Sei sulla pagina 1di 8

Article

pubs.acs.org/ac

Longitudinal On-Column Thermal Modulation for Comprehensive


Two-Dimensional Liquid Chromatography
Mari E. Creese,† Mathew J. Creese,§ Joe P. Foley,†,‡ Hernan J. Cortes,⊥ Emily F. Hilder,†,×
Robert A. Shellie,†,∥ and Michael C. Breadmore*,†

Australian Centre for Research on Separation Science, School of Physical Sciences, University of Tasmania, Private Bag 75, Hobart,
Tasmania 7001, Australia

Department of Chemistry, Drexel University, 3141 Chestnut Street, Philadelphia 19104, United States
§
Allison Laboratories Pty Ltd., Sandy Bay, Tasmania 7005, Australia

Trajan Scientific and Medical, 7 Argent Place, Ringwood, Victoria 3134, Australia

HJ Cortes Consulting LLC, Midland, Michigan 48642, United States
×
Future Industries Institute, University of South Australia, GPO Box 2471, Adelaide, South Australia 5001, Australia

ABSTRACT: Longitudinal on-column thermal modulation for comprehen-


sive two-dimensional liquid chromatography is introduced. Modulation
optimization involved a systematic investigation of heat transfer, analyte
retention, and migration velocity at a range of temperatures. Longitudinal on-
column thermal modulation was realized using a set of alkylphenones and
compared to a conventional valve-modulator employing sample loops. The
thermal modulator showed a reduced modulation-induced pressure impact
than valve modulation, resulting in reduced baseline perturbation by a factor
of 6; yielding a 6−14-fold improvement in signal-to-noise. A red wine sample
was analyzed to demonstrate the potential of the longitudinal on-column
thermal modulator for separation of a complex sample. Discrete peaks in the
second dimension using the thermal modulator were 30−55% narrower than
with the valve modulator. The results shown herein demonstrate the benefits
of an active focusing modulator, such as reduced detection limits and increased total peak capacity.

M odulation is the core process in any comprehensive two-


dimensional chromatographic separation. The modulator
consecutively and systematically collects analytes as they are
switching, which also affects the second dimension column and
reduces its expected lifetime.
Within valve based LC × LC modulators, solvent dilution has
eluted from the first dimension column and delivers them to the been demonstrated as an effective method for active focus-
second dimension column.1 There are many modulation ing.18,19 Dilution of the 1D effluent allows an improvement of the
possibilities for comprehensive two-dimensional gas chromatog- on-column focusing at either the head of the 2D column or the
raphy (GC × GC): resistive heating of painted zones of trap column. For example, the effect of introducing 10−20% less
modulator capillary,2 rotational movement of a slotted heater,3,4 organic solvent to the 2D allowed on-column focusing and
enhanced resolution.20−25 Dilution is achieved by a connection
longitudinal movement of a cold trap,5−7 cryogen jets with or
of a weak solvent to the switching valve modulator and is easily
without the combination of a hot jet,8 resistive heating with or
implemented. The limitation of any valve-based modulation lies
without a cryogen trap,9−11 and more.8 In contrast, compre- within the momentary blocking of flow as the valve is switched
hensive two-dimensional liquid chromatography (LC × LC) has from position a to b. This is particularly critical when operating at
not been demonstrated in online-mode without use of switching or close to the pressure limit as the pressure increase, however
valves. LC × LC using switching valves may be performed with or short in duration, can exceed the limitations of the system and
without trap columns.12−16 A trap column offers a focusing may result in termination of the analysis.
mechanism; however, re-equilibration of the trap column is not High temperatures are commonly employed in LC to achieve
always reached and band compression efficiency is reduced faster separations, improve selectivity and resolution, and reduce
throughout the analysis unless the mobile phase strength of the backpressure,4,26−30 although typically at isothermal conditions.
second dimension gradient greatly exceeds that of the first Since the adaptation of low thermal mass (LTM) resistive
dimension gradient or if an equilibration step is accounted for in
the modulation period.16 Further, switching valves have negative Received: August 21, 2016
effects on the 2D system17 such as pressure pulses (<10% of Accepted: December 5, 2016
operating pressure for 2DLC specific valve) induced by Published: December 5, 2016

© 2016 American Chemical Society 1123 DOI: 10.1021/acs.analchem.6b03279


Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

heating from GC5−7,10 to LC,31 there have been a number of cryogenic system (LMCS) was used to hold the LTM sleeve. A
studies on thermal gradient elution9−11,31−33 and temperature brace was added to mechanically stabilize the trap in order to
pulsing for manipulating discrete sections of the separation.34,35 keep the alignment and allow free movement of the sleeve. The
The LTM technology was also employed as a modulator for LTM was temperature controlled via an LTM module (Agilent
offline LC × LC by Verstraeten et al.36 where a capillary trap Technologies, Santa Clara, CA) and the LMCS was triggered by
segment was temperature cycled to facilitate modulation and remote start-out. Cold jets were employed to bring temperature
fraction collection from the first dimension. down to trapping conditions between mobilization steps in the
Temperature can also be used for peak focusing by inducing a modulation process. Compressed air (6 bar) was passed through
compression effect, exploiting the change in retention with an ice bath, split into two jets, and directed at two positions on
temperature and changing from low temperature for capture to the modulator trap column. The temperature at the points of
higher temperatures for release. Subambient temperatures were cooling were recorded to ∼13 °C. The jets were purging cooled
used for large volume injections where a cold zone is created at air continuously throughout the analysis. A thermal camera
the head of the column and the remaining part of the column is (FLIR E40), set to a temperature range of 0−650 °C ± 0.07 °C,
kept either at ambient or elevated temperatures.37,38 Remobiliza- was employed to record thermal changes across the trap column
tion of the large volume injection zone is realized by moving the when modulating.
column into the hot zone or by assisted heating. Similarly, Instrumentation and Methods. The analyses were
subambient temperature was used for peak compression prior to performed using an Agilent 1290 Infinity 2D-LC solution system
detection39 where an alternating cold and hot zone is created by (Wilmington, DE). The valve modulator was a four-port duo
flooding a reservoir surrounding the column. Gradual and valve (2DLC valve, Agilent Technologies) equipped with two
segmented temperature control for resolution manipulation has 60-μL loops in a countercurrent flow configuration. The diode
also been achieved by a multiple Peltier element design where array detector was set to 254, 260, 275, and 280 nm at a collection
each unit was individually controlled.40 rate of 20 Hz and 4 nm bandwidth. The injection volume was 1.5
The literature above suggests it is theoretically feasible to μL of 50 μg/mL standards and 5 μL for red wine analysis. The
create a thermal modulator for LC × LC. The physicochemical first dimension (1D) separation took place on a C18 column
differences between GC and LC, such as the incompressibility of (Zorbax Eclipse Plus RRHD, 2.1 mm × 150 mm, 1.8 μm; Agilent
liquids compared to gases and slow heat transfer,41,42 are the Technologies, Mulgrave, Victoria, Australia) at a flow rate of
main challenges for nonvalve based modulators in LC × LC. The 0.025 mL/min with a gradient from 55 to 85% B from 15.5 to 150
approach undertaken in this work is inspired by the min (for alkylphenones) and 5−80% B from 15.5 to 215.5 min
longitudinally modulated cryogenic system (LMCS) developed (for separation of red wine). The second dimension (2D) column
by Marriott and Kinghorn for GC × GC modulation5,6,43,44 and was titanium tubing (0.3 mm × 34 mm) packed with 5 μm PGC
the temperature cycled modulator for offline LC × LC by particles (Thermo Scientific) similar to the modulator trap
Vertstraeten et al.36 Marriott and Kinghorn used a double walled column. The 2D pump delivered 0.125 mL/min to a T-piece
hollow cylinder fitted over the modulation segment of the adding to the flow from the first dimension, giving a total of 0.15
capillary column and purged the hollow with cryogenic liquid for mL/min, which was then connected to the modulator trap
trapping. When the trap was moved away from the column, the column. The second dimension mobile phase program was 73−
temperature of the GC column oven quickly heated the focused 97% B (giving 70−95% after mixing with the 1D mobile phase)
band inside the cold zone of the capillary. During this process, a for separation of alkylphenones. The 2D method used for the
zone further upstream (toward the first dimension) was cooled valve modulation setup was set to 0.15 mL/min with a gradient of
before the column was moved back to the original position. 70−95% methanol from 15.5 to 150 min. The method for
In this work, we explore the possibility to employ longitudinal separating the compounds in the red wine sample was 5 μL
on-column thermal modulation (LOCTM) using LTM resistive injected on the 1D column, which was operated at 0.025 mL/min
heating for LC × LC, using its performance toward a set of at a gradient from 5 to 85% methanol from 15.5 to 215.5 min.
alkylphenones as a proof of concept. The 2D pump delivered 0.125 mL/min giving effectively 0.15

■ EXPERIMENTAL SECTION
Chemicals. Alkylphenones (Acetophenone (BDH); propio-
mL/min through the modulator and the second dimension
column. The gradient was 15−73% methanol (effective 15−
70%). The second dimension when employing the valve
phenone, butyrophenone (Aldrich)); and valerophenone (Alfa modulator was operated at 0.15 mL/min at a gradient from 15
Aesar) were prepared to 500 μg/mL stock solutions in 100% to 70% methanol from 15.5 to 215.5 min. A modulation period of
methanol. The standards were diluted to 50 μg/mL with 100% 60 s was employed for all methods.
water for analysis. The mobile phase was water (A) from an in-
house Millipore system (Millipore, North Ryde, NSW, Australia)
and methanol (B) from VWR (Tingalpa, QLD, Australia) both
■ RESULTS AND DISCUSSION
A homologous series of alkylphenones was used for development
with 0.1% formic acid from Sigma-Aldrich (Castle Hill, NSW, and evaluation of the new modulator due to the well-defined
Australia). Red wine was purchased from a local store. characteristics of these analytes allowing predictable retention in
Modulator. The modulator consisted of three parts: (i) a trap both dimensions. Figure 1 shows a schematic of the LOCTM
column of 0.8 mm i.d. titanium tubing cut to 81 mm, slurry system. Modulation is achieved by moving a resistively heated
packed with 5 μm porous graphitic carbon (PGC) particles sleeve (200 °C) from the inlet half to the outlet half of the PGC
(Thermo Scientific, U.K.) in methanol/water using a packing trap column by aid of an LMCS controller system. Compressed
pump (Haskell, Burbank, CA). The column outlet was fitted with air is blown through two ports, one directly onto the trap column
a stainless steel frit (diameter 0.038 in. × 0.030 in. × 0.062 in., 0.5 to enhance cooling and the second onto a shield around the
μm pore size; Kinesis, Redland Bay, QLD, Australia). (ii) LTM heated sleeve. The air was cooled by first flowing through tubing
resistively heated sleeve modified to 2.1 mm × 20 mm with total placed in an ice bath before the flow was split and directed to the
resistance of 97 Ω, and (iii) a modified longitudinally modulated trap column. The modulation column (PGC) was always
1124 DOI: 10.1021/acs.analchem.6b03279
Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

sleeve moved away from the initial position, at 1, 2, 3, 4, and 5 s


and then at 5 s increments until 120 s. A thermal camera was
chosen to record this data as moving the heating sleeve could
have damaged a thermocouple if placed in the small gap (∼0.3
mm) between the heating sleeve and the trap column. As seen in
Figure 2, immediately after the trap is moved, the temperature of
the trap decreases until it reaches thermal equilibrium, which
occurs within 10 s with cooling and about 30 s without cooling.
Also, the equilibrium temperature is about 20−40 °C cooler
when active cooling is employed. The figure also shows that the
equilibrium temperature reached with active cooling for the
outlet section is slightly higher than the inlet section. This is due
to the warmer incoming mobile phase, which has passed through
the heated inlet half of the trap, as opposed to the 25 °C mobile
phase that enters the inlet of the trap column. This higher outlet
Figure 1. Setup of the LOCTM showing the resistive heating sleeve and temperature allows more efficient release of the captured targets
the longitudinal modulator controller: (a) trapping at inlet, remobilizing into the second dimension column. As a result, with active
at outlet, (b) trapping at outlet, remobilizing at inlet. (1) T-piece from cooling, the minimum modulation period is approximately 40 s
1
D column and 2D pump, (2) fitting (stainless steel at inlet, PEEK at based on a two-stage modulation movement (inlet−outlet). On
outlet), (3) PGC trap column (0.8 mm × 81 mm, 5 μm), (4)
the basis of Figure 2, it is beneficial to implement a cooling
longitudinal modulator, (5) resistively heated low thermal mass (LTM)
sleeve, (6) column brace, (7) to 2D column, (8) cooled compressed air. process to rapidly lower the temperature back to improve
trapping of the analytes. In this proof-of-concept study, cooled
compressed air was employed. Further investigation on efficient
shielded from the cooling jet that was directed to the alternating cooling mechanisms would be of benefit to minimize the
halves of the column devoted to mobilization, so as to avoid modulation period and improve the focusing effect.
lowering the temperature and increasing the retention in the It is important that analytes are trapped and eluted from the
mobilization region. The use of a second pump to mix the trap within a single modulation cycle in order for the LOCTM to
effluent before the LMCS allowed independent control of the function properly. As the trap column capacity and retention may
flow rate in both dimensions as well as some manipulation of the be significant, modulation periods must be optimized not only
chemistry for trapping and separation on the 2D column. based on heat transfer rates, but also based on the Murphy-
Understanding the thermal behavior of the heated sleeve and Schure-Foley rule,45 and according to the analyte velocity at the
trap column is critical for constructing a suitable modulator for modulation temperature. The retention and migration velocity of
comprehensive LC. For the modulator to work most effectively, alkylphenones as a function of temperature (up to 150 °C) are
there needs to be a large and rapid thermal change between the given in Table 1. Retention factor values were obtained by
heating and cooling stages of the modulator. To enhance cooling, heating the whole trap column using a 50 mm long resistively
jets are positioned to blow cold air over the trap column when it heated sleeve. The retention data and column void were obtained
is not being heated. To examine the influence of cooling, the by elution at isotherms ranging from 25 to 150 °C and were
heating sleeve was kept at the inlet (Figure 2, left) and outlet corrected for the effect of extra column times (measured at the
(Figure 2, right) position for 600 s before it was moved to the same temperature) by subtraction of the latter from the former.
alternate position. Temperatures were measured as the heating Triplicate analyses were performed. As seen in Table 1, the

Figure 2. External cooling rates at (a) inlet and (b) outlet of the modulator trap column. Impact of active cooling is seen by the black, red, and blue trace
and passive cooling by the yellow, green, and pink trace. Positions 1−6 are illustrated on the modulator trap inset to the right. The infrared camera (FLIR
E40) was recording from −10 to 120 s after the heating sleeve, kept at 200 °C, was moved away from the position of measurement. The heating sleeve
had been held at the measuring position for 600 s before cooling rates were recorded. For active cooling the cold jets, ∼13 °C, were directed at positions
1 (for cooling at inlet) and 4 (for cooling at outlet).

1125 DOI: 10.1021/acs.analchem.6b03279


Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

Table 1. Temperature Dependence of Alkylphenone


Retention Factor and Analyte Velocity on PGC Trap
Columna
alkyl- temperature retention factor analyte velocity
phenone (°C) (k) (mm/s)
aceto- 25 0.51 ± 0.04 3.7
150 0.01 ± 0.01 8.8
propio- 25 1.95 ± 0.06 1.9
150 0.05 ± 0.02 7.5
butyro- 25 3.96 ± 0.07 1.1
150 0.07 ± 0.03 6.5
valero- 25 10.26 ± 0.11 0.5
150 0.17 ± 0.03 4.7
a
Values showing 95% confidence interval (n = 3). Flow rate 0.15 mL/
min.

Figure 3. Modulation optimization showing peak elution at 60 s


retention factor is reduced to approximately 0 for the all of the modulation period for acetophenone (green), propiophenone (orange),
test alkylphenones and analyte velocities increased by a factor of butyrophenone (blue), and valerophenone (pink). The vertical dashed
2−9. The large change in retention factors from 25 to 150 °C lines show when the heating sleeve (200 °C) was moved along the trap
should be sufficient for modulation as both trapping and thermal column and the associated pressure fluctuation (black) due to
remobilization are essential for the performance of the temperature change. The analytes were individually injected onto the
modulator, although slight retention remains at 150 °C. The modulator with the heating sleeve at the outlet position from 0 to 30 s, at
consequence of excessive retention at the remobilization the inlet position from 30 to 60 s, and back to the outlet position from 60
condition will cause the fraction to elute in a later cycle (wrap- to 90 s (when the peak is required to be eluted). This was repeated until
5 min (5 modulation cycles). Note the void time of 12 s between trap
around) and potentially interfere with the early eluting
and detector.
components from the next fractions. Ultimately, this can lead
to a reduction of the 2D column’s peak capacity and loss of overall back to the temperature dependent retention study to estimate
resolution. the effective modulation temperature (heating sleeve pro-
Initial optimization of the modulation period was performed grammed to 200 °C). Fitting the converted retention factors
by injecting individual analytes into the system without the 1D (0.03, 0.15, 0.35, and 0.59) to the trend lines of a thermal
and 2D columns and with well-defined movement of the heated retention study of aceto-, propio-, butyro-, and valerophenone
sleeve. This allows direct observation of the dual stage trapping gives an estimated internal trap column temperature of
and remobilization capability of the modulator. The analyte was approximately 105 °C.
initially trapped on the first half of the modulator trap with the Comparison of modulation at different conditions was
heated sleeve at the outlet position. Then the analyte was performed, and a score table (Table 2) for each of the
mobilized onto the second half of the trap by moving the heated
sleeve to the inlet position. Finally, the heated sleeve was moved
Table 2. Score Table of Modulation Period Effect on
back to the outlet position and the peak was eluted. The trap
Alkylphenone Elutiona
column was kept at isocratic conditions with 70% methanol at
0.15 mL/min, and the modulation period was explored in the system constants effect on alkylphenone
range of 45−150 s. The 60 s modulation period is shown in modulation period (s) aceto propio butyro valero
Figure 3 where the analytes are eluted within one modulation
150 − −−− + +
period except valerophenone. It can be seen from the figure that
120 − + + +
the analytes are injected and trapped during the first 30 s when
90 −−− + + +
the heating sleeve is in the outlet position, transferred to the
60 + + + −
second half of the trap from 30 to 60 s when the heating sleeve is
45 + + − −−−
in the inlet position, and eluted from 60−90 s when the heating a
sleeve was back at the outlet position. Moving the heating sleeve Peak not eluted within one modulation cycle (−), peak splitting (−
−), and peak not eluting within one modulation cycle and splitting (−
up and down the trap column causes pressure fluctuation, seen as
− −).
the black trace, due to thermal changes in the system. At face
value, it would be anticipated that the pressure should not
change; however, the system is asymmetric with the temperature alkylphenones at the different modulation times was constructed
of the liquid entering the cold half of the trap different depending with negatives of (i) peak eluted outside the modulation cycle
on whether the heating sleeve is at the inlet or outlet. The pump (−), (ii) peak splitting (− −), and (iii) the combination of (i)
pressure transducer recognizes this as a change in thermal and (ii) (− − −). The successful elution within one modulation
expansion due to local hot and cold spots. There is a void cycle was annotated with a positive (+).
between the column and detector due to a resistor coil needed for As can be seen from the longest modulation time of 150 s, this
the prevention of solvent evaporation, as such the elution times is too long for acetophenone and propiophenone, which were
in Figure 3 are 15 s later than the actual elution from the trap both eluted before the modulation cycle was complete, which is
column. The actual elution times are acetophenone = 60 s, expected due to their lower retention than butyro- and
propiophenone = 69 s, butyrophenone = 81 s, and valerophenone. The shortest modulation time of 45 s is also
valerophenone = 96 s. These elution times were translated inadequate, being too short for butyro- and valerophenone,
1126 DOI: 10.1021/acs.analchem.6b03279
Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

connected, high elutropic strength causes breakthrough and


incomplete sampling of the first dimension peaks whereas a
mobile phase that is too weak will induce wrap-around and
excessive retention. Wrap-around is an effect of excessive elution
times in the second dimension resulting in broad peaks or in the
worst-case scenario “streaking”.46 This is of particular concern
with the trap column setup used herein.
The physical dimensions of the trap column were chosen on
the basis of direct connection with standard size HPLC fittings
and the smallest practical size of the heating sleeve (20 mm). The
diameter of the trap column was justified based on the thin
column wall for better heat transfer and for the retention
capacity. The 1D column had to offer broad peaks to allow
sampling; a long column operated at a low flow rate could offer
this. The 2D column had to be operated fast, have low
Figure 4. Raw chromatogram of butyrophenone showing the number of backpressure due to the accumulation of pressure through the
cuts of the 1D peak with 60 s modulation period.
serially coupled setup, and express moderate retention at the
operating temperature (75 °C) to avoid extensive wrap-around,
which are eluted in subsequent modulations. This provides thus a short capillary titanium column was packed with PGC.
opportunity for components from different cuts in the first Online comprehensive two-dimensional separations were
dimension to be coeluted in the second, against the rules defined achieved by connecting the 2D column (PGC, 0.3 mm × 34
by Giddings.1 With the modulation periods from 60 to 120 s, mm, 5 μm) to the modulator trap column via a stainless-steel
there is better coverage, with shorter periods not ideal for transfer capillary (0.12 mm × 20 mm) and operated at the same
valerophenone and longer periods not ideal for acetophenone elution conditions as the modulator, gradient 70−95% methanol
Having demonstrated the ability to trap and elute the at 0.15 mL/min. The 2D contour plots of alkylphenones by
alkylphenones using the LOCTM, it was tested for the capability longitudinal on-column thermal modulation and an equivalent
to modulate online with the first dimension at gradient separation by valve modulation are shown in Figure 5a,b,
conditions. The1D column, second dimension pump and the respectively. Although 40 s modulation is possible with LOCTM
modulator trap column were connected to a T-piece. The mobile (Figure 2), a 60 s modulation was found to be more suitable for
phase flow delivered from the 2D pump was matched to the 1D the separation of the alkylphenones on the columns used to
mobile phase to deliver 0.15 mL/min to the modulator trap. The demonstrate the first thermal modulator for LC × LC.
percentage methanol from the 2D pump was adjusted to give an The four alkylphenones fall on a straight line indicating
effective gradient from 70% to 95% methanol assuming good correlated retention in the 1D and 2D columns and demonstrates
mixing in the T-piece. The repeatability of the modulator’s that the modulator is working correctly, as each homologue in
performance is good where 95% confidence intervals (n = 3) the same series is expected to behave similarly within the two
were found to be less than ±1 s for gradient 2D elution. The raw reversed phase separation mechanisms, C 18 and PGC,
chromatogram in Figure 4 shows that this modulator clearly has respectively. The 2D peak width is one metric with which to
the capability to perform according to the 3−4 cuts per 1D peak quantitatively compare the modulator performance. For the
rule.45 alkylphenones, the longitudinal thermal modulator shows wider
2
A suitable gradient in the 2D is important not only for the D peaks than the valve-based modulator. This is caused by the
separation on the 2D column but also for the modulator relatively low retention on the trap and 2D column, which limits
performance. Considering the columns in this setup are serially compression on the modulator column. For acetophenone, this

Figure 5. Separation of alkylphenones using (a) longitudinal on-column thermal modulator and (b) valve modulator.

1127 DOI: 10.1021/acs.analchem.6b03279


Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

Figure 6. Separation of red wine on (a) thermal modulator and (b) valve modulator. Numbered peaks are considered for peak width comparison in the
second dimension (Table 3).

Figure 7. Contour plot of discrete peak (1) from red wine analysis by (a) thermal modulation and (b) valve modulation.

results in breakthrough, as the modulation period is too long. LOCTM to valve modulation induces a time shift in the separation
The time lag in the first dimension experienced with the LOCTM space.
arises from the difference in the operating pressure of the two One advantage of the thermal modulation are the minimized
columns. In the valve interface, the pressure is essentially that pressure fluctuations. The valve modulator used here produced a
from the analytical column alone, while in the LOCTM, the 10% increase in pressure upon switching, compared to a 2%
pressure is higher due to the contribution of the trap and the 2D decrease in pressure due to the thermal modulator. The resulting
column. This change in pressure has an effect on the first baseline noise, as an effect of change in refractive index of the
dimension interaction due to the change in solute partial molar mobile phase observed by the detector,51 translates to improved
volume.47−50 The cumulative backpressure with thermal signal-to-noise by a factor of up to 14 for the thermal modulator.
modulation experienced by the 1D column is 7 times larger The longitudinal movement of the heated sleeve and the
compared to the backpressure when using the valve setup. A online implementation are the fundamental differences between
difference of 26% for acetophenone, 22% for propiophenone, LOCTM and the approach by Verstaeten et al.36The offline
18% for butyrophenone, and 15% for valerophenone is not approach relied on the rapid temperature transfer to and from the
unreasonable given that McGuffin and Evans reported an trap column, which was packed with PGC as in this study, by
increase of 9 to 24% for a homologous series of fatty acid temperature cycling the heating sleeve. LOCTM is a continuous
derivatives under isocratic conditions.47 The pressure-dependent system where temperature is kept isothermal, circumventing the
retention observed here is presumably due to the decreasing time lag for heat transfer by temperature programming,
backpressure as the viscosity of the mobile phase decrease during particularly in this case where a 0.8 mm i.d. trap column is
gradient elution, i.e., the smaller the homologue, the earlier it employed and as cryogenic cooling was not available. LOCTM
elutes during the gradient when the backpressure is greater and may also be gentler on the trap column stationary phase as it is
thus the usual pressure-dependent effect on homologue operated at a lower temperature than the maximum according to
retention is inverted. Overall, the larger total pressure by stationary phase specifications. The improvement in signal-to-
1128 DOI: 10.1021/acs.analchem.6b03279
Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

noise ratio corresponds to the improvement of signal response overall performance of the modulation technique. This shows
reported by Verstraeten et al.36 that much work remains in further developments and
In a complex sample, the range of molecules with different optimization of LOCTM before it can be implemented as a
retention on both the first and trap/second dimension stationary routine approach to LC × LC to meet the fast modulation
phase will give a variety of relative retentions (kPGC/kC18), as such periods required for narrow first dimension peaks. While these
breakthrough and wrap-around will be observed. There is also results show the first experimental implementation of a thermal
potential for focusing through the thermal modulator, which is modulator for LC × LC, there are several aspects that are not
not possible with the simple valve approach. Red wine was ideal. The first is the discriminating nature of the trap column,
analyzed as an example of a complex sample using both where the analyte interaction on the trap phase may lead to
modulation techniques with the chromatograms shown in Figure breakthrough or wrap-around and in the worst-case scenario
6. The separation is not intended as a quantitative study but to induce significant retention due to polar retention interaction on
illustrate the performance of the modulators and to provide some graphite.52−55 An analyte specific trap column phase may be
evaluation of the potential of the LOCTM as a modulator for required to obtain a desirable peak focusing effect, keeping in
comprehensive two-dimensional LC. mind the operating temperature using LOCTM. In the current
Both modulation techniques show a comparable peak pattern system, the 2D time is limited by both the speed of the 2D
in the 2D space and both show streaks. The effect of baseline separation and the modulator. New materials, particularly those
noise in Figure 6b is seen as a blue smear along the bottom length that are thermally responsive, may be required to obtain a
of the plot. The streaks are attributed to the 1D effluent fraction desirable peak focusing effect, keeping in mind the operating
migrating through the modulator and second dimension column temperature using LOCTM. The second limitation is the speed of
without being modulated. Wrap-around is also seen for both the modulation cycle. A third shortcoming is the accumulation of
modulation techniques for the wider second dimension bands; in pressure, which restricts column selection and flow rates. The
cases of excessive wrap-around, this results in streaking. The entire system can be improved through addressing improve-
vertical lines in the contour plot using the valve modulator ments in the 1D and 2D column and mobile phase chemistry, as
(Figure 6 b) are due to wrap-around and/or sample well as a more rapid thermo-cycling interface. Despite the
concentration overload. To provide some quantitative compar- drawbacks, this work has demonstrated equal repeatability to
ison of performance, an initially well-resolved peak was selected valve modulation and reduced baseline distortion. The
for analysis, Figure 7. The peak width using LOCTM gave 50% longitudinal on-column thermal modulator shows promise for
narrower bands in the second dimension compared to the valve improvement of comprehensive two-dimensional liquid chro-
matography as a valveless, online, dual stage focusing modulator.


modulator. On the basis of this result, three more peaks were
carefully selected based on the difference in retention, the 2D
CONCLUSION
peak widths of the longitudinal thermal modulator and the valve
modulator shown in Table 3. An important condition for successful compression by
modulation is the retention on the modulator trap relative to
Table 3. Peak Widths, With 95% Confidence Interval (n = 3), the first dimension stationary phase. Also, the rapid and large
in the Second Dimension for Longitudinal Thermal and Valve changes of analyte retention and migration velocity on the
Modulation of Red Wine, Peak Identification As in Figure 6 modulator trap at trapping (low) and remobilizing (high)
temperatures are critical parameters. The longitudinal on-
peak LOCTM tw (s) valve tw (s) column thermal modulator has paved a new path for
1 11 ± 2 23 ± 2 comprehensive two-dimensional liquid chromatography. The
2 19 ± 3 41a ± 3 recognition of a dual stage active modulator is of utmost
3 18 ± 1 25 ± 2 importance for the performance of LC × LC. Peak capacity is
4 13 ± 1 26a ± 7 critical and reducing 2D peak widths is one of the possibilities to
a
Baseline noise/peak overlap interferes with peak width determi- achieve this. The LC × LC separation of red wine using the
nation. An estimated peak boundary is used to give an approximate longitudinal on-column thermal modulator achieved up to 60%
peak width value. narrower bands of discrete peaks in the second dimension
compared to the valve-based modulator at otherwise identical
The data show that the peaks widths are reduced by 53% for conditions. Furthermore, the signal-to-noise ratios were up to an
peak 1, 55% for peak 2, 28% for peak 3, and 48% for peak 4 by the order of magnitude larger by the longitudinal thermal modulator.
LOCTM compared to the valve. This is a good indication of the
benefit of active focusing modulation. Peaks 2 and 4 are partly
incorporated in the baseline noise when employing valve
■ AUTHOR INFORMATION
Corresponding Author
modulation, making it difficult to determine the exact second *Phone: + 61 3 6226 2154. Fax: + 61 3 6226 2858. E-mail:
dimension peak width. Michael.Breadmore@utas.edu.au.
Like solvent-dilution active modulation,18,19 improved second ORCID
dimension peak widths are obtained with LOCTM. However,
solvent-dilution modulation outperforms LOCTM in terms of
Michael C. Breadmore: 0000-0001-5591-4326
reduced modulation periods and shorter total analysis times. The Notes
The authors declare no competing financial interest.


20 s modulation period and half the analysis time were obtained
with active solvent-dilution modulation compared to passive
valve-modulation,19 allowing the first dimension to operate at ACKNOWLEDGMENTS
closer to optimum conditions provided the 2D separation is fast M.E.C. would like to acknowledge the University of Tasmania for
enough. Cooling rates and the physical dimensions of the trap receiving a graduate research scholarship. J.P.F. gratefully
column currently restricts the speed of the LOCTM and thus the acknowledges the 2015−2016 sabbatical leave provided by
1129 DOI: 10.1021/acs.analchem.6b03279
Anal. Chem. 2017, 89, 1123−1130
Analytical Chemistry Article

Drexel University and the Visiting Fellowship provided by the (31) Gu, B.; Cortes, H.; Luong, J.; Pursch, M.; Eckerle, P.; Mustacich,
University of Tasmania and the Australian Centre for Research R. Anal. Chem. 2009, 81 (4), 1488−1495.
on Separation Science (ACROSS). M.C.B. is the recipient of an (32) Verstraeten, M.; Pursch, M.; Eckerle, P.; Luong, J.; Desmet, G. J.
ARC Future Fellowship (Grant FT130100101). The authors Chromatogr. A 2011, 1218 (16), 2252−2263.
(33) De Pauw, R.; Pursch, M.; Desmet, G. J. Chromatogr. A 2015, 1420,
would also like to thank Dr. Jim Luong (Dow Canada, Analytical 129−134.
Technology Center, Fort Saskatchewan, AB, Canada) for (34) Causon, T. J.; Cortes, H. J.; Shellie, R. A.; Hilder, E. F. Anal. Chem.
valuable discussions and technical guidance and Thermo 2012, 84 (7), 3362−3368.
Scientific for the porous graphitic carbon. (35) Pursch, M.; Eckerle, P.; Gu, B.; Luong, J.; Cortes, H. J. J. Sep. Sci.

■ REFERENCES
(1) Giddings, J. C. Anal. Chem. 1984, 56 (12), 1258A−1270A.
2013, 36 (7), 1217−1222.
(36) Verstraeten, M.; Pursch, M.; Eckerle, P.; Luong, J.; Desmet, G.
Anal. Chem. 2011, 83 (18), 7053−7060.
(37) Holm, A.; Molander, P.; Lundanes, E.; Greibrokk, T. J. Sep. Sci.
(2) Liu, Z.; Phillips, J. B. J. Chromatogr. Sci. 1991, 29 (6), 227−231. 2003, 26 (12−13), 1147−1153.
(3) Phillips, J. B.; Beens, J. J. Chromatogr. A 1999, 856 (1), 331−347. (38) Wilson, R. E.; Groskreutz, S. R.; Weber, S. G. Anal. Chem. 2016,
(4) Marriott, P. J.; Kinghorn, R. M.; Ong, R.; Morrison, P.; Haglund, P.; 88, 5112−5120.
Harju, M. J. High Resolut. Chromatogr. 2000, 23 (3), 253−258. (39) Eghbali, H.; Sandra, K.; Tienpont, B.; Eeltink, S.; Sandra, P.;
(5) Marriott, P. J.; Kinghorn, R. M. Anal. Chem. 1997, 69 (13), 2582− Desmet, G. Anal. Chem. 2012, 84 (4), 2031−2037.
2588. (40) Collins, D.; Nesterenko, E.; Connolly, D.; Vasquez, M.; Macka,
(6) Kinghorn, R. M.; Marriott, P. J. J. High Resolut. Chromatogr. 1998, M.; Brabazon, D.; Paull, B. Anal. Chem. 2011, 83 (11), 4307−4313.
21 (11), 620−622. (41) Goodwin, R. D. J. Phys. Chem. Ref. Data 1987, 16 (4), 799−892.
(7) Marriott, P.; Kinghorn, R. TrAC, Trends Anal. Chem. 1999, 18 (2), (42) Billen, J.; Broeckhoven, K.; Liekens, A.; Choikhet, K.; Rozing, G.;
114−125. Desmet, G. J. Chromatogr. A 2008, 1210 (1), 30−44.
(8) Edwards, M.; Mostafa, A.; Górecki, T. Anal. Bioanal. Chem. 2011, (43) Marriott, P. J.; Kinghorn, R. M. Anal. Sci. 1998, 14 (4), 651−659.
401 (8), 2335−2349. (44) Kinghorn, R. M.; Marriott, P. J. J. High Resolut. Chromatogr. 1998,
(9) Libardoni, M.; Fix, C.; Waite, J. H.; Sacks, R. Anal. Methods 2010, 2 21 (1), 32−38.
(7), 936. (45) Murphy, R. E.; Schure, M. R.; Foley, J. P. Anal. Chem. 1998, 70,
(10) Luong, J.; Gras, R.; Mustacich, R.; Cortes, H. J. Chromatogr. Sci. 1585−1594.
2006, 44 (5), 253−261. (46) Guiochon, G.; Marchetti, N.; Mriziq, K.; Shalliker, R. A. J.
(11) Jacobs, M. R.; Hilder, E. F.; Shellie, R. A. Anal. Chim. Acta 2013, Chromatogr. A 2008, 1189 (1−2), 109−168.
803, 2−14. (47) McGuffin, V. L.; Evans, C. E. J. Microcolumn Sep. 1991, 3, 513−
(12) Shellie, R. A.; Haddad, P. R. Anal. Bioanal. Chem. 2006, 386 (3), 520.
405−415. (48) Guiochon, G.; Sepaniak, M. J. J. Chromatogr. 1992, 606 (2), 248−
(13) Kivilompolo, M.; Pól, J.; Hyötyläinen, T. LC GC Eur. 2011, 24 250.
(5), 232−237. (49) Martin, M.; Guiochon, G. J. Chromatogr. A 2005, 1090 (1−2),
(14) Malerod, H.; Lundanes, E.; Greibrokk, T. Anal. Methods 2010, 2 16−38.
(2), 110−122. (50) Åsberg, D.; Samuelsson, J.; Leśko, M.; Cavazzini, A.; Kaczmarski,
(15) Shi, X.; Wang, S.; Yang, Q.; Lu, X.; Xu, G. Anal. Methods 2014, 6, K.; Fornstedt, T. J. Chromatogr. A 2015, 1401, 52−59.
7112−7123. (51) Evans, C. E.; Shabushing, J. G.; McGuffin, V. L. J. Chromatogr.
(16) Egeness, M. J.; Breadmore, M. C.; Hilder, E. F.; Shellie, R. A. LC 1988, 459, 119−138.
GC Eur. 2016, 5 (29), 268−276. (52) Knox, J. H.; Kaur, B.; Millward, G. R. J. Chromatogr. A 1986, 352,
(17) Talus, E. S.; Witt, K. E.; Stoll, D. R. J. Chromatogr. A 2015, 1378, 3−25.
50−57. (53) Vial, J.; Hennion, M.-C.; Fernandez-Alba, A.; Agüera, A. J.
(18) van de Ven, H. C.; Gargano, A. F. G.; van der Wal, S.; Chromatogr. A 2001, 937 (1), 21−29.
Schoenmakers, P. J. J. Chromatogr. A 2016, 1427, 90−95. (54) Pereira, L. J. Liq. Chromatogr. Relat. Technol. 2008, 31 (11−12),
(19) Gargano, A. F. G.; Duffin, M.; Navarro, P.; Schoenmakers, P. J. 1687−1731.
Anal. Chem. 2016, 88 (3), 1785−1793. (55) West, C.; Elfakir, C.; Lafosse, M. J. Chromatogr. A 2010, 1217
(20) Groskreutz, S. R.; Swenson, M. M.; Secor, L. B.; Stoll, D. R. J. (19), 3201−3216.
Chromatogr. A 2012, 1228, 31−40.
(21) Groskreutz, S. R.; Swenson, M. M.; Secor, L. B.; Stoll, D. R. J.
Chromatogr. A 2012, 1228, 41−50.
(22) Larson, E. D.; Groskreutz, S. R.; Harmes, D. C.; Gibbs-Hall, I. C.;
Trudo, S. P.; Allen, R. C.; Rutan, S. C.; Stoll, D. R. Anal. Bioanal. Chem.
2013, 405 (13), 4639−4653.
(23) De Vos, J.; Desmet, G.; Eeltink, S. J. Chromatogr. A 2014, 1360,
164−171.
(24) De Vos, J.; Eeltink, S.; Desmet, G. J. Chromatogr. A 2015, 1381,
74−86.
(25) De Vos, J.; Desmet, G.; Eeltink, S. J. Chromatogr. A 2016, 1455,
86−92.
(26) Greibrokk, T.; Andersen, T. J. Chromatogr. A 2003, 1000 (1),
743−755.
(27) McNeff, C. V.; Yan, B.; Stoll, D. R.; Henry, R. A. J. Sep. Sci. 2007,
30 (11), 1672−1685.
(28) Cabooter, D.; Heinisch, S.; Rocca, J. L.; Clicq, D.; Desmet, G. J.
Chromatogr. A 2007, 1143 (1−2), 121−133.
(29) Heinisch, S.; Desmet, G.; Clicq, D.; Rocca, J.-L. J. Chromatogr. A
2008, 1203 (2), 124−136.
(30) Wiese, S.; Teutenberg, T.; Schmidt, T. C. Anal. Chem. 2011, 83
(6), 2227−2233.

1130 DOI: 10.1021/acs.analchem.6b03279


Anal. Chem. 2017, 89, 1123−1130

Potrebbero piacerti anche