Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User Manual
OMICRON Test Universe
2
Table of Contents
Table of Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.1 EnerLyzer Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2 Test Sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.1 BINARY/ANALOG INPUT (1 - 10). . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.2 General Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3
OMICRON Test Universe
Index ................................................................................................................................... 83
4
Introduction
1 Introduction
EnerLyzer is a test module that allows to control the measuring features of
CMC test sets. It runs as a stand-alone test module. It has four modes of
operation: multimeter mode, transient recording mode, harmonic analysis
mode and trend recording mode.
These four modes are mutually exclusive. You can not run the multimeter
mode at the same time that you perform a transient recording or run
harmonic analysis. Switching between the four modes can be done via
toolbar icons.
The optional EnerLyzer package consists of:
• The CD-ROM that contains the EnerLyzer test module installation, the
license file and the EnerLyzer documentation (Help and PDF manual).
• The printed EnerLyzer user manual.
5
OMICRON Test Universe
6
Test Sets
2 Test Sets
Each of the ten binary inputs in the BINARY/ANALOG INPUT section of a a
CMC 256, a CMC 256plus or a CMC 356 test set can be configured through
the EnerLyzer test module to be an analog input for DC and AC voltages up
to 600 V.
Because the analog inputs of the test set deal with voltage inputs, active
current probes with voltage outputs have to be used to measure currents. All
probes must be active current probes with voltage output or current probes
with a shunt.
OMICRON offers the C-PROBE1 as a suitable current probe. This current
probe is not included in the deliverables of EnerLyzer and must be ordered
separately. The test set also supports other current probes than the named
C-PROBE1.
1
The binary inputs 1 - 10 can also be used as counter inputs for input frequencies up to 3 kHz.
7
OMICRON Test Universe
8
Test Sets
Time
The maximum length of the recording depends on the settings for the sample
rate and the number of channels to be recorded.
9
OMICRON Test Universe
10
The EnerLyzer Test Module
11
OMICRON Test Universe
Function
The channel usage can be specified in the function row with drop down
menus at each channel 1 through 10. Channels can be used as binary inputs,
counter inputs, voltage inputs, or current inputs.
Nominal Range
The nominal range can be specified for each voltage or current channel. The
nominal range is the dynamic maximum (peak value) that is expected to
appear on the channel.
12
The EnerLyzer Test Module
Clamp Ratio
For each current channel, the current clamp ratio can be specified. This
setting in the Hardware Configuration dialog box should match the settings
of the switches on the actual current clamp.
The inputs of the CMC test set were intended for sinusoidal signals. As such,
the nominal range can be considered Vrms. Together with a clamp ratio, it
specifies how large the peak values can be before clipping occurs. This is
important to remember for non-sinusoidal signals that may be within the Vrms
nominal range but that may have larger peak values which get clipped.
Specifying a larger range can prevent clipping. However, doing so results in
a loss of resolution, because the analog-to-digital conversion only has 12-bits
to represent the entire range.
The left-hand column of the Hardware Configuration dialog box is used to
specify the test module input signal. The index numbers n on Vn and In
represent a logical connection in subsequent power dialog boxes in order to
determine real, apparent, and reactive Power n. In other words, Power2 is
calculated from V2 and I2. It is not possible to monitor the results of, say, V1
and I3.
The second column from the left in the Hardware Configuration dialog box
is used to enter a Display Name for subsequent dialog boxes. In this example
for monitoring voltages and currents, the display names chosen are V a, V b,
and V c for the input voltages and I a, I b, and I c for the input currents, as is
shown in Figure 3-1.
The remaining portion of the Hardware Configuration dialog box is a table
for establishing logical connections between the input signals or display
names and the physical channels on the CMC test set. Each input signal can
be assigned to only one channel and vice versa.
Figure 3-1 shows an example of the logical connections and the intended
physical wiring of the voltage probes and current clamps to the CMC test set.
If the actual wiring of the current clamps is different (for instance, if the I
inputs are exchanged for the V inputs), the logical connections can be used
to make the associations instead of physically moving cables on the front
panel of the CMC test set.
13
OMICRON Test Universe
In the F1 drop-down menu, select the first channel that measures frequency.
This channel is also used as reference channel for phase measurement. (By
default, it always shows 0° phase).
In the F2 drop-down menu, you can select a second channel for frequency
measurement. The averaging factor helps smooth out noisy signals. If
measuring a steady state signal, setting the averaging factor to medium or
high increases the accuracy. If the signal varies rapidly or has a complicated
wave form, it's better to switch this off.
The averaging applies to all readings (i.e., RMS, phase, frequency, and
power). The averaging works by "remembering" previous values. It uses the
equation:
result = newValue * averfact + oldValue * (1 - averfact)
newValue value of the new sample
oldValue value of the previous samples (also averaged)
result newly calculated value
averfact averaging factor.
1 for no averaging
0.6 for medium averaging, and
0.1 for high averaging.
The sampling frequency determines the accuracy of the recorded signal. If
the measured signal varies significantly in time or has a complicated wave
form, increasing the sampling rate will help measure the signal accurately.
The 9 kHz sampling frequency is adequate for normal sinusoidal signals.
14
The EnerLyzer Test Module
In the Refresh rate edit box, enter a value which determines how often the
information on the screen gets updated. The refresh rate refers to the screen
parameters and not to the actual sampling of signals.
The phase, voltage, and current are updated at the specified refresh rate. If
the signal being measured is greater than the maximum range then an
"Overload" message is displayed in the corresponding cell.
If no measurement results are available, then a blank is displayed in the
corresponding cell. This happens for a short period when the measurement
is started. It also can occur for the phase and frequency if the signal is too
weak to measure.
15
OMICRON Test Universe
The real and apparent power values are correct for all waveforms (i.e.,
including harmonics). Reactive power will show only the correct values for
pure sine waves.
If there is more than one power system, the real, apparent and reactive
power and the DC component of each system are added together and
displayed in a separate row.
The power sum is only correct under the following conditions:
• the real and reactive powers are of the same sign for each power system,
• the waveforms are sinusoidal,
• the frequency of all power systems are the same.
A maximum of four power signals can be measured simultaneously.
If no power systems are assigned then this grid is not shown at all.
16
The EnerLyzer Test Module
17
OMICRON Test Universe
A right click any of the diagrams pops up a context menu at which the Auto
Zoom feature can be turned on or off, and the display range be zoomed in or
out manually.
18
The EnerLyzer Test Module
The Transient Recording mode lets you to capture a transient record in the
CMC test set. You can then display that record.
Trigger conditions
Set either an immediate, a power quality, or a basic trigger condition. For
power quality triggers, there are 6 types available to choose from: Sag, Swell,
Harmonic, Frequency, Frequency Change and Notch.
Note: CMB IO-71 or ISIO 200 binary input channels can be used for
recording, only, not for triggering. For triggering, use
CMC 356/CMC 256plus/CMC 256 inputs.
1
CMB IO-7 is a meanwhile discontinued product
19
OMICRON Test Universe
Acquisition parameters
Here you specify the parameters for the signal recording to a COMRADE
file.
Saving data
Specify how and where to save the records to the COMTRADE file.
For each record
• Create new file:
Creates a new COMTRADE file for each transient record uploaded
from the PC. In this mode, each new file name has an integer
appended to the file name to make it unique (e.g., Event1.cfg,
Event2.cfg, etc.).
• Overwriting existing file:
Uses the same existing COMTRADE file each time a record is
uploaded from the CMC test set.
Saving records
• First only:
Saves the first record sent to the PC, only. After saving the record,
recording automatically stops.
• Automatically:
Continuously records and uploads the records to the PC. Each
uploaded transient record is converted to a COMTRADE file, and
saved under a unique name.
20
The EnerLyzer Test Module
• On demand:
The Save now button is enabled whenever there is a record available
in the CMC test set. Click Save now to upload and save that record to
the PC.
File specification
Either enter the file and the location of the generated COMTRADE file, or
click Browse to search for the file and its location.
The field below then displays the transient records. This is useful for
Advanced TransPlay and TransView.
Input configuration
Input Configuration provides a way of configuring the binary/analog inputs
of the CMC test set, of the 2 & 3-phase systems, and of the power systems.
Consider it a replacement of the Hardware Configuration, which is still
available, though. To change the input configuration, double-click any of the
ten binary/analog inputs. This opens a dialog box where you can configure
the inputs.
Virtual Inputs
In Transient Recording mode, EnerLyzer supports the use of virtual inputs.
Virtual inputs are a "virtual extension" of the number of available binary
inputs. Physically they do not exist (as an input connector at the CMC test
set); internally, however, they are handled like real inputs by the
Test Universe software. You can use these virtual inputs, for example, to
map GOOSE messages in cases where the regular BINARY INPUTs of your
CMC test set are already used up by analog input signals.
→ See the Hardware Configuration Help topic General Tab (subheading
Virtual Inputs/Outputs) for more information about defining and mapping
virtual inputs/outputs...
EnerLyzer displays used virtual inputs with an additional string, such as "+ 2
virtual binary inputs", next to the BINARY INPUT 10 symbol at the right of the
Input configuration group.
21
OMICRON Test Universe
22
The EnerLyzer Test Module
23
OMICRON Test Universe
24
The EnerLyzer Test Module
25
OMICRON Test Universe
1. = 20 ms duration 2. = 10 % magnitude
3. = RMS = 100 % 4. = RMS = 90 %
26
The EnerLyzer Test Module
Harmonic Trigger:
The CMC calculates the harmonics by means of a DFT (Discrete Fourier
Transformation) every 2 cycles and then, if the specified level is exceeded, a
trigger occurs.
The THD is calculated with the following formula:
THD = √(URMS2 / Uh12 -1)
where is Uh1 is the fundamental.
Harmonic: You can choose the specific harmonic (2nd - 7th), or the
THD (total harmonic distortion) to trigger on.
Level: This trigger activates when the harmonics distortion exceeds
this lever. It is expressed in terms of percentage of the
fundamental.
The minimum level for the harmonics is 1 %.
The minimum level for the THD trigger is 3 %, but it is
recommended to enter at least 5 % because the trigger is not
very accurate below this level.
Note: If the THD trigger is used then it is recommended that the
nominal current / voltage must be equal or above 30 % of the
input range, otherwise the measurement may be inaccurate.
Frequency Trigger:
The Frequency trigger activates when the measured frequency goes outside
the specific deviation. For example, if the nominal frequency is 50 Hz and the
deviation is 1 Hz, then it will trigger if the frequency drops below 49 Hz or
rises above 51 Hz. The frequency is measured by measuring the time of the
zero crossing for several cycles (determined by the resolution) and
calculating the average.
Deviation: The deviation from the nominal frequency that causes a
trigger.
Resolution: The resolution is the amount of cycles over which the
frequency is measured. Entering a small value allows you to
detect short variations in frequency, while entering a large
value makes the measurements more accurate and makes
it less likely to trigger on short variations.
27
OMICRON Test Universe
28
The EnerLyzer Test Module
Typical notch:
1. = Duration 2. = Depth
Duration: The duration of the notch (see diagram). The value must be
between 106 μs and 2 ms.
Depth: The depth of the notch (see diagram).
Quantity: The trigger activates after this amount of notches have
occurred within a ten-cycle period.
Indication of invalid parameters:
Invalid parameters are marked with a red border around the
edge of the cell.
29
OMICRON Test Universe
30
The EnerLyzer Test Module
31
OMICRON Test Universe
32
The EnerLyzer Test Module
33
OMICRON Test Universe
34
The EnerLyzer Test Module
35
OMICRON Test Universe
36
The EnerLyzer Test Module
The bottom row of the grid has a button "Calculate Average". When this
button is pressed the average value of the selected signal is calculated. Only
the sample points that fall between the two cursors are used to calculate the
average.
The total recording time is shown to the right of the grid.
37
OMICRON Test Universe
Reactive power:
Select this check box to record the reactive power for all configured power
systems.
Cos Phi:
Select this check box to record the power systems.
Note: Cos Phi also displays the power factor for a three-phase system. If
more than two power systems are configured than the power sum will also
be plotted for real, apparent, reactive and Cos Phi.
Capacity remaining:
This shows you how many resources are available to make measurements.
There are 13 tasks available initially and this number will be reduced
depending on the amount of channels configured and the amount
measurements selected. Also the 28 kHz sampling frequency allows more
tasks to be used.
Measurement rate:
This is the rate at which the data is plotted. For example if the measurement
rate is one second, then a new point will be plotted on the graph each
second.
Sampling frequency:
The sampling frequency determines the accuracy of the recorded signal.
Normally the 9 kHz sampling frequency is sufficient, however for phase
measurement at a 60 Hz nominal frequency, the 28kHz sampling frequency
is recommended.
38
TransView
4 TransView
The TransView test tool assists you with the analysis of fault records. In
addition to the values measured and recorded in the fault record, the program
can also be used to calculate other variables, such as impedances, RMS
values, etc.
The TransView test tool is accessible from EnerLyzer only after a record has
been acquired in the transient recording mode of EnerLyzer.
Figure 4-1:
Transient Recording
dialog box
Analyze
The Analyze button is used to start up the TransView test tool with the
selected record.
39
OMICRON Test Universe
The TransView test tool offers a graphical display of the measured and
calculated values and the binary signals.
These include:
• Time signal diagrams (refer to section 4.2 on page 45).
• Harmonics charts (refer to section 4.3 on page 46).
• Vector diagrams (refer to section 4.4 on page 47).
• Impedance “Circle” diagrams (refer to section 4.5 on page 48).
It is possible to assign any number of diagrams to the different views and any
number of signals to the different diagrams. Signal assignment is performed
with the help of the Assign Signals matrix.
40
TransView
41
OMICRON Test Universe
42
TransView
43
OMICRON Test Universe
44
TransView
45
OMICRON Test Universe
4.3 Harmonics
The Harmonics view shows the RMS values of harmonics of selected
measuring values in the form of bar charts. The harmonics are determined
with the help of a DFT (Discrete Fourier Transformation).
Figure 4-3:
Harmonic diagrams in the
TransView test tool
The measuring window is always placed to the left of the reference point
(cursor position) and its length corresponds to one period of the nominal
frequency TN (e.g., 20 ms at 50 Hz, 16.6 ms at 60 Hz).
The calculated variables are valid only if no status change (fault inception,
disconnection, etc.) occurs within the measuring window. A separate
diagram is created for each measured and calculated variable to be
analyzed.
The RMS value and the percentage of the fundamental component are
displayed above each bar.
In a table you can read the current position of cursor 1 on the time axis. To
position the cursor at a specific point in time, enter the desired time value in
this table. TransView then moves the cursor automatically to the
corresponding point on the time axis.
46
TransView
The measuring window is always placed to the left of the reference point
(cursor position) and its length corresponds to one period of the nominal
frequency TN (e.g., 20 ms at 50 Hz, 16.6 ms at 60 Hz).
The calculated variables are valid only if no status change (fault inception,
disconnection) occurs within the measuring window.
The vector angle for currents and voltages always refers to a standard vector
e j2πfN rotating at nominal frequency (fN).
In addition to the graphical display, individual values can be entered in a
table. In this table you can also read the current positions of cursor 1 and
cursor 2 on the time axis.
The data at specific points in time can be recorded by either moving the scroll
bar cursors to the time or by entering the time in the respective time fields.
The signal name, the value, and the point in time are then shown in this table.
47
OMICRON Test Universe
48
Working with Current Clamps
49
OMICRON Test Universe
50
Working with Current Clamps
To analog
input of
Shunt R CMC test
set
Current clamp
WARNING
Death or severe injury caused by high voltage or current
► Do not open the current loop!
51
OMICRON Test Universe
Example:
The clamp data is often specified as a real current ratio.
Therefore, 500 A:5 A means r = 100. Let’s say the nominal burden is
R = 200 mΩ. The clamp ratio in EnerLyzer then is
200 mV/A / 100 = 2 mV/A.
The optimal voltage range for a 50 A current range would then be 100 mV.
52
Working with Current Clamps
5. Function
Specify in the function row that the channel is being used to measure
currents. This is accomplished using a drop-down menu as is shown for
binary channel 5 in Figure 5-3.
6. Nominal Range
Enter the nominal range for the current clamp. The nominal range is the
dynamic maximum (peak value) that is expected to appear on the channel
from the current clamp.
7. Clamp Ratio
Enter the current clamp ratio. This setting in the Hardware Configuration
dialog box should match the settings of the switches on the actual current
clamp. The clamps used in this example have a clamp ratio of 100 mV/A
nominal range, as is shown for channel 6 in Figure 5-3.
53
OMICRON Test Universe
54
Working with Precision Shunts
55
OMICRON Test Universe
DANGER
Death or severe injury caused by high voltage or current
► Do not disconnect the measuring circuit from its load while current is still
flowing!
► A typical use case is working at the secondary of a transformer.
Opening the measuring circuit at the secondary of a transformer, while
there is current in the primary, will cause the transformer secondary to
continue driving current across the effectively infinite impedance up to
its core saturation voltage. This may produce a high voltage across the
open secondary into the range of several kilovolts, causing arcing,
compromising operator and equipment safety, or permanently affecting
the accuracy of the transformer.
56
Working with Precision Shunts
4. Function
Specify in the function row that the channel is being used to measure
currents. This is accomplished using a drop-down menu as is shown for
binary channel 5 in Figure 6-3.
5. Nominal Range
Enter the nominal range for the C-Shunt. The nominal range is the
dynamic maximum (peak value) that is expected to appear on the channel
from the C-Shunt.
6. Clamp Ratio
Enter the current ratio in Clamp Ratio.
7. The left-hand column of the Hardware Configuration dialog box is used
to specify the test module input signal. Select the name from the drop-
down menus.
8. The second column from the left in the Hardware Configuration dialog
box is used to enter a “Display Name” for subsequent dialog boxes. Enter
the display name.
9. Assign the signals. The remaining portion of the Hardware
Configuration dialog box is a table for establishing logical connections
between the input signals or display names and the physical channels on
the CMC test set. Each input signal can be assigned to only one channel
and vice versa.
10.Physically connect the C-Shunt connectors to the binary/analog inputs on
the front panel of the CMC test set.
57
OMICRON Test Universe
58
EnerLyzer Multimeter Mode Example
Fixed connection
Current probe
If the physical wiring to the various channels differs from the wiring here, the
logical connections can be changed in the Hardware Configuration dialog
box (refer to Figure 7-4 on page 63).
59
OMICRON Test Universe
5. Repeat these steps for each of the three required current clamps.
6. When finished with all current clamps, exit the QuickCMC test module.
60
EnerLyzer Multimeter Mode Example
5. Function
For each input channel 1 through 3, specify in the function row that the
channel is being used to measure voltage (instead of as a binary input,
counter input, or current input). This is accomplished using a drop-down
menu as is shown for binary channel 1 in Figure 7-2.
6. Nominal Range
For each voltage channel, enter the nominal range for the voltage. The
nominal range is the dynamic maximum that is expected to appear on the
channel. The voltage channels In this example have a 600 V nominal
range, as is shown for channel 2 in Figure 7-2.
61
OMICRON Test Universe
Figure 7-3:
Partial view of Hardware
Configuration dialog box
for analog current channel
7. Function
For each input channel 5 through 7, specify in the function row that the
channel is being used to measure currents. This is accomplished using a
drop-down menu as is shown for binary channel 5 in Figure 7-3.
8. The binary/analog channel 4 is not being used in this example. These
channels are divided into five groups of two with respect to being
galvanically isolated from one another. Therefore, channels 3 and 4 share
the same ground. In order to keep the three phase voltages separated
from the ground used to measure the currents, the current inputs are
placed on channels 5 through 7 and channel 4 and channel 8 are not
used.
9. Nominal Range
For each current channel, enter the nominal range for the current clamp.
The nominal range is the dynamic maximum that is expected to appear
on the channel from the current clamp. The clamps used in this example
have a 1 V nominal range, as is shown for channel 6 in Figure 7-3.
10.Clamp Ratio
For each current channel, enter the current clamp ratio. This setting in the
Hardware Configuration dialog box should match the settings of the
switches on the actual current clamp. The clamps used in this example
have a clamp ratio of 100 mV/A nominal range, as is shown for channel 6
in Figure 7-3.
11.The inputs of the CMC test set were intended for sinusoidal signals. As
such, the nominal range can be considered Vrms. Together with a clamp
ratio, it specifies how large peak values can be before clipping occurs.
This is important to remember for non-sinusoidal signals that may be
within the Vrms nominal range but that may have larger peak values which
get clipped.
62
EnerLyzer Multimeter Mode Example
63
OMICRON Test Universe
Figure 7-5:
Multimeter Configuration
dialog box
64
EnerLyzer Multimeter Mode Example
WARNING
Death or severe injury caused by high voltage or current
► Do not attach wires between the CMC test set and the motor while
power is still being supplied to the motor on its input lines.
► Turn off voltage and currents to the motor before attaching probes.
In this example, the plan was to monitor the voltages and currents going into
a three phase, grounded motor. Figure 7-6 shows the intended wiring for the
front panel BINARY/ANALOG INOUTs of the CMC test set. Channels 1
through 3 are to measure the phase voltages going into the motor, while
channels 5 through 7 have use current clamps to measure the currents going
into each of the motor’s windings.
Figure 7-6:
Measuring currents and
voltages on a three-phase
motor
Fixed connection
Current probe
Note that channel 4 is not being used, because it shares a ground with
channel 3. Likewise, channel 8 is also not being used. If the physical wiring
to the various channels differs from the wiring here, the logical connections
can be changed in the Hardware Configuration dialog box (refer to Figure
7-4 on page 63).
65
OMICRON Test Universe
66
EnerLyzer Transient Mode Example
Figure 8-1 shows the schematic of the YD5 transformer and its differential
relay. According to the specifications for the YD5 transformer, the output
currents are 5 x 30° (or 150°) out of phase with the input currents in addition
to any transformer current scaling.
The relay is attached to the input and output side of the transformer using
current transformers. The input currents of the differential relay IB1, IB2, and
IB3 are proportional to the input currents of the transformer. Likewise, the
input currents on the other side of the differential relay IA1, IA2, and IA3 are
proportional to the output currents of the transformer.1
Under normal operating conditions, the differential relay trips when an input
current, such as IB2, is different from its associated output current, in this
case IA2. The differential relay already takes into consideration the effects of
transformer current scaling and phase shifts between input and output.
Differences that trip the relay can be the result of, say, additional ground
currents in a phase or line to line currents, neither of which should be present
for sound operation of the transformer.
1
The currents in Figure 8-1 are shown going into the differential relay. Actual current direction
depends on wiring for the current transformer and settings for differential relay.
67
OMICRON Test Universe
68
EnerLyzer Transient Mode Example
5. Repeat these steps for each of the six required current clamps.
6. When finished with all current clamps, exit the QuickCMC program.
69
OMICRON Test Universe
5. Function
For each input channel 1 through 6, specify in the function row that the
channel is being used to measure currents (instead of as a binary input,
counter input, or voltage input). This is accomplished using a drop-down
menu as is shown for binary channel 1 in Figure 8-2.
6. Nominal Range
For each current channel, enter the nominal range for the current clamp.
The nominal range is the dynamic maximum that is expected to appear
on the channel from the current clamp. The clamps used in this example
have a 1 V nominal range, as is shown for channel 2 in Figure 8-2.
7. Clamp Ratio
For each current channel, enter the current clamp ratio. This setting in the
Hardware Configuration dialog box should match the settings of the
switches on the actual current clamp. The clamps used in this example
have a clamp ratio of 100 mV/A, as is shown for channel 2 in Figure 8-2.
The inputs of the CMC test set were intended for sinusoidal signals. As
such, the nominal range can be considered Vrms. Together with a clamp
ratio, it specifies how large peak values can be before clipping occurs.
This is important to remember for non-sinusoidal signals that may be
within the Vrms nominal range but that may have larger peak values which
get clipped.
Specifying a larger range can prevent clipping. However, doing so results
in a loss of resolution, because the analog-to-digital conversion only has
12 bits to represent the entire range.
Figure 8-3:
The completed
configuration of the input
current clamps
70
EnerLyzer Transient Mode Example
9. The second column from the left in the Hardware Configuration dialog
box is used to enter a “Display Name” for subsequent dialog boxes. In this
example for measuring the in-rush currents, the display names chosen
are IB1 through IB3 for the input currents and IA1 through IA3 for the
output currents, as is shown in Figure 8-3.
10.The remaining portion of the Hardware Configuration dialog box is a
table for establishing logical connections between the input signals or
display names and the physical channels on the CMC test set. Each input
signal can be assigned to only one channel and vice versa.
Figure 8-3 shows the logical connections for this example and the
intended physical wiring of the current clamps to the CMC test set. If the
actual wiring of the current clamps is different (for instance, if the IA inputs
are exchanged for the IB inputs), the logical connections can be used to
make the associations instead of physically moving cables on the front
panel of the CMC test set.
71
OMICRON Test Universe
72
EnerLyzer Transient Mode Example
Figure 8-4:
Transient Recording dialog
box
73
OMICRON Test Universe
• Enter a value for the recording length in the Acquisition Length edit
box. This defines the duration for the recording. The acquisition length
is 1000 ms for this example. Note that the acquisition length includes
the pre-trigger length.
• Select the desired sampling frequency. The sampling frequency has
an effect on the maximum acquisition length. The higher the sampling
frequency, the higher the resolution of the recording at the expense of
a shorter maximum recording time. On the other hand, the lower the
sampling frequency, the longer the maximum recording time at the
expense of a lower resolution. The maximum recording time can only
be modified by changing the sampling frequency. The sampling
frequency is 9 kHz for this example.
5. In the “Saving records” section of the transient recorder interface, choose
one of the following options:
First only Saves only the first record sent to the PC. After
saving the record, recording is automatically
stopped.
Automatically Records and uploads the records to the PC
continuously. Each transient record uploaded is
converted to a comtrade file and saved to a unique
name.
On demand Records and uploads a record to the PC only when
the “Save now” button is activate. The “Save now”
button is only active when there is a record
available in the CMC.
The settings should be for First only for this example.
6. The “For each record” section of the dialog box has mutually exclusive
radio buttons that can be set:
Create new file Creates a new comtrade file for each transient
record uploaded from the computer. In this mode
each new file name has an integer appended to the
file name to make it unique (e.g., Event1.cfg,
Event2.cfg).
Overwriting
existing file Uses the same existing comtrade file each time a
record is uploaded from the CMC.
The settings should be for “Create new file” for this example.
7. In the File specification edit field, either enter the file and the location of
the generated comtrade file or click Browse to search for the file name
and location. For this example, the location and file name are:
C:\Program Files\OMICRON\Test Universe\Test Library\
Samples\Record.cfg.
74
EnerLyzer Transient Mode Example
75
OMICRON Test Universe
Note: The TransView test tool is accessible from the Transient Recording
dialog box after a record has been acquired in the EnerLyzer test tool.
TransView offers a graphical display of the measured and calculated values
and the binary signals. These include time signal diagrams, vector diagrams,
harmonics charts and impedance “circle” diagrams
76
EnerLyzer Transient Mode Example
77
OMICRON Test Universe
78
EnerLyzer Transient Mode Example
79
OMICRON Test Universe
80
Support
Support
When you are working with our products we want to provide you with the
greatest possible benefits. If you need any support, we are here to assist you!
OMICRON electronics GmbH, Oberes Ried 1, 6833 Klaus, Austria, +43 59495
81
Support
82
Index
acquisition
of transient records . . . . . . . . . . . . . . . . . 30
setting parameters . . . . . . . . . . . . . . . . . . 30
Adv. TransPlay test module . . . . . . . . . . . . . 79
D
analog input Depth (power quality trigger) . . . . . . . . . . . . 29
configuring current signals . . . . . . . . . 53, 57 Deviation (power quality trigger) . . . . . . . . . . 27
analyze DFT . . . . . . . . . . . . . . . . . . . . . . . . . 46, 47, 78
the Analyze button . . . . . . . . . . . . . . . . . . 39 Diagram elements
Apparent power (Trend Recording Conf.) . . . 37 zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
ASCII (Comtrade option) . . . . . . . . . . . . . . . . 44 Diagram Error Status (Trend Recording
Auto scale (Trend Recording mode) . . . . . . . 35 Mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Auto-adjust resolution (power quality trigger) . . 25 Diagram line (diagram element) . . . . . . . . . . 40
available (power quality trigger) . . . . . . . . . . 25 Diagram menu (Trend Recording Mode) . . . 35
averaging (Multimeter Conf.) . . . . . . . . . . . . 14 Diagram properties (diagram element) . . . . . 41
Diagram stop/start indication (Trend
Recording Mode) . . . . . . . . . . . . . . . . . . . . . 36
Diagrams
B impedance circle . . . . . . . . . . . . . . . . . . . 48
Duration (power quality trigger) . . . . . . . . . . 29
basic trigger . . . . . . . . . . . . . . . . . . . . . . . . . 24
Binary (Comtrade option) . . . . . . . . . . . . . . . 44
BINARY/ANALOG INPUT of CMC . . . . . . . . . 7
E
error
C display of error conditions (Harmonic
Analysis mode) . . . . . . . . . . . . . . . . . . . . 34
Capacity remaining (Trend Recording display of error conditions (Multimeter
Configuration) . . . . . . . . . . . . . . . . . . . . . . . . 38 mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Channel (trigger conditions) . . . . . . . . . . . . . 24 overload error . . . . . . . . . . . . . . . . . . . . . 35
Chart cursors (Trend Recording Mode) . . . . 36
Clamp ratio . . . . . . . . . . . . . . . . . . . . . . . 13, 52
clamp ratio . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Clear (Trend Recording Mode) . . . . . . . . . . . 35
CMC (suitable test sets for EnerLyzer) . . . . . . 5
F
Comtrade . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Frequency (Trend Recording Conf.) . . . . . . . 37
comment (Comtrade options) . . . . . . . . . 44 Frequency (trigger conditions) . . . . . . . . . . . 24
format (Comtrade options) . . . . . . . . . . . . 44 Frequency and phase measurement (Trend
options . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Recording Mode) . . . . . . . . . . . . . . . . . . . . . 36
configuration Frequency change (trigger conditions) . . . . . 24
analog input . . . . . . . . . . . . . . . . . . . . 53, 57 Frequency change trigger (power quality
test hardware . . . . . . . . . . . . . . . . . . . . . . 12 trigger) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
trend recording . . . . . . . . . . . . . . . . . . . . . 37 frequency trigger (power quality trigger) . . . . 27
Configuration change (Trend Recording
Mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
83
OMICRON Test Universe
H O
Hardware Configuration . . . . . . . . . . . . . . . . 12 Optimize
Harmonic (power quality trigger) . . . . . . . . . . 27 x-axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Harmonic (trigger conditions) . . . . . . . . . . . . 24 y-axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
harmonic trigger (power quality trigger) . . . . 27 overload error . . . . . . . . . . . . . . . . . . . . . . . . 35
I P
Input configuration Phase (Trend Recording Conf.) . . . . . . . . . . 37
creating 2 & 3 phase systems . . . . . . . . . 22 playback
creating power systems . . . . . . . . . . . . . . 22 transient record (Advanced TransPlay) . . 30
floating configure input . . . . . . . . . . . . . . . 22 power
Invalid parameters (power quality trigger) . . . 29 meas. real/apparent/reactive power . . . . 59
Power (Multimeter vector diagrams) . . . . . . 18
power quality trigger . . . . . . . . . . . . . . . . . . . 25
Primary transformer ratio (Input
L Configuration) . . . . . . . . . . . . . . . . . . . . . . . . 23
M
Measurement rate (Trend Recording Conf.) . 38
Multimeter
R
configuration . . . . . . . . . . . . . . . . . . . 14, 53 Range (Input Configuration) . . . . . . . . . . . . . 23
power grid . . . . . . . . . . . . . . . . . . . . . . . . 16 Rate (power quality trigger) . . . . . . . . . . . . . 28
signals grid . . . . . . . . . . . . . . . . . . . . . . . . 15 ratio
vector diagrams . . . . . . . . . . . . . . . . . . . . 18 current clamp ratio . . . . . . . . . . . . . . . . . 23
Multimeter mode . . . . . . . . . . . 8, 11, 12, 18, 53 primary transformer ratio . . . . . . . . . . . . . 23
secondary transformer ratio (Input Conf.) . . 23
Reactive power (Trend Recording Conf.) . . . 38
Real power (Trend Recording Conf.) . . . . . . 37
N recording
of transient records . . . . . . . . . . . . . . . . . 30
Name (input configuration) . . . . . . . . . . . . . . 23 setting parameters . . . . . . . . . . . . . . . . . 30
Nominal Recording limit (Trend Recording Mode) . . . 36
frequency (Comtrade options) . . . . . . . . . 44 Reference channel (Trend Recording Conf.) . . 37
frequency (power quality trigger) . . . . . . . 25 Resolution (power quality trigger) . . . . . . 27, 28
power quality trigger . . . . . . . . . . . . . . . . . 25 RMS (Trend Recording Conf.) . . . . . . . . . . . 37
range (Hardware Configuration) . . . . . . . 12
Notch trigger (power quality trigger) . . . . . . . 28
Notches (trigger conditions) . . . . . . . . . . . . . 25
84
Index
S W
sag weak signal error (Trend Recording Mode) . 36
typical sag . . . . . . . . . . . . . . . . . . . . . . . . 26
Sag and swell (trigger conditions) . . . . . . . . . 24
sampling frequency (record accuracy) . . . . . 14
Sampling frequency (Trend Recording
Conf.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Y
secondary transformer ratio (Input Conf.) . . . 23 YD5 transformer (Transient mode example) 67
Selected grid (power quality trigger) . . . . . . . 25
Signal properties (diagram element) . . . . . . . 41
Signals (Multimeter vector diagrams) . . . . . . 18
Slope (trigger conditions) . . . . . . . . . . . . . . . 24 Z
Snapshot view . . . . . . . . . . . . . . . . . . . . . . . . 33 zoom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Status (Trend Recording Mode) . . . . . . . . . . 35 Zoom (diagram element) . . . . . . . . . . . . . . . 41
Swell/Sag (power quality trigger) . . . . . . . . . 26
Symmetrical components (Multimeter vector
diagrams) . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
T
Table (diagram element) . . . . . . . . . . . . . . . . 40
test sets suitable for EnerLyzer . . . . . . . . . . . . 5
Threshold (Input Configuration) . . . . . . . . . . 23
Time Signals
viewing time signals in TransView test
tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Time started (Trend Recording Mode) . . . . . 35
transient record playback (Advanced
TransPlay) . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
transient recording . . . . . . . . . . . . . . . . . . 9, 11
TransView . . . . . . . . . . . . . . . . . . . . . 40, 77, 79
Trend recording mode . . . . . . . . . . . . . . . 11, 35
trigger
basic trigger . . . . . . . . . . . . . . . . . . . . . . . 24
power quality trigger . . . . . . . . . . . . . . . . 25
Trigger channel (power quality trigger) . . . . . 25
Typical notch (power quality trigger) . . . . . . . 29
V
Vector diagrams . . . . . . . . . . . . . . . . . . . . . . 47
View properties (diagram element) . . . . . . . . 40
85
OMICRON Test Universe
86