Data Sheet
Analog Digital Multimeter
with Insulation resistance
Measuring Measuring Resolution Discharge U0 max. Intrinsic error of digital display Over load capacity 1)
function range resistance ± (…% of rdg + …digit) at Over load Overload
reference condition value duration
30.00 nF 10 pF 250 KΩ 2.5 V 1.0 + 3 2) 1000 V
300.0 nF 100 pF 250 KΩ 2.5 V 1.0 + 3 DC 10 sec
Farad 3.000 μF 1 nF 25 KΩ 2.5 V 1.0 + 3 AC
30.00 μF 10 nF 25 KΩ 2.5 V 3.0 + 3 eff / rms
Sine
Measuring Measuring Resolution Discharge U0 max. Intrinsic error of digital display Over load capacity 1)
function range resistance ± (…% of rdg + …digit) at Over load Overload
reference condition value duration
f min V dc f min V ~
300.0 Hz 0.1 Hz 1 Hz 45 Hz ≤ 3 kHz
3.000 KHz 1 Hz 1 Hz 45 Hz 0.5 + 1 3) 1000 v Continu-
Hz 30.00 KHz 10 Hz 10 Hz 45 Hz ≤ 30 kHz; ously
100.0 KHz 100 Hz 100 Hz 100 Hz 300 V
≤100 kHz
2 Hz… 1kHz ± 5 Digit 4) 30 V
% 2.0....98.0% 0.1 % 2 Hz -- 1 kHz … 10 kHz; ± 5 Digit /
kHz 4)
-200.0...
Pt +200.0 ºC 0.1 ºC – -- 2 Kelvin + 5 Digit 5)
1000 V
ºC 100 +200.0... 0.1 ºC 1.0 + 5 5)
DC
+850.0 ºC AC 10 sec
-100.0... eff / rms
Pt +200.0 ºC 0.1 ºC 2 Kelvin + 2 Digit 5) Sine
1000 +200.0...
+850.0 ºC 0.1 ºC – -- 1.0 + 2 5)
Environmental
Operating temperature -20 to +50°C
Storage temperature -25 to +70°C
Relative humidity <75% non condensing.
Terminal Protection IP20 for terminals
Altitude Up to 2000 m
Battery
Battery Voltage 6 x 1.5 V Cells
Battery type Alkaline manganese Dioxide cell as per IEC LR 03 , ANSI 24A (Size AAA)
Battery Life Minimum 600 hours on Vdc, Adc ,240 hours on Vac, Aac,
For MΩISO @1000 V, 800 Measurements possible with nominal current
MΩISO @500,250V, 100V, 50 V, 2400 Measurements possible with nominal current.
Interface
Type RS232C, serial, as per DIN 19241.
Data transmission Optically with infrared light through the case.
Baud rate 8192 bits/s.
30/300 mV dc 1.0 + 3
3...300 V dc 0.15 + 1
1000 V dc 0.2 + 1
V~ 0.4 + 1
300μA … 300mA DC 0.5+1
A AC+DC 0.75+3
30 Ω 2) 0.15 + 2
Temperature 0 ºC 300 Ω 0.25 + 2
+21 ºC 3 KΩ – 3 MΩ 0.15 + 1
and 30 MΩ 1.0 + 1
+25 ºC…+40ºC 30 nF 2) – 3 μF 0.5 + 2
MΩIT 0.25 + 2 30 μF 2.0 + 2
Hz 0.5 + 1
% ± 5 digits
-200…+200 ºC 0.5 K + 2
+200…+850ºC 0.5 + 2
15 Hz....< 30 Hz 3…1000 V ~ 1.0 + 3
30 Hz....< 45 Hz 0.5 + 3
Frequency of the measured > 65 Hz... 400 Hz 2.0 + 3
quantity 3….300 V ~ 3.0 + 3
>400 Hz…1 KHz 1000 V ~ 3.0 + 7
20Hz …< 45 Hz A~ 2.0 + 3
>66 Hz… 1 kHz 3.0 + 3
Wave form of the 1….3 ± 1 % of rdg
measured quantity 3) Crest factor CF V ~ 4) , A~ 4)
3….5 ± 3 % of rdg
V DC 2 Digit
V~, ADC 4 Digit
A AC+DC 6 Digit
5)
Battery Voltage …< 7.9 V 30Ω / 300 Ω/˚C 4 Digit
> 8.1 V …10.0 V 3 kΩ – 30MΩ, MΩIT 3 Digit
nF, μF,
1 Digit
Hz 1 Digit
% 1 Digit
Relative humidity
3 Days Ω 1 х intrinsic error
Hz
Meter off ˚C
DATA - % ± 1 digits
Applicable Standards:
For Use as a Insulation Measuring IEC 61557: Devices for testing, measuring and monitoring protective safety measures
Instrument. in system with voltages of up to 1000 V A.C. and 1500 V D.C.
IEC 61557- 1: For general requirements
IEC 61557- 2: For Insulation resistance measuring instruments
EMC IEC 61326: Class B
Immunity IEC 61000-4-2 8 KV atmosphere discharge, 4 KV contact discharge
IEC 61000-4-3 : 3 V/m