300| Dräger-Tubes used with Dräger Aerotest

Impactor, Measurement of Oil Mist


Order No. 81 03 560

Allgemeine Daten
Standard Measuring Range:
0.1 mg/m3, 0.5 mg/m3,

ST-357-2008
1.0 mg/m3 Oil mist
(Oil Aerosols)
Detection Limit: 0,05 mg/m3 Oil mist Dräger Impactor
Test Volume: 20 L
Volumenstrom: 4 L/min
Time for Measurement: 5 min

ST-1230-2008
Evaluation: see details in operating
instructions for Impactor
0.1 mg/m3
Ambient Operating Conditions
Temperature: 10 to 30 °C
Humidity: max. 60 % r. h.

ST-1231-2008
Pressure: only to be used for unstressed
compressed air
0.5 mg/m3
Reaction Principle
Compressed air is guided through the Impactor vertically onto a
baffle plate made of cut glass. A 90 ° re-direction of the air flow in

ST-1232-2008
the Impactor separates the oil aerosols. The aerosols flow directly
onto the glass plate caused by the high inertia of the aerosols. The
recesses in the glass are filled with the oil aerosols and the light
1.0 mg/m3
dispersed by the glass grinding is compensated.

Cross Sensitivity
The measurement result is not dependent on the oil grade.
ST-604-2008

However, it must be noted that oil aerosols evaporate at higher


temperatures. Oil vapor is not displayed.
Adapter of the Impactor
Additional Information
The Impactor has to be used together with the Adapter of the
Impactor (Order No. 81 03 557) in conjunction with the Dräger
ST-602-2008

Aerotest Simultan.

Adapter with Impactor


connected in Dräger
Aerotest Simultan