Sei sulla pagina 1di 27

MineARC

Aura-FX
Operation Manual
Aura-FX
Operation Manual

DISCLAIMER
The MineARC® Aura-FX Fixed Gas Monitor requires regular maintenance checks
to function correctly. Please consult the manual for periodical maintenance
requirements or contact MineARC’s service department to ensure the system is
being monitored correctly.

MineARC® shall not be liable for failure to adhere to the guidelines, warnings
and precautions provided in this manual.

MineARC® warrants its products against defective parts and workmanship for a
period of 12 months from date of purchase.

MineARC will repair or (at its option) replace any goods which are or may
become defective under proper use where such defects solely arise from faulty
design, materials or workmanship provided such defects are made aware to
MineARC within the requisite time limit.
Aura-FX
Operation Manual

SAFETY PRECAUTIONS
Read all instructions and warnings carefully. Follow all safety precautions to
avoid personal injury or property damage during system operation. MineARC ®
is not responsible for damage or injury resulting from unsafe product use, lack
of maintenance or incorrect product and/or system operation. Contact
MineARC® Systems when in doubt as to the safety precautions and operations.
Failure to comply with the following warnings could cause personal injury.

WARNING: Read and understand manual before operating


Aura-FX.

CAUTION: Only personnel trained in electrical equipment or


certified by MineARC® Systems should maintain Aura-FX.

CAUTION: 24-48 VDC power supply required. Disconnect


power from UPS Power source or power supply when
servicing electrical components.

CAUTION: This equipment is designed to operate only from


a fused 24-48 VDC power supply.

CAUTION: All sensing equipment will be subjective to some


fluctuation in sensitivity and require calibration. Calibration
shall be undertaken in the periods outlined in the manual for
each sensor type.
Aura-FX
Operation Manual
Chamber or safe haven. Alarms generated by
INDEX: the Aura-FX are intended to provide clear
DESCRIPTION ............................................ 1 instruction to refuge chamber occupants when
SENSORS................................................... 8 applicable. For example, a high carbon dioxide
(CO2) alarm will elicit a voice command
OPERATION ............................................. 11 instructing occupants to commence scrubbing
or replace MARCISORB CO2 cartridge(s).
INSTALLING & CHANGING SENSORS .... 14
CALIBRATION .......................................... 16 1.1 Layout
TROUBLESHOOTING .............................. 21 The main components of the Aura-FX Rack
WARRANTY.............................................. 23 Mount (RM) and Aura-FX Compact Model
(CM) are the same. The display board module,
as shown in Figure 1, contains the majority of
sub-components that the user will interface
with. Any one module can house up to two
DESCRIPTION sensors. The Aura-FX-RM can house up to six
display board modules; the standard offering is
The MineARC® Aura-FX Fixed Gas Monitor has
four. The Aura-FX-CM can house two display
been designed to provide a modern and versatile
modules.
gas monitoring platform that can be custom
tailored to site specific requirements at the
request of the client.

The benefits of Aura-FX over standard digital gas


monitoring equipment are that it requires
significantly less maintenance time in the form of
calibration and servicing, it provides a real time
display of gas concentrations to chamber
occupants, and it can be integrated into the
MineARC® GuardIAN system to allow facility
management to remotely check conditions inside
of a refuge chamber or safe haven.

The Aura-FX fixed gas monitoring system


measures gases that are specific to the
breathable atmosphere within a refuge chamber
or safe haven. Gases and atmospheric conditions
measured allow for informed decisions to be
made towards managing livable conditions within
a refuge

Figure 1: Aura-FX Display Board Module

US_Rev2.0 1
Aura-FX
Operation Manual
1.1.1 AURA-FX-RM
The Aura-FX-RM is housed in a standard 19” rack each – Apparent Temperature, O2, CO2, and
mount enclosure, taking up 4U (4 rack units). It can CO – though many combinations can be
house up to 6 displays with up to two sensors on configured including the measurement of
each display. The standard configuration consists external gases (refer to Figure 2 & Figure 3).
of four displays with one sensor on

Figure 2: Aura-FX-RM Layout

US_Rev2.0 2
Aura-FX
Operation Manual

Figure 3: AURA-FX-RM model exploded view

AURA-FX-RM
EXTERNAL OPERATING
MATERIAL HEIGHT DEPTH WEIGHT VOLTAGE
POWER DRAW
WIDTH
Housing:
MAX 33W cont.
Aluminum 482 mm 178 mm 158 mm 5 kg
24-48 VDC 1.37A @
Faceplate: 19 in 7 in 6.2 in 11 lbs.
24 VDC
Acrylic

US_Rev2.0 3
Aura-FX
Operation Manual
1.1.2 AURA-FX-CM
The AURA-FX-CM is housed in a compact configuration consists of two displays with two
enclosure with an adjustable mounting bracket sensors on each – Apparent Temperature, O2,
allowing it to be installed in different orientations. It CO2, and CO – though many combinations can
can house up to two displays with up to two be configured including the measurement of
sensors on each display. The standard external gases (refer to Figure 4 & Figure 5).

Figure 4: AURA-FX-CM Layout

US_Rev2.0 4
Aura-FX
Operation Manual

Figure 5: AURA-FX-CM model exploded view

AURA-FX-CM
EXTERNAL OPERATING
MATERIAL HEIGHT DEPTH WEIGHT VOLTAGE
POWER DRAW
WIDTH
Housing:
MAX 33W cont.
Aluminum 210 mm 270 mm 190 mm 2.3 kg
24-48 VDC 1.37A @
Faceplate: 8.3 in 10.6 in 7.5 in 5 lbs.
24 VDC
Acrylic

US_Rev2.0 5
Aura-FX
Operation Manual
1.1.3 External Gas Monitoring (Optional)
Both the Aura-FX-RM and Aura-FX-CM have over the intended sensors. A switch is mounted
optional manifolds that enable the measurement of on the front face to turn on a sampling pump to
external gases while maintaining a seal against the pull external air over the sensors; the pump
outside environment. The manifold is placed over exhausts to the exterior (refer to Figure 6,
pair of display board modules, sealing directly Figure 7 & Figure 8).

Figure 6: Aura-FX-RM External Gas Monitoring Gen 1

Figure 7: Aura-FX-RM External Gas Monitoring Gen 2

US_Rev2.0 6
Aura-FX
Operation Manual

Figure 8: Aura-FX-CM External Gas Monitoring

US_Rev2.0 7
Aura-FX
Operation Manual

SENSORS Alarm 1: T <10°C/50°F or >35°C/95°F


The alarm 1 sound is a chime that draws
2.1.1 Standard Sensors the occupant’s attention to the sensor
output where a decision on action should
The standard sensors are Apparent Temperature,
be determined by the trend line displayed
oxygen, carbon dioxide, and carbon monoxide;
on the screen.
they are the critical gases to monitor while
managing the livable atmosphere within a refuge Alarm 2: T <0°C/32°F or >40°C/104°F
chamber or safe haven. The selection of sensors This alarm will sound a voice prompt to
may differ from one application to the next. ensure air conditioning system has been
activated. This alarm warrants immediate
STANDARD SENSORS attention.
SENSOR
ALARM STATES 2.1.1.2 Oxygen (O2) Sensor
TYPES
Alarm 1: Temp < 10°C/50°F
Apparent Temp > 35°C/95°F
Temperature Alarm 2: Temp < 0°C/32°F
Temp > 40°C/104°F

Alarm 1: O2 < 19.5%


Oxygen Sensor O2 > 22%
(O2) Alarm 2: O2 < 18.5%
O2 > 23%

Carbon Dioxide Alarm 1: CO2 > 0.5%


This optical sensor is used to measure
(CO2) Alarm 2: CO2 > 1% oxygen levels within the chamber. Its
output is required to remain within the
Carbon
Alarm 1: CO > 15 ppm refuge chambers general operating
Monoxide
Alarm 2: CO > 25 ppm procedures guidelines of 18.5%-23%
(CO)
oxygen in the atmosphere. High and low
2.1.1.1 Apparent Temperature Sensor readings require immediate attention.

Alarm 1: O2 < 19.5% or O2 > 22%


This alarm sound is a chime that alerts the
occupants to the sensors output. Again, the
sensor output is displayed as a trend line
that can be reviewed before action is taken.

Alarm 2: O2 < 18.5% or O2 > 23%


This alarm will sound a voice prompt to
This sensor is used to measure Apparent control the breathing source. If the oxygen
Temperature. Apparent Temperature is a key levels are too low, the addition of oxygen is
indicator of the survivability of the chamber. When required using the regulated compressed
temperature is too high, the occupants must take oxygen source within the chamber. If the
measures to counteract it by activating the air oxygen is too high the oxygen source will
conditioning system. need to be throttled to limit the flow, usually
by one notch on the pre-defined regulator.

US_Rev2.0 8
Aura-FX
Operation Manual

2.1.1.3 Carbon Dioxide (CO2) Sensor 2.1.1.4 Carbon Monoxide (CO) Sensor

This optical sensor is used to measure This chemical sensor is used to measure
carbon dioxide levels within the chamber. the toxic gas carbon monoxide within the
Carbon dioxide will usually be an internal chamber. Carbon monoxide will be an
threat from the chamber occupants. Alarms external threat, usually on entry. Alarms
from this sensor will require immediate from this sensor will require immediate
action using the scrubbing equipment and action using the scrubbing equipment and
associated chemical. associated chemical.

Alarm 1: CO2 > 0.5% Alarm 1: CO > 15 ppm


The alarm 1 sound is a chime that draws The alarm 1 sound is a chime that draws
the occupant’s attention to the sensor the occupant’s attention to the sensor
output where a decision on action should output where a decision on action should
be determined by the trend line displayed be determined by the trend line displayed
on the screen. on the screen. As this breathing
contaminant is a toxic threat, any alarm
Alarm 2: CO2 > 1% level must be investigated immediately with
This alarm will sound a voice prompt to the preparation for scrubbing to start.
activate or replace CO2 cartridge(s) on the
scrubbing system. The external Alarm 2: CO > 25 ppm
compressed air source will now be isolated
This alarm will sound a voice prompt to
if applicable.
activate or check that CO cartridge is
seated correctly on the scrubbing system.
The external compressed air source will
now be isolated if applicable.

US_Rev2.0 9
Aura-FX
Operation Manual

2.1.2 Toxic Gas Sensors 2.1.3 Calibration Gases


The toxic gas sensors are typically part of an All sensors are calibrated before delivery of the
external gas monitoring configuration, where a Aura-FX and often once again during
sampling pump pulls external air in for monitoring. commissioning to maximize the life of the
The external gas readings are used to determine sensor.
exit strategies. The sensors chosen are based on
the potential hazards present at the installed site. Functional testing is recommended to be
undertaken at least every 4-6 months.
TOXIC GAS SENSORS Optimally, this will coincide with the standard
SENSOR refuge chamber scheduled servicing but if this
ALARM STATES
TYPES is not possible, MineARC Systems
Hydrogen Sulfide Alarm 1: H2S > 5 ppm recommends a test and (calibration if needed)
(H2S) Alarm 2: H2S > 10 ppm
be completed to ensure that Aura-FX is in good
Ammonia Alarm 1: NH3 > 25 ppm working order should an emergency occur.
(NH3) Alarm 2: NH3 > 50 ppm
Functional testing consists of exposing the
Chlorine Alarm 1: Cl2 > 0.5 ppm sensors to known gas sources, verifying the
(Cl2) Alarm 2: Cl2 > 1.0 ppm
sensor responds as shown on the display, and
Methane Alarm 1: CH4 > 1.1% that alarms are heard.
(CH4) Alarm 2: CH4 > 2.2%
Nitric Oxide Alarm 1: NO > 12 ppm The following table shows the recommended
(NO) Alarm 2: NO > 25 ppm calibration gases used for initial and
Nitrogen Dioxide Alarm 1: NO2 > 1 ppm subsequent calibrations. Other gases used
(NO2) Alarm 2: NO2 > 5 ppm should be evaluated against the operating
range and other sensor characteristics that
may affect operation.

CALIBRATION GASES
ITEM NO. DESCRIPTION COMPOSITION
20.9% Oxygen (O2)
1 Zero Air
Balance Nitrogen (N2)
15% Oxygen (O2)
2% Carbone Dioxide (CO2)
2 Standard Mixture
100 PPM Carbon Monoxide (CO)
Balance Nitrogen (N2)
25 PPM Hydrogen Sulfide (H2S)
3 Hydrogen Sulfide
Balance Nitrogen (N2)
50 PPM Ammonia (NH3)
4 Ammonia 20.9% Oxygen (O2)
Balance Nitrogen (N2)
10 PPM Chlorine (Cl2)
5 Chlorine Gas
Balance Nitrogen (N2)
18% Oxygen (O2)
100 PPM Carbon Dioxide (CO)
6 Methane Mixture
2.5% Methane (CH4)
Balance Nitrogen (N2)
100 PPM Nitric Oxide (NO)
7 Nitric Oxide
Balance Nitrogen (N2)
10 PPM Nitrogen Dioxide (NO2)
8 Nitrogen Dioxide
Balance Air

US_Rev2.0 10
Aura-FX
Operation Manual

OPERATION The trend history is updated approximately


every minute, adding a line of data. This screen
Aura-FX is not required to be turned on at all displays readings and trend on the graph
times. It is however recommended to be left on marked with the upper and lower alarm levels
where ever possible. Aura-FX connected to (if applicable). A green circle indicates
GuardIAN can only be viewed via the remote acceptable levels, a yellow triangle will indicate
monitoring if left on. In addition, some toxic the pre-alarms (chime), and a red square will
sensors may require a warm-up time before they indicate levels outside of the acceptable range
are fully operational, so these should be (voice prompt). Recorded data mimics the
considered as well. Consult with MineARC color of their status indicator (refer to Figure
Systems for any application specific 10).
considerations.

Emergency periods are defined as periods in


which the chamber is occupied. These periods
require the Aura-FX to be powered and gas
readings monitored. Actions shall be taken from
the information being displayed from each of the
screens.

3.1 Monitoring Gases and Critical


Information
Each module display will present measured
values in real time and trended over time in the
following manner (refer to Figure 9).

Figure 10: Aura-FX Main Display; Alarm Indications

Figure 9: Aura-FX Main Display Layout

US_Rev2.0 11
Aura-FX
Operation Manual
For display modules with multiple sensors, the Standard sensors in an alarm state will prompt
display is automatically split in two with the voice commands giving instructions to
same layout. This is typical with the standard occupants for necessary actions. Toxic and
Aura-FX-CM units (refer to Figure 11). external measured gases may not require
action from occupants. The standard voice
commands are shown in the below table.

Figure 11: Aura-FX Dual Sensor Layout

ALARM
ALARM VOICE COMMAND
SETPOINT

Apparent Temperature 104°F Temperature level high. Ensure air conditioner is turned on
High (40°C) and set to 30°C or 86°F.
Oxygen levels high. Decrease oxygen flow rate by one
Oxygen High 23%
notch.

Oxygen Low 18.50% Oxygen levels low. Increase oxygen flow rate by one notch.

Carbon Dioxide levels high. Commence scrubbing or replace


Carbon Dioxide High 1%
MARCISORB CO2 cartridge.
Carbon Monoxide levels high. Commence scrubbing or
Carbon Monoxide High 25 ppm
check if the MARCISORB CO cartridge is seated correctly.

US_Rev2.0 12
Aura-FX
Operation Manual
3.1.1 GuardIAN Remote Monitoring
The Aura-FX can be connected to GuardIAN, monitored above ground or from any remote
MineARC System’s browser-based monitoring location that has access to the GuardIAN
interface. All Aura-FX come with an ethernet cable network (refer to Figure 13). Features include
which simply plugs into the ethernet switch as part push e-mail notifications for sensors falling into
of the overall GuardIAN architecture (refer to alarm levels.
Figure 12). This allows gas levels and Apparent
Temperature to be

Figure 12: GuardIAN Browser Interface

Figure 13: AURA-FX-RM Connections

Please refer to the GuardIAN Installation Manual for more details.


US_Rev2.0 13
Aura-FX
Operation Manual
INSTALLING & CHANGING 4.1 Inserting Sensors
SENSORS
4.1.1 Sensor Pin Configuration
Each display can hold two sensors except for Every sensor board has the same pin
displays used for external monitoring. The life of configuration so that it will work on any module.
each sensor is as follows: Each board has a memory device that will allow
Aura-FX to identify it and its age before
REPLACEMENT displaying its information on the corresponding
SENSOR
INTERVAL display screen.
Apparent Temperature 3 Year
The sensor should be pushed in all the way into
Oxygen 3 Year the display module. Occasional misalignment
of pins may result in a blank screen displaying
Carbon Dioxide 3 Year
the MineARC Systems logo only. If the sensor
Carbon Monoxide 1 Year has been installed and is not displaying any
information, it is advised that the sensor be
Toxic Gas (various) 1 Year
removed and repositioned (refer to Figure 14).

Figure 14: Sensor Pin configuration (back). Green dots indicate pins.

Note: The green dots on Figure 14 are to indicate the sensor is positioned in the correct orientation.

US_Rev2.0 14
Aura-FX
Operation Manual
4.1.2 Display Module Configuration Once the cover is removed, the ports on the
module are separated in to two
The display module can remain in place while
independent slots. Currently it’s advised to
a sensor change is undertaken. Figure 15
operate one gas per module using SLOT 1
shows how sensors are changed on both a
as indicated in Figure 16 below. With the
Standard and Compact Aura-FX with the
introduction of further gases or the
sensor module remaining affixed to the
requirement for isolated external gas
faceplate and the quick-snap sensor cover
sensing, further gases can be added to
removed allowing access to the sensor ports.
SLOT 2 on each module (refer to Figure 15
& Figure 16).

Figure 15: Shows how a sensor change is


removed or installed Figure 16: Close up view of display module –
Slot 1 and 2 with ports

US_Rev2.0 15
Aura-FX
Operation Manual
CALIBRATION “FRESH AIR” as the Oxygen sensor’s “ZERO”
which corresponds to 20.9% O2.
For multi-year life sensors, it is recommended
they are calibrated annually. See below for the recommended “ZERO” and
“SPAN” settings for each sensor. The
Calibration can be performed from the Aura-FX calibration gas mixtures can be found in
display interface with the use of calibrated gas section 2.1.3 of this document. Different gases
sources. The calibration process requires two can be used for calibration so long as the
different, known points of measurement. These correct corresponding value is entered during
two points are typically referred to as ZERO and calibration. However, the calibration points
SPAN, although ZERO does not necessarily recommended below have been evaluated as
coincide with 0% or 0 ppm with respect to gas suitable with respect to linearity and practical
content. For example, we treat “ZERO AIR” or range.

ZERO SPAN
CALIBRATION GAS SETPOINT CALIBRATION GAS SETPOINT
O2 [1] ZERO AIR 20.9% [2] STANDARD MIXTURE 15%
CO2 [1] ZERO AIR 0.01% [2] STANDARD MIXTURE 2.00%
AURA-FX SENSORS

CO [1] ZERO AIR 0.0 PPM [2] STANDARD MIXTURE 100 PPM
H2S [1] ZERO AIR 0.0 PPM [3] HYDROGEN SULFIDE 25 PPM
NH3 [1] ZERO AIR 0.0 PPM [4] AMMONIA 50 PPM
CL2 [1] ZERO AIR 0.0 PPM [5] CHLORINE GAS 10 PPM
CH4 [1] ZERO AIR 0.0% [6] METHANE MIXTURE 2.5%
NO [1] ZERO AIR 0.0 PPM [7] NITRIC OXIDE 100 PPM
NO2 [1] ZERO AIR 0.0 PPM [8] NITROGEN DIOXIDE 10 PPM

Specifically, for this procedure, the “ZERO” will always be set first and then “SPAN”.

In addition to a calibrated gas source as outlined in section 2.1.3, you will need a few components
to calibrate efficiently. The components are as follows:
1. Gas Cylinder Flow regulator – 0.3-0.5 LPM
2. Plastic hose
3. Calibration cup
MineARC offers service start-up kits which include all the contents above as well as the necessary
calibration cylinders for a given unit. Contact MineARC Systems for further information (contact details
provided on the last page of this manual).

US_Rev2.0 16
Aura-FX
Operation Manual
6.1 Calibration Procedure This will bring you to Calibration Mode as
shown below. Press the amber crosshair
An Oxygen (O2) sensor will be used as an to continue. Alternatively, to exit this
example to demonstrate the calibration process. screen, tap the top left of the screen to
return to the normal operating display
To access the calibration screen from the normal (refer to Figure 19).
display screen, tap the bottom right hand corner
of the display, then tap the top left corner of the
display. Long taps may ensure good contact with
the screen, as these are invisible buttons. The
buttons should not be held at the same time (refer
to Figure 17).

Figure 19: Calibration Mode

The sensor(s) installed on that display


module will show up on the SENSOR
SLOT screen. Select the sensor you wish
Figure 17: Normal Display to calibrate by tapping within the
associated box for that SLOT. In this
This will bring you to the Aura-FX Information
example, we’ll be calibrating Oxygen (O2)
screen. From here, tap the top left corner again
(refer to Figure 20).
(refer to Figure 18).

Figure 18: Information Screen Figure 20: Sensor Slots


US_Rev2.0 17
Aura-FX
Operation Manual
The SENSOR TYPES screen is used to confirm After selecting the SENSOR TYPE, you’ll
the type of sensor you will be calibrating. For this be brought to your first calibration screen.
example, select “O2”. Sensors that are not shown This is where you will set your “ZERO” for
on this screen (i.e. NH3, CH4, etc.) can be the corresponding gas. For Oxygen, our
accessed by pressing “OTHER” (refer to Figure “ZERO” is 20.9%.
21).
Note: The raw reading shown: this is the
uncalibrated number. It is useful to monitor
the raw reading to ensure the reading is
stable and in the expected range. For some
sensors (O2, CO2), the raw reading will
appear in the expected engineering units.
For other sensors (toxic gas sensors), the
reading may show the millivolt output.

The large number in the center of the


screen corresponds to the calibration point.
Often, the value will not need to be
changed. However, if the value does need
to be adjusted, use the arrows above and
Figure 21: Sensor Types below each displayed digit to adjust and
input the desired value matching the
Once the correct gas is confirmed, the calibration calibration gas’s known value. Press “OK”
can begin. To prepare, MineARC recommends at the bottom right of the screen when
removing the sensor cover and using a calibration ready (refer to Figure 23).
cup to ensure the sensor is flooded with the
known gas without environmental influence (refer
to Figure 22).

Figure 23: Calibration - ZERO

Figure 22: Sensor cover removed and


calibration cup with plastic hose connected to
calibration gas cylinder.

US_Rev2.0 18
Aura-FX
Operation Manual
The next screen gives you an opportunity to After the ZERO calibration point has finished
prepare the calibration gas and flush the sensor. writing, it will automatically proceed to the next
The screen recommends 15 seconds of flushing; screen for setting the SPAN. The ZERO
however, the sensor should be flushed until the calibration gas cylinder is no longer required
raw reading becomes stable. Place the calibration at this point. From this screen, you will set the
cup over the sensor and open the flow regulator to corresponding SPAN value for the selected
allow between 0.3 and 0.5 LPM of gas to flood the gas sensor. For Oxygen, it is 15%. The next
calibration cup cavity (refer to Figure 24). calibrated gas source for flushing should be
prepared with plastic hose and calibration cup
before continuing. Adjust the value if
necessary and press OK to continue (refer to
Figure 26).

Figure 24: Calibration – ZERO; Flush

Once the raw reading has appeared to stabilize,


press “OK” at the bottom right of the screen.
Continue to expose sensor to calibration gas by Figure 26: Calibration - SPAN
holding the calibration cup over the sensor. This
will proceed to take 5 measurements and write the
Using the associated calibration gas for
values for this calibration point (refer to Figure 25). SPAN, flush the sensor for 15 seconds or
until the raw reading becomes stable (refer to
Figure 27).

Figure 25: Calibration – ZERO; Write


Figure 27: Calibration – SPAN; Flush
US_Rev2.0 19
Aura-FX
Operation Manual
Once the raw reading has appeared to stabilize, 6.2 Testing and Inspection
press “OK” at the bottom right of the screen.
Continue to expose sensor to calibration gas, It is recommended that a visual inspection and
holding the calibration cup over the sensor (refer “bump” test is carried out after every calibration
to Figure 27). and at every service interval.

The Aura-FX will proceed to take 5 measurements A visual inspection should look for a few basic
things:
and write the values for this calibration point (refer
to Figure 28) • All displays powered on and information is
legible.
• Gas levels are as reading within expected
range. NAN (not a number) or INF (infinity)
may be sign of sensor failure.
• If Aura-FX has an external monitoring
manifold, turn on sampling pump briefly to
ensure proper operation.
• Calibration due date stickers on Aura-FX
and individual sensors are as expected and
within the valid date range.
Bump testing is an informal test to verify the
responsiveness and accuracy of a sensor by
introducing calibration gas (or other known air
source) with a calibration cup and plastic hose
(similar to calibration) and forcing sensor
response. It is typical to bump test the sensors
using the SPAN calibration gas. This typically
induces alarms. During a bump test, check for
Figure 28: Calibration – SPAN; Write the following:

Once the measurements have been recorded, • Sensor values respond in the appropriate
there will be a brief screen that reads “Writing direction towards the expected value
calibration…” before returning to the normal (based on content of gas introduced).
operating display. The calibration gas cylinder is • Audible alarm chime or voice prompts can
no longer required at this point. After completing a be heard when expected.
calibration, the written date on the AuraFX and All sensors undergo calibration at 1-year
individual sensor stickers should be replaced or intervals. Toxic gas sensors are replaced on 1-
updated. The embedded calibration date on the year intervals; calibration can be waived for
sensor board will also update automatically after a sensors that will be discarded. Replacement
new calibration to reflect the latest date calibrated. sensors should be calibrated upon
commissioning.

US_Rev2.0 20
Aura-FX
Operation Manual
TROUBLESHOOTING
The following is a collection of common troubleshooting tips while operating the Aura-FX. For all issues that
are not covered below, please contact MineARC Systems for support.

Identified Problem
No information is displayed on the screen when a sensor is installed
Background
Sensors installed into the ports of a module (in either slot 1 or slot 2) will display its readings on that
associated display screen. If no information is being displayed the issue may be -
Component/Issue Possible Solution
The sensor hasn’t been correctly Remove the sensor and confirm that all pins are aligned, and that the
installed into the slot. sensor is completely pushed in.
The incoming power can be checked to confirm it meets the
The sensor may be drawing too specifications. A quick solution is to arrange the sensors in a different
much power from the supply order. Often the oxygen sensor will draw the most and can be
source. positioned on the module that’s on the far right. The other sensors can
be shifted to suit.
Cycle power to the Aura-FX or press the display module RESET
The sensor may have had an button (may require removal of sensor board depending on which slot
issue and requires to be reset. the sensor is installed). This only has to be pushed once and will
display as a screen wipe.

Other Contact MineARC systems for assistance, replacement, or warranty.

Identified Problem
Known false reading (for example – extremely low or unrealistic gas readings)
Background
Usually this error will occur during start up while the system is experiencing power distribution and
processing sensor information. A delay has been added to the software to minimize any short spikes in this
reading.
Component/Issue Possible Solution
Ensure that the system isn’t being exposed to occupants’ direct
External influence on gas
breath. Other external influences such as cleaning materials on an
readings
Ammonia sensor should be considered.
Confirm how long and at what point these false readings occur. If
Standard sensor warm-up time during start up periods, the will be considered as common fluctuation
and no action is necessary.
Calibrated improperly or values
have drifted out of calibration due Ensure the system is correctly calibrated.
to other factors.
Other Contact MineARC Systems for assistance, replacement, or warranty.

US_Rev2.0 21
Aura-FX
Operation Manual
Identified Problem
No sound for Pre-Alarm/Alarm
Background
A chime alerts occupants to the first alarm set point, accompanied by a yellow triangle and yellow trend
lines. Voice prompting alerts occupants to the second alarm set point, accompanied by a red square status
symbol and red trend lines.

Not all alarm setpoints have a voice prompt alert; these will continue to chime in both pre-alarm and alarm
states.
Component/Issue Possible Solution
Speaker cable connection is Confirm the speaker cable is connected. Please contact MineARC
loose/disconnected. Systems for further instructions for installing speakers.
SD card which stores sound files Confirm the SD card with all stored sounds is present and not
has become disconnected or damaged. Contact MineARC Systems for instructions on adjusting or
damaged. installing the SD card.
The Aura-FX display automatically rounds based on the extended
decimal values. If a value has rounded up to an alarm setpoint but the
Rounding error with displayed actual value is less than the setpoint, the alarm will not respond. This
value. can be verified by noting the sensor status symbol (circle, square, or
triangle) as this is driven by the actual value and not the displayed
value.
An error has occurred on the
Cycle power to the Aura-FX or press the masterboard RESET button
master circuit board and requires
(requires some disassembly).
a reset.
Other Contact MineARC Systems for assistance, replacement, or warranty.

Identified Problem
No information being displayed on the GuardIAN home screen
Background
The most likely cause for this issue will be the connection of the Ethernet cable to the switch. Any further
issues and troubleshooting guide for aspects or components of the GuardIAN system can be found in that
manual.
Component/Issue Possible Solution
Ethernet cable is not connected Confirm Ethernet cable is connected to switch. Replace switch if
or is damaged. necessary.
Contact MineARC Systems for further details on GuardIAN or for
Other
assistance, replacement, or warranty.

Identified Problem
Broken sensor covers
Background
During the installation of sensors, sensor covers need to be removed.
Component/Issue Possible Solution
The snap fit connection is No significant impact will result in removing this clip to the operation of the
damaged unit. Replacement covers can be obtained from MineARC Systems.

US_Rev2.0 22
Aura-FX
Operation Manual
WARRANTY
MineARC® Systems shall not be liable for any
failure to adhere to guidelines, warnings, and
precautions provided in this manual.

The MineARC® Aura-FX is designed and


manufactured to the latest internationally
recognized standards. As such MineARC®
warrants its products against defective parts and
workmanship for a period of 12 months from date
of purchase. MineARC® will repair or (at its option)
replace any goods which are or may become
defective under proper use where such defects
solely arise from faulty design materials or
workmanship provided such defects are made
aware to MineARC® within the requisite time limit.

US_Rev2.0 23

Potrebbero piacerti anche