1.0 PURPOSE

1.1 This procedure describes the operation and routine maintenance of the
CLiMET® CI-450 Laser Particle Counter.

2.0 SCOPE

2.1 This SOP includes the procedure for operation and routine maintenance of
the CLiMET® CI-450 Laser Particle Counter.

3.0 REFERENCES

3.1 None

4.0 DEFINITIONS

4.1 None

5.0 RESPONSIBILITIES

5.1 It is the responsibility of the person/s operating and performing routine


maintenance on the CLiMET® CI-450 Laser Particle Counter to adhere to
this SOP.

6.0 PROCEDURE/PROCESSES

6.1 Sample Volume

6.1.1 Turn the particle counter on by pressing “POWER” button.

6.1.2 From the Main Menu, press “SET UP”.

6.1.3 Press the “VOLUME” key at the top of the menu area. The sample
volume entry screen will appear.
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.1.4 Press the “CUBIC METER” key, and then press “1”.

6.1.5 Press the enter key to return to the main setup screen.

6.2 Display

6.2.1 From the main setup screen press the “MORE” key. Select “0.5
µm” and “5.0 µm” by pressing those keys. The keys will turn
green.

6.2.2 Press “SET DATE” key. The entry screen will appear. Enter the
date by month/day/year. Then press “ENTER”.

6.2.3 Select the “D/M/Y” key. It will turn green.

6.2.4 Press “SET TIME” key. The entry screen will appear. Enter the
time by XX:XX (hour : minute). This is the 24 hours display of
time (no AM/PM). Then Press “ENTER”.

6.2.5 Press “EXIT” key to come back to main setup screen.

6.3 Initial Delay

6.3.1 From main setup screen press the “INITIAL DELAY” key. The
initial delay entry screen will appear.

6.3.2 Enter 30 for 30 seconds delay, then press “ENTER”.

6.3.3 Press the enter key to return to the main setup screen.
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.4 Number of Samples

6.4.1 To set the CI-450 to take a fixed number of samples:

6.4.1.1 Press the “STOP AFTER N” key from the sample group.
It will turn green.

6.4.1.2 Press the “NUMBER ‘N’” key and enter the desired
number of samples (a minimum of 1 sample of 1 m3 of
air will be sampled).

6.4.2 To set the particle counter to take continuous samples, one sample
after another, until the “STOP” key is pressed:

6.4.2.1 From the main setup screen, press the “NUMBER ‘N’”
key. Press the “INF” key.

6.4.2.2 Press the “ENTER” key.

6.5 Auto-average

6.5.1 Press the “PRINT AVERAGES AFTER N” key. The key will turn
green.

6.5.2 The number of samples to be averaged is determined by the


number of samples entered.

6.6 Total mode

6.6.1 From the main setup screen, press “TOTAL”. The “TOTAL” key
will turn green.

6.6.2 Press “COUNTS”. The “COUNTS” key will turn green.


PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.7 End of sample printouts

6.7.1 From the main setup screen, press the “PRINT END OF
SAMPLE” key. It will enable when the key is green.

6.8 Press “SAVE SETTING” to save the setting, and press “EXIT” to exit the
main setup screen.

6.9 Sampling

6.9.1 From the main menu, press “PAPER FEED” to feed the paper up.
Deleted: probe
6.9.2 Attach a 0.2 µm filter into the probe tubing. Connect the tubing
into the inlet. The probe should be vertical to the sampling
surface. Press “START” to purge the particle counter for 1 cycle
(1 cubic meter).
Note: Run the purge cycle (Zero Count) only when sample taken is in an
ISO class 5 room.
Deleted: purge cycle
6.9.2.1 The “Zero Count” must have a zero count on both 0.5
µm and 5.0 µm.

6.9.2.2 If any counts are obtained, repeat the purge cycle until a
zero count is obtained.

6.9.3 After purge cycle is completed, remove the filter.

6.9.4 Press “START” to run the particle counter. The printing result will
automatically print out at the end of sampling.
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.10 Security

6.10.1 Level 4: allows free access to all menus and functions.

6.10.2 Level 3: prevents entry into the set up screen.

6.10.3 Level 2: prevents entry into the set up and load settings screens.
This level allows access to the report menu screens.

6.10.4 Level 1: prevents entry into the set up, load settings, and report
menu screens.

6.11 Changing Paper

6.11.1 Turn the unit on. Remove the printer paper compartment cover in
the upper right corner of the rear panel by unscrewing the two
thumbscrews by hand.

6.11.2 Cut or tear any remaining paper near the compartment, and then
hold down the paper advance key to remove old paper still in the
printer.

6.11.3 Hold the new roll of paper in the palm of your hand so that the end
of the paper extends from the bottom of the roll out to your
fingertips.

6.11.4 Use your fingertips to insert the end of the paper into the printer
mechanism located at the upper rear of the compartment.

6.11.5 Insert the paper behind the visible horizontal lip. The printer will
automatically sense and draw the paper through the mechanism.
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.11.6 If the paper does not feed through after the paper advance stops
running, pull the paper back out, and cut a fresh edge on the end of
the paper. Make sure that the paper is straight rather than curled so
that it does not curl in front of the lip.

6.12 Calibration

6.12.1 The CLiMET® CI-450 Laser Particle Counter should be calibrated


annually by external qualified service.

6.13 Cleaning

6.13.1 Cleaning will be performed when placing the CI-450 in a control


area.

6.13.2 Turn the unit off and remove the power cord.

6.13.3 Cap the inlet or attach a sample probe to the inlet to prevent fluid
from entering the inlet.

6.13.4 Cover the electrical connectors on the rear panel with the caps
provided.

6.13.5 Wipe the surface of the unit with disinfectants. Do not allow the
disinfectant to saturate the printer mechanism.

6.14 Storage

6.14.1 If the CLiMET® CI-450 is to be stored for three months or longer


without use, the battery should be disconnected.

6.14.1.1 Open the battery access door at the rear of the unit.
Locate the battery connector and disconnect it from the
battery by separating the connector.
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

6.14.1.2 It is adequate to store the disconnected battery in the


unit.

6.14.1.3 Reconnect the battery again before use. Make sure that
the battery wires are not pinched against the metal edges
of the enclosure. The battery typically lasts 6 months to
a year. Failure to hold a charge for more than two hours
is an indication that the battery is failing and should be
replaced.

7.0 ATTACHMENTS

7.1 None

8.0 REVISION HISTORY


DCR # Revision Effective Date Reason
00 New Doc
PHARMACEUTICAL SERVICES CORPORATION
STANDARD OPERATING PROCEDURE

SOP Number:
Operation and Maintenance of the CLiMET® CI-450
Title/Description:
Laser Particle Counter
Revision No.

Approved:

Nicole C. Canto, PSC Validation Engineer date

Approved:

Kevin Hines, Sr. Project Manager date

Potrebbero piacerti anche