Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
64 Viscometers—Industrial AR
AE
VISC
Flow Sheet Symbol
Types: A. Capillary
A1. Differential pressure
A2. Backpressure
B. Falling element
B1. Falling ball
B2. Falling piston
B3. Falling needle
C. Float in variable-area flow meter
C1. Single-float
C2. Two-float
C3. Concentric
D. Oscillating element
D1. Oscillating blade
D2. Oscillating piston
D3. Torsional oscillating element
E. Plastometers
E1. Cone and plate
E2. Kneader
E3. Capillary extrusion
F. Rotational element
F1. Rotating cone
F2. Agitation power
F3. Double cylinder
G. Vibrating
G1. Reed
G2. Ultrasonic
G3. Rod
H. Coriolis mass flow meter
H1. Measuring tube torsional movement
1723
© 2003 by Béla Lipták
1724 Analytical Instrumentation
Costs: A1. Balanced dual-capillary element without d/p transmitter for fuel oil service, $3000
to $4000
B1. Falling ball viscometer, from $7000 to $16,000; high-pressure design with digital
readout of ball roll time corrected for density, $23,000
B2. From $2500 to $8000; a driven-piston element, $4000
C1 and C2. Two-float indicators with brass or stainless steel fittings in sizes from 0.5
to 1.5 in. (12 to 38 mm), from $1500 to $3000
C3. 1.5-in. (38-mm) transmitter in stainless steel with 300 PSIG flanges, $8000
D2. Sensor with explosion-proof transmitter, $6000 to $8000
D3. Microprocessor-based temperature-compensated transmitter, $15,000
E. $10,000 to $60,000, depending on features and on degree of automation
F1. Variable-speed microprocessor-based units from $15,000 to $25,000
G. $3000 to $5000, depending materials of construction and transmitter features
H. $400 to $20,000, depending on line size for the in-line treatment
INTRODUCTION asphalts. They are also used in fuel oil viscosity control to
optimize atomization in the power industry.
This section covers viscometers that can be installed in a tank As shown in Equation 8.62(2), the pressure drop under
or pipe in an industrial environment. The portable and bench- viscous flow conditions is linearly proportional to viscosity.
top laboratory viscometers are discussed in the previous sec- Therefore, at constant flow rate, the pressure drop in a cap-
tion. The distinction between the two groups is not a sharp illary can be used as a measure of viscosity.
one, and some of the rotary, capillary, cone-and-plate, and
piston designs are actually used in both services. Sensor for Newtonian Fluids
In this section, centipoise (cP) and centistokes (cSt) are
the viscosity units used. For conversion between these Newton’s hypothesis assumed that the viscosity of a fluid is
and others, refer to Tables 8.62d, 8.62e, A.2p, A.2q, and A.2r. independent of the rate of shear or of the shearing force of
A summary of industrial viscometer features and capa- deformation, as long as the flowing temperature and pressure
bilities is given in the Orientation Table 8.64a. In the follow- are fixed. Fluids that behave in this manner are called
ing paragraphs, the various viscometer designs are discussed newtonian. Based on Poiseuille’s law [Equation 8.62(2)] and
in the same order as listed in the Orientation Table. Newton’s hypothesis, we can observe that, if the flow rate
through a fixed capillary tube is held constant, the absolute
viscosity of the flowing liquid will be a linear function of
CAPILLARY VISCOMETERS the capillary pressure drop.
Continuous capillary viscometers are primarily designed
Continuous capillary viscometers are successfully used in oil to measure the viscosity of newtonian liquids. Because d/p
refineries to control various products such as fuel oils, transmitters are used to measure the viscosity, they are also
hydraulic oils, lubricating oils, fuels, and various grades of readily adaptable to the automatic control of processes.
Laboratory Device
Minimum Sample
Maximum design
Inaccuracy (±%)
Temperature, °F
Non-Newtonian
Pressure, PSIG
Local Readout
In-line Device Applicable Viscosity Ranges
Newtonian
in Centipoise
Gas
−2 −1 1 2 3 4 5 6 7 8 9 10
Type of Design 10 10 1 10 10 10 10 10 10 10 10 10 10
8.64 Viscometers—Industrial
Plastometer Cone and plate 100 400 0.5(1) 25 CC Arbitrary Mooney 0–200 Points
Kneader ATM. 570 1.0(1) 80 CC Units %Scale 0–1,000 Division
Capillary 5,000 570 2.0(1) 0.6 #/HR Are Used MI & CIL 0–200 & 0–100
Rotational Cone disc sphere 1,000 650 1.0(1) —
Agitator power 125 200 ~5.0(1) —
Double cylinder 2465 572 1.0(1) —
Vibrational Reed 3,000 392 1.0(1) —
Rod 1,485 500 1.0(1) —
Normal Range. With Special Modifications.
1725
© 2003 by Béla Lipták
1726 Analytical Instrumentation
RELIEF
VALVE
Limitations
VIC
DP
VT TRANSMITTER
OIL TO FUEL
PUMPS OR BURNERS
STEAM VISCOMETER
CONTROL
VALVE GEAR PUMP
HOUSING DISCHARGE SIDE
OF GEAR PUMP
+
CONNECTIONS TO THE
OIL FUEL DIFFERENTIAL PRESSURE
STEAM PREHEATER OUTLET TRANSMITTERS
SUPPLY OIL
FUEL −
INLET
EQUAL CAPILLARY SUCTION SIDE
TUBES OF GEAR PUMP
FUEL
INLET
FIG. 8.64d
Balanced dual-capillary viscometer used to control oil viscosity, which guarantees proper atomization and therefore improved combustion
efficiency. (Courtesy of Conameter Corp.)
The range and span of the d/p transmitter determine the REPEATER DIFFERENTIAL
PRESSURE
range of a single capillary tube. Overall inaccuracy is about TRANSMITTER
±1% of full scale, with repeatability being about the same. TO CONTROLLER
PCV
PIT FALLING-PISTON VISCOMETER
PCV
SAMPLE POV
The falling-piston viscometer has been used in paper sizing,
RETURN printing, coating, polymerization, starch conversion, textile siz-
ing, and blending process applications. The working principle
FIG. 8.64f of the falling-piston viscometer is quite similar to the falling-
Schematic system diagram of a backpressure-type continuous cap-
ball viscometer, discussed in the previous section in connection
illary viscometer.
with Figure 8.63n. Because of its excellent reproducibility, the
falling piston viscometer can be used for the measurement of
As shown in Figure 8.64f, the sample fluid is continuously the viscosity of both newtonian and non-Newtonian fluids.
fed to the instrument from the process line or from a vessel. The measuring element consists of a piston inside a mea-
The sample temperature is maintained by flowing through a suring tube as shown in Figure 8.64g. The measuring element
heat exchanger, which is immersed in a constant-temperature can be installed in a tank (open or closed) or in a liquid-filled
bath. The sample then passes through a pressure regulator pipeline, as long as the measuring tube is completely
followed by a flow controller. Under these conditions of immersed in the fluid.
constant flow rate and constant temperature, the sample pres- During the filling phase, the piston, which is resting at
sure at the entrance to the measuring capillary tube is linearly the bottom of the tube, is automatically raised by an air lifting
proportional to the viscosity of the liquid. mechanism or by a motor-cam mechanism. As the piston is
The inlet pressure to the capillary is sensed by a strain raised, a sample of the liquid is drawn in through the openings
gauge. The strain gauge signal is converted to the desired in the sides of the tube and fills the measuring tube as the
units of viscosity and is transmitted to indicate, record, or piston is withdrawn.
control the process viscosity. Because this viscometer mea-
sures only the inlet pressure to the capillary tube, it is Process Pressure-Operated Design
extremely important to maintain the outlet side at constant
pressure by discharging the capillary to atmosphere or to a In some in-line units, the piston can be raised by the process
pressure-regulated vessel or pipeline. fluid itself. In such designs, the measuring tube is filled
ATTACHED TO LIFTING
MECHANISM
MAGNET
PROXIMITY SWITCH
SWITCH HOUSING
PISTON ROD
LIQUID LEVEL
TUBE OPENING
MEASURING TUBE
PISTON
FIG. 8.64g
The falling-piston viscometer.
through an opening at the bottom of the tube, and the filling 4. The measuring unit should be cleaned at regular
and measuring cycles are controlled by the periodic opening intervals—frequency depends on the rate of material
and closing of a valve. The pressure drop across the measur- buildup.
ing tube is about 5 PSI (35 kPa). 5. An in-line filter should be used to remove any larger
During the measuring phase, the piston assembly is allowed sized solids or foreign materials. The maximum allow-
to fall by gravity, expelling the sample out of the tube through able size of solids remaining should be small enough
the same route that it entered. The time of fall is a measure that they will not interfere with the measurement.
of viscosity, using the clearance between the piston and the 6. Sensitivity requirements should be known for blending
inside wall of the measuring tube as the measuring orifice. The process applications.
timed interval is then displayed on an indicator or recorded. 7. Do not use this viscometer if the process liquids have
For process control or for high-low alarm purposes, the inter- poor flow characteristics.
mittent time signal is converted to a continuous signal by a 8. Erratic readings may result if this detector is subjected
simple clutch mechanism or by a sling-wire mechanism that to severe vibration.
is linked to the recorder pen.
When high-viscosity materials are to be detected, and it Falling-Slug or Falling-Ball Viscometers
is desired to make the measurement within a reasonable
cycle, a two-way air cylinder is used to lift and force down This instrument automatically measures the time required for
the piston. This design also accommodates the use of any a cylindrical slug of a specific density to fall a given distance
mounting position that is desired. in a vertical tube, which is filled with the process liquid at a
6 − constant temperature.
The viscosity range of this sensor is from 0.1 to 10 cP (10
4 3
to 10 Pa⋅s). Each piston has a 100:1 range. The inaccuracy As shown in Figure 8.64h, this viscometer operates by
of this viscometer is 1% of full scale, and the reproducibility the sample pump first purging the system of the previous
and sensitivity are 0.25% and 0.1% of full scale, respectively. sample and then introducing a fresh one. Two separate ther-
Because the cycle frequency of the measurement is from mostats control the purge and recirculation cycles by chang-
once every two minutes up to two cycles per minute, it is ing the direction of flow through a three-way valve.
applicable to both batch and continuous process applications, In the recirculation phase, the sample pump flow raises
with or without automatic process-control capabilities. Pres- the slug to the top of the fall tube. When the sample temper-
sure and temperature ratings of the standard in-line unit are ature has stabilized and matches the thermostat set point, the
from full vacuum to 300 PSIG (2 MPa) and for up to 650°F pump and the sample flow stops, thereby permitting the slug
(340°C). A special high-pressure unit is available for operat- to fall. As it does, it actuates two magnetic switches that are
ing pressures up to 500 PSIG (3.5 MPa). attached to the side of the fall tube.
The first switch starts the recorder motor. When the slug
sinks to and actuates the second magnetic switch, located at
Precautions an adjustable distance (1 to 20 in., or 25 to 500 mm) below
the first, this switch stops the recorder motor. The resultant
As in the case of all other viscometers, operating temperature
time measurement is directly proportional to the viscosity of
and pressure should be specified and kept constant when
the sample. Actuation of the lower switch also initiates the
using this sensor. The error caused by small variations in the
system purge phase.
process fluid temperature can be substantial, so the use of either
a temperature-controlled sampling system or a temperature-
compensated design should be used. Naturally, if compensa- PROCESS
STREAM
PROCESS
STREAM
tion is the choice, the viscosity vs. temperature relationship
must be accurately known. TEMPERATURE TEMPERATURE
M M
Because this viscometer operates in a batch manner, it WELL WELL
should not be used where fast (less than one minute) response
OUTLET THREE-WAY THREE-WAY
time is required. Where a sampling loop is used, the rate of VALVE VALVE
MAGNETIC MAGNETIC
sample pumping should be set to minimize the lag time. To SWITCH SWITCH
FO
obtain reasonable reproducibility, the following should be FALL FO
FALL
TUBE TUBE
observed: MAGNETIC MAGNETIC
SWITCH SWITCH
INLET
1. Avoid any vapor entrainment in the sample liquid GEAR TYPE INLET GEAR TYPE
SAMPLE PUMP SAMPLE PUMP
caused by agitation or boiling.
PURGE PHASE RECIRCULATION PHASE
2. Avoid turbulence.
3. Calibrate the instrument regularly to correct for mea- FIG. 8.64h
surement drift caused by gradual material buildup or The purge and the recirculating (measuring) cycle of the falling-
for wear of the piston and tube. slug viscometer.
The viscosity range of this detector is 10 to 1,000,000 2. The instrument should be installed in a vibration-free
3
cP (0.01 to 10 Pa⋅s). The specific ranges for particular appli- location.
cations are field selectable by adjusting the distance between 3. The process fluid temperature should be carefully
the two magnetic switches on the fall tube. Full scale indi- controlled, or temperature compensation should be
cation can be selected from 10 to 250 sec in five steps. The employed. If no automatic compensation is provided,
inaccuracy and the reproducibility of this viscometer are both viscosity vs. temperature curves should be available
about ±1% full scale, depending on the precision of the for use by operating personnel.
thermostat and of the recorder. Errors caused by small vari- 4. The flow rate and the pressure of the process fluid
ations in the process fluid temperature can be substantial. should both be constant and smooth, pulsating meter-
The viscometer is designed to operate at temperatures up ing pumps should not be used.
to 300°F (150°C) and at pressures up to 200 PSIG (1.4 MPa). 5. The process fluid should be free of foreign material
This detector is suited for continuous viscosity measurement to prevent plugging of the small orifice (0.1 in., or
applications in which it is sufficient to make a reading once 2.5 mm, dia. min.) inside the viscometer. The use of
every 3 min. The falling-slug viscometer is recommended for in-line filters is recommended to remove all foreign
use only on clean process fluids that are not shear sensitive. materials.
6. Install the viscometer in a bypass line to permit the
flushing of the main pipeline and to facilitate viscom-
FLOAT VISCOMETERS
eter maintenance and service.
7. Make certain that the specified operating conditions
The float viscometer has been successfully used to maximize
are not exceeded.
combustion efficiency by controlling the viscosity of fuel oils
8. The viscometer should be periodically recalibrated
in marine and stationery boilers. Other areas of application
using a known viscosity fluid. At the time of calibra-
have been in measuring the viscosity of cement slurry, starch,
tion, the tube and float should also be cleaned.
glue, and petroleum products (motor oils).
9. The sample fluid flow rate through the instrument
These viscometers are used both industrially and in the pilot
should be sufficient to give good speed of response
plant to measure the viscosity of process fluids and to contin-
and sensitivity, but the rate of flow should not cause
uously indicate, record, or control the process. This type of
turbulence within the viscometer, which would result
viscometer is used in a closed-flow system. The operating prin-
in erratic readings.
ciple is similar to that of the variable area flow meter (Section
10. All air and vapor entrainment should be removed
2.27), where the viscous drag force on a float is proportional
from the sample liquid.
to the orifice opening required (between float and tapered tube)
to move the fluid through that orifice at a constant flow rate. 11. When the process temperature is other than ambient,
In a rotameter-type flow meter, the forces acting on the make sure the viscometer is insulated or stream traced,
float are affected by the flow rate, by the float and liquid as required.
specific gravity, and by the viscosity of the fluid being
metered. For flow metering applications, the floats are Single-Float Design
designed so that the viscous drag area is relatively small, so The single-float viscometer is a continuous and direct reading
the float is relatively insensitive to viscosity while being viscosity instrument. As illustrated in Figure 8.64i, a positive
sensitive to flow rate and density changes. displacement pump (other flow control devices can also be used)
In the viscometer version of this design, the flow rate provides the constant sample flow rate through the instrument.
through the variable area meter is held constant. Therefore,
if the position of the float changes, that change is an indica-
THERMOMETER
tion of a change in fluid viscosity and density or in kinematic MAIN LINE
viscosity. To increase its sensitivity, the viscometer float is
designed with a large viscous drag area. To obtain accurate
THERMOMETER
viscosity measurements with float-type viscometers, the flow
rate must be constant. Because this required flow control can
be obtained in three different ways, there are three different FLOAT
The recommended flow rate is between 0.75 and 2.0 GPM pressure differential to provide sufficient flow through the
(2.9 and 7.6 l/m). Automatic temperature compensation is instrument. The required flow rate is 0.25 to 2.5 GPM (0.95
not available for this type of direct reading viscometer. If a to 9.5 l/m), depending on size.
metering pump is used to generate the constant flow rate, the The throttling valve or orifice plate is not required if the
temperature rise through the pump should be measured and outlet can be discharged to a location where the pressure is
corrected for by the use of a viscosity vs. temperature curve. lower than that of the main pipe. A needle valve should be
The single-float viscometer is also available as a transmit- used on the viscometer inlet to allow sensitive and accurate
ter that is used when remote displays or controls are required. flow rate adjustment. As a function of size, the unit is rated
In this case, the float position is detected by the use of an for pressures up to 300 PSIG at 450°F (2 MPa at 232°C),
armature attached to the float extension rod with a magnetic depending on size.
sensing device around its outer periphery. The two-float viscometer can measure the viscosity of
−4
The glass tube viscometer is rated for 450°F (232°C) newtonian fluids from 0.3 to 250 cP (10 to 0.25 Pa⋅s). It
temperature and for 90 PSIG (621 kPa) pressure. With steel- has a rangeability of 10:1 and an inaccuracy of ±4% of
tube viscometers, the pressure rating depends on the operat- indication when the viscosity is higher than 35 cP (0.035
ing temperature, such as a unit for 650 PSIG service at 450°F Pa⋅s). When measuring lower viscosities, the error is ±2%.
(4.5 MPa at 232°C).
The single-float viscometer can be used to measure the Concentric Design
viscosity of non-Newtonian fluids at less that 400 cP (0.4 Pa·s)
The concentric viscometer (Figure 8.64k) consists of a dif-
and can handle newtonian fluids up to 10,000 cP (10 Pa⋅s).
ferential pressure regulator that maintains a constant pressure
The rangeability of this viscometer is between 3:1 and 6:1.
drop across the meter, and a variable area flow meter, which
The inaccuracy of this sensor is ±4% of actual indication,
is provided with a viscosity-sensitivity float. As the fluid
and the reproducibility is ±4% of indication.
enters the instrument, it splits into two streams.
The portion of the fluid that flows upward on the outside,
Two-Float Design
around the differential pressure float, is used to control the
This is a relatively low-cost viscometer designed to provide pressure drop across the meter. The upper end of the differ-
intermittent viscosity measurement in the laboratory, on the ential pressure float acts as a control valve so that, when the
test bench, or in industry. It is only for local indication, not
for transmission.
The design incorporates two floats. The upper float is EXTENSION WELL
sensitive to fluid flow rate, and the lower is sensitive to FLOAT EXTENSION
viscosity. When making a measurement, the fluid flow rate
is first manually adjusted to a constant value as indicated by
the position of the upper float. Under such conditions, but
only while maintaining the volumetric flow constant, the
position of the other float indicates the viscosity of the fluid OUTLET
on a direct reading scale. VISCOSITY FLOAT
The recommended piping configuration for a two-float
viscometer is shown in Figure 8.64j. In the main pipeline, a
throttling valve or an orifice plate can be used between the
viscometer inlet and outlet connections to produce the required
DIFFERENTIAL PRESSURE
FLOAT
INLET
VISCOMETER TUBE
FLOW RATE FLOAT
NEEDLE
VALVE VISCOSITY FLOAT
TWO-FLOAT VISCOMETER
THERMOMETER ORIFICE
OSCILLATING VISCOMETERS
INDICATING
TRANSMITT
Oscillating Blade
PLUNGER COIL
OPTICAL SENSOR
BEAM INTERRUPTER SOLENOID HOUSING
225
MECHANICAL 160 9" 65
STOPS 6.3" 2.5"
BLADE
FIG. 8.64n
Oscillating-blade-type viscosity transmitter. (Courtesy of BTG Inc.)
11/4" NPT
1/ " PIPE 1/ " NPT
2 2
WITH ELBOW
1.48"
(37.6 mm)
11/4" NPT(M)
1.75"
(44.5 mm)
2.6"
(66 mm)
MIXING BLADE
RESERVOIR
COILS SENSOR
A B
WIRES
A B
DEFLECTOR
PLATE OSCILLATING
PISTON
FIG. 8.64o
As the piston is made to oscillate by coils A and B, which are alternately energized, the time of piston travel is registered as viscosity.
(Courtesy of Cambridge Applied Systems Inc.)
This viscometer is provided with an integral resistance It can be provided with a choice of pistons to cover
temperature detector that provides temperature compensa- viscosity ranges from 0.1 to 2 cP up to 1000 to 20,000 cP.
tion. The oscillating piston viscometer can operate at high To use this sensor, the process fluid should be free of solid
pressures (up to 10,000 PSIG, or 700 barg) and at high particles that are larger than about 25 µm in size. The inac-
temperatures (600°F, or 315°C). The wetted parts of the curacy claimed is 2% of actual reading or better. The elec-
sensor are made of stainless steel and Teflon. tronics are available in explosion-proof construction and can
PROCESS
FLUID 11/4" (32 mm)
TEE
PROCESS
FLUID
FIG. 8.64p
Pipeline installation of the oscillating-piston viscometer. (Courtesy of Cambridge Applied Systems Inc.)
PUSH PULL
DRIVER AND
PICKUP
DETECTOR DRIVER
COIL COIL
SENSOR THERMOCOUPLE
FIG. 8.64q
Torsional oscillation-type viscometers are available with spherical and rod-type probe sensors. (Courtesy of Nametre Co.)
be either the local indicator or the transmitter type. The pipe- the prevailing viscosity, the error can be reduced to 2% of
line installation of this detector is illustrated in Figure 8.64p. actual reading. In addition, a compensating filter is provided
to minimize the effect of pipeline vibrations.
Torsional Oscillation Design The viscometer reads the product of absolute viscosity
(cP) and density. To correct for density variations, the density
The sensor in a torsional oscillation-type viscometer can either of the process fluid can be manually dialed in or automati-
be spherical or rod-shaped. The spherical elements can be used cally measured and compensated. These transmitters are
between 0.1 and 200,000 cP, and the rod-shaped ones from 10 available in explosion-proof construction and can be provided
to 2,000,000 cP. Both can be exposed to high pressures (5000 with threaded (1.5-in. NPT) or flanged (2-in. Class 150 ANSI)
PSIG, or 352 barg) and high temperatures (850°F, or 454°C). connections.
As shown in Figure 8.64q, the sensor is electromechan-
ically oscillated in torsion by a rigid inner shaft around its
central axis. The amplitude of oscillation is only 1 µ m, and
the power required to maintain that amplitude is proportional PLASTOMETERS
to viscosity. To provide temperature compensation, the tem-
perature of the process fluid is detected by a thermocouple, Plastometers are employed to study the plastic behavior (melt
which is inserted through a central hole in the shaft. flow) of plastic materials or the molecular weight distribution
To minimize inaccuracy (error), automatic range switching of polymers. Some materials do not begin to flow unless the
is available. By always selecting the smallest span to measure stress has exceeded their yield stress value. Such materials
behave as elastic bodies that yield with time as the stresses THERMOCOUPLE OR DRIVING SHAFT
THERMOMETER
exceed the yield limit. This phenomenon is called plastic HEATER
behavior.
UPPER DIE PLATEN
ROTOR
Plastic Behavior LOWER DIE PLATEN
SAMPLE CHAMBER
Prof. Eugene Bingham assumed that the flow rate for such
materials was zero when they experienced stresses below a TORQUE SENSOR
critical value and was linear with stresses above this limit
value. A body that behaves in such a manner is called a FIG. 8.64r
Bingham solid or an ideal plastic. Extremely few solids Cone-and-plate plastometer.
behave in this manner. For most plastic materials, the yield
value is not sharp and shows a nonlinear behavior with a After the warm-up period has been completed, the plastic
delayed and imperfect recovery. The different types of plas- behavior of the sample is tested by driving the rotor at either
tics behave differently, and their behavior also changes with a constant speed (normally at 2 r/min) or at various speeds
temperature. between 0.05 to 2 r/min to test for relative molecular weight.
For this reason, a plastometer gives only a relative indi- The shearing action that takes place between the rotor and
cation of polymer behavior, thus characterizing a plastic the die cavity is measured by the deflection of a calibrated
material. Because the shear stress/shear rate relationship is U-spring, which is attached to the torque sensing rotor.
based on the molecular behavior of materials, the plasticity The deflection of the U-spring is directly proportional to
can be related to the molecular weight of a material or to its the shearing torque of the specimen being tested. An elec-
distribution. The plasticity of plastic materials is expressed tronic strain gauge may be attached to the U-spring to trans-
in arbitrary units such as ASTM melt index (MI), Canadian mit this signal for continuous recording. The plastometer
Industries Limited (CIL) flow index, Mooney points, or per- operates within a range of 0 to 200 (Mooney) units and has
cent of full scale. a repeatability of ±0.5 units.
In general, the plastometer consists of a heating chamber
to keep the sample at constant temperature, a mechanism to The Unit of Mooney Dr. Mooney gives the following equa-
apply a high shear force, and a device to measure torque or tion to determine the average viscosity ( µ ) of solids:
flow rate of the sample material and thereby detect the plas-
ticity of the sample. Plastometers can be used in a laboratory
G 188.44gθ
(Section 8.63) to study polymeric behavior or used on line µ= 8.64(3)
to control processes manually or automatically. This type of 2πωR 3 ( R/2a + h/b)
rheometer is an invaluable tool in manufacturing plastics and
synthetic rubbers. where
G = gauge reading
Cone-and-Plate Plastometer g = acceleration from gravity
θ = pitch radius of worn gear
The cone-and-plate plastometer is used to conduct tests in ω = angular velocity of rotor, radians/sec
evaluating crude rubber, rubber compounds, or reclaims; in R = radius of rotor, inches
the control of mill breakdown of polymer molecules; in the a = vertical clearance between rotor and stator above or
determination of time to scorch; in the calculation of optimal below the rotor, inches
cure time; and for the evaluation of the processing charac- h = thickness of rotor, inches
teristics of plastics. b = effective radial clearance between rotor and stator
This plastometer incorporates the features of the Mooney
plastometer and is designed to meet the requirements of Kneader Plastometer
ASTM standard test method D-1646. The working principle
of this sensor is the same as discussed in connection with This instrument is widely used in measuring the plastic char-
laboratory-type cone-and-plate viscometers (Figure 8.63r). acteristics of plastic, rubber, food, pigment, cement, paint,
As shown in Figure 8.64r, this detector is designed to cosmetics, and coating products. This type of plastometer is
eliminate polymer “ball-up” and slippage tendencies by con- equally suitable for both laboratory and on-line measurement
fining the sample in a disc-shaped cavity. Machined serra- and the control of highly viscous flowing materials. The chief
tions provided on all of the die and rotor surfaces prevent advantage of this instrument is that it measures plastic behav-
slippage. In operation, the sample is introduced into a cylin- ior of melt viscosity under conditions that are very similar
drical test chamber where integral heaters are used to bring to those that exist in the processing equipment.
it to the predetermined test temperature (up to 400°F, or Many different shapes of measuring heads are available
204°C). (see Figure 8.64s) to accommodate a wide range of viscosities.
THERMOSTATING JACKET
HEATING
MEDIUM
EXTRUDER FLOW
OUT
HEATING
MEDIUM
SELECTOR
IN
SIGMA BLADE DELTA BLADE CAM BLADE VALVE
FOR PLASTIC FOR FOOD, FOR RUBBERS &
DOUGHTY MATERIALS TOOTH PASTE, PLASTICS
FATS AND
METERING
VASELINE TYPE
PUMP
DC
FIG. 8.64s MOTOR
Measuring heads of kneader plastometer.
EXTRUDATE
GEAR
REDUCER
STRANDING DIE TO RECEIVER
TACHOMETER
EXTRUDER FLOW
FIG. 8.64u
CIL PRESSURE Flow-sensing capillary plastometer.
TRANSDUCER
SS
MI PRESSURE
polyethylene through an open-ended capillary at 190°C
TRANSDUCER METERING [374°F] and 43.2 PSIG [298 kPa] pressure). It can also record
PUMP
CIL flow index (190°C and 1500 PSIG [10 MPa] for polypro-
PRC DC pylene) or can alternately record both ASTM and CIL melt
MOTOR
MI & CIL indexes.
SELECTOR These two measurements for a particular polymer can be
VALVE
GEAR
interpreted as both a molecular weight parameter and as a
REDUCER molecular weight distribution parameter. The capillary-type
TO plastometer is not as versatile as the previously described
TACHOMETER
RECEIVER
plastometers.
In feeding the capillary type plastometer, the polymer
FIG. 8.64t melt is first conditioned to be at some specific temperature
Pressure-controlled capillary plastomer. and pressure. It is then extruded through a capillary of suit-
able dimensions. The rate of flow of the polymer through the
capillary is measured by detecting the output of the tachom-
In this plastometer, the jacket heater keeps the sample at a eter on the metering pump and is recorded in units appropri-
constant temperature (up to 570°F, or 299°C) by the jacketed ate for the test (Figure 8.64u).
heater. The kneader is driven by a dynamometer (available The operating range of this plastometer is 0 to 200 MI
with variable-speed and fixed-speed drives). The resistance and 0 to 100 CIL units, with a repeatability within ±2% of
encountered by the mixing blades is detected and transmitted full scale. The unit is designed to operate at up to 570°F
to the dynamometer housing, which tends to rotate in a direc- (299°C) and up to 5000 PSIG (35 MPa) in pressure.
tion opposite to that of the driving shaft.
The torque is transmitted for recording or to automati- Advantages One advantage of this instrument is that it can
cally control processes with medium-viscosity, free-flowing be used as an automatic on-line process control device for
fluids. The readings and recordings are on a 1000-unit divi- both viscous fluids and plastic solids. Especially for the solu-
sion arbitrary scale, used for indicating shearing torque of tion polymerization processes, the polymer solution can be
the specimen being tested. Repeatability is ±1% of full scale. directly analyzed in this plastometer as the sample is received
The measurement range can be adjusted. from the processing reactors through an autosampling device
that flashes the solvent and the unreacted monomers, and
Capillary Plastometer melts the polymer prior to sending it to the plastometer.
For plastic solids, the die unit (capillary) can be mounted
This plastometer is designed for use in polymer manufactur- to the pelletizing extruder of the process stream for measure-
ing plants and is based on the capillary-tube viscometer prin- ment. The capillary type plastometer is ideally suited for
ciple. The capillary plastometer shown in Figure 8.64t is study of plastic behavior of materials that are processed
calibrated to record either ASTM melt index (flow rate of through injection molding or finishing operations.
Limitations The following conditions might degrade the accu- the viscous drag on the spindle. At this point, the cage and
racy of the measurements obtained using capillary-extrusion spindle both rotate at the same speed but with a definite angular
viscometers: relationship to each other. This angular relationship is pro-
portional to the torque on the spring.
1. Nonuniformity of sample shear rate Two methods are used to convert the angular relationship
2. Entrance effects—energy losses at the capillary into a viscosity reading. One method is shown in Figure 8.64v,
entrance and discharge points where one side of a variable capacitor is attached to the cage
3. Compressibility of fluids and the other to the spindle arm. A capacitance is thus made
4. Pressure loss produced by the flow in the sample proportional to the angular relationship between cage and
chamber spindle.
5. Temperature gradient created by shear-induced heat The other method is based on a potentiometer mounted
in the cage with its free member connected to the spindle
When performing scientific research, great care should arm. In this case, a resistance is made that is directly propor-
be taken to correct for these errors. If that is done, capillary- tional to the angular relationship between cage and spindle.
extrusion viscometers have excellent reproducibility and are In either case, the measured signal is transmitted to a receiver
well suited for routine industrial and scientific work. and is converted into viscosity units.
When measuring low viscosities, the variable capacitor
type design is preferred. This type of viscometer can measure
viscosity ranges from 0–10 cP (0–0.01 Pa⋅s) to 0–50,000 cP
ROTARY VISCOMETERS
(0–50 Pa⋅s). Its inaccuracy is ±1% of full scale between the
ranges of 0 to 100 and 0 to 50,000 cP (0 to 0.1 and 0 to
Rotating-cone viscometers have been applied chiefly to mea- 50 Pa⋅s). Its repeatability is ±0.3% of full scale for all ranges.
sure non-Newtonian fluids such as paints, printing ink, starch,
The response time of this viscometer is less than 30 sec
size solution, and varnishes. They are designed to operate
for detecting a full scale change in viscosity. The measure-
continuously in industrial environments. They measure vis-
ment range of this viscometer can be varied over a 30:1 ratio
cosity by detecting the torque required to continuously rotate
by changing the size or shape of the spindle. The standard
a cone, disk, sphere, or spindle in a liquid. In this type of
spindle speed is 50 r/min. In some installations, the maximum
process viscometer application, the sample is continuously
range can be increased to 0 to 100,000 cP (0 to 100 Pa⋅s).
replaced and is subjected to a constant shear rate; thus, the
measurement of non-Newtonian apparent viscosity is possi-
Installation This viscometer may be installed on a tank, in
ble, as well as absolute viscosity measurement of newtonian
an open vessel, or in a sampling line (Figure 8.64w). For in-
fluids.
line installations, the flow rate of the process liquid should
be constant, nonpulsating, and laminar (less than 3 GPM
Rotary Spindle Design [11.4 l/m] for a standard 4-in. [101.6-mm] pipe). Swirling
In the design illustrated in Figure 8.64v, a synchronous induc- motion in the measurement chamber may be reduced by
tion motor (for safety) drives a cage coupled through a cal- installation of a spider-type deflector in the inlet port.
ibrated spring to a spindle arm that supports the spindle or The viscometer head is designed to operate at tempera-
cylinder in the fluid being measured. During measurement, tures up to 160°F (71°C) and at pressure ranges of 8 mm Hg
the spring tends to wind up until the force it generates equals absolute to 100 PSIG (1 to 690 kPa). The head should be
3
purged with dry air or other inert gases at a rate of 0.5 ft /min
SPINDLE
SPINDLE EXTENSION
FLOW
SPINDLE
(A) ON A REACTOR, TANK OR VESSEL (B) IN A SAMPLING LINE
−4 3
(2.4 × 10 m /s) or greater. A positive pressure differential CONNECTING ROD
TO MEASURING
of 3 PSI (21 kPa) should be maintained between the viscom- POTENTIOMETER
eter and the process vessel to prevent liquid or vapor pene-
SYNCHRONOUS
tration into the head. MOTOR
The temperature of the mounting flange on which the
viscometer is placed should not exceed 600°F (316°C).
Above 220°F (104°C), it is necessary to use a cooling pad BALL BEARING
STAINLESS STEEL/
and place it between the flange and the viscometer. The SAPPHIRE BEARING
MAGNET (ROTATING)
process liquid temperature can be up to 2500°F (1370°C). NICKEL ROD
(ARMATURE OF
COUPLING)
Performance The rotating-cone viscometer is not affected
by normal industrial vibration. It is provided with overrange SAMPLE IN
protection through a friction clutch on the spindle extension.
It is simple to clean the spindle. The sample fluid should be
maintained at a constant temperature, or the viscosity mea- MEASURING BODY
(ROTATING DISC
surement should be compensated for temperature variations. MEASURING IN LIQUID)
Logarithmic temperature-compensating units are available CELL
LOWER BODY BEARING
and are matched to the viscosity–temperature relationship
shown by the particular process fluid. Logarithmic compen- FIG. 8.64x
sation to within 1% error over a span of 10°F (6°C) can be Magnetically coupled rotational industrial viscometer.
provided.
A baffle tube should always be installed when this viscom-
eter is used in an agitated tank or reactor to obtain laminar
flow condition. However, if very violent agitation or bubbling
conditions exist, this viscometer should not be used at all.
The spindle should always be immersed, and the liquid level
should be kept constant to obtain reproducible readings.
The frequency of the power supply should be checked.
Wide deviations from the specified frequency will cause spin-
dle speed changes and can introduce errors in the viscosity
measurement of non-Newtonian fluids, because their appar-
ent viscosity is related to shear rate.
The rotating-cone viscometer is an electromechanical
device, so a monthly preventive maintenance and cleaning FIG. 8.64y
schedule is recommended. If the unit is operated continu- Pipeline and vessel installations of the gyrating element viscometer.
ously, the speed reduction gears will need to be replaced on (Courtesy of C. W. Brabender Instruments Inc.)
a yearly basis.
Because of its simple design, ease of cleaning, and non- The magnetically coupled viscometer should not be used
clogging features, this viscometer has been successfully used to measure fluids that contain fiber, ferrite, or abrasive mate-
in monitoring solid and liquid blending processes. Rotating- rials, because they interfere with the operation of the mag-
cone viscometers are quite versatile in their application over netic coupling or can damage the stainless steel–sapphire
wide ranges of viscosities and are equally applicable for bearings.
measuring the viscosities of newtonian and non-Newtonian
fluids. Gyrating-Element Viscometer
RECEIVER
PROCESS FLUID
SEALING
ELEMENT
OUTLET HOLE
APERTURES ON
EXTERNAL CYLINDER
PROPELLER
MEASURING
ZONE
ROTATING
MEASURING
STATIONARY CYLINDER
EXTERNAL
CYLINDER
FIG. 8.64aa
Self-cleaning cylinder viscometer. (Courtesy of BIG Inc.)
DRIVE COIL
PICKUP END
DETECTOR OUTPUT SIGNAL
800
DRIVER END
3/4"
N.P.T. 600
(MVAC)
9/16" 200
1 10 100 1k 10k 100k
NODE LOG. VISCOSITY (CENTIPOISE)
POINT DRIVE
1/2" CONDUIT ARMATURE SOFT IRON FIG. 8.64cc
HUB STATOR
The response curves of four vibrating-reed viscometers.
FIG. 8.64bb
Vibrating-reed viscometer.
Typical response curves of four different detector units are
shown in Figure 8.64cc.
The magnitude of the voltage generated in the pickup coil is Because the viscosity of fluids depends on their temper-
a measure of the process viscosity, because it is proportional ature and pressure, it is mandatory to maintain the process
to the pickup armature vibration. The 120-Hz output signal pressure and temperature constant. As shown in Figure
from the detector is converted into a 0 to 10 mV DC signal 8.64dd, if the fluid temperature must be allowed to fluctuate,
to indicate, record, or control the viscosity of newtonian or temperature compensation is required. Before ordering a vis-
non-Newtonian fluids. cometer for in-process use, the fluid’s viscosity must be mea-
sured as a function of temperature so that this characteristic
Performance Inaccuracy of the instrument is about ±1% of curve can be used in the converter.
full scale, and reproducibility is a little better. There are five For accurate measurements, laminar flow conditions should
different detector units to cover the viscosity ranges from be maintained. Therefore, the installation of this type of
−1 5 −4 2
10 to 10 cP (10 to 10 Pa⋅s) at relatively low shear rate. viscometer directly in an agitated vessel is not recommended.
VISCOSITY DETECTOR vibration frequency is close to 120 Hz, it can interfere with
the reliability of viscosity measurement by this sensor.
CONVERTER RECORDER
Chu, B. and Wang, J., Magnet enhanced optical falling needle/sphere rhe- Khachatryan, G. M. et al., Device for the automatic continuous measurement
ometer, Rev. Sci. Instrum., 63(4), April 1992. of the dynamic viscosity of polymers, Fibre Chem. (UK), May–June
Cox, C. J., Park, M. J., and Wilson, C. A., Use of the capillary tube viscom- 1983.
eter for the determination of the flow behavior of tall oil soap, Trans. Kurano, Y., Accurate measurement of viscosity under high pressure with a
New Zealand Inst. Prof. Eng., March 1983. falling-sphere viscometer, High Temp.—High Press., 26(1), 1994.
Cutrone, L., Use of the Brookfield viscometer to predict rheological perfor- Ohene, F., Livingston, C., Matthews, C., and Rhone, Y., Study of Pressure
mance of coatings, J. Coatings Tech., January 1984. Drop in a Capillary Tube Viscometer for a Two-Phase Flow, Depart-
Dutka, A. P., Crisalle, O. D., Fricke, A. L., Kalotay, P., Evaluation of a ment of Energy Report #DOE/PC/91292-T17, NTIS accession number
capillary-coriolis instrument for online viscosity and density measure- DE96000151 1995.
ments of kraft black liquor, Proc. TAPPI Process Control, Electrical Oppliger, H. R., Matusik, F. J., and Fitzgerald, J. V., New technique accu-
and Instrumentation Conf. (ISA), March 1997. rately measures low viscosity on-line, Control Eng., July 1985.
Ferry, J. D., Oscillation viscometry—effects of shear rate and frequency, Roger, U., Hessenkemper, H., and Roth, P., Glass Conditioning by Viscosity
Meas. Control, September–October 1977. Control, Glass Sci. Tech., 69(8), August 1996.
Fitzgerald, J. V., Matusik, F. J., and Walsh, T. M., Inline viscometry, Meas. Sawhney, I. K., et al., Development of capillary tube viscometer for mea-
Control, December 1987. suring flow properties of khoa, J. Institute Eng. (India), August 1984.
Sexton, W. C., Rotating spindle viscometer, International and 42nd Annual
Fournier-Villalon, P., Bertrand, J., and Couderc, J. P., Using a Brookfield
Symposium and Exposition on Scientific Glassblowing, NTIS acces-
viscometer for determining rheological properties of newtonian, plastic
sion number DE970610182/XAB, June 1997.
and pseudoplastic media, Entropie, 20(120), 1984.
Sheen, S. H., Chien, H. T., and Raptis, A. C., In-Line Ultrasonic Viscometer,
Frisman, M. L., Rotation viscometer for readily unmixing suspensions,
1994 Review of Progress in Quantitative Nondestructive Evaluation
Indust. Lab. (USSR), October 1984.
Conference, Sponsored by Department of Energy, NTIS accession
Galvin, G. D., Hultin, J. F., and Jones, B., Development of a High-Pressure,
number DE95002982, October 1994.
High-Shear Rate Capillary Viscometer, J. Non-Newtonian Fluid
Smith, R. E., Brookfield viscometers for determination of low-shear viscos-
Mechanics, January 1981.
ity and leveling behavior, J. Coatings Tech., January 1984.
Gillis, K. A., Mehl, J. B., and Moldover, M. R., Greenspan Acoustic Vis- Villemaire, J. P. and Agassant, J. F., Apparent viscosity measurements using
cometer for Gases, Rev. Sci. Inst., 67(5), NTIS accession number a capillary viscometer, Polymer Process Eng., 1(3), 1983–84.
PB96–204417, May 1996. Viscometers, Meas. Control, June 1993.
Gimera, M., Software Quality Assurance Plan for Viscometer, Department Vrentas, J. S., Mechanical energy balances for a capillary viscometer, J.
of Energy Report, NTIS accession number DE95002145, October Non-Newtonian Fluid Mech., March 1983.
1994. Walsh, T. M., Continuous viscosity control improves quality, Adhesives Age,
Grupp, J., Torsion Viscometer for Liquid Crystals, Rev. Sci. Inst., June 1983. December 1989.
Helle, H., Valimaki, H., and Lekkala, J., Comparing a 10 MHz thickness- Webber, P. J. and Savage, J. A., Measurement of the viscosity of chalco-
shear mode quartz resonator with a commercial process viscometer in genide glasses by parallel plate technique, J. Mater. Sci., March 1981.
monitoring resol manufacturing process, Sensors and Actuators B Wunderlich, T. and Brunn, P. O., Ultrasound pulse doppler method as a
(Chemical), B81(2–3), January 2002. viscometer for process monitoring, Flow Meas. and Instrum., 10(4),
Jethra, R., Viscosity measurement, ISA Trans., 33(3), September 1994. December, 1999.
Kalotay, P., Online viscosity measurement using coriolis mass flowmeters, Zhang, Z. and Reddy, R. G., Viscosities of lead silicate slags, Minerals
Flow Meas. Instrum., 5(4), October 1994. Metall. Process., 19(1), February 2002.