NEONATAL INCUBATOR

Osorio Nicolás, Ospina Daniela, Pérez Alejandro,


niko9306@hotmail.com, Ingdanny89@hotmail.com, alejandro-p91@hotmail.com
UNIVERSIDAD MANUELA BELTRAN

Abstract
II. THEORETICAL FRAMEWORK
An incubator is a treatment equipment specially
designed for premature babies this equipment
allows to control the temperature and humidity for INCUBATOR: Incubator is a medical treatment
babies, the temperature control is diverse and team that has a camera, into which is placed the
sophisticated control systems to not affect the infant in order to provide a controlled
integrity of the baby and not have problems in environment. Depending on the type of incubator,
equipment since today have been implemented you can control the temperature, humidity
advanced monitoring systems where weight surrounding the patient. Incubator patient transfers
control is included, respiration depending on the heat mainly by convection principle, that is, heat
complexity of the unit this laboratory is intended transfer is performed by the movement of a fluid
to design an incubator that controls the variables in this case moving air. The heater control unit or
of temperature and moisture through an interface heat activated by an electrical signal, which is
graph where specified temperature and humidity
directly proportional to the difference between the
inside the device.
measured temperature and the reference value
Key Words: incubator, equipment, weight, preset by the operator. The control variable may
temperature, humidity. be the temperature of the air produced by the
heating of the incubator or the skin temperature of
I. INTRODUCTION the infant. Temperature control can be performed
in two ways: Air Control Mode: The nurse sets a
value of temperature control a benchmark of air
The incubator is a method of artificial treatment
inside the incubator. The heating system will work
which supplies the baby or premature through
so that the temperature inside reaches the
apparatus stability in temperature and humidity by
means of control and monitoring when they can’t reference value. Generally, the nurse has
continue the cycle of normal gestation the main monitoring and display of reference temperatures,
purpose of such equipment is precisely control the the interior of the incubator and the patient (for
growth and control of vital signs of a baby this which, a temperature sensor is used). In the case
computer uses thermodynamic principles as is the that the control system detects the incubator
convection heating implemented with a series of temperature differences between air control and
circuits and control stages is achieved that the camera and that these differences exceed the
baby can be well. previously established range, equipment alarms
will skyrocket; in some cases they are quite
The VM must be able to monitor the temperature complicated when the heating system and security
control can be performed by means of on / off a control is disabled. Control mode or servo-
when a certain temperature reaches a set point controlled patient: The nurse establishes a
activate a series of fans expel this temperature to reference value of the desired temperature in the
prevent the glass from fogging in some equipment
neonate; thus the heating system will be in
controls vital signs is implemented to have at most
operation until the patient reaches the desired
this is premature monitored by sensors previously
calibrated and adjusted. temperature previously scheduled. To achieve
this, the incubator needs to take samples
continuously the actual temperature of the patient,
which is achieved by placing a temperature
sensor; this continuous monitoring is the source
that feeds the system to direct the operation of the
heater.

Components of a neonatal incubator:

 Control: Temp. And Humidity.


 Skin temperature.
Image 2. Example about Prototype of Neonatal
 Patient temperature sensor. incubator
 General system failure.
 Air intake filters.
 Alarms.
IV. MATERIALS AND METHODS

In the design of the incubator both electronic,


III. METHODOLOGY
electrical and mechanical, different materials and
mechanisms used to complement the physical
For our prototype implement a neonatal incubator design of biomedical equipment, to implement
that allows control through an interface in this and the following materials were used with
LabVIEW to achieve interact with the user of the following functions are used:
biomedical equipment, in addition to this are to
include the steps of temperature, humidity and • Heat resistance (110V): heats the air
control them, in order to make a efficient and entering the incubator, allowing heat
servo-controlled system, according to that in inside the device to maintain a set
picture 1, we can find a block diagram as going to temperature.
implement the device, which will be according to • WATER HEATER (110V): Allows
what will develop. humidify the environment to maintain the
relative humidity it takes hold inside the
incubator according to the need required
by the patient.
• RELAYS: Control voltage step enabling
the passage of current according to the
need required by the control.
• FANS (12V): A fan type air into the
machine and other handles expel air
flowing into the incubator.
• TEMPERATURE SENSOR (LM35):
performs temperature census, according
to their values change voltage delivery.
• HUMIDITY SENSOR (HS1101):
According to its variation must be
linearized by a 555, he gives us a
frequency which according to the
information contained in the datasheet
values are set humidity variation.
• ARDUINO UNO: Works as the
Image 1. Block Diagram of Neonatal Incubator. microcontroller of the device, for it is
responsible for controlling all the
electronics, receive and transmit
information for controlling the computer, incubator (location of sensors for correct reading
also behaves as the brain of the device. of information).
• ACRYLIC DOME: Incubators as To vary a friendly interface was developed with
biomedical equipment according to the user so that it could develop proper
technical regulations, regulations that communication with the interface and so does the
come with the dome must be user has the possibility to enter faster and much
manufactured in acrylic anticorrosive. more easily information, so that it no mistakes in
• BASE WOOD: For the prototype timber scheduling variables when entering a patient
was used to be a material that resists the newborn in picture 4, we show the interface with
weight and temperature supports it which the user interacts with the biomedical
without changing its physical properties device and likewise in the image 5, a part of the
to a certain value, was painted white to programming and control shown temperature
simulate the color is handled in the where the integration is performed of temperature,
Neonatal ICU. this interface was designed in LabVIEW software.
• Diodes, transistors, resistors and H
bridge to handle fan power.
• LABVIEW: This software perform all
the programming interface and
standardization of sensors according to
reference values, according to this whole
interaction was performed with the user
where a simulation where the user could
freely choose the values held temperature
and relative humidity, as well as display
the same value that was obtained from
the incubator.

V. RESULTS

As obtained in the laboratory and in practice the Image 4. Interface of neonatal incubator.
following prototype according to the preset was
made in the methodology:

Image 5. Control PID of temperature


Image 3. Protoype of Neonatal Incubator in (integration)
Simulation Laboratory UMB.
VI. ANALYSIS OF RESULTS
Regarding the design it was modified and was
designed from the beginning all the aspects that With respect to that obtained in practice, the
make the machine out as similar to a patentable biomedical device could be made, taking into
medical device, for it took into account the account the real aspects of a real team, for that
manufacturing materials, electronic devices used sensors were installed on the surface of the drink,
were taken into account and protocols using an also the mechanisms correctly were located to
control flow air inside the incubator, also the
design of the machine with the most tightness was • It was identified that the sensor currently used
generated, to prevent leakage of temperature and is the HS1101 sensor, for making relative
humidity. humidity, in the industry is the sensor used
and therefore knew how to linearize it and
Control: operate it according to their variability
The control is made by a servo-controlled system, frequency.
so that executes an accumulation of the
temperature data and makes integration to give as
result the temperature value, i.e. which it behaves BIBLIOGRAPHY
as a system similar to PID .

Temperature: [1] Dinerstein N., Alejandro; González, María


The incubator in control temperature according to Aurelia; Brundi, Mónica Humanizacion en la
a maximum value and a minimum value from atención neonatal Revista del Hospital Materno
entering the patient, so it could control the Infantil Ramón Sardá, vol. 19, núm. 4, 2000, pp.
temperature according to the weight of the 173-176 Hospital Materno Infantil Ramón Sardá
newborn, according to that you can control the Buenos Aires, Argentina
temperature in two different ways with the same
goal, also the heater was intended to heat the air to [2] Dávalos Pinto, J.; Ávila Córdova, E.
Identificación del mezclador de gases de la nueva
a temperature to control average was found.
incubadora neonatal BAN Industrial Data, vol. 12,
núm. 1, enero-junio, 2009, pp. 21-26 Universidad
Humidity: Nacional Mayor de San Marcos Lima, Perú.
To control humidity with HS1101 sensor
linearization according to not deliver voltage [3] Buscaglia, Juan Carlos; Carduzo, Carlos
values but delivers a frequency value, which Proyecto de incubadora para prematuros de muy
combined with 555 changes the frequency and bajo peso de nacimiento Revista del Hospital
likewise that frequency is the result for the Materno Infantil Ramón Sardá, vol. 19, núm. 1,
algorithm is performed which makes me that 2000, pp. 18-20 Hospital Materno Infantil Ramón
frequency in a preset percentage of relative Sardá Buenos Aires, Argentina
humidity.

VI. CONCLUTIONS

• It was determined that to establish a proper


sealing of the device should provide for
specific measures, which do not allow the
escape of heat or moisture, since this affects
the reading of the data and comparing with
the thermo-hygrometer will not be possible
correct comparison.

• It was established that for a servo control,


must implement a system that performs a
comparison of the same information and
integrates it or derive depending on the
information received.

• It is understood that for a modern incubator


closed servo control should be performed,
because if control is performed ON-OFF can
affect the integrity of the newborn, so it may
possibly be proposed as an adverse event in
the ICU Neonates.
BIBLIOGRAPHY

➢ http://www.udb.edu.sv/udb/archivo/guia/biomedica-tecnologico/equipo-medico-de-
tratamiento/2012/ii/guia-6.pdf

http://www.draeger.com/sites/assets/PublishingImages/Segments/CSA/Hospital/Downloa
d -Center/Cuidados%20Neonatales/NICU_Booklet_Preview.pdf
➢ http://www.bioingenieria.edu.ar/grupos/geic/biblioteca/Trabypres/P06EPPe02.pdf
https://cdn.fbsbx.com/v/t59.2708-
21/14736554_102026364381…oh=8b17c1980be6b6ddbdde73b7a64996fb&oe=580971FF&dl=1 19/10/
8:22 a.m.

a
3

Potrebbero piacerti anche