LICNAME INSTITUT MARINA 18/04/2018

Fluke Biomedical Ansur


Test and Inspection Procedure
Copyright © 2000 - 2014 Fluke Biomedical

Test Record
TEST PASSED
Test Performed Ansur Components Used
Date: 18/04/2018 Ansur Version 3.0.0
TEST ELE
Record: Plug-In: ESA620 Version 1.1.11
INCUBADORA.mtr
IEC 62353 Alternative
Template:
Leakage - CL2.mtt
Template Version: 1.0.0

Test Setup
Selections
Service Events Performed Standards Performed
IEC 62353 (CL2)

Device under test


Serial Number HCDG50344 Type CLASE II
Appliance Code - Model CARE PLUS
Group INCUBADORAS Location -
Status - Address 1 -
Manufacturer OHMEDA MEDICAL Address 2 -

MTI Data
Test Instrument Serial Number Firmware Version
ESA 620 3617067 v2.11

Signatures

Test Record 1 Page 1 of 2


Fluke Biomedical Ansur Test Report LICNAME INSTITUT MARINA 18/04/2018

Test Result
Test Element Test Type Fail
IEC 62353 Alternative Leakage - Class II Auto Sequence

Mains Voltage Mains Voltage

Mains Voltage
Live to Neutral Live to Neutral
Result: Value Unit High Limit Low Limit Standard
Live to Neutral 229,7 V IEC 62353 (CL2)

Alternative Equipment Leakage Alternative Equipment Leakage

Alternative Equipment Leakage


Open Earth Open Ground
Result: Value Unit High Limit Low Limit Standard
Open Earth 0,6 uAAC 500 IEC 62353 (CL2)

Insulation Resistance Insulation Resistance


Configuration:
Test Voltage: 500V

Test Record 1 Page 2 of 2