Maintenance Manual

09/2002 63595-4A
Copyright © 2002 by Instrumentarium Imaging

Documentation, trademark and the software are copyrighted with all rights
reserved. Under the copyright laws the documentation may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or
machine readable form in whole or part, without the prior written permission
of Instrumentarium Imaging.

Orthopantomograph® and Orthoceph® are registered trademarks of Instru-


mentarium Corporation. U.S. patents 4,641,336; 5,016,264; 5,425,065 and
5,444,754. German patent 4,344,745. Orthopantomograph® OP100 D and
Orthoceph® OC100 D comply with UL and C-UL (File E218408).

The original language of this manual is English.

Instrumentarium Imaging reserves the right to revise this publication from


time to time and to make changes in the content of it without obligation to
notify any person of such revision or changes.

Manufactured by Instrumentarium Imaging


P.O. Box 20
FIN-04301 Tuusula
FINLAND
Tel. +358 10 394 6500
Fax. +358 10 394 6501
E-mail: dental@fi.instrumentarium.com
Internet: http://www.InstrumentariumImaging.com
Table of Contents
1 Preventive maintenance ..................................................................................... 1
2 Maintenance program........................................................................................ 3
2.1 Product life time ...................................................................................................... 4
3 Preventive maintenance reminder display ....................................................... 5
4 OP100D configuration sheet .............................................................................. 7
5 Checklist for Digital Dental X-ray installations............................................... 9
5.1 OP 100D .................................................................................................................. 9
5.2 OC 100D................................................................................................................ 10

63595-4A Instrumentarium Imaging i


ii Instrumentarium Imaging 63595-4A
1 Preventive maintenance

1 Preventive maintenance
This unit is designed to provide reliable performance and many years of
customer satisfaction. In order to assure safe performance of this X-ray
equipment, a preventive maintenance program must be established. It is the
owner’s responsibility to supply or arrange for this service. Consult your
Orthopantomograph dealer to arrange for this service. Maintenance service
for Orthopantomograph OP100D is recommended every year or 2000
exposures, whichever comes first.

NOTE
For USA: to keep the equipment in compliance with the DHHS Performance
Standard the following maintenance schedule shall be observed: up to 40
exposures per week: perform maintenance every 12 months. At 40 - 100
exposures per week, perform maintenance every 6 months.

63595-4A Instrumentarium Imaging 1


1 Preventive maintenance

2 Instrumentarium Imaging 63595-4A


2 Maintenance program

2 Maintenance program
This periodic maintenance for Orthopantomograph OP100D should include:

1 Check the steel wires, which carry the weight of the vertical carriage and
counterweight. If there is any sign of wearing, the steel wires must be
replaced.
2 Check that the unit is well fixed to the wall and floor
3 Check that the carriage motor clutch adjustment is correct.
4 Check kVp / mA / exposure times according to each procedure.
5 Grounding: Check that GND in Filament Control Board is connected to
the power line ground. Check that the top of the tube head assembly is
grounded (yellow-green wire on the top of the tubehead assembly).
Resistance must be less than 0.2 ohm.
6 Clean the rotation friction surface from debris.
7 Run the “Sr 80 ro-”, “Sr 81 LI-” and “Sr82CE-”(if you have OC 100
D)movement test programs in the Service mode to verify that each
movement is OK.
8 Make a normal exposure @ 57kV and 2mA, with a needle test phantom
to verify that the image layer is in the correct position, and that there are
no vertical stripes in the image.

If there are stripes, use Service programs (Sr 83 ro=, and Sr 84 LI=) to
find the problem.
9 Check that there are no oil leaks in the Tube Head Assembly.
10 Verify the collimator microswitch operation in both direction: QA - PAN
- CEPH. Program selections in the control panel should change
accordingly. Check also that the switch in the cephalostat head changes
the LAT and PA mode.
11 OP/OC100 Beam collimation test: Check that the panoramic x-ray beam
is aligned and that it is in the middle of the secondary slot and display
test screen.
12 OC100 Cephalometric beam collimation test: Check that the x-ray beam
is aligned and that the x-ray beam stays within the fluorecent tool
markings. Check the patient collimator alignment.Verify that earposts
are aligned.
13 Check the drive belt in the Cephalostat
14 Check several examples of the customers radiographs to verify the
correct operation of the unit.
15 Check the operation of the control panel and patient positioning
panel(s), and check that the vertical carriage moves smoothly, and stop
at the limit switches.
16 Check the positioning light adjustments. In Cephalostat also the laser
positioning light.
17 Make the Quality Assurance test and record the AEC and Gain settings
from “Pr 68 INS, FrE AEC” and the density settings from “Pr 52 CCO”.
18 Record the cumulative exposure counter value from “Pr 61 CLC”.
19 Reset the “PSE” counter in the program Pr 59 PSE.

63595-4A Instrumentarium Imaging 3


2 Maintenance program

20 To keep a record of unit’s parameters, fill out the OP100 Configuration


Form. Keep it for future use.

The above maintenance procedures require the services of a qualified


technician. In addition to periodic maintenance any deviation from normal
performance should be immediately reported to your dealer.

WARNING
Only trained and qualified personnel should be permitted access to the
internal parts of the equipment.

2.1 Product life time


If the product cannot meet the Manufacturer’s specifications, despite of
maintenace and/or repair, its life time has ended.

4 Instrumentarium Imaging 63595-4A


3 Preventive maintenance reminder display

3 Preventive maintenance reminder display


The equipment has a special feature that displays a message “CH 8 PSE” on
the exposure time display after every 2000 exposures. This code is displayed
for few seconds when the power is switched on. This code indicates that the
owner should carry the preventive maintenance for equipment. However, it is
owners responsibility to decide whether service will be done. This code has
no effects to the operation of equipment.

63595-4A Instrumentarium Imaging 5


3 Preventive maintenance reminder display

6 Instrumentarium Imaging 63595-4A


4 OP100D configuration sheet

4 OP100D configuration sheet


The OP100D can be configured in many different ways depending on the site
requirements and customer needs. The following tables summarize the
important settings that are useful to record in the case of future service calls.

63595-4A Instrumentarium Imaging 7


4 OP100D configuration sheet

8 Instrumentarium Imaging 63595-4A


5 Checklist for Digital Dental X-ray installations

5 Checklist for Digital Dental X-ray installations

NOTE
Pay extra attention to the ones on gray backround as they have an impact on
the image quality!!

5.1 OP 100D

Value/Status Value/Status
# Item to check before the after the
adjustment adjustment

Unit levelled and securely


1 bolted
Vertical movements smooth-
2 ness & limits
Rotation and linear move-
3 ment
Panoramic collimator adjust-
4 ment
AEC frequency (PAN term.
5 software ver.___)
Image layer (ball-pin phan-
6 tom image)
7 Midsagittal light
8 Frankfort light
9 Layer light
10 QA image
Dose settings, Pr 52 Cco
11 (P1-P9)
Existence of User Manual
12 and all accessories
Application training given
13 properly

Name of technician: Unit serial #:


Number of exposures taken:

63595-4A Instrumentarium Imaging 9


5 Checklist for Digital Dental X-ray installations

5.2 OC 100D

Value/Status Value/Status
# Item to check before the after the
adjustment adjustment

1 Cephalostat head levelled


Cephalostat collimator
2 adjustment
Cephalostat drive belt ten-
3 sion
4 Beam at CCD camera
Secondary collimator adjust-
5 ment
6 Ear rod alignment
7 FH laser light adjustment
Nasio frequency, (far, end &
8 close posit.)

CliniView

Value/Status Value/Status
Item to check
# before the after the
adjustment adjustment

PC specifications fulfil the


requirements;
Processor, RAM, HDD, display
1 adapter

CliniView installation;
Version#, file locations, network
2 paths, gainfiles

PC settings;
3 screen resolution, power saving off

Cliniview settings;
4 histograms, gamma value

10 Instrumentarium Imaging 63595-4A


5 Checklist for Digital Dental X-ray installations

NOTES:

63595-4A Instrumentarium Imaging 11


5 Checklist for Digital Dental X-ray installations

12 Instrumentarium Imaging 63595-4A


09/2002 © Instrumentarium Imaging
Instrumentarium Imaging Instrumentarium Imaging France S.A.R.L. Instrumentarium Imaging Singapore
P.O.Box 20, FIN-04301 Tuusula, Finland 4, Avenue des Roses 152 Beach Road
Tel. +358 10 394 6500 · Fax +358 10 394 6501 94386 Bonneuil Sur Marne Cedex, France #12-03A Gateway East
dental@fi.instrumentarium.com Tel. +33 1 43 39 51 51 · Fax +33 1 43 39 75 75 189721 Singapore
instrumentarium.imaging@wanadoo.fr Tel. +65 6391 8600 · Fax +65 6396 3009
Instrumentarium Imaging Inc. singapore@sg.instrumentarium.com
300 West Edgerton Avenue, Milwaukee Instrumentarium Imaging Italia S.R.L.
Wisconsin 53207, USA Via Cassanese, 100 Instrumentarium Imaging Dental GmbH
63595-4A

Tel. +1 800 558 6120, +1 414 747 1030 20090 Segrate (MI), Italy P.O.Box 2044, 77680 Kehl am Rhein, Germany
Fax +1 414 481 8665 Tel. +39 02 21 30 28 1 · Fax +39 02 21 30 28 60 Tel. +49 7851 932 90 · Fax +49 7851 932 930
info@usa.instrumentarium.com instrumentarium@it.instrumentarium.com kontakt@instrumentarium-imaging.de

Instrumentarium Imaging is constantly improving its products and reserves the right to
change these specifications without notice. Distributor:

www.InstrumentariumImaging.com