Opto Electronic Level Switch

Amplifier Type 720.0250


Transducer Type 720.06XX
Transducer Type 720.0032
Transducer Type 720.0042

INSTRUCTION MANUAL


Made by INCATRONIC
TABLE OF CONTENTS
Page

1. WORKING PRINCIPLE 4

1.2 Installation hints 6


1.2.1 Transducer Type 720.0042, connections for glass apparatus 6
1.2.2 Transducer Type 720.0032 6

1.3 CE-Marking 8

1.4 Quality 8

1.5 Test certificate 8

2. SELECTION TABLE 9

2.1 Multi-part Transducer Type 720.....06XX 10


2.1.1 Multi-part Transducer Type 720.0660 10
2.1.3 Multi-part Transducer Type 720.0689 10
2.1.4 Multi-part Transducer Type 720.06XX 11
2.1.5 Construction details of Level and Interface sensors 11

2.2 One-piece Transducer Type 720.....0032 12

2.3 One-piece Transducer Type 720.....0042, completely made of glass 12

2.4 Controller 19” Type 720.X6027X...0250 13

2.5 Controller built in housing Type 720.X6021X...0250 13

3. PUTTING INTO OPERATION 14

3.1 Mechanical installation of the Transducers 14

3.2 Mechanical Installation of Controller 15

3.3 Electrical connection of Transducer and Controller 15


3.3.1 Connection diagram Transducer and Controller for 230 V∼/24V∼/24 VDC 16
3.3.2 Electrical connection Transducer 16
3.3.3 Electrical connection Controller 17

3.4 Mains connection 17

3.5 Connection of Relays 18

3.6 Function test 18

4. OPERATION 18

4.1 Setting of the fail-safe direction with Switch S1 18

4.2 Calibration with CAL 20

4.3 Adjusting of delay time for Signal relay with switch S2, Potent. P1 and P2 20

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
2
4.4 TEST-Procedure 21

4.5 Operational characteristics of the relay SIGNAL 22

4.6 Operational characteristics of the relay STÖRUNG (FAILURE) 22

5. MAINTENANCE 22

6. GUARANTEE 22

7. REPAIR 23

7.1 Exchange of the Glass light guide 23


7.1.1 Disassembly 23
7.1.2 ASSEMBLING 24

7.2 Exchange of sensor head 24

7.3 Replacement of fuse 25

7.4 Miscellaneous 25

8. TROUBLE SHOOTING 25

9. TECHNICAL DATA 27

9.1 Transducer 27
9.1.1 One-piece Transducer 27
9.1.2 Multi-part Transducer 28

9.2 Controller 30

10. ELECTRICAL CIRCUIT AND LAYOUT 32

10.1 Position-plan 32

10.2 Electronic circuit, power supply 33

10.3 Electronic circuit, Switch amplifier 34

11. SELECTION CODE 35

11.1 Selection code Type 720...0032 35

11.2 Selection code Type 720...0042 completely glass 36

11.3 Selection code Type 720...06XX 37

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
3
1. WORKING PRINCIPLE
These devices are used for level detection. The conical tip of the sensor ensures
really precise on-off performance level sensing. Its reaction is independent of
various physical parameters of the liquid such as density, dielectric constant,
conductivity, colour or refractive index. This allows the use for safe level switching
as well as for precise level control applications. If there is any foam you even can
decide whether to use the sensor for foam level or liquid level control, suppressing
the influence of foam on switching.
If the sensor is supplied with a U-tip, it is able to sense changes in refractive
index. So it can be used for example as an interface level switch for liquids.

LEVEL, conic tip INTERFACE, U-tip




A B
A A
A B

oil
liquid
water
WETTED

WETTED
GN-WS

GN-WS
INDICATION

INDICATION
U

Trigger level Trigger level


DRY

DRY
Voltage

Voltage

air liquid liquid 1 zone of mixture liquid 2

refractive index refractive index

Fig. 1

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
4
1.1 Range of Application, Selection of products

400 400
Design Temperature [°C]

Design Temperature [°C]


250 250

720.06X0 720.06X9
High Pressure version High Presure version

0 0
0 300 500 300 500
-60 -60 0
250 Design Pressure [bar] 250 Design Pressure [bar]

-200
-269 -269
Typ 720.06X0 Typ 720.06X9 (with heat-sink)

400 400
Design Temperature [°C]

Design Temperature [°C]

250 250

95
0 0
-30 0 250 300 500 250 300 500
-60 -60 0
50 Design Pressure [bar] 10* Design Pressure [bar]

-269 -269
Typ 720.0032 Typ 720.0042

* depending on connection type


Fig. 2

All these transducers need a switch amplifier type 720.0250 for power supply and
switch state signalling.

For the multi-part transducers type 720.06XX the following certificates apply:

• Ex Zone 0 + Zone 1: PTB Nr. Ex 84/2127


EEx ib IIC T6 (up to 60 °C) rsp. T5 (up to 75 °C),

• Overfill prevention device for storage of inflammable liquids in accordance


with VbF PTB Nr. 06/III B/S 1866F

• Overfill prevention device for storage of water polluting liquids in


accordance with “Wasserhaushaltsgesetz (WHG) §19,
DIBt approval no. Z-65.14-1

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
5
1.2 Installation hints
The sensors may be installed in any direction, i. e. from top, bottom, sideways or
inclined. In some special applications it is recommended to use following hints:
Mounting suggestions in the following circumstances:
• High viscosity: sideways or from bottom
• Dry Running protection for pumps: If the pipes to or from the pump are
installed horizontally use top mounting for fast detection of liquid
• Overfill prevention devices: Usually vertically from above

1.2.1 Transducer Type 720.0042, connections for glass apparatus


ML

ML

ML

Rubber plug Spherical flange, KF... Conical plug, NS.../...


Fig. 3

1.2.2 Transducer Type 720.0032

ML
ML

>10 mm 1)
1) This distance may vary
depending on surface and
material properties!
>DN25

Installation from top for storage tanks Installation sideways into pipe or
ML max. 2000 mm, if required fixing tank wall, tip has to protrude beyond
near the sensor tip is possible union to avoid sensing of gas bubbles
Fig. 4

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
6
1.2.2 Transducer Type 720.06XX

Installation of Type 720.0680


from top with flange connection
e. g. as Overfill Prevention
Device.

EL is measured between flange


face and sensor tip.

Minimum flange size is DN25


PN6 or ANSI 1” 150 lbs and all
face types may be used.
Flange and sensor housing may
be welded together especially
for dangerous or toxic media. In

EL
ML
this case there is no hexagon at
the sensor housing.
Fig. 5

Installation sideways into pipe or


wall of Type 720.0660, e. g. in a
pump’s suction pipe.

Sieb It may be equipped with a sieve


to reject gas bubbles from the
sensor tip which ensures
switching only when the pipe
runs empty.
The Type 720.0660 shown here
ML has a fixed measure ML of 25
mm.

Fig. 6

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
7
1.3 CE-Marking
These devices have been tested to fulfill all requirements relevant for CE-Marking.
The basic rules are given by the Council Directives and the technical requirements
are covered by the relevant harmonised European Standards.

The relevant Council Directives are:

89/336/EEC: EMC-Directive (Electromagnetic compatibility)

89/392/EWG: Machinery Directive


Optoelectronic level switches are not machines in the
sense of this directive, but may be used in Machinery
which has to comply with.

73/23/EWG: Low Voltage Directive

1.4 Quality
All devices are produced within an approved QM-System under DIN EN ISO 9001.

1.5 Test certificate

Each switch amplifier is tested individually and is delivered together with a copy of
this test protocol.

Ex-Type: Standard-Type:

INCATRONIC-Phönix Messtechnik GmbH INCATRONIC-Phönix Messtechnik GmbH

Typ: ............................ Ex Prüfer: ...... Typ: ............................ STD Prüfer: ......


Ser.Nr.: ............ Datum: .......... Ser.Nr.: ............ Datum: ..........
Komm.Nr.: ........... Komm.Nr.: ...........
0.5 kV Isolation
eigensicher- prim.- Kurzzeichen
ERDE sek.

Platine ok
Ik zw. gn/bn-ws: Ueigens.gn/ws: Ueigens.: bn-ws:
........................A .....................V ........................V Platine ok

Fig. 7

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
8
2. SELECTION TABLE

Each Unit consists of a Transducer and a Controller (Switch Amplifier). The


Transducer may be of one-piece or multi-part design. Ex types are generally
constructed as multi-part types. Multi-part Transducers can be equipped with a
heat-sink to expand the temperature range.

Zone Design Data Measuring Transducer Controller


Temp. Pressure length ML multi-part 720...06XX
one-piece 720...0032
[°C] [MPa/bar] [mm] 720...0042
STD -60/+250 0,5/5 50 - 250 720.111X000XX.0042 720.1602XX000.
0250

STD -30/+95 5/50 18 - 49 720.11X300000.0032 720.1602XX000.


720.11X300XXX.0032 0250
-30/+95 5/50 30 -1500

STD 60/+250 25/250 25 720.11X300000.0660 720.1602XX000.


-60/+250 25/250 50 - 960 720.11X300XXX.0680 0250
-269/+400 25/250 25 720.11X300000.0669
-269/+400 25/250 50 - 960 720.11X300XXX.0689

Ex -60/+250 25/250 25 720.21X300000.0660 720.2602XX000.


-60/+250 25/250 50 - 960 720.21X300XXX.0680 0250
-269/+400 25/250 25 720.21X300000.0669 EEx ib IIC
-269/+400 25/250 50 - 960 720.21X300XXX.0689
EEx ib IIC T6

STD = non hazardous areas


Ex = Ex-approved
For X in Type-numbers see numbering scheme page 37 - 41
Tab. 1

Controllers may be constructed as 19" plug-in modules or built into a plastic


housing with a clear cover.

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
9
2.1 Multi-part Transducer Type 720.....06XX

2.1.1 Multi-part Transducer Type 720.0660

29
G1/2A
SW27 Sensor head
Fixed measuring length ML,
ø18

i. e. no extension possible.

Protection fingers are part of

ø45
the sensor housing.

PG13,5
ML
= 25 mm Sensor housing
89 Fig. 8

2.1.2 Multi-part Transducer


Type 720.0680
Extension and protection tube Measuring length ML from 50 to
960 mm.
G1/2A

Glass light guide SW27 Sensor head


ø16
ø7

Extension tube screwed into


sensor housing.
ø45

Preferred lengths for ML: 50, 60,


PG13,5 80, 90, 100, 120, 150, 200, 300,
ML 600 and 800 mm. Other lengths
max 960 mm Sensor housing can also be supplied.
ML + 4 89
Fig. 9

2.1.3 Multi-part Transducer Type 720.0689


Extension and protection tube
Heat-sink
G1/2A

Glass light guide SW27 Sensor head


ø16

Extended temperature range with


ø7

heat-sink. This may be combined


ø45

with fixed ML 25 mm and with


extended types.
ML
Sensor housing
max 960 mm PG13,5
ML + 4 149 with heat-sink Fig. 10

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
10
2.1.4 Multi-part Transducer Type 720.06XX

Connection with flanges from


DN25 PN6 rsp. ANSI 1” 150 lbs
and all types of face.
optional
Tightness welding for dangerous
tightness welding
or toxic media is always combined
welding with a sensor housing without
hexagon.

MLF
flange
face to
sensor tip
ML
Fig. 11
sealing surface of sensor to sensor tip

2.1.5 Construction details of Level and Interface sensors

Level sensor with conical tip and Extension tube, measuring


length ML 50 - 960 mm
Type 720.068X

Level sensor with conical tip, measuring


length ML 25 mm
Type 720.066X

Interface sensor with U-tip and Extension tube, measuring


length ML 50 - 960 mm
Type 720.068X

Interface sensor with U-tip, measuring


length ML 25 mm
Type 720.066X

Fig. 12

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
11
2.2 One-piece Transducer Type 720.....0032
Interface
design

ø12
Plug connection
Glass light guide G 1/2 A (optional)
ø7

Level
design
Cable connection
ML=18 - 2000 mm

SL = 100 - 2050 mm

Fig. 13

2.3 One-piece Transducer Type 720.....0042, completely made of glass

Connection types

Interface
Plug
ø7 (8)

Level
Cable
for rubber plugs

ML
ø7

Spherical flange
from KF15 on

ML
ø7

Conical plug
from NS19/26 on

ML

ML max for all types 720...0042 = 500 mm

Fig. 14

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
12
2.4 Controller 19” Type 720.X6027X...0250

H S1 S2

Messtromkreis
Eigensicherer

safe output)
GN BN WS
CAL

(Intrinsic
L
CAL P2 P1
128 (3 HE)

DRY TR (DRY) DELAY SIGNAL

DIN 41612 F (32-pol.)


BEN (WET)

(failure)
Störung
WET

4 5 6
Rel

100
FAILURE
ST (FAILURE)

3
Signal
Rel

2
TEST

1
TEST Sicherung (fuse)

50 Hz
220 V 24 V
L N
Trafo

Netz
T50mA/T0,4A

40.6 (8 TE) 160

Fig. 15

2.5 Controller built in housing Type 720.X6021X...0250

H S1 S2
Messtromkreis
2

Eigensicherer

safe output)
GN BN WS
4

(Intrinsic

L
6

CAL
P2 P1 PG11
18 16 14 12 10 8

TR (DRY) DELAY SIGNAL (13.5/Ex)


BEN (WET)
Störung
5 6
4(failure)
120

ST (FAILURE) Rel PG11


32 30 28 26 24 22 20

1 2 3
Signal

Rel

TEST Sicherung PG11


50 Hz
220 V 24 V
L N
(mains)

Trafo
Netz

(fuse)

200
75 height

Fig. 16

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
13
3. PUTTING INTO OPERATION

It is recommended to check all parts for visible defects while unpacking. Before
installation it is useful to check all functions. For this purpose the devices are
connected temporarily, then the sensor can be tested by dipping the sensor tip
into a glass with liquid (note chapter 4.2 for calibration).Electrical connections
have to be done only by technical personnel by taking into account the local
regulations.

3.1 Mechanical installation of the Transducers


After final installation the sensor tip must not be less than 10 mm from any
opposite surface. This may vary in respect of shape and surface properties of this
surface.

One-piece Transducer - use a -screwed pipe e. g. G12S with thread


Type 720.0032: G 1/2 A to screw into a union

One-piece Transducer - use the appropriate glass connection


Type 720.0042:

Multi-part Transducer - screw it directly into a ½” union by using a


Type 720.06XX: metallic seal or in case of flanges mount it
onto the appropriate counterpart. Please
ensure that the spanner SW27 does not
stress the
transducer’s screw heads ( Fig. 23 no. 8 rsp. 9)

!"
in case of a flange and sensor welded
together
!"
do not try to tighten or loosen the transducer
!"
itself from the sensor housing (Fig. 23 no. 1)
!"
or from the sensor head (Fig. 23 no. 2)!

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
14
3.2 Mechanical Installation of Controller

Controller - prepare mounting holes acc. to Fig. 17


in Makrolonhousing: - detach the clear cover from the housing
- insert mounting screws from cover side into the
mounting holes
- tighten the mounting screws
188,7
120

Mounting holes

88
200

Fig. 17

3.3 Electrical connection of Transducer and Controller

Transducer and Controller connections are coded at the screw terminals with the
colours WS (white) rsp. 2d, BR (brown) rsp. 4d and GN (green) rsp. 6d.
It is not necessary to use shielded cables, but you should avoid laying wires near
strong EMI-sources. The maximum wiring distance and wiring resistance are
shown by following table:

wire cross section wiring distance wiring resistance


[mm2] [m] [Ω]
0,5 175 6,3
0,75 300 7,2
1,0 400 7,2
1,5 600 7,2
Tab. 2
Do not exceed a total resistance of 9Ω incl. Contact resistance, otherwise the
controller will be non-permissive and indicate FAILURE.
For Ex-instrumentation you also have to consider the max. Inductance of La ≤0,5
mH and max. Capacity of Ca ≤380 nF.

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
15
∼/24V∼
3.3.1 Connection diagram Transducer and Controller for 230 V∼ ∼/24 VDC

L1/~/+
Controller N/~/-
Type PE
Transducer 720.X602X...0250
Type
720...0032
220/230/24 V AC
720...0042
720.X1X...06XX 24 V DC

WS (d2)
(d32)
WS
BR (d4)
BR (d28) N
GN (d6)
GN (d30) L
Signal Failure
r r
(d22)

(d24)

(d16)
(d20)

(d14)
(d18)
Fig. 18

For Ex-instrumentation the cable between Controller and Transducer has to be


marked blue (intrinsically safe circuit). The Controller has to be installed outside
Ex-Zones whereas the sensor head may be in Zone 1 and the sensor housing in
Zone 0.

3.3.2 Electrical connection Transducer

WS (SW)
4 GN

BR (BR) 3
1
2
WS BR GN WS BR
GN (RT)

Transducer Transducer
Typ 720.0032 Typ 720.0032
Transducer Typ 720.0042 Typ 720.0042
Typ 720.06XX Cable connection Plug connection

Fig. 19

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
16
3.3.3 Electrical connection Controller

MESSWANDLER
(TRANSDUCER)

GN BR WS
View of female connector
d(c) Row z(a)
TRANSDUCER

WS 2
BR 4
GN 6

STÖRUNG
(FAILURE)

6
8
10

5
r
12
FAILURE

4
14
r 16

3
SIGNAL
18
SIGNAL

20

2
r
r 22

1
24
26

220/230/24 V AC
N/~/- 28
MAINS

N/~/-

(MAINS)

24 V DC
L/~/+24V 30
NETZ

50 Hz
32 L/~/+

Controller19"
Typ 720.X6027X...0250
Plug type DIN 41612 F
Controller Makrolon housing
Typ 720.X6021X...0250
Screw terminal connector

Fig. 20

3.4 Mains connection

First you should check if power supply Voltage of mains and controller are the
same. This can be done by checking the printing on the pc-board as follows:.

220 V ∼ 720.X602X1000.0250
24 V ∼ 720.X602X3000.0250
24 V = 720.X602X4000.0250
24 V = without transformer 720.1602X7000.0250

The version with a housing additionally is marked at the screw terminal block
"NETZ" with the voltage for power supply and wiring hints. For wiring see Fig. 18.
The connectors of the 19” plug-in module are to be wired as follows:
d 28 (c28) = N (-)
d 30 (c30) = L (+)
d 32 (c32) = earth

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
17
3.5 Connection of Relays

A separate relay with SPDT-contacts for "SIGNAL" and "STÖRUNG (FAILURE)"


may be used. These are potential free and so can be used for any application.
The version with a housing is marked with the connection scheme.

The connector of the 19” plug-in module are to be connected as follows:

STÖRUNG (FAILURE) SIGNAL CONTACT


4 / d18 1 / d24 normally closed
5 / d16 2 / d22 normally open
6 / d14 3 / d20 common

Tab. 3

3.6 Function test

After checking the complete wiring, the power supply can be switched on. One of
the green LEDs should flash up now. Switch off S2 (signal delay off!) and push
and hold button "TEST”. One of the two green LEDs should be on for about 1
second and after that it changes to the other green LED which starts to flash as
long as the TEST-button is held down. You now know that wiring, Transducer and
Controller are working correctly.
If there is any deviation, first check the calibration (Chapter 4.2) and if necessary
use the Troubleshooting hints in chapter 7.

4. OPERATION

4.1 Setting of the fail-safe direction with Switch S1

Switch S1 is a wire-hook-switch, which may be soldered or cut for a final setting.

H = High-alarm: - S1 open,
- Relay SIGNAL released when Sensor is immersed
L = Low-alarm: - S1 closed
- Relay SIGNAL released when Sensor is dry
Works setting: - S1 open

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
18
Working conditions and signalling

GREEN GREEN SIGNAL-


S1 LED LED RELAY
Working WET DRY
condition

open
Immersed H flashing off released
(or U-tip in media High-alarm
with higher
refractive index) closed
L on off energised
Low-alarm

open
Dry H off on energised
(or U-tip in media High-alarm
with lower
refractive index) closed
L off flashing released
Low-alarm

Tab. 4

H S1 S2
Messtromkreis
Eigensicherer
32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2

safe output)
GN BN WS

(intrinsic

L
CAL P2 P1
TR (DRY) DELAY SIGNAL
BEN (WET)
Störung
(failure)
5 6

ST (FAILURE) Rel
1 2 3 4
Signal

Rel

TEST Sicherung
(fuse)
50 Hz
220 V 24 V
(mains)

L N

Trafo
Netz

T50mA/T0,4A

Fig. 21

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
19
4.2 Calibration with CAL

This is only necessary during initial setup to compensate for all variables such as
cabling and installation characteristics (e. g. distance to wall) or if, during interface
detection the refractive index of liquids changes.
The sensor should be immersed in the optically more dense medium, i. e.

- Level sensing: liquid/gas = liquid


- Interface sensing: e. g. water/oil = oil

The Voltage across the sensor’s Phototransistor can be measured between "GN"
and "WS" of the Transducer connections. You should use a Voltmeter with an
input resistance Ri ≥ 10 MΩ and a range of 10 VDC.

Adjust CAL (P3) Adjust CAL (P3)


Measurement of with Voltmeter without Voltmeter
at GN/WS turn CAL-potentiometer

LEVEL +0,5 V 1) switching point


(conical tip) 5V + 1 turn
2)
-0,1 V -> BENETZT (WET)

INTERFACE trigger level switching point


(U-tip) + 0,2 V (ca. 4,2 V) 3) + 1/3 turn
-> BENETZT (WET)

1) Bubbling or boiling liquids


2) no reaction with drops and foam
3) Temperature changes up to 25 °C, +0,5 V for changes up to +40 °C
Tab. 5

4.3 Adjusting of delay time for Signal relay with switch S2, Potent. P1 and P2

Switch S2 is a wire-hook-switch for switching signal delay on or off. The delay time
may be adjusted separately for activating and releasing of the “Signal” relay. It can
be used with turbulent or bubbling liquids.

No delay: - close S2 or solder it for permanent closing

With delay: - open S2 or cut off for permanent opening


- delay for activating ( _↑ ) adjust with P1 as
needed (0-8s)
- delay for release ( ↓_ ) adjust with P2 as needed
(0-8s)
Works setting: - S2 open, P1 and P2 middle position (ca. 3 - 4 s)

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
20
4.4 TEST-Procedure

Test results for different settings and conditions:


Push Test Release Test
button button

LED
BEN (WET)
S1 open LED
S2 open

TR (DRY)
LED
BEN (WET)
Sensor wetted

S1 closedLED
TR (DRY)
LED
S2 closed

S1 open BEN (WET)


LED
TR (DRY)
LED
S1 closedBEN (WET)
LED
TR (DRY)
LED
S1 open BEN (WET)
LED
S2 open

TR (DRY)
LED
BEN (WET)
S1closed LED
TR (DRY)
Sensor dry

LED
S2 closed

S1 open BEN (WET)


LED
TR (DRY)
LED
BEN (WET)
S1 closedLED
TR (DRY)

1s
tD tD

Note: BEN=WET, TR=DRY, tD=delay time adjusted


with P1 and P2
=LED off
=LED flashing
=LED on

Fig. 22
Conditions:
- P1 and P2 in middle position, ca. 3 - 4 s delay time if P2 is open
- Transducer in operation
- no failure condition
If S2 is open i. e. delay on “TEST”-button has to be kept pressed longer than
delay time to get a complete test.
If S1 and S2 are open, Sensor tip immersed, the start period of the test is
suppressed. The same situation will apply with S1 closed, S2 open and Sensor
tip dry. No test procedure can be generated when FAILURE is signalised with
flashing red LED.

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
21
This Test-function allows to check independently from the operational situation the
complete signal chain from controller to signalising equipment or actuators. As
long as the automatic internal test gives a permissive signal for a properly working
Transducer, the test also checks the function of the controller.

4.5 Operational characteristics of the SIGNAL relay


see Tab. 4.

4.6 Operational characteristics of the STÖRUNG (FAILURE)relay

In normal operation the FAILURE relay is activated and is released automatically


together with the SIGNAL relay in following cases (fail-safe-behaviour):

- external power supply is interrupted for more than 100 ms


- internal power supply is interrupted or severely reduced
- short-circuit (BR-WS) or failure of Infrared-LED or interruption (BR)
- short-circuit (GN-WS) or interruption to Photo-Transistor (GN)
Supplementary information see Tab. 4. (GR=green, BR=brown, WS=white)

5. MAINTENANCE

Under normal conditions the optoelectronic level switches are maintanance-free. If


there are severe problems with grime and crusting, it may be necessary to
introduce a maintenance-instruction. This can be done by a regularly
measurement of the Voltage between GN/WS (Transducer connection):

- Immersed (BENETZT) 5 V +0,5 V / -0,1 V


- Dry (TROCKEN) 0,1 - 3 V *)

*) depending upon Sensor design; with elongation and/or Heat-sink this Voltage
may reach value of up to 3 V.

If the difference between Immersed and dry is continuously reducing or the above
suggested values can not be measured, the Sensor tip should be cleaned.

6. GUARANTEE
We grant a guarantee period of 12 months, from 1st of January 2002 24 months
for our products, provided that they have been handled and operated under
conditions described in the Operating Manual. In case of wear and spare parts we
only guarantee for failures in construction and material.

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
22
7. REPAIR

If there is any defect, it is normally necessary to return the unit for repair to the
manufacturer:

7.1 Exchange of the Glass light guide


Sensors up to 90 mm measuring length may be repaired by the customer. In all
other cases repair can be done only at the producer’s workshop.

cover screws (12)


FEP-foil (15) sealing (13)
protection tube spacer (5) heat sink (3)
(14)
packing screw (6)
glass rod (7) 15 Nm sensor body (2)

sensor housing (1)


M4x25 L
screws (8) FEP-foil (15)
packing (10) packing ring (11) screws (9)
M4x16

ca. 20 13 locking ring (17)

protection rings (16)


(obove 90 mm measuring length)

Fig. 23

ATTENTION: Before dismantling release pressure and reduce level below


sensor’s pipe union position!

7.1.1 Dismantling

- Unscrew the 4 screws (8) and detach Sensor body (2) - and Heat-sink (3) if
applicable

ATTENTION: The Spacer rings (5) are loose and may fall apart!

- Unscrew Packing screw (6) with spanner SW 17


With ML >90 mm unscrew Extension tube (14) and remove protection rings
- Remove the worn glass rod (7) with rubber pliers applied at plain end
and pull lengthways

- remove Packing (10) and Packing ring (11)

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
23
If the glass light guide (7) is stuck, then the sensor (1) has to be unscrewed and
the glass light guide (7) can be pushed out from the tip side.

7.1.2 ASSEMBLING

- Slide packing ring (11) and packing (10) over the tip onto glass light guide (7)
Note: Glass light guides (7) longer than 90 mm are equipped with protection
ring each 50 mm and are fixed with a special ceramic cement.

- Insert glass light guide (7) into sensor housing (1) from packing screw side and
tighten packing screw (6) manually.

- Insert screws (8) into sensor body (1), s. Fig. 23, attach spacer rings (5)

- Push glass light guide so that it juts out about 0,5 - 1 mm above height "L" of
spacer rings (5).

- Push down glass light guide (7) using sensor body (2) or heat-sink until spacer
rings (5) are fixed in position.

- Remove sensor body (2) again and tighten packing screw (6) with a torque of
15 Nm.

- Mount sensor body (2) or heat-sink (3) again and fix it with screws (8).

FINAL PRESSURE TEST

Any pressure and tightness test shall only be done with fixed sensor head (2) or
heat-sink (3). Test pressure: Max. 1,5 x nominal pressure

7.2 Exchange of sensor body

ATTENTION: Before dismantling always release system pressure!

Dismantling

- Unscrew cover screws (12) and remove cover with sealing (13).
Disconnect cable and unscrew screws (9 rsp. 8).

- Exchange sensor body.

Assembly is done in the reverse sequence to dismantling.

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
24
7.3 Replacement of fuse

Disconnect power supply.

Remove fuse. Insert new fuse only with the same fuse rating as that printed on pc-
board.

Power supply 720.1...0250 720.2...0250


230 V AC T 50 mA T 50 mA
24 V AC T 400 mA T 400 mA
24 V DC w. transformer M 0,16 A M 0,16 A
24 V DC without transf. T 400 mA -

Tab. 6
Connect power supply.

7.4 Miscellaneous

All electronic parts are maintenance-free when used properly.

In the case of an Ex-device, if there is any defect the repair has to be done by
PHÖNIX-ANALYTEC (Address see bottom of any page). If there is an authorised
works specialist, works repairs can be approved by this person.

When ordering spare parts please always state exact type, manufacturer’s serial-
no. and commission-no..

8. TROUBLE SHOOTING

FAILURE REASON REMEDIAL MEASURE


No function no power supply Measure supply voltage,
Check fuse,
Tighten screw terminal
screws. (Do not solder
cable ends!)

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
25
Flashing of red and a Transducer not Connect Transducer
green LED connected

Wire to Transducer Check connections and


interrupted cabling

Wires to Transducer Check connections and


short-circuited cabling

Cal. P3 not correctly set Adjust acc. to chapter 4.2


Indication by LED correct, Contacts of relay REL 1 Check relay contacts
Signal relay state do not open or close
changes , but no reaction
of external circuit
In spite of level change P3 not correctly calibrated Adjust acc. to chapter 4.2
no reaction of SIGNAL- (below trigger-point)
relay
Transducer faulty
(mechanical defect) Effect TEST-routine. If
result is o.k., check glass
light guide for damage or
if surface is filthy. If
necessary exchange or
clean the glass light
guide.
(electrical defects)
Exchange Transducer
Sensor tip too near to
opposite wall s. chapter 3.1.
Inverse reaction S1 in opposite position Change position of S1
Apparently no or too slow S2 open, P1, P2 at right Close S2 or turn P1, P2
reaction after level end position, delay Signal counter-clockwise or wait
change relay on to the end of delay time
No reaction using TEST- S2 open, P1, P2 at right Close S2 or turn P1, P2
button end position, delay Signal counter-clockwise, or
relay on open S2 and keep TEST-
button pressed until delay
time is over

Tab. 7

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
26
9. TECHNICAL DATA

9.1 Transducer

9.1.1 One-piece Transducer

Unit 720......0032 720......0042


Repeatability
- Level (conical tip) mm +0,5 +0,5
- Refractive index(U-tip) R.I. +0,01 +0,01
Temperature 1)
- Product °C -30/+95 -60/+250
- Ambient °C -25/+95 -25/+95
1)
Oper. pressure MPa/bar 0-5/0-50 0-0,5/0-5
Measuring length
- Standard mm 18-49 50
- maximum mm 2050 500
Mechanical conn.
- Standard pipe ∅12 Glass plug
for pipe union G ½ A KF, NS
or acc. to customer’s spec.
Material
- Sensor housing 1.4571 Quartz-glass
- Sensor tip, see Quartz Quartz-glass
Fig. 13, Fig. 14
- Sealing Graphite Quartz-glass
Weight kg 0,1 0,1
+5g/cm elongation +1,3g/cm elongation
electr. connection 3 m PVC-cable 3 m PVC-cable
2 2
3x0,14 mm 3x0,14 mm
or 4-pole plug or 4-pole plug
Cable outlet rubber socket rubber socket
Ingress protection EN 60529 IP 67 IP 67
Certificates none none
Controller type 720.0250 720.0250

1) Special designs may deviate


Tab. 8

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
27
9.1.2 Multi-part Transducer

Unit 720......0660 720......0669


720......0680 720......0689
Repeatability
- Level (conical tip) mm +0,5 +0,5
- Refractive index(U-
tip) R.I. +0,02 +0,03
Temperature 2)
- Product °C -60/+250 -269/+400
- Ambient °C -40/+95 -40/+95
1) 1)
Oper. pressure MPa/bar 0-25/0-250 0-25/0-250
Measuring length
- Standard mm 25 25
- maximum mm 960 960
Mechanical conn.
- Standard G½A G½A
or customers spec. or customers spec.
Material
- Sensor housing 1.4571 1.4571
- Sensor body ˜1.4301 ˜1.4301
- Heat-sink ˜1.4301 ˜1.4301
- Sensor tip, see Cladd core glass Cladd core glass
Fig. 13 Fig. 14
- Sealing
Weight kg 0,77 1,07
+9,3g/cm elongation +9,3g/cm elongation
electr. Screw terminals with Screw terminals with
connection wire protection wire protection
2
2,5 mm 2,5 mm
Cable outlet PG 13,5 PG 13,5
blue for Exi blue for Exi
Ingress protect. EN 60529 IP 65 IP 65
3) 3)
Certificates PTB Ex
4)
PTB Ex
4)
VbF VbF
5) 5)
WHG§19 WHG§19
Controller type Controller Controller
6) 6)
Typ 720.0250 Typ 720.0250

Tab. 9

1) Special designs may deviate (up to 500 bar) 6) These certificates are only valid in
2) Note Derating-Diagram Fig. 24, Fig. 25 conjunction with Controller-certificate
3) PTB Ex 84/2127 Zone 0 and Zone 1 PTB Ex 81/2136X
4) Overfill prevention device for storage of 7) e. g. Flanges DIN, ANSI, ..., also with
inflammable liquids tightness welding Sensor housing/flange
06/PTB III B/S 1866F (VbF) 8) all metallic materials acc. to customer’s
5) Overfill prevention device for storage of not specification and non-metallic materials like
inflammable water polluting liquids z. B. PVC, PP, PTFE, ...
DIBt Z-65.14-1 (WHG §19)

Rights for Technical changes reserved!

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
28
product temperature [°C]
400

350
300
250
200
150
100
50

-50 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 100

-60 -100 95
ambient temperature [°C]
-150

-196 -200
-250
-273

Derating-Diagram Transducer Type 720.06X0, without heat-sink


Fig. 24
product temperature [°C]

400

350
300
250
200
150
100
50

-50 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 100


95
-60 -100 ambient temperature [°C]
-150

-196 -200
-250
-273

Derating-Diagram Transducer Type 720.06X0, with heat-sink

Fig. 25

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
29
9.2 Controller

Units 720.1....0250 720.2....0250


LED-Indicators
- WET X X
- DRY X X
- FAILURE X X
Functions
- Alarm, WET/DRY flashes, X X
reversible
- Monitoring of short- and open- X X
circuit Transducer-connection
with fail-safe function
LED FAILURE flashes
- TEST-button for electronic X X
circuit and relay test
- Delay for relay Signal s 0-8 0-8
activating- or release
Ambient temperature
- 19”-plug-in module °C -25 - +70 -25 - +70
- Housing, Makrolon °C -40 - +40 -40 - +40
Power supply
- Alternating current V AC 24, 230 ±15% 24, 230 ±15%
- Direct current with Transformer V DC 24 ±15% 24 ±15%
- Direct current without
Transformer V DC 24 ±25% -
Power consumption VA 2,8 2,8
Outputs
- Signal, relay, 1 x UM V/A 250/4 250/4
- Failure, relay, 1 x UM V/A 250/4 250/4
at switch current cosϕ ≥ 0,7
Housing
- Dimensions H x B x T mm 200 x 120 x 75 200 x 120 x 75
- Cable outlet 3 x PG11 2 x PG11, 1 x
PG13,5 blue
- electrical connection Screw terminals Screw terminals
with wire protection with wire
2,5 mm2 protection
- Ingress protection EN60529 IP 65 2,5 mm2
- Weight kg 0,6 IP 65
0,73
19”-plug-in module
- Pc-board, Dimensions mm 160 x 100 160 x 100
- Face plate 3HE, 8TE 3HE, 8TE
- Connector male plug DIN 41612 DIN 41612
Form F (z, d) Form F (z, d)
or or
Form D (a, c) Form D (a, c)
with coding
use only in

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
30
- Ingress protection EN60529 - housing IP 20
- Weight kg 0,31 0,36
Certificates
Ex - PTB Ex 81/2136X

VbF - together with


Transducer
Overfill prevention
device for storage
of inflammable
liquids
06/PTB III B/S
1866F
WHG§19 together with
Transducer Overfill
prevention device
WHG§19
DIBt Z-65.14-1

Tab. 10
Rights for Technical changes reserved!

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
31
10. ELECTRICAL CIRCUIT AND LAYOUT

10.1 Component side


H

R31 2x1M

270R

68R
L

15k
10k
10k
10k

16k
10k P2

15k
P1

R12
R13
R14
R15
R16
R17
R19
R18
1k5
S1 C4 0.22uF S2

GN BR WS
T6 R30

32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2
R35 BC183R28
R29
P3 470R
27
IC4 4052

9x47k
D15
2x1N4148D16

26

IC2 4081

R10
+

IC1 324

R9
R8
25
CAL C5 24
D19 D18 D17

C3 + 23
IC3

1uF + 22
TR C2
21
20
R33
R32
2x10uF

82R
R34 R11

D7
D8
16k
15k
7k5

D5
D6
1458 T5 15k
IC6 4011

BEN R36 470R 2x10k BC183

Störung
IC9 4x1N4148

1 2 3 4 5 6
1M
7808
ST C6 D4 Rel 2
0.22uF 2x1N4148
R37
IC6 4011

Signal
1M D2 Rel 1
T

TEST GL B40C900/600 R1 5,1/1R


R38 ~ + ~ -

50 Hz
IC7 4052

L N
100k

220 V 24 V
+ 2200uF

Netz
R39 TR
100k Sicherung
7812

C1 T50mA/T0,4A
IC8

BV EL30-3 2,8 VA M 0,16 A


720.1602XX000.0250

Assembling print Standard execution (not Ex)


Fig. 26
D14
D13
D12

Messtromkreis
Eigensicherer

2x1N5346B
D11
D10
D9

R6 R7
4xBZX83
910R 6,8R C9V1

T4
T2 T3 BC182
2N3055 2N3055 D4
1N4148
K1 K2 D1
R4 R5 1N4148
6,8R R2 910R
910R
T1 R2
BC182 910R

720.2602XX000.0250

additional assembly for Ex, printed blue


Fig. 27

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
32
T2 T3 IC9
BV EI 30 - 3/29 R4 R7 7808
R1 TR 2N3055 2N3055
6R8 6R8
N (c28) 1 +15,8V/124mA
5,1R (+8V) eigensicher
15 V
230 V~ 2,8 VA

SI T1 R5 R6 T4
R3 D14
L (c30) 2 BC182 910R 910R BC182
910R D9 1N5352B
T 50 mA
(c32) R2 1N5352B
1 910R
BV EI 30 - 3/29 IC8
R1 TR 7812
N (c28) 1 +12V
1R
15 V 2
24 V~ 2,8 VA + C1
GL 2200uF/40V
SI B40
L (c30) 2 C900/600 Type 720.2602XX000.0250
T 0,4 A
10.2 Electronic circuit, power supply

(c32)
IC9

33
7808

Fig. 28
SI DC/DC-Converter
+ (c30) PTB-No.Ex-80/2112 U
M 0,16 A +8V
1 3
+ = + 1 Ua =
24 V= 2 4 ca. 15 V
= 2
- -
R1
- (c28)
1R
(c32)
1
SI IC8
+ (c30) 1 7812
T 0,4 A +12V
Comments:
1. Resistors R2, R3, R5 and R6
24 V= only 720.1602X7000.0250 2 C7
are 0,66W + C1 0.22uF
R1 2. With exception of GL, C1, IC8 and
GL 2200uF/40V

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
- (c28) 2 IC9 no components with different
values may be used! B40
1R C900/600
see PTB-Certification for
(c32) type 720.2602XX000.0250 Typ 720.1602XX000.0250
monitoring
+12 V R20 R23 intrinsic safe
D5 7k5 4k7 circuit
1N4148 IC1 IC2 IC2 IC2
R22 47k 10 4081 4081 4081
+ 324
C 8 9 10 6 4 2 3
9 - IC5
& & T5 D3 D4 & T6 D3 D4
R21 8 5 BC183 1 BC183 4011 R35
D6 12
1N4148 R14 R13 47k 470R
10k 10k C B A 11
R33
& 13
+15,8 V / 124 mA Rel2 R31 Rel1 D19
10k D
intrinsic safe (+8 V) 1N4148 10k 1N4148
FAILURE SIGNAL IC5
FAILURE

CQV21
IC3 4011
monitoring 9
LED-circuit 1458 6
R2 P3 R8 R9 - 10
82R 1M D7 15k 7
16k R26
IC1 B &
AA118 47k IC2 5 8
324 4081 +
12 + R11
D 14 12 11 6 7 C
15k R25 + R32 + P1 P2
R30 13 - C3 10k
C2
47k 4 IC4 10uF 10uF 1M 1M S2
1k5 & 3Y 6 7
D8 R27 13 4052
D11 D12 D10 D13 AA118 R18 R19 47k 2
2Y D16 D15 4 IC4 R34
R15 16k 15k D 5 IC3 3Y
1Y 4052 470R
100R Y 3 3 + 1458 2
1 2Y
DRY

IC1 R12 C3 + 0Y A 1 5 3
47k 324 15k 11 1Y D17
2 - 1N4148
10.3 Electronic circuit, Switch amplifier

3 + 3X X 13 Y
1

34
R28 A 1 10uF 15 0Y

4 x C9V1 1.5W
2X 11 D18

Fig. 29
2 -
14 3X X
R16 signal 1X 15
multiplexer 12 2X
WET

10k IC1 0X R38 R39 13


47k 100k 14 CQV25
5 324 100k 1X
+ 12
R29 B 7 9 0X
IC5 IC5 10
6 -
4011 4011
R17 2 5
R24 47k C5 9 10
Logic

68R +
Display

monitoring 3 4
Photo-Transistor & & TEST S1
R40 R42 circuit 1 6
A 1uF B IC6 IC6
62R 620R 4011 4011
R36 BOARD HIGH/LOW 8 5
Comments: 1M TEST ALARM
T7 10 4
D20 & &
R41 1. Metal film resistors are marked with "O" 9 C3 6
39R 2. For Type 720.1.....0250 D1, D3, D10-D13 C B
are not used relay contacts R37
3. For Type 720.2.....0250 [Ex ib] IIC IC6
internal supply voltage + FAILURE SIGNAL 4011 1M
PTB-No. Ex 81/2136 X 12 10uF
TSUS3400 BPW22 components R8-R10, R20-R30, 1
IC1 - 324 4 11 5 2

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
D1-D4, D10-D13, Rel1 and Rel2 must 14 7 11 3
not be changed in value IC2 -4081 r (c16) r (c22) & &
Transducer IC3 -1458 8 4 6 3 13 2
16 8 (c4) 4 (c20) 1 D A
IC4,7 -4052
IC5,6 -4011 14 7 (c18) (c24) IC6
4011
11. Selection code

11.1 Selection code Type 720...0032


one-piece Transducer

T -30 to +95 °C
P 0 - 5 MPa (0 - 50 bar)

720. 1 1 X3 00 XXX. 00 3 2

Measuring length 000 = Standard length 100 mm


XXX in cm
e.g.: 150 mm = 015

Measuring principle 1 = Level (K)


2 = Interface (U)

Fig. 30
Level (conical tip)
length of Sensor SL [mm] Type Order-No.
100 720.111300000.0032 173799
150 720.111300010.0032 256776
250 720.111300020.0032 256784
550 720.111300050.0032 256792
1050 720.111300100.0032 256800

Interface (U-tip)
length of Sensor SL [mm] Type Order-No.
100 720.112300000.0032 173807
150 720.112300010.0032 256883
250 720.112300020.0032 256891
550 720.112300050.0032 256909
1050 720.112300100.0032 256917

Tab. 11

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
35
11.2 Selection code Type 720...0042 completely glass
One-piece Transducer

T -60 to +250 °C
P 0 to 1 MPa (0 to 10 bar)

720. 1 1 X3 0 0 XXX. 0 0 4 2

Measuring length XXX in cm


e.g.: 150 mm = 015

Measuring principle 1 = Level (K)


2 = Interface (U)

Fig. 31

Level (conical tip)


length of Sensor SL [mm] Type Order-No.
50 720.111300000.0042 173809
80 720.111300010.0042 173809
100 720.111300020.0042 173809
150 720.111300050.0042 173809
250 720.111300100.0042 173809

Interface (U-tip)
length of Sensor SL [mm] Type Order-No.
50 720.112300000.0042 173808
80 720.112300010.0042 173808
100 720.112300020.0042 173808
150 720.112300050.0042 173808
250 720.112300100.0042 173808

Tab. 12

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
36
11.3 Selection code Type 720...06XX
multi-part Transducer

T without Heat-sink -60 to +250 °C


T with Heat-sink -273 to +400 °C
P 0 to 25 MPa (250 bar)

720. X1 X3 00 XXX. 06 XX
Temperature 0 = without Dissipater
9 = with Dissipater

Elongation 6 = Standard length 25 mm


8 = with elongation
90 - 960 mm

Measuring length XXX in cm


z.B.: 150 mm = 015

Measuring principle 1 = Level (K)


2 = Interface (U)

Application 1 = Standard (not Ex)


2 = Ex

Fig. 32
ff. Selection code Transducer Type 720...06XX

Level (conical tip), Standard (not Ex)


Measuring length ML Type Order-No.
[mm]
25 720.111300000.0660 256941
50 720.111300005.0680 257477
60 720.111300006.0680 257485
80 720.111300008.0680 257493
90 720.111300009.0680 257501
100 720.111300010.0680 257519
120 720.111300012.0680 257527
150 720.111300015.0680 257535
200 720.111300020.0680 257543
300 720.111300030.0680 257550
600 720.111300060.0680 257568
800 720.111300080.0680 257576

Level (conical tip), Ex


Measuring length ML mm] Type Order-No.
25 720.211300000.0660 257808
50 720.211300005.0680 258368

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
37
60 720.211300006.0680 257709
80 720.211300008.0680 257717
90 720.211300009.0680 257725
100 720.211300010.0680 257733
120 720.211300012.0680 257741
150 720.211300015.0680 257758
200 720.211300020.0680 257766
300 720.211300030.0680 257774
600 720.211300060.0680 257782
800 720.211300080.0680 257790

Interface (U-tip), Standard (not Ex)


Measuring length ML[mm] Type Order-No.
25 720.112300000.0660 256958
50 720.112300005.0680 257584
60 720.112300006.0680 257592
80 720.112300008.0680 257600
90 720.112300009.0680 257618
100 720.112300010.0680 257626
120 720.112300012.0680 257634
150 720.112300015.0680 257642
200 720.112300020.0680 257659
300 720.112300030.0680 257667
600 720.112300060.0680 257675
800 720.112300080.0680 257683
ff. Selection code Transducer Type 720...06XX

Interface(U- tip), Ex
Measuring length ML[mm] Type Order-No.
25 720.212300000.0660 256974
50 720.212300005.0680 257816
60 720.212300006.0680 257824
80 720.212300008.0680 257832
90 720.212300009.0680 257840
100 720.212300010.0680 257857
120 720.212300012.0680 257865
150 720.212300015.0680 257873
200 720.212300020.0680 257881
300 720.212300030.0680 257899
600 720.212300060.0680 257907
800 720.212300080.0680 257915

Tab. 13

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
38
11.4 Selection code Type 720...0250
Controller

720. X6 0 2 XX0 00 . 0 2 50 Std Ex


1 = 230 V/50 Hz x x
3 = 24 V/50 Hz x x
Power supply 4 = 24 V= x x
7 = 24 V= without x -
transformer

Housing type 1 = Housing Makrolon


7 = plug-in module19"

Application 1 = Standard (not Ex)


2 = Ex

Fig. 33
Housing Makrolon, Standard (not Ex)
Power supply Type Order-No.
230 V AC 720.160211000.0250 193201
24 V AC 720.160213000.0250 193219
24 V DC with transformer 720.160214000.0250 193235
24 V DC without transformer 720.160217000.0250 193227

Housing Makrolon, Ex
Power supply Type Order-No.
230 V AC 720.260211000.0250 193177
24 V AC 720.260213000.0250 193185
24 V DC with transformer 720.260214000.0250 193193

Plug-in module 19”, Standard (not Ex)


Power supply Type Order-No.
230 V AC 720.160271000.0250 193136
24 V AC 720.160273000.0250 193144
24 V DC with transformer 720.160274000.0250 193161
24 V DC without transformer 720.160277000.0250 193151

Plug-in module 19”, Ex


Power supply Type Order-No.
230 V AC 720.260271000.0250 192948
24 V AC 720.260273000.0250 193110
24 V DC with transformer 720.260274000.0250 193128
Tab. 14

PHÖNIX-ANALYTEC Marketing GmbH, Salzschlirferstr.13, D-60386 Frankfurt/M., Tel.: 069/419099 Fax: 69/410334
39
PHÖNIX-ANALYTEC Marketing GmbH
Salzschlirfer Straße 13
D-60386 Frankfurt
Tel. +49/069/419099
Fax +49/069/410334
Internet: http://www.phoenix-mt.com
e-mail: sales@phoenix-mt.com

Weitere Produkte der


PHÖNIX-Messtechnik Gruppe:

Further products of the


PHÖNIX-Messtechnik Group:

Schauglasanzeiger
Sight Glass Level Gauges

Magnetanzeiger
Magnetic Level Gauges

Schwimmmerschalter
Float Switches

Schwimmer Füllstandmesser
Float Level Gauges

Verdränger Füllstandmesser
Displacer Level Gauges

Ultraschall Füllstandmesser
Ultrasonic Level Gauges

Optoelektronische Grenzwertgeber
Optoelectronic Level Switches

Kapazitive Grenzschalter
Capacitance Level Switches

Ultraschall Grenzschalter
Ultrasonic Switches

DR7200250ENG REV 5 12/01