Sei sulla pagina 1di 21

8.

64 Viscometers—Industrial AR

C. H. KIM (1969, 1982) B. G. LIPTÁK (1995) J. E. JAMISON (2003) AIT

AE
VISC
Flow Sheet Symbol

Types: A. Capillary
A1. Differential pressure
A2. Backpressure
B. Falling element
B1. Falling ball
B2. Falling piston
B3. Falling needle
C. Float in variable-area flow meter
C1. Single-float
C2. Two-float
C3. Concentric
D. Oscillating element
D1. Oscillating blade
D2. Oscillating piston
D3. Torsional oscillating element
E. Plastometers
E1. Cone and plate
E2. Kneader
E3. Capillary extrusion
F. Rotational element
F1. Rotating cone
F2. Agitation power
F3. Double cylinder
G. Vibrating
G1. Reed
G2. Ultrasonic
G3. Rod
H. Coriolis mass flow meter
H1. Measuring tube torsional movement

Design Pressures, See Orientation Table 8.64a


Temperatures, Viscosity
Ranges, Errors:

Materials of Construction: A, B. Hardened stainless steel or other corrosion-resistant metals


C. Same choices as for variable-area flow meters
D1. Wetted parts, stainless steel; sealing ring, silicone rubber
®
D2. Stainless steel and Teflon
D3. E, F. Hardened stainless steel
G, H. Wide selection of corrosion-resistant materials and coatings

Speed of Response: A. From 1 sec to about 5 min


B2. Cycle time from 30 sec to 3 min
C, E, F, H. A few seconds
D1. Strokes every 2 sec, readings averaged over an adjustable number of strokes
D2. About 30 sec
H. Less than 1 sec

1723
© 2003 by Béla Lipták
1724 Analytical Instrumentation

Costs: A1. Balanced dual-capillary element without d/p transmitter for fuel oil service, $3000
to $4000
B1. Falling ball viscometer, from $7000 to $16,000; high-pressure design with digital
readout of ball roll time corrected for density, $23,000
B2. From $2500 to $8000; a driven-piston element, $4000
C1 and C2. Two-float indicators with brass or stainless steel fittings in sizes from 0.5
to 1.5 in. (12 to 38 mm), from $1500 to $3000
C3. 1.5-in. (38-mm) transmitter in stainless steel with 300 PSIG flanges, $8000
D2. Sensor with explosion-proof transmitter, $6000 to $8000
D3. Microprocessor-based temperature-compensated transmitter, $15,000
E. $10,000 to $60,000, depending on features and on degree of automation
F1. Variable-speed microprocessor-based units from $15,000 to $25,000
G. $3000 to $5000, depending materials of construction and transmitter features
H. $400 to $20,000, depending on line size for the in-line treatment

Partial List of Suppliers: Automation Products (G) (www.dynatrolusa.com)


Benz Co. Inc., Edwin H. (B1), (www.benztesters.com)
Bohlin Rheologi (E) (www.bohlinusa.com)
C. W. Brabender Instruments Inc. (E, F1) (www.cwbrabender.com)
Brookfield Engineering Laboratories Inc. (F1) (www.brookfieldengineering.com)
BTG Inc. (D1, F3) (www.btghome.com)
Cambridge Applied Systems Inc. (D2) (www.cambridge-applied.com)
CB Engineering Ltd.(www.cbeng.com)
Conameter Corp. (A1); Dynatrol Viscosity Systems (G3) (www.dynatrolusa.com)
Endress + Hauser (H) (www.endress.com)
Extech Instruments (B2) (www.extech.com)
Fischer & Porter, a Unit of ABB Inc. (C1, C2, C3) (www.abb.com/us)
Gam Rad West Inc. (B1, E) (1-800-321-8452)
Nametre Co. (D3, F1) (www.nametre.com)
Norcross Corp. (B2) (www.viscosity.com)
Petroleum Analyzer Co. (PAC) (A, B2) (www.petroleum-analyzer.com)
Ruska Instrument Corp. (B1) (www.ruska.com)
Solartron Mobrey (G1) (www.solartronusa.com)
Stony Brook Scientific Ltd. (B3) (www.stonybrooksci.com)
Techne Inc. (E3) (www.techneusa.com).

INTRODUCTION asphalts. They are also used in fuel oil viscosity control to
optimize atomization in the power industry.
This section covers viscometers that can be installed in a tank As shown in Equation 8.62(2), the pressure drop under
or pipe in an industrial environment. The portable and bench- viscous flow conditions is linearly proportional to viscosity.
top laboratory viscometers are discussed in the previous sec- Therefore, at constant flow rate, the pressure drop in a cap-
tion. The distinction between the two groups is not a sharp illary can be used as a measure of viscosity.
one, and some of the rotary, capillary, cone-and-plate, and
piston designs are actually used in both services. Sensor for Newtonian Fluids
In this section, centipoise (cP) and centistokes (cSt) are
the viscosity units used. For conversion between these Newton’s hypothesis assumed that the viscosity of a fluid is
and others, refer to Tables 8.62d, 8.62e, A.2p, A.2q, and A.2r. independent of the rate of shear or of the shearing force of
A summary of industrial viscometer features and capa- deformation, as long as the flowing temperature and pressure
bilities is given in the Orientation Table 8.64a. In the follow- are fixed. Fluids that behave in this manner are called
ing paragraphs, the various viscometer designs are discussed newtonian. Based on Poiseuille’s law [Equation 8.62(2)] and
in the same order as listed in the Orientation Table. Newton’s hypothesis, we can observe that, if the flow rate
through a fixed capillary tube is held constant, the absolute
viscosity of the flowing liquid will be a linear function of
CAPILLARY VISCOMETERS the capillary pressure drop.
Continuous capillary viscometers are primarily designed
Continuous capillary viscometers are successfully used in oil to measure the viscosity of newtonian liquids. Because d/p
refineries to control various products such as fuel oils, transmitters are used to measure the viscosity, they are also
hydraulic oils, lubricating oils, fuels, and various grades of readily adaptable to the automatic control of processes.

© 2003 by Béla Lipták


TABLE 8.64a
Orientation Table for Industrial Viscometers

Provides Continuous Signal

(2) Based on Measurement


Features

Remote Readout Trans.

(1) Based on Full Scale


Temp. Compensation

[°C = (°F −32)/1.8]


Fluids

Laboratory Device

(1 Bar = 14.2 PSI)

Size or Flow Rate


Maximum Design

Minimum Sample
Maximum design

Inaccuracy (±%)
Temperature, °F
Non-Newtonian

Pressure, PSIG
Local Readout
In-line Device Applicable Viscosity Ranges

Newtonian
in Centipoise

Gas
−2 −1 1 2 3 4 5 6 7 8 9 10
Type of Design 10 10 1 10 10 10 10 10 10 10 10 10 10

Continuous Differential       670 900 1–2(1) 1–4 GPH


capillary pressure
Back pressure       500 210 1.0(1) 1 GPH
Coriolis Torsional        1,440 302 5.0(2) —
or ± 0.5 cP
Falling element Ball or slug        300 350 1.0(1) —
Piston        500 650 1.0(1) —
Needle        2000 662 1.0(1) —
Float Single float        650 450 4.0(2) 0.75–2 GPM
Two-float         300 450 2–4(2) 0.25–2.5 GPM
Concentric      650 450 2–4(2) 2 GPM
Oscillating Blade       375 150 1.0(1) Up to 6.5 fps (2 m/s)
Piston         10,000 700 1.0(2)
Torsional         5,000 850 2.0(2)

8.64 Viscometers—Industrial
Plastometer Cone and plate     100 400 0.5(1) 25 CC Arbitrary Mooney 0–200 Points
Kneader       ATM. 570 1.0(1) 80 CC Units %Scale 0–1,000 Division
Capillary       5,000 570 2.0(1) 0.6 #/HR Are Used MI & CIL 0–200 & 0–100
Rotational Cone disc sphere         1,000 650 1.0(1) —
Agitator power       125 200 ~5.0(1) —
Double cylinder      2465 572 1.0(1) —
Vibrational Reed         3,000 392 1.0(1) —
Rod        1,485 500 1.0(1) —
Normal Range. With Special Modifications.

1725
© 2003 by Béla Lipták
1726 Analytical Instrumentation

In most capillary-type viscometers, the temperature effect SAMPLE PNEUMATIC OR


SWEEPSTREAM ELECTRICAL SIGNAL
on the measurement is eliminated not by temperature com- TO A RECEIVER
CONSTANT DIFFERENTIAL
pensation but by temperature controls. This is done by a STRAINER SPEED PRESSURE
thermostatic device that keeps the sample at a constant ref- MOTOR TRANSMITTER SAMPLE OUTLET

erence temperature as it enters the capillary tube.


This type of viscometer can measure viscosities up to CONSTANT
HEAT
15,000 poise (1500 Pa⋅s) and can operate at temperatures EXCHANGER
TEMPERATURE
BATH
CAPILLARY

up to 900°F (480°C). The viscosity span of this sensor is a


FILTER
function of the bore and the length of the capillary. To minimize
METERING
the end effects (the nozzle effect at the tip of the capillary), PUMP
the use of large-diameter bores and long capillary tubes is TEMPERATURE HEAT
EXCHANGER
recommended. CONTROLLER

RELIEF
VALVE
Limitations

Although this type of continuous viscometer is simple enough FIG. 8.64b


Schematic flow diagram of a differential-pressure-type continuous
to be field fabricated (rather than purchased), one of its chief
capillary viscometer.
disadvantages is that, for accurate and reliable measurements,
the capillary tube must be kept absolutely clean. But fouling
DIFFERENTIAL PRESSURE
is likely to occur, because the capillary tube diameter is very TRANSDUCER
PNEUMATIC OR
FLOW ELECTRICAL SIGNAL
small—usually in the range of 0.05 to 0.2 in. (1.25 to 5 mm). CONTROLLER SAMPLE RETURN LINE TO RECEIVER
Because the sample flow rates must be low, this require- P1
ment limits the continuous capillary viscometer to by-pass P1 P1
installations, where the goal is to automatically control the T1

processes with minimal time lag. MAIN LINE


FLOW IN FLOW OUT
Other potential error sources include (1) fluctuating flow
rates through the capillary tube, (2) a dirty or plugged cap- CAPILLARY METERING PIPE
illary, (3) leakage in the viscometer, (4) incorrectly set or
drifting transducer zero and span, (5) fluctuating fluid pres- FIG. 8.64c
sure, (6) insufficient sample supply pressure, and (7) fluid Schematic diagram of an in-line, differential-pressure-type contin-
temperature fluctuations due to thermostat malfunctions or uous capillary viscometer.
to variation in sample supply temperature.
Differential Pressure Type
Calibration Figure 8.64b is a schematic flow diagram of a differential
pressure viscometer, which requires a bypass installation. An
The continuous capillary viscometer can be calibrated using external strainer is provided to remove any solids from the
the following equations: sample before it enters the viscometer. A constant sample flow
rate (at about 1 GPH or 3.8 l/h) is maintained by a precision
6 µVL
d=4 8.64(1) metering pump, which is driven by a synchronous motor.
∆P × 10 9 Two heat exchangers, before and after the metering pump,
are used to keep the sample fluid in thermal equilibrium with
∆Pd 4 × 10 9
µ= 8.64(2) the thermostatic bath. A relief valve protects against damage
6VL caused by excessive pressures that may occur in a blocked
Equation 8.64(1) is used to calculate the diameter of the capillary.
capillary tube when the viscosity of the process fluid is The pressure drop across the capillary is measured with
known. Equation 8.64(2) is used to prepare the calibration a differential pressure transducer, which is connected to the
curve of viscosity versus pressure drop using several fluids inlet and outlet sides of the capillary. The d/p transmitter
of known viscosity. Such a calibration should produce a output is a linear indication of the process viscosity and is
straight line that goes through the zero point of the viscosity used for indicating, recording, or controlling the process.
and pressure drop coordinates. This type of viscometer can measure viscosities up to
There are basically two types of viscometers that utilize 2500 cP (2.5 Pa⋅s) and can operate at temperatures up to
the continuous capillary principle. One type measures the 240°F (116°C). If high-pressure capillaries and high-pressure
pressure drop across the capillary tube, and the other mea- metering pumps are used, the viscosity measurement can go
sures the upstream pressure as the sample flows through the up to 15,000 P (1,500 Pa⋅s) and the line pressures can reach
capillary tube. 670 PSIG at 900°F (4.6 MPa at 480°C).

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1727

VIC

DP
VT TRANSMITTER
OIL TO FUEL
PUMPS OR BURNERS

STEAM VISCOMETER
CONTROL
VALVE GEAR PUMP
HOUSING DISCHARGE SIDE
OF GEAR PUMP
+
CONNECTIONS TO THE
OIL FUEL DIFFERENTIAL PRESSURE
STEAM PREHEATER OUTLET TRANSMITTERS
SUPPLY OIL
FUEL −
INLET
EQUAL CAPILLARY SUCTION SIDE
TUBES OF GEAR PUMP
FUEL
INLET

FIG. 8.64d
Balanced dual-capillary viscometer used to control oil viscosity, which guarantees proper atomization and therefore improved combustion
efficiency. (Courtesy of Conameter Corp.)

The range and span of the d/p transmitter determine the REPEATER DIFFERENTIAL
PRESSURE
range of a single capillary tube. Overall inaccuracy is about TRANSMITTER
±1% of full scale, with repeatability being about the same. TO CONTROLLER

The average response time is about 2 min (0.6-min time


constant), but it varies with the length of the sample loop.

In-Line Design To reduce the response time of the viscometer


to less than 1 sec, the capillary tube can be inserted into the
main process stream, and sample flow rate can be increased to FIG. 8.64e
about 4 GPH (15.2 l/h), as shown in Figure 8.64c. In this design, Differential pressure measurement, using extended diaphragm sen-
the constant flow rate through the capillary tube is maintained sors across any section of process pipe to detect viscosity.
by a pressure-regulated, diaphragm-type flow controller.
The error in this measurement is higher than with the and discharge pressures is a linear function of viscosity. This
external sampling design, and inaccuracy is about ±2.0% of pressure difference is measured by a d/p cell and is kept
full scale when measuring viscosity in the range of 5 to 30 constant by adjusting the heat input into the oil preheater.
−6 −6
cSt (5 × 10 to 30 × 10 Pa⋅s). This in-line viscometer is provided with 2-in. (50-mm)
One advantage of this design is its increased speed of DIN flanges, and the flow through it can be up to 100 GPM
response. Another advantage is that, because the viscosity mea- (380 l/m). The fuel oil pressure can be up to 500 PSIG (35 barg),
surements are made at the process stream temperature, this and the temperature can be up to 300°F (149°C). The differ-
viscometer can be used as a viscosity controller that main- ential pressure range of the d/p cell is 0 to 150 in. (0 to 3.8 m)
tains constant fluid viscosity by adjusting temperature. Such of water, which in units of viscosity can correspond to 0 to
applications include controlling fuel oil viscosity to main- 300 SUS, 0 to 200 Redwood 1, or 0 to 6° Engler.
taining optimal atomization patterns in industrial furnaces,
heating plants, steam power stations, and marine boilers. Pipe Section as Viscometer If the flow is constant or is mea-
sured, and therefore it is possible to compensate for the varia-
Dual-Capillary Design Figure 8.64d illustrates a dual-capillary tions in flow rate, the pressure drop through any pipe section
viscometer that is used for controlling the viscosity of fuel in the plant (Figure 8.64e) can be used to detect viscosity. As
oil to boilers and other burners. The dual-capillary design is long as the inner walls of the pipe section are clean, and as long
superior to the single-capillary units, because it is unaffected as the temperature and flow variations are compensated for, the
by variations in fuel line pressure. detected pressure drop will be linearly proportional to viscosity.
Dual-capillary viscometers are installed because keeping
the viscosity constant guarantees consistent atomization of Backpressure Type
the fuel, which, in turn, improves combustion efficiency. In
this viscometer, a gear pump, which is inserted into the fuel The operation of this viscometer is quite similar to the differ-
oil pipe, takes a constant flow-rate sample of the oil and ential pressure type except that it measures only the upstream
discharges it continuously back into the oil stream. As the pressure to a capillary tube that discharges to the atmosphere
flow rate is constant, the difference between pump suction or returns the sample to a pressure-regulated process line.

© 2003 by Béla Lipták


1728 Analytical Instrumentation

PROCESS LINE OR VESSEL


SAMPLE
The backpressure-type viscometer can measure viscosi-
−6 −4
T1
OUTLET ties within the range of 5 to 500 cSt (5 × 10 to 5 × 10
Pa⋅s) at temperatures up to 210°F (99°C). Its overall inaccu-
HEATING
CAPILLARY racy and repeatability are ±1% of full scale, and its response
COIL TO
RECEIVER
time is 3 to 6 min.
P1

PCV
PIT FALLING-PISTON VISCOMETER
PCV
SAMPLE POV
The falling-piston viscometer has been used in paper sizing,
RETURN printing, coating, polymerization, starch conversion, textile siz-
ing, and blending process applications. The working principle
FIG. 8.64f of the falling-piston viscometer is quite similar to the falling-
Schematic system diagram of a backpressure-type continuous cap-
ball viscometer, discussed in the previous section in connection
illary viscometer.
with Figure 8.63n. Because of its excellent reproducibility, the
falling piston viscometer can be used for the measurement of
As shown in Figure 8.64f, the sample fluid is continuously the viscosity of both newtonian and non-Newtonian fluids.
fed to the instrument from the process line or from a vessel. The measuring element consists of a piston inside a mea-
The sample temperature is maintained by flowing through a suring tube as shown in Figure 8.64g. The measuring element
heat exchanger, which is immersed in a constant-temperature can be installed in a tank (open or closed) or in a liquid-filled
bath. The sample then passes through a pressure regulator pipeline, as long as the measuring tube is completely
followed by a flow controller. Under these conditions of immersed in the fluid.
constant flow rate and constant temperature, the sample pres- During the filling phase, the piston, which is resting at
sure at the entrance to the measuring capillary tube is linearly the bottom of the tube, is automatically raised by an air lifting
proportional to the viscosity of the liquid. mechanism or by a motor-cam mechanism. As the piston is
The inlet pressure to the capillary is sensed by a strain raised, a sample of the liquid is drawn in through the openings
gauge. The strain gauge signal is converted to the desired in the sides of the tube and fills the measuring tube as the
units of viscosity and is transmitted to indicate, record, or piston is withdrawn.
control the process viscosity. Because this viscometer mea-
sures only the inlet pressure to the capillary tube, it is Process Pressure-Operated Design
extremely important to maintain the outlet side at constant
pressure by discharging the capillary to atmosphere or to a In some in-line units, the piston can be raised by the process
pressure-regulated vessel or pipeline. fluid itself. In such designs, the measuring tube is filled

ATTACHED TO LIFTING
MECHANISM

MAGNET

PROXIMITY SWITCH

SWITCH HOUSING

PISTON ROD

LIQUID LEVEL

TUBE OPENING

MEASURING TUBE

PISTON

FIG. 8.64g
The falling-piston viscometer.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1729

through an opening at the bottom of the tube, and the filling 4. The measuring unit should be cleaned at regular
and measuring cycles are controlled by the periodic opening intervals—frequency depends on the rate of material
and closing of a valve. The pressure drop across the measur- buildup.
ing tube is about 5 PSI (35 kPa). 5. An in-line filter should be used to remove any larger
During the measuring phase, the piston assembly is allowed sized solids or foreign materials. The maximum allow-
to fall by gravity, expelling the sample out of the tube through able size of solids remaining should be small enough
the same route that it entered. The time of fall is a measure that they will not interfere with the measurement.
of viscosity, using the clearance between the piston and the 6. Sensitivity requirements should be known for blending
inside wall of the measuring tube as the measuring orifice. The process applications.
timed interval is then displayed on an indicator or recorded. 7. Do not use this viscometer if the process liquids have
For process control or for high-low alarm purposes, the inter- poor flow characteristics.
mittent time signal is converted to a continuous signal by a 8. Erratic readings may result if this detector is subjected
simple clutch mechanism or by a sling-wire mechanism that to severe vibration.
is linked to the recorder pen.
When high-viscosity materials are to be detected, and it Falling-Slug or Falling-Ball Viscometers
is desired to make the measurement within a reasonable
cycle, a two-way air cylinder is used to lift and force down This instrument automatically measures the time required for
the piston. This design also accommodates the use of any a cylindrical slug of a specific density to fall a given distance
mounting position that is desired. in a vertical tube, which is filled with the process liquid at a
6 − constant temperature.
The viscosity range of this sensor is from 0.1 to 10 cP (10
4 3
to 10 Pa⋅s). Each piston has a 100:1 range. The inaccuracy As shown in Figure 8.64h, this viscometer operates by
of this viscometer is 1% of full scale, and the reproducibility the sample pump first purging the system of the previous
and sensitivity are 0.25% and 0.1% of full scale, respectively. sample and then introducing a fresh one. Two separate ther-
Because the cycle frequency of the measurement is from mostats control the purge and recirculation cycles by chang-
once every two minutes up to two cycles per minute, it is ing the direction of flow through a three-way valve.
applicable to both batch and continuous process applications, In the recirculation phase, the sample pump flow raises
with or without automatic process-control capabilities. Pres- the slug to the top of the fall tube. When the sample temper-
sure and temperature ratings of the standard in-line unit are ature has stabilized and matches the thermostat set point, the
from full vacuum to 300 PSIG (2 MPa) and for up to 650°F pump and the sample flow stops, thereby permitting the slug
(340°C). A special high-pressure unit is available for operat- to fall. As it does, it actuates two magnetic switches that are
ing pressures up to 500 PSIG (3.5 MPa). attached to the side of the fall tube.
The first switch starts the recorder motor. When the slug
sinks to and actuates the second magnetic switch, located at
Precautions an adjustable distance (1 to 20 in., or 25 to 500 mm) below
the first, this switch stops the recorder motor. The resultant
As in the case of all other viscometers, operating temperature
time measurement is directly proportional to the viscosity of
and pressure should be specified and kept constant when
the sample. Actuation of the lower switch also initiates the
using this sensor. The error caused by small variations in the
system purge phase.
process fluid temperature can be substantial, so the use of either
a temperature-controlled sampling system or a temperature-
compensated design should be used. Naturally, if compensa- PROCESS
STREAM
PROCESS
STREAM
tion is the choice, the viscosity vs. temperature relationship
must be accurately known. TEMPERATURE TEMPERATURE
M M
Because this viscometer operates in a batch manner, it WELL WELL

should not be used where fast (less than one minute) response
OUTLET THREE-WAY THREE-WAY
time is required. Where a sampling loop is used, the rate of VALVE VALVE
MAGNETIC MAGNETIC
sample pumping should be set to minimize the lag time. To SWITCH SWITCH
FO
obtain reasonable reproducibility, the following should be FALL FO
FALL
TUBE TUBE
observed: MAGNETIC MAGNETIC
SWITCH SWITCH

INLET
1. Avoid any vapor entrainment in the sample liquid GEAR TYPE INLET GEAR TYPE
SAMPLE PUMP SAMPLE PUMP
caused by agitation or boiling.
PURGE PHASE RECIRCULATION PHASE
2. Avoid turbulence.
3. Calibrate the instrument regularly to correct for mea- FIG. 8.64h
surement drift caused by gradual material buildup or The purge and the recirculating (measuring) cycle of the falling-
for wear of the piston and tube. slug viscometer.

© 2003 by Béla Lipták


1730 Analytical Instrumentation

The viscosity range of this detector is 10 to 1,000,000 2. The instrument should be installed in a vibration-free
3
cP (0.01 to 10 Pa⋅s). The specific ranges for particular appli- location.
cations are field selectable by adjusting the distance between 3. The process fluid temperature should be carefully
the two magnetic switches on the fall tube. Full scale indi- controlled, or temperature compensation should be
cation can be selected from 10 to 250 sec in five steps. The employed. If no automatic compensation is provided,
inaccuracy and the reproducibility of this viscometer are both viscosity vs. temperature curves should be available
about ±1% full scale, depending on the precision of the for use by operating personnel.
thermostat and of the recorder. Errors caused by small vari- 4. The flow rate and the pressure of the process fluid
ations in the process fluid temperature can be substantial. should both be constant and smooth, pulsating meter-
The viscometer is designed to operate at temperatures up ing pumps should not be used.
to 300°F (150°C) and at pressures up to 200 PSIG (1.4 MPa). 5. The process fluid should be free of foreign material
This detector is suited for continuous viscosity measurement to prevent plugging of the small orifice (0.1 in., or
applications in which it is sufficient to make a reading once 2.5 mm, dia. min.) inside the viscometer. The use of
every 3 min. The falling-slug viscometer is recommended for in-line filters is recommended to remove all foreign
use only on clean process fluids that are not shear sensitive. materials.
6. Install the viscometer in a bypass line to permit the
flushing of the main pipeline and to facilitate viscom-
FLOAT VISCOMETERS
eter maintenance and service.
7. Make certain that the specified operating conditions
The float viscometer has been successfully used to maximize
are not exceeded.
combustion efficiency by controlling the viscosity of fuel oils
8. The viscometer should be periodically recalibrated
in marine and stationery boilers. Other areas of application
using a known viscosity fluid. At the time of calibra-
have been in measuring the viscosity of cement slurry, starch,
tion, the tube and float should also be cleaned.
glue, and petroleum products (motor oils).
9. The sample fluid flow rate through the instrument
These viscometers are used both industrially and in the pilot
should be sufficient to give good speed of response
plant to measure the viscosity of process fluids and to contin-
and sensitivity, but the rate of flow should not cause
uously indicate, record, or control the process. This type of
turbulence within the viscometer, which would result
viscometer is used in a closed-flow system. The operating prin-
in erratic readings.
ciple is similar to that of the variable area flow meter (Section
10. All air and vapor entrainment should be removed
2.27), where the viscous drag force on a float is proportional
from the sample liquid.
to the orifice opening required (between float and tapered tube)
to move the fluid through that orifice at a constant flow rate. 11. When the process temperature is other than ambient,
In a rotameter-type flow meter, the forces acting on the make sure the viscometer is insulated or stream traced,
float are affected by the flow rate, by the float and liquid as required.
specific gravity, and by the viscosity of the fluid being
metered. For flow metering applications, the floats are Single-Float Design
designed so that the viscous drag area is relatively small, so The single-float viscometer is a continuous and direct reading
the float is relatively insensitive to viscosity while being viscosity instrument. As illustrated in Figure 8.64i, a positive
sensitive to flow rate and density changes. displacement pump (other flow control devices can also be used)
In the viscometer version of this design, the flow rate provides the constant sample flow rate through the instrument.
through the variable area meter is held constant. Therefore,
if the position of the float changes, that change is an indica-
THERMOMETER
tion of a change in fluid viscosity and density or in kinematic MAIN LINE
viscosity. To increase its sensitivity, the viscometer float is
designed with a large viscous drag area. To obtain accurate
THERMOMETER
viscosity measurements with float-type viscometers, the flow
rate must be constant. Because this required flow control can
be obtained in three different ways, there are three different FLOAT

designs: single-float, two-float, and concentric.


SINGLE FLOAT
VISCOMETER
Precautions

When using float type viscometers, the following recommen- PRECISION


POSITIVE DISPLACEMENT
dations should be observed: METERING PUMP

1. The float viscometer must be installed vertically, with FIG. 8.64i


the outlet at the top. Typical piping arrangement of a single-float viscometer.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1731

The recommended flow rate is between 0.75 and 2.0 GPM pressure differential to provide sufficient flow through the
(2.9 and 7.6 l/m). Automatic temperature compensation is instrument. The required flow rate is 0.25 to 2.5 GPM (0.95
not available for this type of direct reading viscometer. If a to 9.5 l/m), depending on size.
metering pump is used to generate the constant flow rate, the The throttling valve or orifice plate is not required if the
temperature rise through the pump should be measured and outlet can be discharged to a location where the pressure is
corrected for by the use of a viscosity vs. temperature curve. lower than that of the main pipe. A needle valve should be
The single-float viscometer is also available as a transmit- used on the viscometer inlet to allow sensitive and accurate
ter that is used when remote displays or controls are required. flow rate adjustment. As a function of size, the unit is rated
In this case, the float position is detected by the use of an for pressures up to 300 PSIG at 450°F (2 MPa at 232°C),
armature attached to the float extension rod with a magnetic depending on size.
sensing device around its outer periphery. The two-float viscometer can measure the viscosity of
−4
The glass tube viscometer is rated for 450°F (232°C) newtonian fluids from 0.3 to 250 cP (10 to 0.25 Pa⋅s). It
temperature and for 90 PSIG (621 kPa) pressure. With steel- has a rangeability of 10:1 and an inaccuracy of ±4% of
tube viscometers, the pressure rating depends on the operat- indication when the viscosity is higher than 35 cP (0.035
ing temperature, such as a unit for 650 PSIG service at 450°F Pa⋅s). When measuring lower viscosities, the error is ±2%.
(4.5 MPa at 232°C).
The single-float viscometer can be used to measure the Concentric Design
viscosity of non-Newtonian fluids at less that 400 cP (0.4 Pa·s)
The concentric viscometer (Figure 8.64k) consists of a dif-
and can handle newtonian fluids up to 10,000 cP (10 Pa⋅s).
ferential pressure regulator that maintains a constant pressure
The rangeability of this viscometer is between 3:1 and 6:1.
drop across the meter, and a variable area flow meter, which
The inaccuracy of this sensor is ±4% of actual indication,
is provided with a viscosity-sensitivity float. As the fluid
and the reproducibility is ±4% of indication.
enters the instrument, it splits into two streams.
The portion of the fluid that flows upward on the outside,
Two-Float Design
around the differential pressure float, is used to control the
This is a relatively low-cost viscometer designed to provide pressure drop across the meter. The upper end of the differ-
intermittent viscosity measurement in the laboratory, on the ential pressure float acts as a control valve so that, when the
test bench, or in industry. It is only for local indication, not
for transmission.
The design incorporates two floats. The upper float is EXTENSION WELL

sensitive to fluid flow rate, and the lower is sensitive to FLOAT EXTENSION
viscosity. When making a measurement, the fluid flow rate
is first manually adjusted to a constant value as indicated by
the position of the upper float. Under such conditions, but
only while maintaining the volumetric flow constant, the
position of the other float indicates the viscosity of the fluid OUTLET
on a direct reading scale. VISCOSITY FLOAT
The recommended piping configuration for a two-float
viscometer is shown in Figure 8.64j. In the main pipeline, a
throttling valve or an orifice plate can be used between the
viscometer inlet and outlet connections to produce the required

THROTTLING VALVE OR ORIFICE


MAIN LINE

DIFFERENTIAL PRESSURE
FLOAT

INLET
VISCOMETER TUBE
FLOW RATE FLOAT

NEEDLE
VALVE VISCOSITY FLOAT

TWO-FLOAT VISCOMETER

THERMOMETER ORIFICE

FIG. 8.64j FIG. 8.64k


Typical bypass piping arrangement of a two-float viscometer. Cross-sectional view of a concentric viscometer.

© 2003 by Béla Lipták


1732 Analytical Instrumentation

the valve until the viscometer reading levels off. Further


100
opening will cause the reading to start to fall off and then,
% CHART INDICATION

80 when turbulent conditions evolve, rise sharply, as illustrated


in Figure 8.641.
60
RECOMMENDED The valve openings that produce steady viscosity read-
RANGE ings are in the operating range. This procedure should be
40
carried out when the main line flow is at a maximum; after
20
that, the valve should be locked in place to prevent unautho-
0 rized adjustments. If the main line flow rate is lower than 2
0.1 0.2 0.4 0.6 1 2 4 6 10 20 40 60 100
TOTAL FLOW RATE, GPM GPM (7.6 l/m), a booster pump should be used to provide
adequate flow rate through the meter. The unit is rated for
FIG. 8.64l 650 PSIG pressure at a temperature of 450°F (4.5 MPa at
The characteristics viscosity reading of a concentric viscometer as 232°C).
a function of flow rate through the meter. The concentric viscometer can measure viscosities in the
−4
range of 0.5 to 550 cP (5 × 10 to 0.55 Pa⋅s) and is available
RESTRICTION (ORIFICE OR VALVE) with a span range of 10:1. Its inaccuracy is ±2% of indication
for viscosities lower than 35 cP (0.035 Pa⋅s), and it is ±4%
MAIN LINE
for higher viscosities.

OSCILLATING VISCOMETERS
INDICATING
TRANSMITT
Oscillating Blade

Figure 8.64n illustrates the oscillating blade viscometer in


BY-PASS which, as the blade moves between the fixed side walls, it
VALVE
VISCOMETER
discharges the process fluid through the openings of one wall
of the measuring chamber while pulling in a fresh sample
through the other. The blade is rotated around its fulcrum by
a plunger/solenoid mechanism as its tip travels the distance
between points A and B. When the stroke is completed, the
FIG. 8.64m polarity of the current to the plunger coil is reversed, and the
Bypass installation of a concentric viscometer.
blade is returned to its original position.
Adjustable mechanical stops are provided to limit the
flow rate changes, it throttles the flow to maintain a constant stroke. The time of travel between points A and B is detected
pressure drop. The pressure drop across this portion of the optically as, upon the completion of the stroke, a beam inter-
meter is determined by the weight of the d/p float. rupter blocks a light beam. The total stroke time is about
The portion of the fluid that flows downward enters the 0.2 sec, and the blade makes a stroke about every 2 sec. The
viscometer tube through an orifice and then passes the viscosity- average time of several strokes (adjustable) is measured as
sensitive float. The flow rate through the viscometer tube is an indication of viscosity.
maintained constant, because the fluid passes through the This intrinsically safe, indicating in-line transmitter can
orifice at a fixed pressure drop. Thus, constant flow rate is detect viscosities between 10 and 200,000 cP. It has a mini-
maintained, which is a requirement for the measurement of mum span of 20 cP and a maximum span of 200,000 cP. The
viscosity. maximum allowable liquid velocity in the process pipe can
An extension attached to the viscosity float transmits its be up to 6.5 ft/sec (2 m/s), but the fluid should not contain
movement through a magnetic coupling to a receiver for particles that are larger than 0.04 in. (1 mm) in diameter.
display, recording, or automatic process control, as desired.
The unit may be placed in either the main process line Oscillating Piston
or in a sample line, depending on the process flow rates. As
shown in Figure 8.641, this viscometer does have a flow- The unit illustrated in Figure 8.64o can be used as an on-line
insensitive range, so the flow rate through it should not industrial viscometer. In this design, the alternate energization
exceed 30 GPM (114 l/m). of two coils (A and B) causes a piston, located in the inner
The recommended piping configuration for bypass instal- cavity of the probe, to move back and forth within the cavity.
lation is shown in Figure 8.64m. The restriction orifice shown The viscosity of the process fluid resists the movement of the
in this figure should be sized for approximately 4 PSI (28 kPa) piston, so the travel time of the piston increases as the process
pressure drop. If a valve is used in place of an orifice, the viscosity rises. This travel time is measured as an indication
correct valve position can be determined by slowly opening of viscosity. About 30 sec is needed to make a measurement.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1733

PLUNGER COIL
OPTICAL SENSOR
BEAM INTERRUPTER SOLENOID HOUSING

225
MECHANICAL 160 9" 65
STOPS 6.3" 2.5"

SEALING RING FULCRUM

BLADE

FIXED SIDE WALLS Min 100


4"

FIG. 8.64n
Oscillating-blade-type viscosity transmitter. (Courtesy of BTG Inc.)

11/4" NPT
1/ " PIPE 1/ " NPT
2 2
WITH ELBOW

1.48"
(37.6 mm)

11/4" NPT(M)
1.75"
(44.5 mm)

2.6"
(66 mm)
MIXING BLADE
RESERVOIR

COILS SENSOR

A B

WIRES

A B

DEFLECTOR
PLATE OSCILLATING
PISTON

FIG. 8.64o
As the piston is made to oscillate by coils A and B, which are alternately energized, the time of piston travel is registered as viscosity.
(Courtesy of Cambridge Applied Systems Inc.)

This viscometer is provided with an integral resistance It can be provided with a choice of pistons to cover
temperature detector that provides temperature compensa- viscosity ranges from 0.1 to 2 cP up to 1000 to 20,000 cP.
tion. The oscillating piston viscometer can operate at high To use this sensor, the process fluid should be free of solid
pressures (up to 10,000 PSIG, or 700 barg) and at high particles that are larger than about 25 µm in size. The inac-
temperatures (600°F, or 315°C). The wetted parts of the curacy claimed is 2% of actual reading or better. The elec-
sensor are made of stainless steel and Teflon. tronics are available in explosion-proof construction and can

© 2003 by Béla Lipták


1734 Analytical Instrumentation

PROCESS
FLUID 11/4" (32 mm)
TEE
PROCESS
FLUID

SMALL LINES LARGE LINES


AND TANKS

FIG. 8.64p
Pipeline installation of the oscillating-piston viscometer. (Courtesy of Cambridge Applied Systems Inc.)

PUSH PULL
DRIVER AND
PICKUP

DETECTOR DRIVER
COIL COIL

SENSOR THERMOCOUPLE

FIG. 8.64q
Torsional oscillation-type viscometers are available with spherical and rod-type probe sensors. (Courtesy of Nametre Co.)

be either the local indicator or the transmitter type. The pipe- the prevailing viscosity, the error can be reduced to 2% of
line installation of this detector is illustrated in Figure 8.64p. actual reading. In addition, a compensating filter is provided
to minimize the effect of pipeline vibrations.
Torsional Oscillation Design The viscometer reads the product of absolute viscosity
(cP) and density. To correct for density variations, the density
The sensor in a torsional oscillation-type viscometer can either of the process fluid can be manually dialed in or automati-
be spherical or rod-shaped. The spherical elements can be used cally measured and compensated. These transmitters are
between 0.1 and 200,000 cP, and the rod-shaped ones from 10 available in explosion-proof construction and can be provided
to 2,000,000 cP. Both can be exposed to high pressures (5000 with threaded (1.5-in. NPT) or flanged (2-in. Class 150 ANSI)
PSIG, or 352 barg) and high temperatures (850°F, or 454°C). connections.
As shown in Figure 8.64q, the sensor is electromechan-
ically oscillated in torsion by a rigid inner shaft around its
central axis. The amplitude of oscillation is only 1 µ m, and
the power required to maintain that amplitude is proportional PLASTOMETERS
to viscosity. To provide temperature compensation, the tem-
perature of the process fluid is detected by a thermocouple, Plastometers are employed to study the plastic behavior (melt
which is inserted through a central hole in the shaft. flow) of plastic materials or the molecular weight distribution
To minimize inaccuracy (error), automatic range switching of polymers. Some materials do not begin to flow unless the
is available. By always selecting the smallest span to measure stress has exceeded their yield stress value. Such materials

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1735

behave as elastic bodies that yield with time as the stresses THERMOCOUPLE OR DRIVING SHAFT
THERMOMETER
exceed the yield limit. This phenomenon is called plastic HEATER
behavior.
UPPER DIE PLATEN

ROTOR
Plastic Behavior LOWER DIE PLATEN

SAMPLE CHAMBER
Prof. Eugene Bingham assumed that the flow rate for such
materials was zero when they experienced stresses below a TORQUE SENSOR
critical value and was linear with stresses above this limit
value. A body that behaves in such a manner is called a FIG. 8.64r
Bingham solid or an ideal plastic. Extremely few solids Cone-and-plate plastometer.
behave in this manner. For most plastic materials, the yield
value is not sharp and shows a nonlinear behavior with a After the warm-up period has been completed, the plastic
delayed and imperfect recovery. The different types of plas- behavior of the sample is tested by driving the rotor at either
tics behave differently, and their behavior also changes with a constant speed (normally at 2 r/min) or at various speeds
temperature. between 0.05 to 2 r/min to test for relative molecular weight.
For this reason, a plastometer gives only a relative indi- The shearing action that takes place between the rotor and
cation of polymer behavior, thus characterizing a plastic the die cavity is measured by the deflection of a calibrated
material. Because the shear stress/shear rate relationship is U-spring, which is attached to the torque sensing rotor.
based on the molecular behavior of materials, the plasticity The deflection of the U-spring is directly proportional to
can be related to the molecular weight of a material or to its the shearing torque of the specimen being tested. An elec-
distribution. The plasticity of plastic materials is expressed tronic strain gauge may be attached to the U-spring to trans-
in arbitrary units such as ASTM melt index (MI), Canadian mit this signal for continuous recording. The plastometer
Industries Limited (CIL) flow index, Mooney points, or per- operates within a range of 0 to 200 (Mooney) units and has
cent of full scale. a repeatability of ±0.5 units.
In general, the plastometer consists of a heating chamber
to keep the sample at constant temperature, a mechanism to The Unit of Mooney Dr. Mooney gives the following equa-
apply a high shear force, and a device to measure torque or tion to determine the average viscosity ( µ ) of solids:
flow rate of the sample material and thereby detect the plas-
ticity of the sample. Plastometers can be used in a laboratory
G 188.44gθ
(Section 8.63) to study polymeric behavior or used on line µ= 8.64(3)
to control processes manually or automatically. This type of 2πωR 3 ( R/2a + h/b)
rheometer is an invaluable tool in manufacturing plastics and
synthetic rubbers. where
G = gauge reading
Cone-and-Plate Plastometer g = acceleration from gravity
θ = pitch radius of worn gear
The cone-and-plate plastometer is used to conduct tests in ω = angular velocity of rotor, radians/sec
evaluating crude rubber, rubber compounds, or reclaims; in R = radius of rotor, inches
the control of mill breakdown of polymer molecules; in the a = vertical clearance between rotor and stator above or
determination of time to scorch; in the calculation of optimal below the rotor, inches
cure time; and for the evaluation of the processing charac- h = thickness of rotor, inches
teristics of plastics. b = effective radial clearance between rotor and stator
This plastometer incorporates the features of the Mooney
plastometer and is designed to meet the requirements of Kneader Plastometer
ASTM standard test method D-1646. The working principle
of this sensor is the same as discussed in connection with This instrument is widely used in measuring the plastic char-
laboratory-type cone-and-plate viscometers (Figure 8.63r). acteristics of plastic, rubber, food, pigment, cement, paint,
As shown in Figure 8.64r, this detector is designed to cosmetics, and coating products. This type of plastometer is
eliminate polymer “ball-up” and slippage tendencies by con- equally suitable for both laboratory and on-line measurement
fining the sample in a disc-shaped cavity. Machined serra- and the control of highly viscous flowing materials. The chief
tions provided on all of the die and rotor surfaces prevent advantage of this instrument is that it measures plastic behav-
slippage. In operation, the sample is introduced into a cylin- ior of melt viscosity under conditions that are very similar
drical test chamber where integral heaters are used to bring to those that exist in the processing equipment.
it to the predetermined test temperature (up to 400°F, or Many different shapes of measuring heads are available
204°C). (see Figure 8.64s) to accommodate a wide range of viscosities.

© 2003 by Béla Lipták


1736 Analytical Instrumentation

THERMOSTATING JACKET
HEATING
MEDIUM
EXTRUDER FLOW
OUT

HEATING
MEDIUM
SELECTOR
IN
SIGMA BLADE DELTA BLADE CAM BLADE VALVE
FOR PLASTIC FOR FOOD, FOR RUBBERS &
DOUGHTY MATERIALS TOOTH PASTE, PLASTICS
FATS AND
METERING
VASELINE TYPE
PUMP

DC
FIG. 8.64s MOTOR
Measuring heads of kneader plastometer.
EXTRUDATE

GEAR
REDUCER
STRANDING DIE TO RECEIVER
TACHOMETER
EXTRUDER FLOW

FIG. 8.64u
CIL PRESSURE Flow-sensing capillary plastometer.
TRANSDUCER
SS

MI PRESSURE
polyethylene through an open-ended capillary at 190°C
TRANSDUCER METERING [374°F] and 43.2 PSIG [298 kPa] pressure). It can also record
PUMP
CIL flow index (190°C and 1500 PSIG [10 MPa] for polypro-
PRC DC pylene) or can alternately record both ASTM and CIL melt
MOTOR
MI & CIL indexes.
SELECTOR These two measurements for a particular polymer can be
VALVE
GEAR
interpreted as both a molecular weight parameter and as a
REDUCER molecular weight distribution parameter. The capillary-type
TO plastometer is not as versatile as the previously described
TACHOMETER
RECEIVER
plastometers.
In feeding the capillary type plastometer, the polymer
FIG. 8.64t melt is first conditioned to be at some specific temperature
Pressure-controlled capillary plastomer. and pressure. It is then extruded through a capillary of suit-
able dimensions. The rate of flow of the polymer through the
capillary is measured by detecting the output of the tachom-
In this plastometer, the jacket heater keeps the sample at a eter on the metering pump and is recorded in units appropri-
constant temperature (up to 570°F, or 299°C) by the jacketed ate for the test (Figure 8.64u).
heater. The kneader is driven by a dynamometer (available The operating range of this plastometer is 0 to 200 MI
with variable-speed and fixed-speed drives). The resistance and 0 to 100 CIL units, with a repeatability within ±2% of
encountered by the mixing blades is detected and transmitted full scale. The unit is designed to operate at up to 570°F
to the dynamometer housing, which tends to rotate in a direc- (299°C) and up to 5000 PSIG (35 MPa) in pressure.
tion opposite to that of the driving shaft.
The torque is transmitted for recording or to automati- Advantages One advantage of this instrument is that it can
cally control processes with medium-viscosity, free-flowing be used as an automatic on-line process control device for
fluids. The readings and recordings are on a 1000-unit divi- both viscous fluids and plastic solids. Especially for the solu-
sion arbitrary scale, used for indicating shearing torque of tion polymerization processes, the polymer solution can be
the specimen being tested. Repeatability is ±1% of full scale. directly analyzed in this plastometer as the sample is received
The measurement range can be adjusted. from the processing reactors through an autosampling device
that flashes the solvent and the unreacted monomers, and
Capillary Plastometer melts the polymer prior to sending it to the plastometer.
For plastic solids, the die unit (capillary) can be mounted
This plastometer is designed for use in polymer manufactur- to the pelletizing extruder of the process stream for measure-
ing plants and is based on the capillary-tube viscometer prin- ment. The capillary type plastometer is ideally suited for
ciple. The capillary plastometer shown in Figure 8.64t is study of plastic behavior of materials that are processed
calibrated to record either ASTM melt index (flow rate of through injection molding or finishing operations.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1737

Limitations The following conditions might degrade the accu- the viscous drag on the spindle. At this point, the cage and
racy of the measurements obtained using capillary-extrusion spindle both rotate at the same speed but with a definite angular
viscometers: relationship to each other. This angular relationship is pro-
portional to the torque on the spring.
1. Nonuniformity of sample shear rate Two methods are used to convert the angular relationship
2. Entrance effects—energy losses at the capillary into a viscosity reading. One method is shown in Figure 8.64v,
entrance and discharge points where one side of a variable capacitor is attached to the cage
3. Compressibility of fluids and the other to the spindle arm. A capacitance is thus made
4. Pressure loss produced by the flow in the sample proportional to the angular relationship between cage and
chamber spindle.
5. Temperature gradient created by shear-induced heat The other method is based on a potentiometer mounted
in the cage with its free member connected to the spindle
When performing scientific research, great care should arm. In this case, a resistance is made that is directly propor-
be taken to correct for these errors. If that is done, capillary- tional to the angular relationship between cage and spindle.
extrusion viscometers have excellent reproducibility and are In either case, the measured signal is transmitted to a receiver
well suited for routine industrial and scientific work. and is converted into viscosity units.
When measuring low viscosities, the variable capacitor
type design is preferred. This type of viscometer can measure
viscosity ranges from 0–10 cP (0–0.01 Pa⋅s) to 0–50,000 cP
ROTARY VISCOMETERS
(0–50 Pa⋅s). Its inaccuracy is ±1% of full scale between the
ranges of 0 to 100 and 0 to 50,000 cP (0 to 0.1 and 0 to
Rotating-cone viscometers have been applied chiefly to mea- 50 Pa⋅s). Its repeatability is ±0.3% of full scale for all ranges.
sure non-Newtonian fluids such as paints, printing ink, starch,
The response time of this viscometer is less than 30 sec
size solution, and varnishes. They are designed to operate
for detecting a full scale change in viscosity. The measure-
continuously in industrial environments. They measure vis-
ment range of this viscometer can be varied over a 30:1 ratio
cosity by detecting the torque required to continuously rotate
by changing the size or shape of the spindle. The standard
a cone, disk, sphere, or spindle in a liquid. In this type of
spindle speed is 50 r/min. In some installations, the maximum
process viscometer application, the sample is continuously
range can be increased to 0 to 100,000 cP (0 to 100 Pa⋅s).
replaced and is subjected to a constant shear rate; thus, the
measurement of non-Newtonian apparent viscosity is possi-
Installation This viscometer may be installed on a tank, in
ble, as well as absolute viscosity measurement of newtonian
an open vessel, or in a sampling line (Figure 8.64w). For in-
fluids.
line installations, the flow rate of the process liquid should
be constant, nonpulsating, and laminar (less than 3 GPM
Rotary Spindle Design [11.4 l/m] for a standard 4-in. [101.6-mm] pipe). Swirling
In the design illustrated in Figure 8.64v, a synchronous induc- motion in the measurement chamber may be reduced by
tion motor (for safety) drives a cage coupled through a cal- installation of a spider-type deflector in the inlet port.
ibrated spring to a spindle arm that supports the spindle or The viscometer head is designed to operate at tempera-
cylinder in the fluid being measured. During measurement, tures up to 160°F (71°C) and at pressure ranges of 8 mm Hg
the spring tends to wind up until the force it generates equals absolute to 100 PSIG (1 to 690 kPa). The head should be
3
purged with dry air or other inert gases at a rate of 0.5 ft /min

SPEED HOUSING FOR


REDUCTION SYNCHRONOUS MOTOR DRIVER & CAPACITOR
GEARS
ROTOR & STATOR BRUSHES
SIGNAL
VARIABLE CAPACITOR PLATES
TO RECEIVER
CALIBRATED SPRING

CAGE BAFFLE TUBE


FRICTION CLUTCH

SPINDLE
SPINDLE EXTENSION

FLOW
SPINDLE
(A) ON A REACTOR, TANK OR VESSEL (B) IN A SAMPLING LINE

FIG. 8.64v FIG. 8.64w


The operation of the rotating-spindle-type industrial viscometer. Tank and pipeline installation of the rotating spindle viscometer.

© 2003 by Béla Lipták


1738 Analytical Instrumentation

−4 3
(2.4 × 10 m /s) or greater. A positive pressure differential CONNECTING ROD
TO MEASURING
of 3 PSI (21 kPa) should be maintained between the viscom- POTENTIOMETER
eter and the process vessel to prevent liquid or vapor pene-
SYNCHRONOUS
tration into the head. MOTOR
The temperature of the mounting flange on which the
viscometer is placed should not exceed 600°F (316°C).
Above 220°F (104°C), it is necessary to use a cooling pad BALL BEARING
STAINLESS STEEL/
and place it between the flange and the viscometer. The SAPPHIRE BEARING
MAGNET (ROTATING)
process liquid temperature can be up to 2500°F (1370°C). NICKEL ROD
(ARMATURE OF
COUPLING)
Performance The rotating-cone viscometer is not affected
by normal industrial vibration. It is provided with overrange SAMPLE IN
protection through a friction clutch on the spindle extension.
It is simple to clean the spindle. The sample fluid should be
maintained at a constant temperature, or the viscosity mea- MEASURING BODY
(ROTATING DISC
surement should be compensated for temperature variations. MEASURING IN LIQUID)
Logarithmic temperature-compensating units are available CELL
LOWER BODY BEARING
and are matched to the viscosity–temperature relationship
shown by the particular process fluid. Logarithmic compen- FIG. 8.64x
sation to within 1% error over a span of 10°F (6°C) can be Magnetically coupled rotational industrial viscometer.
provided.
A baffle tube should always be installed when this viscom-
eter is used in an agitated tank or reactor to obtain laminar
flow condition. However, if very violent agitation or bubbling
conditions exist, this viscometer should not be used at all.
The spindle should always be immersed, and the liquid level
should be kept constant to obtain reproducible readings.
The frequency of the power supply should be checked.
Wide deviations from the specified frequency will cause spin-
dle speed changes and can introduce errors in the viscosity
measurement of non-Newtonian fluids, because their appar-
ent viscosity is related to shear rate.
The rotating-cone viscometer is an electromechanical
device, so a monthly preventive maintenance and cleaning FIG. 8.64y
schedule is recommended. If the unit is operated continu- Pipeline and vessel installations of the gyrating element viscometer.
ously, the speed reduction gears will need to be replaced on (Courtesy of C. W. Brabender Instruments Inc.)
a yearly basis.
Because of its simple design, ease of cleaning, and non- The magnetically coupled viscometer should not be used
clogging features, this viscometer has been successfully used to measure fluids that contain fiber, ferrite, or abrasive mate-
in monitoring solid and liquid blending processes. Rotating- rials, because they interfere with the operation of the mag-
cone viscometers are quite versatile in their application over netic coupling or can damage the stainless steel–sapphire
wide ranges of viscosities and are equally applicable for bearings.
measuring the viscosities of newtonian and non-Newtonian
fluids. Gyrating-Element Viscometer

The viscometer illustrated in Figure 8.64y differs from the


previously described designs in that its element gyrates
Magnetically Coupled Viscometer
instead of rotates. This gyrating motion is the result of the
In the rotating-cone viscometer shown in Figure 8.64x, a drive shaft not being straight. The drive and the torque-
magnetic coupling is provided between the electronic detec- sensing mechanism are isolated from the process side by
tor (which is at atmospheric pressure) and the rotating sensor stainless steel bellows that are capable of operating at up to
in the measuring cell (which is operated at the process pres- 1000 PSIG (70 barg) operating pressures and 572°F (300°C)
sure). With this separation between atmospheric and pressur- operating temperatures.
ized areas, the requirement for purging is eliminated and, in The drive motor speed is adjustable from 15 to 120 r/min,
the measuring cell, operating conditions can be up to 2850 and the unit can detect viscosities between 10 and 10 million
PSIG (19.7 MPa) at 20°C or 1620 PSIG (11.2 MPa) at 300°C. cP (the overall range is subdivided into several working ranges).

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1739

AGITATOR HORSE POWER LIMIT Double-Cylinder Viscometer

Figure 8.64aa illustrates the double-cylinder type of in-line


viscometer. It consists of a stationary outer and a rotating
inner cylinder. This viscometer provides self-cleaning oper-
ation and fast measurement.
The process fluid is drawn into the measuring chamber
VISCOSITY

through the apertures in the stationary outer cylinder by the


propeller. The propeller pumps the process fluid downward
RECOMMENDED into the measurement zone, where the sensor cylinder is
OPERATING RANGE rotated inside the stationary external cylinder. The measuring
zone is the clearance space between the two cylinders. The
torque required to rotate the measuring cylinder is propor-
tional to viscosity.
TORQUE The self-cleaning nature of this viscometer is partially
created by the lip on the upper edge of the measuring cylin-
FIG. 8.64z der, which serves to prevent the entry of larger particles into
Relationship between viscosity and torque in agitator power the measurement zone. In addition, the solids that enter the
viscometer. measurement zone can also leave through two slots that are
provided in the outer cylinder. Large particles will not enter
the measuring zone at all but will leave through the outlet
For pipeline installation, the minimum pipe size is 4 in.
holes that are provided on the top of the rotating cylinder.
(100 mm).
Viscometers are available for in-line installation in pipe
sizes from 3 to 10 in. (76 to 250 mm) in diameter. The
Agitator Power
viscometer can detect viscosities from 5 to 350 cP at pres-
This type of instrument is widely used in the paper industry sures up to 145 PSIG (10 barg) and temperatures up to 300°F
to control and measure the consistency of paper pulp slurries. (150°C). The wetted parts can be made out of stainless steel
In this viscometer, instead of measuring the torque exerted on or Hastelloy C with Viton, EPDM, or silicone O-ring seals.
a rotating cone, a transmitting wattmeter is used to detect the The mechanical seal can be cooled and lubricated by water.
torque exerted on the process agitator. It measures the power
consumed as the agitator is being driven in a mixing tank.
As shown in Figure 8.64z, because most of the industrial VIBRATING-REED VISCOMETER
agitator motors are oversized, the torque response of this
sensor is poor at low viscosities. The size of the motor and The vibrating-reed viscometer is designed for continuous
impeller should be selected so that the agitator will operate measurement. It can be installed directly into the process
in a region where the viscosity-vs.-torque relationship is lin- vessels or pipelines. As illustrated in Figure 8.64bb, it con-
ear. The motor, reducer, bearing assembly, and pressure seal sists of a frequency generator, a vibrating spring rod, a probe,
are all part of the viscometer, and any change in character- and a pickup unit to complete a measurement loop through
istics of these components would also affect the power con- the process material.
sumption of the motor and, in turn, the viscosity reading. The drive coil is excited at a frequency of 120 Hz from
This instrument is simple and easy to install, but it has a 60-Hz input frequency. This produces a pulsating magnetic
low sensitivity and a low ratio between agitator power con- field that causes the drive armature to vibrate at the same
sumption and changes in viscosity of the fluid. Many different frequency. The mechanical vibration of the drive armature is
impeller designs are available to improve the sensitivity and transmitted to the probe through an all-welded, pressure-
rangeability. Because it is a self-cleaning and agitating sealed node that is located where the amplitude of vibration
design, it is ideal for materials that have a tendency to cling is zero.
to slow-moving parts or to settle out from suspensions. The principle of operation is based on the fact that the
This type of instrument should not be used with fluids amplitude of probe vibration depends on the viscosity of the
that are thixotropic or rheopectic (materials whose viscosity process media. The resistance to the shearing action caused
changes with duration of agitation or shear). However, it can by the probe’s vibration increases with an increase in the
be used in in-line applications for all non-Newtonian fluids, viscosity of the process media. The amplitude of probe vibra-
because the process fluid is continuously replaced by a fresh tion is sent through a second welded node point by the upper
sample at a reasonably constant flow rate. spring rod to the pickup end of the detector.
The accuracy and sensitivity of this method of viscosity The pickup end is similar to that of the driver end, except
detection are both poor, but its repeatability is reasonable at that a permanent magnet is used to induce a 120-Hz AC voltage
about ±1% of full scale. in the pickup coil from the vibration of the pickup armature.

© 2003 by Béla Lipták


1740 Analytical Instrumentation

RECEIVER

CHECK VALVE WATER FOR COOLING AND


FLUSH
LUBRICATION OF MECHANICAL SEAL
FLUSHING OF BY-PASS PIPE

PROCESS FLUID

SEALING
ELEMENT

OUTLET HOLE
APERTURES ON
EXTERNAL CYLINDER

PROPELLER

MEASURING
ZONE

ROTATING
MEASURING
STATIONARY CYLINDER
EXTERNAL
CYLINDER

FIG. 8.64aa
Self-cleaning cylinder viscometer. (Courtesy of BIG Inc.)

DRIVE COIL
PICKUP END
DETECTOR OUTPUT SIGNAL

800
DRIVER END
3/4"
N.P.T. 600
(MVAC)

PROBE (1) (2) (3) (4)


400

9/16" 200
1 10 100 1k 10k 100k
NODE LOG. VISCOSITY (CENTIPOISE)
POINT DRIVE
1/2" CONDUIT ARMATURE SOFT IRON FIG. 8.64cc
HUB STATOR
The response curves of four vibrating-reed viscometers.
FIG. 8.64bb
Vibrating-reed viscometer.
Typical response curves of four different detector units are
shown in Figure 8.64cc.
The magnitude of the voltage generated in the pickup coil is Because the viscosity of fluids depends on their temper-
a measure of the process viscosity, because it is proportional ature and pressure, it is mandatory to maintain the process
to the pickup armature vibration. The 120-Hz output signal pressure and temperature constant. As shown in Figure
from the detector is converted into a 0 to 10 mV DC signal 8.64dd, if the fluid temperature must be allowed to fluctuate,
to indicate, record, or control the viscosity of newtonian or temperature compensation is required. Before ordering a vis-
non-Newtonian fluids. cometer for in-process use, the fluid’s viscosity must be mea-
sured as a function of temperature so that this characteristic
Performance Inaccuracy of the instrument is about ±1% of curve can be used in the converter.
full scale, and reproducibility is a little better. There are five For accurate measurements, laminar flow conditions should
different detector units to cover the viscosity ranges from be maintained. Therefore, the installation of this type of
−1 5 −4 2
10 to 10 cP (10 to 10 Pa⋅s) at relatively low shear rate. viscometer directly in an agitated vessel is not recommended.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1741

VISCOSITY DETECTOR vibration frequency is close to 120 Hz, it can interfere with
the reliability of viscosity measurement by this sensor.

CONVERTER RECORDER

CORIOLIS MASS FLOW METER


TEMPERATURE
ELEMENTS Measuring Tube Torsional Movement
SPAN ADJUSTOR
ZERO ADJUSTOR Coriolis mass flow meters currently are well established in
process industry applications and have extremely good func-
FIG. 8.64dd tionality and performance. They have proven multivariable
Complete loop diagram of a temperature-compensated vibrating- measurement capabilities, including mass flow, density, and
reed-type viscosity recorder.
temperature, that can be transmitted via various methods such
as 4 to 20 mA signals, pulse, and various fieldbus links. Ease
FLUID OUT of installation and maintenance are also well established.
Further details on these flow meters are available in another
section of this volume.
Very recently (at the time of the writing of this volume),
VISCOSITY a new development was announced by a manufacturer of
DETECTOR
Coriolis meters. Viscosity can now be added to the list of
multivariable measurements. Viscosity is obviously a key
process variable, and to be able to acquire it via an existing
proven technique is significant. Four significant process vari-
TEMPERATURE ables can be measured simultaneously with proven in-line
SENSING
ELEMENT technology.
FLUID IN
The measurement principle for this viscosity process
FIG. 8.64ee variable is the torsional movement of the measuring tube.
Installation recommendation for an in-line vibrating-reed viscometer. For an optimal balance, and for isolation from external influ-
ences, a pendulum is mounted on the single measuring tube
that causes an additional torsional movement on it, as shown
If it must done, the viscosity reading should be taken only
in Figures 8.64ff and 8.64gg. Based on this technique, tor-
when the agitator is off.
sional movements are created in addition to the lateral oscil-
lation movements of the measuring tube.
Installation and Calibration Complete immersion of the
A velocity profile is created on the tube cross section by
probe in the process fluid is essential. Figure 8.64ee illus-
the torsional and lateral oscillation movements. This is shown
trates the recommended pipeline installation of this viscom- in Figure 8.64hh. Viscosity measurement is based on the
eter probe in vertical upward flow. Complete immersion of shape of the velocity profile, shear forces, and corresponding
the probe is required at all times. To improve the accuracy
of measurement, the recommended viscometer installation is
in a temperature, pressure, and flow rate controlled sample
loop.
The amplitude of vibration will change with condition
and age of the probe. It is good practice to calibrate the
instrument regularly when it is used for critical operations.
Material build-up on the probe can occur if the process mate-
rial has a tendency to adhere or if it contains long fibrous
materials. Slight or loose buildup can be removed by periodic
in-line purging. If the buildup can be severe, this type of
viscometer should not be used.
Because the detector unit transmits a low-voltage signal,
the use of shielded wires is recommended. The wires should
be grounded at both ends to minimize noise pickup. Supply Pendulum
voltage variations between 105 and 125 V will have no harm- FIG. 8.64ff
ful effects. The frequency of supply power should be checked, Mass flowmeter/measuring tube torsional movement type (pendulum
because the probe vibration frequency is controlled by the on the measuring tube balancing the system for immunity to external
line frequency. If the vessel or pipeline vibrates, and this forces). (Courtesy of Endress + Hauser.)

© 2003 by Béla Lipták


1742 Analytical Instrumentation

of a sample from the process (capillary, float, rotational), but


others can be inserted directly into the process pipe or vessel.
Because sampling systems usually increase installation and
maintenance costs, and because sampling also adds dead time
(transportation lag) to the measurement, the in-line designs
are generally preferred.
Pendulum
Viscous process fluids usually also contain solids. For this
reason, the designs that cannot tolerate the presence of solids
Measuring tube (capillary, float) are less desirable, and the self-cleaning designs
(oscillating, vibrational, double-cylinder) are preferred.
In some process applications, the operating viscosity can
FIG. 8.64gg change over a wide range. If that is the case, measurement error
Mass flowmeter/measuring tube torsional movement type (oscilla- can be reduced by using viscometer designs that are capable of
tion movement of the measuring tube and pendulum). (Courtesy of automatic range changing, e.g., the torsional oscillation design.
Endress + Hauser.) The torque or shear force measured by most viscometers
will also vary with changes in temperature and density. There-
fore, viscometer designs that provide automatic temperature
compensation and allow for manual or automatic density
correction are likely to provide more accurate readings.
A particularly appealing new approach involves using
existing Coriolis mass flow technology and extending the prin-
ciples and techniques to extract process viscosity. This yields
a proven industrial in-line continuous measurement technique
with three multivariable measurements that can spin out an
additional key process variable, viscosity. Installation is easy,
and no additional independent complex sensors are required.
FIG. 8.64hh In addition, one must also be concerned with the viscosity
Mass flowmeter/measuring tube torsional movement type (cross sec- range, design pressure, design temperature, and required pre-
tion of the tube, showing the velocity profile of the fluid). (Courtesy cision of the measurement. Table 8.64a provides a summary
of Endress + Hauser.) of the design features of the viscometers discussed in this
section.
shear rates developed in the fluid. This is similar to the
rotational element viscometers in which shear rates are
related to viscosity. Bibliography
The shear forces dampen the tube oscillation, so a larger
drive force is required to maintain tube oscillation. Thus, vis- Abnett, A. and Garvey, R., New orbital ball type of viscometer, Integrated
Monitoring, Diagnostics, and Failure Prevention—Proc. Joint Confer-
cosity can be determined by measuring the required power or
ence, Mobile, AL, NTIS Accession Number AD-P010 179/0/XAB,
excitation current for tube oscillation and torsional movement. April 1996.
Repeatability and measurement inaccuracy are both ±5% ASTM Standard, ASTM-D-445–97/XAB, Kinematic Viscosity of Transparent
of actual reading or ±0·5 cP, whichever is greater. The in-line and Opaque Liquids (the Calculation of Dynamic Viscosity), September
meter is 2 in. in diameter and is provided with 600-lb flanges. 1998.
Bandrup, J., Immergut, E. H., Grulke, E. A., and Bloch, D., Eds., Polymer
The output signal can be 4 to 20 mA or digital. The speed
Handbook, 4th ed., John Wiley & Sons, New York, 1999.
of response of the meter is high and therefore suited for Basker, V. R., Dutka, A. P., Crisalle, O. D., and Fricke, A.L., Evaluation of
closed-loop process control applications. an online torsional oscillatory viscometer for kraft black liquor, Paperi
ja Puu/Pulp and Timber, 82(7), October 2000.
Batra, R. L., Falling Cylinder Viscometer for Casson Fluids, Jamshedpur
CONCLUSIONS Regional Institute of Technology (India), 1981.
Black, S. S., Automated differential viscometry leads to 300 percent increase
in analysis rate, Sci. Comput. Autom., 5, April 1996.
Each viscometer exposes the process fluid to a different shear Bullon, H. H., Shear Rate Determination in a Concentric Cylinder Viscom-
rate, so, when the process fluid is non-Newtonian, the appar- eter, Defence Science and Technology Organization, NTIS accession
ent viscosities will vary with shear rate, and the readings will number AD-A315 366/5, July 1996.
vary with the viscometer used. Carr, K. R., Rotational Viscometer for High-Pressure, High-Temperature
Some viscometer designs provide intermittent readings Fluids, Patent Application 6–501–313, June 6, 1983.
Chang, V., Zambrano, A., Mena, M., and Millan, A., A sensor for online
(falling-ball or falling-piston, oscillating-blade, etc.), and measurement of the viscosity of non-Newtonian fluids using a neural
others are continuous. In closed-loop control, the continuous network approach, Sensors and Actuators A (Physical), A47(1–3),
ones are much preferred. Some viscometers require the taking March–April 1995.

© 2003 by Béla Lipták


8.64 Viscometers—Industrial 1743

Chu, B. and Wang, J., Magnet enhanced optical falling needle/sphere rhe- Khachatryan, G. M. et al., Device for the automatic continuous measurement
ometer, Rev. Sci. Instrum., 63(4), April 1992. of the dynamic viscosity of polymers, Fibre Chem. (UK), May–June
Cox, C. J., Park, M. J., and Wilson, C. A., Use of the capillary tube viscom- 1983.
eter for the determination of the flow behavior of tall oil soap, Trans. Kurano, Y., Accurate measurement of viscosity under high pressure with a
New Zealand Inst. Prof. Eng., March 1983. falling-sphere viscometer, High Temp.—High Press., 26(1), 1994.
Cutrone, L., Use of the Brookfield viscometer to predict rheological perfor- Ohene, F., Livingston, C., Matthews, C., and Rhone, Y., Study of Pressure
mance of coatings, J. Coatings Tech., January 1984. Drop in a Capillary Tube Viscometer for a Two-Phase Flow, Depart-
Dutka, A. P., Crisalle, O. D., Fricke, A. L., Kalotay, P., Evaluation of a ment of Energy Report #DOE/PC/91292-T17, NTIS accession number
capillary-coriolis instrument for online viscosity and density measure- DE96000151 1995.
ments of kraft black liquor, Proc. TAPPI Process Control, Electrical Oppliger, H. R., Matusik, F. J., and Fitzgerald, J. V., New technique accu-
and Instrumentation Conf. (ISA), March 1997. rately measures low viscosity on-line, Control Eng., July 1985.
Ferry, J. D., Oscillation viscometry—effects of shear rate and frequency, Roger, U., Hessenkemper, H., and Roth, P., Glass Conditioning by Viscosity
Meas. Control, September–October 1977. Control, Glass Sci. Tech., 69(8), August 1996.
Fitzgerald, J. V., Matusik, F. J., and Walsh, T. M., Inline viscometry, Meas. Sawhney, I. K., et al., Development of capillary tube viscometer for mea-
Control, December 1987. suring flow properties of khoa, J. Institute Eng. (India), August 1984.
Sexton, W. C., Rotating spindle viscometer, International and 42nd Annual
Fournier-Villalon, P., Bertrand, J., and Couderc, J. P., Using a Brookfield
Symposium and Exposition on Scientific Glassblowing, NTIS acces-
viscometer for determining rheological properties of newtonian, plastic
sion number DE970610182/XAB, June 1997.
and pseudoplastic media, Entropie, 20(120), 1984.
Sheen, S. H., Chien, H. T., and Raptis, A. C., In-Line Ultrasonic Viscometer,
Frisman, M. L., Rotation viscometer for readily unmixing suspensions,
1994 Review of Progress in Quantitative Nondestructive Evaluation
Indust. Lab. (USSR), October 1984.
Conference, Sponsored by Department of Energy, NTIS accession
Galvin, G. D., Hultin, J. F., and Jones, B., Development of a High-Pressure,
number DE95002982, October 1994.
High-Shear Rate Capillary Viscometer, J. Non-Newtonian Fluid
Smith, R. E., Brookfield viscometers for determination of low-shear viscos-
Mechanics, January 1981.
ity and leveling behavior, J. Coatings Tech., January 1984.
Gillis, K. A., Mehl, J. B., and Moldover, M. R., Greenspan Acoustic Vis- Villemaire, J. P. and Agassant, J. F., Apparent viscosity measurements using
cometer for Gases, Rev. Sci. Inst., 67(5), NTIS accession number a capillary viscometer, Polymer Process Eng., 1(3), 1983–84.
PB96–204417, May 1996. Viscometers, Meas. Control, June 1993.
Gimera, M., Software Quality Assurance Plan for Viscometer, Department Vrentas, J. S., Mechanical energy balances for a capillary viscometer, J.
of Energy Report, NTIS accession number DE95002145, October Non-Newtonian Fluid Mech., March 1983.
1994. Walsh, T. M., Continuous viscosity control improves quality, Adhesives Age,
Grupp, J., Torsion Viscometer for Liquid Crystals, Rev. Sci. Inst., June 1983. December 1989.
Helle, H., Valimaki, H., and Lekkala, J., Comparing a 10 MHz thickness- Webber, P. J. and Savage, J. A., Measurement of the viscosity of chalco-
shear mode quartz resonator with a commercial process viscometer in genide glasses by parallel plate technique, J. Mater. Sci., March 1981.
monitoring resol manufacturing process, Sensors and Actuators B Wunderlich, T. and Brunn, P. O., Ultrasound pulse doppler method as a
(Chemical), B81(2–3), January 2002. viscometer for process monitoring, Flow Meas. and Instrum., 10(4),
Jethra, R., Viscosity measurement, ISA Trans., 33(3), September 1994. December, 1999.
Kalotay, P., Online viscosity measurement using coriolis mass flowmeters, Zhang, Z. and Reddy, R. G., Viscosities of lead silicate slags, Minerals
Flow Meas. Instrum., 5(4), October 1994. Metall. Process., 19(1), February 2002.

© 2003 by Béla Lipták

Potrebbero piacerti anche