Olympus AU480 ® Chemistry Analyzer

General Characteristics ISE Module


Test menu selection: >125 Analytical System: Vertical block flow ion selective electrode cell
On-board Parameters: 60 measured tests + 3 ISEs Analytical Tests: Na, K, Cl
On-board Pre-programmed tests: 120 Sample Throughput: 200 per hour
Maximum Tests/Hour: 400 photometric tests/hr and Max Tests/Hr: 600 if ISE mode only
800 tests/hr with ISE Sample Type: Serum, plasma, urine
Analytical System: Fully automated random access chemistry analyzer Sample Volume: 20 µL serum, 20 µL urine
Methods: Endpoint Final Dilution Ratio: 1:32.4 (serum), 1:32.4 (urine)
Endpoint with self blanking
Endpoint with sample blanking Measuring Range: Serum and plasma:
Kinetic Sodium 50 – 200 mEq/L
Fixed rate Potassium 1 – 10 mEq/L
ISE Chloride 50 - 200 mEq/L
Urine:
Sample Processing Selection: Single, multiple or profile Sodium 10 – 400 mEq/L
Sample Volume: 1-25 µL in 0.1 µL steps Potassium 2 – 300 mEq/L
Sample Type: Serum, plasma, urine, CSF Chloride 15 - 400 mEq/L
Sample Tray Capacity: 80 samples by rack, 22 position STAT carousel
Sample Container: 12 x 75 – 16 x 100 mm tubes Electrode Composition: Sodium Crown-ether membrane
3, 5, 7, 10 mL primary collection tubes Potassium Crown-ether membrane
Nested micro sample cup Chloride Molecular oriented PVC membrane
Sample Transport: 10 position bar code racks Reference Liquid on liquid membrane
STAT Sample Transport: Up to 22 positions user-definable
for STAT, QC, Cal, routine samples
Sample Quality Analysis: Lipemia, Hemolysis, Icterus Optical Characteristics
Clot detection
Probe crash protection Photometer: Concave diffraction grating as the dispersive element
Sample Identification: Bar code reader [ Maximum of 26 digits ] Light Source: Tungsten-halogen lamp
Sample Bar Code Formats: NW7, CODE 39, CODE 128, ISBT-CODE 128, Photometric Modes: Monchromatic or bichromatic
2 of 5 interleaved, 2 of 5 standard. 2 of 5 is Wavelengths: 340, 380, 410, 450, 480, 520, 540,
the specified digit number. Mixed readable 570, 600, 660, 700, 750, 800 nm
(maximum of 4 types at the same time) Light Path: 5 mm
Reagent Identification: Bar code reader Photodetector: Silicone photodiode array (built-in preamplifier)
Reagent Configuration: Normal strength and concentrated Photometric Range: 0~3.0 Abs (light path length converted to 10 mm)
Reagent Volume: R1. 15-120 µL; R2. 15-250 µL; (in 1 µL increments) Photometric Sensitivity: 0.0001 Abs
Reagent Delivery: One, or two reagents, probe dispensed with
probe crash prevention
Reagent Storage: 4 - 12 °C Data Processor
Reagent Bottle Capacity: 15, 30, 60, mL bottles
Reagent Storage Capacity: R1/R2 = 76 positions
Interface: RS232C, bidirectional
Patient storage: Sample number: 100,000 samples
Reaction Mixture: 90 – 350 µL
Reaction data: 200,000 tests
Reaction Temperature: 37 °C
Data Input: Keyboard,mouse, touchscreen monitor
Reaction Incubation Method: Dry bath
Display: 19” touchscreen monitor
Reaction Cell: Permanent square glass cuvette
Printer: Color laser
(inside dimension 5x5 mm)
Operating System: Windows XP®
Reaction Time: 8 minutes, 37.5 seconds
Repeat Run: Automatic with sample volume increase
or reduction
Reflex Testing: User defined
Sample Pre-dilution: Dilution factors of 3,5,10,15,20,25,50,75,100 fixed
Olympus AU480 ® Chemistry Analyzer

Electrical Requirements Installation Requirements


Electrical Consumption: 3.5 KVA
Current: 20 Amp
Circuit: Dedicated noise free
Voltage: 208/220 VAC [±10%] single phase 19.68”
19.68”
Frequency: 60 Hz (within ± 1 Hz)
Ground Requirements: Resistance < 100 ohms
Location: Power cable length 30 ft.
29.92”

Environmental Requirements
Average heat output: 5500 BTU/Hr. 57.08” 19.68”
Ambient temp: 18 ~ 32 °C (fluctuations during 19.68”
analysis shall be within ± 2 °C)
Ambient humidity: 40 ~ 80% RH (No condensation)
Water consumption: 20 L/hour average
Water type: Deionized CAP Type II, bacteria free
Supply: Continuous flow
Resistivity: Less than 2.0 µS/cm filtered with a 0.5 µm filter
Pressure: 7 - 57 psi Cuvette
Control: Shut off valve within 30 ft. Reagent Probe Washer/Dryer
Mixer Cuvettes

Drain Requirements
Waste: Maximum height – 59 in. Sample Probe

Dimensions Reagent
Width: 57.1 in. Compartment Rack Loader
Depth: 30 in.
Height: 47.5 in.
Weight: 926 lbs.

ISE Module STAT/QC/Cal/Routines

Olympus America Inc.


3500 Corporate Parkway
PO Box 610
AU480SPECV1 09.04

Center Valley, PA 18034-0610


Tel: 1.800.223.0125
www.olympusamerica.com