the configuration menu, you can select between three different modes in
the brightness menu: AUTO - LOW AUTO - MANUAL.
AUTO
In AUTO mode, the camera automatically controls the brightness of the video
image to maintain the value selected using the and control keys. Expo-
sure control affects the exposure time of the camera.
The nominal brightness is set in the AUTO mode. The camera controls the ex-
posure time in the range between 1/50(60)s and 1/10000s accordingly. The
nominal brightness can be adjusted between 0 and 100%. Changes are made
in 5% increments.
A nominal value in the range between 50% and 70% usually provides op-
timum exposure results.
LOW AUTO
This has the same function as AUTO Mode, however, exposure times longer
than 1/50(60)s are allowed, which can result in blurred movements.
MANUAL
In MANUAL mode, the selected exposure time is displayed by pressing the
control keys. The setting range extends from 1/6 to
1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500, 1/250, 1/
120, 1/60, 1/30, 1/15 and 1/8
Large exposure times may result in blurred movements.
The selected exposure time disappears again from the monitor if the control
keys or have not been pressed for more than 2 seconds.
Version 4.0
Page 218 G-30-1781-en
S100 / OPMI® pico Operation
The "freeze" key allows freezing the current live video image and displaying it
as a full-screen freeze frame on the video monitor. During this time, the text
"freeze" is displayed in the upper left corner.
When you press the "live" key, the current live video image is displayed again
on the monitor, and the text "freeze" disappears.
If the current live video image is displayed and the (foot switch) key connected
to the remote port is pressed and FREEZE/LIVE is set in the CONFIG MENU
under SERVICE MENU - REMOTE SETUP - REMOTE FUNCTION, the live video
image is frozen and displayed in freeze frame on the monitor. This corre-
sponds to the "freeze" function. If you press the connected switch (foot con-
trol panel) once again while the freeze frame is displayed, the live video image
appears again on the monitor. This corresponds to the "live"function.
Version 4.0
G-30-1781-en Page 219
Operation S100 / OPMI® pico
Version 4.0
Page 220 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
Troubleshooting ............................................................................226
For your safety ........................................................................................226
Malfunctions of the system......................................................................226
Malfunctions of S100 suspension systems with video (option) .................228
Malfunctions of S100 suspension systems with video and image storage op-
tion .........................................................................................................231
Version 4.0
G-30-1781-en Page 221
What to do in the event of malfunctions S100 / OPMI® pico
The lamp housing contains a backup lamp which can be manually swung into
the illumination beam path when the first lamp fails. Open flap (1) indicates
that the backup lamp is operative.
Version 4.0
Page 222 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
Version 4.0
G-30-1781-en Page 223
What to do in the event of malfunctions S100 / OPMI® pico
The lamp housing contains a backup lamp which can be manually swung into
the illumination beam path when the first lamp fails. The backup lamp is in
operation if the red segment (3) is lit in button (2).
Version 4.0
Page 224 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
1
2
Version 4.0
G-30-1781-en Page 225
What to do in the event of malfunctions S100 / OPMI® pico
Troubleshooting
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the system
stating it is out of order and contact our service representative.
Version 4.0
Page 226 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
Version 4.0
G-30-1781-en Page 227
What to do in the event of malfunctions S100 / OPMI® pico
Version 4.0
Page 228 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
Version 4.0
G-30-1781-en Page 229
What to do in the event of malfunctions S100 / OPMI® pico
Version 4.0
Page 230 G-30-1781-en
S100 / OPMI® pico What to do in the event of malfunctions
Malfunctions of S100 suspension systems with video and image storage op-
tion
Version 4.0
G-30-1781-en Page 231
What to do in the event of malfunctions S100 / OPMI® pico
Version 4.0
Page 232 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
System maintenance......................................................................234
Maintenance intervals..............................................................................234
Safety check ............................................................................................235
Changing the halogen lamp ....................................................................236
Changing the lamp module of the xenon illumination system ..................238
Changing the fuse - halogen illumination system .....................................240
Version 4.0
G-30-1781-en Page 233
Care and Maintenance S100 / OPMI® pico
System maintenance
Maintenance intervals
Version 4.0
Page 234 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
Safety check
Version 4.0
G-30-1781-en Page 235
Care and Maintenance S100 / OPMI® pico
Version 4.0
Page 236 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
1
2
3
7 8 9
Version 4.0
G-30-1781-en Page 237
Care and Maintenance S100 / OPMI® pico
Version 4.0
Page 238 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
1
3
Version 4.0
G-30-1781-en Page 239
Care and Maintenance S100 / OPMI® pico
Fuse rating
For 100...120 V~ T 6.3 A/H 250 V
For 220..0.240 V~ T 3.15 A/H 250 V
Version 4.0
Page 240 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
Cleaning
Prevention of fogging
To protect the eyepiece optics from fogging, we recommend using an anti-
fogging agent. Anti-fogging agents provided by eyecare professionals for use
with eyeglass lenses are also suitable for Zeiss eyepieces.
• Please observe the instructions for use supplied with each anti-fogging
agent.
Anti-fogging agents do not only ensure fog-free eyepiece optics. They also
clean the eyepiece optics and protect them from dirt, grease, dust, fluff and
fingerprints.
Version 4.0
G-30-1781-en Page 241
Care and Maintenance S100 / OPMI® pico
Sterilization
The asepsis sets available from Carl Zeiss contain rubber caps and hand grips
which can be sterilized in autoclaves. For detailed information on sterilization
please see the enclosed instructions "Preparation of resterilizable products"
for the respective asepsis set.
Sterile single-use drapes are available to cover the system.
• When draping the system, make sure there is enough slack in the drapes
to allow for movement of the microscope carrier and surgical microscope.
• Do not cover any ventilation slots to permit sufficient cooling of the lamps
and avoid lamp failure.
Disinfection
Version 4.0
Page 242 G-30-1781-en
S100 / OPMI® pico Care and Maintenance
Note on disposal
User information
on the disposal of electrical and electronic devices
This symbol means that the product must not be disposed of as normal do-
mestic waste.
The correct disposal of electrical or electronic devices helps to protect the en-
vironment and to prevent potential hazards to the environment and/or human
health which may occur as a result of improper handling of the devices con-
cerned.
For detailed information on the disposal of the product, please contact your
local dealer or the device manufacturer or its legal successor. Please also note
the manufacturer's current information on the Internet. In the event of resale
of the product or its components, the seller is required to inform the buyer
that the product must be disposed of in accordance with the applicable na-
tional regulations currently in force.
Version 4.0
G-30-1781-en Page 243
Care and Maintenance S100 / OPMI® pico
Version 4.0
Page 244 G-30-1781-en
S100 / OPMI® pico System Data
System Data
Version 4.0
G-30-1781-en Page 245
System Data S100 / OPMI® pico
Technical data
Component Features
Magnification Manual magnification changer
Five settings at
γ = 0.4x, γ = 0.6x, γ = 1.0x, γ = 1.6x, γ = 2.5x
Focusing Focusing is performed by setting the working dis-
tance required from the surgical field.
Object lenses with a focusing mechanism are fo-
cused mechanically using a button.
Focusing range 13 mm
Tubes / eyepieces Optionally diagonal, straight, or 180° inclined binoc-
ular tube.
10x widefield eyepieces, alternatively: 12.5 x
Object lenses Options: f = 200 mm, f = 250 mm, f = 300 mm
(non-focusable) f = 350 mm, or f = 400 mm
Object lenses Options: f =200 mm, f = 250 mm or f = 300 mm
(focusable)
Weight approx. 1.7 kg (without CCU, Endoport, tube, object
lens, eyepieces and MORA interface)
Version 4.0
Page 246 G-30-1781-en
S100 / OPMI® pico System Data
Component Features
Suspension arm Length: 600 mm
Swivel angle about carrier arm: ± 150°
Lifting range: ± 300 mm
Carrier arm Length: 500 mm
Stand height 1730 mm
Base 625 x 625 mm
(dimensions)
Weight approx. 90 kg
(including
microscope)
± 300 mm
1730 mm
360° ± 150°
1470 mm
Version 4.0
G-30-1781-en Page 247
System Data S100 / OPMI® pico
Component Features
Suspension arm Length: 950 mm
Lifting range: ± 400 mm
Swivel angle about carrier arm: ± 150°
Carrier arm Length: 500 mm
Swivel angle about mount column: ± 105°
Stand height 1570 mm in horizontal position
Weight approx. 45 kg
(including
microscope)
520 mm
± 400 mm
±105° ±150°
Version 4.0
Page 248 G-30-1781-en
S100 / OPMI® pico System Data
Components Features
Suspension arm Length: 950 mm
Swivel angle about carrier arm: ± 150°
Lifting range: ± 400 mm
Carrier arm Length: 500 mm
Swivel angle about mount column: ± 163°
Stand height 1570 mm in horizontal position
Weight approx. 55 kg
(including
microscope)
500 mm 950 mm
425 mm
± 400 mm
± 163° ± 150°
Version 4.0
G-30-1781-en Page 249
System Data S100 / OPMI® pico
Component Features
Suspension arm Length: 950 mm
Swivel angle on carrier arm: 150°
Stroke: ± 400 mm
Carrier arm Length: 500 mm
Installation height 1570 mm in horizontal position
(recommended)
Weight approx. 30 kg
(including microscope
and illumination
system)
500 mm 950 mm
425 mm
±150° ± 400 mm
Version 4.0
Page 250 G-30-1781-en
S100 / OPMI® pico System Data
Component Features
Suspension arm Length: 950 mm
Swivel angle about carrier arm: ± 150°
Stroke: ± 400 mm
Carrier arm Length: 500 mm
Weight approx. 35 kg
(including microscope and
illumination)
275 mm
± 400 mm
± 150°
Version 4.0
G-30-1781-en Page 251
System Data S100 / OPMI® pico
Component Features
Suspension arm Length: 950 mm
Swivel angle about carrier arm: 150°
Lifting range: ± 400 mm
Carrier arm Length: 500 mm
Installation height 1570 mm in horizontal position
(recommended)
Weight (including approx. 25 kg
microscope)
275 mm
± 400 mm
± 150°
Version 4.0
Page 252 G-30-1781-en
S100 / OPMI® pico System Data
Component Features
Suspension arm Length: 600 mm
Swivel angle about carrier arm: ± 160°
Stroke: ± 300 mm
Carrier arm Length: 500 mm
Weight approx. 34 kg
(including microscope and
illumination)
275 mm
* 330 mm
±300 mm
± 160°
Version 4.0
G-30-1781-en Page 253
System Data S100 / OPMI® pico
Component Features
Suspension arm Length: 950 mm
Swivel angle about carrier arm: ± 160°
Stroke: ± 400 mm
Carrier arm Length: 500 mm
Weight approx. 35kg
(including microscope and
illumination)
500 mm 950 mm
* 330 mm
425 mm
± 400 mm
± 160°
Version 4.0
Page 254 G-30-1781-en
S100 / OPMI® pico System Data
Component Features
Suspension arm Length: 950 mm
Swivel angle about carrier arm: 150°
Stroke: ± 400 mm
Carrier arm Length: 500 mm
Installation height 1570 mm in horizontal position
(recommended)
Weight (including approx. 25 kg
microscope)
275 mm
± 400 mm
± 150°
Version 4.0
G-30-1781-en Page 255
System Data S100 / OPMI® pico
Component Features
Power connection Only connect the system to wall outlets which
are provided with a properly connected pro-
tective ground conductor.
Rated voltage 115 V~ (100 - 120 V~)
220 VAC (230 - 240 VAC)
Rated frequency 50 - 60 Hz
Electrical outputs SD USB socket for USB stick (optional)
video control units
DV interface IEEE 1394 (fireWire) output port
(optional)
Video output port Y/C
Video output BNC
Remote socket
External switching contact input
Electrical outputs HD Video output HDMI
video control units
Video output YPbPr
Video output BNC
Remote port
External switching contact input
Halogen illumination system
Current consumption 115 V ~ max. 2.0 A
230 V ~ max. 1.0 A
Fiber optic illumination 12 V, 100 W halogen lamp,
Backup lamp can be manually swung into beam
path
Fuses Fuse link
115 V ~ T 6.3 A/H 250 V
230 V ~ T 3.15 A/H 250 V
Xenon illumination system
Current consumption 115 V ~ max. 5.0 A
230 V ~ max. 2.5 A
Version 4.0
Page 256 G-30-1781-en
S100 / OPMI® pico System Data
Component Features
Fiber optic illumination Xenon short-arc reflector lamp,
Color temperature: approx. 5000 K
Rated power: approx. 180 W
Backup lamp can be manually swung into beam
path
Fuses Automatic circuit breaker
Version 4.0
G-30-1781-en Page 257
System Data S100 / OPMI® pico
When using the device, the EMC precautions specified below must be ob-
served.
• Only use spare parts approved by Carl Zeiss for this device.
• Do not use any portable or mobile RF communication equipment in the
vicinity of the device as this may impair the device's function.
• Do not use a mobile phone in the vicinity of the equipment because the
radio interference can cause the equipment to malfunction. The effects of
radio interference on medical equipment depend on a number of various
factors and are therefore entirely unforeseeable.
• Please note the EMC guidelines on the following pages.
Version 4.0
Page 258 G-30-1781-en
S100 / OPMI® pico System Data
Electromagnetic interference
Version 4.0
G-30-1781-en Page 259
System Data S100 / OPMI® pico
< 5 % UT < 5 % UT
(95 % dip of UT) for (95 % dip of UT) for
5s 5s
Version 4.0
Page 260 G-30-1781-en
S100 / OPMI® pico System Data
Version 4.0
G-30-1781-en Page 261
System Data S100 / OPMI® pico
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Version 4.0
Page 262 G-30-1781-en
S100 / OPMI® pico System Data
Version 4.0
G-30-1781-en Page 263
System Data S100 / OPMI® pico
d = 1, 17 P d = 1, 17 P d = 2, 33 P
Version 4.0
Page 264 G-30-1781-en
S100 / OPMI® pico System Data
System upgrades
The device S100 / OPMI® pico can be upgraded to an already verified system
with the accessories offered by Carl Zeiss (please refer to the order data). This
device combination is approved by Carl Zeiss.
Additional devices that are connected to electronic medical devices shall
comply with the relevant standards and guidelines
(for example IEC 60950 for data processing devices).
Furthermore, any configurations shall comply with normative requirements
for electronic medical devices /please refer to IEC 60601-1-1:2000 or IEC
60601-1:2005, chapter 16).
The one who is connecting additional devices to electronic medical devices
acts as system configurer and is responsible for the system complying with the
normative requirements for electronic medical devices.
Please note that local legislation takes priority over the above-mentioned nor-
mative requirements. If you have any questions, please contact your local
dealer or Carl Zeiss Service.
Version 4.0
G-30-1781-en Page 265
System Data S100 / OPMI® pico
Ordering data
Only operate the system with the accessories included in the delivery package
and approved by Carl Zeiss. You will find the contact responsible for orders in
your country on this website:
www.meditec.zeiss.com
The product configurations, accessories and spare parts that are applicable for
authorized Carl Zeiss partners are stipulated by contract.
For more information, please contact your authorized Carl Zeiss partner.
Illumination
Description Order No.
12 V, 100 W halogen lamp 380079-9040-000
Tinted UV protective eyewear 000000-0480-028
N-xenon lamp module incl. 2 xenon lamps in trans- 304977-9036-700
port case for Superlux 180
Couplings
Version 4.0
Page 266 G-30-1781-en
S100 / OPMI® pico System Data
Tubes
Eyepieces
Description Cat. No.
10x push-in widefield eyepiece 305542-0000-000
12.5x push-in widefield eyepiece, asph. 305543-9901-000
Objective lenses
Version 4.0
G-30-1781-en Page 267
System Data S100 / OPMI® pico
Hand grips
Splash protection
Description Cat. No.
Drape adapter ring for VisionGuard 000000-1366-769
VisionGuard replacement lenses (sterile, 20/box) 306001-0000-000
Vision Guard lens protector (non-sterile, 10/box) 306002-0000-000
Asepsis
Description Quantity Cat. No.
Sterile hand grip sleeves for OPMI pico 2 pairs 305810-9004-000
Sterile hand grips (rotatable through 360°) pack of 2 305810-9016-000
12 mm asepsis caps pack of 6 305810-9002-000
22 mm asepsis caps pack of 6 305810-9001-000
Asepsis caps PD 180° tiltable tube pack of 6 305810-9003-000
27 mm asepsis caps pack of 6 305810-9008-000
49 mm asepsis caps pack of 2 305810-9007-000
Version 4.0
Page 268 G-30-1781-en
S100 / OPMI® pico System Data
Video accessories
Video accessories can be found in the separate product overview G-30-1888.
Foot switch
Version 4.0
G-30-1781-en Page 269
System Data S100 / OPMI® pico
Accessories
Version 4.0
Page 270 G-30-1781-en
S100 / OPMI® pico System Data
These upgrade kits require a service call which must be coordinated with your
local Carl Zeiss service partner.
Version 4.0
G-30-1781-en Page 271
System Data S100 / OPMI® pico
Environmental conditions
For operation
Version 4.0
Page 272 G-30-1781-en
S100 / OPMI® pico System Data
Symbol Meaning
This way up
Indicates the correct upright position of the
package.
Fragile, handle with care
Keep dry
Do not stack
1
Permissible range of relative humidity
Humidity limits for storage and operation
Permissible range of air pressure
Air pressure limits for storage and operation
Version 4.0
G-30-1781-en Page 273
System Data S100 / OPMI® pico
Regulatory information
Version 4.0
Page 274 G-30-1781-en
S100 / OPMI® pico System Data
Devices S100 / OPMI® pico (Support unit for secondary light source) (000000-1095-
094)
S100 / OPMI® pico (Support unit 600) (000000-1781-138)
S100 / OPMI® pico (Support unit 950) (000000-1781-141)
S100 / OPMI® pico (Support unit Atmos) (000000-1080-349)
Version 4.0
G-30-1781-en Page 275
System Data S100 / OPMI® pico
Version 4.0
Page 276 G-30-1781-en
S100 / OPMI® pico Indexes
Indexes
Term Explanation
Apochromatic optics Optical system with maximum correction of chro-
matic aberration
Drapes Sterile covers for surgical microscopes
FlexioStill Adapter for the attachment of digital compact
cameras to Carl Zeiss surgical microscopes
FlexioMotion Adapter for the attachment of camcorders to Carl
Zeiss surgical microscopes
Interlace (i) In interlace mode, the even and uneven-numbered
lines are alternatively set up. Todays video stand-
ards are based on this procedure.
MediLive Primo Documentation system for Carl Zeiss surgical mi-
croscopes for all disciplines
MediLive Trio Dent Video camera system for documentation purposes
on Carl Zeiss surgical microscopes
MEDIALINK 100 Documentation system for data recording and
output
Pico HD Integrated high-resolution video camera for the
OPMI pico
Progressive (p) In Progressive mode, images are displayed in full
frames. This way, the line structure cannot be rec-
ognized and interline flicker is prevented.
Radiation intensity Radiation incident or emitted at right angles per
unit time and unit area.
TRIO 610 with Video camera system for documentation purposes
CCU TRIO 600 on Carl Zeiss surgical microscopes
VisionGuard Lens guard for Carl Zeiss surgical microscopes
Version 4.0
G-30-1781-en Page 277
Indexes S100 / OPMI® pico
List of abbreviations
Term Explanation
AG Aktiengesellschaft (German acronym for "stock
company")
BEV German acronym for "special emergency backup
power supply"
BNC Bayonet Neill Concelmann
(coaxial bayonet-locking connector for high fre-
quencies, named after its inventor)
CCU Camera Control Unit
(Video camera control unit)
CE Communauté Européenne (European Community)
- The manufacturer declares that the device com-
plies with the directives of the European Union.
DIN Deutsches Institut für Normung (German stand-
ards association)
D Diopter (dioptric power of optical systems)
DVI Digital Video Interface
(digital interface for video data)
EMC Electromagnetic Compatibility - defines non-inter-
ference of electrical and electronic devices with
their environment
EN European Norm (European Standard)
H Breaking capacity of the instrument fuse
HD High Definition
(providing high resolution)
HDMI High Definition Multimedia Interface
(digital interface for video and audio data)
RF Radio frequency
i interlace (see technical term index)
IEC International Electrotechnical Commission
IOL Intraocularlens - artificial lens for the eye
Version 4.0
Page 278 G-30-1781-en
S100 / OPMI® pico Indexes
Term Explanation
JPEG JPEG File Interchange Format
(file format for the storage of images)
LED Light Emitting Diode
(LED)
NTSC National Television Systems Committee
(US institution defining the first color television
standard)
OR Surgical operation
OPMI Surgical microscope
PAL Phase Alternating Line
(color encoding system in analog TV systems)
p progressive (see technical term index)
PD Pupillary distance
SD Standard Definition
SDTV Standard Definition TeleVision
(TV standard providing a lower resolution than
HDTV, e.g. PAL or NTSC)
T* T* multi-coating - minimizes reflection
TIFF Tagged Image File Format
(file format for the storage of images)
UV UltraViolet
VESA Video Electronics Standards Association
(Device to mount flat screen monitors on the wall
or ceiling in accordance with the VESA standard.
Y/C Analog interface that transmits separate brightness
signals (luminance Y) and color signals (chromi-
nance C).
YPbPr Color model that transmits brightness information
Y and color difference information Pb and Pr sep-
arately.
Version 4.0
G-30-1781-en Page 279
Indexes S100 / OPMI® pico
Version 4.0
Page 280 G-30-1781-en
S100 / OPMI® pico
Keyword index
Numerics
1080 Mode
Interlace ......................................................................................207
Progressive ..................................................................................207
720 Mode ..............................................................................................207
Version 4.0
G-30-1781-en Page 281
S100 / OPMI® pico
Version 4.0
Page 282 G-30-1781-en
S100 / OPMI® pico
Electrical
safety ............................................................................................12
Electrical connections, making ................................................................136
Electrical standard ..................................................................................246
Electrician ...............................................................................................136
EMC (electromagnetic compatibility) .......................................................258
Emergency backup power supply .............................................................12
Enter key ..........................................................................................96, 152
Entering a file name for images ..............................................................182
Ergonomic position .................................................................................132
Exposure measurement ..........................................................................194
Eyecup .....................................................................................................76
Eyepiece optics, fog-free .........................................................................112
Eyepieces
technical data .............................................................................246
Version 4.0
G-30-1781-en Page 283
S100 / OPMI® pico
Version 4.0
Page 284 G-30-1781-en
S100 / OPMI® pico
Lamp failure
Halogen ......................................................................................222
xenon .........................................................................................224
Lamp module ...........................................................................................86
Lamp module, change of ........................................................................238
Legal regulations ......................................................................................12
Letterbox ................................................................................................211
Light guide socket ..............................................................................88, 92
Light guide, socket .......................................................................72, 88, 92
Light Source submenu, HD video ............................................................213
Lightmeter sub menu SD Video ......................................................162, 194
Line fuses ...............................................................................................240
live .........................................................................................................156
Live video image, HD video .....................................................................219
Locking pedal ...........................................................................................84
Locking screws .......................................................................................132
Low Light Auto (semi-automatic brightness setting) ...............................218
Version 4.0
G-30-1781-en Page 285
S100 / OPMI® pico
Navigation ..............................................................................................185
Navigation using the control keys ...........................................................152
No Confirm .............................................................................................172
No Confirmation .....................................................................................172
Normal use ...............................................................................................10
Number of images ..................................................................................174
Version 4.0
Page 286 G-30-1781-en
S100 / OPMI® pico
Version 4.0
G-30-1781-en Page 287
S100 / OPMI® pico
Version 4.0
Page 288 G-30-1781-en
S100 / OPMI® pico
Posterior .....................................................................................214
Treatment ...................................................................................214
User sub menu SD Video ........................................................................159
User submenu, HD video ........................................................................190
Version 4.0
G-30-1781-en Page 289
S100 / OPMI® pico
Version 4.0
Page 290 G-30-1781-en
S100 / OPMI® pico
Version 4.0
G-30-1781-en Page 291
Carl Zeiss Meditec AG
Goeschwitzer Strasse 51-52 Fax: + 49 (0)7364 - 20 4823
07745 Jena Email: surgical@meditec.zeiss.com
Germany Internet: www.meditec.zeiss.com